Accelerate Pheno System, Accelerate PhenoTest BC Kit
Applicant
Accelerate Diagnostics, Inc.
Product Code
PRH · Microbiology
Decision Date
Sep 15, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1650
Device Class
Class 2
Attributes
AI/ML
Intended Use
The Accelerate PhenoTest™ BC kit is a multiplexed in vitro diagnostic test utilizing both qualitative nucleic acid fluorescence in situ hybridization (FISH) identification and quantitative, antimicrobial susceptibility testing (AST) methods and is intended for use with the Accelerate Pheno™ system. The Accelerate PhenoTest™ BC kit is capable of simultaneous detection and identification of multiple microbial targets followed by susceptibility testing of the appropriate detected bacterial organisms. The Accelerate PhenoTest™ BC kit is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system. Results are intended to be interpreted in conjunction with Gram stain results.
Device Story
The Accelerate Pheno System is an automated in vitro diagnostic platform for rapid identification (ID) and antimicrobial susceptibility testing (AST) of pathogens directly from positive blood culture samples. The system uses a disposable test cassette and reagent cartridge. Input samples undergo automated gel electrofiltration to isolate microbial cells, which are then captured on a flowcell surface using an electric field. Identification is performed via FISH using fluorescently-labeled DNA probes targeting ribosomal RNA, imaged by fluorescence microscopy. AST is performed using Morphokinetic Cellular Analysis (MCA), a computer vision technology that tracks individual live cells and microcolonies over time in the presence of antimicrobials. The system produces MIC values and categorical interpretations (S, I, R). Used in clinical laboratories, the system provides results in approximately 7 hours. The output assists clinicians in selecting appropriate antimicrobial therapy, potentially improving patient outcomes by enabling faster targeted treatment.
Clinical Evidence
No clinical testing was performed. Performance was established via non-clinical internal evaluation using 144 characterized P. aeruginosa isolates. AST performance was compared to broth microdilution (BMD). Results showed essential agreement (EA) ranging from 93.8% to 96.5% and categorical agreement (CA) from 88.2% to 96.5% across five beta-lactam antimicrobials. Adjusted major and very major error rates were calculated where applicable due to lack of intermediate breakpoints.
Technological Characteristics
System uses gel electrofiltration (GEF) for sample prep and indium tin oxide (ITO) coated flowcells for cell capture. Imaging uses fluorescence and dark-field microscopy with ATTO-532 and ATTO-647 probes. Connectivity includes web-based software accessible via touchscreen or network. Software is moderate level of concern. Hardware includes ID/AST modules, computing system (Interface PC/Analysis module), and touchscreen. Sterilization is not specified; materials include polymer microspheres and gel wells.
Indications for Use
Indicated for use as an aid in the diagnosis of bacteremia and fungemia and for susceptibility testing of specific pathogenic bacteria (Staphylococcus spp., Enterococcus spp., Streptococcus spp., Pseudomonas aeruginosa, Acinetobacter baumannii, Klebsiella spp., E. coli, Enterobacter spp., Proteus spp., Citrobacter spp., Serratia marcescens) and yeasts (Candida albicans, Candida glabrata) in positive blood culture samples.
Regulatory Classification
Identification
A cellular analysis system for multiplexed antimicrobial susceptibility testing is a multiplex qualitative and/or quantitative in vitro diagnostic device intended for the identification and determination of the antimicrobial susceptibility results of organisms detected in samples from patients with suspected microbial infections. This device is intended to aid in the determination of antimicrobial susceptibility or resistance when used in conjunction with other laboratory findings.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) Detailed device description documentation, including the device components, ancillary reagents required but not provided, a detailed explanation of the methodology, including primer/probe sequence, design, rationale for sequence selection, and details of the antimicrobial agents, as applicable.
(ii) Detailed documentation from the following analytical and clinical performance studies: limit of detection, inclusivity, precision, reproducibility, interference, cross-reactivity, carryover, and cross-contamination, quality control and additional studies, as applicable to specimen type and assay intended use.
(iii) Detailed documentation from an appropriate clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to results obtained from well-accepted reference methods.
(iv) Detailed documentation for device software, including software applications and hardware-based devices that incorporate software.
(2) The labeling required under § 809.10(b) of this chapter must include:
(i) Limitations and protocols regarding the need for correlation of results by standard laboratory procedures, as applicable.
(ii) A detailed explanation of the interpretation of results and acceptance criteria.
(iii) A detailed explanation of the principles of operation and procedures for assay performance and troubleshooting.
