MICROSCAN RAPID/S PLUS GRAM-NEGATIVE MIC/COMBO PANELS WITH CEPHALOTHIN (2-64 UG/ML)

K020410 · Dade Behring, Inc. · LON · Jun 11, 2002 · Microbiology

Device Facts

Record IDK020410
Device NameMICROSCAN RAPID/S PLUS GRAM-NEGATIVE MIC/COMBO PANELS WITH CEPHALOTHIN (2-64 UG/ML)
ApplicantDade Behring, Inc.
Product CodeLON · Microbiology
Decision DateJun 11, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1645
Device ClassClass 2

Indications for Use

The MicroScan® rapID/S plus" Gram-Negative MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic Gram-Negative bacilli (Enterobacteriaceae, glucose non-fermenters, and non-Enterobacteriaceae glucose fermenters. After inoculation, panels are read on the WalkAway® SI System or equivalent (upgraded WalkAway® 40 or WalkAway® 96) according to the Package Insert. This particular submission is for the antimicrobial Cephalothin on the rapID/S plus " Gram-Negative MIC/Combo Panels. The Gram-Negative organisms which may be used for Cephalothin susceptibility testing in this panel are: Klebsiella pneumoniae Proteus mirabilis The MicroScan® rapID/S plus" Gram-Negative Cephalothin is not intended for use with: Escherichia coli Klebsiella oxytoca Pseudomonas aeruginosa and all Non-Enterobacteriacea

Device Story

MicroScan® rapID/S plus™ Gram-Negative MIC/Combo Panels are miniaturized broth dilution susceptibility tests. Panels contain antimicrobial agents in Mueller-Hinton Broth in dried form. Clinical laboratory technicians inoculate and rehydrate panels with standardized organism suspensions; panels are incubated at 35°C for 4.5-18 hours within the WalkAway® SI System (or upgraded WalkAway® 40/96). The system measures growth inhibition to determine the Minimum Inhibitory Concentration (MIC). Results assist clinicians in selecting appropriate antimicrobial therapy for patients with Gram-negative bacterial infections.

Clinical Evidence

Bench testing only. Performance evaluated by comparing the rapID/S plus™ panel against an NCCLS frozen Reference Panel using fresh/stock Efficacy isolates and stock Challenge strains. Results demonstrated >96% Essential Agreement for Cephalothin. Reproducibility and precision were confirmed using the WalkAway® SI System with turbidity inoculum preparation. Quality control testing met acceptance criteria.

Technological Characteristics

Miniaturized broth dilution susceptibility test; dried antimicrobial agents in micro-titer wells; Mueller-Hinton Broth; automated incubation and reading via WalkAway® SI System (or upgraded WalkAway® 40/96); turbidity-based inoculum preparation.

Indications for Use

Indicated for antimicrobial susceptibility testing of rapidly growing aerobic and facultative anaerobic Gram-Negative bacilli (Enterobacteriaceae, glucose non-fermenters, and non-Enterobacteriaceae glucose fermenters). Specifically for Cephalothin (2-64 µg/ml) testing of Klebsiella pneumoniae and Proteus mirabilis. Contraindicated for Escherichia coli, Klebsiella oxytoca, Pseudomonas aeruginosa, and all Non-Enterobacteriaceae.

