CLEARVIEW EXACT PBP2A TEST, MODEL 891-000

K091766 · Binax, Inc. · MYI · Feb 24, 2010 · Microbiology

Device Facts

Record IDK091766
Device NameCLEARVIEW EXACT PBP2A TEST, MODEL 891-000
ApplicantBinax, Inc.
Product CodeMYI · Microbiology
Decision DateFeb 24, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K091766 · Feb 24, 2010CLEARVIEW EXACT PBP2A TEST, MODEL 891-000Binax, Inc.Retrospective clinical isolates from hospital laboratory repositoriesRetrospective S. aureus isolates were tested alongside prospective isolates to evaluate the clinical performance (sensitivity and specificity) of the Clearview® Exact PBP2a Test compared to the 30 μg cefoxitin disk diffusion method.Retrospective isolates; Clinical performance; Hospital laboratory repositories

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Clinical evaluation of Clearview® Exact PBP2a Test; Multi-site clinical study using prospective and retrospective isolatesStaphylococcus aureus isolates; Sample Size: 516 total isolates (including retrospective stock isolates); Number of Sites: 330 μg cefoxitin disk diffusionSensitivity and specificity for detection of PBP2a

Indications for Use

The Clearview® Exact PBP2a Test is a qualitative, in vitro immunochromatographic assay for the detection of penicillin-binding protein 2a (PBP2a) in isolates identified as Staphylococcus aureus, as an aid in detecting methicillin-resistant Staphylococcus aureus (MRSA). The Clearview® Exact PBP2a Test is not intended to diagnose MRSA nor to quide or monitor treatment for MRSA infections.

Device Story

Rapid immunochromatographic membrane assay; detects PBP2a protein directly from bacterial isolates. Input: bacterial isolates sampled from culture plates. Process: isolates eluted into Reagent 1; Reagent 2 added; dipstick inserted. Principle: monoclonal antibodies immobilized on nitrocellulose membrane bind PBP2a; visual result read at 5 minutes. Used in clinical laboratories; operated by laboratory technicians. Output: visual qualitative result. Assists healthcare providers in identifying MRSA; does not diagnose or guide treatment. Benefits: rapid identification of PBP2a, a marker for methicillin resistance.

Clinical Evidence

Multicenter study (2009) with 457 S. aureus samples. Compared to 30 ug cefoxitin disk diffusion (CLSI standards). Sensitivity ranged 98.1-99.5% and specificity 98.8% across three plate types (Tryptic Soy Agar, Columbia Agar, Mueller Hinton). Reproducibility study at 3 sites showed 97.3% agreement.

Technological Characteristics

Immunochromatographic membrane assay. Components: nitrocellulose membrane, sample pad, blue conjugate pad, absorption pad. Detection: monoclonal antibodies against PBP2a. Form factor: dipstick. Reagents: Reagent 1 and Reagent 2 for sample elution. No electronic components or software.

