K151688 · bioMerieux, Inc. · JSO · Mar 11, 2016 · Microbiology
Device Facts
Record ID
K151688
Device Name
chromID MRSA
Applicant
bioMerieux, Inc.
Product Code
JSO · Microbiology
Decision Date
Mar 11, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1700
Device Class
Class 2
Intended Use
chromID™ MRSA agar is a selective and differential chromogenic medium for: A. The qualitative detection of nasal colonization of methicillin-resistant Staphylococcus aureus (MRSA), to aid in the prevention and control of MRSA in healthcare settings. The test is performed on anterior nares swab specimens from patients and healthcare workers to screen for MRSA colonization. chromID™ MRSA when used to detect nasal colonization is not intended to diagnose, guide, or monitor therapy for MRSA infections, or provide results of susceptibility to methicillin. B. The qualitative detection of MRSA from skin and skin structure infections. chromID™ MRSA is indicated for use in conjunction with other laboratory tests and clinical data available to aid in the identification and diagnosis of MRSA infections. Concomitant cultures for skin and skin structure infections are necessary to recover organisms for further microbiological susceptibility testing or epidemiological typing. A negative result does not preclude MRSA infection. chromID™ MRSA is not intended to monitor treatment for MRSA infections, or provide results of susceptibility to methicillin.
Device Story
chromID™ MRSA is a selective/differential chromogenic culture medium; utilizes nutrient agar base with peptones, antibiotics (e.g., cefoxitin), and α-glucosidase chromogenic substrate. Used in clinical laboratories; manual operation. Input: anterior nares swabs or skin/skin structure infection specimens. Principle: α-glucosidase produced by S. aureus cleaves chromogenic substrate, producing green colonies; antibiotics inhibit non-Staphylococcus bacteria and yeasts. Output: visual identification of green colonies indicating MRSA presence. Healthcare providers use results alongside clinical data to aid MRSA identification/diagnosis; negative results do not preclude infection. Benefits: facilitates rapid screening/identification of MRSA to support infection control and clinical management.
Clinical Evidence
Clinical study evaluated 680 valid skin/skin structure infection (SSSI) specimens across four sites. Compared to reference culture method (TSB enrichment followed by BAP subculture). Sensitivity 93.8% (95% CI: 89.2-96.5%); Specificity 97.4% (95% CI: 95.6-98.5%). Analytical studies included reproducibility (100%), reactivity with mecA/mecC strains, and interference testing. Bench testing only for non-clinical aspects.
Technological Characteristics
Selective chromogenic agar medium; contains antibiotic mixture for inhibition of non-Staphylococcus species and yeasts. Identification via chromogenic substrate cleavage by S. aureus (green colonies). Manual reading; aerobic incubation at 35-37°C. Standalone diagnostic medium.
Indications for Use
Indicated for qualitative detection of MRSA nasal colonization in patients and healthcare workers, and for detection of MRSA from skin and skin structure infections. Not for diagnosing, guiding, or monitoring MRSA therapy or determining methicillin susceptibility.
Regulatory Classification
Identification
A culture medium for antimicrobial susceptibility tests is a device intended for medical purposes that consists of any medium capable of supporting the growth of many of the bacterial pathogens that are subject to antimicrobial susceptibility tests. The medium should be free of components known to be antagonistic to the common agents for which susceptibility tests are performed in the treatment of disease.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 11, 2016
BIOMERIEUX, INC. JEWELL COULSON SR. REGULATORY AFFAIRS SPECIALIST 595 ANGLUM RD. HAZELWOOD MO 63042
Re: K151688
Trade/Device Name: chromID™ MRSA Regulation Number: 21 CFR 866.1700 Regulation Name: Culture medium for antimicrobial susceptibility tests Regulatory Class: II Product Code: JSO Dated: February 4, 2016 Received: February 8, 2016
Dear Ms. Coulson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Ribhi Shawar -S
For Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVIGES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K151688
Device Name
chromIDTM MRSA
Indications for Use (Describe)
chromID™ MRSA agar is a selective and differential chromogenic medium for :
A. The qualitative detection of nasal colonization of methicillin-resistant Staphylococcus aureus (MRSA), to aid in the prevention and control of MRSA in healthcare settings.
