← Product Code [LON](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON) · K232756

# Colibrí (K232756)

_Copan Wasp Srl · LON · Dec 27, 2023 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON/K232756

## Device Facts

- **Applicant:** Copan Wasp Srl
- **Product Code:** [LON](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON.md)
- **Decision Date:** Dec 27, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1645
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

The Colibrí is an automated in vitro diagnostic specimen preparation system for use with WASPLab to prepare MALDI-TOF targets for the bioMérieux VITEK MS systems or Bruker MALDI Biotyper CA mass spectrometry systems for qualitative identification and microbial suspension for the bioMérieux VITEK 2 systems or Beckman Coulter MicroScan WalkAway Antimicrobial Susceptibility Testing (AST) systems for qualitative testing of isolated colonies of gram-negative and gram-positive bacterial species grown on solid culture media. The Colibrí is an automated pre-analytical processor that picks isolated colonies designated by the operator and uses a pipetting system to prepare MALDI-TOF MS (Matrix-Assisted Laser Desorption/lonization-Time of Flight Mass Spectrometry) target slides for bacterial identification and microbial suspension at known concentration for Antimicrobial Susceptibility Testing and purity assessment. The Colibrí software records the identity of each sample and its position on the target slide and communicates this information electronically to the MALDI-TOF MS analyzers. Bacterial suspensions for AST and purity plates are identified by barcode label. The Colibrí is intended for use by trained healthcare professionals in clinical laboratories in conjunction with other clinical and laboratory findings, including Gram staining, to aid in the diagnosis of bacterial infections. The Colibrí has not been validated for use in the identification or processing of yeast species, molds, Nocardia, or mycobacteria.

## Device Story

Colibrí is an automated pre-analytical processor integrated with WASPLab; it automates colony picking, MALDI-TOF target slide preparation, and microbial suspension preparation for AST. Input: isolated bacterial colonies from solid agar plates identified via WASPLab barcode and digital imaging. Operation: operator selects colonies on a GUI; instrument uses a pipetting system to pick colonies and transfer them to target slides (for ID) or primary/secondary tubes (for AST). Nephelometer on-board ensures correct microbial concentration for AST suspensions. Output: prepared MALDI-TOF targets and AST microbial suspensions. Used in clinical microbiology laboratories by trained personnel. Output affects clinical decision-making by providing standardized, automated sample preparation for downstream ID and AST analyzers, reducing manual variability and potential for human error. Benefits include increased laboratory workflow efficiency and standardized sample quality for diagnostic testing.

## Clinical Evidence

Bench testing only. Reproducibility study (n=2,619 MIC results) demonstrated combined best-case reproducibility ≥95% and worst-case ≥89%. Accuracy study compared Colibrí-prepared samples to manual preparation across 110 clinical isolates (Enterobacterales, Staphylococci, Streptococci, Enterococci, non-fermenters) using 29 antimicrobial agents. Results showed 100% Essential Agreement (1232/1232) and 98.4% Category Agreement (4187/4254) with no major or very major errors. Microbial suspension concentration accuracy was 98.5% (130/132) within expected ranges.

## Technological Characteristics

Automated robotic pipetting system; on-board nephelometer for turbidity measurement; integrated camera units for positioning. Connectivity: Electronic communication with MALDI-TOF MS analyzers. Software-controlled workflow. Calibration: Auto-calibration of nephelometer and cameras; daily nephelometer verification. Materials: Pipette tips, primary/secondary tubes, spreader. Sterilization: On-board incinerator for spreader.

## Regulatory Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

## Predicate Devices

- Colibrí ([K223245](/device/K223245.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

INSTRUMENT ONLY

## I Background Information:

A 510(k) Number

K232756

B Applicant

Copan WASP S.r.l.

C Proprietary and Established Names

Colibrí

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  LON | Class II | 21 CFR 866.1645 - Fully automated short-term incubation cycle antimicrobial susceptibility system | MI - Microbiology  |
|  QQV | Class II | 21 CFR 866.3378 - Clinical mass spectrometry microorganism identification and differentiation system | MI - Microbiology  |
|  QBN | Class II | 21 CFR 866.3378 - Clinical mass spectrometry microorganism identification and differentiation system | MI - Microbiology  |

## II Submission/Device Overview:

### A Purpose for Submission:

To obtain a substantial equivalence determination for the Colibrí for use with the Beckman Coulter MicroScan WalkAway AST system and update the package insert AST data summary tables to be consistent across both claimed AST devices (Beckman Coulter MicroScan WalkAway, subject of this submission, and bioMérieux VITEK 2, cleared for use with the Colibrí in K220546).

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

K232756 - Page 2 of 16

## B Type of Test:

Qualitative *in vitro* diagnostic device for identification of bacteria cultured from human specimens by automation of target preparation for mass spectrometry analysis and antimicrobial susceptibility test (AST) assessment of bacteria cultured from human specimens by automation of culture suspensions for AST analysis.

## III Intended Use/Indications for Use:

### A Intended Use(s):

See Indications for Use below.

### B Indication(s) for Use:

The Colibrí is an automated *in vitro* diagnostic specimen preparation system for use with WASPLab to prepare MALDI-TOF targets for the bioMérieux VITEK MS systems or Bruker MALDI Biotyper CA mass spectrometry systems for qualitative identification and microbial suspension for the bioMérieux VITEK 2 systems or Beckman Coulter MicroScan WalkAway Antimicrobial Susceptibility Testing (AST) systems for qualitative testing of isolated colonies of gram-negative and gram-positive bacterial species grown on solid culture media.

