← Product Code [LON](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON) · K161510

# VTEK 2 AST-GN Ceftolozane/Tazobactam (<=0.25 - >= 32 mcg/mL) (K161510)

_bioMerieux, Inc. · LON · Feb 13, 2017 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON/K161510

## Device Facts

- **Applicant:** bioMerieux, Inc.
- **Product Code:** [LON](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON.md)
- **Decision Date:** Feb 13, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1645
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

VITEK® 2 Gram Negative Ceftolozane Tazobactam is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK® 2 Gram Negative Ceftolozane/Tazobactam is a quantitative test. Ceftolozane Tazobactam has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for this antimicrobial. Active in vitro and in clinical infections Enterobacter cloacae Escherichia coli Klebsiella oxytoca Klebsiella pneumoniae Proteus mirabilis Pseudomonas aeruginosa In vitro data available but clinical significance is unknown Citrobacter freundii Citrobacter koseri Enterobacter aerogenes Enterobacter cloacae Proteus vulgaris Providencia stuartii Serratia liquefacians The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococus spp., Enterococcus spp. and clinically significant yeast

## Device Story

VITEK 2 AST-GN Ceftolozane/Tazobactam is an automated, miniaturized, growth-based antimicrobial susceptibility test (AST) system. It uses 64-microwell cards containing premeasured antimicrobial concentrations (0.5/4 to 32/4 µg/mL). Input: pure culture organism suspension in saline, standardized to McFarland 0.5 via DensiCHEK Plus. Operation: system automatically fills, seals, and incubates cards; optical scanner measures light transmittance periodically over 4-18 hours to monitor growth. Output: MIC value and interpretive category (S, I, R). Used in clinical laboratories by technicians. Results assist clinicians in selecting appropriate antibiotic therapy for Gram-negative infections. Benefits include standardized, automated, and rapid susceptibility determination compared to manual broth microdilution.

## Clinical Evidence

Performance evaluated using 1,064 clinical and challenge isolates (Enterobacteriaceae and P. aeruginosa). Overall performance (automatic dilution): 94.8% Essential Agreement (EA), 98.4% Category Agreement (CA). Minor discrepancy rate 1%, major error 0.4%, very major error 1.4%. Reproducibility was 100%. Trending analysis for P. aeruginosa showed a bias toward higher MIC values compared to CLSI reference method, addressed via labeling.

## Technological Characteristics

Miniaturized broth microdilution card; contains antimicrobial agents in concentrations equivalent to standard mcg/ml methods. Automated incubation and optical growth monitoring system. Compatible with VITEK 2 and VITEK 2 Compact platforms. Software-driven analysis of growth kinetics.

## Regulatory Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

## Predicate Devices

- VITEK®2 AST-GN Doxycycline ([K121546](/device/K121546.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
K161510

B. Purpose for Submission:
To obtain a substantial equivalence determination for Ceftolozane/Tazobactam for testing non-fastidious gram negative organisms on the VITEK®2 and VITEK®2 Compact Antimicrobial Susceptibility Test (AST) Systems

C. Measurand:
The VITEK 2 AST-Gram Negative card contains the following concentrations of Ceftolozane/Tazobactam: 0.5/4, 1/4, 4/4, 8/4, and 32/4 µg/mL (equivalent standard method concentration by efficacy in µg/mL). The MIC result reporting range for the card is ≤0.25 - ≥32 µg/mL.

D. Type of Test:
Automated quantitative or qualitative antimicrobial susceptibility test for Ceftolozane/Tazobactam

E. Applicant:
bioMérieux, Inc.

F. Proprietary and Established Names:
VITEK 2 AST-GN Ceftolozane/Tazobactam (≤0.25 - ≥32 µg/mL)

G. Regulatory Information:
1. Regulation section:
221 CFR 866.1645 Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility System
2. Classification:
Class II
3. Product code(s):

{1}

LON - Fully automated short-term incubation cycle antimicrobial susceptibility system.
LTW - Susceptibility Test Cards, Antimicrobial
LTT - Panels, Test, Susceptibility, Antimicrobial

4. Panel:
Microbiology (83)

H. Intended Use:

1. Intended use(s):
The VITEK®2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK®2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.

