The ARIES® MRSA Assay is an integrated, real-time, polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection of methicillinresistant Staphylococcus aureus (MRSA) DNA from nasal swabs in patients at risk for nasal colonization. The ARIES® MRSA Assay is intended to aid in the prevention and control of MRSA infections in healthcare settings. The assay is not intended to guide, diagnose, or monitor treatment for MRSA infections. It is not intended to provide results of susceptibility to oxacillin/methicillin. A negative result does not preclude MRSA nasal colonization. Concomitant cultures are necessary to recover organisms for epidemiological typing or for further susceptibility testing. The ARIES® MRSA Assay is indicated for use with ARIES® Systems.
Device Story
The ARIES MRSA Assay is a qualitative, real-time PCR-based in vitro diagnostic test for detecting MRSA DNA in nasal swabs. The device uses the ARIES System or ARIES M1 System to automate sample processing, nucleic acid extraction, and PCR amplification. The assay targets mecA/mecC genes, the orfX complex gene, and the SCCmec junction region. A disposable, single-use cassette contains reagents and an internal sample processing control (SPC). The system uses magnetic beads for extraction and monitors fluorescence during PCR amplification and subsequent melting curve analysis. Results are determined by the ARIES System software and reported to healthcare providers to aid in MRSA infection prevention and control. The assay is not for diagnosing or monitoring MRSA treatment. It provides rapid detection compared to traditional culture methods, though concomitant cultures are required for epidemiological typing or susceptibility testing.
Clinical Evidence
Prospective clinical study (n=1,762 eligible specimens) compared ARIES® MRSA Assay to direct and enriched bacterial culture. Sensitivity: 93.3% (95% CI: 87%-97%); Specificity: 93.5% (95% CI: 92%-95%). Positive percent agreement against direct culture: 93.5%; Negative percent agreement: 92.9%. Analytical studies included precision, reproducibility, LoD, inclusivity (55 strains), challenge study (high/low MIC strains), cross-reactivity (99 organisms), and interference testing.
Technological Characteristics
The device is a qualitative real-time PCR assay using magnetic bead-based nucleic acid extraction. It utilizes an integrated, single-use disposable cassette containing lyophilized PCR reagents, primers, and probes for mecA/mecC, orfX, SCCmec, and an internal SPC. Detection is performed via fluorescence monitoring and melting curve analysis on the ARIES System. The system is networked/standalone capable. Sterilization is not specified for the instrument; cassettes are single-use disposables.
Indications for Use
Indicated for patients at risk for nasal colonization with methicillin-resistant Staphylococcus aureus (MRSA). For prescription use only.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
{0}
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K191742
B Applicant
Luminex Corporation
C Proprietary and Established Names
ARIES MRSA Assay
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| NQX | Class II | 21 CFR 866.1640 - Antimicrobial Susceptibility Test Powder | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain a Substantial Equivalence Determination for the ARIES MRSA Assay
B Measurand:
Target DNA sequences for
(1) mecA and mecC genes for methicillin resistance
(2) orfX complex gene of Staphylococcus aureus
(3) SCCmec junction region
C Type of Test:
Qualitative Real Time Polymerase Chain Reaction (PCR)
K191742 - Page 1 of 19
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
The ARIES MRSA Assay is an integrated, real-time, polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection of methicillin-resistant Staphylococcus aureus (MRSA) DNA from nasal swabs in patients at risk for nasal colonization.
The ARIES MRSA Assay is intended to aid in the prevention and control of MRSA infections in healthcare settings.
The assay is not intended to guide, diagnose, or monitor treatment for MRSA infections. It is not intended to provide results of susceptibility to oxacillin/methicillin. A negative result does not preclude MRSA nasal colonization. Concomitant cultures are necessary to recover organisms for epidemiological typing or for further susceptibility testing.
The ARIES MRSA Assay is indicated for use with ARIES Systems.
B Indication(s) for Use:
Same as Intended Use
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
For use with ARIES Systems
IV Device/System Characteristics:
A Device Description:
The ARIES MRSA Assay is a polymerase chain reaction (PCR)-based qualitative in vitro diagnostic test system which consists of the ARIES System or the ARIES M1 System with their included ARIES Software, a sample processing tube, an assay-specific cassette, and an assay-specific protocol file. The ARIES MRSA Assay cassette is a disposable, single-use cassette containing nucleic acid purification reagents, internal sample process control (SPC), and an assay-specific master mix capable of performing the designated assay on one sample. The ARIES MRSA Assay cassette directly detects methicillin-resistant Staphylococcus aureus (MRSA) from nasal swabs in patients at risk of nasal colonization. Specifically, the ARIES MRSA Assay cassette detects the methicillin resistance genes (mecA and mecC), Staphylococcus aureus orfX gene, the SCCmec junction region, and a DNA Sample Processing Control.
Nasal swab specimens are collected using the Liquid Amies Elution Swab (ESwab) Collection and Transport System, or equivalent. A portion of the sample is transferred to the provided 2 mL ARIES MRSA Sample Processing Tube and vortexed. The processed sample is then transferred to the ARIES MRSA Assay cassette.
K191742 - Page 2 of 19
{2}
The specimen is lysed and nucleic acid is extracted using the ARIES System. An extractable sample processing control (SPC) target present in the ARIES MRSA Assay cassette is processed with the specimen. The SPC controls for recovery of extracted nucleic acid, for inhibitory substances and for PCR reagent and instrument integrity. The Ct value of the SPC is designed to verify nucleic acid extraction, to identify PCR inhibition, if any, and verify proper function of the extraction system and real-time instrument.
The extracted nucleic acid and SPC are transferred via magnetic beads through the cassette to the ARIES MRSA Assay lyophilized PCR reagents in the PCR tube that contain primer pairs and probes specific to mecA/mecC, orfX, SCCmec and the SPC sequence. Each probe is labeled with a distinct fluorophore and detected in a distinct channel of an ARIES System. PCR amplification is performed and assay fluorescence is monitored. Hybridization of a fluorescently labeled probe to the amplified target results in the release of quenching and generation of fluorescence signal that is indicative of PCR generated amplicon. Following amplification, the reaction is heated to separate the fluorescent labeled probe from the amplified target, a process that results in a decrease in the fluorescence signal. The reaction fluorescence is measured during this process and the temperature at which the change in fluorescence is the maximum is the Tm of the amplicon. The instrument fluorescence output is analyzed and test results are determined using the ARIES System software and the ARIES MRSA Assay protocol and run files. ARIES MRSA Assay results may be reported from the ARIES Software or from the optional SYNCT Software.
