← Product Code [LON](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON) · K193299

# VITEK 2 AST-Gram Negative Ceftazidime (=<0.5 - =>32 µg/mL) (K193299)

_bioMerieux, Inc. · LON · Sep 28, 2020 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON/K193299

## Device Facts

- **Applicant:** bioMerieux, Inc.
- **Product Code:** [LON](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON.md)
- **Decision Date:** Sep 28, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1645
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

VITEK® 2 AST-Gram Negative Ceftazidime is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 AST-Gram Negative Ceftazidime is a quantitative test. Ceftazidime has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections: Citrobacter species Enterobacter species Escherichia coli Klebsiella species Proteus mirabilis Proteus vulgaris Pseudomonas aeruginosa Serratia species In vitro data are available, but clinical significance is unknown: Acinetobacter species Citrobacter koseri (formerly Citrobacter diversus) Citrobacter freundii Providencia species (including Providencia rettgeri) Salmonella species Shigella species Yersinia enterocolitica The VITEK® 2 Gram-Negative Susceptibility Card is intended for use with the VITEK® 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram-negative bacilli to antimicrobial agents when used as instructed.

## Device Story

VITEK 2 AST-Gram Negative Ceftazidime is an automated in vitro diagnostic test for determining minimum inhibitory concentrations (MIC) of ceftazidime against aerobic Gram-negative bacilli. The device uses miniaturized, 64-well test cards containing premeasured antimicrobial concentrations and nutrient media. Clinical laboratories prepare a standardized saline suspension of the patient isolate (via auto-dilution or manual dilution), which rehydrates the card. The VITEK 2 system incubates the card and uses an optical scanner to measure light attenuation every 15 minutes, monitoring bacterial growth. The system's software analyzes growth patterns to calculate MIC values. Results are reported as quantitative MICs and interpretive categories (Susceptible, Intermediate, Resistant). The device aids clinicians in selecting appropriate antimicrobial therapy for infections. Limitations exist for specific organism/antibiotic combinations (e.g., P. aeruginosa at 8 µg/mL) requiring alternative testing methods.

## Clinical Evidence

Bench testing only. Performance evaluated using 981 clinical and 118 challenge isolates. Comparison against CLSI broth microdilution reference method showed acceptable essential agreement (EA) and category agreement (CA). For P. aeruginosa, adjusted very major error rate was 2.6% after accounting for MIC values one doubling dilution from reference. Trending analysis identified a bias toward lower MIC values for Yersinia enterocolitica.

## Technological Characteristics

Automated growth-based detection system using light transmittance optics. 64-well test card format. Energy source: electrical (VITEK 2/Compact systems). Connectivity: networked/standalone. Software: VITEK 2 Systems 9.02 or later. Standardized inoculum preparation via DensiCHEK Plus. Materials: nutrient media and premeasured antimicrobial concentrations.

## Regulatory Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

## Predicate Devices

- VITEK® 2 AST-GN Eravacycline ([K191766](/device/K191766.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number

K193299

B Applicant

bioMérieux, Inc.

C Proprietary and Established Names

VITEK 2 AST-Gram Negative Ceftazidime (≤0.5 - ≥32 μg/mL)

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  LON, LTT, LTW | Class II | 21 CFR 866.1645 - Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System | MI - Microbiology  |

## II Submission/Device Overview:

A Purpose for Submission:

To obtain a substantial equivalence determination for ceftazidime for testing of Gram-negative bacilli on the VITEK 2 and VITEK 2 Compact Antimicrobial Susceptibility Test (AST) Systems

B Measurand:

Ceftazidime ≤0.5 – ≥32 μg/mL

C Type of Test:

Automated quantitative or qualitative antimicrobial susceptibility test

## III Intended Use/Indications for Use:

A Intended Use(s):

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

{1}

The VITEK 2 Gram-Negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram-negative bacilli to antimicrobial agents when used as instructed.

## B Indication(s) for Use:

VITEK 2 AST-Gram Negative Ceftazidime is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 AST-Gram Negative Ceftazidime is a quantitative test. Ceftazidime has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.

