BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - CONFIRMATORY ESBL TEST
K033458 · Becton, Dickinson & CO · LON · May 28, 2004 · Microbiology
Device Facts
Record ID
K033458
Device Name
BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - CONFIRMATORY ESBL TEST
Applicant
Becton, Dickinson & CO
Product Code
LON · Microbiology
Decision Date
May 28, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Intended Use
The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.
Device Story
System uses sealed, self-inoculating polystyrene trays with 136 micro-wells containing dried reagents; requires pure culture isolates suspended in broth (0.5 McFarland standard). Instrument continuously incubates panels at 35°C; performs automated colorimetric oxidation-reduction growth-based detection. Readings taken every 20 minutes via barcode scanning. Algorithm compares relative growth rates in test wells (containing cephalosporins +/- clavulanate) against growth control wells using a decision tree; interprets ESBL presence/absence. Expert system applies NCCLS-derived rules for final categorical interpretation. Output provides ID and ESBL status; if positive, beta-lactam drugs (except carbapenems) are reported as resistant. Used in clinical microbiology labs; operated by laboratory technicians. Benefits include rapid (4-16 hours) automated ESBL detection compared to manual methods.
Clinical Evidence
Performance evaluated using clinical and challenge isolates across multiple US sites. Confirmatory ESBL test accuracy: Positive Percent Agreement 96.8% (183/189), Negative Percent Agreement 96.1% (780/812), Overall Percent Agreement 96.2% (963/1001). Reproducibility studies showed intra-site reproducibility >90% and inter-site reproducibility >95%. Comparison made against NCCLS reference broth microdilution method.
Technological Characteristics
Polystyrene trays with 136 micro-wells; dried reagents; cation-adjusted Mueller-Hinton broth with 0.01% Tween 80; redox indicator (blue to pink/colorless). Automated incubation at 35°C. Colorimetric oxidation-reduction sensing. Software-driven decision tree algorithm and Expert system. No manual readings.
Indications for Use
Indicated for in vitro quantitative determination of antimicrobial susceptibility (MIC) of Gram-negative aerobic/facultative anaerobic bacteria (Enterobacteriaceae and Non-Enterobacteriaceae) and Gram-positive bacteria (Staphylococcus and Enterococcus) from pure culture. Includes confirmatory ESBL test for detection of extended spectrum beta-lactamase producing Escherichia coli, Klebsiella pneumoniae, and Klebsiella oxytoca.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
Sensititre® ESBL confirmatory test plate (K031545)
BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321)
BD Phoenix™ Automated Microbiology System with Ofloxacin (K020323)
BD Phoenix™ Automated Microbiology System with Levofloxacin (K020322)
Related Devices
K031429 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-MEROPENEM 0.25-16 UG/ML · Bd Diagnostic Systems · Jun 6, 2003
K021954 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-CLINDAMYCIN 0.125-8.0MCG/ML · Becton, Dickinson & CO · Aug 1, 2002
K060214 — BD PHOENIX AUTOMATED MIRCROBIOLOGY SYSTEM TETRACYCLINE-GN 0.5-16 UG/ML AND GP 0.5-16 UG/ML · Becton, Dickinson & CO · Mar 9, 2006
K031984 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-TICARCILLIN-1-128 UG/ML · Becton, Dickinson & CO · Aug 27, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
# K033458
| MAY 2 8 2004 | 510(K) SUMMARY |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTED BY: | Becton, Dickinson and Company<br>7 Loveton Circle<br>Sparks, MD 21152<br>Phone: (410) 316-4287<br>Fax: 410-316-4499 |
| CONTACT NAME: | Monica Evelyn Giguere<br>Regulatory Affairs Specialist |
| DATE PREPARED: | May 27, 2004 |
| DEVICE TRADE NAME: | BD Phoenix™ Automated Microbiology System -<br>confirmatory ESBL Test |
| DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation |
| DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation Cycle<br>Antimicrobial Susceptibility Device, 21 CFR 866.1645 |
| PREDICATE DEVICES: | Sensititre® ESBL confirmatory test plate (K031545) and<br>BD Phoenix™ Automated Microbiology System with<br>Gatifloxacin (K020321, May 23, 2002), Ofloxacin<br>(K020323, April 14, 2002), and Levofloxacin (K020322,<br>March 27, 2002). |
| INTENDED USE: | The BD Phoenix™ Automated Microbiology System is<br>intended for the rapid identification and in vitro<br>antimicrobial susceptibility testing of isolates from pure<br>culture of most aerobic and facultative anaerobic Gram-<br>negative and Gram-positive bacteria of human origin. |
{1}------------------------------------------------
### DEVICE DESCRIPTION:
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- BD Phoenix instrument and software. .
- BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial . agents for AST determinations.
- BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
- BD Phoenix AST Broth used for performing AST tests only. .
- BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination.
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID Broth.
The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S. I, or R (sensitive, intermediate, or resistant).
#### DEVICE COMPARISON:
The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the NCCLS reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Gram-negative ID/AST or AST only Phoenix panels with the confirmatory ESBL test.
#### SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:
The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the NCCLS reference broth microdilution method (AST panels prepared according to NCCLS M7).
{2}------------------------------------------------
#### Site Reproducibility
Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of Gram-negative isolates. Each site tested the isolates in triplicate on three different days using one lot of Gram-negative Phoenix panels containing this test and associated reagents.
The results of the study demonstrate for the this test there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the Gram-negative isolates tested.
#### Clinical Studies
Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Gram-negative Phoenix panel format containing this test. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the NCCLS reference broth microdilution method.
#### Confirmatory ESBL Test Performance
To determine the accuracy of the Phoenix Confirmatory ESBL test, accuracy testing was performed at multiple sites using Clinical and Challenge isolates. The results from the ESBL test resident on the Phoenix panels were compared to the results obtained from the reference confirmatory ESBL test.
For Challenge organisms this result is an expected result and for Clinical isolates this result was obtained from concurrent testing in the NCCLS reference broth microdilution method. Additionally, a challenge set of 30 previously characterized organisms was tested at one site.
> Positive Percent Agreement = 183/189 = 96.8% Negative Percent Agreement = 780/812 = 96.1% Overall Percent Agreement = 963/1001 = 96.2%
{3}------------------------------------------------
## Conclusions Drawn from Substantial Equivalence Studies
The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this test is substantially equivalent. Technological characteristics of this system are substantially equivalent to those used in the Sensititre® ESBL confirmatory test plates, which received clearance by the FDA under K031545 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular fashion around the left side of the logo.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAY 2 8 2004
Ms. Monica E. Giguere Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152
k033458 Re: Trade/Device Name: BD Phoenix™ Automated Microbiology System Confirmatory ESBL Test Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: April 6, 2004 Received: April 7, 2004
Dear Ms. Giguere:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (sec above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
#### Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincercly yours.
Sagahts
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of 1
510(k) Number: K033458
Device Name: BD Phoenix™ Automated Microbiology System for use with the confirmatory ESBL Test - Gram-negative ID/AST or AST only Phoenix panels.
Indications for Use:
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anacrobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.
This premarket notification is for the addition of the confirmatory ESBL Test to Gram-negative ID/AST or AST only Phoenix panels. The Phoenix confirmatory ESBL Test is a confirmatory test for the detection of the organisms that produce extended spectrum 3-lactamses (ESBL) in Escherichia coli, Klebsiella pneumoniae, and Klebsiella oxytoca.
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
SAKWA
Division Siging
Office of the Millettral Diginostic Device Evaratizitation rath Safety
510(k) \$\frac{K03245K}{}\$
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.