← Product Code [LON](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON) · K190616

# VITEK 2 AST-Gram Positive Dalbavancin (<=  0.015 - >= 1 µg/mL) (K190616)

_bioMerieux, Inc. · LON · Jun 6, 2019 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON/K190616

## Device Facts

- **Applicant:** bioMerieux, Inc.
- **Product Code:** [LON](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON.md)
- **Decision Date:** Jun 6, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1645
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

VITEK 2 AST-Gram Positive Dalbavancin is designed for antimicrobial susceptibility testing of Gram positive microorganisms and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The VITEK 2 Gram-positive Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed.

## Device Story

VITEK 2 AST-Gram Positive Dalbavancin is an automated, growth-based antimicrobial susceptibility test (AST) system. It uses miniaturized cards containing 64 wells with premeasured antimicrobial concentrations and nutrient media. Input: bacterial isolate suspension (saline-diluted). Operation: VITEK 2 system automatically fills, seals, and incubates cards; VITEK 2 Compact requires manual loading. Principle: optical scanner measures light attenuation (transmittance) every 15 minutes to monitor bacterial growth. Output: MIC values and interpretive category results. Used in clinical laboratories by technicians. Results assist clinicians in determining appropriate antimicrobial therapy for Gram-positive infections.

## Clinical Evidence

Performance evaluated using fresh and stock clinical isolates and challenge strains compared to CLSI broth microdilution reference method. Results: 98.1% Essential Agreement (EA), 100.0% Categorical Agreement (CA), 0.0% Very Major Errors (VME), and 0.0% Major Errors (ME). Reproducibility and Quality Control testing demonstrated acceptable results.

## Technological Characteristics

Miniaturized broth microdilution card containing antimicrobial agents in specific concentrations. Sensing principle: optical growth monitoring. Connectivity: integrated with VITEK 2 and VITEK 2 Compact systems. Software: automated incubation and analysis.

## Regulatory Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

## Predicate Devices

- VITEK 2 AST-GP Ceftaroline ([K141149](/device/K141149.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Page 1 of 13

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
K190616

B. Purpose for Submission:
To obtain a substantial equivalence determination for dalbavancin for testing of Gram-positive organisms on the VITEK 2 and VITEK 2 Compact Antimicrobial Susceptibility Test (AST) Systems

C. Measurand:
Dalbavancin ≤0.015 – ≥1 μg/mL for Staphylococcus aureus (including methicillin-resistant isolates), Enterococcus faecalis (vancomycin-susceptible isolates only), and Streptococcus agalactiae.

D. Type of Test:
Automated quantitative or qualitative antimicrobial susceptibility test.

E. Applicant:
bioMérieux, Inc.

F. Proprietary and Established Names:
VITEK 2 AST-Gram Positive Dalbavancin (≤0.015 – ≥1 μg/mL)

G. Regulatory Information:
1. Regulation section:
21 CFR 866.1645: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System

{1}

2. Classification:

Class II

3. Product code(s):

LON – Fully automated short-term incubation cycle antimicrobial susceptibility system
LTW – Susceptibility Test Cards, Antimicrobial
LTT – Panels, Test, Susceptibility, Antimicrobial

4. Panel:

Microbiology, 83

H. Intended Use/Indications for Use:

1. Intended Use (s):

The VITEK 2 Gram-positive Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant Gram positive microorganisms to antimicrobial agents when used as instructed.

2. Indications for Use:

VITEK 2 AST-Gram Positive Dalbavancin is designed for antimicrobial susceptibility testing of Gram positive microorganisms and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 AST-Gram Positive Dalbavancin is a quantitative test. Dalbavancin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.

Active both in vitro and in clinical infections
Staphylococcus aureus (including methicillin-resistant isolates)
Enterococcus faecalis (vancomycin-susceptible isolates only)
Streptococcus agalactiae

The VITEK 2 Gram-positive Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant Gram positive microorganisms to antimicrobial agents when used as instructed.