The Colibrí is an automated pre-analytical processor that picks isolated colonies designated by the operator and uses a pipetting system to prepare MALDI-TOF MS (Matrix-Assisted Laser Desorption/Ionization-Time of Flight Mass Spectrometry) target slides for bacterial identification and microbial suspension at known concentration for Antimicrobial Susceptibility Testing and purity assessment.

The Colibrí software records the identity of each sample and its position on the target slide and communicates this information electronically to the MALDI-TOF MS analyzers.

Bacterial suspensions for AST and purity plates are identified by barcode label.

The Colibrí is intended for use by trained healthcare professionals in clinical laboratories in conjunction with other clinical and laboratory findings, including Gram staining, to aid in the diagnosis of bacterial infections.

The Colibrí has not been validated for use in the identification or processing of yeast species, molds, Nocardia, or mycobacteria.

### C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

IVD - For In Vitro Diagnostic Use Only

Special Instruments for Use:

- Bruker MALDI Biotyper for Clinical Applications (MBT-CA)
- bioMerieux VITEK MS
- bioMerieux VITEK 2
- Beckman Coulter MicroScan WalkAway
- WASPLab

{2}

[Refer to the K193138 Decision Summary for Special Conditions for Use Statements that are only applicable to the sample preparation for bacterial identification and MALDI-TOF MS workflow.]

The following Special Conditions for Use Statements related to microbial suspension preparation for AST were modified from K220546 to include the MicroScan:

- Use of this device is permitted only in association with bioMérieux VITEK MS systems or Bruker MALDI Biotyper CA System for microbial identification and in association with bioMérieux VITEK 2 or Beckman Coulter MicroScan WalkAway systems for AST.
- The performance of Colibrí in conjunction with the VITEK 2 and MicroScan WalkAway systems was evaluated with Trypticase Soy Agar + 5% sheep blood (BD), Columbia Agar + 5% sheep blood (bioMerieux), MacConkey Agar (bioMerieux), Trypticase Soy Agar + 5% sheep blood / MacConkey Agar (BD). The use of other types of culture media has not been validated.
- Preparation of bacterial suspensions from unclaimed species from VITEK 2 or MicroScan WalkAway systems has not been evaluated.

The following Special Conditions for Use Statements related to microbial suspension preparation for AST were included in K220546 and are still applicable – the trade name was updated as needed:

- This product is intended for target slides preparation for identification and suspensions preparation for susceptibility testing from colonies grown on solid agar media plates. Do not use for identification from liquid cultures.
- For AST application, make sure that the selected colonies are morphologically similar.
- Results obtained using the Colibrí for sample preparation with the compatible analyzers should be used as an adjunct to clinical observations and other information available to the physician.
- The ability of the Colibrí to prepare samples for analysis with the compatible analyzers was evaluated using only the species listed in the Inclusivity and AST Challenge studies, described in the Performance Characteristics section of this Package Insert. The ability of the Colibrí to prepare samples of other species for mass spectrometry analysis and AST has not been evaluated.
- The Colibrí should be used only for preparation of target slides and microbial suspension from isolated colonies of Gram-negative and Gram-positive bacterial species grown on solid culture media. Colibrí is not intended for processing of yeasts, moulds, Nocardia or Mycobacteria.

## IV Device/System Characteristics:

### A Device Description:

The Colibrí is an instrument designed to be used as an accessory for downstream matrix assisted laser desorption/ionization time of flight mass spectrometry (MALDI-TOF MS) and antimicrobial susceptibility testing (AST) analyzers. The Colibrí is comprised of the Colibrí instrument (including on-board pipetting system, spreader, and nephelometer), software, primary tubes for microbial suspension, and daily verification kit which automatically prepares target

K232756 - Page 3 of 16

{3}

slides for ID and microbial suspensions and purity plates for AST from isolated colonies. The Colibrí is designed to be used in conjunction with the Copan WASPLab to automate the plate incubation, plate management, image acquisition, colony selection, target slide preparation (for ID), microbial suspension preparation (for AST), and purity plate preparation steps of the ID and/or AST sample preparation workflow. Refer to the K223245 Decision Summary for the device description and performance characteristics of the Colibrí when used with the WASPLab.

The Colibrí automates the preparation of MALDI target slides for the bioMérieux VITEK MS or the Bruker MALDI Biotyper CA systems that are used in clinical laboratories for identification (ID) of gram-negative and gram-positive bacterial species grown on plated media by MALDI-TOF MS. Refer to the K193138 Decision Summary for the device description and performance characteristics of the sample preparation for bacterial identification and MALDI-TOF MS workflow.