The test is performed on anterior nares swab specimens from patients and healthcare workers to screen for MRSA colonization. chromID™MRSA when used to detect nasal colonization is not intended to diagnose, guide, or monitor therapy for MRSA infections, or provide results of susceptibility to methicillin.
B. The qualitative detection of MRSA from skin structure infections.
chromID™ MRSA is indicated for use in conjunction with other laboratory tests and clinical data available to aid in the identification and diagnosis of MRSA infections. Concomitant cultures for skin structure infections are necessary to recover organisms for further microbiological susceptibility testing or epidemiological typing.
A negative result does not preclude MRSA infection. chromID™MRSA is not intended to monitor treatment for MRSA · infections, or provide results of susceptibility to methicillin.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for bioMerieux. The logo consists of the word "BIOMERIEUX" in a sans-serif font, with a stylized graphic above it. The graphic is a circle bisected by a curved line, with one half of the circle containing a series of horizontal lines.
# 510(k) SUMMARY
| Submitter's Name: | bioMérieux, Inc. |
|----------------------|--------------------------------------------------------|
| Address: | 595 Anglum Road<br>Hazelwood, MO 63042 |
| Contact Person: | Jewell Coulson<br>Senior Regulatory Affairs Specialist |
| Phone Number: | 314-731-7342 |
| Fax Number: | 314-731-8689 |
| Date of Preparation: | February 4, 2016 |
| <b>Device :</b> | |
| Formal/Trade Name: | chromID™ MRSA |
| Common Name: | Culture media |
| Classification | II |
| Regulation Name | Culture Media for Antimicrobial Susceptibility Tests |
| Regulation Number | 21 CFR 866.1700 |
| Product Code | JSO |
| Predicate Device: | Bio-Rad MRSA Select™ (K100589) |
| 510(k) Summary: | |
#### Intended Use:
chromID™ MRSA agar is a selective and differential chromogenic medium for :
- A. The qualitative detection of nasal colonization of methicillin-resistant Staphylococcus aureus (MRSA), to aid in the prevention and control of MRSA in healthcare settings.
The test is performed on anterior nares swab specimens from patients and healthcare workers to screen for MRSA colonization. chromID™ MRSA when used to detect nasal colonization is not intended to diagnose, guide, or monitor therapy for MRSA infections, or provide results of susceptibility to methicillin.
- B. The qualitative detection of MRSA from skin structure infections.
chromID™ MRSA is indicated for use in conjunction with other laboratory tests and clinical data available to aid in the identification and diagnosis of MRSA infections. Concomitant cultures for skin and skin structure infections are necessary to recover organisms for further microbiological susceptibility testing or epidemiological typing. A negative result does not preclude MRSA infection. chromID™ MRSA is not intended to monitor treatment for MRSA infections, or provide results of susceptibility to methicillin.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for bioMerieux. The logo consists of a stylized globe with a curved line running through it. Below the globe is the company name, "BIOMERIEUX", in a sans-serif font.
# Device Description:
chromID™ MRSA agar consists of a rich nutritive base combining different peptones. It also contains a chromogenic substrate of α-glucosidase and a combination of several antibiotics (cefoxitin, etc.) that favor the growth and direct detection of MRSA, including hetero-resistant strains, by revealing orglucosidase activity (patent registered) through the appearance of green colonies. The α-glucosidase produced by S. aureus cleaves the chromogenic substrate, which gives a green color to the S. aureus colonies growing on the medium. The MRSA strains are identified by the presence of green colonies that result from the chromogen incorporated in the medium. The selective mixture of antibiotics inhibits most bacteria not belonging to the genus Staphylococcus, as well as yeasts.
