← Product Code [NGZ](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NGZ) · K151923

# VITEK 2 AST-Yeast Micafungin (0.06-8 mcg/mL) (K151923)

_bioMerieux, Inc. · NGZ · Mar 25, 2016 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NGZ/K151923

## Device Facts

- **Applicant:** bioMerieux, Inc.
- **Product Code:** [NGZ](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NGZ.md)
- **Decision Date:** Mar 25, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1640
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** AI/ML

## Intended Use

VITEK® 2 Yeast Micafungin is designed for antifungal susceptibility testing of Candida species. VITEK® 2 Yeast Micafungin is a quantitative test intended for use with the VITEK® 2 COMPACT Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Yeast Micafungin has been shown to be active against most isolates of the microorganisms listed below, according to the FDA label for this antifungal. The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp., S. pneumoniae, and clinically significant yeast.

## Device Story

VITEK® 2 AST-YS Micafungin is a miniaturized broth microdilution test card for determining minimum inhibitory concentrations (MIC) of micafungin against Candida species. Input: isolated yeast colonies diluted in 0.45-0.50% saline. Operation: card rehydrated with sample; VITEK® 2 system automatically fills/seals/incubates (VITEK® 2 COMPACT requires manual fill/seal); system monitors well growth via optical reader for up to 18 hours. Output: quantitative MIC value and interpretive category result. Used in clinical microbiology laboratories; results assist clinicians in selecting appropriate antifungal therapy. Benefits: automated, standardized susceptibility testing compared to manual methods.

## Clinical Evidence

Performance evaluated using 727 clinical and 97 challenge isolates across four sites. Comparison against CLSI broth microdilution reference method. Combined essential agreement (EA) was 98.9% and category agreement (CA) was 96.6% for the automatic dilution method. Minor discrepancies were observed; very major errors were mitigated by labeling limitations. MIC trending was identified: C. albicans/C. glabrata tended to read higher, while C. krusei/C. parapsilosis tended to read lower than reference methods, addressed via labeling footnotes.

## Technological Characteristics

Miniaturized broth microdilution card; contains antimicrobial medium in concentrations equivalent to standard mcg/ml. Sensing: optical growth monitoring. Connectivity: VITEK® 2 / VITEK® 2 COMPACT instrument platforms. Software: automated incubation and analysis. Sterilization: not specified.

## Regulatory Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Predicate Devices

- VITEK® 2 AST-YST Flucytosine ([K133952](/device/K133952.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 25, 2016

BIOMERIEUX, INC. CHERECE JONES STAFF REGULATORY AFFAIRS SPECILAIST 595 ANGLUM ROAD HAZELWOOD MO 63042

Re: K151923

Trade/Device Name: Vitek 2 AST-Yeast Micafungin (0.06-8 ug/ml) Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Susceptibility Test Powder Regulatory Class: II Product Code: NGZ, LRG, LTT, LTW Dated: March 7, 2016 Received: March 8, 2016

Dear Ms. Jones:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Ribhi Shawar -S

For Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K151923

Device Name VITEK® 2 AST-Yeast Micafungin (0.06 - 8 ug/mL)

#### Indications for Use (Describe)

VITEK® 2 Yeast Micafungin is designed for antifungal susceptibility testing of Candida species. VITEK® 2 Yeast Micafungin is a quantitative test intended for use with the VITEK® 2 COMPACT Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Yeast Micafungin has been shown to be active against most isolates of the microorganisms listed below, according to the FDA label for this antifungal.

The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp., S. pneumoniae, and clinically significant yeast.

| Type of Use (Select one or both, as applicable)                                                                                                               |                                                                                                                                               |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif; font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)  </span> | <span style="font-family: Arial, sans-serif; font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)  </span> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) SUMMARY

## VITEK® 2 AST AST-YS Micafungin

#### A. 510(k) Submission Information:

| Submitter's Name:    | bioMérieux, Inc.                                                                |
|----------------------|---------------------------------------------------------------------------------|
| Address:             | 595 Anglum Road<br>Hazelwood, MO 63042                                          |
| Contact Person:      | Cherece Jones<br>Staff Regulatory Affairs Specialist                            |
| Phone Number:        | 314-731-8684                                                                    |
| Fax Number:          | 314-731-8689                                                                    |
| Date of Preparation: | February 5, 2016                                                                |
| Device Name:         |                                                                                 |
| Formal/Trade Name:   | VITEK® 2 AST-Yeast Micafungin (0.06 - 8<br>µg/mL)                               |
| Classification Name: | 21 CFR 866.1640<br>Antimicrobial susceptibility test powder<br>Product Code NGZ |
| Common Name:         | VITEK® 2 AST-YS Micafungin                                                      |
| Predicate Device:    | VITEK® 2 AST-YS Flucytosine (K133952)                                           |

#### D. 510(k) Summary:

B.

C.

VITEK® 2 Yeast Micafungin is designed for antifungal susceptibility testing of Candida species. VITEK® 2 Yeast Micafungin is a quantitative test intended for use with the VITEK® 2 and VITEK® 2 COMPACT Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Yeast Micafungin has been shown to be active against most isolates of the microorganisms listed below, according to the FDA label for this antifungal.

Active in vitro and clinical infections: Candida albicans Candida glabrata Candida guilliermondii Candida krusei

{4}------------------------------------------------

### Candida parapsilosis Candida tropicalis

The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp., S. pneumoniae, and clinically significant yeast.

The antimicrobial presented in VITEK® 2 AST-YS Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/ml. The VITEK® 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.

The isolate to be tested is diluted to a standardized concentration in 0.45 - 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK® 2 automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 COMPACT has a manual filling and sealing operation. The VITEK® 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.

VITEK® 2 AST-YS Micafungin demonstrated substantially equivalent performance when compared with the CLSI broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, Issued August 28, 2009.

This 510(k) presents data in support of VITEK® 2 AST-YS Micafungin. An external evaluation was conducted with fresh and stock clinical isolates, as well as a set of challenge strains. The external evaluations were designed to confirm the acceptability of VITEK® 2 AST-YS Micafungin by comparing its performance with the CLSI broth microdilution reference method incubated at 24 hrs. The data is representative of performance on both the VITEK®2 and VITEK® 2 COMPACT instrument platforms. VITEK® 2 AST-YS Micafungin demonstrated acceptable performance of 98.9% for overall Essential Agreement and 96.6% for overall Category Agreement for Candida species. Reproducibility and Quality Control demonstrated acceptable results.

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NGZ/K151923](https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/NGZ/K151923)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).