## B Principle of Operation:
The ARIES MRSA Assay PCR amplification and detection reagents contain primers and probes specific to the methicillin resistance genes (mecA and mecC), a Staphylococcus aureus complex gene (orfX), the SCCmec region, and the Sample Processing Control (SPC) sequence. Each of the probes are labeled with a distinct fluorophore and detected in distinct channels of the ARIES System. The probes contain a fluorophore on the 5' end of the oligonucleotide sequence and a quencher at the 3' end of the oligonucleotide sequence such that in random coil state when the probe is not hybridized to the target sequence, the fluorescent signal is quenched. PCR amplification is performed and assay fluorescence is monitored. Hybridization of a fluorescently labeled probe to the amplified target results in the release of quenching and generation of fluorescence signal that is indicative of PCR generated amplicon.
## V Substantial Equivalence Information:
A Predicate Device Name(s): Bd Max Mrsa Xt, Bd Max Instrument
B Predicate 510(k) Number(s): K133605
C Comparison with Predicate(s):
K191742 - Page 3 of 19
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device (K191742) | Predicate (K133605) |
| | ARIES MRSA Assay | BD Max MRSA XT |
| Intended Use | The ARIES MRSA Assay is an integrated, real-time, polymerase chain reaction (PCR) based qualitative in vitro diagnostic test for the direct detection of methicillin-resistant Staphylococcus aureus (MRSA) DNA from nasal swabs in patients at risk for nasal colonization.The ARIES MRSA Assay is intended to aid in the prevention and control of MRSA infections in healthcare settings.The assay is not intended to guide, diagnose, or monitor treatment for MRSA infections.It is not intended to provide results of susceptibility to oxacillin/methicillin. A negative result does not preclude MRSA nasal colonization.Concomitant cultures are necessary to recover organisms for epidemiological typing or for further susceptibility testing.The ARIES MRSA Assay is indicated for use with ARIES Systems. | The BD MAX MRSA XT assay performed on the BD MAX System is an automated qualitative in vitro diagnostic test for the direct detection of methicillin-resistant Staphylococcus aureus (MRSA) DNA from nasal swabs in patients at risk for nasal colonization. The test utilizes real-time polymerase chain reaction (PCR) for the amplification of MRSA DNA and fluorogenic target specific hybridization probes for the detection of the amplified DNA.The BD MAX MRSA XT assay is intended to aid in the prevention and control of MRSA infections in healthcare settings. It is not intended to diagnose MRSA infections nor guide or monitor treatment for MRSA infections. A negative result does not preclude nasal colonization.Concomitant cultures are necessary to recover organisms for epidemiological typing or for further susceptibility testing. |
| Assay Targets | (1) mecA and mecC genes for methicillin resistance(2) orfX complex gene of S. aureus(3) SCCmec junction region | (1) SCCmec/orfX junction area of methicillin-resistant Staphylococcus aureus (i.e., MREJ for SCCmec Right Extremity Junction). The BD MAX MRSA XT assay is designed to detect MREJ types i, ii, iii, iv, v, vi, vii, ix, xiii, xiv, and xxi; and(2) mecA and mecC genes for methicillin resistance. |
| Sample Types | Nasal Swabs | Same |
| Assay Type | Real-time PCR | Same |
| Assay Results | Qualitative | Same |
| Assay Controls | Sample Processing Control (SPC) | Specimen Processing Control (SPC) |
K191742 - Page 4 of 19
{4}
| Differences | | |
| --- | --- | --- |
| Item | Device (K191742) | Predicate (K133605) |
| | ARIES MRSA Assay | BD Max MRSA XT |
| Extraction Method | Automated by the ARIES Systems | Automated by the BD MAX System |
| Assay Instrument | ARIES Systems | BD MAX System |
| Detection | Fluorescent reporter probes for each target, and melting curve analysis | Fluorogenic target-specific hybridization |
VI Standards/Guidance Documents Referenced:
CLSI. Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition. CLSI document EP05-A3. Wayne, PA: Clinical and Laboratory Standards Institute; 2014.
CLSI. Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition. CLSI document EP07-A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2005.
CLSI. User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline - Second Edition. CLSI document EP12-A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2008.
CLSI. User Verification of Precision and Estimation of Bias; Approved Guideline - Third Edition. CLSI document EP15-A3. Wayne, PA: Clinical and Laboratory Standards Institute; 2014.
CLSI. Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition. CLSI document EP17-A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2012.
CLSI. Assessment of the Diagnostic Accuracy of Laboratory Tests Using Receiver Operating Characteristic Curves; Approved Guideline - Second Edition. CLSI document EP24-A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2011.
CLSI. Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline. CLSI document EP25-A. Wayne, PA: Clinical and Laboratory Standards Institute; 2009.
CLSI. Supplemental Tables for Interference Testing in Clinical Chemistry - First Edition. CLSI document EP37. Wayne, PA: Clinical and Laboratory Standards Institute; 2018.
CLSI. Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically - $11^{th}$ Edition. CLSI document M07. Wayne, PA: Clinical and Laboratory Standards Institute; 2018.
CLSI. Abbreviated Identification of Bacteria and Yeast: Approved Guideline - Second Edition. CLSI document M35-A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2008.
CLSI. Performance Standards for Antimicrobial Susceptibility Testing - $28^{th}$ Edition. CLSI document M100. Wayne, PA: Clinical and Laboratory Standards Institute; 2018.
K191742 - Page 5 of 19
{5}
ISO 14971. Medical devices - Application of risk management to medical devices.
