Prospective clinical rectal swab specimens from hospitalized or long-term care patients
The study evaluated the clinical performance (PPA and NPA) of the Xpert Carba-R Assay relative to a reference method of culture and DNA sequence analysis using prospectively collected clinical specimens.
Rectal swab specimens; Clinical performance; Prospective collection; Infection control
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Multi-site investigational study; Prospective, multi-site clinical study
Patients hospitalized or in a long-term care facility; Sample Size: 755 eligible specimens; Number of Sites: 5 (3 in the US, 2 in Europe)
Reference method of culture (MacConkey enrichment broth) and PCR/bi-directional DNA sequence analysis
Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) for detection of blaKPC, blaNDM, blaVIM, and blaOXA-48 gene sequences
Indications for Use
The Xpert® Carba-R Assay, performed on the GeneXpert® Instrument Systems, is a qualitative in vitro diagnostic test designed for the detection and differentiation of the blakpc, blaNDM, blaym, blaoxA-48, and blamp gene sequences associated with carbapenem-non-susceptibility. The test utilizes automated real-time polymerase chain reaction (PCR). The Xpert Carba-R Assay is intended as an aid to infection control in the detection of carbapenem-non-susceptible bacteria that colonize patients in healthcare settings. The Xpert Carba-R Assay is not intended to guide or monitor treatment for carbapenem-nonsusceptible bacterial infections. A negative Xpert Carba-R Assay result does not preclude the presence of other resistance mechanisms. The Xpert Carba R-Assay is for use with the following sample types: Rectal Swab Specimens The assay is performed on rectal swab specimens from patients at risk for intestinal colonization with carbapenem-non-susceptible bacteria. Concomitant cultures are necessary to recover organisms for epidemiological typing, antimicrobial susceptibility testing, and for further confirmatory bacterial identification. Pure Colonies The assay is performed on carbapenem-non-susceptible pure colonies of Enterobacteriaceae, Acinetobacter baumannii, or Pseudomonas aeruginosa, when grown on blood agar or MacConkey agar. For testing pure colonies, the Xpert Carba-R Assay should be used in conjunction with other laboratory tests including phenotypic antimicrobial susceptibility testing.
Device Story
Automated real-time PCR assay for qualitative detection of carbapenemase gene sequences (blaKPC, blaNDM, blaVIM, blaOXA-48, blaIMP). Input: rectal swab specimens resuspended in sample reagent. Process: single-use, multi-chambered fluidic cartridge performs automated sample preparation, DNA extraction, amplification, and real-time detection on GeneXpert Instrument Systems. Output: detected/not detected results for each target gene. Used in healthcare settings for infection control monitoring. Healthcare providers use results to identify colonized patients; does not guide treatment. Benefits: rapid (approx. 50 min) identification of resistance markers to support infection control measures.
Clinical Evidence
Multi-site prospective clinical study (n=755 eligible specimens) and contrived specimen study (n=432). Compared against culture and bi-directional DNA sequencing. Prospective PPA ranged 60.0%-100% and NPA 98.6%-99.9% across targets. Contrived PPA 95%-100%, NPA 100%. Reproducibility study demonstrated high agreement across sites, operators, and lots.
Technological Characteristics
Automated real-time PCR; disposable multi-chambered fluidic cartridges; ultrasonic horn for cell lysis; I-CORE® thermocycler; TaqMan® probes; internal Sample Processing Control (SPC) and Probe Check Control (PCC). Standalone instrument system (GeneXpert Dx/Infinity).
Indications for Use
Indicated for patients at risk for intestinal colonization with carbapenem-non-susceptible bacteria. Used for qualitative detection and differentiation of blaKPC, blaNDM, blaVIM, blaOXA-48, and blaIMP gene sequences from rectal swab specimens. Not for bacterial identification or reporting susceptibility status. Not for guiding/monitoring treatment.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
K160901
B. Purpose for Submission:
To obtain a substantial equivalence determination for the Xpert® Carba-R Assay on the Cepheid GeneXpert Instrument Systems (GeneXpert Dx, GeneXpert Infinity-48, GeneXpert Infinity-48s, and GeneXpert Infinity-80 systems) in the qualitative detection of the $bla_{\mathrm{KPC}}$, $bla_{\mathrm{NDM}}$, $bla_{\mathrm{VIM}}$, $bla_{\mathrm{OXA-48}}$, and $bla_{\mathrm{IMP}}$ gene sequences from rectal swab specimens.
C. Measurand:
Target DNA sequence of the following genes:
$bla_{\mathrm{KPC}}$, $bla_{\mathrm{NDM}}$, $bla_{\mathrm{VIM}}$, $bla_{\mathrm{OXA-48}}$, and $bla_{\mathrm{IMP}}$
D. Type of Test:
Qualitative real-time polymerase chain reaction (PCR) assay
E. Applicant:
Cepheid
F. Proprietary and Established Names:
Proprietary Name: Xpert® Carba-R
Common Name: Xpert Carba-R Assay
G. Regulatory Information:
1. Regulation section:
21 CFR 866.1640 (Antimicrobial susceptibility test powder)
2. Classification:
Class II
{1}
3. Product code:
POC—System, nucleic acid amplification test, DNA, antimicrobial resistance marker, direct specimen
OOI—Real-time nucleic acid amplification system
4. Panel:
83-Microbiology
H. Intended Use:
1. Intended use(s):
The Xpert® Carba-R Assay, performed on the GeneXpert® Instrument Systems, is a qualitative in vitro diagnostic test designed for the detection and differentiation of the $bla_{\mathrm{KPC}}$, $bla_{\mathrm{NDM}}$, $bla_{\mathrm{VIM}}$, $bla_{\mathrm{OXA-48}}$, and $bla_{\mathrm{IMP}}$ gene sequences associated with carbapenem-non-susceptibility. The test utilizes automated real-time polymerase chain reaction (PCR).
The Xpert Carba-R Assay is intended as an aid to infection control in the detection of carbapenem-non-susceptible bacteria that colonize patients in healthcare settings. The Xpert Carba-R Assay is not intended to guide or monitor treatment for carbapenem-non-susceptible bacterial infections. A negative Xpert Carba-R Assay result does not preclude the presence of other resistance mechanisms.
The Xpert Carba R-Assay is for use with the following sample types:
Rectal Swab Specimens
The assay is performed on rectal swab specimens from patients at risk for intestinal colonization with carbapenem-non-susceptible bacteria. Concomitant cultures are necessary to recover organisms for epidemiological typing, antimicrobial susceptibility testing, and for further confirmatory bacterial identification.
Pure Colonies
The assay is performed on carbapenem non-susceptible pure colonies of Enterobacteriaceae, Acinetobacter baumannii, or Pseudomonas aeruginosa, when grown on blood agar or MacConkey agar. For testing pure colonies, the Xpert Carba-R Assay should be used in conjunction with other laboratory tests including phenotypic antimicrobial susceptibility testing.
2. Indication(s) for use:
Same as the Intended Use.
{2}
3. Special conditions for use statement(s):
For prescription use only.
The Xpert Carba-R Assay detects $bla_{\mathrm{KPC}}$, $bla_{\mathrm{NDM}}$, $bla_{\mathrm{VIM}}$, $bla_{\mathrm{OXA-48}}$, and $bla_{\mathrm{IMP}}$ gene sequences from rectal swab specimens and is not for bacterial identification or to report susceptibility status.
The detection of assay targets by the Xpert Carba-R Assay does not indicate the presence of viable organisms containing the resistance marker.
The Xpert Carba-R Assay is not a sub-typing tool and does not report variants of the $bla_{\mathrm{KPC}}$, $bla_{\mathrm{NDM}}$, $bla_{\mathrm{VIM}}$, $bla_{\mathrm{OXA-48}}$, and $bla_{\mathrm{IMP}}$ genes.
Detection of $bla_{\mathrm{KPC}}$, $bla_{\mathrm{NDM}}$, $bla_{\mathrm{VIM}}$, $bla_{\mathrm{OXA-48}}$, and/or $bla_{\mathrm{IMP}}$ gene sequences does not indicate the presence of viable organisms with these markers in the specimen.
4. Special instrument requirements:
The Xpert Carba-R Assay uses PCR technology on the GeneXpert Instrument Systems, which extract, amplify, and detect the target DNA.
I. Device Description:
The Xpert Carba-R Assay is an automated real-time polymerase chain reaction (PCR) in vitro diagnostic test for the qualitative detection of the $bla_{\mathrm{KPC}}$, $bla_{\mathrm{NDM}}$, $bla_{\mathrm{VIM}}$, $bla_{\mathrm{OXA-48}}$, and $bla_{\mathrm{IMP}}$ gene sequences from rectal swab specimens. The Xpert Carba-R Assay is intended as an aid for infection control to monitor the spread of carbapenem-non-susceptible organisms in healthcare settings. The assay is performed on the Cepheid GeneXpert Instrument Systems (GeneXpert Dx, GeneXpert Infinity-48, GeneXpert Infinity-48s, and GeneXpert Infinity-80 systems). The GeneXpert Instrument Systems utilize single-use, disposable cartridges (Xpert Carba-R cartridges) containing PCR reagents that allow for automated sample preparation, amplification, and real-time detection of gene targets in approximately 50 minutes. A Sample Processing Control (SPC) and a Probe Check Control (PCC) have been incorporated into the assay design to address key failure modes that could result in a false negative determination.
The GeneXpert Instrument Systems (GeneXpert Dx Systems and the GeneXpert Infinity Systems) have 1 to 80 randomly accessible modules, depending upon the instrument, that are each capable of performing separate sample processing and real-time PCR tests. Because the cartridges are self-contained and specimens never come into contact with working parts of the instrument modules, cross-contamination between samples is minimized.
