K020321 · Becton, Dickinson & CO · LON · May 23, 2002 · Microbiology
Device Facts
Record ID
K020321
Device Name
BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM
Applicant
Becton, Dickinson & CO
Product Code
LON · Microbiology
Decision Date
May 23, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Indications for Use
The BD Phoenix™ Automated Microbiology System is intended for the in vitro rapid identification and antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic gram negative and gram positive bacteria of human origin.
Device Story
The BD Phoenix System is an automated microbiology platform for rapid identification (ID) and antimicrobial susceptibility testing (AST) of bacterial isolates. The system uses sealed, self-inoculating polystyrene panels containing biochemicals for ID and antimicrobial agents for AST. Input consists of a pure culture bacterial inoculum (0.5 McFarland standard) prepared in Phoenix ID broth, then transferred to Phoenix AST broth containing an indicator solution. Panels are loaded into the instrument, which incubates them at 35°C and performs optical readings every 20 minutes. The system measures changes in a redox indicator and bacterial turbidity to determine growth. Software interprets these measurements to provide organism identification, MIC values, and categorical interpretations (S, I, or R). Used in clinical laboratory settings by trained personnel, the system automates the broth microdilution process, providing rapid results to assist clinicians in selecting appropriate antimicrobial therapy for patients with bacterial infections.
Clinical Evidence
Clinical studies conducted at ten diverse U.S. sites evaluated the Phoenix System using Gram-negative and Gram-positive panels. Performance was assessed by comparing results against the NCCLS reference broth microdilution method. Essential Agreement (EA) and Category Agreement (CA) were calculated. For Gatifloxacin, the system demonstrated >90% EA and CA across tested isolates. Intra- and inter-site reproducibility exceeded 95%.
Technological Characteristics
Automated broth microdilution system. Components: instrument, software, polystyrene panels (136 micro-wells), ID/AST broths, and redox indicator. Sensing principle: optical measurement of redox indicator changes and bacterial turbidity. Incubation: 35°C. Connectivity: automated instrument-based processing. Software: automated interpretation of growth kinetics to determine MIC and susceptibility categories.
Indications for Use
Indicated for in vitro quantitative determination of antimicrobial susceptibility by MIC of gram-negative aerobic/facultative anaerobic bacteria (Enterobacteriaceae and Non-Enterobacteriaceae) and gram-positive bacteria (Staphylococcus and Enterococcus) from pure culture. Gatifloxacin results for Enterobacteriaceae are indicated only for urinary tract isolates.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
VITEK® System (PMA No. N50510)
Related Devices
K031912 — BD PHOENIX AUTOMATED MICRBIOLOGY SYSTEM AMOXICILLIN/CLAVUALANTE 0.5/0.25-32/16 UG/ML · Becton, Dickinson & CO · Jul 21, 2003
K033560 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - CEFTAZIDIME 0.5-64 UG/ML-GRAM-NEGATIVE ID/AST OR AST ONLY PHOENIX PANELS · Becton, Dickinson & CO · Jan 14, 2004
K041572 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-PIPERCILLIN (GN) 0.5-128 UG/ML · Becton, Dickinson and Co. · Aug 12, 2004
K023895 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - CIPROFLOXACIN GN · Becton, Dickinson & CO · Jan 8, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
MAY 2 3 2002
K020321
Attachment 2 510(k) SUMMARY Becton, Dickinson and Company SUBMITTED BY: 7 Loveton Circle Sparks, MD 21152 Phone: 410-316-4206 410-316-4499 Fax: Bradford Spring CONTACT NAME: Manager, Regulatory Affairs May 6, 2002 DATE PREPARED: BD Phoenix™ Automated Microbiology System DEVICE TRADE NAME: Antimicrobial susceptibility test system-short incubation DEVICE COMMON NAME: In accordance with FDA's reclassification order issued DEVICE CLASSIFICATION: December 28, 2001, Docket # 97P-0313, the BD Phoenix™ Automated Microbiology System has been classified as a Class II device, Automated Antimicrobial Susceptibility System Test, short incubation (Product Code LON) VITEK® System (PMA No. N50510) PREDICATE DEVICES: The BD Phoenix™ Automated Microbiology System is INTENDED USE: intended for the in vitro rapid identification and antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic gram negative and gram positive bacteria of human origin.
#### DEVICE DESCRIPTION:
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- . BD Phoenix instrument and software.
- BD Phoenix panels containing biochemicals for organism ID testing and . antimicrobial agents for AST determinations.
- . BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum.
- . BD Phoenix AST Broth used for performing AST tests only.
- BD Phoenix AST Indicator solution added to the AST broth to aid in bacterial growth . determination.
{1}------------------------------------------------
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a gram-negative or gram-positive isolate. For each isolate, an inoculum suspension equivalent to a 0.5 McFarland standard is prepared in the Phoenix ID broth. Prior to inoculating the Phoenix AST broth, one drop of Phoenix AST indicator solution is added to the uninoculated AST broth. The AST broth is then inoculated using the organism suspension from the ID broth. The Phoenix antimicrobial susceptibility test is inoculated with the AST broth. Inoculated panels are loaded into the Phoenix instrument.
The Phoenix AST method is a broth based microdilution test. The Phoenix system utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations. S. I. or R (sensitive, intermediate, and resistant).