Predicate Devices
Accelerate PhenoTest™ BC kit/Accelerate Pheno™ system (DEN160032)
Related Devices
DEN160032 — Accelerate Pheno system, Accelerate Phenotest BC Kit · Accelerate Diagnotics · Feb 23, 2017
K221688 — ASTar BC G- Kit and ASTar Instrument · Q-Linea AB · Apr 26, 2024
K223493 — PBC Separator with Selux AST System · Selux Diagnostics, Inc. · Feb 15, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 15, 2020
Accelerate Diagnostics, Inc. Carrene Plummer Director, Regulatory Affairs 3950 S. Country Club Road #470 Tucson, Arizona 85714
## Re: K192665
Trade/Device Name: Accelerate Pheno System, Accelerate PhenoTest BC Kit Regulation Number: 21 CFR 866.1650 Regulation Name: Positive Blood Culture Identification and AST Kit Regulatory Class: Class II Product Code: PRH, NSU, PEO, PAM, PEN, LON Dated: September 22, 2019 Received: September 25, 2019
Dear Carrene Plummer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ribhi Shawar, Ph.D. (ABMM) Chief General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192665
Device Name Accelerate PhenoTest™ BC kit Accelerate Pheno™ system
#### Indications for Use (Describe)
The Accelerate PhenoTest™ BC kit is a multiplexed in vitro diagnostic test utilizing both qualitative nucleic acid fluorescence in situ hybridization (FISH) identification and quantitative, antimicrobial susceptibility testing (AST) methods and is intended for use with the Accelerate Pheno™ system. The Nccelerate PhenoTest™ BC kit is capable of simultaneous detection and identification of multiple microbial targets followed by susceptibility testing of the appropriate detected bacterial organisms. The Accelerate PhenoTest™ BC kit is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system. Results are intended to be interpreted in conjunction with Gram stain results.
The Accelerate PhenoTest™ BC kit identifies the following Gram-positive and Gram-negative bacteria and yeasts utilizing FISH probes targeting organism-specific ribosomal RNA sequences:
Staphylococcus aureus, Staphylococcus lugdunensis, Coagulase-negative Staphylococcus species (i.e., Staphylococcus epidermidis, Staphylococcus haemolyticus, Staphylococcus capits, Staphylococcus lugdunensis, Staphylococcus warneri, not differentiated), Enterococcus faecum, Streptococus spp. (i.e., Streptococus mitis, Streptococcus oralis, Streptococcus gallolyticus, Streptococcus agalactiae, Streptococus pneumoniae, not differentiated), Pseudomonas aeruginosa, Acinetobacter baumannii, Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter cloacae, Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter koseri, not differentiated), Serratia marcescens, Candida albicans and Candida glabrata.
The Accelerate PhenoTest™ BC kit tests the following antimicrobial agents with the specific target organisms identified below:
Amikacin: Pseudomonas aeruginosa, Acinetobacter baumannii, Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter cloacae, Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Ampicillin: Enterococcus faecalis and Enterococcus faecium
Ampicillin/Sulbactam: Escherichia coli, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), and Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated)
Aztreonam: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Ceftazidime: Pseudomonas aeruginosa, Klebsiella pneumoniae, Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Ceftaroline: Staphylococcus aureus
Cefepime: Pseudomonas aeruginosa, Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter freundii,
{3}------------------------------------------------
Citrobacter koseri, not differentiated) and Serratia marcescens
Ceftriaxone: Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Ciprofloxacin: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Daptomycin: Staphylococcus aureus, Coagulase-negative Staphylococcus species (i.e., Staphylococcus epidermidis, Staphylococcus haemolyticus, Staphylococcus hominis, Staphylococcus capitis, Staphylococcus lugdunensis, Staphylococcus warneri, not differentiated), Enterococcus faecalis and Enterococcus faecium
Ertapenem: Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Gentamicin: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Linezolid: Staphylococcus aureus, Enterococcus faecalis and Enterococcus faecium
Meropenem: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Piperacillin/Tazobactam: Pseudomonas aeruginosa, Acinetobacter baumannii, Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter cloacae, Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter koseri, not differentiated) and Seratia marcescens Tobramycin: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Vancomycin: Staphylococcus aureus. Staphylococcus lugdunensis. Coagulase- negative Staphylococcus species (i.e., Staphylococcus epidermidis, Staphylococcus haemolyticus, Staphylococus hominis, Staphylococcus capitis, Staphylococus lugdunensis, Staphylococus warneri, not differentiated), Enterococcus faecius and Enterococus faccium
The following resistance phenotype is reported based on qualitative tests: Methicillin-resistance (S. aureus S. Jugdunensis, coagulase negative staphylococci).
The Accelerate PhenoTest™ BC kit is indicated as an aid in the diagnosis of bacteremia and fungemia. It is also indicated for susceptibility testing of specific pathogenic bacteria as identified above commonly associated with or causing bacteremia. Results are intended to be used in conjunction with other clinical and laboratory findings.
Standard laboratory protocols for processing positive blood cultures should be followed to ensure availability of isolates for supplemental testing as needed. Additionally, subculture of positive blood culture is necessary for the identification and susceptibility testing of: organisms not identified by the Accelerate PhenoTest™ BC kit, organisms present in polymicrobial samples, organisms for which species identification is critical for patient care (e.g. speciation of Streptococus spp.), samples for which an "indeterminate" result for any probe was obtained, for testing antimicrobial agents not included on the Accelerate panel and for epidemiologic testing.
Type of Use (Select one or both, as applicable)
|× Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
{4}------------------------------------------------
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo consists of a blue geometric shape on the left, followed by the word "ACCELERATE" in gray, block letters. Below "ACCELERATE" is the word "DIAGNOSTICS" in a smaller, blue font with a trademark symbol.
K192665
September 10, 2020
# 510(k) Summary
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo features a blue geometric shape on the left, resembling a stylized diamond or hexagon. To the right of the shape is the word "ACCELERATE" in a bold, sans-serif font. Below "ACCELERATE" is the word "DIAGNOSTICS" in a smaller, lighter blue font.