# Substantial Equivalence
The similarities of ChromID™ MRSA agar when compared to the predicate device are described in the following table.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for bioMérieux, a French multinational biotechnology company. The logo features the company name in a sans-serif font, with the word "BIO" stacked above "MÉRIEUX". Above the company name is a circular graphic that is bisected by a thin vertical line. The left half of the circle has several curved lines, while the right half is solid.
| | Device | Predicate |
|--------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item | chromID™ MRSA Agar | Bio-Rad MRSASelect™ |
| Similarities | | |
| Intended Use | chromID™ MRSA agar is a selective and<br>differential chromogenic medium for : | MRSASelect™ is a selective and<br>differential chromogenic medium for: |
| | A. The qualitative detection of nasal colonization<br>of methicillin-resistant Staphylococcus<br>aureus (MRSA), to aid in the prevention and<br>control of MRSA in healthcare settings.<br>The test is performed on anterior nares swab<br>specimens from patients and healthcare<br>workers to screen for MRSA colonization.<br>chromID™ MRSA when used to detect nasal<br>colonization is not intended to diagnose,<br>guide, or monitor therapy for MRSA<br>infections, or provide results of susceptibility<br>to methicillin. | A) The qualitative detection of nasal<br>colonization of methicillin resistant<br>Staphylococcus aureus (MRSA) to aid<br>in the prevention and control of MRSA<br>infections in healthcare settings. The<br>test can be performed on anterior<br>nares specimens from patients and<br>healthcare workers to screen for<br>MRSA colonization. MRSASelect™ is<br>not intended to diagnose MRSA<br>infection nor to guide or monitor<br>treatment of infection. Results can be<br>interpreted after 18 to 28 hours<br>incubation. |
| | B. The qualitative detection of MRSA from skin<br>and skin structure infections.<br>chromID™ MRSA is indicated for use in<br>conjunction with other laboratory tests and<br>clinical data available to aid in the<br>identification and diagnosis of MRSA<br>infections. Concomitant cultures for skin and<br>skin structure infections are necessary to<br>recover organisms for further microbiological<br>susceptibility testing, or epidemiological<br>typing.<br>A negative result does not preclude MRSA<br>infection. chromID™ MRSA is not intended to<br>monitor treatment for MRSA infections, or<br>provide results of susceptibility to methicillin. | B) The qualitative detection of methicillin<br>resistant Staphylococcus aureus<br>(MRSA) from skin and soft-tissue<br>wound specimens. The<br>MRSASelect™ is indicated for use in<br>conjunction with other laboratory tests<br>and clinical data available to aid in the<br>identification and diagnosis of MRSA<br>from patients with skin and soft-tissue<br>infections. Concomitant cultures and<br>susceptibility testing are necessary for<br>all skin and soft-tissue wound<br>specimens. MRSASelect™ is not<br>intended to guide, or monitor treatment<br>for MRSA infection, or provides results<br>of susceptibility to methicillin. Results<br>can be interpreted after 18 to 28 hours<br>incubation. |
| Test method | Manual | Manual |
| Inoculum | Direct Specimen | Direct Specimen |
| Specimen | Anterior nares specimens | Anterior nares specimens |
| | Skin and skin structure specimens | Skin and soft-tissue wound specimens |
The chromID™ MRSA agar utilizes nutrient agar medium that contains selective and differential agents and is very similar to the MRSASelect™ agar. Both devices incorporate a chromogenic substrate that is enzymatically cleaved by S. aureus and result in colored MRSA colonies. Both devices incorporate selective agents in the agar to inhibit most bacteria not belonging to the genus Staphylococcus, as well as yeasts. The significant technological difference between the two media is the type of chromogenic substrate incorporated in the media to indicate the presence of MRSA colonies. The safety and effectiveness of the chromID™ MRSA agar is not impacted by the technology differences.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for bioMerieux. The logo consists of the company name in a sans-serif font, with a stylized globe above it. A curved line extends from the top of the globe, arcing over the company name. The globe is divided into two halves, one with vertical lines and the other solid.
# Performance Characteristics:
Non-clinical (analytical) and clinical studies were performed for the chromID™ MRSA agar following the FDA Draft Guidance "Establishing the Performance Characteristics of In Vitro Diagnostic Devices for the Detection of Methicillin-Resistant Staphylococcus aureus (MRSA) for Culture Based Systems", issued June 15, 2011.
