← Product Code [LON](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON) · K053186

# VITEK 2 GRAM POSITIVE TELITHROMYCIN FOR STREPTOCOCCUS PNEUMONIAE (K053186)

_bioMerieux, Inc. · LON · Dec 1, 2005 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON/K053186

## Device Facts

- **Applicant:** bioMerieux, Inc.
- **Product Code:** [LON](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON.md)
- **Decision Date:** Dec 1, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1645
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

VITEK® 2 Gram Positive Telithromycin for Streptococcus pneumoniae is designed for antimicrobial susceptibility testing of Streptococcus pneumoniae. VITEK 2 Gram Positive Telithromycin for Streptococcus pneumoniae is a qualitative test. It is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 and VITEK® 2 Compact Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gramnegative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae.

## Device Story

Device performs antimicrobial susceptibility testing (AST) for Streptococcus pneumoniae; utilizes VITEK 2 or VITEK 2 Compact systems. Input: bacterial isolate diluted in 0.45% saline; rehydrates antimicrobial medium in AST card. Operation: system automatically (VITEK 2) or manually (VITEK 2 Compact) fills/seals card; incubates/monitors growth in wells for up to 18 hours. Output: report containing MIC value and interpretive category result. Used in clinical laboratory settings by trained personnel. Output assists healthcare providers in selecting appropriate antimicrobial therapy for patients with S. pneumoniae infections.

## Clinical Evidence

Bench testing only. External evaluation conducted using fresh and stock clinical isolates and stock challenge strains. Performance compared to CLSI microbroth dilution reference method. Results demonstrated 99.5% overall Category Agreement. Reproducibility and Quality Control testing yielded acceptable results.

## Technological Characteristics

Miniaturized version of doubling dilution technique for MIC determination. AST cards contain antimicrobial medium. System uses automated incubation and optical growth monitoring. Compatible with VITEK 2 and VITEK 2 Compact instrument platforms.

## Regulatory Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

## Predicate Devices

- VITEK 2 Gram Positive Moxifloxacin for Streptococcus pneumoniae ([K031865](/device/K031865.md))

## Reference Devices

- VITEK 2 Compact ([K050002](/device/K050002.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo for BIOMÉRIEUX and the date DEC 1 2005. The logo is in a sans-serif font and is centered below a black circle. The date is printed above the logo in a smaller font size.

K 053/86

### 510(k) SUMMARY

## VITEK® 2 Gram Positive Telithromycin for Streptococcus pneumoniae

#### 510(k) Submission Information:

| Submitter's Name:    | bioMérieux, Inc.                                                                                    |
|----------------------|-----------------------------------------------------------------------------------------------------|
| Address:             | 595 Anglum Road<br>Hazelwood, MO 63042                                                              |
| Contact Person:      | Nancy Weaver<br>Staff Regulatory Affairs Specialist                                                 |
| Phone Number:        | 314-731-8695                                                                                        |
| Fax Number:          | 314-731-8689                                                                                        |
| Date of Preparation: | November 10, 2005                                                                                   |
| B. Device Name:      |                                                                                                     |
| Formal/Trade Name:   | VITEK® 2 Gram Positive Telithromycin for Streptococcus<br>pneumoniae                                |
| Classification Name: | Fully Automated Short-Term Incubation Cycle<br>Antimicrobial Susceptibility Device, 21 CFR 866.1645 |
| Common Name:         | VITEK 2 AST-GP Telithromycin for Streptococcus<br>pneumoniae                                        |
| C. Predicate Device: | VITEK 2 Gram Positive Moxifloxacin for Streptococcus<br>pneumoniae (K031865).                       |

VITEK® 2 Gram Positive Telithromycin for Streptococcus pneumoniae is designed for antimicrobial susceptibility testing of Streptococcus pneumoniae. It is intended for use with the VITEK 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The antimicrobial presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/ml. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.

The bacterial isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK 2 Compact has a manual filling and sealing operation. The VITEK 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.

VITEK 2 Gram Positive Telithromycin for Streptococcus pneumoniae demonstrated substantially equivalent performance when compared with the CLSI broth microdilution reference method, as

#### bioMérieux, Inc.

595 Anglum Road, Hazelwood, Missouri 63042-2320, USA Phone: 314/731-8500 800/638-4835 Fax: 314/731-8700

{1}------------------------------------------------

defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, Issued Feb. 5, 2003.

The Premarket Notification (510[k]) presents data in support of VITEK 2 Gram Positive Telithromycin for Streptococcus pneumoniae. An external evaluation was conducted with fresh and stock clinical isolates and stock challenge strains. The external evaluations were designed to confirm the acceptability of VITEK 2 Gram Positive Telithromycin for Streptococcus pneumoniae by comparing its performance with the CLSI microbroth dilution reference method. The data is representative of performance on both the VITEK 2 System and VITEK 2 Compact instrument platforms, as evidenced in the AST equivalency study presented in the VITEK 2 Compact 510(k), K050002. VITEK 2 Gram Positive Telithromycin for Streptococcus pneumoniae demonstrated acceptable performance of 99.5% overall Category Agreement. Reproducibility and Quality Control demonstrated acceptable results.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes extending from its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

DEC 1 2005 Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Nancy Weaver Staff Regulatory Affairs Specialist bioMérieux, Inc. 595 Anglum Road Hazelwood, MO 63042-2320

Re: k053186 Trade/Device Name: VITEK 2® Gram Positive Telithromycin for Streptococcus pneumoniae (<0.25 - >4 ug/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: November 10, 2005 Received: November 14, 2005

Dear Ms. Weaver:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

#### Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

Sales, a ton

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K053186

Device Name: VITEK® 2 Gram Positive Telithromycin for Streptococcus pneumoniae (< 0.25 - > 4 µg/ml)

Indications For Use:

VITEK® 2 Gram Positive Telithromycin for Streptococcus pneumoniae is designed for antimicrobial susceptibility testing of Streptococcus pneumoniae. VITEK 2 Gram Positive Telithromycin for Streptococcus pneumoniae is a qualitative test. It is intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents.

The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 and VITEK® 2 Compact Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gramnegative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae.

Prescription Use x (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Freddie A. Cook

Vision Sign-Off

Office of In Vitze Diagnostic Device Evaluation and Salety

510(k) K053181

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON/K053186](https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/LON/K053186)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