#### 510(k) SUMMARY
#### Submitter Information:
Submitter: Accelerate Diagnostics, Inc.
Address: 3950 S. Country Club Road #470 Tucson, AZ 85714
Establishment Registration No: 3010671651
Contact Person: Carrene Plummer
Phone: (520) 405-1462
Fax: (520) 269-6580
E-mail: cplummer@axdx.com
Date Prepared: September 22, 2019
#### Name of Device and Classification:
Trade Name: Accelerate PhenoTest™ BC kit, Accelerate Pheno™ system
Classification Name: A cellular analysis system for multiplexed antimicrobial susceptibility testing
Product Code: PRH, NSU, PEO, PAM, PEN, LON
#### Predicate Device:
Accelerate PhenoTest™ BC kit/Accelerate Pheno™ system (DEN160032)
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo consists of a blue geometric shape on the left, followed by the word "ACCELERATE" in gray, sans-serif font. Below "ACCELERATE" is the word "DIAGNOSTICS" in a smaller, blue, sans-serif font with a trademark symbol.
## Device Description
## Assay Description
There are no changes to the description of the assays in the Accelerate PhenoTest™ BC kit description provided in DEN160032, except for updated time to result and addition of an optical filter to assist in removal of interfering sample debris.
The Accelerate PhenoTest™ BC kit contains a sample vial, a 48-channel disposable test cassette and a reagent cartridge. All identification (ID) and antimicrobial susceptibility testing (AST) is performed in individual flowcells of the test cassette. The reagent cartridge contains gel electrofiltration (GEF) stations, fluorescence in situ hybridization (FISH) probes, antimicrobials, and reagents for automated sample preparation, identification of bacterial and fungal target organisms, and antimicrobial susceptibility and resistance marker detection testing for bacterial target organisms. The user loads the sample into the sample vial, places the test cassette, reagent cartridge and sample vial into an Accelerate Pheno™ system module, then presses the module button to close the module door and start the run. The rest of the operations are automated as described below.
#### Automated Sample Preparation
Automated sample preparation is performed using gel electrofiltration (GEF) which is based on gel electrophoresis principles. Sample is automatically transferred to a gel well containing pores smaller than bacterial or yeast cells. Application of an electric field causes lysed blood cells and/or other sample debris to pass into the gel wall while bacterial/yeast cells remain inside the gel well. The electric field is briefly reversed to dislodge bacterial/yeast cells from the gel wall prior to removal.
## Cell Capture
Following sample preparation, recovered cells are automatically pipetted into multiple flowcell channels of the test cassette. Conductive layers of transparent indium tin oxide (TTO) coat the top and bottom inner surfaces of each flowcell channel and act as electrodes. An additional cationic poly-L-lysine layer on the bottom of each flowcell acts as a capture surface. When a voltage is applied, the negatively-charged bacterial/yeast cells migrate to the positively-charged capture surface where they are captured prior to imaging.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo features a blue and gray geometric shape on the left, followed by the word "ACCELERATE" in gray, sans-serif font. Below "ACCELERATE" is the word "DIAGNOSTICS" in a smaller, blue, sans-serif font with a trademark symbol.
## Fluorescence in situ Hybridization (FISH) for Identification
Cocktails of ATTO-532 (green) fluorescently-labeled DNA probes bind to the ribosomal RNA of target organisms following permeabilization. Each cocktail also includes ATTO-647 (red) labeled universal bacterial probe that binds to the ribosomal RNA of all clinically relevant bacteria (bacterial ID channels) or universal eukaryotic probe that binds to the ribosomal RNA of all clinically relevant yeast (yeast ID channels). The system images each flowcell using a fluorescence and dark-field microscope with camera and a filter set that captures emission from the FISH ID probes at 532 nm. 647 nm and in dark-field. An additional filter, capturing emission at 720 nm, is utilized prior to FISH ID for removal of interfering sample debris. To further exclude debris, only dark-field objects colocalized with universal probe signal are included in analysis. Colocalization of target probe signal and universal probe signal identifies a target organism.
The software also quantitates the total number of organisms present in a sample using a nucleic acid stain in a separate control flowcell. Comparing the relative numbers of each target organism to the number of objects lit up by the universal probes allows the system to differentiate bacteria/yeast from debris. FISH ID results are reported approximately 2 hours after loading the sample, and the ID result determines the selection of appropriate antimicrobials for subsequent antimicrobial susceptibility testing.
## Morphokinetic Cellular Analysis (MCA) for Antimicrobial Susceptibility Testing (AST)
The Accelerate Pheno™ system leverages Morphokinetic Cellular Analysis (MCA) technology to measure distinct morphokinetic features of live microbial cells responding to antimicrobials to generate susceptibility results.
MCA is a computer vision based analytical method that uses digital microscopy inputs and machine learning technology to observe individual live cells and microcolonies (or clones) and recognize patterns of change over time. This technology tracks and analyzes multiple morphological and kinetic changes of individual cells and microcolonies under a variety of conditions. These changes include morphokinetic features such as cell morphology, mass as measured by light intensity of a growing microcolony, division rate, anomalous growth patterns, and heterogeneity.