Active in vitro and in clinical infections:
- Citrobacter species
- Enterobacter species
- Escherichia coli
- Klebsiella species
- Proteus mirabilis
- Proteus vulgaris
- Pseudomonas aeruginosa
- Serratia species

In vitro data are available, but clinical significance is unknown:
- Acinetobacter species
- Citrobacter koseri (formally Citrobacter diversus)
- Citrobacter freundii
- Providencia species (including Providencia rettgeri)
- Salmonella species
- Shigella species
- Yersinia enterocolitica

The VITEK 2 Gram-Negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram-negative bacilli to antimicrobial agents when used as instructed.

## C Special Conditions for Use Statement(s):

- Rx - For Prescription Use Only
- The lack of an intermediate category has shown very major discrepancies when compared to the reference method. Testing should be repeated using an alternative method prior to reporting results for the following antibiotic/organism combination(s): Ceftazidime: Pseudomonas aeruginosa when the VITEK 2 MIC is 8 µg/ml.
- Perform an alternative method of testing prior to reporting for the following antibiotic / organism combination(s): Ceftazidime: Morganella morganii

K193299 - Page 2 of 13

{2}

- The ability of the AST card to detect resistance with the following combination(s) is unknown because an insufficient number of resistant strains were available at the time of comparative testing:
- Ceftazidime: *Citrobacter koseri*, *Yersinia enterocolitica*

## D Special Instrument Requirements:

VITEK 2 and VITEK 2 Compact Systems using VITEK 2 Systems 9.02 software or later versions

## IV Device/System Characteristics:

### A Device Description:

The VITEK 2 AST card is a miniaturized, abbreviated and automated version of the doubling dilution technique for determining the minimum inhibitory concentration (MIC). Each VITEK 2 AST card contains 64 wells. A control well(s) which contain only nutrient medium is resident on all cards. The remaining wells contain premeasured portions of antimicrobials combined with the nutrient media. The isolate to be tested is diluted to a standardized concentration with 0.45% to 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System will automatically (or allow operator to manually) dilute the bacterial suspension to prepare an inoculum for susceptibility cards. Then, the VITEK 2 will fill, seal and place the card into the incubator/reader. The VITEK 2 Compact has a manual filling, sealing, and loading operation. The VITEK 2 Systems monitor the growth of each well in the card over a defined period of time. The analysis program determines when a well demonstrates growth based on attenuation of light measured by an optical scanner. This data is used to determine the minimum inhibitory concentration or "MIC" values for the antimicrobial agent. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.

VITEK 2 AST-Gram Negative Ceftazidime has the following concentrations in the card: 1, 2, 4, 8 and 32 µg/mL (equivalent standard method concentration by efficacy in µg/mL). The ceftazidime MIC result range for the VITEK 2 is ≤0.5 to ≥32 µg/mL. For all species, the VITEK 2 system is capable of reporting the following MIC results: ≤0.5, 1, 2, 4, 8, 16 and ≥32 µg/mL for the AST-Gram Negative Ceftazidime test.

### B Principle of Operation:

The VITEK 2 and VITEK 2 Compact Systems utilize automated growth-based detection using attenuation of light measured by an optical scanner. The optics in the systems use visible light to directly measure organism growth within each of the 64 micro-wells. Transmittance optics is based on an initial light reading of a well before significant growth has begun. Every 15 minutes throughout the incubation cycle (defined period of time based on the VITEK 2 card), light transmittance readings of each well determine organism growth by the amount of light that is prevented from passing through the well. At the completion of the incubation period, the MIC values and their associated interpretive category results for each antimicrobial on the test card are displayed in an automatically generated report.

### V Substantial Equivalence Information:

K193299 - Page 3 of 13

{3}

A Predicate Device Name(s):

VITEK 2 AST-Gram Negative Eravacycline  $(\leq 0.12 - \geq 4\mu \mathrm{g / mL})$

B Predicate 510(k) Number(s):

K191766

C Comparison with Predicate(s):