Regulatory Classification

Identification

A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUN 1 1 2002 Image /page/0/Picture/1 description: The image shows a sequence of characters that appear to be handwritten. The characters are 'K020410'. The characters are written in a rough, slightly distressed style, giving them a textured appearance. The overall impression is that of a handwritten code or identifier. # 510(k) Summary ### 510(k) Submission Information: | Device Manufacturer: | Dade Behring Inc. | |----------------------|------------------------------------------------------------| | Contact name: | Maureen Mende, Group Manager Regulatory Affairs | | Fax: | 916-374-3144 | | Date prepared: | February 4, 2002 | | Product Name: | Microdilution Minimum Inhibitory Concentration (MIC) Panel | | Trade Name: | MicroScan® rapID/S plus™ Gram-Negative MIC/Combo panel | | Intended Use: | To determine antimicrobial agent susceptibility | | 510(k) Notification: | Antimicrobials: Cephalothin | | Predicate device: | MicroScan Dried Gram Negative MIC/Combo Panels | #### 510(k) Summary: MicroScan® rapID/S plus™ Gram-Negative MIC/Combo Panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-negative bacilli. The MicroScan® rapID/S plus™ Gran-Negative MIC/Combo Panels are read on the WalkAway® SI System or equivalent (upgraded WalkAway® 40 or WalkAway® 96 instruments). The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in Mueller-Hinton Broth to concentrations bridging the range of clinical interest and are presented in micro-titer wells in dried form. rapID/S plus™ panels are inoculated and rehydrated with a standardized suspension of the organism and incubated at 35°C in the WalkAway® SI System or equivalent for 4.5 - 18 hours. The minimum inhibitory concentration (MIC) for the test organism is determined by the lowest antimicrobial concentration showing inhibition of growth. The proposed MicroScan® rapID/S plus™ Gram-Negative MIC/Combo Panel demonstrated substantially equivalent performance when compared with an NCCLS frozen Reference Panel, as defined in the FDA DRAFT document "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", dated March 8, 2000. The Premarket Notification (510[k]) presents data in support of the MicroScan® rapID/S plus™ Gram-Negative MIC/Combo Panel Cephalothin. The external evaluation was conducted with fresh and stock Efficacy isolates and stock Challenge strains. The external evaluations were designed to confirm the acceptability of the proposed rapID/S p/us™ Gram-Negative Panel by comparing its performance with an NCCLS frozen Reference panel. Challenge strains were compared to Expected Results determined prior to the evaluation. The rapID/S plus™ Gram-Negative Panel demonstrated acceptable performance with an overall Essential Agreement of greater than 96% for Cephalothin when compared with the frozen Reference panel. Instrument reproducibility testing demonstrated acceptable reproducibility and precision with Cephalothin with Turbidity inoculum preparation method and the WalkAway® SI System or equivalent (upgraded WalkAway® 40 or WalkAway® 96 instruments). Quality Control testing demonstrated acceptable results for Cephalothin. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a serif font. The text is centered in the image and is the only element present. Image /page/1/Picture/1 description: The image shows the seal for the Department of Health & Human Services USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter of the circle. In the center of the seal are three stylized human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest. Food and Drug Administration -2098 Gaither Road Rockville MD 20850 # JUN 1 1 2002 Ms. Maureen Mende Group Manager Regulatory Affairs Dade Behring Inc. 1584 Enterprise Boulevard West Sacramento, CA 95691 Re: k020410 > Trade/Device Name: MicroScan® RapID/S plus™4 Gram-Negative MIC/Combo Panels with Cephalothin (2-64 µg/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: May 3, 2002 Received: May 7, 2002 Dear Ms. Mende: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and v additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entifled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Butman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT . Page 1 of 1 . ### 510(k) Number (if known): K020410 Device Name: MicroScan® rapID/S plus" Gram-Negative MIC/Combo Panels with Cephalothin (2 - 64 ug/ml) ### Indications For Use: The MicroScan® rapID/S plus" Gram-Negative MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic Gram-Negative bacilli (Enterobacteriaceae, glucose non-fermenters, and non-Enterobacteriaceae glucose fermenters. After inoculation, panels are read on the WalkAway® SI System or equivalent (upgraded WalkAway® 40 or WalkAway® 96) according to the Package Insert. This particular submission is for the antimicrobial Cephalothin on the rapID/S plus " Gram-Negative MIC/Combo Panels. The Gram-Negative organisms which may be used for Cephalothin susceptibility testing in this panel are: Klebsiella pneumoniae Proteus mirabilis The MicroScan® rapID/S plus" Gram-Negative Cephalothin is not intended for use with: Escherichia coli Klebsiella oxytoca Pseudomonas aeruginosa and all Non-Enterobacteriacea (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | Over-The-Counter Use | |------------------|----------------------| |------------------|----------------------| (Per 21 CFR 801.109) OR (Optional Format 1-2-96) ![signature](signature.png) (Division Sign-Off) Division of Clinical Laboratory Devices | 510(k) Number | K02 0410 | |---------------|----------| |---------------|----------|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...