Prior to AST, the remaining sample is combined with growth media and undergoes a pregrowth step during the FISH ID assay to normalize growth rates. Following automated sample preparation, the cells are quantitated and dynamically diluted to the appropriate concentration for AST testing. The cells are then captured in flowcell channels and
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo consists of a blue geometric shape on the left, followed by the word "ACCELERATE" in gray, block letters. Below "ACCELERATE" is the word "DIAGNOSTICS" in a smaller, lighter blue font. The logo is simple and modern, with a focus on the company name.
immobilized when growth media containing single concentrations of each test antimicrobial are added to separate flowcell channels. The bacteria in each flowcell are imaged every 10 minutes for up to 4.5 hours, creating a time-lapse record of bacterial growth from individual progenitor cells into clones of daughter cells.
During this period, morphokinetic features are measured and used for analysis. The precise quantitative measurement of individual clone growth rate over time is a powerful indicator of antimicrobial efficacy. Onboard software algorithms derive minimum inhibitory concentration (MIC) values from the measured features, and apply appropriate expert rules for proper interpretation and reporting of categorical interpretations - S, I or R (susceptible, intermediate, or resistant).
The Accelerate Pheno™ system is designed to perform Accelerate PhenoTest™ BC kit identification (ID) of bacterial and yeast cells and antimicrobial susceptibility testing (AST) in approximately 7 hours directly from positive blood culture samples. Depending on the computer configuration, up to eight ID/AST modules can be operated concurrently. Analysis time may increase when four or more tests are performed on ID/AST modules simultaneously. Other factors, such as samplexity, the number of organisms and/or antimicrobials available in the panel, may also increase time to result.
## Quality Control
The Accelerate PhenoTest™ BC ID and AST OC tests automate the external QC testing procedure, removing the manual standardized inoculum preparation and manual dispensing of the McFarland standardized inoculum (0.5 for bacteria and 2.0 for fungi). This reduces the complexity of QC testing, eliminating the need for a clinical scientist to perform the test. Furthermore, the stability of the analyte is increased through the use of complementary sequences coupled to polymer microspheres for each of the target probes in the Accelerate PhenoTest™ BC kit.
As accessories to the Accelerate PhenoTest™ BC kit and Accelerate Pheno™ system, the Accelerate PhenoTest™ BC ID and AST QC tests have their own instructions for use.
## Instrument Description
Changes to the instrument description provided in the de novo submission (DEN160032) include the addition of an alternate computing system (Interface PC/Analysis module) that supports up to 8 modules and change in time to result.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo consists of a blue geometric shape on the left, followed by the word "ACCELERATE" in gray, block letters. Below "ACCELERATE" is the word "DIAGNOSTICS" in smaller, blue letters with a trademark symbol.
The Accelerate Pheno™ system is a fully-integrated in vitro diagnostic system comprised of one to eight ID/AST module(s), a computing system, touchscreen monitor and Accelerate Pheno™ system software for use with Accelerate PhenoTest™ kits. It is designed to perform identification (ID) of bacterial and yeast cells and antimicrobial susceptibility testing (AST) in approximately 7 hours directly from positive blood culture samples. Depending on the computer configuration, up to eight ID/AST modules can be operated concurrently. Analysis time may increase when four or more tests are initiated on ID/AST modules simultaneously. Other factors, such as sample complexity, the number of organisms and/or antimicrobials available in the assay kit panel, may also increase time to result.
Identification uses fluorescence in situ hybridization (FISH) and susceptibility testing uses microscopic observation of individual, live, growing bacterial cells in near real time (approximately every 10 minutes) in the presence of antimicrobial agents.
The Accelerate Pheno™ system is comprised of the following hardware:
- Accelerate Pheno™ system ID/AST modules (Up to 4 or 8 depending on . computing system architecture)
- Computing system, either: ●
- Control PC/Analysis PC setup (supports up to 4 ID/AST modules): o
- ▪
- 1 Analysis PC
- Interface PC/Analysis module setup (supports up to 8 ID/AST modules): o
- 1 Interface PC .
- . 1 Analysis module
- Touchscreen monitor ●
- Keyboard ●
- Mouse ●
- Power cords ●
- . Cables
- Uninterruptible Power Supply (1 UPS for up to 4 ID/AST modules and 1 UPS per ● computing system)
Each system contains up to 4 or 8 ID/AST modules (depending on computing system) and each ID/AST module can run one patient sample at a time. Each ID/AST module may be started or stopped at any time, independent of the other ID/AST modules.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo features a blue geometric shape on the left, followed by the word "ACCELERATE" in gray, sans-serif font. Below "ACCELERATE" is the word "DIAGNOSTICS" in a smaller, blue, sans-serif font with a trademark symbol.
#### Software Description
There are no changes to the description of the Accelerate Pheno™ system software provided in DEN160032. The Accelerate Pheno™ system uses web-based software that controls system functions. The software can be accessed via the touchscreen monitor interface or remotely on a computer or device that has a web browser with network access to the host PC.
#### Intended Use/Indications for Use
The following changes have been made to the intended use/indications for use of the Accelerate PhenoTest™ BC kit since DEN160032: addition of the Pseuodomonas aeruginosa aztreonam assay, addition of newly recognized nomenclature for Enterobacter aerogenes, removal of macrolide-lincosamide-streptogramin B resistance (MLSb), and removal of erythromycin.