J. Substantial Equivalence Information:
1. Predicate device name(s):
{3}
Xpert Carba-R Assay
2. Predicate 510(k) number(s):
K152614
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate Device |
| | Cepheid Xpert Carba-R Assay (K160901) | Cepheid Xpert Carba-R Assay (K152614) |
| Intended Use | The Xpert® Carba-R Assay, performed on the GeneXpert® Instrument Systems, is a qualitative in vitro diagnostic test designed for the detection and differentiation of the blaKPC, blaNDM, blaVIM, blaOXA-48, and blaIMP gene sequences associated with carbapenem-non-susceptibility. The test utilizes automated real-time polymerase chain reaction (PCR).The Xpert Carba-R Assay is intended as an aid to infection control in the detection of carbapenem-non-susceptible bacteria that colonize patients in healthcare settings. The Xpert Carba-R Assay is not intended to guide or monitor treatment for carbapenem-non-susceptible bacterial infections. A negative Xpert Carba-R Assay result does not preclude the presence of other resistance mechanisms.The Xpert Carba R-Assay is for use with the following sample types:Rectal Swab SpecimensThe assay is performed on rectal swab specimens from patients at risk for intestinal colonization with carbapenem-non-susceptible bacteria. Concomitant cultures are necessary to recover organisms for epidemiological typing, antimicrobial susceptibility testing, and for further confirmatory bacterial identification. | The Xpert® Carba-R Assay, performed on the GeneXpert® Instrument Systems, is a qualitative in vitro diagnostic test for the detection and differentiation of the blaKPC, blaNDM, blaVIM, blaOXA-48, and blaIMP gene sequences associated with carbapenem-non-susceptible pure colonies of Enterobacteriaceae, Acinetobacter baumannii, or Pseudomonas aeruginosa grown on blood agar or MacConkey agar. The test utilizes automated real-time polymerase chain reaction (PCR).A negative Xpert Carba-R Assay result does not preclude the presence of other resistance mechanisms. The Xpert Carba-R Assay should be used in conjunction with other laboratory tests including phenotypic antimicrobial susceptibility testing.The Xpert Carba-R Assay is intended as an aid for infection control in detecting and differentiating genetic markers of resistance to monitor the spread of carbapenem-non-susceptible organisms in healthcare settings. The Xpert Carba-R Assay is not intended to guide or monitor treatment for carbapenem-non-susceptible bacterial infections. |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate Device |
| | Cepheid Xpert Carba-R Assay (K160901) | Cepheid Xpert Carba-R Assay (K152614) |
| | Pure Colonies The assay is performed on carbapenem-non-susceptible pure colonies of Enterobacteriaceae, Acinetobacter baumannii, or Pseudomonas aeruginosa, when grown on blood agar or MacConkey agar. For testing pure colonies, the Xpert Carba-R Assay should be used in conjunction with other laboratory tests including phenotypic antimicrobial susceptibility testing. | |
| Technological Principles | Same | Fully-automated nucleic acid amplification (DNA); real-time PCR |
| Test Cartridge | Same | Disposable single-use, multi-chambered fluidic cartridge |
| Detection Probes | Same | TaqMan® Probes |
| Controls | Same | Internal sample processing control (SPC); Probe check control (PCC); External controls available |
| Assay Targets | Same | blaKPC, blaNDM, blaVIM, blaOXA-48, and blaIMP gene sequences |
| Instrument System | Same | GeneXpert Instrument System (includes GeneXpert Dx, Infinity-48, Infinity-48s, and Infinity-80) |
| Time to obtain test results | Same | Approximately 50 minutes to results |
| Interpretation of test results | Same | Diagnostic software of the GeneXpert Instrument System |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Sample Type | Rectal swab specimens | Bacterial isolates from culture |
| Organism(s) | N/A | Carbapenem non-susceptible Enterobacteriaceae, Pseudomonas aeruginosa, Acinetobacter baumannii |
# K. Standard/Guidance Document Referenced (if applicable):
1. ASTM D4169-09, Standard Practice for Performance Testing of Shipping Containers and Systems.
2. CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline—Second Edition, 2004
3. CLSI EP15-A2, User Verification of Performance for Precision and Trueness; Approved Guideline—Second Edition, 2006
{5}
4. CLSI M02-A11. Performance standards for Antimicrobial Disk Susceptibility Tests; Eleventh Edition, 2012
5. CLSI M07-A9. Methods for dilution antimicrobial susceptibility tests for bacteria that grow aerobically; Approved standard—Ninth Edition, 2012
6. CLSI M07-A10. Methods for dilution antimicrobial susceptibility tests for bacteria that grow aerobically; Approved standard—Tenth Edition, 2015
7. CLSI M100-S24. Performance Standard for Antimicrobial Susceptibility Testing; Approved standard—Twenty-fourth Informational Supplement, 2014
8. CLSI MM3-A2, Molecular Diagnostic Methods for Infectious Disease; Approved Guideline—Second Edition, 2006
9. EN 13640, Stability Testing of in vitro Diagnostic Reagents, June 2002
10. General Principles of Software Validation; Final Guidance for Industry and FDA Staff, issued January 11, 2002
11. Guidance for Industry and FDA Staff—Format for Traditional and Abbreviated 510(k)s, issued August 12, 2005
12. Guidance for Industry and FDA Staff—Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems, issued on March 10, 2005
13. Guidance for Industry and FDA Staff—Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, issued on October 2, 2014
14. Guidance for Industry—Cybersecurity for Networked Medical Devices Containing Off-the-Shelf (OTS) Software, issued January 14, 2005
15. Guidance for Industry and FDA Staff—Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued May 11, 2005
16. Guidance for Industry, FDA Reviewers and Compliance on Guidance for Off-the-Shelf Software Use in Medical Devices; issued September 9, 1999
17. Guidance for Sponsors, Institutional Review Boards, Clinical Investigators and FDA Staff—Guidance on Informed Consent for In Vitro Diagnostic Device Studies Using Leftover Human Specimens that are Not Individually Identifiable, issued April 25, 2006
L. Test Principle:
The Xpert Carba-R Assay cartridges contain reagents for the detection of $bla_{\mathrm{KPC}}$, $bla_{\mathrm{NDM}}$, $bla_{\mathrm{VIM}}$, $bla_{\mathrm{OXA-48}}$, and $bla_{\mathrm{IMP}}$ gene sequences from rectal swab specimens. Each swab is resuspended in $5\,\mathrm{ml}$ of Sample Reagent. The sample is vortexed then transferred $(1.7\,\mathrm{ml})$ to the sample chamber of a disposable Xpert Carba-R Assay cartridge. The user initiates a test from the system-user interface and places the cartridge into the GeneXpert instrument platform, which performs hands-off real-time, multiplex PCR for the detection of target sequences.
Results of the assay run are interpreted by the GeneXpert Instrument System software from measured fluorescent signals and embedded calculation algorithms. The results are automatically generated at the end of the process in a report that can be viewed and printed. Basic users can see test results reported as "red" highlighted for DETECTED results and "green" highlighted for NOT DETECTED results. Additional results that can be reported include: INVALID, ERROR, and NO RESULT.
{6}
# Interpretation of Results
The Xpert Carba-R Assay provides test results for the IMP, VIM, NDM, KPC, and OXA-48 target DNA sequences. A Sample Processing Control (SPC) and a Probe Check Control (PCC) have been designed for the assay as internal controls to enable the GeneXpert Instrument System to detect specific failure modes related to assay performance. The PCC is considered to PASS if the fluorescence generated meets the validated acceptance criteria. If the PCC fails for any of the IMP, VIM, NDM, KPC, and OXA-48 targets, or SPC target, a probe check error is reported and the test will not continue. The assay also reports if the test has an INVALID, ERROR or NO RESULT. Under these conditions, the test will need to be repeated using a new sample, a new cartridge, and/or new reagents. Retest procedures are described in the Xpert Carba-R package insert. An interpretation table for test results is shown in Table 1.
Table 1. Interpretation of Test Results for the Xpert Carba-R Assay
| Result Report | Interpretation of Results |
| --- | --- |
| DETECTED | Target(s): For a valid “DETECTED” test result, PCR amplification of the target DNA sequence gives Ct value(s) within the valid range and a fluorescence endpoint above the threshold setting for IMP, VIM, NDM, KPC, and/or OXA-48; SPC: Not applicable (if at least one target detected); PCC: PASS; all probe check results pass. |
| NOT DETECTED | Target(s): For a valid “NOT DETECTED” test result, no valid Ct(s) are reported for the IMP, VIM, NDM, KPC, and/or OXA-48 target DNA sequences; SPC: PASS, PCR amplification of the SPC DNA sequence gives a Ct value within the valid range and a fluorescence endpoint above the threshold setting; PCC: PASS; all probe check results pass. |
| ERROR | Target(s): Presence or absence of IMP, VIM, NDM, KPC, and OXA-48 target DNA sequences cannot be determined; SPC: NO RESULT; PCC: FAIL*, one or more of the probe check results failed. *If the probe check passed, the error is caused by a system component failure. |
| INVALID | Target(s): Presence or absence of IMP, VIM, NDM, KPC, and OXA-48 target DNA sequences cannot be determined; SPC: FAIL, No PCR amplification of the SPC DNA sequence or the SPC Ct is not within valid range and the fluorescence endpoint is below threshold setting; PCC: PASS; all probe check results pass. |
| NO RESULT | Target(s): Presence or absence of IMP, VIM, NDM, KPC, and OXA-48 target DNA sequences cannot be determined; SPC: NO RESULT; PCC: Not applicable. A “NO RESULT” indicates that insufficient data were collected. For example, the operator stopped a test that was in progress or a power failure occurred. |
M. Performance Characteristics (if/when applicable):
{7}
# 1. Analytical performance:
# a. Precision/Reproducibility:
The reproducibility of the Xpert Carba-R Assay was established in a multi-center study. An 11-member panel was tested that included: 1) two different organisms for each of the five resistance genes detected by the Xpert Carba-R Assay and 2) one negative sample for all five gene targets. Each organism was spiked into a pooled negative rectal matrix at low positive ( $\sim$ 1x LoD) and moderate positive levels (2-3x LoD). To measure site-to-site reproducibility, the 11-member panel was tested in replicates of four each day at three (3) sites over a six day testing period with two operators per site. Three lots of Xpert Carba-R Assay cartridges were used at each testing site. A total of 1584 samples suspended in rectal matrix were evaluated, where 144 replicates for each of the 11 different panel members were tested. In rectal matrix, $99.4\%$ (1574/1584) of samples were successful and produced the expected result on the first attempt. Seven (7) ERROR cases, one (1) INVALID result, and two (2) NO RESULT outcomes were reported. All ten samples yielded valid results upon repeat testing. The results of the reproducibility study are summarized in Table 2 below.
