BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-TICARCILLIN-1-128 UG/ML

K031984 · Becton, Dickinson & CO · LON · Aug 27, 2003 · Microbiology

Device Facts

Record IDK031984
Device NameBD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-TICARCILLIN-1-128 UG/ML
ApplicantBecton, Dickinson & CO
Product CodeLON · Microbiology
Decision DateAug 27, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Indications for Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

Device Story

System performs automated antimicrobial susceptibility testing (AST) for gram-negative bacteria; utilizes molded polystyrene trays with 136 micro-wells containing dried reagents; requires pure culture inoculum suspended in broth and equated to 0.5 McFarland standard. Instrument continuously incubates panels at 35°C; monitors growth via colorimetric oxidation-reduction redox indicator (blue to pink/colorless); readings taken every 20 minutes. Software-driven 'EXPERT' system interprets biochemical reactions and MIC values based on NCCLS rules. Provides rapid ID and MIC results; eliminates manual reading. Used in clinical microbiology laboratories; assists clinicians in selecting appropriate antibiotic therapy by identifying bacterial resistance profiles.

Clinical Evidence

Performance evaluated using clinical, stock, and challenge isolates across multiple US sites. Compared against NCCLS reference broth microdilution method. Results for ticarcillin: 2294 isolates tested, 94.0% Essential Agreement (EA), 94.0% Category Agreement (CA). Reproducibility study (intra-site >90%, inter-site >95%) confirmed system reliability.

Technological Characteristics

Automated microdilution system. Components: instrument, software, polystyrene panels (136 wells), ID/AST broth, redox indicator. Incubation: 35°C. Sensing: optical turbidity and redox indicator color change. Connectivity: integrated system. Software: automated interpretation of growth kinetics.

