← Product Code [JWY](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JWY) · K243201

# The Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC or Breakpoint Susceptibility System with penicillin in the dilution range of 0.015-32 ug/mL (K243201)

_Thermo Fisher Scientific · JWY · Feb 10, 2025 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JWY/K243201

## Device Facts

- **Applicant:** Thermo Fisher Scientific
- **Product Code:** [JWY](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JWY.md)
- **Decision Date:** Feb 10, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1640
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** PCCP

## Indications for Use

The Sensititre Haemophilus influenzae/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenzae, Streptococcus pneumoniae and Streptococcus species.

## Device Story

Sensititre 20-24 hour HP MIC or Breakpoint Susceptibility System performs in vitro antimicrobial susceptibility testing. Device utilizes microbroth dilution method to determine minimum inhibitory concentrations (MIC) or breakpoint susceptibility for fastidious organisms. System provides quantitative results for penicillin (0.015-32 ug/mL). Used in clinical microbiology laboratories by trained technicians. Results assist clinicians in selecting appropriate antibiotic therapy for patients with bacterial infections. System includes a Predetermined Change Control Plan (PCCP) for future modifications.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing and performance characteristics of the antimicrobial susceptibility test system.

## Technological Characteristics

Multi-well plastic microtiter plates; micro-broth dilution method. Energy source: fluorescence-based detection (ARIS HiQ/OptiRead) or visual (Vizion). Inoculation: Sensititre AIM Autoinoculator. Incubation: 34-36°C. Connectivity: Automated reader integration. Software: Automated growth detection via fluorogenic substrate cleavage.

## Regulatory Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number

K243201

B Applicant

Thermo Fisher Scientific

C Proprietary and Established Names

The Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC or Breakpoint Susceptibility System with Penicillin in the dilution range of 0.015-32 µg/mL

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  JWY | Class II | 21 CFR 866.1640 - Antimicrobial Susceptibility Test Powder | MI - Microbiology  |
|  LTT | Class II | 21 CFR 866.1640 - Antimicrobial susceptibility test powder | MI - Microbiology  |
|  LRG | Class II | 21 CFR 866.1640 - Antimicrobial susceptibility test powder | MI - Microbiology  |

## II Submission/Device Overview:

### A Purpose for Submission:

To obtain substantial equivalence determination for The Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC or Breakpoint Susceptibility System with Penicillin in the dilution range of 0.015-32 µg/mL with updated FDA-recognized breakpoints for Streptococcus pneumoniae and an expanded dilution range from 0.03-8 µg/mL cleared in K062816.

Breakpoints for Streptococcus spp. β-hemolytic group and Streptococcus spp. Viridans group (also indicated for use with this device) remain unchanged from those cleared in K062816.

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

{1}

B Measurand:

Penicillin in the dilution range 0.015 to 32 µg/mL

C Type of Test:

Quantitative antimicrobial susceptibility test (AST) growth-based detection

III Intended Use/Indications for Use:

A Intended Use(s):

The Sensititre Haemophilus influenzae/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenzae, Streptococcus pneumoniae and Streptococcus species.

B Indication(s) for Use:

The Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC or Breakpoint Susceptibility System is an in vitro diagnostic product for clinical susceptibility testing of fastidious isolates.

This 510(k) is for penicillin in the dilution range of 0.015-32 µg/mL for testing fastidious isolates on The Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC or Breakpoint Susceptibility System.

Penicillin has been shown to be active both clinically and in vitro against the following organisms according to the FDA drug label:

Streptococcus pneumoniae

Streptococcus spp. β-hemolytic group (Streptococcus agalactiae, Streptococcus dysgalactiae, Streptococcus pyogenes)

Streptococcus spp. Viridans group

C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

The following limitations were applied to penicillin testing in the appropriate sections of the device labeling that references other drugs:

Studies of the following drugs were performed with the AIM Autoinoculator. The use of an alternative inoculation system has not been evaluated.

Studies of the following drugs when tested with Streptococcus species were read using the ARIS HiQ/OptiRead and Vizion. The use of alternative read methods have not been evaluated.

