The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.
Device Story
The BD Phoenix System is an automated microbiology platform for rapid identification and antimicrobial susceptibility testing (AST) of bacterial isolates. The system utilizes sealed, self-inoculating polystyrene panels containing 136 micro-wells with dried biochemicals and antimicrobial agents. Input consists of a pure culture bacterial isolate prepared in broth to a 0.5 McFarland standard. The system continuously incubates panels at 35°C, performing automated optical readings every 20 minutes. It employs a redox indicator and turbidity measurements to detect bacterial growth in the presence of antimicrobial agents. The system software interprets these readings to determine minimum inhibitory concentration (MIC) values and categorical interpretations (S, I, R, or N). Results are provided to clinicians to guide antimicrobial therapy and improve patient outcomes. The system is intended for use in clinical laboratory settings by trained laboratory personnel.
Clinical Evidence
Performance evaluated via clinical, stock, and challenge isolates across multiple U.S. sites. Comparison performed against CLSI reference broth microdilution method. Metrics included Essential Agreement (EA) and Category Agreement (CA). For vancomycin, EA was 98.2% (n=1939). Intra-site reproducibility >90% and inter-site reproducibility >95% for streptococcal isolates.
Technological Characteristics
Automated broth microdilution system. Uses colorimetric oxidation-reduction redox indicator. Materials: molded polystyrene trays with 136 micro-wells. Energy: 35°C incubation. Connectivity: Automated instrument with software-driven expert system. Software: Rule-based interpretation using CLSI standards. Sterilization: Not specified.
Indications for Use
Indicated for in vitro rapid identification and quantitative antimicrobial susceptibility testing (MIC) of Streptococcus species isolates from pure culture. For prescription use only.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
VITEK® System (PMA No. N50510)
BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321)
BD Phoenix™ Automated Microbiology System with Ofloxacin (K020323)
BD Phoenix™ Automated Microbiology System with Levofloxacin (K020322)
Related Devices
K040106 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM -- AMPICILLIN -- GRAM POSITIVE · Becton, Dickinson & CO · Mar 11, 2004
K023895 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - CIPROFLOXACIN GN · Becton, Dickinson & CO · Jan 8, 2003
K061869 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-PIPERACILLIN-TAZOBACTAM (GP) 1/4-128/4 UG/ML · Becton, Dickinson & CO · Aug 18, 2006
K050089 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - AMPICILLIN-SULBACTAM (GP) 2-32 UG/ML · Becton, Dickinson & CO · Feb 28, 2005
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k050745
B. Purpose for Submission:
Addition of vancomycin to the BD Phoenix™ SMIC/ID and SMIC Panels
C. Measurand:
Vancomycin 0.0625 - 32 µg/ml
D. Type of Test:
Antimicrobial Susceptibility Test (AST) (Quantitative and Qualitative) colorimetric
oxidation-reduction, growth-based
E. Applicant:
Becton, Dickinson & Company
F. Proprietary and Established Names:
BD Phoenix™ Automated Microbiology System – Vancomycin 0.0625 - 32 µg/ml
G. Regulatory Information:
1. Regulation section:
21 CFR 866.1645 Fully Automated Short-Term Incubation Cycle Antimicrobial
Susceptibility System
2. Classification:
II
3. Product code:
LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
4. Panel:
83 Microbiology
H. Intended Use:
1. Intended use(s):
The BD Phoenix™ Automated Microbiology System is intended for in vitro
quantitative determination of antimicrobial susceptibility by minimal inhibitory
concentration (MIC) of gram-negative aerobic and facultative anaerobic bacteria
belonging to the family Enterobacteriaceae and non – Enterobacteriaceae and
most gram-positive bacteria isolates from pure culture belonging to the genera
{1}
Staphylococcus, Enterococcus and Streptococcus.
The BD Phoenix™ SMIC/ID and SMIC Panel is intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most bacteria isolates from pure culture belonging to the genera Streptococcus.
2. Indication(s) for use:
This submission is for the addition of the antibiotic vancomycin at concentrations of 0.0625 - 32 µg/mL to the Streptococcus susceptibility panel.
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Not Applicable
I. Device Description:
The BD Phoenix™ Automated Microbiology System includes instrumentation and software, sealed and self-inoculating molded polystyrene trays with 136 micro-wells containing dried reagents, and specific inoculum broth formulations for ID and AST-S Indicator. The organism to be tested must be a pure culture and be preliminarily identified as gram positive or gram negative. Colonies are then suspended in broth, and equated to a 0.5 McFarland with the recommendation to use the BD CrystalSpec™ Nephelometer. A further dilution is made into an AST-S broth, which contains an AST-S indicator, prior to inoculating the panel. The AST-S broth is a non-blood, cation-adjusted broth containing purified water, Tween 80, pancreatic digest of casein, peptones and other additional supplements for optimization of streptococcal growth. After adding the indicator solution to the AST-S inoculum the color is blue and after inoculation and incubation goes to pink to colorless as reduction in the panel well proceeds. Inoculated panels are barcode scanned and loaded into the BD Phoenix™ Automated Microbiology System instrument where the panels are continuously incubated at 35°C. The resulting AST has a final inoculum of 5 × 10⁵ CFU/ml. The instrument incubates, reads and records the results of the biochemical substrates and antimicrobial agents and interprets the reactions to give an ID of the isolate and MIC value and category interpretation of the antimicrobial agents. Organisms growing in the presence of a given antimicrobic agent reduce the indicator, signaling organism growth and resistance to the antimicrobic agent. Organisms killed or inhibited by a given antimicrobic do not cause reduction of the indicator and therefore do not produce a color change. Additional interpretation is done using software driven "EXPERT" System using rules derived from the Clinical and Laboratory Standards Institute (CLSI).