The Accelerate PhenoTest™ BC kit is a multiplexed in vitro diagnostic test utilizing both qualitative nucleic acid fluorescence in situ hybridization (FISH) identification and quantitative, antimicrobial susceptibility testing (AST) methods and is intended for use with the Accelerate Pheno™ system. The Accelerate PhenoTest™ BC kit is capable of simultaneous detection and identification of multiple microbial targets followed by susceptibility testing of the appropriate detected bacterial organisms. The Accelerate PhenoTest™ BC kit is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system. Results are intended to be interpreted in conjunction with Gram stain results.
The Accelerate PhenoTest™ BC kit identifies the following Gram-positive and Gramnegative bacteria and yeasts utilizing FISH probes targeting organism-specific ribosomal RNA sequences:
Staphylococcus aureus, Staphylococcus lugdunensis, Coagulase-negative Staphylococcus species (i.e., Staphylococcus epidermidis, Staphylococcus haemolyticus, Staphylococcus hominis, Staphylococcus capitis, Staphylococcus lugdunensis, Staphylococcus warneri, not differentiated), Enterococcus faecalis, Enterococcus faecium, Streptococcus spp. (i.e., Streptococcus mitis, Streptococcus oralis, Streptococcus gallolyticus, Streptococcus agalactiae, Streptococcus pneumoniae, not differentiated), Pseudomonas aeruginosa, Acinetobacter baumannii, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii,
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo consists of a blue geometric shape on the left, followed by the word "ACCELERATE" in gray, sans-serif font. Below "ACCELERATE" is the word "DIAGNOSTICS" in a smaller, blue, sans-serif font with a trademark symbol.
Citrobacter koseri, not differentiated), Serratia marcescens, Candida albicans and Candida glabrata.
The Accelerate PhenoTest™ BC kit tests the following antimicrobial agents with the specific target organisms identified below:
Amikacin: Pseudomonas aeruginosa, Acinetobacter baumannii, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Ampicillin: Enterococcus faecalis and Enterococcus faecium
Ampicillin/Sulbactam: Escherichia coli, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), and Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated)
Aztreonam: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae. Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Ceftazidime: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Ceftaroline: Staphylococcus aureus
Cefepime: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Ceftriaxone: Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo features a blue and gray geometric shape on the left, followed by the word "ACCELERATE" in gray, sans-serif font. Below "ACCELERATE" is the word "DIAGNOSTICS" in a smaller, blue, sans-serif font with a trademark symbol.
mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Ciprofloxacin: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae. Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Daptomycin: Staphylococcus aureus, Coagulase-negative Staphylococcus species (i.e., Staphylococcus epidermidis, Staphylococcus haemolyticus, Staphylococcus hominis, Staphylococcus capitis, Staphylococcus lugdunensis, Staphylococcus warneri, not differentiated), Enterococcus faecalis and Enterococcus faecium
Ertapenem: Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Gentamicin: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae. Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Linezolid: Staphylococcus aureus, Enterococcus faecalis and Enterococcus faecium
Meropenem: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Piperacillin/Tazobactam: Pseudomonas aeruginosa, Acinetobacter baumannii, Klebsiella spp. (i.e., Klebsiella pneumoniae, Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Tobramycin: Pseudomonas aeruginosa, Klebsiella spp. (i.e., Klebsiella pneumoniae,
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo consists of a blue geometric shape on the left, resembling a stylized molecule or network. To the right of the shape is the company name, "ACCELERATE," in a bold, sans-serif font. Below "ACCELERATE" is the word "DIAGNOSTICS" in a smaller, lighter blue font, with a trademark symbol next to it.
Klebsiella oxytoca, not differentiated), Escherichia coli, Enterobacter spp. (i.e., Enterobacter cloacae, Enterobacter (Klebsiella) aerogenes, not differentiated), Proteus spp. (i.e., Proteus mirabilis, Proteus vulgaris, not differentiated), Citrobacter spp. (i.e., Citrobacter freundii, Citrobacter koseri, not differentiated) and Serratia marcescens
Vancomycin: Staphylococcus aureus, Staphylococcus lugdunensis, Coagulasenegative Staphylococcus species (i.e., Staphylococcus epidermidis, Staphylococcus haemolyticus, Staphylococcus hominis, Staphylococcus capitis, Staphylococcus lugdunensis, Staphylococcus warneri, not differentiated), Enterococcus faecalis and Enterococcus faecium
The following resistance phenotype is reported based on a qualitative test: Methicillinresistance (S. aureus S. lugdunensis, coagulase negative staphylococci).
The Accelerate PhenoTest™ BC kit is indicated as an aid in the diagnosis of bacteremia and fungemia. It is also indicated for susceptibility testing of specific pathogenic bacteria as identified above commonly associated with or causing bacteremia. Results are intended to be used in conjunction with other clinical and laboratory findings.
Standard laboratory protocols for processing positive blood cultures should be followed to ensure availability of isolates for supplemental testing as needed. Additionally, subculture of positive blood culture is necessary for the identification and susceptibility testing of: organisms not identified by the Accelerate PhenoTest™ BC kit, organisms present in polymicrobial samples, organisms for which species identification is critical for patient care (e.g. speciation of Streptococcus spp.), samples for which an "indeterminate" result for any probe was obtained, for testing antimicrobial agents not included on the Accelerate panel and for epidemiologic testing.