K243201 - Page 2 of 13

{2}

Due to the unacceptable major error rate for S. dysgalactiae and the insufficient number of resistant or non-susceptible S. pneumoniae, Streptococcus spp. β-hemolytic group, and Streptococcus spp. Viridans group isolates evaluated, the following limitation was applied to penicillin testing in the appropriate section of the device labeling that references other drugs:

The ability of the Sensititre system to detect resistance or non-susceptibility to antimicrobics as shown below is unknown because an insufficient number of resistant or non-susceptible strains were available at the time of comparative testing. If such a strain is observed, it should be submitted to a reference laboratory.

## D Special Instrument Requirements:

- Sensititre AIM for device inoculation
- Sensititre Vizion digital viewing device
- Sensititre ARIS HiQ/OptiRead automated plate reader

## IV Device/System Characteristics:

### A Device Description:

The Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC or Breakpoint Susceptibility Plate System is an antimicrobial susceptibility test. Each plate is dosed with antimicrobial agents at appropriate dilutions. It is a micro-version of the classic broth dilution method and can provide both qualitative and quantitative susceptibility results. After inoculation, plates are sealed with an adhesive seal, incubated at 34-36°C for 20-24 hours and examined for bacterial growth.

### B Principle of Operation:

The Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC Susceptibility plates are multi-well plastic microtiter plates that contain doubled dilutions of antibacterial agents. Each plate includes antimicrobial agents at appropriate dilutions. Results can be read using the digital viewing device (Vizion) or by use of an automated plate reader (ARIS HiQ/OptiRead).

The Sensititre Vizion digital viewing device allows the panel image to be displayed on a touch screen directly from a video camera and allows the user to visually determine MIC results. The Sensititre OptiRead utilizes fluorescence technology to read the microbroth dilution plates after 20 to 24 hours incubation. The technology involves the detection of bacterial growth by monitoring the activity of specific surface enzymes produced by the test organism. Growth is determined by generating a fluorescent product from a fluorogenic substrate. The non-fluorescent substrate is prepared by conjugating a fluorescent compound to the specific enzyme substrates with a bond which prevents fluorescence. The enzymatic action of the bacterial surface enzymes on the bound non-fluorescent substrate cleaves the bond releasing fluorescence. The amount of fluorescence detected is directly related to the activity of bacterial growth. The MIC is determined by observing the lowest dilution of antimicrobial agent that inhibits growth of the organism. The non-fluorescent (fluorogenic) substrate can be added to the inoculum broth

K243201 - Page 3 of 13

{3}

which is dispensed into the test plate at the same time as the test organism, or the plates can be prepared with the substrate already added to each micro-well.

Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC plates can either be read automatically on an ARIS HiQ/OptiRead using fluorescence or by visual reading of growth on the Vizion digital viewing device.

## V Substantial Equivalence Information:

### A Predicate Device Name(s):

The Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC or Breakpoint Susceptibility System with Imipenem in the dilution range of 0.015-4 µg/mL

### B Predicate 510(k) Number(s):

K240445

### C Comparison with Predicate(s):

|  Device & Predicate Device(s): | Device: K243201 | Predicate: K240445  |
| --- | --- | --- |
|  Device Trade Name | The Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC or Breakpoint Susceptibility System with Penicillin in the dilution range of 0.015-32 µg/mL | The Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC or Breakpoint Susceptibility System with Imipenem in the dilution range of 0.015-4 µg/mL  |
|  General Device Characteristic Similarities  |   |   |
|  Intended Use | The Sensititre Haemophilus influenzae/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenzae, Streptococcus pneumoniae and Streptococcus species. | Same  |
|  Test Panel | Each 96 well plate is precision dosed with selected antimicrobial agents and substrate for the fluorescent reads, then dried. The bacterial suspension in the appropriate broth is used to rehydrate the plate. | Same  |
|  Incubation | 20-24 hours | Same  |
|  Read Method | Results can be read using fluorescence with the ARIS HiQ/OptiRead or by visual reading of growth with the | Same  |

