The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.
Device Story
BD Phoenix™ Automated Microbiology System performs automated antimicrobial susceptibility testing (AST) and identification (ID). System uses sealed, self-inoculating polystyrene trays with 136 micro-wells containing dried reagents. Pure culture isolates are suspended in broth, equated to 0.5 McFarland standard, and inoculated into panels. Instrument continuously incubates panels at 35°C. Growth detection uses redox indicator (colorimetric oxidation-reduction); indicator changes from blue to pink/colorless in presence of bacterial growth. Instrument reads biochemical substrates and antimicrobial agents every 20 minutes. Software-driven 'EXPERT' system interprets reactions to provide MIC values and categorical interpretations (SIR) based on NCCLS rules. System provides results in 4-16 hours. No manual readings possible. Benefits include rapid, automated susceptibility determination to guide clinical antibiotic therapy.
Clinical Evidence
Performance evaluated via multi-site clinical study (6 sites) comparing BD Phoenix to NCCLS broth dilution reference method. Combined dataset (n=838) showed 98.1% Essential Agreement (EA) and 98.3% Category Agreement (CA). Very major discrepancy rate was 0.7% (3/427 resistant isolates). Reproducibility >95% across sites and isolates. Bench testing confirmed QC results within expected ranges.
Technological Characteristics
System uses molded polystyrene panels with 136 micro-wells containing dried reagents. Sensing principle: redox indicator and turbidity measurement via automated optical monitoring. Incubation: 35°C. Connectivity: automated instrument-based processing. Software: automated interpretation of growth kinetics to determine MIC and categorical results.
Indications for Use
Indicated for in vitro quantitative determination of antimicrobial susceptibility (MIC) of gram-negative aerobic and facultative anaerobic bacteria (Enterobacteriaceae and non-Enterobacteriaceae) to ampicillin (0.5 – 32 µg/mL).
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
VITEK® System (PMA No. N50510)
BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321)
BD Phoenix™ Automated Microbiology System with Ofloxacin (K020323)
BD Phoenix™ Automated Microbiology System with Levofloxacin (K020322)
Related Devices
K032131 — BD AUTOMATED MICROBIOLOGY SYSTEM TETRACYCLINE - GP 0.5-16 UG/ML-GRAM POSITIVE ID/AST OR AST ONLY PHOENIX PANELS · Becton, Dickinson & CO · Aug 18, 2003
K031429 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-MEROPENEM 0.25-16 UG/ML · Bd Diagnostic Systems · Jun 6, 2003
K033458 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - CONFIRMATORY ESBL TEST · Becton, Dickinson & CO · May 28, 2004
K060214 — BD PHOENIX AUTOMATED MIRCROBIOLOGY SYSTEM TETRACYCLINE-GN 0.5-16 UG/ML AND GP 0.5-16 UG/ML · Becton, Dickinson & CO · Mar 9, 2006
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
DEVICE ONLY TEMPLATE
A. 510(k) Number:
k032299
B. Analyte:
Ampicillin 0.5 – 32 µg/mL Gram-Negative AST
C. Type of Test:
Antimicrobial Susceptibility Test (Quantitative) colorimetric oxidation-reduction, growth-based
D. Applicant:
Becton, Dickinson & Company
E. Proprietary and Established Names:
BD Phoenix™ Automated Microbiology System – Ampicillin Gram Negative
F. Regulatory Information:
1. Regulation section:
21 CFR 866.1645 Fully Automated Short-Term Incubation Cycle
Antimicrobial
2. Classification:
Class II
3. Product Code:
LON
4. Panel:
83
G. Intended Use:
1. Intended use(s):
BD Phoenix™ Automated Microbiology System:
The BD Phoenix™ Automated Microbiology System is intended for
in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration of gram-negative aerobic and facultative anaerobic bacteria belonging to the family Enterobacteriaceae and
non-Enterobacteriaceae and gram-positive bacteria belonging to the genera Staphylococcus and Enterococcus.
The BD Phoenix™ GN Panel:
The BD Phoenix™ Automated Microbiology System is intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of gram-negative aerobic and facultative anaerobic bacteria belonging to the family Enterobacteriaceae and non-Enterobacteriaceae.
2. Indication(s) for use:
{1}
Page 2 of 5
This submission is for the addition of the antibiotic ampicillin at concentrations of 0.5 – 32 µg/mL to the gram negative susceptibility panel for testing Enterobacteriaceae.
3. Special condition for use statement(s): Not applicable
4. Special instrument Requirements: Not applicable
H. Device Description:
The BD Phoenix™ Automated Microbiology System includes instrumentation and software, sealed and self-inoculating molded polystyrene trays with 136 micro-wells containing dried reagents, and specific inoculum broth formulations for ID and AST Indicator. The organism to be tested must be a pure culture and be preliminarily identified as gram positive or gram negative. Colonies are then suspended in broth, and equated to a 0.5 McFarland with the recommendation to use the BD CrystalSpec™ Nephelometer. A further dilution is made into an AST broth, which contains an AST indicator, prior to inoculating the panel. The AST broth is a cation-adjusted formulation of Mueller-Hinton broth containing 0.01% Tween 80. After adding the indicator solution to the AST inoculum the color is blue and after inoculation and incubation goes to pink to colorless as reduction in the panel well proceeds. Inoculated panels are barcode scanned and loaded into the BD Phoenix™ Automated Microbiology System instrument where the panels are continuously incubated at 35°C. The AST has a final inoculum of 5 × 10⁵ CFU/ml. The instrument incubates, reads and records the results of the biochemical substrates and antimicrobial agents and interprets the reactions to give an ID of the isolate and MIC value and category interpretation of the antimicrobial agents. Organisms growing in the presence of a given antimicrobic agent reduce the indicator, signaling organism growth and resistance to the antimicrobic agent. Organisms killed or inhibited by a given antimicrobic do not cause reduction of the indicator and therefore do not produce a color change. Additional interpretation is done using software driven "EXPERT" System using rules derived from the NCCLS documentation.