2. Indication(s) for use:
VITEK®2 Gram Negative Ceftolozane/Tazobactam is designed for antimicrobial susceptibility testing of Gram negative bacilli and is intended for use with the VITEK®2 and VITEK®2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK®2 Gram Negative Ceftolozane/Tazobactam is a quantitative test. Ceftolozane/Tazobactam has been shown to be active against most strains of the microorganisms listed below, according to the FDA label for the antimicrobial.

Active in vitro and in clinical infections:
Enterobacter cloacae
Escherichia coli
Klebsiella oxytoca
Klebsiella pneumoniae
Proteus mirabilis
Pseudomonas aeruginosa

In vitro data available but clinical significance is unknown:
Citrobacter freundii
Citrobacter koseri
Enterobacter aerogenes
Enterobacter cloacae
Proteus vulgaris
Providencia stuartii
Serratia liquefaciens
The VITEK®2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK®2 Systems for the automated quantitative or qualitative susceptibility testing of

2

{2}

isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast.

3. Special conditions for use statement(s):

Prescription use only

Limitation:

"The ability of the AST card to detect resistance with the following combination(s) is unknown because an insufficient number of resistant strains were available at the time of comparative testing. If such a strain is observed, it should be submitted to a reference laboratory for further testing.

- Ceftolozane/Tazobactam: Citrobacter koseri, Proteus mirabilis, Proteus vulgaris, and Serratia Liquefaciens."

4. Special instrument requirements:

VITEK® 2 and VITEK® 2 Compact Systems

I. Device Description:

The VITEK 2 AST card is a miniaturized, abbreviated and automated version of the doubling dilution technique for determining the minimum inhibitory concentration (MIC). Each VITEK 2 test card contains 64 microwells. A control well containing only culture medium is included on all cards, with the remaining wells containing premeasured amounts of a specific antimicrobial agent in a culture medium base. A suspension of organism from a pure culture is prepared in a tube containing 0.45-0.5% sterile saline and standardized to a McFarland 0.5 using the DensiCHEK Plus™. The VITEK 2 System automatically fills seals and places the card into the incubator/reader; manual methods can also be used for the inoculation of test cards for use in the VITEK 2 System. The VITEK 2 Compact has a manual filling and sealing operation. The VITEK 2 Systems monitor the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antimicrobial contained on the card.

VITEK 2 AST-GN Ceftolozane/Tazobactam has the following concentrations in the card: 0.5/4, 1/4, 4/4, 8/4, and 32/4μg/mL (equivalent standard method concentration by efficacy in μg/mL). The MIC result range for the VITEK 2 card is ≤0.25 - ≥32 μg/mL.

J. Substantial Equivalence Information:

1. Predicate device name(s):

{3}

VITEK®2 AST-GN Doxycycline

2. Predicate 510(k) number(s):

K121546

3. Comparison with predicate:

Table 1: Comparison with the Predicate Device

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device
VITEK 2 AST-GN
Ceftolozane/Tazobactam
K161510 | Predicate
VITEK®2 AST-GN
Doxycycline
K121546  |
|  Intended Use | The VITEK®2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK®2 Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus spp. and clinically significant yeast. | Same  |
|  Test Method | Automated quantitative antimicrobial susceptibility test for use with the VITEK®2 and VITEK®2 Compact Systems to determine the in vitro susceptibility of Gram negative bacilli | Same  |
|  Inoculum | Saline suspension of organisms | Same  |
|  Test Card | VITEK®2 Gram Negative Susceptibility Test Card | Same  |
|  Instrument | VITEK 2 and VITEK 2 Compact Systems | Same  |
|  Detection Method | Light optics using optical scanner | Same  |
|  Report | Automatically generated and contains the MIC value with an interpretive category (S, I, R) for each antimicrobial on the test card | Same  |

4

{4}

|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Antimicrobial | Ceftolozane/Tazobactam | Doxycycline  |
|  Antimicrobial Concentration | 0.5/4, 1/4, 4/4, 8/4, and 32/4 μg/mL | 1, 4, 6 μg/mL  |
|  Algorithm Analysis | Growth pattern analysis-Unique to Ceftolozane/Tazobactam | Discriminate analysis-Unique to Doxycycline  |
|  Reporting Range | ≤0.25 - ≥32 μg/mL | ≤0.5 - ≥16 μg/mL  |

