BD PHOENIX AUTOMATED MICRBIOLOGY SYSTEM AMOXICILLIN/CLAVUALANTE 0.5/0.25-32/16 UG/ML

K031912 · Becton, Dickinson & CO · LON · Jul 21, 2003 · Microbiology

Device Facts

Record IDK031912
Device NameBD PHOENIX AUTOMATED MICRBIOLOGY SYSTEM AMOXICILLIN/CLAVUALANTE 0.5/0.25-32/16 UG/ML
ApplicantBecton, Dickinson & CO
Product CodeLON · Microbiology
Decision DateJul 21, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Indications for Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.

Device Story

The BD Phoenix System is an automated microbiology platform for rapid identification and antimicrobial susceptibility testing (AST) of bacterial isolates. The system utilizes sealed, self-inoculating polystyrene panels containing 136 micro-wells with dried biochemicals and antimicrobial agents. Input consists of pure bacterial cultures prepared in Phoenix ID or AST broth (0.5 McFarland standard). The system performs broth microdilution testing, using a redox indicator and turbidity measurements to detect bacterial growth in the presence of antimicrobial agents. Panels are incubated at 35°C, with the instrument taking readings every 20 minutes. Software interprets these readings to provide organism identification, MIC values, and categorical interpretations (S, I, or R). Used in clinical microbiology laboratories, the system assists clinicians in selecting appropriate antibiotic therapy by providing rapid susceptibility profiles, potentially improving patient outcomes through targeted treatment.

Clinical Evidence

Performance evaluated using clinical, stock, and challenge isolates across multiple U.S. sites. Comparison against NCCLS reference broth microdilution method. Primary endpoints: Essential Agreement (EA) and Category Agreement (CA). For amoxicillin/clavulanate (Gram-negative), EA was 96.5% (n=1919) and CA was 90.3% (n=1919). Reproducibility studies showed >90% intra-site and >95% inter-site agreement.

Technological Characteristics

Automated broth microdilution system. Components: instrument, software, panels (molded polystyrene, 136 wells), ID/AST broths, redox indicator. Sensing: optical detection of redox indicator changes and bacterial turbidity. Incubation: 35°C. Connectivity: automated instrument-based processing. Software: automated interpretation of MIC and categorical results.

Indications for Use

Indicated for in vitro quantitative determination of antimicrobial susceptibility (MIC) of Gram-negative aerobic/facultative anaerobic bacteria (Enterobacteriaceae and Non-Enterobacteriaceae) and Gram-positive bacteria (Staphylococcus and Enterococcus) from pure culture. Specifically for use with amoxicillin/clavulanate (0.5/0.25-32/16 µg/mL) in Gram-negative ID/AST or AST only panels.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(K) SUMMARY | SUBMITTED BY: | Becton, Dickinson and Company<br>7 Loveton Circle<br>Sparks, MD 21152<br>Phone: (410) 316 - 4778<br>Fax: 410-316-4499 | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | CONTACT NAME: | Michelle B. Bandy<br>Regulatory Affairs Specialist | | DATE PREPARED: | June 19, 2003 | | DEVICE TRADE NAME: | BD Phoenix™ Automated Microbiology System -<br>Amoxicillin/Clavulanate 0.5/0.25-32/16 µg/mL | | DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation | | DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation Cycle Antimicrobial<br>Susceptibility Device, 21 CFR 866.1645 | | PREDICATE DEVICES: | VITEK® System (PMA No. N50510) and BD Phoenix™<br>Automated Microbiology System with Gatifloxacin (K020321,<br>May 23, 2002), Ofloxacin (K020323, April 14, 2002), and<br>Levofloxacin (K020322, March 27, 2002). | | INTENDED USE: | The BD Phoenix™ Automated Microbiology System is<br>intended for the rapid identification and in vitro antimicrobial<br>susceptibility testing of isolates from pure culture of most<br>aerobic and facultative anaerobic Gram-negative and Gram-<br>positive bacteria of human origin. | ## DEVICE DESCRIPTION: The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components: - . BD Phoenix instrument and software. - BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . for AST determinations. - BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. ● - BD Phoenix AST Broth used for performing AST tests only. ● - BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth . determination. {1}------------------------------------------------ The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID Broth. The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations. The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant). ## DEVICE COMPARISON: The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the NCCLS reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Gram-negative ID/AST or AST only Phoenix panels with this antimicrobial agent. ## SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING: The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the NCCLS reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000. ## Site Reproducibility Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of Gram-negative isolates. Each site tested the isolates in triplicate on three different days using one lot of Gram-negative Phoenix panels containing this antimicrobial agent and associated reagents. The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the Gram-negative isolates tested. BD Diagnostic Systems Becton, Dickinson and Company Page 1 1 {2}------------------------------------------------ ### Clinical Studies Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Gram-negative Phoenix panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the NCCLS reference broth microdilution method. The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within ± one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant). Table 1 summarizes the performance for the isolates tested in this study. #### Performance of BD Phoenix System for Gram-negative Organisms by Drug Table 1: | and the controlled on the controlled to the controllation of the comments of the comments of the comments of the comments of the comments of the comments of the comments of t<br> Antimicrobial | Concentration | EA (n) | EA (%) CA (n) CA (%) | | | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|--------|--------------------------|------|------| | Amoxicillin/Clavulanate 0.5/0.25-32/16 ug/mL | | 1919 | 96.5 | 1919 | 90.3 | ### Conclusions Drawn from Substantial Equivalence Studies The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002). BD Diagnostic Systems Becton, Dickinson and Company Page 12 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows a logo with a stylized bird in flight. The bird is composed of three curved lines that suggest movement and feathers. The bird is encircled by a dotted line, which gives the impression of a seal or emblem. The logo is simple, yet conveys a sense of freedom and flight. Public Health Service od and Drug Administration 2098 Gaither Road Rockville MD 20850 JUL 2 1 2003 Ms. Michelle B. Bandy Regulatory Affairs Specialist BD Diagnostic Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 Re: k031912 Trade/Device Name: BD Phoenix™ Automated Microbiology Systems Amoxicillin/clavulanate (0.5/0.25-32/16 µg/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: June 19, 2003 Received: June 20, 2003 Dear Ms. Bandy: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ ### Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Page 1 of 1 510(k) Number: Ko 31912 Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent amoxicillin/clavulanate (0.5/0.25-32/16 ug/mL) - Gram-negative ID/AST or AST only Phoenix panels. Indications for Use: The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus. This premarket notification is for the addition of the antimicrobial agent amoxicillin/clavulanate at concentrations of 0.5/0.25-32/16 ug/mL to Gram-negative ID/AST or AST only Phoenix panels. Amoxicillin/clavulanate has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent. ### Active In Vitro and in Clinical Infections Against: Enterobacter species Escherichia coli Klebsiella species Active In Vitro Against: Proteus mirabilis Prescription Use (Per 21 CFR 801.109) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Sagatt Division Sigh-Off 7/17/2003 Over-the-Counter Use e of In Vitro Diagnostic Device Fireliation and No :(n(.) K031912
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...