## VII Performance Characteristics (if/when applicable):
## A Analytical Performance:
### 1. Precision/Reproducibility:
#### Site-to-Site Reproducibility
The reproducibility of the ARIES MRSA Assay between sites was evaluated in a study performed by two operators at each of three sites over a period of five days. Each operator tested a blinded panel of five samples in triplicate: two methicillin-resistant *Staphylococcus aureus* (MRSA) strains each spiked at 1x and 5x Limit of Detection (LoD) in simulated nasal matrix (SNM) in modified Liquid Amies (LA) (spiked SNM+LA) and a negative sample (unspiked SNM+LA) using the same lot of reagents (3 sites X 2 operators X 5 days X 3 replicates = 90 data points per panel member). The results of the study demonstrated acceptable reproducibility from site-to-site at target levels close to the limit of detection (LoD) of the assay. The reproducibility study results are presented in Table 1.
Table 1. Summary of results from the ARIES MRSA Assay Site-to-Site Reproducibility Study, stratified by site and overall
| Strain/Level | Positive/Number (%) | | | |
| --- | --- | --- | --- | --- |
| | Site 1 | Site 2 | Site 3 | Overall |
| MRSA mecA+ | 30/30 | 30/30 | 30/30 | 90/90 |
| Moderate Positive 5x LoD^{1} | (100) | (100) | (100) | (100) |
| MRSA mecA+ | 30/30 | 30/30 | 30/30 | 90/90 |
| Low Positive 1x LoD | (100) | (100) | (100) | (100) |
| MRSA mecC+ | 30/30 | 30/30 | 30/30^{2} | 90/90 |
| Moderate Positive 5x LoD | (100) | (100) | (100) | (100) |
| MRSA mecC+ | 30/30^{2} | 30/30 | 30/30 | 90/90 |
| Low Positive 1x LoD | (100) | (100) | (100) | (100) |
| Negative | 0/30 | 0/30^{3} | 0/30 | 0/90 |
| | (0.0) | (0.0) | (0.0) | (0.0) |
¹LoD; Limit of detection for MRSA Strains BEI NR-46232 and BAA-2313
⁵x LoD MRSA mecA+ strain BEI NR-46232: 9.75E+04 CFU/mL; 1x LoD MRSA mecA+ BEI NR-46232: 1.95E+04 CFU/mL; 5x LoD MRSA strain mecC+ ATCC BAA-2313: 3.88E+05 CFU/mL; 1x LoD MRSA mecC+ BAA-2313: 7.75E+04 CFU/mL
²A single sample was reported as Invalid on initial testing; reported as Positive upon repeat
³A single sample was reported as Invalid on initial testing; reported as Negative upon repeat
### Within Laboratory Precision/Repeatability
Within laboratory precision/repeatability of the ARIES MRSA Assay was evaluated by two operators who tested a panel of samples in triplicate on a single ARIES instrument over a period of 5 days (2 operators X 3 replicates X 5 days = 30 replicates per panel member). The
K191742 - Page 6 of 19
{6}
panel members were the same as those used in the Site-to-Site Reproducibility Study, above. The precision/repeatability study results are presented in Table 2. The results demonstrated acceptable repeatability and precision from day-to-day with target levels close to the LoD of the assay.
Table 2. Summary of results from the Within Laboratory Precision/Repeatability Study for the AIREs MRSA Assay
| Strain/Level | Positive/Tested (%) |
| --- | --- |
| MRSA mecA+ | 30/30 |
| Moderate Positive | (100) |
| 5x LoD^{1} | |
| MRSA mecA+ | 30/30 |
| Low Positive | (100) |
| 1x LoD | |
| MRSA mecC+ | 30/30^{2} |
| Moderate Positive | (100) |
| 5x LoD | |
| MRSA mecC+ | 30/30 |
| Low Positive | (100) |
| 1x LoD | |
| Negative | 0/30 |
| | (0.0) |
$^{1}$ LoD; Limit of detection for MRSA Strains BEI NR-46232 and BAA-2313
5x LoD MRSA mecA+ strain BEI NR-46232: 9.75E+04 CFU/mL; 1x LoD MRSA mecA+ BEI NR-46232: 1.95E+04 CFU/mL; 5x LoD MRSA strain mecC+ ATCC BAA-2313: 3.88E+05 CFU/mL; 1x LoD MRSA mecC+ BAA-2313: 7.75E+04 CFU/mL
$^{2}$ A single sample was reported as Invalid on initial testing; reported as Positive upon repeat
2. Linearity:
Not applicable.
3. Analytical Specificity/Interference:
Cross-reactivity Study
The analytical specificity of the ARIES MRSA Assay was evaluated by testing a panel of 99 organisms that may be found in nasal swab specimens (Table 3). Each organism was tested in triplicate in simulated nasal matrix in Modified Liquid Amies at $10^{6}$ CFU/mL for non-viral organisms, $10^{5}$ TCID50/mL for viruses and $5\mu \mathrm{g} / \mathrm{mL}$ for human genomic DNA. No false positive were obtained.