K243201 - Page 4 of 13

{4}

|   | Vizion. |   |
| --- | --- | --- |
|  General Device Characteristic Differences  |   |   |
|  Antibiotic Range and Dilution Range | Penicillin
0.015-32 μg/mL | Imipenem
0.015-4 μg/mL  |
|  Test Organisms | Streptococcus pneumoniae
Streptococcus spp. β-hemolytic group (Streptococcus agalactiae, Streptococcus dysgalactiae, Streptococcus pyogenes)
Streptococcus spp. Viridans group | Streptococcus pneumoniae
Haemophilus influenzae  |

VI Standards/Guidance Documents Referenced:

CLSI M07, "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard - Eleventh Edition", (January 2018)

CLSI M100, "Performance Standards for Antimicrobial Susceptibility Testing; 33rd Edition", (March 2023)

Guidance for Industry and FDA: Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems, August 28, 2009

VII Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility:

A reproducibility study of The Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC or Breakpoint Susceptibility System with Penicillin was performed at three sites using a panel of fifteen (15) Streptococcus isolates from indicated species (5 S. pneumoniae, 2 S. agalactiae, 1 S. dysgalactiae, 1 S. anginosus, 1 S. anginosus group, 3 S. sanguinis, 2 S. salivarius). In addition, two Streptococcus isolates (1 S. pneumoniae and 1 S. pyogenes) with modes that were off-scale were also tested. All isolates were tested in triplicate over three days with each read method (i.e., automatically with the ARIS HiQ/OptiRead and visually with the Vizion). The Sensititre AIM autoinoculator was used for Sensititre plate inoculation. The mode MIC value was determined, and the reproducibility of the 15 isolates was calculated based on MIC values falling within $\pm 1$ doubling dilution of the mode MIC value. The reproducibility studies for both the ARIS HiQ/OptiRead and Vizion read methods demonstrated acceptable performance of $\geq 95\%$.

2. Linearity:

Not applicable.

K243201 - Page 5 of 13

{5}

3. Analytical Specificity/Interference:

Not applicable.

4. Assay Reportable Range:

Not applicable.

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

The CLSI-recommended quality control (QC) strain S. pneumoniae ATCC 49619 was tested at three sites. The QC strain was tested a minimum of 20 times per site and read automatically with the ARIS HiQ/OptiRead and visually with the Vizion. The QC strain was also tested with the reference method. The results demonstrate that The Sensititre Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC Susceptibility System with Penicillin produced quality control results within the recommended range &gt;95% of the time (Table 1).

Table 1. Quality Control Results for S. pneumoniae with Penicillin with the Reference Method, ARIS HiQ/OptiRead, and Vizion.

|  QC Organism | Expected Range (μg/mL) | Concentration (μg/mL) | Reference | ARIS HiQ/OptiRead | Vizion  |
| --- | --- | --- | --- | --- | --- |
|  S. pneumoniae ATCC 49619 | 0.25-1 μg/mL | ≤0.12 | 1 | 0 | 0  |
|   |   |  0.25 | 58 | 29 | 14  |
|   |   |  0.5 | 12 | 53 | 68  |
|   |   |  1 | 0 | 0 | 0  |
|   |   |  ≥2 | 1 | 1 | 1  |

Inoculum Density: Inoculum density checks were performed for all QC, reproducibility, challenge and clinical isolates tested. Only results from cultures with appropriate inoculum densities were reported.

Purity Checks: Purity checks were performed for all QC, reproducibility, challenge, and clinical isolates tested. Only results from pure cultures were reported.

Growth Failures: There were two growth failures each for S. pneumoniae, S. agalactiae, S. anginosus group, and S. intermedius.

ARIS HiQ/OptiRead Invalid (No fluorescence): There was one invalid for S. pneumoniae, and one invalid for S. intermedius.

6. Detection Limit:

Not applicable.

7. Assay Cut-Off:

Not applicable.

K243201 - Page 6 of 13

{6}

B Comparison Studies:

1. Method Comparison with Predicate Device:

Testing of The Sensititre Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC Susceptibility System with Penicillin was performed at two external sites and one internal site. Results were compared to those obtained with the CLSI broth microdilution reference method. Sensititre panels were inoculated using only the AIM Autoinoculator and results were read automatically by the ARIS HiQ/OptiRead and visually by the Vizion. Reference panels were inoculated according to recommendations in the M07 CLSI document and results were interpreted manually using a mirrored reader.