# K. Standard/Guidance Document Referenced (if applicable):

- CLSI M100-S24: Performance Standards for Antimicrobial Susceptibility Testing; Twenty-fourth Informational Supplement, Vol. 33 No. 1 (January 2014)
- CLSI M07-A9: Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard-Ninth Edition" Vol. 32 No, 2 (January 2012)
- Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009)

# L. Test Principle:

The VITEK 2 and VITEK 2 Compact Systems utilize automated growth-based detection using attenuation of light measured by an optical scanner. The optics used in the systems use visible light to directly measure organism growth. Transmittance optics is based on an initial light reading of a well before significant growth has begun. Periodic light transmittance samplings of the same well measure organism growth by how much light is prevented from going through the well. The VITEK 2 System monitors the growth of each well in the card over a defined period of time. An interpretive call is made between 4 and 16 hours for a "rapid" read but may be extended to 18 hours in some instances. At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic on the card.

# M. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

# a. Precision/Reproducibility:

A reproducibility study was conducted at three sites using ten isolates of gram negative bacilli that were consistent with the intended use. Isolates were tested in triplicate over three days for a total of 270 data points. The isolates tested in the reproducibility study included Pseudomonas aeruginosa (three isolates),  $E.$  coli (one isolate), Enterobacter aerogenes (two isolates), and Klebsiella pneumoniae (three isolates). Inocula were prepared both manually and using automatic dilution for

{5}

testing in the VITEK 2. Inocula were prepared manually for testing in the VITEK 2 Compact. The mode MIC value was determined and the reproducibility was calculated based on MIC values falling within ± 1 dilution of the mode MIC value.

Using VITEK 2 (automatic and manual dilution), and the VITEK 2 Compact (manual dilution), best case and worst case reproducibility was 100%.

The reproducibility results were acceptable.

b. Linearity/assay reportable range:

Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

**Inoculum Density Check:**

The inoculum density was monitored using the DensiCHEK Plus™ instrument during the trial. The DensiCHEK Plus™ was standardized daily with all results recorded at each site. Calibration values were within the expected range

**Purity Check:**

A purity check of all organisms was performed at the time of VITEK2 card inoculation. Only results obtained with pure cultures were evaluated.

**Growth Failure Rate:**

Out of 859 clinical isolates, 840 (97.78%) isolates grew in the VITEK 2 Ceftolozane/Tazobactam test.

All 224 challenge organisms grew in the VITEK 2 automatic dilution method and in the VITEK 2 Compact manual dilution method (100%). In the VITEK 2 manual dilution method, 223 of 224 (99.5%) of the challenge organisms grew.

**Quality Control (QC) Testing:**

Organisms recommended by both the FDA and the CLSI, namely Escherichia coli ATCC 25922, Escherichia coli ATCC 35218, Klebsiella pneumoniae ATCC 700603, and Pseudomonas aeruginosa ATCC 27853 were tested against Ceftolozane/Tazobactam. These recommended QC organisms were tested a minimum of 20 times/site by both the VITEK 2 AST-GN card and with the with CLSI broth microdilution reference method. Both the automatic dilution and the manual dilution methods were used for the VITEK 2 and the manual dilution method was used for the VITEK 2 Compact. The following table (Table 2) provides a summary of the QC results.

6

{6}

Table 2: Quality Control Results for VITEK 2 with Automatic and Manual Dilution Inoculation Methods and for VITEK 2 Compact with the Manual Dilution Inoculation Method.