Page 2 of 13

{2}

Page 3 of 13

3. Special conditions for use statement(s):

For Prescription Use Only

Limitations:

1. If the following antibiotic/organism combination is susceptible and vancomycin is intermediate or resistant, perform an alternative method of testing prior to reporting of results:

Dalbavancin: Staphylococcus aureus

2. The ability of the AST card to detect non-susceptible strains with the following combination(s) is unknown because an insufficient number of non-susceptible strains were available at the time of comparative testing:

Dalbavancin: Staphylococcus aureus, Streptococcus agalactiae, Enterococcus faecalis (vancomycin-susceptible)

4. Special instrument requirements:

VITEK 2 and VITEK 2 Compact Systems, VITEK 2 Systems (PC) version 10.1 or higher

I. Device Description:

The VITEK 2 AST card is a miniaturized, abbreviated and automated version of the doubling dilution technique for determining the minimum inhibitory concentration (MIC). Each VITEK 2 AST card contains 64 wells. A control well(s) which contain only nutrient medium is resident on all cards. The remaining wells contain premeasured portions of antimicrobials combined with the nutrient media. The isolate to be tested is diluted to a standardized concentration with 0.45% to 0.50% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System will automatically dilute the bacterial suspension to prepare an inoculum for susceptibility cards. Then the VITEK 2 will fill, seal and place the card into the incubator/reader. The VITEK 2 Compact has a manual filling, sealing, and loading operation. The VITEK 2 Systems monitor the growth of each well in the card over a defined period of time. The analysis program determines when a well demonstrates growth based on attenuation of light measured by an optical scanner. This data is used to determine the minimum inhibitory concentration or "MIC" values for the antimicrobial agent. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.

VITEK 2 AST-GP Dalbavancin has the following concentrations in the card: 0.0625, 0.125, 0.25, and 0.5 µg/mL (equivalent standard method concentration by efficacy in µg/mL). The Dalbavancin MIC result range for the VITEK 2 is ≤0.015 – ≥1 µg/mL for Staphylococcus aureus, Enterococcus faecalis, Streptococcus agalactiae. For all species, the VITEK 2

{3}

system is capable of reporting the following MIC results:  $\leq 0.015$ , 0.03, 0.06, 0.125, 0.25, 0.5 and  $\geq 1~\mu \mathrm{g / mL}$  for the AST-GP Dalbavancin test.

# J. Substantial Equivalence Information:

1. Predicate device name(s):

VITEK 2 AST-GP Ceftaroline

2. Predicate  $510(\mathrm{k})$  number(s):

K141149

3. Comparison with predicate:

Table 1: Comparison with Predicate Device

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device: VITEK 2 AST-GP Dalbavancin (K190616) | Predicate Device: VITEK 2 AST-GP Ceftaroline (K141149)  |
|  Intended Use/Indications for Use | VITEK 2 AST-Gram Positive Dalbavancin is designed for antimicrobial susceptibility testing of Gram positive microorganisms and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 AST-Gram Positive Dalbavancin is a quantitative test. Dalbavancin has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. | VITEK 2 Gram Positive Ceftaroline is designed for antimicrobial susceptibility testing of Gram positive microorganisms and is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 Gram Positive Ceftaroline is a quantitative test. Ceftaroline has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial.  |
|   |  Active in vitro and in clinical infections Staphylococcus aureus (including methicillin-resistant isolates) Enterococcus faecalis (vancomycin-susceptible isolates only) Streptococcus agalactiae | Active in vitro and in clinical infections Staphylococcus aureus (including methicillin-susceptible and -resistant isolates)  |
|   |  The VITEK 2 Gram-positive Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to | The VITEK 2 Antimicrobial Susceptibility Test (AST) is intended for use with the VITEK 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative  |

{4}

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device: VITEK 2 AST-GP Dalbavancin (K190616) | Predicate Device: VITEK 2 AST-GP Ceftaroline (K141149)  |
|   | determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed. | bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.  |
|  Test Method | Automated quantitative antimicrobial susceptibility test for use with the VITEK 2 and VITEK 2 Compact Systems to determine the in vitro susceptibility of microorganisms | Same  |
|  Inoculum | Saline suspension of organism | Same  |
|  Test Card | Gram Positive (AST-GP) Susceptibility Card | Same  |
|  Instrument | VITEK 2 and VITEK 2 Compact Systems | Same  |
|  Analysis Algorithm | Growth pattern analysis | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Antimicrobial Agent | Dalbavancin | Ceftaroline  |
|  Antimicrobial Concentrations | 0.0625, 0.125, 0.25, 0.5 μg/mL | 0.25, 0.5, 1, 2 μg/mL  |
|  Base Broths | GP13 | GP1  |