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo features a blue and gray geometric shape on the left, followed by the word "ACCELERATE" in a sans-serif font. Below "ACCELERATE" is the word "DIAGNOSTICS" in a smaller, lighter blue font with a trademark symbol.
# Comparison of Technological Characteristics with the Predicate Device
| Similarities | | |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| Item | Device | Predicate |
| | Accelerate PhenoTestTM BC Kit | Accelerate PhenoTestTM<br>BC Kit (DEN160032) |
| Intended Use | The Accelerate PhenoTestTM BC kit is a multiplexed <i>in vitro</i> diagnostic test utilizing both qualitative nucleic acid fluorescence <i>in situ</i> hybridization (FISH) identification and quantitative, antimicrobial susceptibility testing (AST) methods and is intended for use with the Accelerate PhenoTM system. The Accelerate PhenoTestTM BC kit is capable of simultaneous detection and identification of multiple microbial targets followed by susceptibility testing of the appropriate detected bacterial organisms. The Accelerate PhenoTestTM BC kit is performed directly on blood culture samples identified as positive by a continuous monitoring blood culture system. Results are intended to be interpreted in conjunction with Gram stain results. | Same |
| Similarities | | |
| Item | Device | Predicate |
| | Accelerate PhenoTest™ BC Kit | Accelerate PhenoTest™<br>BC Kit (DEN160032) |
| Organisms Identified | The Accelerate PhenoTest™ BC<br>kit identifies the following Gram-<br>positive and Gram-negative<br>bacteria and yeasts utilizing FISH<br>probes targeting organism-<br>specific ribosomal RNA<br>sequences: Staphylococcus<br>aureus, Staphylococcus<br>lugdunensis, Coagulase-negative<br>Staphylococcus species<br>(Staphylococcus epidermidis,<br>Staphylococcus haemolyticus,<br>Staphylococcus hominis,<br>Staphylococcus capitis,<br>Staphylococcus lugdunensis,<br>Staphylococcus warneri, not<br>differentiated), Enterococcus<br>faecalis, Enterococcus faecium,<br>Streptococcus spp. (Streptococcus<br>mitis, Streptococcus oralis,<br>Streptococcus gallolyticus,<br>Streptococcus agalactiae,<br>Streptococcus pneumoniae, not<br>differentiated), Pseudomonas<br>aeruginosa, Acinetobacter<br>baumannii, Klebsiella spp.<br>(Klebsiella pneumoniae,<br>Klebsiella oxytoca, not<br>differentiated), Escherichia coli,<br>Enterobacter spp. (Enterobacter<br>cloacae, Enterobacter | Same |
| Similarities | | |
| Item | Device<br>Accelerate PhenoTest™ BC Kit | Predicate<br>Accelerate PhenoTest™<br>BC Kit (DEN160032) |
| | (Klebsiella) aerogenes, not<br>differentiated), Proteus spp.<br>(Proteus mirabilis, Proteus<br>vulgaris, not differentiated),<br>Citrobacter spp. (Citrobacter<br>freundii, Citrobacter koseri, not<br>differentiated), Serratia<br>marcescens, Candida albicans<br>and Candida glabrata. | |
| Sample | Positive Blood Culture as<br>identified by a continuous<br>monitoring blood culture system | Same |
| Reagent Cartridge | Accelerate PhenoTest™ BC kit | Same |
| Instrument | Accelerate Pheno™ system | Same |
## Table 1: Predicate Device Comparison Similarities
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo consists of a blue geometric shape on the left, followed by the word "ACCELERATE" in gray, block letters. Below "ACCELERATE" is the word "DIAGNOSTICS" in a smaller, light blue font with a trademark symbol.
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo consists of a blue and gray geometric shape on the left, followed by the word "ACCELERATE" in gray, block letters. Below the word "ACCELERATE" is the word "DIAGNOSTICS" in a smaller, blue font with a trademark symbol.