K191742 - Page 7 of 19
{7}
Table 3. Organisms tested for potential cross-reaction in the ARIES MRSA Assay
| Non-Staphylococcal organisms | |
| --- | --- |
| Acinetobacter baumannii | Listeria monocytogenes |
| Acinetobacter haemolyticus | Legionella pneumophila |
| Bacillus cereus | Moraxella catarrhalis |
| Bordetella pertussis | Micrococcus luteus |
| Candida albicans | Mycoplasma pneumoniae |
| Citrobacter freundii | Mycobacterium tuberculosis avirulent |
| Candida glabrata | Neisseria meningitidis |
| Citrobacter koseri | Listeria monocytogenes |
| Chlamydia pneumoniae | Pasteurella aerogenes |
| Corynebacterium bovis | Pseudomonas aeruginosa |
| Corynebacterium flavescens | Pseudomonas fluorescens1 |
| Corynebacterium genitalium | Proteus mirabilis |
| Cryptococcus neoformans | Providencia stuartii |
| Enterobacter aerogenes | Proteus vulgaris |
| Enterobacter cloacae1 | Salmonella enterica subsp. Enterica |
| Enterococcus faecalis | Serratia marcescens |
| Enterococcus faecium | Streptococcus agalactiae |
| Escherichia coli (O157:H7) | Streptococcus anginosus |
| Enterococcus flavescens | Streptococcus mitis |
| Enterococcus gallinarum | Streptococcus mutans |
| Enterococcus hirae | Streptococcus pneumoniae |
| Haemophilus influenzae | Shigella sonnei |
| Klebsiella oxytoca | Streptococcus pyogenes |
| Klebsiella pneumoniae (ESBL-producing) | Streptococcus salivarius |
| Klebsiella pneumoniae (KPC-producing) | Streptococcus sanguinis |
| Lactobacillus crispatus | Streptococcus suis |
| | Yersinia enterocolitica |
| Viruses | |
| Adenovirus Type 40 | Measles virus |
| Adenovirus Type 7 | Mumps virus |
| Coronavirus 229E | Parainfluenza 1 |
| Coronavirus OC43 | Parainfluenza 2 |
| Cytomegalovirus | Parainfluenza 3 |
| Epstein Barr Virus | Rhinovirus type 1A |
| Influenza A | RSV A |
| Influenza B | RSV B |
| Human metapneumovirus | |
| Coagulase Negative Staphylococci (CoNS) | |
| Staphylococcus arlettae | Staphylococcus equorum |
| Staphylococcus captis | Staphylococcus felis |
| Staphylococcus carnosus | Staphylococcus gallinarum |
| Staphylococcus chromogenes | Staphylococcus haemolyticus Z067 |
| Staphylococcus epidermidis (255-01B) | Staphylococcus kloosii |
| Staphylococcus epidermidis (RP12 CIP 106510) | Staphylococcus lentus |
| Staphylococcus epidermidis (MRSE;RP62A) | Staphylococcus pulvereri |
| Staphylococcus epidermidis (CCF 15990) | Staphylococcus simulans Z032 |
| Staphylococcus epidermidis (CCF 16471) | Staphylococcus warneri Z113 |
| Coagulase Positive Staphylococci | |
| Staphylococcus pseudointermedius | Staphylococcus delphini |
K191742 - Page 8 of 19
{8}
K191742 - Page 9 of 19
| Staphylococcus scheleiferi subsp. Coagulans | Staphylococcus intermedius |
| --- | --- |
| Staphylococcus scheleiferi subsp. Schleiferi | Staphylococcus lutrae |
| Methicillin-susceptible Staphylococcus aureus (MSSA) | |
| Staphylococcus aureus BAA-1749 | Staphylococcus aureus 29213 |
| Staphylococcus aureus BAA-1765 | Staphylococcus aureus 0801675 |
| Staphylococcus aureus BAA-1718 | |
| Other | |
| Human genomic DNA | |
¹One replicate was reported as Invalid on initial testing; reported as Negative upon repeat
## Bioinformatic Analysis
In silico analysis was performed to evaluate the potential for cross-reaction of the ARIES MRSA Assay primers and probes. Non-specific amplification and detection of Staphylococcus argenteus by the assay oligos was identified. An investigation into this species indicates high sequence homology between S. argenteus and S. aureus in the orfX, SCCmec and mecA/mecC regions, and that S. argenteus is a member of the S. aureus complex. Due to the high sequence homology in the orfX, SCCmec and mecA/mecC regions between S. argenteus and S. aureus, it would not be possible for the assay oligos to distinguish between methicillin resistant S. argenteus and methicillin resistant S. aureus.
## Carry-over/Contamination Study
The potential for false-positive results with the ARIES MRSA Assay due to within run or between run cross-contamination was evaluated by testing an alternating series of MRSA "high positive" and negative samples in 20 successive instrument runs using two ARIES instruments. The high positive samples contained MRSA at a concentration of $10^{7}$ CFU/mL in simulated nasal matrix in modified Liquid Amies (SNM+LA). Negative samples comprised SNM+LA alone. One replicate was reported Invalid on initial testing; reported as Positive upon repeat testing. The expected results were obtained for all MRSA positive and negative samples (60/60 each).
## Potentially Interfering Substances Study
The potential for interference with the ARIES MRSA Assay was evaluated with endogenous and exogenous substances that may be present in nasal swab specimens (Table 4). Each substance was tested in triplicate in the presence and absence of one strain of mecA+ MRSA (NR-46232) and one strain of mecC+ MRSA (BAA-2313) at 3x LoD prepared in simulated nasal matrix in Modified Liquid Amies (SNM+LA). No false positives or false negatives were obtained.
Table 4. Substances evaluated for potential interference with the ARIES MRSA Assay
| Substance | Test Concentration |
| --- | --- |
| Whole Blood | 5% (v/v) |
| Mucin | 5 mg/mL |
| Phenylephrine | 0.03 μg/mL |
| Drixoral (Oxymetazoline) | 10% (v/v) |
| Benzalkonium chloride | 0.12% |
| Propylene glycol | 20% (v/v) |
| Sorbitol¹ | 6.45% |
| Benzyl alcohol¹ | 0.5% (v/v) |
{9}
| Hypromellose | 0.10% |
| --- | --- |
| Phosphoric acid | 1.282 mg/mL |
| Beclomethasone | 8.4 μg/mL |
| Dexamethasone | 12 μg/mL |
| Flunisolide | 5 μg/mL |
| Triamcinolone | 22 μg/mL |
| Budesonide | 6.30E-03 μg/mL |
| Mometasone | 4.50E-04 μg/mL |
| Flonase (Fluticasone) | 1.26E-03 μg/mL |
| ZICAM² (Galphimia glauca, Histaminum hydrochloricum) | 10% (v/v) |
| Benzocaine | 75 μg/mL |
| Menthol | 0.5 mg/mL |
| Zanamivir³ | 1 mg/mL |
| Mupirocin | 50 μg/mL |
| Tobramycin | 33 μg/mL |
| FluMist (Live intranasal influenza virus vaccine) | 10% (v/v) |
One replicate of SNM+LA negative matrix was reported as Invalid on initial testing; reported as Negative upon repeat
One replicate containing mecA+ MRSA strain was reported as Invalid on initial testing; reported as Positive upon repeat
One replicate containing mecC+ MRSA strain was reported as Invalid on initial testing; reported as Positive upon repeat
## Microbial Interference Study
The potential for interference with the ARIES MRSA Assay by organisms that may be present in nasal swab specimens was investigated using the same list of species that was evaluated for potential cross-reactivity (refer to Table 3). Testing was performed in triplicate with each potentially interfering species in the presence of one strain of mecA+ MRSA (NR-46232) or one strain of mecC+ MRSA (BAA-2313) at 3x LoD. The potentially interfering species were tested in triplicate in simulated nasal matrix in Modified Liquid Amies at $10^{6}$ CFU/mL for non-viral organisms, $10^{5}$ TCID50/mL for viruses and $5\mu \mathrm{g} / \mathrm{mL}$ for human genomic DNA. One replicate containing mecA+ MRSA was reported as Invalid for S. gallinarum on initial testing but was MRSA positive upon repeat testing. One replicate containing mecC+ MRSA strain was reported as Invalid for the specimens containing A. baumannii, Rhinovirus type 1A and S. warneri Z113; however, each specimen was reported as MRSA positive upon repeat testing. These results are acceptable.