# K. Standard/Guidance Documents Referenced (if applicable)

FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009)
- CLSI M07-A10, "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard-Ninth Edition" Vol. 35 No. 2 (January 2015)
- CLSI M100, "Performance Standards for Antimicrobial Susceptibility Testing"; Twenty-eighth Edition (January 2018)

{5}

# L. Test Principle:

The VITEK 2 and VITEK 2 Compact Systems utilize automated growth-based detection using attenuation of light measured by an optical scanner. The optics in the systems uses visible light to directly measure organism growth within each of the 64 micro-wells. Transmittance optics is based on an initial light reading of a well before significant growth has begun. Every 15 minutes throughout the incubation cycle (defined period of time based on the VITEK 2 card), light transmittance readings of each well measures organism growth by the amount of light that is prevented from passing through the well. At the completion of the incubation period, the MIC values and their associated interpretive category results for each antimicrobial on the test card are displayed in an automatically generated report.

# M. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

# a. Precision/Reproducibility:

A reproducibility study for the VITEK 2 AST-GP card with Dalbavancin was conducted at three external clinical sites using a panel of ten isolates of Gram-positive microorganisms consistent with the indications for use. Testing was performed on three separate days and in triplicate for a total of 270 data points. The isolates tested in the reproducibility study included five Staphylococcus aureus and five Enterococcus faecalis isolates. Inocula were prepared using both the auto-dilution and manual dilution methods for testing in the VITEK 2 System. In addition, inocula were prepared by the manual dilution method only for use with the VITEK 2 Compact. The mode MIC value was determined and the reproducibility was calculated based on MIC values that fell within  $+ / -$  one doubling dilution from the mode MIC value. The data was analyzed taking into consideration best case and worst case scenarios as described in the Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems. All reproducibility performance results were considered acceptable; however,  $7 \%$  of the total isolates were off-scale. The reproducibility performance of the on-scale isolates is presented in Table 2.

Table 2: Reproducibility Performance

|  VITEK 2 |   | VITEK 2 Compact  |
| --- | --- | --- |
|  Auto-Dilution | Manual Dilution | Manual Dilution  |
|  99.63% | 99.63% | 97.41%  |

# b. Linearity/assay reportable range:

Not applicable

{6}

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

# Quality Control (QC) Testing

The CLSI recommended QC organisms (Enterococcus faecalis ATCC 29212 and Staphylococcus aureus ATCC 29213) were tested using both the VITEK 2 card and the reference method at each site. Both the automatic dilution and manual dilution methods were used for the VITEK 2 and the manual dilution method was used for the VITEK 2 Compact.

As shown in Table 3, both the auto-dilution and the manual dilution methods for VITEK 2 and the manual dilution for VITEK 2 Compact QC results were within the expected range  $&gt;95\%$  of the time.

Table 3: Quality Control Summary Results for VITEK 2 (Auto-Dilution and Manual Dilution Methods) and VITEK 2 Compact (Manual Dilution Method)