# Table 1: Predicate Device Comparison Differences
| Differences | | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item | Device | Predicate |
| | Accelerate PhenoTestTM BC<br>Kit | Accelerate PhenoTestTM<br>BC Kit (DEN160032) |
| Antimicrobial Agents | Double concentrations of<br>aztreonam, cefepime,<br>ceftazidime and piperacillin<br>tazobactam for <i>Pseudomonas</i><br><i>aeruginosa</i> testing | Single concentrations of<br>aztreonam, cefepime,<br>ceftazidime and piperacillin<br>tazobactam |
| Differences | | |
| Item | Device<br>Accelerate PhenoTest™ BC<br>Kit | Predicate<br>Accelerate PhenoTest™<br>BC Kit (DEN160032) |
| Test Kit | Removal of MLSb and<br>erythromycin | Included MLSb and<br>erythromycin |
| | Accelerate PhenoTest™ BC kit,<br>enhanced wet reagent well<br>(consolidated wells) | Accelerate PhenoTest™ BC<br>kit, wet reagent well |
| External Quality Control<br>Assays | Accelerate PhenoTest™ BC ID<br>and AST QC test(s) | Manual QC assay |
| ATCC QC Organisms<br>Maintenance | No QC organism maintenance<br>required | Users required to maintain<br>all 20 QC strains required<br>for testing |
| QC Sample Preparation | Automated QC inoculum<br>preparation and standardization<br>performed by the Accelerate<br>Pheno™ system | User prepares standardized<br>inoculum of each QC strain<br>and manually dispenses into<br>designated wells in the<br>Accelerate PhenoTest™ BC<br>kit |
| Accessories/Materials<br>Required But Not<br>Provided For External QC<br>Testing | Accelerate PhenoTest™ BC ID<br>and AST QC test(s) | QC Assay Loading<br>"Template Tool" |
| | | ATCC QC organism(s)<br>35°C (+/- 2) Incubator with<br>CO2 |
| | | Trypticase soy agar (TSA)<br>plates containing 5%<br>sheep's blood (BAP) (for<br>bacteria growth)<br>Sabouraud Dextrose plates<br>(for yeast growth) |
| Differences | | |
| Item | Device<br>Accelerate PhenoTest™ BC Kit | Predicate<br>Accelerate PhenoTest™ BC Kit (DEN160032) |
| | | Commercially prepared<br>Trypticase soy broth (TSB) with no additives |
| | | Commercially available, calibrated turbidity meter |
| | | 0.5 and 2.0 McFarland<br>Standards for use with commercially available<br>turbidity meter |
| | | Falcon® 5mL Round<br>Bottom Polystyrene Test<br>Tubes with Snap Cap,<br>Sterile (Corning Product # 352054) or equivalent |
| Test Interpretation and<br>Results Reporting | Modified Expert rules | Original Expert rules from<br>DEN160032 |
| | ID Interpretive Rules increase<br>reporting of the monomicrobial<br>call and decreases incidence of<br>indeterminate calls and<br>ambiguous results due to<br>debris/noise | Original Monomicrobial<br>Indeterminate and false<br>positive in DEN160032 |
| | AO Bright Rule improves<br>overall reportability by<br>decreasing invalid results | Original invalid results in<br>DEN160032 |
| Differences | | |
| Item | Device<br>Accelerate PhenoTestTM BC Kit | Predicate<br>Accelerate PhenoTestTM BC Kit (DEN160032) |
| | Noise Rejection analysis decreases by-run false positive rate of clinical stock isolates | By-run false positive rate for clinical stock isolates in DEN160032 |
| | Improved ID target detection thresholds | Original normal hit, bright rule and high sensitivity detection in DEN160032 |
| | MEM algorithm changes decrease MIN error rate for PAE | Original MIN error rate for PAE |
| Time to Result | ID Results ~2 hours<br>AST Results ~7 hours | ID Results ~1.5 hours<br>AST Results ~6.5 hours |
| Instrument | Optical set submitted in DEN160032 with additional far-red filter | Optical set submitted in DEN160032 |
| Computing System | Add Interface PC/Analysis module setup (supports up to 8 ID/AST modules) | Control PC/Analysis PC setup (supports up to 4 ID/AST modules) |
| Software Algorithms | Add ID interpretive rules, AO bright rule, and ID algorithm improvements. Updated algorithms for aztreonam, cefepime, ceftazidime, meropenem and piperacillin-tazobactam with <i>P. aeruginosa</i> | Original algorithms for aztreonam, cefepime, ceftazidime, meropenem and piperacillin-tazobactam with <i>P. aeruginosa</i> |
| Software Algorithms | Accelerate PhenoTM system software includes noise rejection analysis using far red filter to reject interfering debris | Noise rejection analysis using far red filter to reject interfering debris not performed |
| Differences | | |
| Item | Device<br>Accelerate PhenoTest™ BC<br>Kit | Predicate<br>Accelerate PhenoTest™<br>BC Kit (DEN160032) |
| Software QC Assays | Accelerate Pheno™ system<br>software automates QC testing<br>for ID and AST using the<br>Accelerate PhenoTest™ BC<br>ID/AST QC tests | QC assays performed by<br>Accelerate Pheno™ system<br>requires manual dispensing<br>of standardized inoculum<br>into designated wells of<br>Accelerate PhenoTest™ BC<br>kit |
{18}------------------------------------------------
Image /page/18/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo features a blue geometric shape on the left, resembling a stylized hexagon or diamond. To the right of the shape, the word "ACCELERATE" is written in a bold, sans-serif font. Below "ACCELERATE", the word "DIAGNOSTICS" is written in a smaller, lighter blue font, with a trademark symbol next to it.
{19}------------------------------------------------
Image /page/19/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo features a blue geometric shape on the left, resembling a stylized hexagon or diamond. To the right of the shape, the word "ACCELERATE" is written in a bold, sans-serif font. Below "ACCELERATE", the word "DIAGNOSTICS" is written in a smaller, lighter font.
{20}------------------------------------------------
Image /page/20/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo consists of a blue geometric shape on the left, followed by the word "ACCELERATE" in gray, block letters. Below the word "ACCELERATE" is the word "DIAGNOSTICS" in smaller, blue letters.