No inoculation system other than the AIM Autoinoculator was used in the comparative study. To address the inoculation method limitation, the following limitation was applied to penicillin testing in the appropriate section of the device labeling that references other drugs:

Studies of the following drugs were performed with the AIM Autoinoculator. The use of an alternative inoculation system has not been evaluated.

No read method other than ARIS HiQ/OptiRead and Vizion were used in the comparative study. To address the read method limitation, the following limitation was applied to penicillin testing in the appropriate section of the device labeling that references other drugs:

Studies of the following drugs when tested with Streptococcus species were read using the ARIS HiQ/OptiRead and Vizion. The use of alternative read methods have not been evaluated.

The testing conditions for the reference method consisted of the following:

- Media: per CLSI M07 guidelines for Streptococcus spp.
- Inoculum: Inoculated per CLSI M07 guidelines
- Incubation: 34-36°C in a non-CO₂ incubator for 20 to 24 hours.

Inoculation and incubation procedure for Streptococcus spp.

- Media: cation-adjusted Mueller Hinton broth with TES buffer (CAMHBT) and cation-adjusted Mueller Hinton broth with TES buffer and lysed horse blood (CAMHBT+LHB)
- Inoculum: A suspension approximating a 0.5 McFarland standard was prepared with Streptococcus spp. in 5 mL CAMHBT. A volume of 50 μL of the standardized suspension was added to 11 mL of CAMHBT + LHB. Susceptibility panels were inoculated with 100 μL of the final organism suspension using the Sensititre AIM Autoinoculator.
- Incubation: 34-36°C in a non-CO₂ incubator for 20 to 24 hours.

A total of 455 clinical isolates comprised of 157 S. pneumoniae isolates, 188 Streptococcus spp. β-hemolytic group isolates (74 S. pyogenes, 74 S. agalactiae, 40 S. dysgalactiae) and 110 Streptococcus spp. Viridans group isolates (15 S. anginosus, 18 S. anginosus group (undifferentiated), 8 S. constellatus, 2 S. intermedius, 25 S. mitis, 13 S. salivarius, 11 S.

K243201 - Page 7 of 13

{7}

sanguinis, 18 Streptococcus spp. Viridans group (undifferentiated)) as well as 198 challenge isolates comprised of 72 S. pneumoniae isolates, 67 Streptococcus spp. β-hemolytic group isolates (20 S. pyogenes, 37 S. agalactiae, 10 S. dysgalactiae) and 59 Streptococcus spp. Viridans group isolates (25 S. anginosus, 5 S. constellatus, 5 S. intermedius, 9 S. mitis, 2 S. oralis, 6 S. salivarius, 7 S. sanguinis) were evaluated with the ARIS HiQ/OptiRead and the results are provided in Table 2.

Using meningitis breakpoints for S. pneumoniae read using the ARIS HiQ/OptiRead, the combined clinical and challenge results (229 isolates) were acceptable at 98.3% and 99.1% for EA and CA, respectively. There were two major errors (2/151 = 1.3%) and no very major errors.

Using non-meningitis breakpoints for S. pneumoniae read using the ARIS HiQ/OptiRead, the combined clinical and challenge results (229 isolates) were acceptable at 98.3% and 95.6% for EA and CA, respectively. There were ten minor errors, no major errors and no very major errors.