Table 2. Reproducibility of the 11-Member Sample Panel
| Resistance Gene | Site 1a | | | Site 2 | | | Site 3 | | | % Total Agreement by Sample |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Op 1 | Op 2 | Site | Op 1 | Op 2 | Site | Op 1 | Op 2 | Site | |
| Neg | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (144/144) |
| IMP Mod Pos | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (144/144) |
| IMP Low Pos | 91.7% (22/24) | 87.5% (21/24) | 89.5% (43/48) | 83.3% (20/24) | 87.5% (21/24) | 85.4% (41/48) | 87.5% (21/24) | 79.2% (19/24) | 83.3% (40/48) | 86.1% (124/144) |
| VIM Mod Pos | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (144/144) |
| VIM Low Pos | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (144/144) |
| NDM Mod Pos | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (144/144) |
| NDM Low Pos | 91.7% (22/24) | 95.8% (23/24) | 93.8% (45/48) | 95.8% (23/24) | 95.8% (23/24) | 95.8% (46/48) | 100% (24/24) | 91.7% (22/24) | 95.8% (46/48) | 95.1% (137/144) |
| KPC Mod Pos | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (144/144) |
| KPC Low Pos | 95.8% (23/24) | 100% (24/24) | 97.9% (47/48) | 100% (24/24) | 91.7% (22/24) | 95.8% (46/48) | 95.8% (23/24) | 95.8% (23/24) | 95.8% (46/48) | 96.5% (139/144) |
| OXA-48 Mod Pos | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (24/24) | 100% (24/24) | 100% (48/48) | 100% (144/144) |
| OXA-48 Low Pos | 95.8% (23/24) | 100% (24/24) | 97.9% (47/48) | 95.8% (23/24) | 100% (24/24) | 97.9% (47/48) | 91.7% (22/24) | 100% (24/24) | 95.8% (46/48) | 97.2% (140/144) |
${}^{a}$ Each site tested one GeneXpert Instrument System—GeneXpert Dx,Infinity-80,or Infinity-48.
The reproducibility of the Xpert Carba-R Assay was also evaluated by assessing the fluorescent signal (expressed in Ct values) for each target detected. The mean, standard deviation (SD), and coefficient of variation (CV) between sites, days, and
{8}
operators for each panel member are presented in Table 3 below.
Table 3. Reproducibility of the Fluorescent Signal
| Resistance Gene (sample number) | Assay Channel (Analyte) | Na | Mean Ct | Between Site | | Between Lot | | Between Day | | Between Operator | | Within Assay | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) | SD | CV (%) |
| Neg | SPC | 144 | 32.9 | 0.2 | 0.5 | 0.2 | 0.7 | 0.0 | 0.1 | 0.0 | 0 | 0.6 | 1.8 | 0.7 | 2.0 |
| IMP Mod Pos | IMP | 144 | 34.5 | 0.0 | 0.0 | 0.2 | 0.5 | 0 | 0.0 | 0.1 | 0.2 | 0.7 | 2.0 | 0.7 | 2.1 |
| IMP Low Pos | IMP | 140 | 36.4 | 0.0 | 0.0 | 0.0 | 0.0 | 0.2 | 0.5 | 0.0 | 0 | 1.2 | 3.3 | 1.2 | 3.4 |
| VIM Mod Pos | VIM | 144 | 31.0 | 0.0 | 0.0 | 0.3 | 0.9 | 0 | 0.0 | 0.2 | 0.5 | 0.5 | 1.6 | 0.6 | 1.9 |
| VIM Low Pos | VIM | 144 | 33.8 | 0.0 | 0.0 | 0.6 | 1.8 | 0.3 | 0.9 | 0.3 | 1.0 | 1.4 | 4.0 | 1.6 | 4.6 |
| NDM Mod Pos | NDM | 144 | 33.7 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 | 0.1 | 0.0 | 0.0 | 0.6 | 1.7 | 0.6 | 1.7 |
| NDM Low Pos | NDM | 143 | 36.2 | 0.2 | 0.7 | 0.0 | 0.0 | 0.3 | 0.7 | 0.0 | 0.0 | 0.8 | 2.3 | 0.9 | 2.5 |
| KPC Mod Pos | KPC | 144 | 34.2 | 0.0 | 0.0 | 0.3 | 0.8 | 0.2 | 0.6 | 0.0 | 0.0 | 0.4 | 1.2 | 0.6 | 1.6 |
| KPC Low Pos | KPC | 141 | 35.8 | 0.0 | 0.0 | 0.5 | 1.5 | 0.0 | 0.0 | 0.3 | 0.9 | 0.7 | 1.9 | 0.9 | 2.6 |
| OXA-48 Mod Pos | OXA-48 | 144 | 34.3 | 0.0 | 0.0 | 0.2 | 0.5 | 0.2 | 0.5 | 0.1 | 0.3 | 0.5 | 1.6 | 0.6 | 1.7 |
| OXA-48 Low Pos | OXA-48 | 143 | 36.1 | 0.0 | 0.0 | 0.0 | 0.0 | 0.2 | 0.6 | 0.0 | 0.0 | 0.8 | 2.3 | 0.9 | 2.4 |
${}^{a}$ Results with non-zero Ct values out of 144.
At $\sim 1\mathrm{x}$ LoD, the expected target was not detected in some samples (See Table 2 above). The lowest $\%$ total agreement, $86.1\%$ (124/144), was observed with the low positive IMP sample. All moderate positive samples (2-3x LoD) gave the expected result in rectal matrix. All negative samples were correctly identified as negative (144/144). A total of 76 control samples were run with only 1 INVALID reported; all remaining positive and negative controls gave the expected results. Agreement between sites, operators, and lots was evaluated using Fisher's Exact test. The data presented in Table 2 and Table 3 demonstrated an acceptable reproducibility for the Xpert Carba-R Assay on the GeneXpert Instrument Systems.
# b. Linearity/assay reportable range:
Not applicable.
# c. Traceability, Stability, Expected values (controls, calibrators, or methods):
# External Controls
Commercially-available external controls can also be run in accordance with local, state, and federal accrediting organizations, as applicable. For each day of the study, an external negative control and two types of experimental positive controls were tested. One external positive control consisted of $E$ coli cells containing a plasmid with an insert carrying amplicon sequences from all five Xpert Carba-R target analyte genes (Multivalent External Positive Control). The second external positive control consisted of individual carbapenemase-producing bacteria, each harboring only one of the Xpert Carba-R target carbapenemase genes. On each testing day, one negative control, the five-gene construct positive control, and two of five individual bacterial controls were tested. Of the 970 external control samples run, $99.2\%$ (962/970) gave a valid result on the first attempt. Re-testing of the eight indeterminate controls gave the expected results.
{9}
External controls include the following bacteria harboring target genes:
Multivalent External Positive Control—External positive control (inactivated Escherichia coli carrying plasmid with KPC, NDM, VIM, IMP, and OXA-48 gene sequences)
Individual Positive Controls
- K. pneumoniae KPC (ATCC BAA-1705)
- K. pneumoniae NDM (ATCC BAA-2146)
- K. pneumoniae VIM (NCTC 13439)
- K. pneumoniae OXA-48 (NCTC 13442)
- Escherichia coli IMP (NCTC 13476)
External Negative Control—Inactivated E. coli containing a plasmid with no resistance gene inserted.
## Internal Control (IC) Reaction Analysis
Internal controls enable the system to detect specific failure modes that could potentially result in an incorrect test result. Each Xpert Carba-R Assay includes a Sample Processing Control (SPC) and Probe Check Control (PCC) pre-loaded in the cartridge and provided with the assay.
### Sample Processing Control (SPC)
The SPC contains Bacillus globigii that is included in each cartridge to verify adequate processing of the sample. The SPC verifies that lysis of bacteria has occurred if the organisms are present and verifies the effectiveness of each sample preparation step—reaction tube filling, reaction components are present and functioning, and monitoring for the presence of potential inhibitor(s) in the PCR assay. Test results are reported as INVALID if the SPC fails to meet the valid minimum or maximum Ct specification.
### Probe Check control (PCC)
The PCC verifies reagent rehydration, real-time PCR tube filling in the cartridge, probe integrity, and dye stability. The PCC is considered to PASS if the fluorescence generated meets the validated acceptance criteria. If the PCC fails for any of the IMP, VIM, NDM, KPC, and OXA-48 targets or SPC target, a probe check error is reported and the test will not continue. If a probe check error is reported, the test may be repeated using a new sample, new cartridge, and new reagents.
## Stability Studies
A number of studies were conducted to establish the specimen stability of rectal swabs with the Xpert Carba-R Assay. These studies were performed to 1) establish the specimen stability for rectal swabs after specimen collection but prior to transfer to the Xpert Carba-R Sample Reagent and 2) establish the specimen stability of rectal swabs eluted into Xpert Carba-R Sample Reagent.
10
{10}
11
# Stability of Rectal Swabs
In order to support a claim for the period of time that rectal swabs could be held before transfer to Xpert Carba-R Sample Reagent and further processing, a mixture of five carbapenemase-producing bacteria (at approximately 3x LoD) was spiked onto negative matrix swabs to create known positive samples. Both positive and negative matrix swab samples were tested with the Xpert Carba-R Assay after being stored for 1, 3, 5, and 7 days at 2°C, 8°C, 15°C and 28°C. Four (4) replicate negative matrix swab samples and 14 replicate positive matrix swab samples were tested per time point and temperature condition. At the time of testing, swabs were then added to Xpert Carba-R Sample Reagent. Eight replicate positive and negative matrix swab samples were tested at time t=0.
Under the conditions of this study, positive and negative specimens at all storage conditions and temperatures tested were correctly identified using the Xpert Carba-R Assay. The data supports that rectal swabs are stable from the time of collection prior to transfer to Xpert Carba-R Sample Reagent for up to 7 days when stored at 2°C –28°C before testing with the Xpert Carba-R Assay.
# Stability after Rectal Swabs Resuspended in Xpert Carba-R Sample Reagent
To establish a claim for the stability of rectal swabs in the Xpert Carba-R Sample Reagent, pooled negative rectal swab matrix was spiked at approximately 3x LoD with a mixture of five carbapenemase-producing bacteria that harbored targets of the Xpert Carba-R Assay. Aliquots of the positive swab matrix (with bacteria) in Sample Reagent were prepared to create positive samples. Negative samples in the study consisted of aliquots of negative pooled swab matrix in Xpert Carba-R Sample Reagent. Positive and negative matrix samples were tested with the Xpert Carba-R Assay after being stored for 1, 3, 5 and 7 days at 2°C, 8°C, 15°C, and 28°C. Four (4) replicate negative matrix samples and 14 replicate positive matrix samples were tested per time point and temperature condition. Positive and negative matrix swab replicates were also tested at time t=0.
Under the conditions of this study, positive and negative samples at all storage conditions and temperatures tested were correctly reported using the Xpert Carba-R Assay. The data supports that sample material collected from rectal swabs are stable in Xpert Carba-R Sample Reagent for up to 7 days when stored at 2°C–28°C prior to testing with the Xpert Carba-R Assay.