|  Organism | VITEK 2 Result Range | MIC Result (μg/mL) | VITEK 2 Auto Dilution | Ref | VITEK 2 Manual Dilution | Ref | VITEK 2 Compact Manual Dilution | Ref  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  E. faecalisATCC 29212Expected Result: 0.03-0.125 μg/mL |  | ≤ 0.008 |  |  |  |  |  |   |
|   |  * | 0.015 |  |  | 1 |  | 4 |   |
|   |  * | 0.03 | 181 | 72 | 181 | 66 | 182 | 65  |
|   |  * | 0.06 | 25 | 90 | 9 | 83 | 2 | 81  |
|   |  * | 0.125 |  | 37 |  | 35 |  | 35  |
|   |  * | 0.25 |  | 7 |  | 7 |  | 7  |
|   |  * | 0.5 |  |  |  |  |  |   |
|   |  * | 1 |  |  |  |  |  |   |
|   |   | 2 |  |  |  |  |  |   |
|   |   | ≥ 4 |  |  |  |  |  |   |
|  S. aureusATCC 29213Expected Result: 0.03-0.125 μg/mL |  | ≤ 0.008 |  |  |  |  |  |   |
|   |  * | 0.015 |  | 1 |  | 1 |  | 1  |
|   |  * | 0.03 | 134 | 123 | 141 | 113 | 147 | 113  |
|   |  * | 0.06 | 65 | 41 | 44 | 38 | 39 | 38  |
|   |  * | 0.125 | 1 | 35 |  | 33 |  | 33  |
|   |  * | 0.25 |  | 1 |  | 1 |  | 1  |
|   |  * | 0.5 | 1 |  | 1 |  |  |   |
|   |  * | 1 |  |  |  |  |  |   |
|   |   | 2 |  |  |  |  |  |   |
|   |   | ≥ 4 |  |  |  |  |  |   |

* denotes MIC result range produced by AST-GP Dalbavancin test Ref, CLSI reference method

# Inoculum Density Control

The DensiCHEK Plus was used to standardize the inoculum to a 0.5 McFarland standard. The instrument was standardized daily with all results recorded at each site. Calibration values were within the expected range.

{7}

Page 8 of 13

# Purity Check

A purity check of all organisms was performed on the dilution tube used to prepare the VITEK 2 card inoculum. Only those cultures that were pure were evaluated in the study.

# Growth Failure Rate

A total of 593 clinical isolates were evaluated. All 593 isolates grew in the VITEK 2 AST-GP Dalbavancin test. Complete test results are available for 593 isolates.

A challenge set of 75 isolates was evaluated at one site. All 75 challenge organisms grew in the VITEK 2 GP card with Dalbavancin using both the auto-dilution and manual dilution methods for the VITEK 2 and manual inoculation for the VITEK 2 Compact System.

A total of 668 VITEK 2 AST results were available.

d. Detection limit:

Not applicable

e. Analytical Specificity:

Not applicable

f. Assay cut-off:

Not applicable

2. Comparison studies:

a. Method comparison with predicate device:

Testing of Dalbavancin on the VITEK 2 AST-Gram Positive card was performed at three external sites and an internal site. Results obtained with the bioMérieux VITEK 2 AST-Gram Positive card with Dalbavancin were compared to results obtained with the CLSI broth microdilution reference panel. The MIC result range for the VITEK 2 AST-GP Dalbavancin is ≤0.015 – ≥1 μg/mL for all species. The reference panel contained two-fold serial dilutions with a range of ≤0.008 to ≥8 μg/mL. The testing conditions for the reference method consisted of the following:

- Medium – Cation Adjusted Mueller Hinton broth with appropriate dilutions of antimicrobial solution added
- Inoculum – Direct colony suspension
- Incubation – 35°C ambient air incubator; 20-24 hours Streptococci, 16-20 hours Staphylococci and Enterococci

{8}

The VITEK 2 AST cards were inoculated with test organisms using the auto-dilution method (VITEK 2 System) and using the manual dilution method (VITEK 2 System and VITEK 2 Compact). All test inocula used for the VITEK 2 AST cards and the reference method were standardized using the DensiCHEK Plus instrument.

A total of 593 clinical isolates were evaluated. The majority of isolates were recently isolated from clinical specimen (559 isolates, 94.3%). The remainder were stock isolates (34 isolates, 5.7%).

A total of 75 challenge organisms (41 Staphylococcus aureus, 22 Enterococcus faecalis, 12 Streptococcus agalactiae) were evaluated at one site. The challenge set was tested with the auto-dilution and manual dilution options of the VITEK 2 System and with the manual dilution method on the VITEK 2 Compact System.