For Streptococcus spp. β-hemolytic group read using the ARIS HiQ/OptiRead, the combined clinical and challenge results (255 isolates) were acceptable at 98.0% and 98.4% for EA and CA, respectively. There were three potential major errors (3/254 = 1.2%) and one potential very major error (1/1 = 100%). Since no category is defined other than "susceptible" for penicillin when testing Streptococcus spp. β-hemolytic group, isolates for which the MIC values are above the susceptible breakpoint are reported as non-susceptible. When categorical errors occur, these are considered potential errors. Additionally, due to the lack of a category other than "susceptible" for penicillin when testing Streptococcus spp. β-hemolytic group, further analysis of the errors is performed, and adjustments are made by considering the MIC values where the errors occurred. The potential very major error had an MIC value that was not in essential agreement with the reference MIC value; therefore, the adjusted potential very major error rate is 100% (1/1) and is not acceptable. When evaluating results by individual species, S. pyogenes had an adjusted potential very major error rate of 100% (1/1). Due to the limited number of non-susceptible isolates tested, this error was considered random and a footnote to address the potential very major error rate is included below. S. dysgalactiae had two potential major errors (2/50 = 4.0%), which is not acceptable. The two potential major errors had MIC values that were not in essential agreement with the reference MIC values; therefore, the adjusted potential major error rate is 4.0% (2/50) and is not acceptable. A limitation to address the unacceptable potential major error rate for S. dysgalactiae and the insufficient number of resistant or non-susceptible S. pneumoniae, Streptococcus spp. β-hemolytic group, and Streptococcus spp. Viridans group isolates evaluated, the following limitation was applied to penicillin testing in the appropriate section of the device labeling that references other drugs:

The ability of the Sensititre system to detect resistance or non-susceptibility to antimicrobials as shown below is unknown because an insufficient number of resistant or non-susceptible strains were available at the time of comparative testing. If such a strain is observed, it should be submitted to a reference laboratory.

For Streptococcus spp. Viridans group read using the ARIS HiQ/OptiRead, the combined clinical and challenge results (169 isolates) were acceptable at 97.6% and 95.9% for EA and CA, respectively. There were six minor errors, no major errors and no very major errors.

Table 2. Penicillin Performance of S. pneumoniae, Streptococcus spp. β-Hemolytic Group, and Streptococcus spp. Viridans Group Read by ARIS HiQ/OptiRead

K243201 - Page 8 of 13

{8}

|   | Tot | No. EA | EA % | Eval EA Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R/NS | No. S | min | major | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Streptococcus pneumoniae [≤0.06 (S), ≥0.12 (R)] meningitis breakpoints  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 157 | 155 | 98.7 | 59 | 57 | 96.6 | 156 | 99.4 | 42 | 115 | NA | 1 | 0  |
|  Challenge | 72 | 70 | 97.2 | 42 | 40 | 95.2 | 71 | 98.6 | 36 | 36 | NA | 1 | 0  |
|  Combined | 229 | 225 | 98.3 | 101 | 97 | 96.0 | 227 | 99.1 | 78 | 151 | NA | 2 | 0  |
|  Streptococcus pneumoniae [≤2 (S), 4 (I), ≥8 (R)] non-meningitis breakpoints  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 157 | 155 | 98.7 | 59 | 57 | 96.6 | 154 | 98.1 | 1 | 149 | 3 | 0 | 0  |
|  Challenge | 72 | 70 | 97.2 | 42 | 40 | 95.2 | 65 | 90.3 | 2 | 67 | 7 | 0 | 0  |
|  Combined | 229 | 225 | 98.3 | 101 | 97 | 96.0 | 219 | 95.6 | 3 | 151 | 10 | 0 | 0  |
|  Streptococcus spp. β-hemolytic group [≤0.12 (S)]  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 188 | 184 | 97.9 | 75 | 71 | 94.7 | 185 | 98.4 | 1 | 187 | NA | 2 | 1  |
|  Challenge | 67 | 66 | 98.5 | 40 | 39 | 97.5 | 66 | 98.5 | 0 | 67 | NA | 1 | 0  |
|  Combined | 255 | 250 | 98.0 | 115 | 110 | 95.7 | 251 | 98.4 | 1 | 254 | NA | 3 | 1  |
|  Streptococcus spp. Viridans group [≤0.12 (S), 0.25-2 (I), ≥4 (R)]  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 110 | 108 | 98.2 | 93 | 91 | 97.9 | 106 | 96.4 | 0 | 88 | 4 | 0 | 0  |
|  Challenge | 59 | 57 | 96.6 | 51 | 49 | 96.1 | 56 | 94.9 | 9 | 43 | 3 | 0 | 0  |
|  Combined | 169 | 165 | 97.6 | 144 | 140 | 97.2 | 162 | 95.9 | 9 | 131 | 7 | 0 | 0  |

NA - Not applicable
EA - Essential Agreement
CA - Categorical Agreement
S - Susceptible
NS - Non-susceptible
Maj - Major Discrepancies
SDD - Susceptible-dose dependent
EVAL - Evaluable MICs
R - Resistant
min - Minor Discrepancies
vmj - Very Major Discrepancies

Essential agreement (EA) occurs when the result of the reference method and that of the Sensititre panel are within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the reference method and the Sensititre panel or those in which an off-scale result is at least two doubling dilutions from the on-scale result. Category agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the Sensititre panel.