Table 1. Device Comparison with Predicate

|  Device & Predicate Device(s): | Device: K193299 | Predicate: K191766  |
| --- | --- | --- |
|  Device Trade Name | Vitek 2 AST-Gram Negative Ceftazidime (≤0.5 – ≥32 μg/mL) | Vitek 2 AST-Gram Negative Eravacycline (≤0.12 – ≥4 μg/mL)  |
|  General Device Characteristic Similarities  |   |   |
|  Intended Use | The VITEK 2 Gram-Negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic Gram-negative bacilli to antimicrobial agents when used as instructed. | Same  |
|  Test Method | Automated quantitative antimicrobial susceptibility test for use with the VITEK 2 and VITEK 2 Compact Systems to determine the in vitro susceptibility of Gram-negative bacilli | Same  |
|  Inoculum | Standardized saline suspension of test organism | Same  |
|  Test Card | VITEK 2 Gram Negative Susceptibility Test Card | Same  |
|  Instrument | VITEK 2 and VITEK 2 Compact Systems | Same  |
|  Analysis Algorithm | Growth pattern analysis | Same  |
|  General Device Characteristic Differences  |   |   |
|  Antimicrobial Agent | Ceftazidime | Eravacycline  |
|  Antimicrobial Concentration | 1, 2, 4, 8 and 32 μg/mL | 0.25, 1, 2 and 4 μg/mL  |
|  Reporting Range | ≤0.5 to ≥32 μg/mL | ≤0.12 to ≥4 μg/mL  |
|  Indicated Organisms | Citrobacter species, Enterobacter species, Escherichia coli, Klebsiella species, Proteus mirabilis, Proteus vulgaris, Pseudomonas aeruginosa, Serratia species, Acinetobacter species, Citrobacter koseri (formally Citrobacter diversus), Citrobacter | Citrobacter freundii, Enterobacter cloacae, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Citrobacter koseri, Klebsiella (Enterobacter) aerogenes  |

K193299 - Page 4 of 13

{4}

|  Device & Predicate Device(s): | Device: K193299 | Predicate: K191766  |
| --- | --- | --- |
|   | freundii, Providencia species (including Providencia rettgeri), Salmonella species, Shigella species, Yersinia enterocolitica |   |

VI Standards/Guidance Documents Referenced:

- Guidance for Industry and FDA - Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems – August 28, 2009.
- CLSI M07, 10th ed., "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard, 2015".
- CLSI M100, 29th ed., "Performance Standards for Antimicrobial Susceptibility Testing; Twenty-Ninth Informational Supplement, January 2019".

VII Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility:

Reproducibility testing for the VITEK 2 AST-GN card with Ceftazidime was conducted at three external clinical sites using a panel of ten Gram-negative bacilli consistent with the indications for use (i.e., four Pseudomonas aeruginosa, three Enterobacter spp., two Serratia marcescens and one Klebsiella pneumoniae pneumoniae). Each isolate was tested in triplicate over three days for a total of 270 data points. Inocula were prepared using both the auto-dilution and manual dilution methods for testing in the VITEK 2 System. In addition, inocula were prepared by the manual dilution method for use with the VITEK 2 Compact. The mode or median of MIC values was determined for each isolate and the reproducibility was calculated based on the number of MIC values that fell within ±1 doubling dilution of the mode/median MIC value. The data was analyzed taking into consideration best-case and worst-case scenarios as described in the Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems. The reproducibility performance is summarized in Table 2.

Table 2. Reproducibility of VITEK 2 AST-GN with Ceftazidime

|   | VITEK 2 |   | VITEK 2 Compact  |
| --- | --- | --- | --- |
|   |  Auto-Dilution | Manual Dilution | Manual Dilution  |
|  Best Case | 262/270 (97.0%) | 261/270 (96.7%) | 263/270 (97.4%)  |
|  Worst Case | 250/270 (92.6%) | 247/270 (91.5%) | 254/270 (94.1%)  |

All reproducibility performance results were considered acceptable.

2. Linearity:

Not applicable

K193299 - Page 5 of 13

{5}

3. Analytical Specificity/Interference:
Not applicable

4. Assay Reportable Range:
Not applicable

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

Quality Control (QC) Testing:
The CLSI recommended QC strains, namely *E. coli* ATCC 25922 and *P. aeruginosa* ATCC 27853 were tested a sufficient number of times (i.e., at least 20/site) at each testing site using both the VITEK 2 card and broth microdilution (BMD) reference methods. Both the automatic dilution and manual dilution methods were used for the VITEK 2 and the manual dilution method was used for the VITEK 2 Compact. The results for all sites combined are summarized in Table 3.

The quality control results were within the expected results range &gt; 95% of the time which is acceptable.