|   |  | VITEK 2 Automatic-Dilution |   | VITEK 2 Manual Dilution |   | VITEK 2 Compact Manual Dilution  |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Organism | Conc.a(μg/mL) | Test | Ref. | Test | Ref. | Test | Ref.  |
|  E. coli ATCC 25922Expected Range(0.12-0.5 μg/mL) | ≤0.03 |  |  |  |  |  |   |
|   |  0.06 |  |  |  |  |  |   |
|   |  0.125 |  | 121 |  | 70 |  | 70  |
|   |  ≤0.25* | 225 | 101 | 112 | 42 | 112 | 42  |
|   |  0.5 |  | 3 |  |  |  |   |
|   |  1 |  |  |  |  |  |   |
|   |  2 |  |  |  |  |  |   |
|   |  4 |  |  |  |  |  |   |
|   |  8 |  |  |  |  |  |   |
|   |  16 |  |  |  |  |  |   |
|   |  32 |  |  |  |  |  |   |
|   |  ≥64 |  |  |  |  |  |   |
|  E. coliATCC 35218Expected Range(0.06-0.25 μg/mL) | ≤0.03 |  |  |  |  |  |   |
|   |  0.06 |  |  |  |  |  |   |
|   |  0.125 |  | 208 |  | 103 |  | 103  |
|   |  ≤0.25* | 220 | 12 | 107 | 4 | 107 | 4  |
|   |  0.5 |  |  |  |  |  |   |
|   |  1 |  |  |  |  |  |   |
|   |  2 |  |  |  |  |  |   |
|   |  4 |  |  |  |  |  |   |
|   |  8 |  |  |  |  |  |   |
|   |  16 |  |  |  |  |  |   |
|   |  32 |  |  |  |  |  |   |
|   |  ≥64 |  |  |  |  |  |   |
|  K. pneumoniaeATCC 700603Expected Range(0.5-2 μg/mL) | ≤0.03 |  |  |  |  |  |   |
|   |  0.06 |  |  |  |  |  |   |
|   |  0.125 |  |  |  |  |  |   |
|   |  0.25 |  |  |  |  |  |   |
|   |  0.5 | 0 | 25 | 0 | 9 | 0 | 9  |
|   |  1 | 225 | 193 | 112 | 98 | 112 | 98  |
|   |  2 | 0 | 7 | 0 | 5 | 0 | 5  |
|   |  4 |  |  |  |  |  |   |
|   |  8 |  |  |  |  |  |   |
|   |  16 |  |  |  |  |  |   |
|   |  32 |  |  |  |  |  |   |
|   |  ≥64 |  |  |  |  |  |   |
|  P. aeruginosaATCC 27853Expected Range | ≤0.03 |  |  |  |  |  |   |
|   |  0.06 |  |  |  |  |  |   |
|   |  0.125 |  |  |  |  |  |   |
|   |  ≤0.25 | 0 | 3 | 0 | 3 | 0 | 3  |

{7}

|  0.25- 1 μg/mL | 0.5 | 216 | 209 | 103 | 97 | 102 | 97  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |  1 | 4 | 8 | 4 | 7 | 4 | 7  |
|   |  2 |  |  |  |  |  |   |
|   |  4 |  |  |  |  |  |   |
|   |  8 |  |  |  |  |  |   |
|   |  16 |  |  |  |  |  |   |
|   |  32 |  |  |  |  |  |   |
|   |  ≥64 |  |  |  |  |  |   |

$^{a}$ MIC for Ceftolozane in the presence of a fixed concentration of 4 mg/L of Tazobactam.
*The lowest dilution of the VITEK 2 Ceftolozane/Tazobactam MIC range is 0.25 μg/mL. Obtaining ≤0.25 μg/mL was considered as an indicator that the quality control results were acceptable.

The expected range for *E. coli* ATCC 25922 with Ceftolozane/Tazobactam is 0.12 – 0.5 μg/mL. Even though the Ceftolozane/Tazobactam concentrations included in the VITEK 2 AST-Gram Negative card are 0.5/4, 1/4, 4/4, 8/4, and 32/4 μg/mL, the reporting range is ≤0.25 - ≥32 μg/mL. Therefore, all results for the QC strain were off-scale for the VITEK 2 and VITEK 2 Compact Systems as both VITEK systems report the lowest end of the scale as ≤0.25 μg/mL (See Table 2 above). *E. coli* ATCC 35218 was also tested to verify the performance of the device. Many results of this QC strain were also off scale (≤0.25 μg/mL). However, *Klebsiella pneumoniae* ATCC 700603 and *Pseudomonas aeruginosa* ATCC 27853 were also tested to verify the performance of the device and all results were on-scale (See Table 2 above).

The quality control results were acceptable.