The Colibrí also automates the preparation of gram-negative and gram-positive bacterial suspensions at a known concentration from isolated colonies for the bioMérieux VITEK 2 and Beckman Coulter MicroScan WalkAway that are used in clinical laboratories for AST analyses. The Colibrí is also used for purity plate preparation for purity assessments. Refer to the K220546 Decision Summary for the device description and performance characteristics of the sample preparation for AST workflow with the bioMérieux VITEK 2. The workflow for both AST systems is similar and both rely on isolated colony samples, although there are differences in the Primary and Secondary Tubes. For use with the VITEK 2, the Primary Tube and Secondary Tube are filled with saline for all organisms. For use with the Beckman Coulter MicroScan WalkAway, the operator will need to input the correct Primary Tube (which will contain either saline or water) in accordance with the MicroScan WalkAway IFU. The Secondary Tube will be automatically filled with either water with PLURONIC or Mueller-Hinton Broth with 3% lysed horse blood to prepare the final suspension in accordance with the MicroScan WalkAway IFU. The performance characteristics of the sample preparation for AST from isolated colony workflow with the Beckman Coulter MicroScan WalkAway are described in this submission.

## B Instrument Description Information:

1. Instrument Name:
Colibrí

2. Specimen Identification:
Culture plates for processing are identified by the WASPLab by scanning a manually applied linear barcode on the side of each plate. The loading conveyor moves the plate inside the Imaging Module in the WASPLab, where the plate is checked in to the system through a barcode reader. The barcode is again scanned when the plate is manually transferred to the Colibrí. The barcode is used to orientate the plate and, together with the plate's geometric center, also used to define the Cartesian coordinates of each of the colonies that are designated for picking. The designated colonies are then picked for the assigned downstream activities.

K232756 - Page 4 of 16

{4}

# 3. Specimen Sampling and Handling:

Plates labeled with a barcode on the side are loaded onto a conveyor and loaded into the WASPLab. The barcode is scanned, and the plate is incubated within the WASPLab. After appropriate incubation, each plate is photographed, and the image is saved on the WASPLab server and is displayed to the operator on the WebApp interface. After selecting isolated colonies and assigning the downstream task, the user manually transfers the plate to the Colibri. Within the Colibri, the plates are prepared for ID (as described in the K193138 Decision Summary) or AST (as described in the K220546 Decision Summary).

# 4. Calibration:

Colibri requires four different calibrations: one on the on-board nephelometer and three on the cameras. No changes are made in set-up calibration, auto-calibration, and run-time calibration checks which are performed as described in the K193138 Decision Summary. No changes are made to daily nephelometer verification, which is performed as described in the K220546 Decision Summary.

The WASPLab does not require calibration.

# 5. Quality Control:

Quality Control for AST testing was performed in accordance with the manufacturer's instructions for the Beckman Coulter MicroScan. CLSI-recommended reference strains Escherichia coli ATCC 25922, Pseudomonas aeruginosa ATCC 27853, Staphylococcus aureus ATCC 29213, and Enterococcus faecalis ATCC 29212 were grown on trypticase soy agar  $+5\%$  sheep blood. Microbial suspensions were prepared using three separate Colibri instruments and loaded in the appropriate MicroScan panel. MIC values for each drug/organism combination were compared to the established ranges reported in the MicroScan panel labeling. Purity of all the suspensions was confirmed by purity plates prepared by the Colibri, with  $100\%$  of purity plates showing monomicrobial growth. Quality control testing was performed daily during analytical performance testing using the quality control bacterial strains in rotation, and all results obtained were in-range. Results are summarized in Table 1. The quality control testing is acceptable.

Quality control for bacterial identification is described in the K193138 Decision Summary. Quality control for AST with the bioMérieux VITEK 2 was conducted in a similar manner and is described in the K220546 Decision Summary. Additional quality control measures are not needed for the WASPLab.

Table 1. Quality Control Organisms and Antimicrobial Agents Tested

|  QC organism | Antimicrobial agent | QC range (CLSI M100) | On panel values* (panel IFU) | No. MIC within QC Range  |
| --- | --- | --- | --- | --- |
|  Escherichia coli ATCC 25922 | Amikacin | 0.5 - 4 | ≤4 - 8 | 36/36 (100%)  |
|   |  Ampicillin/Sulbactam | 2/1 - 8/4 | ≤4/2 | 36/36 (100%)  |
|   |  Aztreonam | 0.06 - 0.25 | ≤0.5 | 36/36 (100%)  |