Readings are taken every 20 minutes with an ID result available between 2-12 hours and an AST result available between 4-16 hours. This is only an autoread result; there are no manual readings possible.
{2}
J. Substantial Equivalence Information:
1. Predicate device name(s): VITEK® System
2. Predicate 510(k) number(s): N50510
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| 1. Intended Use | Intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most bacteria. | Same |
| 2. Isolates | Isolated colonies from culture used | Isolated colonies from culture used |
| 3. Result Reported | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) |
| 4. Incubation Time | <16 hours | <16 hours |
| 5. Type of Test | Automated | Automated |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| 1. Results achieved | Results are determined from serial twofold dilutions of antimicrobial agents | Results are determined from extrapolation of doubling dilutions |
| 2. Sample Preparation | Inoculum density equated to 0.5 McFarland standard | Inoculum density equated to 1.0 McFarland standard |
| 3. Technology | Automated growth based enhanced by use of a redox indicator (colorimetric oxidation-reduction) to detect organism growth. | Automated growth based with detection using an attenuation of light measured by an optical scanner. |
{3}
K. Standard/Guidance Document Referenced (if applicable):
“Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test Systems; Guidance for Industry and FDA”; CLSI M7 (M100-S15) “Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard.”
L. Test Principle:
The system employs conventional, colorimetric, fluorogenic and chromogenic substrates to identify the genus and species of the isolate. The AST portion of the BD Phoenix™ Automated Microbiology System is a broth based microdilution method that utilizes a redox indicator (colorimetric oxidation-reduction) to enhance detection of organism growth. The MIC is determined by comparing growth in wells containing serial two-fold dilutions of an antibiotic to the growth in “growth control wells” which contains no antibiotic.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Thirty three gram-positive on-scale organisms were evaluated for site to site and inter site reproducibility demonstrating >95% reproducibility. The ten isolate study described in the guidance document was used (10 organisms tested 3 times on 3 days at 3 sites).
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The CLSI recommended QC isolate, S. pneumoniae ATCC 49619 was tested on every test occasion with the reference method and the BD Phoenix™. The reference method QC results were in range for every day tested. The BD Phoenix™ was tested a sufficient number of times to demonstrate that the system can produce QC results in the CLSI recommended ranges most of the time. Both QC organisms had the same mode with the reference and the BD Phoenix™.
Quality Control Table
| ORGANISM | conc. | Reference | | | Phoenix | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | | | | |
| S. pneumoniae | 0.125 | | | | | 28 | |
| ATCC 49619 | 0.25 | | 123 | | | 95 | |
| Expected Range: | 0.5 | | 2 | | | | |
| 0.125 – 0.5 μg/mL | | | | | | | |
| | | | | | | | |
{4}
Inoculum density control: The organism suspension density of the ID broth was equivalent to a 0.5 McFarland standard using the BBL™ CrystalSpec™ Nephelometer which was verified each day of testing. Internal data was used to demonstrate that the use of the BBL™ CrystalSpec™ Nephelometer would produce reproducible results. Five different instruments were used. Five Streptococcal strains were evaluated to demonstrate acceptable reproducibility performance.
d. Detection limit: Not Applicable
e. Analytical specificity: Not Applicable
f. Assay cut-off: Not Applicable
2. Comparison studies:
a. Method comparison with predicate device: The CLSI recommended broth dilution reference panel was prepared according to the CLSI recommendation. Clinical testing was performed at four sites. The broth reference panel was set up on MH supplemented with 2 - 5% lysed horse blood as recommended by CLSI. The test device had a growth rate of >95%. The testing included both fresh clinical isolates and stock isolates along with a challenge set with known results. A comparison was provided to the reference method with the following agreement.
Summary Table for Streptococcus spp.
| | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | NS |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Clinical | 1815 | 1782 | 98.2 | 1810 | 1782 | 98.5 | 1812 | 99.8 | 1 |
| Challenge | 124 | 123 | 99.2 | 124 | 123 | 99.2 | 124 | 100 | 0 |
| Combined | 1939 | 1905 | 98.2 | 1934 | 1905 | 98.5 | 1936 | 99.8 | 1 |
EA-Essential Agreement
NS-Not Susceptible
CA-Category Agreement
Essential agreement (EA) is when the BD Phoenix™ panels agree with the reference test panel results exactly or within one doubling dilution of the reference method. Category agreement (CA) is when the BD Phoenix™ panel result interpretation agrees exactly with the reference panel result interpretation. Eval EA is when the MIC result is on scale for both the BD Phoenix™ and the reference and have on-scale EA.
There appears to be a trend where the reference device is slightly more resistant than the test device as reflected in the Accuracy and QC studies however results are
{5}
still within essential agreement.
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
Streptococcus pneumoniae ≤1 (S)
Streptococcus spp. other than S. pneumoniae ≤1 (S)
The ability of the BD Phoenix™ system to detect resistance to vancomycin in Streptococcus organisms is unknown because resistant organisms were not available at the time of comparative testing.
The Interpretative criteria, QC isolates and the expected ranges are the same as recommended by CLSI. All values will be included in the package insert.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.