The study supports the following product claim specimen storage conditions prior to testing with the Xpert Carba-R Assay: 1) up to 5 days at 15°C–28°C for rectal swab specimens in transport tubes before transfer to Xpert Carba-R Sample Reagent and 2) up to 5 days at 2°C–28°C for rectal swab specimens suspended in Xpert Carba-R Sample Reagent.
## d. Detection limit:
A study was conducted to determine the Limit of Detection (LoD) of the Xpert Carba-R Assay for carbapenemase-producing organisms harboring the blaKPC, blaNDM, blaVIM, blaOXA-48, and blaIMP gene sequence targets seeded into rectal swab
{11}
background matrix. The LoD was defined as the lowest concentration (reported as cells/swab and cells/ml in Sample Reagent) of sample that can be reproducibly distinguished from negative samples with $95\%$ confidence. Pre-screened negative rectal swab background matrix was used to confirm that the Limit of Blank (LoB) was zero. Two strains harboring each assay target were tested. Point estimates and two-sided $95\%$ confidence intervals for the analytical LoD (reported as CFU/swab) were determined using probit regression analysis. The LoD estimate was verified by preparing two independent dilutions of each bacterial culture to the point estimate LoD value. Twenty measurements (10 from each dilution) were tested, and the study was completed using at least two unique lots of Xpert Carba-R Assay reagents.
Of the 2,774 runs in the LoD study to estimate the LoD of various targets in rectal matrix (excluding the external controls), 32 runs $(1.2\%)$ provided indeterminate GeneXpert results (2 NO RESULT, 9 ERROR, 21 INVALID). All 32 runs were repeated yielding valid Carba-R results as expected. A total of 195 of 198 external control runs provided valid GeneXpert results.
All the data was collected with the GeneXpert Dx software version 4.4a on the GeneXpert Dx GX-IV and GX-XVI instruments and with the Infinity Xpertise software version 6.1 on the Infinity-80 instrument. Table 4 shows the estimated LoD results by probit analysis for the test panel.
Table 4. LoD for Organisms in Rectal Matrix Harboring Carbapenemase Genes using the Xpert Carba-R Assay in $5\mathrm{ml}$ Sample Reagent
| Organism | Target Gene | Rectal Matrix | |
| --- | --- | --- | --- |
| | | LOD Claim CFU/swaba | Estimated LoD In Sample Reagent CFU/mlb |
| Acinetobacter baumannii | IMP-1 | 174 | 35 |
| Klebsiella pneumoniae | IMP-1 | 306 | 61 |
| Klebsiella pneumoniae | VIM-1 | 305 | 61 |
| Escherichia coli | VIM-4 | 815 | 163 |
| Klebsiella pneumoniae | NDM-1 | 251 | 50 |
| Klebsiella pneumoniae | NDM | 74 | 15 |
| Klebsiella pneumoniae | KPC-3 | 373 | 75 |
| Enterobacter cloacae | KPC | 779 | 156 |
| Enterobacter cloacae | OXA-48 | 154 | 31 |
| Escherichia coli | OXA-48 | 104 | 21 |
{12}
a. Colony counts per swab confirmed by plating of each organism.
b. Estimated LoD for swab in 5 ml of Sample Reagent using 0.2 conversion factor: $0.2 \times$ CFU/swab
# Analytical Reactivity
An Inclusivity Study was conducted to test reactivity of the Xpert Carba-R Assay with 72 well-characterized bacterial isolates. The panel consisted of the following molecular resistance marker groups: (11) $bla_{\mathrm{KPC}}$ isolates, (13) $bla_{\mathrm{NDM}}$ isolates, (11) $bla_{\mathrm{VIM}}$ isolates, (8) $bla_{\mathrm{OXA-48}}$ isolates, (5) $bla_{\mathrm{NDM/OXA-181}}$ isolates, (6) $bla_{\mathrm{OXA-181}}$ , (17) $bla_{\mathrm{IMP}}$ isolates, and (1) $bla_{\mathrm{KPC/VIM}}$ isolate. Strains were seeded at 3x LoD and tested in pooled negative rectal swab matrix. For a list of strains tested during the Analytical Reactivity Study, please refer to Table 5 below.
Table 5. Strains Tested in Analytical Reactivity
| Strain ID | Organism | Confirmed Genetic Resistance Marker | Concentration Tested (CFU/ml) | Rectal Matrix (Mean Ct) |
| --- | --- | --- | --- | --- |
| KPC Isolates | | | | |
| NCTC 13438 | Klebsiella pneumoniae | KPC-3 | 153 | 34.4 |
| 31551 | Klebsiella pneumoniae | KPC-4 | 50 | 34.5 |
| ATCC BAA-1705 | Klebsiella pneumoniae | KPC-2 | 130 | 35.9 |
| CFVL | Enterobacter cloacae | KPC-2 | 160 | 34.4 |
| KBM18 | Enterobacter aerogenes | KPC-2 | 250 | 33.7 |
| COL | Escherichia coli | KPC-2 | 147 | 33.1 |
| BM9 | Klebsiella pneumoniae | KPC-3 | 330 | 34.6 |
| CGNC | Serratia marcescens | KPC-2 | 300 | 33.9 |
| PA3 | Klebsiella pneumoniae | KPC-2 | 100 | 34.1 |
| PA-COL | Pseudomonas aeruginosa | KPC-2 | 250 | 34.4 |
| 164-3 | Klebsiella oxytoca | KPC | 70 | 33.7 |
| NDM Isolates | | | | |
| NCTC 13443 | Klebsiella pneumoniae | NDM-1 | 80 | 33.5 |
| ATCC BAA-2146 | Klebsiella pneumoniae | NDM-1 | 80 | 33.3 |
| 34262 | Klebsiella pneumoniae | NDM | 80 | 33.6 |
| GEN | Acinetobacter baumannii | NDM-1 | 130 | 33.4 |
| 3047 | Enterobacter cloacae | NDM-1 | 70 | 36.2 |
| 7892 | Proteus mirabilis | NDM-1 | 30 | 32.6 |
| CAN | Salmonella spp. | NDM-1 | 70 | 33.7 |
| EGY | Acinetobacter baumannii | NDM-2 | 40 | 35.0 |
| I5 | Escherichia coli | NDM-4 | 30 | 33.4 |
| 405 | Escherichia coli | NDM-5 | 30 | 33.7 |
| CF-ABE | Citrobacter freundii | NDM | 30 | 33.6 |
| 73999 | Pseudomonas aeruginosa | NDM | 50 | 33.1 |
| 39365 | Providencia rettgeri | NDM-1 | 70 | 33.3 |
| VIM Isolates | | | | |
| NCTC 13437 | Pseudomonas aeruginosa | VIM-10 | 500 | 31.9 |
{13}
| Strain ID | Organism | Confirmed Genetic Resistance Marker | Concentration Tested (CFU/ml) | Rectal Matrix (Mean Ct) |
| --- | --- | --- | --- | --- |
| NCTC 13439 | Klebsiella pneumoniae | VIM-1 | 130 | 32.6 |
| NCTC 13440 | Klebsiella pneumoniae | VIM-1 | 70 | 32.9 |
| 758 | Pseudomonas aeruginosa | VIM | 250 | 32.7 |
| PA-87 | Klebsiella pneumoniae | VIM | 200 | 33.3 |
| B92A | Pseudomonas aeruginosa | VIM | 2000 | 31.4 |
| Col1 | Pseudomonas aeruginosa | VIM-2 | 500 | 32.5 |
| BM19 | Serratia marcescens | VIM-2 | 250 | 33.6 |
| KOW7 | Escherichia coli | VIM-4 | 250 | 33.0 |
| DIH | Klebsiella pneumoniae | VIM-19 | 250 | 33.4 |
| MSH2014-3 | Enterobacter cloacae | VIM | 500 | 31.8 |
| OXA-48 and OXA-181 Isolates | | | | |
| NCTC 13442 | Klebsiella pneumoniae | OXA-48 | 40 | 33.2 |
| OM11 | Klebsiella pneumoniae | OXA-48 | 60 | 33.4 |
| 501 | Enterobacter cloacae | OXA-48 | 80 | 34.2 |
| DUW | Klebsiella pneumoniae | OXA-48 | 120 | 33.9 |
| OM22 | Escherichia coli | OXA-48 | 80 | 33.5 |
| BOU | Enterobacter cloacae | OXA-48 | 80 | 33.6 |
| TUR | Enterobacter cloacae | OXA-48 | 120 | 33.9 |
| 11670 | Escherichia coli | OXA-48 | 100 | 33.3 |
| MSH2014-64 | Klebsiella pneumoniae | OXA-181 | 280 | 32.0 |
| MSH2014-72 | Escherichia coli | OXA-181 | 100 | 34.7 |
| 166643 | Klebsiella pneumoniae | OXA-181 | 20 | 34.1 |
| 42194 | Klebsiella pneumoniae | OXA-181 | 20 | 33.5 |
| 74 | Escherichia coli | OXA-181 | 100 | 33.4 |
| CDC0051 | Klebsiella ozaenae | OXA-181 | 250 | 34.2 |
| IMP Isolates | | | | |
| NCTC 13476 | Escherichia coli | IMP-1 | 250 | 34.2 |
| 695 | Acinetobacter baumannii | IMP-1 | 1720 | 33.5 |
| 2340 | Enterobacter cloacae | IMP-1 | 250 | 34.6 |
| IMPBMI | Klebsiella pneumoniae | IMP-1 | 100 | 32.5 |
| 6852 | Klebsiella pneumoniae | IMP-1 | 100 | 33.5 |
| Yonsei_1 | Acinetobacter baumannii | IMP-1 | 1000 | 33.9 |
| Yonsei_2 | Acinetobacter baumannii | IMP-1 | 500 | 34.4 |
| 70450-1 | Pseudomonas aeruginosa | IMP-1 | 250 | 33.7 |
| 3994 | Pseudomonas spp. | IMP-10 | 250 | 34.5 |
| MKAM | Pseudomonas aeruginosa | IMP-1 | 500 | 33.4 |
| 5344 | Pseudomonas aeruginosa | IMP-2 | 60 | 33.9 |
| CDC0161 | Enterobacter aerogenes | IMP-4 | 50,000 | 36.6 |
| 3985 | Pseudomonas aeruginosa | IMP-11 | 2000 | 35.5 |
| 4032 | Pseudomonas aeruginosa | IMP-6 | 80 | 32.7 |
| 3424 | Pseudomonas aeruginosa | IMP-7 | 1x10^6 | 0 |
| 32443 | Klebsiella pneumoniae | IMP-13 | 1x10^6 | 0 |
{14}
| Strain ID | Organism | Confirmed Genetic Resistance Marker | Concentration Tested (CFU/ml) | Rectal Matrix (Mean Ct) |
| --- | --- | --- | --- | --- |
| 0092 | Pseudomonas aeruginosa | IMP-14 | \( 1 \times 10^6 \) | 0 |
| Isolates with more than one genetic marker target | | | | |
| GR-04/KP-69 | Klebsiella pneumoniae | KPC-2/VIM | 80 | 33.9 |
| B108A | Klebsiella pneumoniae | OXA-181/NDM | 10 | 32.6, 35.7 |
| KP-OMA3 | Klebsiella pneumoniae | OXA-181/NDM | 60 | 32.6, 34.3 |
| 1300920 | Klebsiella pneumoniae | OXA-181/NDM | 15 | 32.8, 36.5 |
| MSH2014-69 | Klebsiella pneumoniae | OXA-181/NDM | 20 | 33.8,34.2 |
| C10192-DISCS | Enterobacter aerogenes | OXA-181/NDM | 10 | 33.6, 35.8 |
Under the conditions of this study, 69 of 72 carbapenemase-producing bacterial strains were detected with the Xpert Carba-R Assay. Three carbapenemase-producing bacterial strains (IMP-7, IMP-13, and IMP-14) were not detected with the Xpert Carba-R Assay even at $10^{6}\mathrm{CFU / ml}$ . For a summary of these results, please refer to Table 6 below.