### Analytical Performance
- Reproducibility A set of ten mecA-positive or mecA-negative Staphylococcus aureus isolates . were tested in triplicate each day for five days at three of the clinical study sites using multiple lots of chromID™ MRSA media. Isolates were tested at 103 CFU/mL and plates were read at 24 hours. The overall reproducibility rate was 100% (450/450).
- Quality Control Two quality control organisms were tested at each study site by chromlD™ ● MRSA on each day of testing. The organisms tested were:
| Staphylococcus aureus | ATCC 29213 |
|-----------------------|------------|
| Staphylococcus aureus | ATCC 43300 |
The results of QC testing with chromID™ MRSA agar met the pre-defined acceptance criteria (≥ 95% agreement) when compared to expected results as determined by oxacillin MIC and mecA PCR.
- Analytical Reactivity - A challenge set composed of 80 mecA MRSA strains and 5 mecC MRSA strains was inoculated on the chromID™ MRSA medium with an inoculum equivalent to 10° CFU/mL. After 24 hours of incubation, 58/80 mecA MRSA strains and 4/5 mecC MRSA strains were detected on the chromID™ MRSA medium.
- . Expression of Resistance - Twenty-eight well-characterized S. aureus strains were tested on chromID™ MRSA (10 low level methicillin-resistant, 10 high level methicillin-resistant, 5 borderline oxacillin-resistant, and 3 methicillin-susceptible) using inoculum concentrations between 100 -10° CFU/mL. All low level and high level methicillin-resistant strains were detected at an inoculum of >10° CFU/mL. The 5 border-line oxacillin-resistant and 3 methicillin-susceptible strains did not grow, as expected, on chromID™ MRSA at inoculum concentrations as high as 10° CFU/mL.
- Mixed Infection Study Ten MRSA strains (5 low level methicillin-resistant and 5 high level ● methicillin-resistant) at a concentration of 103 CFU/mL were inoculated on chromID™ MRSA with one of three non-MRSA strains (methicillin-susceptible S. aureus, methicillin-resistant coaqulase negative Staphylococcus epidermidis, and Escherichia coli) at 10° – 10° CFU/mL to evaluate the recovery of MRSA on chromID™ MRSA media in the presence of non-target organisms. The presence of these non-MRSA organisms did not negatively impact chromID™ MRSA performance.
#### Recovery Study .
Two well-characterized MRSA strains (ATCC® 43300™ and CDC Mu3-BR) were tested to determine the lowest number of CFU/mL detected on chromID™ MRSA. At 24 hours the lowest dilution for detection on chromID™ MRSA for ATCC® 43300™ was 103 CFU/mL and 10° CFU/mL for CDC Mu3-8R.
- Cross Reactivity To evaluate the analytical specificity of the chromID™ MRSA media, 71 non-● MRSA strains representing bacterial and fungal species potentially encountered in skin and skin structure infections were inoculated onto chromID™ MRSA medium at a high inoculum level (10° CFU/mL). After 24 hours of incubation, 44 strains did not grow and 20 strains grew colonies without green pigment. Green colonies (cross reactivity) were observed for 3 Klebsiella
{7}------------------------------------------------
pneumoniae and one Enterobacter cloacae strains. All 4 were carbapenemase producing (KPC). Green colonies were also observed for one S. pseudintermedius and 2 S. sciuri strains. All 3 were oxacillin-resistant strains.
- Interference Interfering substances were evaluated at physiologically or biologically relevant ● concentrations and mixed with bacterial suspensions (9:1, v:v). No interference was observed for human blood, mucin, anticoagulants, plasma, and buffy coat. Some topical antibiotics and antiseptics interfered with MRSA detection on chromID™ MRSA due to their antibacterial activity. Use of compounds containing the active ingredients listed below may have an inhibitory effect on MRSA growth that is unrelated to chromID™ MRSA medium performance: colistimethate sodium 25 MUI/100mL, bacitracin 50,000 Ul/100mL, dexamethasone 0.1g/100mL, neomycin sulfate 350,000 Ul/100mL, polymixin B sulfate 600,000 Ul/100mL, benzalkonium chloride 500mg/100mL, chlorhexidine digluconate 200mg/100mL, 1% hydrocortisone, 10% povidone iodine, 2% mupirocin, hydrogen peroxide 8 mg/g, ethanol 380 mg/g, 70% isopropyl alcohol. Benzocaine (14 g/L) may delay colony coloration or inhibit growth of MRSA on chromID™ MRSA.