Table 3: Quality Control Result Frequencies for VITEK 2 (Auto-Dilution and Manual Dilution Methods) and VITEK 2 Compact (Manual Dilution Method)

|  Organism | VITEK 2 Result Range^{1} (μg/mL) | BMD Result Range (μg/mL) | VITEK 2 Auto-Dilution | BMD | VITEK 2 Manual Dilution | BMD | VITEK 2 Compact Manual Dilution | BMD  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  E. coli ATCC 25922
Expected Result: 0.06 – 0.5 μg/mL |  | ≤0.03 |  |  |  |  |  |   |
|   |   | 0.06 |  |  |  |  |  |   |
|   |   | 0.125 |  | 153 |  | 118 |  | 114  |
|   |   | 0.25 |  | 110 |  | 65 |  | 65  |
|   |  ≤ 0.5 | 0.5 | 266 | 3 | 185 | 2 | 181 | 2  |
|   |  1 | 1 |  |  |  |  |  |   |
|   |  2 | 2 |  |  |  |  |  |   |
|   |  4 | 4 |  |  |  |  |  |   |
|   |  8 | 8 |  |  |  |  |  |   |
|   |  16 | 16 |  |  |  |  |  |   |
|   |  ≥ 32 | 32 |  |  |  |  |  |   |
|   |   | 64 |  |  |  |  |  |   |
|   |   | ≥128 |  |  |  |  |  |   |
|  P. aeruginosa ATCC 27853
Expected Result: 1 - 4 μg/mL |  | ≤0.03 |  |  |  |  |  |   |
|   |   | 0.06 |  |  |  |  |  |   |
|   |   | 0.125 |  |  |  |  |  |   |
|   |   | 0.25 |  |  |  |  |  |   |
|   |  ≤ 0.5 | 0.5 |  |  |  |  |  |   |
|   |  1 | 1 |  | 181 |  | 126 |  | 125  |
|   |  2 | 2 | 267 | 84 | 185 | 57 | 182 | 55  |
|   |  4 | 4 |  | 1 |  | 1 |  | 1  |

K193299 - Page 6 of 13

{6}

|  Organism | VITEK 2 Result Range^{1} (μg/mL) | BMD Result Range (μg/mL) | VITEK 2 Auto-Dilution | BMD | VITEK 2 Manual Dilution | BMD | VITEK 2 Compact Manual Dilution | BMD  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   | 8 | 8 |  | 1^{2} |  | 1^{2} |  | 1^{2}  |
|   |  16 | 16 |  |  |  |  |  |   |
|   |  ≥ 32 | 32 |  |  |  |  |  |   |
|   |   | 64 |  |  |  |  |  |   |
|   |   | ≥128 |  |  |  |  |  |   |

1 Does not include the full CLSI/FDA-recommended dilution range for QC testing. For *E. coli*, an in-range VITEK result will be ≤ the lowest dilution on the card (i.e., ≤ 0.5 μg/mL).
2 Two replicates of this organism were tested; one result was out of range. Since the results for the other replicate and the other QC organisms were within the acceptable range, the reference results were accepted and testing was not repeated. QC results were within the expected range on the next testing day.

## Inoculum Density Check:

The DensiCHEK Plus was used to standardize the inoculum to a 0.5 McFarland standard. The instrument was standardized daily with all results recorded at each site. Calibration values were within the expected range.

## Purity Check:

A purity check of all organisms was performed on the dilution tube used to prepare the VITEK 2 card inoculum. Only those cultures that were pure were evaluated in the study.

## Growth Failure Rate:

A total of 981 clinical isolates were evaluated. A total of 977 organisms grew in the VITEK 2 AST-Gram Negative Ceftazidime test using the auto-dilution method which is acceptable (&lt;10% growth failure).

A total of 118 challenge isolates were evaluated at one site. All 118 challenge organisms grew in the VITEK 2 AST-Gram Negative Ceftazidime test using both the auto-dilution and manual dilution methods for the VITEK 2 and manual dilution method for the VITEK 2 Compact.

A total of four isolates did not grow in the VITEK 2 AST-Gram Negative Ceftazidime test so complete test results are available for 1095 isolates.

