BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM TIGECYCLINE (0.25-16 UG/ML)
K132909 · Becton, Dickinson and Company · LON · Feb 14, 2014 · Microbiology
Device Facts
Record ID
K132909
Device Name
BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM TIGECYCLINE (0.25-16 UG/ML)
Applicant
Becton, Dickinson and Company
Product Code
LON · Microbiology
Decision Date
Feb 14, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Intended Use
The BD Phoenix™ Automated Microbiology System is intended for in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram negative aerobic and facultative anaerobic bacteria belonging to the family Enterobacteriaceae and Non-Enterobacteriaceae.
Device Story
System performs rapid identification and antimicrobial susceptibility testing (AST) of bacterial isolates. Input: pure culture bacterial isolates inoculated into sealed, 136-well polystyrene panels containing biochemicals and antimicrobial agents. Operation: panels incubated at 35°C; instrument measures redox indicator changes and bacterial turbidity every 20 minutes. Output: MIC values and categorical interpretations (S, I, R, N). Used in clinical microbiology laboratories; operated by laboratory technicians. Results assist clinicians in selecting appropriate antimicrobial therapy for patients with bacterial infections.
Clinical Evidence
Clinical study compared Phoenix System results to CLSI reference broth microdilution method using clinical, stock, and challenge isolates. Performance metrics for Tigecycline (0.25-16µg/mL) on Gram-negative panels: Essential Agreement (EA) 97.5% (n=884) and Category Agreement (CA) 97.4% (n=884). Reproducibility study across three sites showed >95% agreement (+/- 1 dilution).
Technological Characteristics
Automated microbiology system; molded polystyrene panels with 136 micro-wells; redox indicator-based growth detection; turbidity measurement; incubation at 35°C +/- 1°C; automated or manual inoculum preparation; software-based interpretation of growth kinetics.
Indications for Use
Indicated for in vitro quantitative determination of antimicrobial susceptibility (MIC) of Gram-negative aerobic/facultative anaerobic bacteria (Enterobacteriaceae and Non-Enterobacteriaceae) and Gram-positive bacteria (Staphylococcus, Enterococcus, Streptococcus) from pure culture. Prescription use only.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
Vitek® Antimicrobial Susceptibility Test System (N50510)
K131331 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - VANCOMYCIN (0.5-32 UG/ML) · Becton, Dickinson & CO · Jul 10, 2013
K151320 — BD Phoenix Automated Microbiology System-Ertapenem 0.0625-8 mcg/ml · Becton, Dickinson and Company · Jan 15, 2016
K061355 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - AMPICILLIN (GN) 0.5-32 UG/ML · Becton, Dickinson & CO · Jun 7, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
\$\Upsilon 132909\$
# 510(k) SUMMARY
FEB 1 4 2014
- SUBMITTED BY: Becton, Dickinson and Company 7 Loveton Circle Sparks. MD 21152 Phone: 410-316-4975 Fax: 410-316-4188
- CONTACT NAME: Katie Covle Requlatory Affairs Specialist
DATE PREPARED: February 5, 2014
DEVICE TRADE NAME: BD Phoenix™ Automated Microbiology System -Tigecycline (0.25 - 16ug/mL)
- DEVICE COMMON NAME: Automated microbiology system
#### DEVICE CLASSIFICATION: 21 CFR 866.1645, Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Device, Product Code (LON)
#### PREDICATE DEVICES: VITEK® System (PMA No. N50510)
INTENDED USE: The BD Phoenix Automated Microbiology System is intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of Gram Negative aerobic and facultative anaerobic bacteria belonging to the family Enterobacteriaceae and non-Enterobacteriaceae.
# DEVICE DESCRIPTION:
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- . BD Phoenix instrument and software.
- . BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents or AST determinations.
- BD Phoenix ID Broth used for performing ID tests and preparing AST Broth . inoculum.
- . BD Phoenix AST Broth used for performing AST tests only.
- BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial . growth determination.
{1}------------------------------------------------
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. Phoenix panels are inoculated with a specified organism density and placed into the instrument. Inoculum for use with the Phoenix system may be prepared either manually or may be automated using the BD Phoenix™ AP System.
The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial aqents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35° ± 1°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, R or N (susceptible, intermediate, resistant or not susceptible).
# DEVICE COMPARISON:
The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance for inoculum prepared manually and inoculum prepared with the BD Phoenix™ AP instrument when compared with the CLSI reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Gram negative ID/AST or AST only Phoenix panels with this antimicrobial agent.
#### SUMMARY OF TESTING:
The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to CLSI M7). The system has been evaluated as defined in the FDA guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", August 28, 2009. Shelf life (stability data) for the drug is ongoing and will be maintained on file at BD. There have been no previous submissions for this formulation of drug for which FDA has provided feedback related to data or information needed to support substantial equivalence.
{2}------------------------------------------------
### Site Reproducibility
Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of Gram-negative isolates. Each site tested the isolates in triplicate on three different days using one lot of Gram Negative Phoenix panels containing this antimicrobial agent and associated reagents.
The results of the study demonstrate that for this antimicrobial agent and the Gramnegative organisms tested there was an overall reproducibility across test sites of greater than 95%
(+/- 1 dilution) agreement when compared to the test mode.
# Clinical Studies
Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with a Gram Negative Phoenix Panel containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.
The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within + one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, resistant or not susceptible).
The following table summarizes the performance for the isolates tested in this study.
| Antimicrobial | Concentration | EA (n) | EA (%) | CA (n) | CA (%) |
|---------------|----------------|--------|--------|--------|--------|
| Tigecycline | 0.25 - 16µg/mL | 884 | 97.5 | 884 | 97.4 |
#### Performance of BD Phoenix System for Gram-Negative Organisms by Drug
#### Conclusions Drawn from Substantial Equivalence Studies
The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems: Guidance for Industry and FDA", August 28, 2009. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three wavy lines that form a stylized human figure. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the symbol.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
BECTON, DICKINSON AND COMPANY KATIE COYLE 7 LOVETON CIRCLE MC 614 SPARKS MD 21152
February 14. 2014
Re: K132909
Trade/Device Name: BD Phoenix Automated Microbiology System Tigecycline (0.25-16 ug/ml)
Regulation Number: 21 CFR 866.1645 Regulation Name: Fully automated short-term incubation cycle antimicrobial susceptibility system Regulatory Class: II Product Code: LON Dated: December 19, 2013 Received: December 20, 2013
Dear Ms. Coyle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{4}------------------------------------------------
Page 2-Ms. Coyle
electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Uwe Scherf -S for
Sally A. Hojvat. M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known): K132909
#### Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent Tigecycline (0.25 - 16ug/mL) - Gram-negative ID/AST or AST only Phoenix Panels.
Indications for Use:
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus and Streptococcus.
This premarket notification is for the addition of the antimicrobial agent Tigecycline at concentrations of 0.25 - 16pg/mL to Gram-negative ID/AST or AST only Phoenix panels. Tigecycline has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
#### Active In Vitro and in Clinical Infections Against:
Citrobacter freundii Enterobacter cloacae Escherichia coli
Klebsiella oxvtoca Klebsiella pneumoniae
#### Active In Vitro
Serratia marcescens Citrobacter koseri Enterobacter aerogenes
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Image /page/5/Picture/16 description: The image shows the words "Ribhi Shawar" and the letters "FDA" in a stylized font. Below this is the date and time "2014.02.13 11:48:21 -05'00'". The letters "FDA" are in a larger font and are partially obscured by the date and time.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches