← Product Code [NQX](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NQX) · K190771

# Xpert MRSA/SA Blood Culture, GeneXpert Dx System, GeneXpert Infinity-48s System, GeneXpert Infinity-80 System (K190771)

_Cepheid · NQX · Apr 25, 2019 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NQX/K190771

## Device Facts

- **Applicant:** Cepheid
- **Product Code:** [NQX](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NQX.md)
- **Decision Date:** Apr 25, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 866.1640
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The Cepheid® Xpert® MRSA/SA Blood Culture test, performed on the GeneXpert® Instrument Systems, is a qualitative in vitro diagnostic test intended for the detection of Staphylococcus aureus (SA) and methicillin-resistant Staphylococcus aureus (MRSA) DNA directly from positive blood cultures. The assay utilizes automated real-time polymerase chain reaction (PCR) for the amplification of MRSA/SA specific DNA targets and fluorogenic target-specific hybridization probes for the real-time detection of the amplified DNA. The assay is performed directly on positive blood culture samples using BD BACTECTM Plus Aerobic/F, BacT/ALERT® SA (Standard Aerobic) or VersaTREK REDOX 1® (aerobic) blood culture bottles that are determined as Gram Positive Cocci in Clusters (GPCC) or as Gram Positive Cocci in singles (GPC) by Gram stain. The Xpert MRSA/SA Blood Culture test is indicated for use in conjunction with other laboratory tests, such as culture, and clinical data available to the clinician as an aid in the detection of MRSA/SA from positive blood cultures. Subculturing of positive blood cultures is necessary to recover organisms for susceptibility testing or for epidemiological typing. The Cepheid Xpert MRSA/SA Blood Culture test is not intended to monitor treatment for MRSA/SA infections.

## Device Story

Device performs rapid, automated qualitative detection of MRSA and SA DNA from positive blood culture bottles; utilizes real-time PCR; targets spa, mecA, and SCCmec genes. Sample preparation, amplification, and detection integrated within single-use disposable fluidic cartridge; processed on GeneXpert Instrument Systems (Dx, Infinity-48s, Infinity-80). Workflow: aliquot of positive blood culture transferred to elution reagent, then to cartridge; instrument automates cell lysis via ultrasonic horn, fluid movement via syringe drive, and thermocycling via I-CORE module. Results provided in ~60 minutes. Used in clinical laboratories; operated by trained personnel. Output aids clinicians in rapid identification of MRSA/SA, facilitating timely clinical decision-making and patient management. Benefits include faster detection compared to traditional culture methods.

## Clinical Evidence

No new clinical trials conducted. Substantial equivalence supported by re-analysis of original clinical performance data and a subset of original analytical performance data (LoD, inclusivity, exclusivity, potential interfering substances, reproducibility, and precision) to validate the updated Assay Definition File and software.

## Technological Characteristics

Nucleic acid amplification test (NAAT) using real-time PCR. Disposable, multi-chambered fluidic cartridges. Internal controls: Sample Processing Control (SPC) and Probe Check Control (PCC). Targets: spa, mecA, and SCCmec genes. Automated cell lysis (ultrasonic) and thermocycling (I-CORE). Software-based rules-based algorithms for result interpretation. Connectivity via GeneXpert Instrument Systems (Dx, Infinity-48s, Infinity-80).

## Regulatory Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Predicate Devices

- Xpert MRSA/SA Blood Culture Assay ([K130894](/device/K130894.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov

# SPECIAL 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

## I Background Information:

A 510(k) Number
K190771

B Applicant
Cepheid

C Proprietary and Established Names
Xpert MRSA/SA Blood Culture, GeneXpert Dx System, GeneXpert Infinity-48s System, GeneXpert Infinity-80 System

D Regulatory Information
|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  NQX | Class II | 21 CFR 866.1640 - Antimicrobial Susceptibility Test Powder | IM - Immunology & MI - Microbiology  |
|  OOI | Class II | 21 CFR 862.2570 - Instrumentation for clinical multiplex test systems | CH - Clinical Chemistry  |

## II Review Summary:

This 510(k) submission contains information/data on modifications made to the submitter's own CLASS II device requiring 510(k). The following items are present and acceptable

1. The name and 510(k) number of the SUBMITTER'S previously cleared device:
Xpert MRSA/SA Blood Culture Assay
510(k) number: K130894

K190771 - Page 1 of 6

{1}

2. Submitter's statement that the INDICATIONS FOR USE/INTENDED USE of the modified device as described in its labeling HAS CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).

This change was for the device name changed from "Xpert MRSA/SA Blood Culture Assay" to "Xpert MRSA/SA Blood Culture".

3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.