## Competitive Interference Study
Competitive interference was tested with methicillin-resistant Staphylococcus aureus (MRSA) at 1x the limit of detection (LoD) and the co-infecting agent, methicillin-susceptible Staphylococcus aureus (MSSA) or methicillin-resistant coagulase-negative Staphylococci (MRCoNS), at increasing concentrations. Each combination was tested in triplicate. The results showed that the ARIES MRSA Assay detected MRSA at 1x LoD in the presence of high concentration of MSSA ( $\sim2\mathrm{E}+04\mathrm{x}$ LoD) or MRCoNS ( $\sim1\mathrm{E}+05\mathrm{x}$ LoD). No competitive interference in the ARIES MRSA Assay was observed for co-infections of MRSA with MSSA or MRCoNS.
K191742 - Page 10 of 19
{10}
4. Assay Reportable Range:
Not applicable.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Specimen Stability
The stability of nasal swabs for use with the ARIES MRSA Assay was evaluated analytically by testing six replicates of each of two methicillin-resistant Staphylococcus aureus (MRSA) strains prepared at 3x and 10x Limit of Detection (LoD) in Natural Nasal Matrix (NNM) and stored under different conditions. Unseeded negative sample (NNM only) was included to assess the effect of specimen storage on the performance of the Sample Processing Control (SPC). Specimens stored at 2±2°C were tested at 5 time points over 10 days and specimens stored at 30±1°C were tested at 7 time points over 10 days. The results showed that MRSA specimens stored at both temperatures (2°C and 30°C) generated 100% expected MRSA positive results across all time points tested. An overall cassette invalid rate of 0.53% (4/757) was observed for the testing of MRSA strains and NNM samples in this study. The results of these studies support the stability of nasal swabs for use with the ARIES MRSA Assay when collected using the Liquid Amies Elution Swab (Eswab) Collection and Transport system for up to 10 days at 2-30°C.
Reagent Stability
The shelf-life of the ARIES MRSA Assay cassettes was evaluated in a real-time stability study performed on three lots of reagents that were stored either refrigerated (2-8°C) or at room temperature (15-30°C). The results from the study support assignment of an expiration date 19 months from the day of manufacture for the assay cassettes when stored under the recommended conditions.
Cassette open box stability evaluated the performance of the ARIES MRSA Assay Cassettes after removal from the cassette pouch and exposed to ambient temperatures, humidity and light for 10 hours. Over the course of 10 hours, all three lots of cassettes produced expected results, showing that ARIES MRSA Assay Cassettes are stable in ambient temperatures for up to 10 hours after they have been removed from the storage pouch.
Sample Processing Control
Each ARIES MRSA Assay cassette contains a Sample Processing Control (SPC) that is designed to verify nucleic acid extraction, and proper reagent, cassette, ARIES System, and assay protocol performance. The SPC has a known melting temperature (Tm) range and Ct range. Each time an assay is run, the system measures the temperature and fluorescence intensity of the SPC control to ensure the thermal and optical subsystems have remained in calibration.
External Controls
External controls should be tested according to guidelines or requirements of local, provincial and/or federal regulations or accreditation organizations. A reference methicillin-resistant Staphylococcus aureus strain or well characterized methicillin-resistant Staphylococcus aureus clinical isolates may be used as positive controls. The ARIES MRSA Assay Kit does not include external positive or negative controls.
K191742 - Page 11 of 19
{11}
External Positive and Negative Controls were tested on a daily basis during the prospective Clinical Study using a total of five ARIES systems and three ARIES MRSA Assay cassette lots. The Positive External Control comprised a standardized suspension of a strain of MRSA at $1.95\mathrm{E} + 05$ CFU/mL (10x LoD). The Negative External Control comprised a standardized suspension of a strain of S. epidermis at $4.23\mathrm{E} + 04$ CFU/mL (10x LoD). On initial testing, 182/183 (99.5%) Positive and 178/183 (97.3%) Negative External Controls produced the expected results. Upon repeat testing, all controls produced the expected results.
# 6. Detection Limit:
The Limit of Detection (LoD) of the ARIES MRSA Assay was estimated for two strains of MRSA by testing various dilutions of enumerated cell stocks in natural nasal matrix (NNM). The LoD for each strain was then confirmed by testing a further 20 replicates at the lowest target level that produced $100\%$ positive results. The LoD was defined as the lowest concentration tested at which $\geq 95\%$ of assay replicates produced positive results. For MRSA mecA+ BAA-2312, the LoD was determined to be $1.55 \times 10^{4} \mathrm{CFU/mL}$ and for MRSA mecC+ NRA-46232, it was $7.75 \times 10^{4} \mathrm{CFU/mL}$ .
# Inclusivity (Analytical Reactivity)
The inclusivity of the ARIES MRSA Assay was evaluated by testing 55 strains of MRSA in simulated nasal matrix in Modified Liquid Amies (SNM+LA), which include those tested in the LoD Study. All 55 strains produced $3/3$ positive results at $3\mathrm{x}$ LoD $(2.32\times 10^{5}\mathrm{CFU / mL}$ for mecC+ strains BAA-2312 and BAA-2313; $4.65\times 10^{4}\mathrm{CFU / mL}$ for all other strains). Analytical reactivity study results are presented in Table 5.
