HardyDisk Meropenem/Vaborbactam 20/10µg (MEV30)

K172621 · Hardy Diagnostics · JTN · Oct 3, 2017 · Microbiology

Device Facts

Record IDK172621
Device NameHardyDisk Meropenem/Vaborbactam 20/10µg (MEV30)
ApplicantHardy Diagnostics
Product CodeJTN · Microbiology
Decision DateOct 3, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1620
Device ClassClass 2

Intended Use

Use of HardyDisk™ Meropenem/Vaborbactam 20/10ug (MEV30) for in vitro agar diffusion susceptibility testing is indicated when there is the need to determine the susceptibility of bacteria to Meropenem/Vaborbactam. Meropenem/Vaborbactam has been shown to be active against susceptible isolates of the following microorganisms both in vitro and in clinical infections: Enterobacter cloacae species complex Escherichia coli Klebsiella pneumoniae Meropenem/Vaborbactam has been shown to be active in vitro against susceptible isolates of the following microorganisms: Citrobacter freundii Citrobacter koseri Enterobacter aerogenes Klebsiella oxytoca Morganella morganii Proteus mirabilis Providencia spp. Pseudomonas aeruginosa Serratia marcescens. HardyDisk™ AST Disks are used for semi-quantitative in vitro susceptibility testing by the agar diffusion test procedure (Kirby-Bauer) of rapidly growing and certain fastidious bacterial pathogens. Standardized methods for agar diffusion testing have been described for Enterobacteriaceae, Staphylococcus spp., Acinetobacter spp., Listeria monocytogenes, Enterococcus spp., and by modified procedures, Haemophilus spp., Neisseria gonorrhoeae, N. meningitidis and Streptococcus spp., including Streptococcus pneumoniae.

Device Story

HardyDisk Meropenem/Vaborbactam 20/10ug (MEV30) is an antimicrobial susceptibility test (AST) disk used for semi-quantitative in vitro testing via the Kirby-Bauer agar diffusion method. The device consists of paper disks impregnated with the antibiotic combination Meropenem/Vaborbactam. In a clinical laboratory setting, the disk is placed on an agar plate inoculated with a bacterial isolate. As the antibiotic diffuses into the agar, it creates a concentration gradient. After incubation, the zone of inhibition around the disk is measured. The size of this zone correlates with the susceptibility of the bacteria to the antibiotic. Healthcare providers use these results to guide antibiotic therapy decisions for patients with infections caused by susceptible organisms, such as Enterobacter cloacae, E. coli, and K. pneumoniae. The device aids in selecting appropriate antimicrobial treatment, potentially improving patient outcomes by ensuring targeted therapy.

Clinical Evidence

No clinical data. Performance characteristics rely on breakpoints and quality control ranges established by CDER for the pharmaceutical drug.

Technological Characteristics

Paper-based AST disk impregnated with 20/10ug Meropenem/Vaborbactam. Operates via passive diffusion in agar medium. Class II device, product code JTN. No software or electronic components.

Indications for Use

Indicated for in vitro agar diffusion susceptibility testing of bacteria to Meropenem/Vaborbactam. Active against Enterobacter cloacae complex, E. coli, K. pneumoniae (in vitro and clinical infections) and Citrobacter spp., Enterobacter aerogenes, K. oxytoca, M. morganii, P. mirabilis, Providencia spp., P. aeruginosa, and S. marcescens (in vitro only).

Regulatory Classification

Identification

An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure by a disc-agar diffusion technique or a disc-broth elution technique the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents. In the disc-agar diffusion technique, bacterial susceptibility is ascertained by directly measuring the magnitude of a zone of bacterial inhibition around the disc on an agar surface. The disc-broth elution technique is associated with an automated rapid susceptibility test system and employs a fluid medium in which susceptibility is ascertained by photometrically measuring changes in bacterial growth resulting when antimicrobial material is eluted from the disc into the fluid medium. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, resembling a bird-like shape. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 3, 2017 HARDY DIAGNOSTICS ANDRE HSIUNG DIRECTOR OF TECHNICAL SERVICES 1430 WEST MCCOY LANE SANTA MARIA CA 93455 Re: K172621 Trade/Device Name: HardyDisk Meropenem/Vaborbactam 20/10ug (MEV30) Regulation Number: 21 CFR 866.1620 Regulation Name: Antimicrobial susceptibility test disc Regulatory Class: II Product Code: JTN Dated: August 30, 2017 Received: August 31, 2017 Dear Mr. Hsiung: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ## Ribhi Shawar -S For Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K172621 Device Name HardyDiskTM Meropenem/Vaborbactam 20/10ug (MEV30) Indications for Use (Describe) Use of HardyDisk™ Meropenem/Vaborbactam 20/10ug (MEV30) for in vitro agar diffusion susceptibility testing is indicated when there is the need to determine the susceptibility of bacteria to Meropenem/Vaborbactam. Meropenem/Vaborbactam has been shown to be active against susceptible isolates of the following microorganisms both in vitro and in clinical infections: Enterobacter cloacae species complex Escherichia coli Klebsiella pneumoniae Meropenem/Vaborbactam has been shown to be active in vitro against susceptible isolates of the following microorganisms: Citrobacter freundii Citrobacter koseri Enterobacter aerogenes Klebsiella oxytoca Morganella morganii Proteus mirabilis Providencia spp. Pseudomonas aeruginosa Serratia marcescens. HardyDisk™ AST Disks are used for semi-quantitative in vitro susceptibility testing by the agar diffusion test procedure (Kirby-Bauer) of rapidly growing and certain fastidious bacterial pathogens. Standardized methods for agar diffusion testing have been described for Enterobacteriaceae, Staphylococcus spp., Acinetobacter spp., Listeria monocytogenes, Enterococcus spp., and by modified procedures, Haemophilus spp., Neisseria gonorrhoeae, N. meningitidis and Streptococcus spp., including Streptococcus pneumoniae. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------------------| | <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) | | <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...