Detection limit:

Not applicable

e. Analytical specificity:

Not applicable

f. Assay cut-off:

Not applicable

2. Comparison studies:

a. Method comparison with predicate device:

Results obtained with the bioMérieux VITEK 2 AST - Gram Negative card with Ceftolozane/Tazobactam were compared to results obtained with the CLSI broth microdilution reference panel. The VITEK 2 AST-Gram Negative card with Ceftolozane/Tazobactam contains the following concentrations of Ceftolozane/Tazobactam: 0.5/4, 1/4, 4/4, 8/4, and 32/4μg/mL (equivalent standard

{8}

method concentration by efficacy in $\mu \mathrm{g} / \mathrm{mL}$ and the reporting range is $\leq 0.25 - \geq 32$ $\mu \mathrm{g} / \mathrm{mL}$. The reference panel contained two-fold serial dilutions with a range of 0.0313 to $128~\mu \mathrm{g} / \mathrm{mL}$.

Test inocula were standardized using the DensiCHEK Plus instrument. VITEK 2 AST – Gram Negative cards were inoculated using automatic dilution (for reading on the VITEK 2 instrument) or using a manual dilution method (for reading on the VITEK 2 instrument or on the VITEK 2 COMPACT instrument). Reference panels were inoculated as outlined in the CLSI document M07-A9.

A total of 859 (Enterobacteriaceae and Pseudomonas aeruginosa) clinical isolates were evaluated at three sites with VITEK 2 AST – Gram Negative cards inoculated by automatic dilution and interpreted using the VITEK 2 instrument. 840 isolates grew in the VITEK 2 Ceftolozane/Tazobactam test. Complete test results are available for 840 in Table 3. The majority of isolates were fresh specimens (779/859 isolates, $90.68\%$); 80 of the clinical isolates tested (80/859 $9.3\%$) were stock isolates.

A total of 76 challenge isolates were tested at one external site. In response to a request from FDA, 148 additional challenge isolates were evaluated internally at bioMérieux as well as at an external site which resulted in a total of 224 challenge isolates. The challenge set was tested with both card inoculation options (automatic dilution and manual dilution) on the VITEK 2 System and with the manual dilution on the VITEK 2 COMPACT system.

Results from the 1064 clinical and challenge isolates are summarized in Table 3. The overall performance using the VITEK 2 System and inoculated using the automatic dilution method was acceptable, with an EA of $94.8\%$ and CA of $98.4\%$, and a minor discrepancy rate of $1\%$ (11/1064), a major error discrepancy rate of $0.4\%$ (4/907) and a very major error discrepancy rate of $1.4\%$ (2/141).

9

{9}

Table 3: Performance of Clinical and Challenge Isolates, VITEK 2 Automatic Dilution Method

|   | Tot | No. EA | EA % | Eval Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Enterobacteriaceae  |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 694 | 671 | 96.7 | 104 | 82 | 78.8 | 685 | 98.7 | 31 | 7 | 2 | 0  |
|  Challenge | 156 | 135 | 86.5 | 35 | 27 | 77.1 | 152 | 97.4 | 83 | 2 | 1 | 1  |
|  Combined | 850 | 806 | 94.8 | 139 | 109 | 78.4 | 837 | 98.5 | 114 | 9 | 3 | 1  |
|  Pseudomonas aeruginosa  |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 146 | 138 | 94.5 | 140 | 132 | 94.3 | 145 | 99.3 | 3 | 1 | 0 | 0  |
|  Challenge | 68 | 65 | 95.6 | 47 | 45 | 95.7 | 65 | 95.6 | 24 | 1 | 1 | 1  |
|  Combined | 214 | 203 | 94.9 | 187 | 177 | 94.7 | 210 | 98.1 | 27 | 2 | 1 | 1  |
|  All Organisms  |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 840 | 809 | 96.3 | 244 | 214 | 87.7 | 830 | 98.8 | 34.0 | 8 | 2 | 0  |
|  Challenge | 224 | 200 | 89.3 | 82 | 72 | 87.8 | 217 | 96.9 | 107 | 3 | 2 | 2  |
|  Combined | 1064 | 1009 | 94.8 | 326 | 286 | 87.7 | 1047 | 98.4 | 141 | 11 | 4 | 2  |

EA - Essential Agreement (+/- 2 dilutions)
CA - Category Agreement
EVAL - Evaluable isolates
R or NS - Resistant or non-susceptible isolates

min - minor discrepancies

maj - major discrepancies

vmj - very major discrepancies

Essential Agreement (EA) occurs when there is agreement between the result of the reference method and that of VITEK 2 test card within plus or minus one serial two-fold dilution of the antibiotic.

