The MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-positive staphylococci and enterococci. After inoculation, panels are incubated for 4.5 - 18 hours at 35°C +/- 1°C, in a WalkAway® SI, or equivalent, and read by the MicroScan® Instrumentation. Additionally, the panels may be incubated in a non-CO2 incubator and the AST portions can be read visually, according to the Package Insert.
Device Story
MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels are used for antimicrobial susceptibility testing (AST) of gram-positive cocci and Listeria. The device utilizes dehydrated antibiotic panels rehydrated with an inoculated broth suspension (turbidity method). The WalkAway®-SI system uses a colorimetric optics system (color wheel/lamp assembly and photosensor) to detect bacterial growth by measuring optical changes over time. The system performs periodic readings starting at 4.5 hours up to 16 hours for rapid results, or 16-20 hours for overnight results. The device provides quantitative MIC values and qualitative interpretations. It is intended for use in clinical laboratories by trained personnel. The output assists clinicians in selecting appropriate antimicrobial therapy for patients with bacterial infections.
Clinical Evidence
Bench testing only. Performance evaluated using fresh and stock clinical isolates and challenge strains. Compared against CLSI frozen reference panels. Essential Agreement for Trimethoprim/Sulfamethoxazole was >98%. Reproducibility and precision were confirmed using the WalkAway® System and turbidity inoculum preparation. Quality control testing met acceptable criteria.
Technological Characteristics
Miniaturized broth dilution panels; dehydrated antimicrobial agents in wells. Rehydrated with Synergies plus™ Pos Broth. Incubation at 35°C +/- 1°C. Compatible with WalkAway® SI automated instrumentation or visual reading. Utilizes rapid fluorogenic identification and dried overnight antimicrobial susceptibility testing technologies.
Indications for Use
Indicated for testing of Staphylococcus spp. with Trimethoprim/Sulfamethoxazole at concentrations between 0.5/9.5 to 8/152 µg/mL on gram-positive test panels for testing at 4.5-16 hours or 16-20 hours for overnight reading.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
{0}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K051350
B. Purpose for Submission:
To add Trimethoprim/Sulfamethoxazole to the MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels
C. Measurand:
Trimethoprim/Sulfamethoxazole at 0.5/9.5 to 8/152 µg/mL
D. Type of Test:
Quantitative and Qualitative growth based detection algorithm using optics light detection
E. Applicant:
Dade Behring Inc,
MicroScan®
F. Proprietary and Established Names:
MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels
G. Regulatory Information:
1. Regulation section:
866.1645 - Fully automated short-term incubation cycle antimicrobial susceptibility system
866.1640 - Antimicrobial Susceptibility Test Powder
2. Classification:
Class II
3. Product code:
LON – Automated AST system short incubation
{1}
LRG-Instrument for Auto Reader & Interpretation of Overnight Antimicrobial Susceptibility Systems
JWY - Manual Antimicrobial Susceptibility Test Systems
LTT – Panels, Test, Susceptibility, Antimicrobial
4. Panel:
83 Microbiology
H. Intended Use:
1. Intended use(s):
For use with MicroScan® Synergies plus™ Panels read on the WalkAway® -SI System (including upgraded WalkAway® -40 or WalkAway® -96 to meet WalkAway® SI equivalence). MicroScan® panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification to the species level of colonies, grown on solid media, of rapidly growing aerobic and facultative anaerobic gram-positive cocci and Listeria.
2. Indication(s) for use:
The testing of *Staphylococcus spp.* with Trimethoprim/Sulfamethoxazole at concentrations between 0.5/9.5 to 8/152 µg/mL on the gram-positive test panel for testing at 4.5-16 hours or 16-20 hours for an overnight reading.
3. Special conditions for use statement(s):
- Turbidity method of inoculum preparation only.
- For prescription use only.
4. Special instrument requirements:
Not Applicable
I. Device Description:
Each panel contains two control wells: a negative control well, and a growth control well (contains test medium without antibiotic). Antibiotics are diluted in water, buffer, or minute concentrations of broth to selected concentrations prior to dehydration of the panels. The panel is rehydrated and inoculated at the same time with 0.1 ml of suspension prepared by the turbidity method (inoculum prepared in 0.4% saline with PLURONIC®, then 0.1ml transferred to 25ml of inoculum Synergies plus Pos Broth with PLURONIC®) for a final inoculum concentration of 3-7 X 10⁵ CFU/ml. Panels are incubated in a Walk-Away® System and read periodically starting at 4.5 hours until sufficient growth to determine the MIC. Alternately the panels may be incubated at 35°C in a non-CO₂ for 16-20 hours and read by visual observation of growth.
2
{2}
J. Substantial Equivalence Information:
1. Predicate device name(s):
MicroScan® Dried Gram-Positive and Gram-Negative MIC/Combo Panels
2. Predicate 510(k) number(s):
k862140
k020185
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use | MicroScan® panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification to the species level of colonies, grown on solid media, of rapidly growing aerobic and facultative anaerobic organisms | Same |
| Specimen | Isolated colonies from culture used | Same |
| Inoculum | Inoculum density to 0.5 McFarland standard | Same |
| Incubation | <16 hours
16 – 24 hours | Same |
| Results | Quantitative with qualitative interpretations | Same |
| Technology | Growth based | same |
| Differences | | |
| Item | Device | Predicate |
| Panels | Dried Trimethoprim/Sulfamethoxazole in water | Dried clindamycin or gentamicin in broth |
| Reading | Uses both a <16 hour read capability and overnight read methods in the same system | Overnight system uses only the overnight reading method and <16 hour instruments use only the <16 hour read method |
| Inoculum preparation | Turbidity method of inoculation only. | Inoculum prepared from isolated colonies using either the Turbidity method or Prompt® system |
| Instrument | WalkAway® -SI System or equivalent | autoScan® -4 or WalkAway® |
| Antibiotic | Trimethoprim/Sulfamethoxazole at 0.5/9 to 8/152 μg/mL | Different concentrations depending on the antibiotic |
{3}
4
K. Standard/Guidance Document Referenced (if applicable):
Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA”; Clinical and Laboratory Standards Institute (CLSI) M7 (M100-S15) “Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard”.