The combined data set of clinical and challenge isolates included 668 AST results (346 Staphylococcus aureus, 260 Enterococcus faecalis, 62 Streptococcus agalactiae). A total of 4 (0.6%) non-susceptible isolates were tested among the clinical (n=593) and challenge (n=75) isolates.

The overall performance using the auto-dilution method of the VITEK 2 System demonstrated an EA of 98.1% and a CA of 100%. There were no very major errors and no major errors. With Staphylococcus aureus isolates, performance was acceptable with an EA of 97.7% and a CA of 100%. With Enterococcus faecalis isolates, performance was acceptable with an EA of 99.6% and a CA of 100%. With Streptococcus agalactiae isolates, performance was acceptable with an EA of 93.5% and a CA of 100%. The performance based on combined clinical and challenge data was acceptable. The overall performance is shown in Table 4.

Page 9 of 13

{9}

Table 4. Performance of Clinical and Challenge Isolates, VITEK 2 Auto-Dilution Method

|  Organism Type | EA Total | EA N | EA % | Eval EA Total | Eval EA N | Eval EA % | CA N | CA % | #NS | Min | Maj | Vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  E. faecalis  |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 238 | 237 | 99.6 | 210 | 209 | 99.5 | 238 | 100.0 | 2 | N/A | 0 | 0  |
|  Challenge | 22 | 22 | 100 | 15 | 15 | 100 | 22 | 100.0 | 1 | N/A | 0 | 0  |
|  Combined | 260 | 259 | 99.6 | 225 | 224 | 99.6 | 260 | 100.0 | 3 | N/A | 0 | 0  |
|  S. aureus  |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 305 | 298 | 97.7 | 299 | 292 | 97.7 | 305 | 100.0 | 1 | N/A | 0 | 0  |
|  Challenge | 41 | 40 | 97.6 | 41 | 40 | 97.6 | 41 | 100.0 | 0 | N/A | 0 | 0  |
|  Combined | 346 | 338 | 97.7 | 340 | 332 | 97.6 | 346 | 100.0 | 1 | N/A | 0 | 0  |
|  S. agalactiae  |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 50 | 46 | 92.0 | 7 | 3 | 42.9 | 50 | 100.0 | 0 | N/A | 0 | 0  |
|  Challenge | 12 | 12 | 100 | 0 | 0 | - | 12 | 100.0 | 0 | N/A | 0 | 0  |
|  Combined | 62 | 58 | 93.5 | 7 | 3 | 42.9 | 62 | 100.0 | 0 | N/A | 0 | 0  |
|  All species  |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 593 | 581 | 98.0 | 516 | 504 | 97.7 | 593 | 100.0 | 3 | N/A | 0 | 0  |
|  Challenge | 75 | 74 | 98.7 | 56 | 55 | 98.2 | 75 | 100.0 | 1 | N/A | 0 | 0  |
|  Combined | 668 | 655 | 98.1 | 572 | 559 | 97.7 | 668 | 100.0 | 4 | N/A | 0 | 0  |

EA - Essential Agreement (± 1 dilution)
Min - minor discrepancies
CA - Category Agreement
Maj - major discrepancies
Eval - Evaluable isolates
Vmj - very major discrepancies
NS - Non-susceptible isolates
N/A - Not applicable

Essential agreement (EA) occurs when the result of the reference method and that of the VITEK card are within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the reference method and the VITEK card. Category agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation provided by the VITEK card.

# Challenge Data -Manual Dilution

The challenge data set was also evaluated at one site with the manual dilution options of the VITEK 2 System and VITEK 2 Compact System. The performance was acceptable. Overall performance is shown in Table 5, and performances with  $E$  faecalis, S. aureus and S. agalacticae isolates are shown in Table 6, Table 7 and Table 8, respectively.