A total of 457 clinical isolates comprised of 158 S. pneumoniae isolates, 188 Streptococcus spp. β-hemolytic group isolates (74 S. pyogenes, 74 S. agalactiae, 40 S. dysgalactiae) and 111 Streptococcus spp. Viridans group isolates (15 S. anginosus, 18 S. anginosus group (undifferentiated), 8 S. constellatus, 3 S. intermedius, 25 S. mitis, 13 S. salivarius, 11 S. sanguinis, 18 Streptococcus spp. Viridans group (undifferentiated)) as well as 198 challenge isolates comprised of 72 S. pneumoniae isolates, 67 Streptococcus spp. β-hemolytic group isolates (20 S. pyogenes, 37 S. agalactiae, 10 S. dysgalactiae) and 59 Streptococcus spp. Viridans group isolates (25 S. anginosus, 5 S. constellatus, 5 S. intermedius, 9 S. mitis, 2 S. oralis, 6 S. salivarius, 7 S. sanguinis) were evaluated with the Vizion and the results were provided in Table 3.

Using meningitis breakpoints for S. pneumoniae read using the Vizion, the combined clinical and challenge results (230 isolates) were acceptable at 98.7% and 98.7% for EA and CA, respectively. There were three major errors (3/152 = 2.0%) and no very major errors. Using non-meningitis breakpoints for S. pneumoniae read using the Vizion, the combined clinical and challenge results (230 isolates) were acceptable at 98.7% and 93.9% for EA and CA, respectively. There were 14 minor errors, no major errors and no very major errors.

K243201 - Page 9 of 13

{9}

For Streptococcus spp.  $\beta$ -hemolytic group read using the Vizion, the combined clinical and challenge results (255 isolates) were acceptable at  $98.8\%$  and  $99.2\%$  for EA and CA, respectively. There was one potential major error  $(1/254 = 0.4\%)$  and one potential very major error  $(1/1 = 100\%)$ . These are considered potential errors since no category is defined other than "susceptible". Additionally, due to the lack of a category other than "susceptible" for penicillin when testing Streptococcus spp.  $\beta$ -hemolytic group, further analysis of the errors is performed, and adjustments are made by considering the MIC values where the errors occurred. The potential very major error had an MIC value that was not in essential agreement with the reference MIC value; therefore, the adjusted potential very major error rate is  $100\%$  (1/1) and is not acceptable. When evaluating results by individual species, S. pyogenes had an adjusted potential very major error rate of  $100\%$  (1/1). Due to the lack of non-susceptible S. pyogenes isolates evaluated, the one potential VMJ error is considered random the following performance footnote was added to the device labeling:

The 1 potential very major error observed was considered a random error due to the limited number of non-susceptible S. pyogenes isolates tested with penicillin.

For Streptococcus spp. Viridans group read using the Vizion, the combined clinical and challenge results (170 isolates) were acceptable at  $96.5\%$  and  $6.5\%$  for EA and CA, respectively. There were six minor errors, no major errors and no very major errors.