This change was for the incorporation of enhanced algorithms (rules-based) into the Xpert MRSA/SA Blood Culture assay definition file (ADF) for determining SA positive, MRSA positive, SA negative and MRSA negative test results. This change is a post-PCR rules-based analysis. The minimum software requirement of the GeneXpert Dx software version 5.3, on which the updated ADF is locked and the rules-based, post-PCR analysis settings are supported and compatible with Infinity software version 6.8 and higher.

The Xpert MRSA/SA Blood Culture Package Insert was updated to revise the Results and Interpretations table to reflect the revisions to assay test results reporting and to update the minimum software requirement to GeneXpert Dx software version 5.3 or GeneXpert Infinity Xpertise 6.8.

The Xpert MRSA/SA Blood Culture Package Insert was updated to include performance characteristics with a summary of re-analyses of original data with newly released updated software modifications.

4. Comparison Information (i.e., similarities and differences) to the submitter's legally marketed predicate device including, labeling, intended use, and physical characteristics.

|  Device & Predicate Device(s): | Device: Xpert MRSA/SA Blood Culture (K190771) | Predicate: Xpert MRSA/SA Blood Culture Assay (K130894)  |
| --- | --- | --- |
|  Similarities  |   |   |
|  Intended Use/Indications For Use | The Cepheid Xpert MRSA/SA Blood Culture test, performed on the GeneXpert Instrument Systems, is a qualitative in vitro diagnostic test intended for the detection of Staphylococcus aureus (SA) and methicillin-resistant Staphylococcus aureus (MRSA) DNA directly from positive blood cultures. The assay utilizes automated real-time polymerase chain reaction (PCR) for the amplification of MRSA/SA | Same  |

K190771 - Page 2 of 6

{2}

K190771 - Page 3 of 6
|   | specific DNA targets and
fluorogenic target-specific
hybridization probes for the real-
time detection of the amplified
DNA. The assay is performed
directly on positive blood culture
samples using BD BACTEC Plus
Aerobic/F, BacT/ALERT SA
(Standard Aerobic) or VersaTREK
REDOX 1 (aerobic) blood culture
bottles that are determined as Gram
Positive Cocci in Clusters (GPCC)
or as Gram Positive Cocci in singles
(GPC) by Gram stain. The Xpert
MRSA/SA Blood Culture test is
indicated for use in conjunction with
other laboratory tests, such as
culture, and clinical data available to
the clinician as an aid in the
detection of MRSA/SA from
positive blood cultures. Subculturing
of positive blood cultures is
necessary to recover organisms for
susceptibility testing or for
epidemiological typing. The
Cepheid Xpert MRSA/SA Blood
Culture test is not intended to
monitor treatment for MRSA/SA
infections. |   |
| --- | --- | --- |
|  Specimen Type | Positive Blood Culture | Same  |
|  Technological Principles | Fully automated nucleic acid
amplification (DNA); real-time PCR | Same  |
|  Test Cartridge | Disposable single-use, multi-
chambered fluidic cartridge | Same  |
|  Sample Preparation | Self-contained and automated after
mixed specimen is added to
cartridge. All other reagents are
contained in the cartridge | Same  |
|  Probes | TaqMan Probes | Same  |
|  Internal Controls | Sample processing (SPC) and probe
check control (PCC) | Same  |
|  DNA Target Sequence | Sequence incorporating the insertion
site (attB) of
Staphylococcal Cassette
Chromosome mec (SCCmec) for
detection of MRSA. | Same  |
|  DNA Target Sequence | Sequence specific to
methicillin/oxacillin resistance
(mecA gene) | Same  |
|  DNA Target Sequence | Sequence specific to Staphylococcus | Same  |

{3}

|   | aureus species (spa gene) |   |
| --- | --- | --- |
|  Ability to identify correctly “Empty Cassette Variants” | Yes, sequence specific to Staphylococcus aureus species (mecA gene) | Same  |
|  Time to Result | Approximately 60 minutes to result | Same  |
|  Device & Predicate Device(s): | Device: Xpert MRSA/SA Blood Culture (K190771) | Predicate: Xpert MRSA/SA Blood Culture Assay (K130894)  |
| --- | --- | --- |
|  Differences  |   |   |
|  Trade Name | Xpert MRSA/SA Blood Culture | Xpert MRSA/SA Blood Culture Assay  |
|  Instrument System | Cepheid GeneXpert Dx Systems and GeneXpert Infinity-48s and Infinity-80 Systems | Cepheid GeneXpert Dx Systems, GeneXpert Infinity-48 System, and GeneXpert Infinity-48s and Infinity-80 Systems  |
|  Minimum software requirements | GeneXpert Dx software version 5.3 and higher, GeneXpert Infinity-48s and Infinity-80 Xpertise software version 6.8 and Higher | GeneXpert Dx software version 4.3 and higher, GeneXpert Infinity-48 Xpertise 4.3 and higher, GeneXpert Infinity-48s and Infinity-80 Xpertise software version 6.0 and higher  |
|  Assay Definition File | Rules-based algorithms incorporating delta Ct values between targets within a valid Ct range and algorithms based on the Ct value for the targets falling within a valid Ct range | Algorithms based on the Ct value for the targets falling within a valid Ct range  |

# 5. A Design Control Activities Summary which includes:

a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis.

A formal risk assessment was conducted for the Xpert MRSA/SA Blood Culture test and the GeneXpert Instrument Systems. This assessment included safety Failure Mode Effects and Analyses (sFMEA) and Device Hazard Analysis (DHA) with the updated ADF and software modifications. False negative results i.e., true MRSA not identified as MRSA, were considered the worst case hazard effects to the patient. It was concluded that based on the Risk Management Reviews, discussions, sFMEA and the mitigation

K190771 - Page 4 of 6

{4}

actions taken, the risks identified are acceptable and Xpert MRSA/SA Blood Culture is safe and effective for its intended use.

b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied.

To confirm assay performance was not negatively impacted by the incorporation of the new software algorithms (and corresponding changes to the ADF) for determining SA positive, MRSA positive, SA negative and MRSA negative test results, Cepheid performed the following re-analysis of original raw data:

- LoD
- No new test runs were performed. Data acquired for MRSA and MSSA during the original LoD study were retrieved and reanalyzed using the updated ADF.
- Testing showed that the LoD estimates were not affected.

- Inclusivity
- No new test runs were performed. Data acquired during the original Inclusivity study were retrieved and reanalyzed using the updated ADF.
- Testing showed that all isolates were correctly reported, and the acceptance criteria were met.

- Exclusivity
- No new test runs were performed. Data acquired during the original Exclusivity study were retrieved and reanalyzed using the updated ADF.
- Testing showed that none of the non-Staphylococcus aureus isolates were detected and the acceptance criteria were met.

- Potentially Interfering Substances
- No new test runs were performed. Data acquired during the original Potentially Interfering Substances study were retrieved and reanalyzed using the updated ADF.
- Testing showed no substantial changes to the Ct values from the original report except for two tests with triglycerides and bilirubin that changed from correct test results (MRSA NEGATIVE; SA POSITIVE) to incorrect test results (MRSA POSITIVE; SA POSITIVE) attributed to a contaminant introduced during the original testing as mecA contamination was also observed in a MSSA negative control.

- Precision
- No new test runs were performed. Data acquired during the original Precision study were retrieved and reanalyzed using the updated ADF.
- Testing showed 95.2% of the results were identical. The majority of the exceptions were high negative samples that changed from MRSA NEGATIVE to MRSA POSITIVE due to the original ADF requiring three targets to be positive MRSA call, and the updated ADF requiring two of the three targets to be positive for a positive MRSA call.

- Reproducibility
- No new test results were performed. Data acquired during the original Reproducibility study were retrieved and reanalyzed using the updated ADF.
- Testing showed 97.5% of the results were identical. The majority of the exceptions were high negative samples that changed from MRSA NEGATIVE to MRSA POSITIVE due to the original ADF requiring three

K190771 - Page 5 of 6

{5}

targets to be positive MRSA call, and the updated ADF requiring two of the three targets to be positive for a positive MRSA call.

- Clinical Performance
- No new tests were performed. Data acquired during the original Clinical Performance study were retrieved and reanalyzed using the updated ADF.
- Testing showed similar performance and acceptance criteria were met. See Table 1 below for comparison between the original and updated analyses.

Table 1. Xpert MRSA/SA Blood Culture Performance Comparison

|  Target | Original |   | Updated  |   |
| --- | --- | --- | --- | --- |
|   |  PPA (n/N, 95% CI) | NPA (n/N, 95% CI) | PPA (n/N, 95% CI) | NPA (n/N, 95% CI)  |
|  MRSA | 98.1% (103/105, 93.3-99.8) | 99.6% (684/687, 98.7-99.9) | 99.0% (104/105, 94.8-100.0) | 99.0% (680/687, 97.9-99.5)  |
|  SA | 99.6% (235/236, 97.7-99.9) | 99.5% (553/556, 98.4-99.9) | 100.0% (236/236, 98.4-100.0) | 99.3% (552/556, 98.2-99.8)  |

The firm provided a summary of the results from the verification and validation studies. The results demonstrated that requirements were met and any test failures have been reviewed and determined to be due to software defects that can be prevented or corrected if they are encountered.

The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.

K190771 - Page 6 of 6

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NQX/K190771](https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NQX/K190771)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