6. Detection Limit:

Not applicable

7. Assay Cut-Off:

Not applicable

## B Comparison Studies:

1. Method Comparison with the Reference Method

K193299 - Page 7 of 13

{7}

Testing of ceftazidime on the VITEK 2 AST-Gram Negative card was performed at three external sites and one internal site. There were 981 clinical isolates and 118 challenge isolates tested for a total of 1099 isolates. Results obtained with the VITEK 2 AST-Gram Negative card with ceftazidime were compared to results obtained with the CLSI broth microdilution reference panel. The MIC result range for the VITEK 2 AST-Gram Negative Ceftazidime is ≤0.5 to ≥32 μg/mL for all species. The reference panel contained two-fold serial dilutions with a range of ≤0.0313 to ≥256 μg/mL. The testing conditions for the reference method consisted of the following:

- Medium – Cation Adjusted Mueller Hinton broth
- Inoculum – Direct colony suspension
- Incubation – 35°C; 16-20 hours

The VITEK 2 AST cards were inoculated with test organisms using the auto-dilution method (VITEK 2) and using the manual dilution method (VITEK 2 and VITEK 2 Compact). All test inocula used for the VITEK 2 AST cards and the reference method were standardized using the DensiCHEK Plus instrument.

A total of 981 clinical isolates were evaluated: 67.9% were considered contemporary isolates (isolated from clinical specimens and tested within six months) and 32.1% were stock isolates. Four isolates did not grow in the VITEK 2 AST-Gram Negative Ceftazidime test so complete test results are available for 977 clinical isolates: 968 isolates from indicated species (25 A. baumannii, 13 A. baumannii complex, 1 A. lwoffii, 1 Acinetobacter spp., 1 C. amalonaticus, 2 C. braakii, 12 C. freundii, 3 C. koseri, 1 E. amnigenus, 1 E. asburiae, 27 E. cloacae cloacae, 21 E. cloacae complex, 156 E. coli, 16 K. aerogenes, 27 K. oxytoca, 1 K. pneumoniae pneumoniae, 109 K. pneumoniae, 54 P. mirabilis, 13 P. vulgaris, 7 P. rettgeri, 5 P. stuartii, 403 P. aeruginosa, 13 Salmonella spp., 27 S. marcescens, 1 S. rubidaea, 3 S. boydii, 3 S. dysenteriae, 5 S. flexneri, 4 S. sonnei and 13 Y. enterocolitica) and 9 isolates from non-indicated species (3 B. cepacia, 3 H. alvei, 1 P. agglomerans, 2 R. planticola). The clinical isolates were tested with the auto-dilution option of the VITEK 2.

A total of 118 challenge isolates (10 A. baumannii, 4 A. baumannii complex, 1 A. calcoaceticus, 1 A. johnsonii, 1 A. lwoffii, 1 Acinetobacter spp., 3 C. freundii, 3 C. koseri, 4 E. cloacae cloacae, 20 E. coli, 9 K. aerogenes, 2 K. oxytoca, 6 K. pneumoniae pneumoniae, 9 P. mirabilis, 2 P. vulgaris, 2 P. rettgeri, 2 P. stuartii, 20 P. aeruginosa, 2 Salmonella enterica arizonae, 8 S. marcescens, 1 S. odorifera, 2 S. boydii, 3 S. flexneri and 2 Y. enterocolitica) were evaluated at one site. The challenge set was tested with the auto-dilution and manual dilution options of the VITEK 2 and with the manual dilution method on the VITEK 2 Compact.

At the time of comparative testing, resistant isolates were not available for Citrobacter koseri and Yersinia enterocolitica. The following statement is included in the Limitations section of the device labeling:

The ability of the AST card to detect resistant strains with the following combination(s) is unknown because an insufficient number of resistant strains were available at the time of comparative testing.

Ceftazidime: Citrobacter koseri, Yersinia enterocolitica

K193299 - Page 8 of 13

{8}

# Clinical and Challenge Data – VITEK 2 Auto-Dilution

The results obtained using the auto-dilution method of the VITEK 2 from the 1095 total isolates (977 clinical isolates and 118 challenge isolates) are summarized in Table 4.

Table 4. Performance of All Clinical and Challenge Isolates: VITEK 2 Auto-Dilution

|  Organism Type | EA Tot | EA N | EA % | Eval. EA Tot | Eval. EA N | Eval. EA % | CA N | CA % | #R | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Acinetobacter spp.  |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 40 | 38 | 95.0 | 24 | 22 | 91.7 | 36 | 90.0 | 17 | 4 | 0 | 0  |
|  Challenge | 18 | 16 | 88.9 | 10 | 8 | 80.0 | 16 | 88.9 | 8 | 2 | 0 | 0  |
|  Combined | 58 | 54 | 93.1 | 34 | 30 | 88.2 | 52 | 89.7 | 25 | 6 | 0 | 0  |
|  Enterobacteriaceae^{1}  |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 525 | 514 | 97.9 | 25 | 14 | 58.3 | 514 | 98.3 | 44 | 10 | 1 | 0  |
|  Challenge | 80 | 78 | 97.5 | 6 | 4 | 66.7 | 77 | 96.3 | 2 | 1 | 2 | 0  |
|  Combined | 605 | 592 | 97.9 | 31 | 18 | 58.1 | 591 | 97.7 | 46 | 11 | 3 | 0  |
|  Pseudomonas aeruginosa  |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 403 | 380 | 94.3 | 259 | 236 | 91.1 | 391 | 97.0 | 155 | N/A | 3 | 9  |
|  Challenge | 20 | 20 | 100 | 18 | 18 | 100 | 20 | 100 | 0 | N/A | 0 | 0  |
|  Combined | 423 | 400 | 94.6 | 277 | 254 | 91.7 | 411 | 97.2 | 155 | N/A | 3 | 9  |

Only includes data for Enterobacteriaceae species indicated for VITEK 2 Ceftazidime.

EA – Essential Agreement
min – minor errors
CA – Category Agreement
maj – major errors
EVAL – Evaluable isolates
vmj – very major errors
R – Resistant isolates
N/A – Not applicable due to the lack of an intermediate interpretive criterion for ceftazidime with *P. aeruginosa*

When using the auto-dilution method of the VITEK 2, the overall performance of Acinetobacter spp. is acceptable with an EA of 93.1% and a CA of 89.7%. There were no major or very major errors. The overall performance with all Enterobacteriaceae is acceptable with an EA of 97.9% and a CA of 97.7%. There were three major errors (3/559 = 0.5%) and no very major errors. The overall performance of Pseudomonas aeruginosa is acceptable with an EA of 94.6% and a CA of 97.2%. There were three major (3/268 = 1.1%) and nine (9/155 = 5.8%) very major errors; seven of the nine very major errors had MIC values of 8 µg/mL. This is addressed as a limitation statement in the device labeling:

The lack of an intermediate category has shown very major discrepancies when compared to the reference method. Testing should be repeated using an alternative method prior to reporting results for the following antibiotic/organism combination(s):

Ceftazidime: Pseudomonas aeruginosa when the VITEK 2 MIC is 8 µg/mL

Due to the lack of an intermediate interpretive criteria for ceftazidime with *P. aeruginosa*, further analysis of the errors is performed and adjustments are made by considering the MIC values where the errors occurred. Two of the three major errors had an MIC value that was one doubling dilution from the reference and thus in essential agreement. Therefore, the adjusted number of major errors is reduced to 1 (1/268 = 0.4%). Five of the nine very major errors had an MIC value that was one doubling dilution from the reference and thus in

K193299 - Page 9 of 13

{9}

essential agreement. Therefore, the adjusted number of very major errors is reduced to four $(4/155 = 2.6\%)$. In addition, seven of the nine very major errors had MIC values of $8~\mu \mathrm{g/mL}$. To describe the very major errors, these results are addressed as a footnote to the performance table in the device labeling:

The overall categorical very major error rate for ceftazidime when testing Pseudomonas aeruginosa with the VITEK 2 system was $5.8\%$ (9/155). The MIC values of five of the nine very major errors were one doubling dilution from the MIC value obtained from the reference method. Based on the essential agreement and lack of an intermediate breakpoint for ceftazidime with P. aeruginosa, the adjusted very major error rate for P. aeruginosa is $2.6\%$ (4/155). Seven of the nine very major errors had MIC values of 8 $\mu \mathrm{g/mL}$; therefore, alternative testing is required prior to reporting results for P. aeruginosa when the VITEK 2 MIC value is $8\mu \mathrm{g/mL}$.

## Challenge Data – VITEK 2 and VITEK 2 Compact Manual Dilution

The 118 challenge isolates were also evaluated at one site with the manual dilution options of the VITEK 2 and VITEK 2 Compact systems (summarized in Table 5).