Table 3. Penicillin Performance of S. pneumoniae, Streptococcus spp.  $\beta$  -Hemolytic Group, and Streptococcus spp. Viridans Group Read by Vizion

|   | Tot | No. EA | EA % | Eval EA Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R/NS | No. S | min | major | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Streptococcus pneumoniae [≤0.06 (S), ≥0.12 (R)] meningitis breakpoints  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 158 | 158 | 100 | 58 | 58 | 100 | 157 | 99.4 | 42 | 116 | NA | 1 | 0  |
|  Challenge | 72 | 69 | 95.8 | 42 | 39 | 92.9 | 70 | 97.2 | 36 | 36 | NA | 2 | 0  |
|  Combined | 230 | 227 | 98.7 | 100 | 97 | 97.0 | 227 | 98.7 | 78 | 152 | NA | 3 | 0  |
|  Streptococcus pneumoniae [≤2 (S), 4 (I), ≥8 (R)] non-meningitis breakpoints  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 158 | 158 | 100 | 58 | 58 | 100 | 153 | 96.8 | 1 | 150 | 5 | 0 | 0  |
|  Challenge | 72 | 69 | 95.8 | 42 | 39 | 92.9 | 63 | 87.5 | 2 | 67 | 9 | 0 | 0  |
|  Combined | 230 | 227 | 98.7 | 100 | 97 | 97.0 | 216 | 93.9 | 3 | 217 | 14 | 0 | 0  |
|  Streptococcus spp. β-hemolytic group [≤0.12 (S)]  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 188 | 186 | 98.9 | 75 | 73 | 97.3 | 187 | 99.5 | 1 | 187 | NA | 0 | 1  |
|  Challenge | 67 | 66 | 98.5 | 39 | 38 | 97.4 | 66 | 98.5 | 0 | 67 | NA | 1 | 0  |
|  Combined | 255 | 252 | 98.8 | 114 | 111 | 97.4 | 253 | 99.2 | 1 | 254 | NA | 1 | 1  |
|  Streptococcus spp. Viridans group [≤0.12 (S), 0.25-2 (I), ≥4 (R)]  |   |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 111 | 107 | 96.4 | 92 | 88 | 95.7 | 107 | 96.4 | 0 | 89 | 4 | 0 | 0  |
|  Challenge | 59 | 57 | 96.6 | 51 | 49 | 96.1 | 57 | 96.6 | 9 | 43 | 2 | 0 | 0  |
|  Combined | 170 | 164 | 96.5 | 143 | 137 | 95.8 | 164 | 96.5 | 9 | 132 | 6 | 0 | 0  |

NA - Not applicable
EA - Essential Agreement
CA - Categorical Agreement
S - Susceptible
NS - Non-susceptible
Maj - Major Discrepancies

SDD - Susceptible-dose dependent

EVAL - Evaluable MICs

R - Resistant

min - Minor Discrepancies

vmj - Very Major Discrepancies

Essential agreement (EA) occurs when the result of the reference method and that of the Sensititre panel are within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the

K243201 - Page 10 of 13

{10}

reference method and the Sensititre panel or those in which an off-scale result is at least two doubling dilutions from the on-scale result. Category agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the Sensititre panel.

# MIC Trending

A trending analysis was conducted using the combined data (clinical and challenge) obtained for both the ARIS HiQ/OptiRead and the Vizion for S. pneumoniae, Streptococcus spp.  $\beta$ -hemolytic group, and Streptococcus spp. Viridans group. This trending calculation takes into account MIC values that are determined to be one or more doubling dilutions lower or higher than the reference method irrespective of whether the device MIC values are on-scale or not. Results that are not clearly at least one dilution lower, at least one dilution higher, or in exact agreement with the CLSI reference method are not considered in the trending analysis.

Species for which the difference between the percentage of isolates with higher vs. lower readings was  $&gt;30\%$  and for which the confidence interval was determined to be statistically significant were considered to show evidence of trending. Trending that shows higher or lower MIC values compared to the reference is addressed in the labeling.

Evaluation of results for S. pneumoniae, Streptococcus spp.  $\beta$ -hemolytic group, and Streptococcus spp. Viridans group with penicillin using the ARIS HiQ/OptiRead and the Vizion are summarized in Table 4. A trend toward higher MIC values was observed for S. dysgalactiae using the ARIS HiQ/OptiRead when compared to the CLSI broth microdilution reference method. No trending was observed with the Vizion.

To address the MIC trending, the following footnote was included in the performance table in the device labeling:

For ARIS HiQ/OptiRead:

Sensititre Penicillin MIC values tended to be in exact agreement or at least one doubling dilution higher when testing S. dysgalactiae with the ARIS HiQ/OptiRead compared to the CLSI broth microdilution reference method.

