← Product Code [JWY](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JWY) · K984144

# SYSTEM U3 FOR AMOXICILLIN/CLAVULANIC ACID (K984144)

_Ab Biodisk · JWY · Feb 9, 1999 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JWY/K984144

## Device Facts

- **Applicant:** Ab Biodisk
- **Product Code:** [JWY](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JWY.md)
- **Decision Date:** Feb 9, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1640
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

System U3™ is a qualitative technique for antimicrobial susceptibility testing of nonfastidious Gram negative and Gram positive aerobic bacteria, such as Enterobacteriaceae, Pseudomonas, Staphylococcus and Enterococcus species. The system comprises an antibiotic gradient with MIC break-point concentrations and is used to determine the susceptibility categories of microorganisms to different antimicrobial agents, as tested on agar media using overnight incubation. This 510(k) application is for System U3/Amoxicillin/clavulanic acid susceptibility testing based on the MIC breakpoints of Susceptible ≤ 4/2 µg/ml and Resistant ≥ 8/4 µg/ml for use with Staphylococcus, and Susceptible ≤ 8/4 µg/ml, Intermediate 16/8 µg/ml and Resistant ≥ 32/16 µg/ml for use with non-fastidious Gram negative and Gram positive aerobic bacteria.

## Device Story

System U3 is an in vitro diagnostic device for antimicrobial susceptibility testing. It utilizes an antibiotic gradient on agar media to determine the Minimum Inhibitory Concentration (MIC) of Amoxicillin/clavulanic acid. The system is used by laboratory professionals to categorize microorganisms as susceptible, intermediate, or resistant based on established MIC breakpoints. Following overnight incubation, the clinician observes the growth inhibition pattern to determine the susceptibility category. This information assists healthcare providers in selecting appropriate antibiotic therapy for patients with bacterial infections.

## Clinical Evidence

No clinical data provided; device relies on established MIC breakpoint performance characteristics for antimicrobial susceptibility testing.

## Technological Characteristics

In vitro diagnostic test utilizing an antibiotic gradient on agar media. Qualitative susceptibility testing based on MIC breakpoints. Manual or standard laboratory incubation procedure.

## Regulatory Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

FEB - 9 1999

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Anne Bolmström President AB BIODISK Dalvägen 10 S-169 56 Solna Sweden

Re: K984144

Trade Name: System U3TM for Amoxicillin/clavulanic acid Regulatory Class: II Product Code: JWY Dated: November 13, 1998 Received: November 18, 1998

Dear Ms. Bolmström:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{1}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## 510(k) Number (if known):

Device Name: System U3™ for Amoxicillin/clavulanic acid

## Indications For Use:

For in vitro diagnostic use:

System U3™ is a qualitative technique for antimicrobial susceptibility testing of nonfastidious Gram negative and Gram positive aerobic bacteria, such as Enterobacteriaceae, Pseudomonas, Staphylococcus and Enterococcus species. The system comprises an antibiotic gradient with MIC break-point concentrations and is used to determine the susceptibility categories of microorganisms to different antimicrobial agents, as tested on agar media using overnight incubation.

This 510(k) application is for System U3/Amoxicillin/clavulanic acid susceptibility testing based on the MIC breakpoints of Susceptible ≤ 4/2 µg/ml and Resistant ≥ 8/4 µg/ml for use with Staphylococcus, and Susceptible ≤ 8/4 µg/ml, Intermediate 16/8 µg/ml and Resistant ≥ 32/16 µg/ml for use with non-fastidious Gram negative and Gram positive aerobic bacteria.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of.Device Evaluation (ODE)

|  | <div align="center">Woody Dubes</div>         |
|--|-----------------------------------------------|
|  | <div align="center">(Division Sign-Off)</div> |

Division of Clinical Laboratory Devices

| 510(k) Number | K984144 |
|---------------|---------|
|---------------|---------|

| Prescription Use<br>(Per 21 CFR 801.109) | <div align="center">X</div> | OR | Over-The Counter Use |
|------------------------------------------|-----------------------------|----|----------------------|
|------------------------------------------|-----------------------------|----|----------------------|

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JWY/K984144](https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JWY/K984144)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).