K232756 - Page 5 of 16

{5}

K232756 - Page 6 of 16
|  QC organism | Antimicrobial agent | QC range (CLSI M100) | On panel values * (panel IFU) | No. MIC within QC Range  |
| --- | --- | --- | --- | --- |
|   | Ceftazidime | 0.06 - 0.5 | ≤1 | 36/36 (100%)  |
|   |  Cefotaxime | 0.03 - 0.12 | ≤4 | 36/36 (100%)  |
|   |  Ciprofloxacin | 0.004 - 0.016 | ≤0.12 | 36/36 (100%)  |
|   |  Cefepime | 0.016 - 0.12 | ≤1 | 36/36 (100%)  |
|   |  Gentamicin | 0.25 - 1 | ≤1 - 2 | 36/36 (100%)  |
|   |  Imipenem | 0.06 - 0.5 | ≤0.5 | 36/36 (100%)  |
|   |  Levofloxacin | 0.008 - 0.06 | ≤0.5 | 36/36 (100%)  |
|   |  Meropenem | 0.008 - 0.06 | ≤0.12 | 36/36 (100%)  |
|   |  Minocycline | 0.25 - 1 | ≤2 | 36/36 (100%)  |
|   |  Piperacillin/Tazobactam | 1/4 - 8/4 | ≤8/4 | 36/36 (100%)  |
|   |  Tetracycline | 0.5 - 2 | ≤2 | 36/36 (100%)  |
|   |  Ticarcillin/K Clavulanate | 4/2 - 16/2 | ≤16/2 | 36/36 (100%)  |
|   |  Tobramycin | 0.25 - 1 | ≤2 | 36/36 (100%)  |
|   |  Trimethoprim/Sulfamethoxazole | ≤0.5/9.5 | ≤1/19 | 36/36 (100%)  |
|  Staphylococcus aureus ATCC 29213 | Ciprofloxacin | 0.12 - 0.5 | ≤0.5 | 42/42 (100%)  |
|   |  Clindamycin | 0.06 - 0.25 | ≤0.25 | 42/42 (100%)  |
|   |  Daptomycin | 0.12 - 1 | ≤0.25 - 1 | 42/42 (100%)  |
|   |  Erythromycin | 0.25 - 1 | ≤0.25 - 1 | 42/42 (100%)  |
|   |  Levofloxacin | 0.06 - 0.5 | ≤1 | 42/42 (100%)  |
|   |  Linezolid | 1 - 4 | ≤1 - 4 | 42/42 (100%)  |
|   |  Minocycline | 0.06 - 0.5 | ≤1 | 42/42 (100%)  |
|   |  Moxifloxacin | 0.016 - 0.12 | ≤0.25 | 42/42 (100%)  |
|   |  Nitrofurantoin | 8 - 32 | ≤32 | 42/42 (100%)  |
|   |  Penicillin G | 0.25 - 2 | 0.25 - >8 | 42/42 (100%)  |
|   |  Tetracycline | 0.12 - 1 | ≤1 | 42/42 (100%)  |
|   |  Tigecycline | 0.03 - 0.25 | ≤0.25 | 42/42 (100%)  |
|   |  Vancomycin | 0.5 - 2 | 0.5 - 2 | 42/42 (100%)  |
|  Pseudomonas aeruginosa ATCC 27853 | Amikacin | 1 - 4 | ≤4 - 8 | 30/30 (100%)  |
|   |  Aztreonam | 2 - 8 | 2 - 8 | 30/30 (100%)  |
|   |  Cefepime | 0.5 - 4 | ≤1 - 4 | 30/30 (100%)  |
|   |  Cefotaxime | 8 - 32 | 8 - 32 | 30/30 (100%)  |
|   |  Ceftazidime | 1 - 4 | ≤1 - 4 | 30/30 (100%)  |
|   |  Ciprofloxacin | 0.12 - 1 | ≤0.12 - 1 | 30/30 (100%)  |

{6}

|  QC organism | Antimicrobial agent | QC range (CLSI M100) | On panel values * (panel IFU) | No. MIC within QC Range  |
| --- | --- | --- | --- | --- |
|   | Gentamicin | 0.5 - 2 | ≤1 - 4 | 30/30 (100%)  |
|   |  Imipenem | 1 - 4 | 1 - 4 | 30/30 (100%)  |
|   |  Levofloxacin | 0.5 - 4 | ≤0.5 - 4 | 30/30 (100%)  |
|   |  Meropenem | 0.12 - 1 | ≤0.12 - 1 | 30/30 (100%)  |
|   |  Piperacillin/Tazobactam | 1/4 - 8/4 | ≤8/4 | 30/30 (100%)  |
|   |  Tetracycline | 8 - 32 | 8 - >8 | 30/30 (100%)  |
|   |  Ticarcillin/K Clavulanate | 8/2 - 32/2 | ≤16/2 - 32/2 | 30/30 (100%)  |
|   |  Tobramycin | 0.25 - 1 | ≤2 | 30/30 (100%)  |
|  Enterococcus faecalis ATCC 29212 | Ampicillin | 0.5 - 2 | ≤1 | 30/30 (100%)  |
|   |  Ciprofloxacin | 0.25 - 2 | ≤0.5 - 1 | 30/30 (100%)  |
|   |  Daptomycin | 1 - 4 | 1 - 4 | 30/30 (100%)  |
|   |  Erythromycin | 1 - 4 | 1 - 4 | 30/30 (100%)  |
|   |  Levofloxacin | 0.25 - 2 | ≤1 - 2 | 30/30 (100%)  |
|   |  Linezolid | 1 - 4 | ≤1 - 4 | 30/30 (100%)  |
|   |  Moxifloxacin | 0.06 - 0.5 | ≤0.25 - 0.5 | 30/30 (100%)  |
|   |  Nitrofurantoin | 4 - 16 | ≤32 | 30/30 (100%)  |
|   |  Penicillin G | 1 - 4 | 0.5 - 2 | 30/30 (100%)  |
|   |  Tetracycline | 8 - 32 | 4 - >8 | 30/30 (100%)  |
|   |  Tigecycline | 0.03 - 0.12 | ≤0.25 | 30/30 (100%)  |
|   |  Vancomycin | 1 - 4 | 1 - 4 | 30/30 (100%)  |

*Due to MicroScan panel design, the full QC range was not available for all organism/drug combinations. For the drugs where the concentrations on the MicroScan panel do not cover the full CLSI/FDA-recommended QC range, the reported value was taken as an indication of acceptable QC.