Table 5. MRSA strains used to evaluate the inclusivity of the ARIES MRSA Assay
| Strain Description | Source | Catalog # | Strain/ Location | PFGE Type | SCCmec Type |
| --- | --- | --- | --- | --- | --- |
| MRSA mecA+ | BEI | NR-46232 (LoD strain) | NRS703/ Minnesota | USA300 | IV |
| | ATCC | BAA-38 | N/A | N/A | I |
| | ATCC | BAA-1686 | N/A | N/A | II |
| | ATCC | BAA-16871 | N/A | N/A | II |
| | ATCC | BAA-1692 | N/A | USA100 | II |
| | ATCC | BAA-1681 | N/A | USA100 | II |
| | ATCC | BAA-1682 | N/A | USA100 | II |
| | ATCC | BAA-1699 | N/A | USA100 | II |
| | BEI | NR-46250 | NRS721/ Oregon | USA100 | II |
| | ATCC | BAA-1761 | N/A | USA100 | II |
| | ATCC | BAA-1750 | N/A | USA200 | II |
| | BEI | NR-46251 | NRS722/ Oregon | USA200 | II |
| | ATCC | BAA-39 | N/A | N/A | III |
| | ATCC | BAA-40 | N/A | N/A | IIIa |
| | ATCC | BAA-1717 | N/A | USA300 | IVa |
| | ATCC | BAA-1762 | N/A | USA300 | IVb |
| | ATCC | BAA-1680 | N/A | USA300 | IVa |
| | BEI | NR-46070 | NRS384/ Mississippi | USA300 | IV |
| | ATCC | BAA-1683 | N/A | USA400 | IVa |
| | ATCC | BAA-1707 | N/A | USA400 | IV |
K191742 - Page 12 of 19
{12}
| | ATCC | BAA-1752 | N/A | USA400 | IV |
| --- | --- | --- | --- | --- | --- |
| | ATCC | BAA-1757 | N/A | USA400 | IV |
| | ATCC | BAA-1696 | N/A | USA400 | IVa |
| | BEI | NR-46207 | NRS678/ Connecticut | USA500 | IV |
| | ATCC | BAA-1689 | N/A | USA500 | IV |
| | BEI | NR-46220 | NRS691/ Georgia | USA500 | IV |
| | ATCC | BAA-1688 | N/A | N/A | V |
| | ATCC | BAA-42 | N/A | N/A | VI |
| | BEI | NR-46177 | NRS648/ California | USA600 | II |
| | ATCC | BAA-1751 | N/A | USA600 | II |
| | BEI | NR-46218 | NRS689/ Georgia | USA700 | IV |
| | ATCC | BAA-1766 | N/A | USA700 | V |
| | BEI | NR-46221 | NRS692/ Georgia | USA800 | IV |
| | ATCC | BAA-1771 | N/A | USA800 | IV |
| | BEI | NR-46197 | NRS668/ Colorado | USA800 | N/A |
| | ATCC | BAA-1747 | N/A | USA1000 | IV |
| | ATCC | BAA-1769 | N/A | USA1000 | IV |
| | ATCC | BAA-1764 | N/A | USA1100 | IV |
| | ATCC | BAA-1767 | N/A | USA1100 | IV |
| | ATCC | 33592 | N/A | N/A | III |
| | ATCC | 33593 | N/A | N/A | III |
| | ATCC | 43300 | N/A | N/A | II |
| | ATCC | 700698 | N/A | N/A | II |
| | ATCC | 700699 | N/A | N/A | II |
| | ATCC | 700787 | N/A | N/A | II |
| | ATCC | 700788 | N/A | N/A | II |
| | ATCC | 700789 | N/A | N/A | II |
| | ATCC | BAA-41 | N/A | N/A | II |
| | ATCC | BAA-43 | N/A | N/A | IIIa |
| | ATCC | BAA-44 | N/A | N/A | I |
| ATCC | BAA-1720 | N/A | N/A | II | |
| ATCC | BAA-2094 | N/A | N/A | V | |
| ATCC | BAA-2096 | N/A | N/A | IV | |
| MRSA mecC+ | ATCC | BAA-2313 (LoD strain) | N/A | N/A | XI |
| | ATCC | BAA-2312 | N/A | N/A | XI |
One replicate was reported as Invalid on initial testing; reported as Positive upon repeat
# Bioinformatic Analysis
The inclusivity of the ARIES MRSA primers and probes for the targeted regions of the genome was analyzed in silico using the Basic Local Alignment Search Tool (BLAST). The region was shown to be well conserved, with nearly $100\%$ of sequences with $\geq 85\%$ identity for orfX and mecA/mecC and approximately $93\%$ of sequences with $\geq 85\%$ identity for SCCmec junction. These results are acceptable.
# Challenge Study
An additional analytical study was carried out to evaluate the analytical performance of the ARIES MRSA Assay using a panel of challenge strains (Table 6). The challenge panel
K191742 - Page 13 of 19
{13}
included 16 methicillin-resistant Staphylococcus aureus (MRSA) strains with high minimum inhibitory concentration (MIC) values of $\geq 16~\mu \mathrm{g / mL}$ oxacillin and 17 MRSA strains with low MIC values of $\leq 8~\mu \mathrm{g / mL}$ oxacillin, four borderline oxacillin-resistant Staphylococcus aureus (BORSA) strains, 16 empty cassette variants of Staphylococcus aureus strains, and one methicillin-resistant Staphylococcus epidermidis (MRSE) strain. The strains were tested in triplicate at $3\mathrm{x}$ LoD (MRSA strains) or $10^{6}$ CFU/mL (all other strains). Two strains of MRSA with high MIC values and one strain of MRSA with low MIC value did not generate the expected MRSA positive results when tested at $3\mathrm{x}$ LoD, and were re-tested at $5\mathrm{x}$ LoD and generated the expected MRSA positive results (100% MRSA Positive). The four BORSA strains, the MRSE strain, and all empty cassette variants of Staphylococcus aureus generated the expected 0% MRSA positive results (100% MRSA Negative). Analytical challenge study results are presented in Table 6.
Table 6. ARIES MRSA Assay Analytical Challenge Study Results
| Target Description | Source | Strain ID | ARIES MRSA Positivity (%) |
| --- | --- | --- | --- |
| MRSA with High Oxacillin MIC | ARLG | ARLG-1643 | 3/3 (100) |
| | ARLG | ARLG-1644 | 3/3 (100) |
| | ARLG | ARLG-1645 | 3/3 (100) |
| | ARLG | ARLG-1646 | 3/3 (100) |
| | ARLG | ARLG-1662 | 3/3 (100) |
| | ARLG | ARLG-1669 | 0/31(0) |
| | ARLG | ARLG-1604 | 3/3 (100) |
| | ARLG | ARLG-1613 | 2/31(66.7) |
| | CDC AR Bank | AR-0215 | 3/3 (100) |
| | CDC AR Bank | AR-0218 | 3/3 (100) |
| | CDC AR Bank | AR-0219 | 3/3 (100) |
| | CDC AR Bank | AR-0220 | 3/3 (100) |
| | CDC AR Bank | AR-0223 | 3/3 (100) |
| | CDC AR Bank | AR-0224 | 3/3 (100) |
| | CDC AR Bank | AR-0227 | 3/3 (100) |
| | CDC AR Bank | AR-0228 | 3/3 (100) |
| MRSA with Low Oxacillin MIC | ARLG | ARLG-1642 | 2/31(66.7) |
| | Lyon University | 20130237 | 3/3 (100) |
| | Lyon University | 20130524 | 3/3 (100) |
| | CDC AR Bank | AR-0216 | 3/3 (100) |
| | CDC AR Bank | AR-0217 | 3/3 (100) |
| | CDC AR Bank | AR-0221 | 3/3 (100) |
| | CDC AR Bank | AR-0225 | 3/3 (100) |
| | CDC AR Bank | AR-0226 | 3/3 (100) |
| | ATCC | BAA-1688 | 3/3 (100) |
| | CDC AR Bank | AR-472 | 3/3 (100) |
| | CDC AR Bank | AR-473 | 3/3 (100) |
| | CDC AR Bank | AR-474 | 3/3 (100) |
| | CDC AR Bank | AR-475 | 3/3 (100) |
| | CDC AR Bank | AR-476 | 3/3 (100) |
| | CDC AR Bank | AR-477 | 3/3 (100) |
| | CDC AR Bank | AR-478 | 3/3 (100) |
| | CDC AR Bank | AR-479 | 3/3 (100) |
| BORSA | CDC AR Bank | AR-0489 | 0/3 (0) |
| | CDC AR Bank | AR-0490 | 0/3 (0) |
| | CDC AR Bank | AR-0491 | 0/3 (0) |
| | CDC AR Bank | AR-0492 | 0/3 (0) |
| | Lyon University | 20101270 | 0/3 (0) |
K191742 - Page 14 of 19
{14}
| Empty Cassette Variant of SA | Lyon University | 20112896 | 0/3 (0) |
| --- | --- | --- | --- |
| | Lyon University | 20112911 | 0/3 (0) |
| | Lyon University | 20120556 | 0/3 (0) |
| | Lyon University | 20120844 | 0/3 (0) |
| | Lyon University | 20120871 | 0/3² (0) |
| | Lyon University | 20120984 | 0/3 (0) |
| | Lyon University | 20121469 | 0/3 (0) |
| | Lyon University | 20121544 | 0/3² (0) |
| | Lyon University | 20121635 | 0/3 (0) |
| | Lyon University | 20121891 | 0/3 (0) |
| | Lyon University | 20130769 | 0/3 (0) |
| | Lyon University | 20131190 | 0/3 (0) |
| | Lyon University | 20131273 | 0/3 (0) |
| | Lyon University | 20131727 | 0/3 (0) |
| | Lyon University | 20140852 | 0/3 (0) |
| MRSE | ATCC | 51625 | 0/3 (0) |
$^{1}$ Repeat testing at 5x LoD reported as 3/3 (100%) MRSA Positive
2One replicate was reported as Invalid on initial testing; reported as Negative upon repeat
# 7. Assay Cut-Off:
For the ARIES MRSA Assay, each target (mecA/mecC, orfX, and SCCmec) has a Ct cut-off, $\mathrm{T}_{\mathrm{m}}$ window, and $\mathrm{T}_{\mathrm{m}}$ Peak Threshold. In addition, the internal sample processing control (SPC) also has a corresponding Ct cut-off, $\mathrm{T}_{\mathrm{m}}$ window, and $\mathrm{T}_{\mathrm{m}}$ Peak Threshold. Collectively, the cut-off values compose the assay protocol file parameters, which are used to determine the assay result for the detection of the target as Positive, Negative, or Invalid. These values are hard-coded into the ARIES MRSA Assay Protocol File and are not modifiable. The Assay Protocol File parameters were determined, and their performance in the ARIES MRSA Assay were evaluated according to the following general procedure:
- Initial Assay Protocol File parameters were set during internal optimization and benchmarking studies.
- The final Assay Protocol File parameters were then established during internal verification studies using data from optimization, benchmarking and verification.
- The selected Assay Protocol File parameter values were utilized in the determination of assay performance in the multi-site clinical trial conducted for the ARIES MRSA Assay.
# B Comparison Studies:
# 1. Method Comparison with Predicate Device:
Not applicable.
# 2. Matrix Comparison:
Comparison of Performance with Natural and Simulated Matrices
To provide a sufficient quantity of material for testing, a simulated nasal matrix was used for the majority of Analytical Studies. Testing with simulated matrix was performed in accordance with the standard assay procedure by transferring $200\mu \mathrm{L}$ of the sample to the ARIES MRSA Assay cassette.
K191742 - Page 15 of 19
{15}
The suitability of the simulated matrix for use in analytical testing was evaluated in a comparison study with natural clinical matrix. The two matrices were tested in parallel as part of the LoD Study using MRSA strain NR-46232. The results demonstrated similar analytical sensitivity in both matrices and there were negligible differences in LoD and MRSA targets. The study therefore provided acceptable evidence to support the use of simulated matrix in the Analytical Studies to characterize the performance of the ARIES MRSA Assay.
## Nasal Swab Comparison Study
A nasal swab equivalency study was performed to evaluate the reproducibility of the ARIES MRSA Assay with two different nasal swab types, Regular Nylon Flocked Swab (Copan Catalog Number: 480C) and Flexible Minitip Nylon Flocked Swab (Copan Catalog Number: 482C). The swabs were evaluated using one strain of methicillin-resistant Staphylococcus aureus (MRSA) mecA+ (NR-46232) at three concentrations, as well as a negative sample (Simulated nasal matrix in Modified Liquid Amies, SNM+LA). Samples at intermediate concentrations prepared in SNM+LA were transferred to Modified Liquid Amies using each of the two nasal swab types to reach the final testing concentrations at 3x LoD, 5x LoD and 10x LoD, respectively, and then tested on the ARIES MRSA Assay. The test results demonstrated that both swab types generated 100% expected MRSA positivity for each strain of the concentrations tested. Both swab types also generated 0% positivity (100% negativity) for negative samples. Swab equivalency study results are presented in Table 7.