Evaluable results are those that are on scale for both the VITEK 2 test card and the reference method.

Category Agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the VITEK 2 test card.

# Enzyme Groups Characterization:

Enzyme characterization was conducted for Pseudomonas aeruginosa and for Enterobacteriaceae to cover the majority of beta lactamase enzyme groups and other resistance markers noted in the FDA drug label. The performance of the VITEK 2 and the VITEK 2 Compact (automatic and manual dilution) with

Ceftolozane/Tazobactam was evaluated using the characterized Pseudomonas aeruginosa and Enterobacteriaceae isolates.

# Pseudomonas aeruginosa:

Enzyme characterization was conducted for Pseudomonas aeruginosa isolates for the following ESBLs: chromosomal AmpC, and loss of OprD. Enzyme characterization was also performed for Pseudomonas aeruginosa isolates that produce serine carbapenemases [K. pneumoniae carbapenemase (KPC)] and metallo-beta lactamases.

During the additional testing, one Very Major Error was observed for Pseudomonas aeruginosa with a VMJ rate of  $3.7\%$  (1/27). This  $P_{\cdot}$  aeruginosa isolate lacked the OprD porin and contained OXA-101 enzyme which has been shown to confer

{10}

resistance to Ceftolozane/Tazobactam for *P. aeruginosa*. This VMJ was addressed by adding the following footnote in the labeling (Product Information Manual):

“One Pseudomonas aeruginosa isolate tested during the clinical evaluation lacked the OprD porin and contained the OXA-101 enzyme which has been reported as being resistant to Ceftolozane/Tazobactam and resulted in a Very Major Error for the VITEK® 2 AST-GN Ceftolozane/Tazobactam test”

*Pseudomonas* isolates with MexXY and MexAB up-regulation enzymes were not evaluated as this information was not available at the time of testing. This was addressed by adding the following footnote in the labeling (Product Information Manual):

“Enzyme group characterization was not available for the following organisms at the time of comparative testing, and therefore the performance of the AST card with Ceftolozane/Tazobactam is unknown: Pseudomonas aeruginosa (up-regulation of MexXY and MexAB)”.

## Enterobacteriaceae (ESBLs):

Enzyme characterization was conducted for claimed *Enterobacteriaceae* isolates for the following ESBLs: TEM, SHV and CTX-M. However, information on *Enterobacteriaceae* isolates with OXA enzymes was not available at the time of testing and was not provided. This was addressed by adding the following footnote in the labeling (Product Information Manual):

“Enzyme group characterization was not available for the following organisms at the time of comparative testing, and therefore the performance of the AST card with Ceftolozane/Tazobactam is unknown: Enterobacteriaceae (OXA)”

## Enterobacteriaceae (KPC and β-lactamases):

Enzyme characterization was also conducted for claimed *Enterobacteriaceae* isolates that produce serine carbapenemases [*K. pneumoniae* carbapenemase (KPC)]. However, testing was not conducted against *Enterobacteriaceae* isolates that are known to produce serine metallo-beta lactamases. Therefore, the following footnote was added in the labeling (Product Information Manual):

“Ceftolozane/Tazobactam is not active against Enterobacteriaceae bacteria that produce metallo-beta lactamases”.

## Resistant Organisms:

A total of 141 resistant organisms were identified out of 1064 organisms tested (13.2%) in the combined challenge and clinical study for Ceftolozane/Tazobactam for the VITEK 2 with the automatic dilution. However, the following organisms had an insufficient number of resistant strains available during the comparative study: *Citrobacter koseri*, *Proteus vulgaris* and *Serratia liquefaciens*. This was addressed by adding the following limitation in the package insert:

11

{11}

"The ability of the AST card to detect resistance with the following combination(s) is unknown because an insufficient number of resistant strains were available at the time of comparative testing. If such a strain is observed, it should be submitted to a reference laboratory for further testing.