V Substantial Equivalence Information:

A Predicate Device Name(s): Colibri

B Predicate 510(k) Number(s): K223245

K232756 - Page 7 of 16

{7}

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | DEVICE K232756 | PREDICATE K223245  |
| --- | --- | --- |
|  Device Trade Name | Colibrí | Colibrí  |
|  General Device Characteristic Similarities  |   |   |
|  Intended Use/Indications For Use | The Colibrí is an automated in vitro diagnostic specimen preparation system for use with WASPLab to prepare MALDI-TOF targets for the bioMérieux VITEK MS systems or Bruker MALDI Biotyper CA mass spectrometry systems for qualitative identification and microbial suspension for the bioMérieux VITEK 2 systems or Beckman Coulter MicroScan WalkAway Antimicrobial Susceptibility Testing (AST) systems for qualitative testing of isolated colonies of gram-negative and gram-positive bacterial species grown on solid culture media.The Colibrí is an automated pre-analytical processor that picks isolated colonies designated by the operator and uses a pipetting system to prepare MALDI-TOF MS (Matrix-Assisted Laser Desorption/Ionization-Time of Flight Mass Spectrometry) target slides for bacterial identification and microbial suspension at known concentration for Antimicrobial Susceptibility Testing and purity assessment.The Colibrí software records the identity of each sample and its position on the target slide and communicates this information electronically to the MALDI-TOF MS analyzers. Bacterial suspensions for AST and purity plates are identified by barcode label.The Colibrí is intended for use by trained healthcare professionals in clinical laboratories in conjunction with other clinical and laboratory findings, including Gram staining, to aid in the diagnosis of bacterial infections.The Colibrí has not been validated for use in the identification or processing of yeast species, molds, Nocardia, or mycobacteria. | The Colibrí is an automated in vitro diagnostic specimen preparation system for use with WASPLab to prepare MALDI-TOF targets for the bioMérieux VITEK MS or Bruker MALDI Biotyper CA mass spectrometry systems for qualitative identification and microbial suspensions for the bioMérieux VITEK 2 Antimicrobial Susceptibility Testing (AST) system for qualitative testing of isolated colonies of gram-negative and gram-positive bacterial species grown on solid culture media.The Colibrí is an automated pre-analytical processor that picks isolated colonies designated by the operator and uses a pipetting system to prepare MALDI-TOF MS (Matrix-Assisted Laser Desorption/Ionization-Time of Flight Mass Spectrometry) target slides for bacterial identification and microbial suspension at known concentration for Antimicrobial Susceptibility Testing and purity assessment.The Colibrí software records the identity of each sample and its position on the target slide and communicates this information electronically to the MALDI-TOF MS analyzers. Bacterial suspensions for AST and purity plates are identified by barcode label.The Colibrí is intended for use by trained healthcare professionals in clinical laboratories in conjunction with other clinical and laboratory findings, including Gram staining, to aid in the diagnosis of bacterial infections.The Colibrí has not been validated for use in the identification or processing of yeast species, molds, Nocardia, or mycobacteria.  |
|  Method of Testing | Direct testing from isolated colonies for ID purposes in conjunction with | Direct testing from isolated colonies for ID purposes in conjunction with  |

K232756 - Page 8 of 16

{8}

K232756 - Page 9 of 16
|  Device & Predicate Device(s): | DEVICE K232756 | PREDICATE K223245  |
| --- | --- | --- |
|   | bioMérieux VITEK MS systems or Bruker MALDI Biotyper CA systems. Direct testing from isolated colonies for AST purposes in conjunction with bioMérieux VITEK 2 systems or Beckman Coulter MicroScan WalkAway systems. | bioMérieux VITEK MS systems or Bruker MALDI Biotyper CA systems. Direct testing from isolated colonies for AST purposes in conjunction with bioMérieux VITEK 2 systems.  |
|  Sample/Media Type | Isolated bacterial colonies from any patient source grown on solid media plates included in K193138 and K220546. | Same  |
|  Plate Management | Barcode-labeling plates are manually loaded into the Colibrí after image acquisition. | Same  |
|  Plate Image Acquisition | WASPLab camera | Same  |
|  Colony Selection | Colonies to be picked are identified on a digital plate image using a Graphical User Interface on a dedicated workstation. | Same  |
|  Method of Colony Picking | Automatic picking of colonies using pipette tips. | Same  |
|  ID Target Preparation | When connected with VITEK MS, a portion of microbial colony from an agar plate is automatically spotted on a VITEK MS-DS target slide (VITEK MS DS Target Slides, 48 positions disposable plastic targets) by using the pipetting system. 1μL of matrix is automatically applied to the spot using the pipetting system. The dried target slide is then manually loaded into the VITEK MS. When connected with MALDI Biotyper CA instrument, a portion of microbial colony from an agar plate is automatically spotted on a Bruker Target Plate (IVD 48 Spot Target plate or MBT Biotarget 96 US IVD) by using the pipetting system. 1μL of matrix is automatically applied to the spot using the pipetting system. The dried target slide is then manually loaded into the MALDI Biotyper CA instrument. | Same  |
|  AST Suspension Preparation | Using a pipetting system, a predefined number of morphologically similar colonies are transferred into the Primary Tube that is successively vortexed. A homogenous heavy suspension of organisms is prepared and checked by using on-board Colibrí nephelometer. A variable aliquot of the heavy suspension is automatically transferred into the Secondary Tube to obtain the final | Same  |