Table 7. ARIES MRSA Assay Nasal Swab Equivalency Results
| Assay Target | Part Number | Test Concentration (CFU/mL) | MRSA Positivity (%) | |
| --- | --- | --- | --- | --- |
| | | | Regular swab (Copan 480C) | Flexible mini-tip swab (Copan 482C) |
| MRSA mecA+ | NR-46232 | 3x LoD (5.85E+04) | 6/6 (100) | 6/6 (100) |
| | | 5x LoD (9.75E+04) | 6/6 (100) | 6/6 (100) |
| | | 10x LoD (1.95E+05) | 6/6 (100) | 6/6 (100) |
| Negative (SNM+ LA) | N/A | N/A | 0/6 (0) | 0/6 (0) |
## C Clinical Studies:
### 1. Clinical Sensitivity:
Clinical performance of the ARIES MRSA Assay for nasal swab specimens collected from patients at risk for methicillin-resistant Staphylococcus aureus (MRSA) colonization was established through a clinical study.
Performance of the ARIES MRSA Assay was evaluated prospectively from August 2018 to February 2019 at four (4) geographically distinct clinical sites within the United States using the ARIES System. Specimens included in the clinical study consisted of excess leftover de-identified, nasal clinical specimens collected using the Liquid Amies Elution Swab (Eswab) Collection and Transport system, or equivalent, from patients at risk for nasal colonization. All eligible clinical specimens were tested by both the reference method (direct and enriched bacterial culture) and ARIES MRSA Assay and the results compared. Reference method
K191742 - Page 16 of 19
{16}
testing was performed at a centralized testing facility while ARIES MRSA Assay testing was performed at each clinical site on their own clinical specimens.
A total of 2254 nasal swab specimens from subjects at risk for MRSA nasal colonization were collected. Of these 2,254 specimens, 472 were excluded from the study based on inclusion/exclusion criteria leaving a total of 1,782 unique specimens that met the predetermined eligibility criteria for inclusion in the study.
In total, 1,782 specimens were enrolled in the study and tested for methicillin-resistant Staphylococcus aureus by both the reference method and the ARIES MRSA Assay. There were 20 specimens that when tested with ARIES MRSA Assay yielded an invalid result due to run failure or instrument error giving an invalid rate of 1.1% (20/1,782). None of these specimens were re-tested due to insufficient specimen volume.
For the 1,762 eligible specimens that were included in the device performance calculations, clinical sensitivity of the ARIES MRSA Assay against direct and enriched bacterial culture (Table 8), was 93.3% (97/104) with a lower bound 95% confidence interval of 87%. Clinical specificity of the ARIES MRSA Assay was 93.5% (1550/1658) with a lower bound 95% confidence interval of 92%. Of the 108 specimens that were MRSA-negative by culture but MRSA-positive by the ARIES MRSA Assay, culture showed that 63 specimens were S. aureus and 45 were negative (no growth).
When ARIES MRSA Assay was compared to Direct Culture only (Table 9), the positive percent agreement of the ARIES MRSA Assay was 93.5% (87/103) with a lower bound 95% confidence interval of 87%. Negative percent agreement of the ARIES MRSA Assay was 92.9% (1550/1658) with a lower bound 95% confidence interval of 92%. Overall, performance was determined to be acceptable.
Table 8: ARIES MRSA Assay Performance Compared to Direct and Enriched Culture
| ARIES MRSA Assay | Direct and Enriched Bacterial Culture for MRSA | | |
| --- | --- | --- | --- |
| | Positive | Negative | TOTAL |
| Positive | 97 | 108 | 205 |
| Negative | 7 | 1550 | 1557 |
| TOTAL | 104 | 1658 | 1762 |
| Sensitivity (95% CI) | 93.3% (87% - 97%) | | |
| Specificity (95% CI) | 93.5% (92% - 95%) | | |
K191742 - Page 17 of 19
{17}
Table 9: ARIES MRSA Assay Performance Compared to Direct Culture (Informational Only)
| ARIES MRSA Assay | Direct Bacterial Culture for MRSA | | |
| --- | --- | --- | --- |
| | Positive | Negative | TOTAL |
| Positive | 87 | 118 | 205 |
| Negative | 6 | 1551 | 1557 |
| TOTAL | 93 | 1669 | 1762 |
| Positive Percent Agreement (95% CI) | 93.5% (87% - 97%) | | |
| Negative Percent Agreement (95% CI) | 92.9% (92% - 94%) | | |
2. Clinical Specificity:
See Clinical Sensitivity above.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
The performance of the ARIES MRSA Assay was evaluated in a prospective Clinical Study conducted at four (4) sites in the US. The overall prevalence of MRSA in nasal swab specimens was 5.9% (104/1762) as determined by culture (direct plus enriched) and 11.6% (205/1762) as determined by the ARIES assay. In Table 10, the prevalence of MRSA as determined by the ARIES assay is stratified by the age and gender of the subjects.
Table 10. Prevalence of MRSA positive subjects stratified by age and gender
| | Number of Subjects | ARIES MRSA Assay Positive | % Prevalence^{1} |
| --- | --- | --- | --- |
| Gender | | | |
| Male | 954 | 122 | 12.8 |
| Female | 808 | 83 | 10.3 |
| Overall | 1762 | 205 | 11.6 |
| Age (yrs) | | | |
| <2 | 405 | 21 | 5.2 |
| 2 – 11 | 251 | 21 | 8.4 |
| 12 – 21 | 189 | 19 | 10.1 |
| 22 – 59 | 481 | 74 | 15.4 |
| ≥60 | 436 | 70 | 16.1 |
| Overall | 1762 | 205 | 11.6 |
¹As determined by the ARIES MRSA Assay
K191742 - Page 18 of 19
{18}
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K191742 - Page 19 of 19
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.