{21}------------------------------------------------
Image /page/21/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo features a blue geometric shape on the left, resembling a stylized hexagon or diamond. To the right of the shape, the word "ACCELERATE" is written in a bold, sans-serif font. Below "ACCELERATE", the word "DIAGNOSTICS" is written in a smaller, lighter font.
{22}------------------------------------------------
Image /page/22/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo features a blue geometric shape on the left, resembling a stylized diamond or hexagon made up of smaller shapes. To the right of the shape is the company name, "ACCELERATE," in a bold, sans-serif font. Below "ACCELERATE" is the word "DIAGNOSTICS" in a smaller, sans-serif font, with a trademark symbol next to it.
## Performance Data
Performance data are provided within this premarket notification in support of the substantial equivalence determination.
## Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing is identical to that conducted for the predicate device. The Accelerate Pheno™ system complies with IEC 61010-1:2010, Safety requirements for electrical equipment for measurement, control, and laboratory use- Part 1: General requirements, and also with IEC 60601-1-2:2014, medical electrical equipment: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation is provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered to have a "moderate" level of concern, since a failure or latent flaw could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider.
#### Non-Clinical Test Summaries
#### Performance Evaluation
Performance of the modified Accelerate PhenoTest™ BC kit aztreonam (ATM), ceftazidime (CAZ), cefepime (FEP), meropenem (MEM) and piperacillin-tazobactam (TZP) Pseudomonas aeruginosa assays was established during an internal performance evaluation study at Accelerate Diagnostics, Inc. A total of 131 characterized Pseudomonas aeruginosa isolates were tested at the internal site in two phases along with evaluation of the new algorithms on the original 13 clinical trial isolates for a total of 144 isolates included in the performance evaluation. Phase 1 included testing of 100 challenge and stock isolates (115 total runs), and Phase 2 evaluated an additional 31 onscale isolates. P. aeruginosa isolates were seeded into BD BACTEC™ Plus Aerobic blood culture bottles containing healthy donor blood and incubated until positivity on the BD BACTECTM system. Samples from positive blood culture bottles were Gramstained, run on the Accelerate Pheno™ system, sub-cultured for purity on trypticase soy agar (TSA) plates, and images were taken with Sphere FLASH®. Antimicrobial susceptibility test (AST) results were compared to historical results from the broth
{23}------------------------------------------------
Image /page/23/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo consists of a blue geometric shape on the left, followed by the word "ACCELERATE" in gray, block letters. Below "ACCELERATE" is the word "DIAGNOSTICS" in a smaller, blue font.
microdilution (BMD) reference method. Quality control testing for the Accelerate PhenoTest™ BC kit was performed daily using the Accelerate PhenoTest™ BC ID and AST QC tests.
During the Phase 1 performance evaluation study, a total of 113 out of the 115 tested samples produced a positive identification for Pseudomonas aeruginosa. Two of the P. aeruginosa samples produced negative or off-panel ID calls and were excluded from analysis per study exclusion criteria. Of the 113 positive P. aeruginosa samples, 13 did not proceed to AST testing due to invalid results (e.g., growth control channel failures) and were additionally excluded from data analysis. In Phase 2, aside from five isolates that were retested once each due to ID failures, all experiments resulted in valid data points for all five beta lactam antimicrobials.
The primary objective of the clinical evaluation study for the Accelerate PhenoTest™ BC kit aztreonam, ceftazidime, meropenem and piperacillin-tazobactam Pseudomonas aeruginosa assays was to produce adequate essential and categorical agreement. At least two two-fold dilutions below the susceptible and one two-fold dilution above the resistant threshold as specified in the FDA Class II Special Controls Guidance document were tested over the Accelerate Pheno™ system reportable ranges of ≤2 µg/mL to ≥64 µg/mL (ATM, CAZ, FEP), ≤1 µg/mL to ≥16 µg/mL (MEM), ≤4 ug/mL to ≥256 ug/mL (TZP). Table 3 provides an overall summary of AST performance for the two testing phases plus analysis of the 13 original clinical isolates for the Accelerate PhenoTest™ BC kit aztreonam, cefepime, ceftazidime, meropenem and piperacillin-tazobactam Pseudomonas aeruginosa assays.
{24}------------------------------------------------
Image /page/24/Picture/0 description: The image shows the logo for Accelerate Diagnostics. The logo consists of a blue diamond shape on the left, followed by the word "ACCELERATE" in gray, sans-serif font. Below the word "ACCELERATE" is the word "DIAGNOSTICS" in a smaller, blue, sans-serif font. The logo is clean and modern, and the colors are professional and trustworthy.
K192665 September 9, 2020
| Abx | Reporting<br>Range | N | #EA | %EA | N<br>(Eval) | #EA<br>(Eval) | %EA<br>(Eval) | #CA | %CA | #R | #S | #vmj | %vmj | #maj | %maj | #min | %min |
|----------|--------------------|-----|-----|------|-------------|---------------|---------------|-----|------|----|-----|------|------|------|------|------|------|
| ATMa | 2-64 | 144 | 135 | 93.8 | 73 | 64 | 87.7 | 134 | 93.1 | 35 | 105 | 0 | 0 | 1 | 1.0 | 9 | 6.3 |
| FEPa,b,d | 2-64 | 143 | 136 | 95.1 | 40 | 33 | 82.5 | 132 | 92.3 | 36…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.