{9}

|  Device & Predicate Device(s): | DEVICE K232756 | PREDICATE K223245  |
| --- | --- | --- |
|   | microbial concentration according to IVD package insert. The suspensions prepared by Colibri must be tested in MANUAL MODE on the VITEK 2 or used to rehydrate MicroScan AST panels. |   |
|  Calibration | Colibri requires four different calibrations, one on the nephelometer, three on the cameras. None require user intervention. | Same  |
|  Preparatory Activities | Nephelometer verification by check using Daily Verification Kit. | Same  |
|  Quality Control | Dependent on next-step IVD analyzers. | Same  |
|  General Device Characteristic Differences  |   |   |
|  Primary Tubes | VITEK 2: contains saline solution (with a yellow cap)
MicroScan: contains water (with a red cap) or saline solution (with a blue cap) | VITEK 2: contains saline solution (with a yellow cap)
MicroScan: n/a  |
|  Secondary Tubes | VITEK 2: manually prefilled with PBS solution 0.45% NaCl, pH 4.5-7.0 (sterile); manually loaded
MicroScan: contains Inoculum Water with PLURONIC or Mueller Hinton Broth with 3% Lysed Horse Blood; manually loaded | VITEK 2: manually prefilled with PBS solution 0.45% NaCl, pH 4.5-7.0 (sterile); manually loaded
MicroScan: n/a  |
|  Card Inoculation | VITEK 2: the prepared Secondary Tube is placed in the cassette with a susceptibility card following the warning / recommendation written in the VITEK 2 Instructions for Use.
MicroScan: The prepared Secondary Tube is manually inoculated in a susceptibility panel following the MicroScan Instructions for Use. | VITEK 2: the prepared Secondary Tube is placed in the cassette with a susceptibility card following the warning / recommendation written in the VITEK 2 Instructions for Use.
MicroScan: n/a  |

VI Standards/Guidance Documents Referenced:

The following specific standards and guidances were used to inform device and study design for preparation of samples for AST, as previously described in the Decision Summary for K220546:

- Guidance for Industry and Food and Drug Administration Staff – Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems (August 28, 2009)
- CLSI M100. Performance Standards for Antimicrobial Susceptibility Testing. (33rd Edition)
- CLSI M07. Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically (11th Edition)

Specific standards and guidances used to inform device and study design for the preparation of samples for ID can be found in the K193138 Decision Summary.

K232756 - Page 10 of 16

{10}

# VII Performance Characteristics:

The performance of the Colibri was evaluated in conjunction with the Beckman Coulter MicroScan WalkAway using the criteria for MIC interpretation as described in the MicroScan labeling.

Refer to the K193138 Decision Summary for the performance characteristics of the sample preparation for bacterial identification and MALDI-TOF MS workflow. Refer to the K220546 Decision Summary for the performance characteristics of the sample preparation for the AST workflow with bioMérieux VITEK 2. Refer to the Decision Summary for K223245 for the performance characteristics of sample preparation with WASPLab for downstream ID and AST.

# A Analytical Performance:

# 1. Precision/Reproducibility:

The reproducibility of MIC results obtained using microbial suspensions automatically prepared from isolated colonies by the Colibri was evaluated with the Beckman Coulter MicroScan WalkAway. Inter-operator reproducibility of the Colibri was previously demonstrated in K220546 and K223245. For the reproducibility study with the MicroScan, three Colibri instruments paired with one WASPLab instrument were used to prepare microbial suspensions from overnight (18 to 24 hour) cultures of 18 representative, clinically relevant bacterial isolates grown on trypticase soy agar with  $5\%$  sheep blood. Representative MicroScan AST panels were chosen for analysis and were tested with pure colonies of nine gram-negative isolates and nine gram-positive isolates (summarized in Table 2). Each panel included drugs representing different drug classes.

Table 1. Representative MicroScan Panels Tested in Reproducibility Study

|  MicroScan Panel | Indicated Organisms Groups | Antimicrobial Agents  |
| --- | --- | --- |
|  MSP6 | Streptococcus spp. | Ampicillin, Cefepime, Cefotaxime, Ceftriaxone, Clarithromycin, Clindamycin, Daptomycin, Erythromycin, Levofloxacin, Linezolid, Meropenem, Penicillin G, Tetracycline, Vancomycin  |
|  NM-NF50 | Enterobacterales, Non-fermenters | Amikacin, Gentamicin, Tobramycin, Cefepime, Ceftazidime, Cefotaxime, Meropenem, Imipenem, Ampicillin/Sulbactam, Piperacillin/Tazobactam, Ticarcillin/K Clavulanate, Aztreonam, Levofloxacin, Ciprofloxacin, Minocycline, Tetracycline, Trimethoprim/Sulfamethoxazole  |

K232756 - Page 11 of 16

{11}

Each isolate was tested in triplicate on three separate days using the appropriate MicroScan panel. On each day of testing, one operator designated at least four (for gram negative) or six (for gram positive) colonies for automated picking from digital images acquired by the WASPLab of each of the three culture plates corresponding to each isolate. A minimum of 117 MIC results were analyzed per panel per instrument.

Samples were prepared and assessed for purity consistent with the Colibri instructions for use. The final suspension in the Secondary Tube was used to manually inoculate the appropriate MicroScan AST panel. The secondary tubes and corresponding AST cards were processed by one MicroScan WalkAway instrument. The reported MIC values were used to calculate within-run and between-run reproducibility.