Table 5. Performance of Challenge Isolates, VITEK 2 and VITEK 2 Compact Manual Dilution - All Organisms

|  System | EA Total | EA N | EA % | Eval EA Total | Eval EA N | Eval EA % | CA N | CA % | #NS | Min | Maj | Vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  VITEK 2 | 75 | 75 | 100 | 56 | 56 | 100 | 75 | 100 | 1 | N/A | 0 | 0  |
|  VITEK 2 Compact | 75 | 74 | 98.7 | 54 | 53 | 98.1 | 75 | 98.7 | 1 | N/A | 0 | 0  |

Table 6. Performance of Challenge Isolates, VITEK 2 and VITEK 2 Compact Manual Dilution - E. faecalis

|  System | EA Total | EA N | EA % | Eval EA Total | Eval EA N | Eval EA % | CA N | CA % | #NS | Min | Maj | Vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  VITEK 2 | 22 | 22 | 100 | 15 | 15 | 100 | 22 | 100 | 1 | N/A | 0 | 0  |
|  VITEK 2 Compact | 22 | 22 | 100 | 13 | 13 | 100 | 22 | 100 | 1 | N/A | 0 | 0  |

{10}

Table 7. Performance of Challenge Isolates, VITEK 2 and VITEK 2 Compact Manual Dilution - S. aureus

|  System | EA Total | EA N | EA % | Eval EA Total | Eval EA N | Eval EA % | CA N | CA % | #NS | Min | Maj | Vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  VITEK 2 | 41 | 41 | 100 | 41 | 41 | 100 | 41 | 100 | 0 | N/A | 0 | 0  |
|  VITEK 2 Compact | 41 | 40 | 97.6 | 41 | 40 | 97.6 | 41 | 97.6 | 0 | N/A | 0 | 0  |

Table 8. Performance of Challenge Isolates, VITEK 2 and VITEK 2 Compact Manual Dilution - S. agalactiae

|  System | EA Total | EA N | EA % | Eval EA Total | Eval EA N | Eval EA % | CA N | CA % | #NS | Min | Maj | Vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  VITEK 2 System  |   |   |   |   |   |   |   |   |   |   |   |   |
|  VITEK 2 | 12 | 12 | 100 | 0 | 0 | - | 12 | 100 | 0 | N/A | 0 | 0  |
|  VITEK 2 Compact  |   |   |   |   |   |   |   |   |   |   |   |   |
|  COMPACT | 12 | 12 | 100 | 0 | 0 | - | 12 | 100 | 0 | N/A | 0 | 0  |

# MIC Trends:

An analysis of trending was conducted using the combined clinical and challenge data for each organism group. This trending calculation considers MIC values that are determined to be one or more doubling dilutions lower or higher compared to the reference method regardless of whether the device MIC values are on-scale. Results that are not clearly determined to be at least one dilution lower, at least one dilution higher or in exact agreement with the CLSI reference method are not considered in the trending analysis.

Trending results were stratified by species to assess species-related trends. Species for which the difference between the percentage of isolates with higher or lower readings was  $\geq 30$  with a statistically significant confidence interval were considered to show evidence of trending. There was no trending noted for  $E.$  faecalis,  $S.$  aureus and  $S.$  agalactiae.

b. Matrix comparison:

Not applicable

3. Clinical studies:

Not applicable

4. Clinical cut-off:

Not applicable

5. Expected values/Reference range:

{11}

Table 9: Interpretive Criteria for Dalbavancin (FDA STIC Webpage* and CLSI M100)

|  Organism | FDA Recognized Interpretive Criteria for Dalbavancin MIC (μg/mL)  |   |   |
| --- | --- | --- | --- |
|   |  S | I | R  |
|  Staphylococcus aureus
(including methicillin-resistant isolates) | ≤0.25 | - | -  |
|  Enterococcus spp.
(vancomycin-susceptible isolates only) | ≤0.25 | - | -  |
|  Streptococcus spp. β-Hemolytic Group | ≤0.25 | - | -  |

* FDA STIC Webpage
https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm

## N. Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device

## O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a breakpoint change protocol that was reviewed and accepted by FDA. This protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage (https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm). The protocol outlined the specific procedures and acceptance criteria that bioMérieux intends to use to evaluate the VITEK 2 Gram-positive Susceptibility Card when revised breakpoints for dalbavancin are published on the FDA STIC webpage. The breakpoint change protocol included with the submission indicated that if specific criteria are met, bioMérieux will update the dalbavancin device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.

Page 12 of 13

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON/K190616](https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON/K190616)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