Table 4. Penicillin Trending Analysis for S. pneumoniae, Streptococcus spp.  $\beta$  -Hemolytic Group,and Streptococcus spp. Viridans Group with ARIS HiQ/OptiRead and Vizion

|  Read Method | Organism | Total Evaluable for Trending | ≥1 Dilution Lower No. (%) | Exact No. (%) | ≥1 Dilution Higher No. (%) | Percent Difference (95% CI) | Trending Noted  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  ARIS HiQ/OptiRead | S. pneumoniae | 125 | 24 (19.2) | 65 | 36 (28.8) | 10%, (-1% to 20%) | No  |
|   |  S. agalactiae | 111 | 14 (12.6) | 68 | 29 (26.1) | 14% (3% to 24%) | No  |
|   |  S. dysgalactiae | 6 | 0 (0) | 1 | 5 (83.3) | 83% (28% to 97%) | Yes, high  |
|   |  S. pyogenes | 5 | 2 (40) | 0 | 3 (60) | 20% (-32% to 60%) | No  |
|   |  Streptococcus spp. Viridans group | 157 | 36 (22.9) | 83 | 38 (24.2) | 1% (-8% to 11%) | No  |

K243201 - Page 11 of 13

{11}

|  Vizion | S. pneumoniae | 124 | 24 (19.4) | 55 | 45 (36.3) | 17% (6% to 28%) | No  |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |  S. agalactiae | 111 | 8 (7.2) | 77 | 26 (23.4) | 16% (7% to 26%) | No  |
|   |  S. dysgalactiae | 5 | 2 (40) | 0 | 3 (60) | 20% (-32% to 60%) | No  |
|   |  S. pyogenes | 5 | 2 (40) | 1 | 2 (40) | 0% (-46% to 46%) | No  |
|   |  Streptococcus spp.
Viridans group | 156 | 30 (19.2) | 86 (55.1) | 40 (25.6) | 6% (-3% to 16%) | No  |

## Testing/Reporting MICs for Non-indicated Species.

For this review, the interpretive criteria are applied to the organisms/organism groups according to the FDA STIC website. As required under 511A(2)(2)(B) of the Federal Food, Drug and Cosmetic Act, the following statement is included in the Warnings and Precautions section of the device labeling to address testing and reporting of non-indicated species:

The safety and efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved.

2. **Matrix Comparison:**

Not applicable.

## C Clinical Studies:

1. **Clinical Sensitivity:**

Not applicable.

2. **Clinical Specificity:**

Not applicable.

3. **Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):**

Not applicable.

## D Clinical Cut-Off:

Not applicable.

## E Expected Values/Reference Range:

K243201 - Page 12 of 13

{12}

Table 5. FDA-Recognized Interpretive Criteria for Penicillin

|  Organism | Minimum Inhibitory Concentrations (μg/mL)a  |   |   |
| --- | --- | --- | --- |
|   |  Susceptible | Intermediate | Resistant  |
|  Streptococcus pneumoniae (meningitis) | ≤0.06 | - | ≥0.12  |
|  Streptococcus pneumoniae (non-meningitis) | ≤2 | 4 | ≥8  |
|  Streptococcus spp. β-hemolytic group | ≤0.12 | - | -  |
|  Streptococcus spp. Viridans group | ≤0.12 | 0.25-2 | ≥4  |

a According to the FDA STIC Webpage

## VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

## IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission incorporated by reference a breakpoint change protocol that was reviewed and accepted by FDA in submission K231994 cleared on August 25, 2023. This referenced protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage (https://www.fda.gov/drugs/development-resources/fda-recognized-antimicrobial-susceptibility-test-interpretive-criteria). The referenced protocol outlined the specific procedures and acceptance criteria that Thermo Fisher Scientific intends to use to evaluate The Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC or Breakpoint Susceptibility System with Penicillin when revised breakpoints for penicillin are published on the FDA STIC webpage. The breakpoint change protocol included with the submission indicated that if specific criteria are met, Thermo Fisher Scientific will update the penicillin device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.

K243201 - Page 13 of 13

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JWY/K243201](https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JWY/K243201)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