Reproducibility was calculated as the total number of MIC results that were within one doubling dilution of the mode result divided by total number of results. Both best-case (assumes that off-scale results are within one dilution of the mode) and worst-case (assumes that off-scale results are more than one dilution of the mode) performance was determined for each antimicrobial, as outlined in the AST Special Controls Guidance.

In total, 2,619 MIC results were generated with 2,602 of those results being on-scale. The results for all samples are summarized and stratified by the tested MicroScan panel in Table 3. For antimicrobials across all panels evaluated, the combined best-case reproducibility was $\geq 95\%$ and combined worst-case reproducibility was $\geq 89\%$. All purity plates displayed monomicrobial growth. The reproducibility results are acceptable.

Table 3. Reproducibility Results Stratified by MicroScan Panel

|  MicroScan Panel | Colibrí | Best Case (%) | Worst Case (%)  |
| --- | --- | --- | --- |
|  MSP6 | Instrument 1 | 117/117 (100%) | 113/117 (96.6%)  |
|   |  Instrument 2 | 117/117 (100%) | 110/117 (94.0%)  |
|   |  Instrument 3 | 117/117 (100%) | 108/117 (92.3%)  |
|   |  Combined | 351/351 (100%) | 331/351 (94.3%)  |
|  NM-NF50 | Instrument 1 | 459/459 (100%) | 453/459 (98.7%)  |
|   |  Instrument 2 | 458/459 (99.8%) | 453/459 (99.7%)  |
|   |  Instrument 3 | 459/459 (100%) | 452/459 (98.5%)  |
|   |  Combined | 1376/1377 (99.9%) | 1358/1377 (98.6%)  |
|  PM-E37 | Instrument 1 | 297/297 (100%) | 297/297 (100%)  |
|   |  Instrument 2 | 297/297 (100%) | 295/297 (99.3%)  |
|   |  Instrument 3 | 297/297 (100%) | 297/297 (100%)  |

K232756 - Page 12 of 16

{12}

|  MicroScan Panel | Colibrí | Best Case (%) | Worst Case (%)  |
| --- | --- | --- | --- |
|   | Combined | 891/891 (100%) | 889/891 (99.8%)  |

Reproducibility of MIC results obtained using microbial suspensions automatically prepared by the Colibrí for use with the bioMérieux VITEK 2 was previously demonstrated in K220546. The sponsor updated the reproducibility results tables for the VITEK 2 to the format above to harmonize their data presentation in labeling between both downstream AST analyzers.

2. Linearity:

Not applicable.

3. Analytical Specificity/Interference:

See Section 4 below.

4. Accuracy (Instrument):

## Microbial Suspension Accuracy

The ability of the Colibrí to prepare accurate microbial suspensions in Primary and Secondary Tubes compatible for use with the Beckman Coulter MicroScan WalkAway was assessed using representative organisms. Isolated colonies were used by the Colibrí to automatically prepare microbial suspensions. Briefly, four microorganisms (E. coli, P. aeruginosa, S. aureus and E. faecalis) were individually plated on trypticase soy agar + 5% sheep blood and incubated for 18-24 hours. Streptococcus microbial suspensions were previously evaluated in K220546 and taken together with the AST accuracy study described herein (in which 48 Streptococcus isolates were evaluated), the results were deemed acceptable. Accuracy of microbial suspensions prepared with the Colibrí from MacConkey and Chocolate agar media was evaluated in K220546. One operator viewed the plate images acquired by one WASPLab instrument and designated four (for gram-negative) or six (for gram-positive) colonies for picking. Three Colibrí instruments were used to prepare Primary Tubes and Secondary Tubes. Secondary Tubes were manually diluted, plated in triplicate, and incubated overnight to determine colony counts and calculate concentration (CFU/mL).

Concentrations were deemed acceptable if they were within the expected ranges reported in the AST Special Controls guidance (3-7x10⁵ CFU/mL for E. coli) or CLSI M07 (2-8x10⁵ CFU/mL) documents. Accuracy was calculated as the percentage of Secondary Tubes with microbial concentrations within the expected range. At least 30 microbial suspensions were prepared for each species. For the tested species, Primary Tubes were filled with saline and Secondary Tubes were filled with MicroScan Inoculum Water with PLURONIC, in accordance with the MicroScan IFU. The summary of agreement between the expected concentration and the concentration calculated by colony count is shown in Table 4 below, stratified by Colibrí instrument and microorganism. Agreement for each species and each instrument was ≥95% and overall agreement was ≥98%. The microbial suspension accuracy results are acceptable.

K232756 - Page 13 of 16

{13}

Table 4. Percentage Suspensions with Acceptable Microbial Concentration

|  Microorganism | Avg. calculated concentration (CFU/mL) | Expected concentration range (CFU/mL) | Colibrí #1 | Colibrí #2 | Colibrí #3 | Overall  |
| --- | --- | --- | --- | --- | --- | --- |
|  E. coli | 4.8 x 10^{5} | 3-7x10^{5} | 12/12 | 12/12 | 12/12 | 36/36 (100%)  |
|  P. aeruginosa | 2.9 x 10^{5} | 2-8x10^{5} | 10/10 | 9/10 | 10/10 | 29/30 (96.7%)  |
|  S. aureus | 5.7 x 10^{5} | 2-8x10^{5} | 13/14 | 14/14 | 14/14 | 41/42 (97.6%)  |
|  E. faecalis | 4.2 x 10^{5} | 2-8x10^{5} | 10/10 | 10/10 | 10/10 | 30/30 (100%)  |
|  Total | 4.5 x 10^{5} | 2-8x10^{5} | 43/44 (97.7%) | 43/44 (97.7%) | 44/44 (100%) | 130/132 (98.5%)  |