Table 5. Performance of Challenge Isolates: VITEK 2 and VITEK 2 Compact Manual Dilution

|  System | EA Tot | EA N | EA % | Eval. EA Tot | Eval. EA N | Eval. EA % | CA N | CA % | #R | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Acinetobacter spp.  |   |   |   |   |   |   |   |   |   |   |   |   |
|  VITEK 2 | 18 | 16 | 88.9 | 10 | 8 | 80.0 | 16 | 88.9 | 8 | 2 | 0 | 0  |
|  VITEK 2 Compact | 18 | 16 | 88.9 | 10 | 8 | 80.0 | 16 | 88.9 | 8 | 2 | 0 | 0  |
|  Enterobacteriaceae  |   |   |   |   |   |   |   |   |   |   |   |   |
|  VITEK 2 | 80 | 80 | 100 | 4 | 4 | 100 | 79 | 98.8 | 2 | 1 | 0 | 0  |
|  VITEK 2 Compact | 80 | 80 | 100 | 4 | 4 | 100 | 79 | 98.8 | 2 | 1 | 0 | 0  |
|  Pseudomonas aeruginosa  |   |   |   |   |   |   |   |   |   |   |   |   |
|  VITEK 2 | 20 | 20 | 100 | 18 | 18 | 100 | 20 | 100 | 0 | 0 | 0 | 0  |
|  VITEK 2 Compact | 20 | 20 | 100 | 18 | 18 | 100 | 20 | 100 | 0 | 0 | 0 | 0  |

When using the manual dilution method of the VITEK 2 and VITEK2 Compact, the overall performance of all Acinetobacter spp. challenge isolates had an EA and CA of $88.9\%$. This performance is less than the $90\%$ performance criteria outlined in the AST Special Controls guidance document, likely due to the limited number of challenge isolates available for testing. Of the 18 challenge isolates tested, 8 isolates had MIC values that were not evaluable for EA. Of the 10 evaluable MIC values, there were no major or very major errors. This is addressed in the following footnote to the performance table in the device labeling:

A limited number of Acinetobacter isolates were available for testing with the manual dilution. Of the 18 isolates tested, 8 had MIC values that were not evaluable for essential agreement. Of the 10 evaluable MIC values, there were 2 minor categorical errors that were not in essential agreement.

K193299 - Page 10 of 13

{10}

When using the manual dilution method of the VITEK 2 and VITEK2 Compact, the overall performance of all Enterobacteriaceae challenge isolates is acceptable with an EA of 100% and a CA of 98.8%. There were no major or very major errors. Overall performance of P. aeruginosa challenge isolates with either instrument is acceptable with an EA and CA of 100%. There were no major or very major errors.

## Resistance Mechanism Characterization

Challenge isolates of Enterobacteriaceae and P. aeruginosa harboring various molecular mechanisms of resistance noted in the FDA drug label were tested with ceftazidime. The following resistance mechanisms were evaluated: β-lactamases (OXA, SHV, TEM, AMPC and others).

## MIC Trends

A trending analysis was conducted using the combined data (clinical and challenge) obtained from the VITEK 2 auto-dilution method for each indicated organism species and group. This trending calculation analyzes device MIC values that are determined to be one or more doubling dilutions lower or higher than the reference method. MIC values that are off-scale for both the reference and device are not considered in the trending analysis.

Trending results were stratified by species to determine if species-related trends were observed (Table 6). Species for which the difference between the percentage of isolates with higher or lower MIC values was ≥30% with a statistically significant confidence interval were considered to have evidence of trending.

Table 6. Trending by Species (Clinical and Challenge Isolates)

|  VITEK 2 Auto-Dilution  |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Organism | Total Evaluable for Trending | ≥1 dil. Lower # (%) | Exact # (%) | ≥1 dil. Higher # (%) | Percent Difference (95% CI) | Trending Noted  |
|  Acinetobacter spp^{1} | 34 | 7 (20.59) | 12 (35.29) | 15 (44.12) | 23.53 (1.28 to 42.89) | No  |
|  Pseudomonas aeruginosa | 289 | 48 (16.61) | 153 (52.94) | 88 (30.45) | 13.84 (6.95 to 20.58) | No  |
|  Enterobacteriaceae | 45 | 21 (46.67) | 7 (15.5) | 17 (37.78) | -8.89 (-27.96 to 11.15) | No  |
|  Citrobacter spp^{2} | 2 | 0 | 1 (50.0) | 1 (50.0) | 50.00 (-27.26 to 90.55) | No  |
|  Enterobacter spp^{3} | 8 | 3 (37.50) | 2 (25.00) | 3 (37.50) | 0 (-39.83 to 39.83) | No  |
|  Escherichia coli | 9 | 2 (22.22) | 4 (44.44) | 3 (33.33) | 11.11 (-27.75 to 46.17) | No  |
|  Klebsiella spp^{4} | 11 | 7 (63.64) | 0 | 4 (36.36) | -27.27 (-57.25 to 12.69) | No  |
|  Proteus mirabilis | 4 | 1 (25.00) | 0 | 3 (75.00) | 50.00 (-13.55 to 78.91) | No  |
|  Proteus vulgaris | 1 | 0 | 0 | 1 (100) | 100 (-12.21 to 100) | No  |