Readings are taken every 20 minutes with an ID result available between 2-12 hours and an AST result available between 4-16 hours. This is only an autoread result; there are no manual readings possible.
I. Substantial Equivalence Information:
1. Predicate device name(s): VITEK® System
2. Predicate K number(s): N50510
3. Comparison with predicate:
{2}
Page 3 of 5
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| 1. | Isolated colonies from culture used | Isolated colonies from culture used |
| 2. | Inoculum density equated to 0.5 McFarland standard | Inoculum density equated to 0.5 McFarland standard |
| 3. | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) |
| 4. | <16 hours | <16 hours |
| Differences | | |
| Item | Device | Predicate |
| 1. | Results are determined from serial twofold dilutions of antimicrobial agents | Results are determined from extrapolation of doubling dilutions |
| 2. | Automated growth based enhanced by use of a redox indicator (colorimetric oxidation-reduction) to detect organism growth. | Automated growth based with detection using an attenuation of light measured by an optical scanner. |
J. Standard/Guidance Document Referenced (if applicable):
“Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA”; NCCLS M7 (M100-S13)
“Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard.”
K. Test Principle:
The system employs conventional, colorimetric, fluorogenic and chromogenic substrates to identify the genus and species of the isolate. The AST portion of the BD Phoenix™ Automated Microbiology System is a broth based microdilution method that utilizes a redox indicator (colorimetric oxidation-reduction) to enhance detection of organism growth. The MIC is determined by comparing growth in wells containing serial two-fold dilutions of an antibiotic to the growth in “growth control wells” which contain no antibiotic.
L. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Reproducibility within sites was determined using the QC isolates for >95% reproducibility. Between sites was performed at three sites for three days in triplicate for >95% reproducibility on 16 isolates.
b. Linearity/assay reportable range:
Not applicable
c. Traceability (controls, calibrators, or method):
{3}
Page 4 of 5
The recommended QC isolate was tested a sufficient number of times with acceptable results with the reference method. The Phoenix results demonstrate that the system can produce QC results in the recommended range.
| ORGANISM | conc. | Reference | | Phoenix | |
| --- | --- | --- | --- | --- | --- |
| | | | | | |
| E. coli
ATCC 25922
Expected Range:
2 – 8 μg/mL | 2 | | 3 | | 383 |
| | 4 | | 318 | | 1 |
| | 8 | | 63 | | 3 |
| | >32 | | 1 | | |
| | | | | | |
Inoculum density control: The organism suspension density of the ID broth was equivalent to a 0.5 McFarland standard using the BBL™ CrystalSpec™ Nephelometer which was verified each day of testing. Internal data was used to demonstrate that the use of the BBL™ CrystalSpec™ Nephelometer would produce reproducible results. Five different instruments were used.
d. Detection limit:
Not applicable
e. Analytical specificity:
Not applicable
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
The NCCLS recommended broth dilution reference panel was prepared according to the NCCLS recommendation. Clinical testing was performed at six sites. The testing included both fresh clinical isolates and stock isolates along with a challenge set with known results. The test device had a growth rate of >99%. A comparison was provided to the reference method with the following agreement.
| | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Clinical | 802 | 788 | 98.3 | 342 | 331 | 96.8 | 790 | 98.5 | 398 | 7 | 3 | 2 |
| Challenge | 36 | 34 | 94.4 | 5 | 5 | 100.0 | 34 | 94.4 | 29 | 1 | 0 | 1 |
| Combined | 838 | 822 | 98.1 | 347 | 336 | 96.8 | 824 | 98.3 | 427 | 8 | 3 | 3 |
EA-Essential Agreement
CA-Category Agreement
R-resistant isolates
maj-major discrepancies
vmj-very major discrepancies
min- minor discrepancies
Essential agreement (EA) is when the BD Phoenix™ panels agree with the reference test panel results exactly or within one doubling
{4}
Page 5 of 5
dilution of the reference method. Category agreement (CA) is when the BD Phoenix™ panel result interpretation agrees exactly with the reference panel result interpretation.
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a and b are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
≤8(S), 16(I), ≥32(R)
The expected value range, interpretative criteria and QC are the same as recommended in NCCLS. All values will be included in the package insert.
M. Conclusion:
This demonstrates acceptable performance as described in the FDA guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA" and therefore the testing of ampicillin on the BD Phoenix™ Automated Microbiology System is substantially equivalent to other commercial devices such as bioMerieux VITEK® AST panels.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.