## Accuracy of AST Results

The accuracy of MICs obtained by the Beckman Coulter MicroScan WalkAway with microbial suspensions prepared by the Colibrí was evaluated by testing representative MicroScan panels with isolated colonies of clinically relevant gram-negative and gram-positive bacteria. Three different Colibrí instruments with one WASPLab operated by three different operators were used to prepare suspensions for testing of from colonies of 110 representative isolates of different species of Enterobacterales (n=50), Staphylococcus spp. (n=20), Streptococcus spp. (n=12), Enterococcus spp. (n=18), and non-fermenting gram-negative bacilli (n=10) grown on trypticase soy agar + 5 % sheep blood and/or MacConkey agar. Refer to Table 2 for the panels included in the AST challenge study.

MicroScan panels containing broad concentrations of antimicrobials representative of major drug classes (Table 2) were tested with representative clinical isolates: NM-NF50 covering Enterobacterales and non-fermenters, PM-E37 covering Staphylococci and Enterococci, and MSP6 covering Streptococci. Use of these panels is acceptable because a wider MIC range was validated with the same sample type (i.e., colonies) and determined to be acceptable, as described in the applicable MicroScan decision summaries. Twenty-nine different antimicrobial agents were analyzed, resulting in a total of 1,232 MIC results.

The MIC results obtained with suspensions prepared from isolated colonies by the Colibrí and tested on the MicroScan WalkAway system were compared to the results from manually prepared samples and the MicroScan WalkAway system. Manual samples were prepared from isolated colonies consistent with the MicroScan system instructions for use.

Essential Agreement (EA) was defined as MIC results from Colibrí prepared samples that were within one doubling dilution of the MIC results from the manually prepared samples. Category Agreement (CA) was defined as MIC interpretations (S/I/R) that were the same between the Colibrí-prepared and manually prepared samples. Very major errors were defined as false susceptible results from the Colibrí-prepared samples, major errors were defined as false resistance results from the Colibrí-prepared samples, and minor errors were

K232756 - Page 14 of 16

{14}

defined as results with minor discrepancies (i.e., an intermediate result reported as either resistant or susceptible, or vice versa). Since this is a method-to-method comparison, results were considered acceptable if the EA and CA were $\geq 95\%$ with no major or very major errors. Additionally, no significant differences should be observed between the Colibrí instruments, operators, culture medium, or incubation time.

Overall performance of MicroScan results using microbial suspensions prepared from isolated colonies by the Colibrí demonstrated that 1232/1232 (100%) of MIC results were within EA and 4187/4254 (98.4%) of results were in CA, which is acceptable. No very major or major errors occurred. All purity plates exhibited 100% monomicrobial growth.

Table 5 summarizes the results stratified by antimicrobial/organism group combinations. Due to the high degree of agreement between MICs obtained from manual and Colibri-prepared samples, the AST accuracy was determined to be acceptable.

Table 5. Summary of AST Results, Stratified by MicroScan Panel

|  MicroScan Panel | Organism group | Total tested | # EA | % EA | Total Eval. | # EA Eval. | % EA Eval. | Total cat. | # CA | % CA | # S | # R | # vmj | # maj | # min  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  NM-NF50 | Enterobacterales | 2454 | 2454 | 100% | 577 | 577 | 100% | 2352 | 2304 | 98.0% | 119
7 | 996 | 0 | 0 | 48  |
|  NM-NF50 | Non-fermenters | 294 | 294 | 100% | 165 | 165 | 100% | 291 | 282 | 96.9% | 126 | 120 | 0 | 0 | 9  |
|  PM-E37 | Staphylococcus | 690 | 690 | 100% | 157 | 157 | 100% | 663 | 656 | 98.9% | 465 | 159 | 0 | 0 | 7  |
|  PM-E37 | Enterococcus | 525 | 525 | 100% | 258 | 258 | 100% | 519 | 516 | 99.4% | 354 | 102 | 0 | 0 | 3  |
|  MSP6 | Streptococcus | 429 | 429 | 100% | 75 | 75 | 100% | 429 | 429 | 100% | 381 | 39 | 0 | 0 | 0  |

Accuracy of MIC results obtained using microbial suspensions automatically prepared by the Colibrí for use with the bioMérieux VITEK 2 was previously demonstrated in K220546. The sponsor updated the AST accuracy results tables for the VITEK 2 to the format above to harmonize their data presentation in labeling between both downstream AST analyzers.

5. Carry-Over:

The ability of the on-board incinerator to properly sterilize the spreader after purity plate preparation was described in K220546. Refer to the K220546 Decision Summary for additional details.

B Other Supportive Instrument Performance Characteristics Data:

Refer to the Decision Summaries of K193138, K220546, and K223245 for supportive studies regarding Colibrí use with bacterial ID IVD-analyzers, bioMérieux VITEK 2 AST analyzer, and WASPLab, respectively.

VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

K232756 - Page 15 of 16

{15}

IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K232756 - Page 16 of 16

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON/K232756](https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON/K232756)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
