K981139 · Ab Biodisk · JWY · Jun 1, 1998 · Microbiology
Device Facts
Record ID
K981139
Device Name
ETEST CEFTRIAXONE
Applicant
Ab Biodisk
Product Code
JWY · Microbiology
Decision Date
Jun 1, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1640
Device Class
Class 2
Indications for Use
For in vitro diagnostic use: Etest is a quantitative technique for the determination of antimicrobial susceptibility of both non-fastidious Gram negative and Gram positive aerobic bacteria, such as Enterobacteriaceae, Pseudomonas, Staphylococcus and Enterococcus species and fastidious bacteria, such as anaerobes, Pneumococcus and Haemophilus species. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC) in ug/ml of individual antibiotics against bacteria as tested on agar media by overnight incubation. This Etest 510(k) application is for MIC determination of Ceftriaxone in the range of 0.016 - 256 µg/ml or 0.002 - 32 µg/ml for H. influenzae, N. gonorrhoeae and Streptococcus spp.
Device Story
Etest is a quantitative in vitro diagnostic device for determining antimicrobial susceptibility. It utilizes a predefined antibiotic gradient on a test strip applied to agar media inoculated with bacterial samples. Following overnight incubation, the device allows for the determination of the Minimum Inhibitory Concentration (MIC) of Ceftriaxone. Used in clinical laboratories by trained personnel; results are read visually by observing the inhibition ellipse intersection with the strip. The MIC value assists clinicians in selecting appropriate antibiotic therapy for patients with bacterial infections.
Clinical Evidence
No clinical data provided; device relies on bench testing for performance validation of MIC determination.
Technological Characteristics
In vitro diagnostic test strip containing a predefined antibiotic gradient. Principle: diffusion of antibiotic into agar media to establish a concentration gradient. Form factor: strip for manual application. No electronic components, software, or energy source required.
Indications for Use
Indicated for in vitro diagnostic determination of antimicrobial susceptibility (MIC) of non-fastidious Gram-negative/positive aerobic bacteria (e.g., Enterobacteriaceae, Pseudomonas, Staphylococcus, Enterococcus) and fastidious bacteria (e.g., anaerobes, Pneumococcus, Haemophilus) to Ceftriaxone.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
K980345 — ETEST CEFTRIAXONE · Ab Biodisk · May 27, 1998
K981138 — ETEST CIPROFLOXACIN · Ab Biodisk · Jun 5, 1998
K042390 — ETEST FOR GEMIFLOXACIN · Ab Biodisk · Nov 19, 2004
K990722 — ETEST FOR CEFUROXIME · Ab Biodisk · Apr 22, 1999
K980603 — ETEST TROVAFLOXACIN · Ab Biodisk · May 13, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around an emblem. The emblem is a stylized image of an eagle with three flowing lines extending from its head, representing the department's mission to protect and promote the health and well-being of Americans.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JUN 1 1998
Ms. Anne Bolmstrom AB Biodisk Dalvagen 10,S-169 56 Solna, Sweden
Re: K981139 Trade Name: Etest® Ceftriaxone Regulatory Class: II Product Code: JWY Dated: March 26, 1998 Received: March 30, 1998
Dear Ms. Bolmstrom:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{1}------------------------------------------------
### Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Dutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# 510(k) Number (if known): K945433 98 //3/
Device Name: Etest® Ceftriaxone
## Indications For Use:
For in vitro diagnostic use:
Etest is a quantitative technique for the determination of antimicrobial susceptibility of both non-fastidious Gram negative and Gram positive aerobic bacteria, such as Enterobacteriaceae, Pseudomonas, Staphylococcus and Enterococcus species and fastidious bacteria, such as anaerobes, Pneumococcus and Haemophilus species. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC) in ug/ml of individual antibiotics against bacteria as tested on agar media by overnight incubation.
This Etest 510(k) application is for MIC determination of Ceftriaxone in the range of 0.016 - 256 µg/ml or 0.002 - 32 µg/ml for H. influenzae, N. gonorrhoeae and Streptococcus spp.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Woody Dubois
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K981139
OR
Prescription Use (Per 21 CFR 801.109) Over-The Counter Use __
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.