Table 6. Summary of Variants Detected by Wet Testing or Predicted to be Detected Based on In Silico Analysis.
| Marker (or Traditional Subgroup) | Wet testing | | | Not tested but predicted to be detected based on in silico analysis |
| --- | --- | --- | --- | --- |
| | No. of Samples with Target | Type(s) Detected | Type(s) not Detected | |
| KPC | 12 | KPC-2, 3, 4 | --- | KPC-5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16 |
| NDM | 18 | NDM-1, 2, 4, 5 | --- | NDM-3, 6, 7, 8, 9 |
| VIM | 12 | VIM-1, 2, 4, 10, 19 | --- | VIM-5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 16, 17, 18, 20, 23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33, 34, 35, 36, 37, 38 |
| OXA-48 | 19 | OXA-48, OXA-181 (OXA-48 variant) | --- | OXA-162, 163, 204, 232, 244, 245, 247 |
| IMP | 17 | IMP-1 (9 strains), IMP-2, 4a, 6, 10, 11 | IMP-7b, 13c, 14b | IMP-3, 8, 9, 13c, 19, 20, 21, 22, 24, 25, 27, 28, 30, 31, 33, 37, 40, 42 |
${}^{a}$ IMP-4 gene (Enterobacter aerogenes) was detected at much higher CFU/ml $\left( {5{\times 10}^{4}\mathrm{{CFU}}/\mathrm{{ml}}}\right)$ than other IMP variants detected by the assay.
bIMP-7 and IMP-14 genes (Pseudomonas aeruginosa) were not detected by the assay and were not predicted to be detected by in silico analysis (Limitations in package insert).
IMP-13 gene (Klebsiella pneumoniae): Although predicted to be detected by in silico analysis, the IMP-13 gene was not detected by the assay (Limitation in package insert).
{15}
# e. Analytical specificity:
The analytical specificity of the Xpert Carba-R Assay was evaluated with a panel of 62 well-characterized carbapenem-susceptible bacteria or bacteria with carbapenem non-susceptibility due to genes or mechanisms other than the Xpert Carba-R target genes. Panel members were re-suspended in negative rectal matrix. A set of 31 commensal/enteric microorganisms was also evaluated in the study, as well as human cells. Bacteria were seeded into negative matrix at $\geq 10^{6}$ CFU/ml in triplicate and tested with the Xpert Carba-R Assay. Viruses were tested at $>1\times 10^{5}$ TCID $_{50}$ /ml or greater than $2.5\times 10^{7}$ RNA copies/ml. Human cells were tested at $1\times 10^{5}$ cells/ml. For Analytical Specificity Study panel, please refer to Table 7 and Table 8 below.
Table 7. Analytical Specificity Panel with Organisms Having a Resistance Mechanism other than Targets of the Xpert Carba-R Assay
| Organism | Strain ID | Confirmed Resistance Mechanism(s)a | Carbapenem Susceptibility (S/I/R)b | | |
| --- | --- | --- | --- | --- | --- |
| | | | ETPb | IMPb | MEMb |
| Escherichia coli | NCTC 13441 | CTX-M (-1, -type 15 like); TEM | S | S | S |
| Klebsiella pneumoniae | NCTC 13465 | CTX-M (25) | S | S | S |
| Enterobacter aerogenes | 810 | OmpC/OmpF deficient; TEM | R | R | R |
| Citrobacter freundii | 1698 | TEM (WT+164S) | S | S | S |
| Enterobacter cloacae | 5557 | AmpC (ACT/MIR) | R | R | R |
| Klebsiella pneumoniae | kpn5 | CTX-M-2 | R | S | R |
| Klebsiella pneumoniae | kpn12 | TEM; SHV; CTX-M | R | R | R |
| Escherichia coli | eco1 | TEM; CTX-M-2 | R | R | R |
| Escherichia coli | eco2 | CTX-M (2); TEM | R | S | S |
| Enterobacter cloacae | cor1 | CTX-M (2); TEM | R | R | R |
| Serratia marcescens | hpp21 | CTX-M (2); TEM | S | S | S |
| Morganella morganii | fer29 | CTX-M (2); TEM | S | R | S |
| Proteus mirabilis | gut25 | CTX-M (2); TEM | S | R | S |
| Salmonella spp. | 3209 | CTX-M (2); TEM | S | S | S |
| Shigella flexneri | 3331 | CTX-M (2); TEM | S | S | S |
| Enterobacter cloacae | PA_3 | AmpC; CTX-M-15; TEM | S | S | S |
| Klebsiella pneumoniae | 32189 | SHV | S | S | S |
| Klebsiella pneumoniae | 32443 | CTX-M (1, -type 15 like); SHV | S | S | S |
| Klebsiella pneumoniae | 32598 | CTX-M (-1, -type 15 like); SHV; TEM | R | I | R |
{16}
| Organism | Strain ID | Confirmed Resistance Mechanism(s)a | Carbapenem Susceptibility (S/I/R)b | | |
| --- | --- | --- | --- | --- | --- |
| | | | ETPb | IMPb | MEMb |
| Klebsiella pneumoniae | 33560 | CTX-M (15); SHV-11; TEM-1 | S | S | S |
| Klebsiella pneumoniae | 33603 | SHV-2 | R | I | R |
| Klebsiella pneumoniae | 33617 | SHV-27 | S | S | S |
| Klebsiella pneumoniae | 33643 | SHV (-5, -55); TEM | S | S | S |
| Klebsiella pneumoniae | 34430 | SHV; TEM; CTX-M-15 | S | S | S |
| Klebsiella pneumoniae | 34680 | TEM; CTX-M-2 | R | S | R |
| Klebsiella pneumoniae | 34732 | CTX-M (15); SHV; TEM | R | S | S |
| Enterobacter cloacae | PA_174 | GX-/Culture+; SHV; TEM | S | S | S |
| Enterobacter aerogenes | STU 645 | SHV (WT+238S+240K) | R | S | R |
| Enterobacter aerogenes | STU 669 | SHV (WT+238S+240K) | R | R | R |
| Escherichia coli | C3015 | AmpC (CMY II); TEM | R | R | R |
| Enterobacter aerogenes | RI_100 | AmpC (DHA); SHV | R | R | R |
| Klebsiella pneumoniae | B4A | SHV (WT + 238S +240K) | R | R | R |
| Klebsiella pneumoniae | B13A | SHV (WT + 238S +240K) | R | S | S |
| Enterobacter cloacae | RI_474 | AmpC (ACT/MIR) | R | I | I |
| Enterobacter amnigenus | B71 | AmpC (ACT/MIR) | R | R | R |
| Klebsiella pneumoniae | DD82A | SHV (WT + 238S + 240K) | R | S | R |
| Klebsiella pneumoniae | B100 | CTX-M (-1, type-15 like); SHV (WT+238S); TEM | R | S | R |
| Enterobacter cloacae | 135B | TEM | S | S | S |
| Klebsiella pneumoniae | B157 | SHV; TEM | R | R | R |
| Escherichia coli | T2914280 | CTX-M (-1, -15); TEM | R | S | R |
| Providencia stuartii | DD188 | TEM (104K + 164S) | R | I | I |
| Enterobacter cloacae | DD189 | AmpC (ACT/MIR) | R | S | S |
| Escherichia coli | B198B | CTX-M (-1, type -15 like); TEM | R | S | R |
| Klebsiella pneumoniae | T3019989-1 | CTXM (-1, type-15 like); SHV | R | I | R |
| Klebsiella pneumoniae | T3019989-2 | CTX-M (-1, type-15 like); SHV | R | S | R |
| Enterobacter cloacae | ENC-THAI14 | VEB-1, TEM | S | S | S |
| Escherichia coli | CB154006 | CTX-M (9); TEM | R | I | I |
| Klebsiella pneumoniae | CB154007 | CTX-M (9); TEM | R | S | S |
| Klebsiella pneumoniae | CB154008 | CTX-M (9); TEM | R | S | S |
| Klebsiella pneumoniae | CB154009 | CTX-M (9); TEM | R | S | S |
| Klebsiella pneumoniae | CB154010 | CTX-M (9); TEM | R | S | S |
| Klebsiella pneumoniae | CB154011 | CTX-M (9); TEM | R | S | S |
| Klebsiella pneumoniae | CB154012 | CTX-M (9); TEM | R | S | S |
| Klebsiella pneumoniae | CB154013 | CTX-M (9); TEM | R | S | S |
| Klebsiella pneumoniae | CB154014 | CTX-M (9); TEM | R | S | S |
| Klebsiella pneumoniae | CB154015 | CTX-M (9); TEM | R | S | S |
| Klebsiella pneumoniae | CB154016 | CTX-M (9); TEM | R | S | S |
{17}
Table 8. Panel of Commensal and Other Enteric Microorganisms Tested in this Study and Human Cells
| Organism | Strain ID | Confirmed Resistance Mechanism(s)a | Carbapenem Susceptibility (S/I/R)b | | |
| --- | --- | --- | --- | --- | --- |
| | | | ETPb | IMPb | MEMb |
| Enterobacter cloacae | S35766 | AmpC(ACT/MIR) | S | S | S |
| Enterobacter cloacae | X1856910 | AmpC (ACT/MIR); TEM | R | I | I |
| Klebsiella pneumoniae | W3758164 | CTX-M (-1, -15 like); SHV; TEM. | R | I | R |
| Klebsiella pneumoniae | X2135758 | CTX-M (-1, -15 like); SHV | R | S | S |
| Klebsiella pneumoniae | W3809535 | CTX-M (-1, -15 like); SHV | R | R | R |
| Pseudomonas aeruginosa | CDC0064 | SPM | R | R | R |
| Serratia marcescens | CDC0099 | SME | R | R | R |
| Serratia marcescens | CDC0121 | SME | R | R | R |
| Serratia marcescens | CDC0122 | SME | R | R | R |
| Serratia marcescens | CDC0123 | SME | R | R | R |
| Serratia marcescens | CDC0124 | SME | R | R | R |
| Serratia marcescens | CDC0130 | SME | R | R | R |
| Serratia marcescens | CDC0131 | SME | R | R | R |
| Enterobacter cloacae group | CDC0132 | IMI | R | R | R |
| Enterobacter cloacae complex | CDC0164 | IMI | R | R | R |
${}^{a}$ Presence of these markers of resistance were determined by individual PCR assays, DNA sequence analysis, or by other research-based methods.