K193299 - Page 11 of 13

{11}

|  VITEK 2 Auto-Dilution  |   |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Organism | Total Evaluable for Trending | ≥1 dil. Lower # (%) | Exact # (%) | ≥1 dil. Higher # (%) | Percent Difference (95% CI) | Trending Noted  |
|  Providencia spp^{5} | 3 | 1 (33.33) | 0 | 2 (66.67) | 33.33 (-31.58 to 71.78) | No  |
|  Salmonella spp^{6} | 0 | - | - | - | - | -  |
|  Serratia spp^{7} | 1 | 1 (100) | 0 | 0 | -100 (-100 to 12.21) | No  |
|  Shigella spp^{8} | 0 | - | - | - | - | -  |
|  Yersinia enterocolitica | 7 | 6 (85.71) | 0 | 1 (14.29) | -71.43 (-88.00 to -19.06) | Yes  |

$^{1}$ A. baumannii*, A. baumannii complex*, A. calcoaceticus*, A. johnsonii*, A. lwoffii*, Acinetobacter spp.*
$^{2}$ C. freundii*, C. koseri*
$^{3}$ E. amnigenus, E. asburiae, E. cloacae cloacae*, E. cloacae complex
$^{4}$ K. aerogenes*, K. oxytoca, K. pneumoniae*, K. pneumoniae pneumoniae*
$^{5}$ P. rettgeri*, P. stuartii*
$^{6}$ S. enteritica arizonae, Salmonella spp
$^{7}$ S. marcescens*, S. odorifera, S. rubidaea
$^{8}$ S. boydii, S. dysenteriae, S. flexneri, S. sonnei
*denotes species with MIC results evaluable for trending

A trend toward lower MIC values was observed for Yersinia enterocolitica. The following footnote to the performance table is included in the package insert to address the trending observed for VITEK 2 AST-Gram Negative Ceftazidime:

VITEK 2 AST-Gram Negative Ceftazidime values tended to be in exact agreement or at least one doubling dilution lower when testing Yersinia enterocolitica compared to the CLSI reference broth microdilution.

As required under 511A(b)(2)(C)(ii)(I) of the Federal Food, Drug and Cosmetic Act, the following statement is included in the Precautions section of the device labeling to address testing and reporting of non-indicated species:

Per the FDA-Recognized Susceptibility Test Interpretive Criteria website, the safety and efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved.

2. Matrix Comparison:

C Clinical Studies:

1. Clinical Sensitivity:

K193299 - Page 12 of 13

{12}

Not applicable

2. Clinical Specificity:

Not applicable

3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):

Not applicable

D Clinical Cut-Off:

Not applicable

E Expected Values/Reference Range:

The FDA-identified and recognized susceptibility interpretive criteria for ceftazidime are listed in Table 7.

Table 7: FDA-Identified and Recognized Interpretive Criteria for Ceftazidime (μg/mL)ᵃ

|  Organisms | S | I | R  |
| --- | --- | --- | --- |
|  Acinetobacter spp | ≤8 | 16 | ≥32  |
|  Enterobacteriaceae | ≤4 | 8 | ≥16  |
|  Pseudomonas aeruginosa | ≤8 | - | ≥16  |

ᵃAccording to FDA STIC Website.

VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a breakpoint change protocol that was reviewed and accepted by FDA. This protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage (https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm). The protocol outlined the specific procedures and acceptance criteria that bioMérieux intends to use to evaluate the VITEK 2 AST-GN Ceftazidime when revised breakpoints for ceftazidime are published on the FDA STIC webpage. The breakpoint change protocol included with the submission indicated that if specific criteria are met, bioMérieux will update the ceftazidime device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.

K193299 - Page 13 of 13

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON/K193299](https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON/K193299)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
