MicroScan Dried Gram-Negative MIC/Combo Panels with Aztreonam (AZT) (0.5-64 µg/mL)

K250084 · Beckman Coulter, Inc. · LTT · Jul 18, 2025 · Microbiology

Device Facts

Record IDK250084
Device NameMicroScan Dried Gram-Negative MIC/Combo Panels with Aztreonam (AZT) (0.5-64 µg/mL)
ApplicantBeckman Coulter, Inc.
Product CodeLTT · Microbiology
Decision DateJul 18, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2
AttributesPCCP

Intended Use

For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.

Device Story

MicroScan Dried Gram-Negative MIC/Combo Panels are single-use, miniaturized broth dilution susceptibility tests. Panels contain dehydrated antimicrobial agents (Aztreonam 0.5-64 µg/mL) in broth. Laboratory professionals inoculate panels with standardized suspensions of isolated colonies from solid media. Panels are rehydrated with water and incubated for 16-20 hours at 35°C ± 1°C in a non-CO2 incubator. The minimum inhibitory concentration (MIC) is determined by identifying the lowest antimicrobial concentration inhibiting growth, either visually or via MicroScan instrumentation (WalkAway or autoSCAN-4). Results assist clinicians in selecting appropriate antimicrobial therapy for gram-negative bacterial infections.

Clinical Evidence

Performance evaluated using 622 clinical and 148 challenge isolates. After excluding non-indicated species, 540 clinical and 108 challenge isolates were analyzed. Overall essential agreement (EA) and category agreement (CA) were generally >90%. Specific limitations were established for organism/method combinations showing unacceptable major or very major error rates (e.g., Citrobacter koseri, Proteus mirabilis, E. coli). No clinical data; performance based on bench-top comparative testing against reference broth microdilution.

Technological Characteristics

Miniaturized broth microdilution panel; dehydrated antimicrobial agent (Aztreonam). Incubation: 35°C ± 1°C, aerobic, 16-20 hours. Reading: Visual or automated (WalkAway/autoSCAN-4). Single-use. Complies with Class II Special Controls Guidance for AST Systems.