Indications for Use

Indicated for in vitro quantitative determination of antimicrobial susceptibility (MIC) of gram-negative aerobic and facultative anaerobic bacteria (Enterobacteriaceae and non-Enterobacteriaceae) to the antibiotic ticarcillin at concentrations of 1-128 µg/mL. Results for Proteus vulgaris/penneri are excluded.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: k031984 B. Analyte: Ticarcillin 1 – 128 µg/mL Gram - Negative AST C. Type of Test: Antimicrobial Susceptibility Test (Quantitative) colorimetric oxidation-reduction, growth-based D. Applicant: Becton, Dickinson &amp; Company E. Proprietary and Established Names: BD Phoenix™ Automated Microbiology System – Ticarcillin Gram - Negative F. Regulatory Information: 1. Regulation section: 21 CFR 866.1645 Fully Automated Short-Term Incubation Cycle Antimicrobial 2. Classification: Class II 3. Product Code: LON 4. Panel: 83 G. Intended Use: 1. Intended use(s): BD Phoenix™ Automated Microbiology System: The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration of gram-negative aerobic and facultative anaerobic bacteria belonging to the family Enterobacteriaceae and non-Enterobacteriaceae and gram-positive bacteria belonging to the genera Staphylococcus and Enterococcus. The BD Phoenix™ GN Panel: The BD Phoenix™ Automated Microbiology System is intended for the in vitro rapid identification (ID) and quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of gram-negative aerobic and facultative anaerobic bacteria belonging to the family Enterobacteriaceae and non-Enterobacteriaceae. {1} Page 2 of 7 2. **Indication(s) for use:** This submission is for the addition of the antibiotic ticarcillin at concentrations of $1 - 128~\mu \mathrm{g / mL}$ to the gram negative susceptibility panel. 3. **Special condition for use statement(s):** Results for *Proteus vulgaris/penneri* have been excluded in the BD Phoenix™ therefore no results will be reported. An alternate method should be performed when this combination is identified. 4. **Special instrument Requirements:** Not applicable ## H. Device Description: The BD Phoenix™ Automated Microbiology System includes instrumentation and software, sealed and self-inoculating molded polystyrene trays with 136 micro-wells containing dried reagents, and specific inoculum broth formulations for ID and AST Indicator. The organism to be tested must be a pure culture and be preliminarily identified as gram positive or gram negative. Colonies are then suspended in broth, and equated to a $0.5\mathrm{McFarland}$ with the recommendation to use the BD CrystalSpec™ Nephelometer. A further dilution is made into an AST broth, which contains an AST indicator, prior to inoculating the panel. The AST broth is a cation-adjusted formulation of Mueller-Hinton broth containing $0.01\%$ Tween 80. After adding the indicator solution to the AST inoculum the color is blue and after inoculation and incubation goes to pink to colorless as reduction in the panel well proceeds. Inoculated panels are barcode scanned and loaded into the BD Phoenix™ Automated Microbiology System instrument where the panels are continuously incubated at $35^{\circ}\mathrm{C}$. The AST has a final inoculum of $5\times 10^{5}$ CFU/ml. The instrument incubates, reads and records the results of the biochemical substrates and antimicrobial agents and interprets the reactions to give an ID of the isolate and MIC value and category interpretation of the antimicrobial agents. Organisms growing in the presence of a given antimicrobic agent reduce the indicator, signaling organism growth and resistance to the antimicrobic agent. Organisms killed or inhibited by a given antimicrobic do not cause reduction of the indicator and therefore do not produce a color change. Additional interpretation is done using software driven "EXPERT" System using rules derived from the NCCLS documentation. Readings are taken every 20 minutes with an ID result available between 2-12 hours and an AST result available between 4-16 hours. This is only an autoread result; there are no manual readings possible ## I. Substantial Equivalence Information: 1. **Predicate device name(s):** VITEK® System 2. **Predicate K number(s):** N50510 3. **Comparison with predicate:** {2} Page 3 of 7 | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | 1. | Isolated colonies from culture used | Isolated colonies from culture used | | 2. | Inoculum density equated to 0.5 McFarland standard | Inoculum density equated to 0.5 McFarland standard | | 3. | Results are determined from serial twofold dilutions of antimicrobial agents | Results are determined from serial twofold dilutions of antimicrobial agents | | 4. | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (SIR) | | 5. | <16 hours | <16 hours | | Differences | | | | Item | Device | Predicate | | 1. | Automated growth based enhanced by use of a redox indicator (colorimetric oxidation-reduction) to detect organism growth. | Automated growth based with detection using an attenuation of light measured by an optical scanner. | J. Standard/Guidance Document Referenced (if applicable): “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA”; NCCLS M7 (M100-S13) “Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard.” K. Test Principle: The system employs conventional, colorimetric, fluorogenic and chromogenic substrates to identify the genus and species of the isolate. The AST portion of the BD Phoenix™ Automated Microbiology System is a broth based microdilution method that utilizes a redox indicator (colorimetric oxidation-reduction) to enhance detection of organism growth. The MIC is determined by comparing growth in wells containing serial two-fold dilutions of an antibiotic to the growth in “growth control wells” which contain no antibiotic. L. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Reproducibility within sites was determined using the QC isolates for &gt;95% reproducibility. Between sites was performed at three sites for &gt;95% reproducibility on twelve isolates. {3} b. Linearity/assay reportable range: Not applicable c. Traceability (controls, calibrators, or method): The recommended QC isolate was tested a sufficient number of times with acceptable results with the reference method. The Phoenix results demonstrate that the system can produce QC results in the recommended range. QC Table | ORGANISM | conc. μg/mL | Reference | | | PhoenixTM | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | | | | | E. coliATCC 25922Expected range: 4-16 μg/mL | 4 | | 63 | | | | | | | 8 | | 181 | | | 375 | | | | 16 | | 139 | | | 6 | | | | 32 | | 2 | | | 4 | | | | 64 | | | | | 1 | | | | >128 | | | | | 1 | | | | | | | | | | | | P. aeruginosaATCC 27853Expected Range: 8-32 μg/mL | 8 | | 1 | | | 1 | | | | 16 | | 313 | | | 151 | | | | 32 | | 61 | | | 221 | | | | 64 | | 9 | | | 2 | | | | 128 | | | | | 2 | | | | >128 | | | | | 8 | | | | | | | | | | | | | | | | | | | | Inoculum density control: The organism suspension density of the ID broth was equivalent to a 0.5 McFarland standard using the $\mathrm{BBL}^{\mathrm{TM}}$ CrystalSpec™ Nephelometer which was verified each day of testing. Internal data was used to demonstrate that the use of the $\mathrm{BBL}^{\mathrm{TM}}$ CrystalSpec™ Nephelometer would produce reproducible results. Five different instruments were used. d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable {4} Page 5 of 7 2. Comparison studies: a. Method comparison with predicate device: The NCCLS recommended broth dilution reference panel was prepared according to the NCCLS recommendation. Clinical testing was performed at six sites. The testing included both fresh clinical isolates and stock isolates along with a challenge set with known results. The test device had a growth rate of 99.0%. The NCCLS recommended breakpoints for ticarcillin are different from FDA’s. This is particularly true for the organism, P. aeruginosa. It is for this reason that P. aeruginosa will be presented in a table independently of Enterobacteriaceae and Non-Enterobacteriaceae. A third table with combined results from all three groups will be included as well. The following charts are based on NCCLS interpretive criteria. GN Accuracy MIC Summary for Enterobacteriaceae and Other Non-Enterobacteriaceae Excluding Pseudomonas aeruginosa Interpretation: ≤16 (S); 32 – 64 (I); ≥128(R) | | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Clinical | 1717 | 1622 | 94.5 | 728 | 664 | 91.2 | 1600 | 93.2 | 743 | 102 | 11 | 4 | | Challenge | 60 | 59 | 98.3 | 15 | 14 | 93.3 | 58 | 96.7 | 36 | 2 | 0 | 0 | | Combined | 1777 | 1681 | 94.6 | 743 | 678 | 91.3 | 1658 | 93.3 | 779 | 104 | 11 | 4 | GN Accuracy MIC Summary for Pseudomonas aeruginosa Interpretation: ≤64 (S); ≥128(R) | | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Clinical | 508 | 468 | 92.1 | 418 | 380 | 90.9 | 456 | 89.8 | 138 | NA | 40 | 12 | | Challenge | 9 | 8 | 88.9 | 5 | 4 | 80.0 | 8 | 88.9 | 5 | NA | 0 | 1 | | Combined | 517 | 476 | 92.1 | 423 | 384 | 90.8 | 464 | 89.7 | 143 | NA | 40 | 13 | NA – No intermediate range therefore no minor errors possible There were 40 maj errors and 24 of the maj errors were in essential agreement. Out of the 13 vmj errors encountered, 11 were in essential agreement. GN Accuracy MIC Summary for Enterobacteriaceae, P. aeruginosa and Other Non-Enterobacteriaceae | | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | Min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Clinical | 2225 | 2090 | 93.9 | 1146 | 1044 | 91.1 | 2056 | 92.4 | 881 | 102 | 51 | 16 | | Challenge | 69 | 67 | 97.1 | 20 | 18 | 90.0 | 66 | 95.7 | 41 | 2 | 0 | 1 | | Combined | 2294 | 2157 | 94.0 | 1166 | 1062 | 91.1 | 2122 | 92.5 | 922 | 104 | 51 | 17 | EA-Essential Agreement CA-Category Agreement R-resistant isolates maj-major discrepancies vmj-very major discrepancies min- minor discrepancies {5} Page 6 of 7 Essential agreement (EA) is when the BD Phoenix™ panels agree with the reference test panel results exactly or within one doubling dilution of the reference method. Category agreement (CA) is when the BD Phoenix™ panel result interpretation agrees exactly with the reference panel result interpretation. The following table represents data analysis using FDA interpretations. GN Accuracy MIC Summary for Enterobacteriaceae, P. aeruginosa and Other Non-Enterobacteriaceae Interpretation: ≤64 (S); ≥128(R) | | EA Tot | EA N | EA % | Eval EA Tot | Eval EA N | Eval EA % | CA N | CA % | #R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Clinical | 2225 | 2090 | 93.9 | 1146 | 1044 | 91.1 | 2121 | 95.3 | 881 | NA | 68 | 36 | | Challenge | 69 | 67 | 97.1 | 20 | 18 | 90.0 | 67 | 97.1 | 41 | NA | 1 | 1 | | Combined | 2294 | 2157 | 94.0 | 1166 | 1062 | 91.1 | 2188 | 95.4 | 922 | NA | 69 | 37 | NA – No intermediate range therefore no minor errors possible Using the FDA’s recommended breakpoints, there are 37 vmj errors and 69 maj errors. It is also important to point out that 25 of the vmj errors are within essential agreement and 34 of the maj errors are within essential agreement. The vmj% of 4.0 (37/922) and maj% of 5.0 (69/1372) are both higher than as recommended in the guidance document but these recommendations are based on errors that would include a minor error range. Since this antibiotic has only a Sensitive (S) and a Resistant (R) result all errors are either vmj or maj with no minor errors possible. However, if those errors that are in EA are removed, its vmj rate would be 1.3% (12/922) and maj rate is 2.6% (35/1372), both of which are acceptable. b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a and b are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Enterobacteriaceae and Other Non – Enterobacteriaceae: ≤16 (S); 32 – 64 (I); ≥128(R) Ps. aeruginosa: ≤64 (S); ≥128(R) The expected value range, interpretative criteria and QC are the same as recommended in NCCLS. All values will be included in the package insert. {6} Page 7 of 7 M. Conclusion: This demonstrates acceptable performance as described in the FDA guidance document, “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA” and therefore the testing of ticarcillin on the BD Phoenix™ Automated Microbiology System is substantially equivalent to other commercial devices such as bioMerieux Vitek® AST panels.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...