# DEVICE COMPARISON:
The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the NCCLS broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Gram negative and Gram positive ID/AST or AST only Phoenix panels with Gatifloxacin.
# SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:
The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the NCCLS reference broth microdilution method (AST panels manufactured according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000.
# Analytical Studies
# Site Reproducibility
Intra- and inter-site reproducibility of the Phoenix AST system in determining minimum inhibitory concentration (MIC) results was evaluated at six external sites. Three sites tested a panel of gram-negative isolates and three sites tested a panel of gram-positive isolates. Each site tested the isolates in triplicate on three different days. One lot of Gram Negative Phoenix Panels and one lot of Gram Positive Phoenix panels with Gatifloxacin and associated reagents was tested at the gram-negative and gram-positive sites, respectively.
.
{2}------------------------------------------------
The results of the study demonstrate for the antimicrobial Gatifloxacin an overall intra-site The results of the study actions and an overall inter-site reproducibility greater than 95% for both gram-negative and gram-positive isolates tested.
## Clinical Studies
Clinical studies were conducted at ten geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test. Six study sites acmonorate the Phoenix System using the Gram Negative Phoenix Panel format and four study sites evaluated the Phoenix System using the Gram Positive Phoenix Panel format. The study involved testing Challenge set isolates and clinical isolates. Phoenix System results for If roll cest isolates were compared to the expected results for each organism/antimicrobic combination. Phoenix System results for clinical isolates were compared to the results obtained from the reference broth microdilution method. The reference AST panels were manufactured according to NCCLS M7. Antimicrobial agents in the Phoenix and reference panels had similar dilution ranges. The Phoenix organism identification was used as the basis for susceptibility interpretation for both the Phoenix and reference methods.
Phoenix panels demonstrated essential agreement of > 90% to the expected/reference results for all three drugs. The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within ± on two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant). Table 1 summarizes the performance for the gram-negative isolates tested by drug. Table 2 summarizes the performance for the gram-positive isolates tested by drug.
| Table 1: Performance of Phoenix System for Gram-Negative Organisms by Drux | | | |
|----------------------------------------------------------------------------|--|--|--|
|----------------------------------------------------------------------------|--|--|--|
| Station of the first and the fit in a many help in the fit in a transmission in the same<br>Antimicrobial | Comprehensive program and program and commended to the comments of the comments of the specifically of the station of the first of the first of the first of the first of the<br>Concentration | n.<br>PA | 10/01<br>EA.<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | n<br>All<br>The first of the policy of the problem of the first of the first of the | 10/0 1 |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|--------|
| E SERVEILLE<br>Carlos Career Career Comers of Career Comers of Construction Comers of Construction of Constitution Comers of Constitution Comers of Constitution Comers of Constitution Comer<br>Gatifloxacın | 0 95 0<br>0.25-8 mcg/mL | production and commend an annual as a clober 1977 C.<br>7160<br>GIV. | 98.8 | 2169 | 96.1 |
# Table 2: Performance of Phoenix System for Gram-Positive Organisms by Drug
| Antimicrobial | Concentration | EA (n) | EA (%) | CA (n) | CA (%) |
|---------------|---------------|--------|--------|--------|--------|
| Gatifloxacin | 0.25-8 mcg/mL | 1180 | 98.6 | 1180 | 90.1 |
# Conclusions Drawn from Substantial Equivalence Studies
The data collected in the analytical and clinical studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with Gatifloxacin is substantially equivalent as outlined in the FDA draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized symbol consisting of three abstract human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Bradford M. Spring Manager, Regulatory Affairs BD Diagnostic Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152
k020321 Re:
Rozoser (0.25-8 µg/ml)
Regulation Number: 21 CFR 866.1645
Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices
Regulatory Class: Class II Product Code: LON Dated: May 6, 2002 Received: May 7, 2002
Dear Mr. Spring: .
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Attachment 1
## 510(k) Number: K020321
Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial Gatifloxacin (0.25-8 ug) on Gram negative and Gram positive ID/AST or AST only Phoenix panels.
Indications for Use:
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.
This premarket notification is for the addition of the antimicrobial Gatifloxacin at concentrations of 0.25-8 µg to Gram negative and Gram positive ID/AST or AST only Phoenix panels. Gatifloxacin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobic.
| Active In Vitro and in Clinical Infections Against:<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>STATUTE AND CONSULT CANADIAN OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF CONSULT OF | | | | | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|
Aerobic Gram-positive microorganisms Staphylococcus aureus (methicillin-susceptible strains only)
Aerobic Gram-negative microorganisms Escherichia coli Klebsiella pneumoniae Proteus mirabilis
#### Active In Vitro Against:
Aerobic Gram-positive microorganisms Staphylococcus saprophyticus
- Aerobic Gram-negative microorganisms Acinetobacter lwoffii Citrobacter koseri Citrobacter freundii Enterobacter aerogenes Enterobacter cloacae Klebsiella oxytoca Morganella morganii Proteus vulgaris
Results for Enterobacteriaceae tested with Gatifloxacin should only be reported for isolates recovered from the urinary tract.
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Atteime Hovelz da Freddie M. Poole
Division of Clinical Laboratory Devices
Prescription Use (Per 21 CFR 801.109) 510(k) Number.
Over-The-Counter Use Optional Format 1-2-96
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.