- Ceftolozane/Tazobactam: Citrobacter koseri, Proteus mirabilis, Proteus vulgaris, and Serratia Liquefaciens

## Challenge Data

### VITEK 2 Manual Dilution:

The challenge set of 223 isolates evaluated using the VITEK 2 and inoculated using the manual dilution method demonstrated an EA of 90.1%, a CA of 96.0% and a minor discrepancy rate of 2.7% (6/223), a major error discrepancy rate of 1.8% (2/109) and a very major error discrepancy rate of 0.9% (1/107). A total of 81 isolates were determined to have evaluable results. Of these isolates, 71 were within essential agreement (EA) for a percent EA of evaluable isolates of 87.7%.

The results are summarized in Table 4. The results were acceptable.

Table 4: Performance of Challenge Isolates, VITEK 2 Manual Dilution Method

|   | Tot | No. EA | EA % | Eval Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Challenge | 223 | 201 | 90.1 | 81 | 71 | 87.7 | 214 | 96.0 | 107 | 6 | 2 | 1  |

### VITEK 2 COMPACT Manual Dilution:

The challenge set of 224 isolates was also evaluated using the VITEK 2 Compact and inoculated using the manual dilution method demonstrated an EA of 92.9%, a CA of 96.4% and a minor discrepancy rate of 2.7% (6/224), a major error discrepancy rate of 1.8% (2/109). There were no very major errors. A total of 83 isolates were determined to have evaluable results. Of these isolates, 73 were within essential agreement (EA) for a percent EA of evaluable isolates of 88.0%. The results are summarized in Table 5. The results were acceptable.

Table 5: Performance of Challenge Isolates, VITEK 2 Compact, Manual Dilution Method

|   | Tot | No. EA | EA % | Eval Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Challenge | 224 | 208 | 92.9 | 83 | 73 | 88.0 | 216 | 96.4 | 107 | 6 | 2 | 0  |

Even though the EA of evaluable was low, the overall performance of the VITEK 2 (manual dilution) and the VITEK 2Compact (manual dilution) were considered acceptable based on the acceptable performance in overall EA, the reproducibility study, and QC (Table 2).

{12}

13

# MIC Trends:

Using the combined clinical and challenge data for *Pseudomonas aeruginosa*, an analysis of trending was conducted. This trending calculation takes into account MIC values that are determined to be one or more doubling dilution lower or higher compared to the reference method irrespective whether the device MIC values are on-scale or not. The combined data for 189 results constitute the evaluable data for trend analysis which is presented in Table 6.

Table 6: Trending of Clinical and Challenge Isolate Results for the VITEK 2, Automatic Dilution

|  P. aeruginosa | # isolates | ≤1 dilution lower | Exact | ≥1 dilution higher  |
| --- | --- | --- | --- | --- |
|  Clinical | 141 | 8 | 75 | 58  |
|  Challenge | 48 | 9 | 28 | 11  |
|  Combined Clinical and Challenge | 189 | 17 | 103 | 69  |
|   |   | *8.99%
95% CI: (5.69% to 13.93) |  | *36.51%
95% CI: (29.98% to 43.58%)  |

*Difference: 27.51 95% CI: (19.33 to 35.32)

A higher MIC reading trend was observed in the overall performance of *Pseudomonas aeruginosa* compared to the CLSI broth microdilution reference method, which raises concerns for potential major errors. This trending and the potential for occurrence of major error(s) for Ceftolozane/Tazobactam when testing clinical and challenge isolate results with the VITEK 2 System, was addressed by adding the following footnote in the labeling (Product Information Manual):

“VITEK 2 Ceftolozane/Tazobactam MIC values tended to be in exact agreement or at least one doubling dilution higher when testing *P. aeruginosa* compared to the reference broth micro-dilution”.

b. Matrix comparison:

Not Applicable

3. Clinical studies:

a. Clinical Sensitivity:

Not Applicable

b. Clinical specificity:

Not Applicable

c. Other clinical supportive data (when a. and b. are not applicable):

Not Applicable

{13}

4. Clinical cut-off:

Not Applicable

5. Expected values/Reference range:

Table 7: FDA Interpretive Criteria for Ceftolozane/Tazobactam

|  Organism | MIC (μg/mL)  |   |   |
| --- | --- | --- | --- |
|   |  S | I | R  |
|  Enterobacteriaceae | ≤2/4 | 4/4 | ≥8/4  |
|  Pseudomonas aeruginosa | ≤4/4 | 8/4 | ≥16/4  |

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON/K161510](https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON/K161510)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