Indications for Use

Indicated for qualitative detection of PBP2a protein in Staphylococcus aureus isolates to aid in MRSA detection. Not for diagnosing MRSA or guiding/monitoring treatment.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Predicate Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K091766 B. Purpose for Submission: To determine substantial equivalence for the Clearview® Exact PBP2a Test, Model 891-000 C. Measurand: Penicillin-binding protein 2a (PBP2a) in isolates of methicillin-resistant Staphylococcus aureus (MRSA) D. Type of Test: The Clearview® Exact PBP2a test is a rapid immunochromatographic membrane assay that uses monoclonal antibodies to detect the PBP2a protein directly from Staphylococcus aureus isolates. E. Applicant: Binax, Inc F. Proprietary and Established Names: Clearview® Exact PBP2a Test. Model 891-000 G. Regulatory Information: 1. Regulation section: CFR 866.1640 2. Classification: II 1 {1} 3. Product code: MYI - System, Test, Genotypic Detection, Resistant Markers, Staphylococcus colonies 4. Panel: 83, Microbiology H. Intended Use: 1. Intended use: The Clearview® Exact PBP2a Test is a qualitative, in vitro immunochromatographic assay for the detection of penicillin-binding protein 2a (PBP2a) in isolates identified as Staphylococcus aureus, as an aid in detecting methicillin-resistant Staphylococcus aureus (MRSA). The Clearview® Exact PBP2a Test is not intended to diagnose MRSA nor to guide or monitor treatment for MRSA infections. 2. Indications for use: The Clearview® Exact PBP2a Test is a qualitative, in vitro immunochromatographic assay for the detection of penicillin-binding protein 2a (PBP2a) in isolates identified as Staphylococcus aureus, as an aid in detecting methicillin-resistant Staphylococcus aureus (MRSA). The Clearview® Exact PBP2a Test is not intended to diagnose MRSA nor to guide or monitor treatment for MRSA infections. 3. Special conditions for use statement(s) Prescription use only 4. Special instrument requirements: Not applicable I. Device Description: The Clearview® Exact PBP2a Test is a rapid immunochromatographic membrane assay that uses monoclonal antibodies to detect the PBP2a protein directly from bacterial isolates. These antibodies and a control antibody are immobilized onto a nitrocellulose membrane as two distinct lines and combined with a sample pad, a blue conjugate pad, and an absorption pad to form a test strip. 2 {2} Isolates are sampled directly from the culture plate and eluted into an assay tube containing Reagent 1. Reagent 2 is then added and the dipstick is placed in the assay tube. Results are read visually at 5 minutes. ## Materials Provided in Clearview® Exact PBP2a Test Kit - Test Strips (x 25): A dipstick covered with a clear, plastic overlay with directional arrows to indicate testing position - Reagent 1 (x1): A clear blue alkaline solution (irritant) – 6 mL - Reagent 2 (x1): A clear, slightly acidic solution containing sodium azide, buffer and proteins – 2.5 mL - Assay Tubes - Test Rack ## Materials required but not provided - Clock, timer or stopwatch, applicator sticks (for S. aureus isolates) - 1 μl inoculating loops (for CoNS isolates) - Vortex mixer - External positive and negative controls - The recommended controls are listed below. - Control Strains for S. aureus isolates: - Positive control: Staphylococcus aureus ATCC # BAA44 - Negative control: Staphylococcus aureus ATCC # 25923 ## J. Substantial Equivalence Information: 1. Predicate device name(s): Mueller Hinton Agar w/4% NaCl w/antibiotics (Remel) a. PBP2’ Oxoid PBP2 Latex Agglutination Test (Oxoid) b. Remel Mueller Hinton Agar with 4% NaCl and 6 μg/mL Oxacillin 2. Predicate 510(k) number(s): a. K011710 b. K850291 {3} 3. Comparison with predicate: | Similarities | | | | | --- | --- | --- | --- | | Item | Device | Predicate (K011710) | Predicate (K850291) | | Intended Use | Detects methicillin resistance from Staphylococcus aureus isolates | Same | Same | | Mode of Detection | Detects PBP2a | Detects PBP2a | Detects (resistance to penicillinase-resistant penicillin (i.e. oxacillin) – (May or may not be related to PBP2a) | | Specimen Type | Identifies isolates as Staphylococcus species isolates | Same | Identifies isolates as Staphylococcus species isolates (i.e. S. aureus) | | Differences | | | | | --- | --- | --- | --- | | Item | Device | Predicate (K011710) | Predicate (K850291) | | Technology | Immuno-chromatographic membrane assay | Latex agglutination | Agar dilution | | Detection mechanism | Dipstick | Sensitized latex | Agar with antibiotic (i.e. oxacillin) | | Reading time | Visual read at 5 minutes after elution | Visual read within 3 minutes after extraction | Read after 24 hrs incubation | # K. Standard/Guidance Document Referenced (if applicable): CLSI 2007. Performance standards for antimicrobial susceptibility testing. CLSI approved standard M100-S17. Clinical and Laboratory Standards # L. Test Principle: The Clearview® Exact PBP2a Test is a qualitative, in vitro immunochromatographic assay. The procedure involves adding bacterial colonies to Reagent 1 to lyse the bacterial cell walls and expose the PBP2a protein. Subsequently, Reagent 2 is added to neutralize the sample. Once the test strips is placed in the extraction solution, the extraction solution is absorbed by the sample pad through to the conjugate pad by capillary action. The conjugate pad contains dried blue latex particles sensitized with mouse anti-PBP2a monoclonal antibody. When the sample reaches the conjugate pad, the sensitized latex is dissolved into the sample where an immune complex is formed if PBP2a protein is present in the sample. This mixture then flows into the nitrocellulose membrane. The nitrocellulose membrane is coated with two lines: one of which is mouse anti-PBP2a (the test line) and the other is anti-mouse immunoglobulin (the control line). As the sample moves over the nitrocellulose {4} membrane, a reaction between the immune complex and the coated anti-PBP2a antibodies creates a blue test line in positive samples. Excess conjugate is trapped on the control line. This indicates that the test results are valid. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: Reproducibility study was done at three sites. Testing was done at three sites, using 18-24 hrs isolates that have been sub-cultured, by two operators in duplicates for five days. The reproducibility panel consists of six MRSA, and four MSSA; 10 x (2) duplicates x 2 operators x 5 days x 3 sites = 600 Overall reproducibility was >95%. #### b. Linearity/assay reportable range: Not applicable #### c. Traceability, Stability, Expected values (controls, calibrators, or methods): ##### Storage and Stability Product Stability Studies were performed to generate real time and accelerated stability data to support the expiration date of each lot under stated conditions, (i.e., Clearview® Exact PBP2a Test kits should be stored at room temperature or under refrigeration (2-30 °C). Reagent 1 and Reagent 2 must be stored refrigerated (2-8 °C)). Stability conditions were set up as follows: 1. Accelerated - Clearview® Exact PBP2a devices and Clearview® Exact PBP2a Reagent 1* at 45°C & Clearview® Exact PBP2a Reagent 2 at 22°C 2. Real-time, room temperature - Clearview® Exact PBP2a devices and Clearview® Exact PBP2a Reagent 1* at 30°C & Clearview® Exact PBP2a Reagent 2 at 4°C 3. Real-time, refrigerated - Clearview® Exact PBP2a devices and Clearview® Exact PBP2a Reagent 1 & Clearview® Exact PBP2a Reagent 2 at 4°C Three device lots were used to test all test samples and controls using the Control Assay Procedure in the package insert. (The same lot of devices, liquid controls, and kit reagents are being used for the duration of the 5 {5} stability study). All devices were interpreted at 5 minutes and negative devices were interpreted at 5 and 10 minutes. All samples were run in triplicate using the same plate. Results are outlined in the table below: Stability Testing Summary | Product Stability (with Desiccant) | | | | | --- | --- | --- | --- | | Lot | 4°C | 30 °C | 45°C | | Device, Reagents 1 and 2 – Transfer lot 3 | 5 months | 5 months | 20 weeks | | Device, Reagents 1 and 2 – Transfer lot 4 | 4 months | 4 months | 17 weeks | | Product Stability (without Desiccant) | | | | | Lot | 4°C | 30 °C | 45°C | | Device, Reagents 1 and 2 – Transfer lot 3 | 12 weeks | 12 weeks | 20 weeks | For Transfer Lot 3, the device and reagents remained stable after 20 weeks (with and without desiccant) at 45°C and 5 months (with desiccant) and 12 weeks (without desiccant) of real-time storage at 4°C and 30 °C. For Transfer Lot 4, the device and reagents remain stable after 17 weeks (with desiccant) and 10 weeks (without desiccant) at 45°C and 4 months (with desiccant) and 15 weeks (without desiccant) of real-time storage at 4°C and 30 °C. Stability continued on transfer lots 3 and 4. Stability testing of each individual component will continue until a 2-year expiration date is achieved. d. Detection Limit: A study was conducted using Staphylococcus aureus (ATCC BAA44) and the detection limit was determined to be 1.4 x 10⁷ CFU/mL e. Analytical specificity: A total of 162 strains of methicillin-resistant Staphylococcus aureus (MRSA), 112 strains of methicillin-sensitive Staphylococcus aureus (MSSA), and other non-Staphylococcal strains were tested in the Clearview® Exact PBP2a Test with expected results f. Assay cut-off: Not applicable 2. Comparison studies: A clinical study was conducted at three geographically diverse hospital laboratories (two U.S. sites and one non-U.S. site), using prospective and {6} retrospective isolates identified as Staphylococcus aureus. A total of 516 S. aureus fresh and stock isolates were tested from Tryptic Soy Agar with 5% sheep blood, Columbia agar with 5% sheep blood and Mueller-Hinton Agar induced with a 1 μg oxacillin disk. Results were compared to those of 30 μg cefoxitin disk diffusion, according to Clinical and Laboratory Standards Institute (CLSI) interpretation. Of these, 30% (225/733) were stock isolates. Retrospective S. aureus isolates were also enrolled into the study. Summary of Performance Characteristics: | Plate Type | Sensitivity | 95% C.I. | Specificity | 95% C.I. | | --- | --- | --- | --- | --- | | Tryptic Soy Agar with 5% sheep blood | 98.1% (206/210) | (95.2-99.3%) | 98.8% (244/247) | (96.5-99.6%) | | Columbia Agar with 5% sheep blood | 99.0% (208/210) | (96.6-99.7%) | 98.8% (244/247) | (96.5-99.6%) | | Mueller Hinton with 1 μg oxacillin induction | 99.5% (209/210) | (97.4-99.9%) | 98.8% (244/247) | (96.5-99.6%) | a. Method comparison with predicate device: Cefoxitin disk at concentration of 30 μg is a surrogate for oxacillin resistance and is used in the clinical studies. The CLSI recommendations are: | Interpretive Criteria (zone diameter in mm) for 30 μg Cefoxitin Disk Diffusion Test | | | | | --- | --- | --- | --- | | | Oxacillin Susceptible | Intermediate* | Oxacillin Resistant | | S. aureus and S. lugdunensis | ≥22 mm | N/A | ≤21 mm | b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: {7} Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: In the external clinical evaluation of the Clearview® Exact PBP2a Test, the overall expected rate of PBP2a (MRSA) among S. aureus isolates was 45.9% (210/457), and among the three site populations the expected positive rate ranged from 34.6% to 49.7%. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The information submitted in this premarket notification is complete and supports a substantial equivalence decision 8
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...