Indications for Use

Indicated for quantitative and qualitative antimicrobial susceptibility testing of rapidly growing aerobic and facultative anaerobic gram-negative bacilli (Enterobacterales and Pseudomonas aeruginosa) from solid media. Contraindicated for Enterobacter cloacae complex, Serratia marcescens, Serratia liquefaciens complex, Proteus vulgaris, Providencia stuartii, and Providencia rettgeri.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION July 18, 2025 Beckman Coulter, Inc. Elaine Duncan Staff Regulatory Affairs 1584 Enterprise Blvd. West Sacramento, California 95691 Re: K250084 Trade/Device Name: MicroScan Dried Gram-Negative MIC/Combo Panels with Aztreonam (AZT) (0.5-64 µg/mL) Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Susceptibility Test Powder Regulatory Class: Class II Product Code: LTT, JWY, LRG, LTW Dated: June 20, 2025 Received: June 20, 2025 Dear Elaine Duncan: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, {1} K250084 - Elaine Duncan Page 2 or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part {2} K250084 - Elaine Duncan Page 3 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Ribhi Shawar -S Ribhi Shawar, Ph.D. (ABMM) Branch Chief, General Bacteriology and Antimicrobial Susceptibility Branch Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K250084 | | | Device Name MicroScan Dried Gram-Negative MIC/Combo Panels with Aztreonam (AZT) (0.5-64 μg/mL) | | | Indications for Use (Describe) The MicroScan Dried Gram-Negative MIC/Combo Panel is used to determine quantitative and qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-negative bacilli. After inoculation, panels are incubated for 16-20 hours at 35°C ± 1°C in a non-CO2 incubator, and read either visually or with MicroScan instrumentation, according to the Package Insert. This particular submission is for the addition of the antimicrobial aztreonam at concentrations of 0.5-64 μg/mL to the test panel. Testing is indicated for Enterobacterales and Pseudomonas aeruginosa, as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC) webpage. The MicroScan Dried Gram-Negative MIC/Combo Panels with Aztreonam (AZT) (0.5-64 μg/mL) has demonstrated acceptable performance with the following organisms: Enterobacterales (Citrobacter freundii complex, Citrobacter koseri, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus mirabilis, Morganella morganii, Yersinia enterocolitica) Pseudomonas aeruginosa | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} BECKMAN COULTER # 510(k) Summary ## 510(k) Submission Information: Submitter's Name: Beckman Coulter Contact Person: Elaine Duncan, Staff Regulatory Affairs Address: 1584 Enterprise Blvd. West Sacramento, CA 95691 Phone: 916-318-0652 Date prepared: July 11, 2025 Product Name: Microdilution Minimum Inhibitory Concentration (MIC) Panels Trade Name: MicroScan Dried Gram-Negative MIC/Combo Panels with Aztreonam (AZT) (0.5 – 64 µg/mL) Classification: 21 CFR 866.1640 Antimicrobial Susceptibility Test Powder; Class II Product Code: LTT, JWY, LRG, LTW 510(k) Notification: Updated antimicrobial agent - Aztreonam Predicate device: MicroScan Dried Gram-Negative MIC/Combo Panels Ceftazidime – (0.5 – 64 µg/mL) - (K202343) ## 510(k) Summary: **Device Description:** MicroScan Dried Gram-Negative MIC/Combo Panels are designed for use in determining quantitative and qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-negative bacilli. The principle of MicroScan panels with antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in broth and dehydrated. Various antimicrobial agents are diluted in broth to concentrations bridging the range of clinical interest. Panels are rehydrated with water after inoculation with a standardized suspension of the organism. After incubation in a non-CO₂ incubator for 16-20 hours, the minimum inhibitory concentration (MIC) for the test organism is read by determining the lowest antimicrobial concentration showing inhibition of growth. This product is single-use and intended for laboratory professional use. **Intended Use:** For Use with MicroScan Dried Gram-Negative MIC/Combo and Dried Gram-Negative Breakpoint Combo panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli. **Indications for Use:** The MicroScan Dried Gram-Negative MIC/Combo Panel is used to determine quantitative and qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultatively anaerobic gram-negative bacilli. After inoculation, panels are incubated for 16-20 hours at 35°C ± 1°C in a non-CO₂ incubator, and read either visually or with MicroScan instrumentation, according to the Package Insert. This particular submission is for the addition of the antimicrobial aztreonam at concentrations of 0.5 - 64 µg/mL to the test panel. Testing is indicated for Enterobacterales and Pseudomonas aeruginosa, as recognized by the FDA Susceptibility Test Interpretive Criteria (STIC) webpage. The MicroScan Dried Gram-Negative MIC/Combo Panels with Aztreonam (Azt) (0.5 - 64 µg/mL) has demonstrated acceptable performance with the following organisms: Enterobacterales (Citrobacter freundii complex, Citrobacter koseri, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus mirabilis, Morganella morganii, Yersinia enterocolitica) Pseudomonas aeruginosa 510(k) Summary {5} BECKMAN COULTER Substantial Equivalence Information: The similarities and differences of the MicroScan Dried Gram Negative MIC/Combo Panels with Aztreonam (Azt) (0.5 – 64 µg/mL) compared to the predicate device, MicroScan Dried Gram-Positive MIC/Combo Panels with Ceftazidime – (K202343), are described in Table 1 below. Table 1: Substantial Equivalence | Similarities | | | | --- | --- | --- | | Item | Proposed MicroScan Dried Gram-Negative MIC/Combo Panels – Aztreonam | Predicate MicroScan Dried Gram-Negative MIC/Combo Panels – Ceftazidime (K202343) | | Intended Use | Determination of susceptibility to Aztreonam with gram-negative bacteria | Determination of susceptibility to Ceftazidime with gram-negative bacteria | | Technology | Overnight Microdilution MIC Susceptibility Tests | Same | | Specimen | Isolated colonies from cultures | Same | | Incubation Temperature | 35° C ± 1 ° C | Same | | Incubation Atmosphere | Aerobic | Same | | Incubation Time | 16 – 20 hours | Same | | Reading Method | Automated or Manual | Same | | Differences | | | | Item | Proposed MicroScan Dried Gram-Negative MIC/Combo Panels – Aztreonam | Predicate MicroScan Dried Gram-Negative MIC/Combo Panels – Ceftazidime (K202343) | | Antimicrobial Agent | Dried Aztreonam | Dried Ceftazidime | | | 0.5 – 64 µg/mL | 0.5 – 64 µg/mL | Performance and Conclusion: The proposed MicroScan Dried Gram-Negative MIC/Combo Panel demonstrated substantially equivalent performance when compared with a CLSI frozen Reference Panel, as defined in the FDA document "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", dated August 28, 2009. The Premarket Notification (510[k]) presents data in support of the MicroScan Dried Gram-Negative MIC/Combo Panel with Aztreonam. The external evaluations were conducted with contemporary and stock Efficacy isolates and stock Challenge strains. The external evaluations were designed to confirm the acceptability of the proposed Dried Gram-Negative Panel by comparing its performance with a CLSI frozen Reference panel. The Dried Gram-Negative Panel inoculated with Prompt and read on the WalkAway instrument demonstrated acceptable performance with an Enterobacterales Essential Agreement (EA) of 91.0% and Categorical Agreement (CA) of 93.1%, and Pseudomonas aeruginosa Essential Agreement (EA) of 91.2% and Categorical Agreement (CA) of 86.0%* for Aztreonam when compared with the frozen Reference panel. *Essential agreement of evaluable isolates 90.3% and most of the categorical discrepancies were minor errors. 510(k) Summary {6} BECKMAN COULTER Inoculum and instrument reproducibility testing demonstrated acceptable reproducibility and precision with Aztreonam, regardless of which inoculum method (i.e., Turbidity or Prompt), or instrument (autoSCAN-4 instrument or WalkAway system) was used. Quality Control testing demonstrated acceptable results for Aztreonam. © 2025 Beckman Coulter, Inc. All rights reserved. All other trademarks are the property of their respective owners. May be covered by one or more patent. See www.beckmancoulter.com/patents. Beckman Coulter, the stylized logo, and the Beckman Coulter product and service marks mentioned herein are trademarks or registered trademarks of Beckman Coulter, Inc. in the United States and other countries. 510(k) Summary Page 3 of 3
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...