- Incubation The incubation times required for three MRSA strains, at an organism concentration ● of 103 CFU/mL to produce positive chromID™ MRSA results was 20 hours for two strains (ATCC 43300 and S. aureus 0611169) and 27 hours for one strain (CDC Mu3-8R).
# Clinical Studies
chromID™ MRSA was evaluated at four geographically diverse clinical sites. chromID™ MRSA performance was determined by examining the presence of green colonies on the media after 24 hours. All green colonies were tested by Gram stain, catalase and latex agglutination. All Staphylococcus aureus colonies were tested for resistance to oxacillin by the Cefoxitin Screen test (30ug). All green colonies were also tested for the presence of the mecA gene by PCR and by VITEK® MS for species confirmation.
For the Reference Culture Method, all specimens were enriched in Tryptic Soy broth with 6.5% NaCl (TSB) for 24 hours. Negative broth cultures were incubated an additional 24 hours. Positive cultures were then subcultured to Tryptic Soy agar with 5% sheep blood (BAP). Colonies suggestive of Staphylococcus species were tested by Gram stain, catalase and latex aqqlutination. Staphylococcus aureus isolates were tested for resistance to oxacillin by the Cefoxitin Screen test. All colonies resistant to oxacillin by the Cefoxitin Screen test were tested for the presence of the mecA gene by PCR and by VITEK MS for species confirmation.
A positive result by Reference Culture was defined as the recovery of cefoxitin-resistant Staphylococcus aureus from the specimen. All other results, including the growth of cefoxitinsusceptible Staphylococcus aureus, growth of other species and no growth, were considered negative.
A total of 690 SSSI specimens were collected for the study at 4 external clinical laboratories. Three specimens were excluded because they did not meet specimen acceptance criteria. Seven specimens were removed due to protocol deviations. A total of 680 specimens were valid for comparison data analysis.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for bioMérieux. The logo consists of the company name in all caps, with a stylized graphic above it. The graphic is a circle with lines on the left side and a vertical line going through the middle.
# chromID™ MRSA at 24 Hours
### Compared to Reference Culture Method (TSB) at 48 Hours
| | Reference Culture Method<br>(TSB) Result | | |
|--------------------------------|------------------------------------------|-----|-------|
| chromID™ MRSA<br>Result in 24h | Pos | Neg | Total |
| Pos | 166 | 13* | 179 |
| Neg | 11** | 490 | 501 |
| Total | 177 | 503 | 680 |
*13 discordant specimens (chromID™ MRSA result of MRSA positive: culture-based testing result of MRSA neqative) were observed. 5 of the 13 chromID™ MRSA positive specimens were confirmed as MRSA by Cefoxitin Screen Test. 8 cultures grew green colonies on
chromID™MRSA that were not MRSA.
**11 discordant specimens (chromID™ MRSA result of MRSA negative: culture-based testing result of MRSA positive) were observed. 6 cultures displayed no growth on chromID' MRSA. 4 cultures grew non-green MRSA colonies on chromID™ MRSA. One specimen grew green colonies on chromID™ MRSA that were not identified as MRSA.
Specificity 97.4% (95% CI: [95.6-98.5]) Sensitivity 93.8% (95% CI: [89.2-96.5])
The overall prevalence of MRSA in this study by the Reference Culture Method was 26.0% (177/680). . The prevalence of MRSA as determined by chromID™ MRSA at 24 hours was 26.3% (179/680).
#### Conclusion
A comparison of the intended use and the results of non-clinical performance studies support that chromID™ MRSA agar is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.