${}^{b}S/I/R =$ Susceptible/Intermediate/Resistant; ETP $=$ Ertapenem,IMP $=$ Imipenem, MEM $=$ Meropenem
| Organism | Strain ID |
| --- | --- |
| Escherichia coli | ATCC 25922 |
| Enterococcus faecalis | ATCC 29212 |
| Klebsiella pneumoniae | ATCC 700603 |
| Escherichia coli | ATCC 35218 |
| Staphylococcus aureus | ATCC 25923 |
| Pseudomonas aeruginosa | ATCC 27853 |
| Clostridium difficile | ATCC 9689 |
| Enterobacter cloacae | ATCC 700621 |
| Enterococcus faecium | ATCC 9756 |
| Klebsiella oxytoca | ATCC 13182 |
| Acinetobacter baumannii | ATCC BAA-747 |
{18}
| Organism | Strain ID |
| --- | --- |
| Citrobacter freundii | ATCC 33128 |
| Morganella morganii | ATCC 49948 |
| Stenotrophomonas maltophilia | ATCC 51331 |
| Citrobacter koseri | ATCC 27028 |
| Providencia stuartii | ATCC 49809 |
| Peptostreptococcus anaerobiusa | ATCC 49037 |
| Streptococcus agalactiae | CCUG 29780 / ATCC 12401 |
| Bifidobacterium adolescentis | ATCC 15703 |
| Enterobacter aerogenes | ATCC 51697 |
| Proteus mirabilis | ATCC 43071 |
| Acinetobacter spp. | CCUG 34787 |
| Citrobacter freundii | CCUG 418 |
| Corynebacterium diphtheriae | CCUG 33629 |
| Helicobacter pylori | CCUG 17874 |
| Listeria monocytogenes | CCUG 33548 |
| Providencia alcalifaciens | CCUG 6325 |
| Campylobacter jejuni | CCUG 43594/ATCC 33560 |
| Viruses | |
| Adenovirus B Type 7A/NY | MRVP/ZeptoMetrix |
| Enterovirus Type 71/NY | MRVP/ZeptoMetrix |
| Norovirus GII | Clinical Sample – Cepheid |
| Human Cells | |
| Bladder Cell Carcinoma (hgDNA) | ATCC HTB-4 |
${}^{a}$ Peptostreptococcus anaerobius was tested at $5 \times {10}^{5}$ CFU/ml.
Of the 94 potentially cross-reactive organisms and nucleic acids tested, including organisms exhibiting antibiotic resistance mechanisms other than production of KPC, NDM, VIM, IMP and OXA-48, none were detected with the Xpert Carba-R Assay. Of the 286 tests in rectal matrix, four (1.4%) runs were indeterminate (4 INVALID). All four indeterminate runs were successfully repeated and were reported as NOT DETECTED for all five targets (KPC, NDM, VIM, IMP and OXA-48) as expected. All external positive and negative controls were correctly reported as expected.
All the data were collected on the GeneXpert Dx (GX-IV) instrument using the GeneXpert Dx software version 4.4a.
# f. Assay cut-off:
For IMP, VIM, NDM, KPC, and OXA-48 gene targets, the valid cycle threshold (Ct) range was 3.0 to 38.0. For the SPC, the valid Ct range was set from 3.0 to 40.0. A Ct
{19}
value outside the valid range is reported as NOT DETECTED. The Ct cut-offs are included as automatic calculations in the assay definition file (ADF) provided with the Xpert Carba-R Assay. Assay cut-off values have not changed from those described in K152614.
# g. Interfering substances:
A study was conducted to assess the inhibitory effects of substances potentially encountered in rectal swab specimens on the performance of the Xpert Carba-R Assay. Twenty-four substances were evaluated at "worst case scenario" concentrations. Eight replicate positive samples were tested per substance. Negative samples consisted of pooled negative rectal swab matrix with/without the interfering substance (not seeded with carbapenemase-producing organisms). Controls consisted of positive and negative samples with no interfering substances added. Positive samples were prepared from a mix of five carbapenemase-producing organisms harboring KPC, NDM, VIM, IMP and OXA-48 gene sequences seeded into pooled negative rectal swab matrix to give concentrations that were 3x analytical LoD (Table 9). The list of potentially interfering substances tested is shown in Table 10.
Table 9. Organisms used to Prepare the Mixed Positive Sample
| Organism | Target Gene |
| --- | --- |
| Klebsiella pneumoniae | KPC |
| Klebsiella pneumoniae | NDM |
| Escherichia coli | VIM |
| Enterobacter cloacae | OXA-48 |
| Acinetobacter baumannii | IMP-1 |
Table 10. Potentially Interfering Substances Tested
| Substance ID | Substance/Class | Active Ingredient | Concentration Tested |
| --- | --- | --- | --- |
| Aleve (1) | Non-steroidal anti-inflammatory medication | Naproxen | 0.25% w/v |
| Barium sulfate (2) | Imaging compound | N/A | 0.25% and 0.1% w/v |
| Antimicrobial (3) | Antibiotic (oral) | Cephalexin | 0.25% w/v |
| Antimicrobial (4) | Antibiotic (oral) | Ciprofloxacin | 0.25% w/v |
| Condom (5) | Condom with spermicidal lubricant | Nonoxynol-9 | 1 condoma |
{20}
| Cortizone (6) | Creams/ointment/suppositories | Hydrocortisone | 0.25% w/v |
| --- | --- | --- | --- |
| ExLax (7) | Laxative | Sennosides | 0.25% w/v |
| Fecal Fat (8) | Lipids | Stearic acid/Palmitic acid/Cholesterol | 0.25% w/v |
| Imodium (9) | Anti-diarrheal medication | Loperamide hydrochloride/bismuth subsalicylate | 0.25% w/v |
| Kaopectate (10) | Anti-diarrheal medication | Loperamide hydrochloride/bismuth subsalicylate | 0.25% w/v |
| K-Y Jelly (11) | Topical cream | N/A | 0.25% w/v |
| Milk of Magnesia (12) | Antacids | Calcium carbonate/aluminum hydroxide/magnesium hydroxide/simethicone | 0.25% w/v |
| Mineral Oil-enema (13) | Enemas | N/A | 0.25% w/v |
| Neosporin (14) | Antibiotic (topical) | Polymixin B/Neomycin/Bacitracin | 0.25% w/v |
| Nystatin (15) | Anti-fungal/anti-itch Vaginal | Nystatin | 0.25% w/v |
| Pepcid (16) | Antacid | Famotidine | 0.25% w/v |
| Pepto-Bismol (17) | Anti-diarrheal medication | Loperamide hydrochloride/bismuth subsalicylate | 0.25%, 0.1%, 0.05%, 0.025%, 0.01% w/v |
| Petroleum jelly (18) | Topical cream | N/A | 0.25% w/v |
| Preparation H (19) | Anti-hemorrhoid creams/ointments | Phenylephrine | 0.25% w/v |
| Prilosec (20) | Acid reducer; antacid | Oemprazole | 0.25% w/v |
| Saline-enema (21) | Enemas | N/A | 0.25% w/v |
| Tagamet (22) | Antacid | Cimetidine | 0.25% w/v |
| Vagisil (23) | Anti-fungal/anti-itch Vaginal | Benzocaine, resorcinol | 0.25% w/v |
| Wet Ones (24) | Moist Towelettes | Benzalkonium chloride, ethanol | 1 pieceb |
${}^{a}$ One condom added to ${40}\mathrm{\;{ml}}$ swab matrix.
${}^{b}$ One piece (5 inch x 7-1/2 inch) added to ${40}\mathrm{\;{ml}}$ swab matrix.
{21}
Results showed that assay targets were detected in the presence of 22 of the 24 potentially interfering substances when tested at $0.25\%$ w/v with the Xpert Carba-R Assay. It was noted that for fecal fat diluted into positive rectal matrix at $0.25\%$ w/v, the VIM target showed a 2.475 Ct delay compared to the mean Ct of the control; however, the VIM target was still detected by the assay. In addition, a change in the Ct value was observed with the rectal swab matrix sample for the OXA target, which showed a 1.237 Ct delay compared to the mean Ct of the control in the presence of barium sulfate at $0.1\%$ w/v. Interference with the Xpert Carba-R Assay may be observed with barium sulfate at $>0.1\%$ w/v, Pepto-Bismol at $>0.01\%$ w/v, and fecal fat at $0.25\%$ w/v (for VIM) in tests with rectal swab matrix samples. Of the 404 tests (not including those test concentrations where barium sulfate and Pepto-Bismol by themselves showed high invalid rates), 2 tests provided indeterminate GeneXpert results (2 ERROR). Both indeterminate GeneXpert results were successfully repeated.
External controls gave the expected GeneXpert test result. All the data were collected using the GeneXpert Dx software version 4.4a on the GeneXpert Dx (GX-IV and GX-XVI) instruments.