L. Test Principle:
The WalkAway® SI uses a Colorimetric Optics System consisting of a color wheel/lamp assembly and a Photosensor. There is an initial read at 2.5 hours with a possible final read at 4.5, 5.5, 6.5, 8, 12, 16, or 18 hours (overnight instrument readings, manual readings) depending on the growth rate of the organism being tested. The time of final read is dependent on the growth rate of the organism and the sensitivity of the automatic reader since cell densities below 2 x 10⁷ cells/ml are not detected.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Reproducibility was demonstrated using 10 isolates tested at 3 sites on 3 separate days in triplicate. The study included the testing on the WalkAway® SI read at <16 hours, WalkAway® 16-18 hour readings and manual readings at 16-20 hours incubation. The WalkAway® SI had 2 results that were not readable at <16 hours. All results were >95% reproducible. Although reproducible results were obtained with all methods of reading the <16 hour readings produced a trend for a more susceptible result but with acceptable performance.
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The recommended Quality Control isolates were tested a sufficient number of times with acceptable results on all testing days with the reference method. All results were obtained in the 4.5-16 hour window when read using the <16 hour reading requirements. Quality control results demonstrated the ability of the different reading parameters (manual and instrument) to produce acceptable results. The following provides the frequency of the results in each concentration tested with the expected range stated.
{4}
| Results | | | | | |
| --- | --- | --- | --- | --- | --- |
| Organism | Conc in μg/mL | # reference | MicroScan® | | |
| | | | Manual overnight | Instrument overnight | <16 h instrument |
| E. faecalis ATCC 29212 Range ≤ 0.5/9.5 μg/mL | ≤ 0.5 | 84 | 84 | 84 | 84 |
| | 1 | | | | |
| | 2 | | | | |
| | 4 | | | | |
| | 8 | | | | |
| | >8 | | | | |
| | | | | | |
| S. aureus ATCC 29213 Range ≤ 0.5/9.5 μg/mL | ≤ 0.5 | 84 | 83 | 83 | 84 |
| | 1 | | 1 | 1 | |
| | 2 | | | | |
| | 4 | | | | |
| | 8 | | | | |
| | >8 | | | | |
Inoculum density control: A turbidity meter was used for the turbidity inoculation method. Turbidity inoculum verification provided.
d. Detection limit:
Not Applicable
e. Analytical specificity:
Not Applicable
f. Assay cut-off:
Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
Clinical testing was conducted at three sites using fresh isolates supplemented with stock isolates. A total of 376 gram-positive isolates were tested of which 263 were fresh isolates. There were 75 challenge isolates tested at one site and compared to the reference broth dilution result mode that was determined by previous testing of each isolate multiple times in the recommended reference panel. The Synergies plus™ readings were obtained at times between 4.5 and 16 hours of incubation for >95% of the results. An additional comparison was done with readings on the instrument after overnight incubation and also read manually when incubated 16 - 18 hours. Performance by these alternate reading methods
{5}
was also acceptable with little differences except for a slight trend for the <16 hour readings to be slightly more susceptible. The recommended CLSI reference method was followed with the exception of the use of a small amount (0.1%) of Pluronic in the final inoculum. A validation of the use of Pluronic in the frozen reference panels was conducted, demonstrating little or no difference.
The chart below demonstrates the performance of all three reading methods (Synergies plus™ readings at <16 hours, overnight on the WalkAway® and manually read at 18 hours using the touchScan®-SR) when compared to the reference method. For the <16 hour read results, all of the major and very major errors except for one were within EA. Since there is no intermediate range the category error will either be a major or a very major error. Considering the error of the reference method and test (both with a +/- one dilution) these would be acceptable.
| | total | EA | %EA | Total evaluable | EA of evaluable | %EA | CA | %CA | #R | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| < 16h | 373 | 371 | 99.5 | 51 | 50 | 98.0 | 365 | 97.9 | 70 | NA | 5 | 3 |
| Overnight instrument | 376 | 373 | 99.2 | 49 | 49 | 100 | 371 | 98.7 | 71 | NA | 3 | 2 |
| Overnight manual | 376 | 373 | 99.2 | 48 | 46 | 95.8 | 369 | 98.1 | 71 | NA | 6 | 1 |
EA-Essential Agreement
CA-Category Agreement
R-resistant isolates
NA – Not applicable since there is no intermediate result reported
maj-major discrepancies
vmj-very major discrepancies
min- minor discrepancies
Evaluable results are those that fall within the test range of the reference method and could also be on-scale with the new device if within the plus/minus one dilution variability. EA is when there is agreement between the reference method and the MicroScan® within plus or minus one serial two-fold dilution of antibiotic. CA is when the interpretation of the reference method agrees exactly with the interpretation of the MicroScan® result.
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
{6}
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
Staphylococcus spp. interpretive criteria:
$\leq 2/38$ (Susceptible), $\geq 4/76$ (Resistant)
All values are included in the package insert.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.