# Competitive Interference
To evaluate the potential competitive inhibitory effect of multiple carbapenemase-producing organisms on the performance of the Xpert Carba-R Assay in rectal swab specimens, a study was performed by testing various combinations of carbapenemase-producing organisms seeded at high and low concentrations into natural matrix. High concentrations of organisms corresponded to $\sim 1\mathrm{x}10^{6}$ CFU/ml, and low concentrations corresponded to $2\mathrm{x}$ LoD. An inhibitory effect was observed for three out of five targets (IMP, VIM, and OXA-48) when a low concentration of each target was present in combination with a high concentration of another assay target. Table 11 reports the number of target replicates that were detected in the competitive interference study with rectal swab matrix.
Table 11. Number of Correct Results for Combinations in the Competitive Interference Study with the Xpert Carba-R Assay using a Rectal Swab Matrix
| Sample number | Combination | Number of DETECTED results/Number of replicates | | | | | Inhibiteda(Yes or No) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | IMP | VIM | NDM | KPC | OXA | |
| 1 | High KPC/High NDM/Low VIM | | 7/8 | 8/8 | 8/8 | | No |
| 2 | High KPC/High NDM/Low OXA | | | 8/8 | 8/8 | 8/8 | No |
| 3 | High KPC/High NDM/Low IMP | 5/8 | | 8/8 | 8/8 | | Yes/IMP |
| 4 | High VIM/High OXA/Low KPC | | 8/8 | | 8/8 | 8/8 | No |
| 5 | High VIM/High OXA/Low NDM | | 8/8 | 7/8 | | 8/8 | No |
| 6 | High VIM/High OXA/Low IMP | 6/8 | 8/8 | | | 8/8 | Yes/IMP |
| 7 | High IMP/Low KPC | 8/8 | | | 8/8 | | No |
| 8 | High IMP/Low NDM | 8/8 | | 8/8 | | | No |
| 9 | High IMP/Low VIM | 8/8 | | | | 8/8 | No |
| 10 | High IMP/Low OXA | 8/8 | | | | 6/8 | Yes/OXA |
{22}
For those targets not detected at $2\mathrm{x}$ LoD in $\geq 7/8$ replicates (IMP, VIM, OXA-48), an additional study was performed where the low target concentration was increased to $4\mathrm{x}$ LoD to evaluate the competitive inhibitory effect. No inhibitory effect was observed for the three targets (IMP, VIM and OXA-48) at $4\mathrm{x}$ LoD in the presence of high concentrations of other targets for the Xpert Carba-R Assay (Table 12).
Table 12. Number of Correct Results for Combinations in the Competitive Interference Study with the Xpert Carba-R Assay using a Low Target Concentration of 4x LoD.
| Sample Number | Combination | Number of DETECTED results/Number of replicates | | | | | Inhibited (Yes or No) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | IMP | VIM | NDM | KPC | OXA | |
| 3 | High KPC/High NDM/Low IMP | 8/8 | | 8/8 | 8/8 | | No |
| 6 | High VIM/High OXA/Low IMP | 7/8 | 8/8 | | | 8/8 | No |
| 10 | High IMP/Low OXA | 8/8 | | | | 8/8 | No |
| 11 | High OXA/Low VIM | | 8/8 | | | 8/8 | No |
| 14 | Negative | 0/8 | 0/8 | 0/8 | 0/8 | 0/8 | N/A |
# h. Carry-over:
The purpose of the carry-over study was to determine the carry-over rate of contamination in negative samples due to the nucleic acid extraction and amplification of high positive samples in the GeneXpert cartridge. In this study, a negative sample was tested in a GeneXpert module immediately following the testing of a high titer positive sample in the same GeneXpert module. The high positive sample was composed of inactivated $E$ coli cells containing a plasmid with all five Xpert Carba-R target analyte genes (KPC, NDM, VIM, IMP and OXA-48 gene targets) diluted in Sample Reagent with rectal swab matrix to a concentration of $1 \times 10^{6}$ CFU/ml. The testing pattern was repeated 25 times on two GeneXpert modules for a total of 102 tests (25 high positive samples and 26 negative samples per module) for the positive cells in rectal swab matrix. For each day of the study, an external negative control and two types of external positive controls were tested as described previously. All 52 negative samples reported NOT DETECTED results for all five Xpert Carba-R Assay targets as expected. All positive samples correctly reported all Xpert Carba-R targets as DETECTED. There was one invalid result reported, which gave a valid result upon re-testing. Study results indicated no evidence of sample or amplicon carry-over contamination in the GeneXpert modules.
# 2. Comparison studies:
{23}
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
In a multi-center study, performance characteristics of the Xpert Carba-R Assay were evaluated with rectal swab specimens. The positive percent agreement (PPA) and negative percent agreement (NPA) of the Xpert Carba-R Assay were evaluated relative to a Composite Reference Method (consisting of Reference Culture + PCR and Sequencing of the amplification product). Five geographically diverse sites were selected (three across the United States and two in Europe), and prospectively paired rectal swab specimens were collected from subjects who were hospitalized or in a long-term care facility. In the study, one swab of a pair was used for Xpert Carba-R Assay testing. Highly soiled rectal swabs were excluded from the study. Contrived specimens were included in the study due to the expected low prevalence for some Xpert Carba-R Assay target genes.
For the Reference Culture, the second swab specimen was inoculated into 11 ml MacConkey enrichment broth containing a 10 µg meropenem disk and incubated overnight at 35°C. An aliquot of MacConkey broth culture was spread onto a MacConkey agar plate, and a 10 µg meropenem disk was placed on the plate. After an overnight incubation at 35°C, the zone of clearing was measured. If growth was observed within a 28 mm zone (including the meropenem disk), species identification of the organisms was performed after subculture of colonies to sheep blood agar. Organism susceptibility status (susceptible, intermediate or resistant) to meropenem, ertapenem and imipenem was determined using CLSI standard test methods (M07-A9) and the interpretive criteria found in the FDA drug label and CLSI M100-S24. Carbapenem non-susceptible organisms were then subcultured to sheep blood agar (with a meropenem disk placed between the 1st and 2nd streak) and incubated overnight at 35°C. Three to five well-isolated colonies of the same morphotype, as described in CLSI M07-A9, were collected and sent for sequencing.
DNA from the carbapenem non-susceptible isolates was purified, quantified, and amplified using primers specific to all 5 target genes that were validated and amplify a larger region than the Xpert Carba-R primers. The production of the appropriate sized amplification products was confirmed on Agilent 2100 Bioanalyzer. If no bands were shown on the Bioanalyzer for any of the five target genes, the isolate was not sent for sequence analysis and the Reference Method result was considered negative for the five target genes. PCR and sequencing was not performed for specimens
24
{24}
where there was no growth in the $28\mathrm{mm}$ zone, if the results of antimicrobial susceptibility testing identified a susceptible isolate, or organisms had intrinsic resistance to all the carbapenems tested (e.g. Stenotrophomonas maltophilia).
A total of 802 prospective rectal swab specimens were initially enrolled in the clinical study, of which 785 were eligible for inclusion. The ineligible specimens included:
(14) specimens from subjects with incomplete specimen information
(1) specimen was too soiled
(1) specimen was previously enrolled
(1) specimen from subject not hospitalized or in long-term care facility
An additional 30 specimens were excluded due to:
(10) shipping delay/testing delay
(4) invalid GX control or GX indeterminate/invalid
(16) organisms where there were no zone interpretations listed corresponding to intermediate or resistant to determine non-susceptibility
Thus, 755 rectal swab specimens remained compliant and were included in the final analysis for the prospective study. Performance of the Xpert Carba-R Assay was assessed separately for each type of resistance marker target and compared to Reference Culture + Sequencing result. With the rectal swab specimens, the study showed that $98.9\%$ (747/755) of the specimens produced a valid result on the first run. Four (4) INVALID and Four (4) ERRORS occurred, which upon re-testing yielded a valid result.
In addition to the prospective study, well-characterized isolates carrying each assay target were tested in a contrived study. Strains were re-suspended in rectal matrix before testing with the Xpert Carba-R Assay at 1x, 3x, and 10x LoD concentrations. The following numbers of unique isolates were evaluated in the study spanning multiple gram negative species (Table 13 and Table 14):
Table 13. Number of Unique Bacterial Strains Tested by Target and Concentration Level
| Target | 1x LoD | 3x LoDa | 10x LoDa |
| --- | --- | --- | --- |
| IMP | 30 | 25 | 25 |
| KPC | 30 | 25 | 25 |
| OXA-48 | 29 | 25 | 25 |
| VIM | 30 | 26 | 26 |
| NDM | 30 | 25 | 25 |
| Negativesb | 15 | | |
${}^{a}$ Strains tested at ${3x}$ and ${10x}\mathrm{{LoD}}$ were chosen from the unique strains tested at ${1x}\mathrm{{LoD}}$ .
bThirty negative samples were included in the study(15 x2).
{25}
Table. 14. Various Species Tested in the Contrived Study by Target
| Target | Various Organisms included in the Study |
| --- | --- |
| VIM | Acinetobacter baumannii |
| | Enterobacter cloacae |
| | Enterobacter asburiae |
| | Escherichia coli |
| | Klebsiella pneumoniae |
| | Pseudomonas aeruginosa |
| | Pseudomonas putida |
| | Serratia marcescens |
| IMP | Acinetobacter baumannii |
| | Enterobacter cloacae |
| | Klebsiella pneumoniae |
| | Pseudomonas aeruginosa |
| | Pseudomonas stutzeri |
| NDM | Acinetobacter baumannii |
| | Citrobacter spp. |
| | Empedobacter brevis |
| | Enterobacter cloacae |
| | Escherichia coli |
| | Klebsiella pneumoniae |
| | Morganella morganii |
| | Proteus mirabilis |
| | Pseudomonas oryzihabitans |
| | Salmonella spp. |
| KPC | Citrobacter koseri |
| | Enterobacter cloacae |
| | Enterobacter aerogenes |
| | Escherichia coli |
| | Klebsiella pneumoniae |
| | Serratia marcescens |
| OXA-48 | Enterobacter cloacae |
| | Enterobacter aerogenes |
| | Escherichia coli |
| | Klebsiella pneumoniae |
| No target | Acinetobacter baumannii |
| | Enterobacter cloacae |
| | Enterobacter aerogenes |
| | Klebsiella pneumoniae |
| | Morganella morganii |
{26}
| Target | Various Organisms included in the Study |
| --- | --- |
| | Pseudomonas aeruginosa |
| | Serratia marcescens |
| | Shigella flexneri |
Discordant samples were tested with an alternate PCR method when the Xpert Carba-R Assay result was positive and the Reference Method result was negative. An aliquot of the MacConkey broth was used to extract DNA and amplify any potential targets by PCR before sending for DNA sequencing. These results were not used to change the original performance data (See results in the footnotes to Table 15).
Study results of the Xpert Carba-R Assay compared to the Reference Method are shown in Table 15 stratified by individual target for prospective and contrived studies.
27
{27}
Table 15. Clinical Performance Data for the Xpert Carba-R Assay vs. Reference Culture + Sequencing
| Study | Target | TP | FP | TN | FN | PPA% (95 CI) | NPA% (95 CI) |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Prospective (n=755) | IMP | 0 | 1a | 754 | 0 | N/A | 99.9% (99.3-100.0) |
| | VIM | 6 | 8b | 737 | 4 | 60.0% (31.3-83.2) | 98.9% (97.9-99.5) |
| | NDM | 7 | 3c | 745 | 0 | 100.0% (64.6-100.0) | 99.6% (98.8-99.9) |
| | KPC | 29 | 6d,e | 720 | 0 | 100.0% (88.3-100.0) | 99.2% (98.2-99.6) |
| | OXA-48 | 29 | 10f | 715 | 1 | 96.7% (83.3-99.4) | 98.6% (97.5-99.2) |
| Contrived (n=432) | IMP | 76 | 0 | 352 | 4g | 95.0% (87.8-98.0) | 100.0% (98.9-100.0) |
| | VIM | 81 | 0 | 350 | 1h | 98.8% (93.4-99.8) | 100.0% (98.9-100.0) |
| | NDM | 80 | 0 | 352 | 0 | 100.0% (95.4-100.0) | 100.0% (98.9-100.0) |
| | KPC | 80 | 0 | 352 | 0 | 100.0% (95.4-100.0) | 100.0% (98.9-100.0) |
| | OXA-48 | 79 | 0 | 352 | 1i | 98.8% (93.3-99.8) | 100.0% (98.9-100.0) |
${}^{a}0$ of the 1FPs was determined to be TP after discordant analysis.
${}^{b}2$ of the 8 FPs were determined to be TPs after discordant analysis.
${}^{c}1$ of the 3 FPs was determined to be TP after discordant analysis.
${}^{d}$ 1of the 6 FPs was determined to be TP after discordant analysis.
${}^{e}$ Site reported that subject was on ertapenem during time of specimen collection.
${}^{f}3$ of the 10 FPs were determined to be TPs after discordant analysis.
$^{\mathrm{g}}$ 4FNs were observed with the following organisms: Acinetobacter baumannii [(2) at 1X LoD and (1) at 3X LoD] and Pseudomonas aeruginosa [(1) at 1X LoD]
$^{h}1FN$ was observed with Enterobacter asburiae.
$^{i}1FN$ was observed with Klebsiella pneumoniae.
The Xpert Carba-R Assay performance by specific organism group is shown in Table 16 for the prospective study.
{28}
Table 16. Xpert Carba-R Results (by Organism and Target) vs Reference Culture + Sequencing (Prospective Study)
| PROSPECTIVE STUDY | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| \(Organism^{a,b}\) | Target | TP | FP | TN | FN | PPA(95% CI) | NPA(95% CI) |
| A. baumannii(n=13) | IMP | 0 | 0 | 13 | 0 | NA | 100%(77.2-100) |
| | VIM | 0 | 0 | 13 | 0 | NA | 100%(77.2-100) |
| | NDM | 0 | 0 | 13 | 0 | NA | 100%(77.2-100) |
| | KPC | 0 | 0 | 13 | 0 | NA | 100%(77.2-100) |
| | OXA-48 | 0 | 0 | 13 | 0 | NA | 100%(77.2-100) |
| E. amnigenus 2(n=1) | IMP | 0 | 0 | 1 | 0 | NA | 100%(20.7-100) |
| | VIM | 0 | 0 | 1 | 0 | NA | 100%(20.7-100) |
| | NDM | 0 | 0 | 1 | 0 | NA | 100%(20.7-100) |
| | KPC | 0 | 0 | 1 | 0 | NA | 100%(20.7-100) |
| | OXA-48 | 0 | 0 | 1 | 0 | NA | 100%(20.7-100) |
| E. cloacae(n=4) | IMP | 0 | 0 | 4 | 0 | NA | 100%(51.0-100) |
| | VIM | 1 | 0 | 3 | 0 | 100%(20.7-100) | 100%(43.9-100) |
| | NDM | 0 | 0 | 4 | 0 | NA | 100%(51.0-100) |
| | KPC | 0 | 0 | 4 | 0 | NA | 100%(51.0-100) |
| | OXA-48 | 1 | 0 | 3 | 0 | 100%(20.7-100) | 100%(43.9-100) |
| E. coli(n=10) | IMP | 0 | 0 | 10 | 0 | NA | 100%(72.3-100) |
| | VIM | 0 | 0 | 10 | 0 | NA | 100%(72.3-100) |
| | NDM | 3 | 0 | 7 | 0 | 100%(43.9-100) | 100%(67.6-100) |
| | KPC | 2 | 0 | 8 | 0 | 100%(34.2-100) | 100%(64.6-100) |
| | OXA-48 | 3 | 0 | 7 | 0 | 100%(43.9-100) | 100%(64.6-100) |
| K. oxytoca(n=1) | IMP | 0 | 0 | 1 | 0 | NA | 100%(20.7-100) |
| | VIM | 0 | 0 | 1 | 0 | NA | 100% |
{29}
For Table 17, the results of the contrived study are stratified by the spiking concentration of the carbapenemase-producing organism tested with the Xpert Carba-R Assay.
Table 17. Performance of the Xpert Carba-R Assay in the Contrived Study (by Concentration)
| Contrived Studya,b | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Concentration Tested | Target | N | TP | FP | TN | FN | PPA (95% CI) |
| 1x LoD (n=150) | IMP | 30 | 27 | 0 | 120 | 3 | 90.0% (74.4-96.5) |
| | VIM | 30 | 29 | 0 | 120 | 1 | 96.7% (83.3-99.4) |
| | NDM | 30 | 30 | 0 | 120 | 0 | 100% (88.7-100) |
| | KPC | 30 | 30 | 0 | 120 | 0 | 100% (88.7-100) |
| | OXA-48 | 30c | 29 | 0 | 120 | 1 | 96.7% |
${}^{a}N$ refers to the total numbers of specimens where an organism was identified. This number contains samples that PCR positive and PCR negative for the presence of assay targets.
${}^{b}$ 16 Stenotrophomonas maltophilia isolates recovered by the Reference Method. Specimens where these isolates were identified were not included in the analysis because of the intrinsic resistance to multiple carbapenems.
{30}
Eight isolates were identified in the prospective study where at least two Xpert Carba-R Assay targets were detected. These results are shown in Table 18 below.
Table 18. Prospective Specimens with Multiple Targets Detected
| Targets Detected by Xpert Carba-R Assay | Targets Detected by PCR and Reference Sequencing | Targets Detected by Alternate PCR and Reference Sequencing |
| --- | --- | --- |
| KPC, OXA-48 | NEGa | NEGa |
| VIM, KPC | NEGb | NEGa |
| VIM, OXA-48 | OXA-48 | OXA-48 |
| KPC, OXA-48 | KPC | KPC, OXA-48 |
| NDM, OXA-48 | NDM | NDM, OXA-48 |
| VIM, NDM | NEGb | NEGa |
| NDM, KPC | KPC | NDM, KPC |
| VIM, KPC | VIM | VIM, KPC |
${}^{a}$ PCR yielded no bands for sequencing.
${}^{b}$ An organism was not isolated from reference culture,therefore reference sequencing was not performed.
External controls for the Xpert Carba-R Assay consisted of one negative sample, one sample positive for all (5) targets of the assay, and five different positive controls each containing a single target of the assay. The negative control and five-target positive controls were run on each day that study samples were tested, along with two of the single-target positive controls (on a rotating basis). Study samples were not run until correct results were obtained for each of the four controls. External control data
{31}
were compiled across all sites and overall QC results were acceptable.
# b. Clinical specificity:
See comments in 3a above.
# c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
# 4. Clinical cut-off:
Not applicable
# 5. Expected values/Reference range:
Of the 755 rectal specimens in the study included for analysis, $10.1\%$ of the specimens (76/755) contained a carbapenem non-susceptible organism with at least one of the assay gene targets (IMP, VIM, NDM, KPC, OXA-48) that was recovered by the Reference Method (Table 19). From 755 rectal specimens, a total of 112 carbapenem non-susceptible organisms (I or R to at least one carbapenem) were recovered by the Reference Culture.
Table 19. Detection of Non-susceptible Organisms with the Gene Targets By the Reference Culture + Sequencing Method
| Site Information | | Rectal Specimens | |
| --- | --- | --- | --- |
| Collection Site | Location | Total | # Positive by Reference Culture + Sequencing (NS organism with at least one assay target) |
| Site 1 | US (Midwest) | 16 | 0 |
| Site 2 | US (Midwest) | 130 | 25 |
| Site 3 | US (Southeastern) | 14 | 0 |
| Site 4 | Europe (Spain) | 456 | 39 |
| Site 5 | Europe (Italy) | 139 | 12 |
| Totals | | 755 | 76 |
# N. Instrument Name:
GeneXpert Instrument Systems
# O. System Descriptions:
1. Modes of Operation:
{32}
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ X ☐ or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ or No ☐ X ☐
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐
3. Specimen Identification:
Similar to previously cleared system.
4. Specimen Sampling and Handling:
Specific instructions should be followed for the collection of rectal swab specimens. The user refers to a Reference Diagram supplied by Cepheid to determine the acceptability of swab specimens. Any specimens that are highly soiled per the Reference Diagram must be excluded from the study. Swabs are then placed in the appropriate collection device. One swab is added to Sample Reagent for Xpert Carba-R Assay testing. An aliquot of sample (1.7 ml) is then transferred to the sample chamber of the disposable, single-use fluidic cartridge (Xpert Carba-R cartridge). The user initiates a test from the system user interface and places the cartridge into the GeneXpert Instrument System. Additional sample preparation, amplification, and real-time detection are all fully-automated and completed by the instrument system.
5. Calibration:
The Xpert Check kit is used by the customer or by Cepheid personnel to perform the calibration check of the instrument. The Xpert Check is not provided with the instrument since the instrument is originally calibrated by Cepheid. A calibration check is recommended on an annual basis. In the GeneXpert Operator’s Manual (Calibration Section), the user is instructed to contact Cepheid Technical Support for information about calibration.
6. Quality Control:
Quality control is addressed for each separately cleared assay to be run on the instrument.
33
{33}
See section M1(c) for information on internal and external controls.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Not applicable
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
34
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.