Sensititre 20-24 hour Haemophilus influenzaelStreptococcus pneumoniae MIC or Breakpoint Susceptibility System with Ceftolozane-tazobactam in the dilution range of 0.03/4-64/4ug/ml
Sensititre 20-24 hour Haemophilus influenzaelStreptococcus pneumoniae MIC or Breakpoint Susceptibility System with Ceftolozane-tazobactam in the dilution range of 0.03/4-64/4ug/ml
Applicant
Thermo Fisher Scientific
Product Code
JWY · Microbiology
Decision Date
Oct 31, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1640
Device Class
Class 2
Indications for Use
The Sensititre Haemophilus influenzae/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenzae, Streptococcus pneumoniae and Streptococcus species.
Device Story
The Sensititre system is an in vitro diagnostic device used for antimicrobial susceptibility testing (AST) of fastidious bacteria. It utilizes microbroth dilution to determine the Minimum Inhibitory Concentration (MIC) of antibiotics. The system consists of multi-well plates containing dried antimicrobial agents (Ceftolozane-tazobactam) at specific concentrations. Clinical laboratory personnel inoculate the plates with bacterial isolates; plates are incubated for 20-24 hours. Growth or inhibition is assessed to determine susceptibility. The output provides MIC values, which assist clinicians in selecting appropriate antibiotic therapy for patients with H. influenzae infections. The device is intended for use in clinical laboratory settings.
Clinical Evidence
No clinical data provided. The submission relies on performance data demonstrating the device's ability to provide accurate MIC results for the specified antibiotic-organism combination compared to reference methods.
Technological Characteristics
Microbroth dilution susceptibility system; multi-well plates with dried antimicrobial agents. Dilution range: 0.03/4 - 64/4 µg/ml. System is designed for 20-24 hour incubation of fastidious organisms.
Indications for Use
Indicated for clinical susceptibility testing of fastidious isolates, specifically Haemophilus influenzae, using the Sensititre 20-24 hour MIC or Breakpoint Susceptibility System. For prescription use only.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
Related Devices
K193024 — Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint, Susceptibility System with Lefamulin in the dilution range of 0.008-16ug/mL · Thermo Fisher Scientific · Jan 8, 2020
K062022 — HAEMOPHILUS/STREPTOCOCCUS PNEUMONIAE (HP) MIC PLATE · Trek Diagnostic Systems, Inc. · Sep 11, 2006
K020859 — CEFOTAXIME · Trek Diagnostic Systems, Inc. · May 15, 2002
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K232310
B Applicant
Thermo Fisher Scientific
C Proprietary and Established Names
Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System with Ceftolozane-tazobactam in the dilution range of 0.03/4-64/4ug/ml
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JWY | Class II | 21 CFR 866.1640 - Antimicrobial Susceptibility Test Powder | MI - Microbiology |
| LRG | Class II | 21 CFR 866.1640 - Antimicrobial susceptibility test powder | MI - Microbiology |
| LTT | Class II | 21 CFR 866.1640 - Antimicrobial susceptibility test powder | MI - Microbiology |
## II Submission/Device Overview:
### A Purpose for Submission:
To obtain a substantial equivalence determination for the addition of Ceftolozane-tazobactam on the Sensititre 20-24-Hour MIC Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System
### B Measurand:
Ceftolozane-tazobactam 0.03/4 - 64/4 µg/mL
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
C Type of Test:
Quantitative Antimicrobial Susceptibility Test (AST) growth-based detection
III Intended Use/Indications for Use:
A Intended Use(s):
The Sensititre Haemophilus influenzae/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenzae, Streptococcus pneumoniae and Streptococcus species.
B Indication(s) for Use:
The Sensititre 20-24-hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System is an in vitro diagnostic product for clinical susceptibility testing of fastidious isolates.
This 510(k) is for Ceftolozane-tazobactam in the dilution range of 0.03/4 - 64/4 µg/ml for testing H. influenzae on the Sensititre 20 - 24-hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System.
Ceftolozane-tazobactam has been shown to be active both clinically and in vitro against the following organisms according to the FDA drug label:
Haemophilus influenzae
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
Bold text was used to indicate updates to the limitations to include ceftolozane-tazobactam.
The evaluation of Tedizolid and Dalbavancin, with Streptococcus spp. (Streptococcus pyogenes, S. agalactiae, and S. anginosus), Delafloxacin with Streptococcus pyogenes, S. agalactiae, S. anginosus, S. pneumoniae and H. influenzae, Imipenem-relebactam with H. influenzae, Ceftolozane-tazobactam with H. influenzae, and the evaluation of Oritavancin with Streptococcus spp. (Streptococcus pyogenes, S. agalactiae, S. dysgalactiae, and S. anginosus) was performed using the AIM autoinoculator. The use of an alternative inoculation system when testing Tedizolid, Dalbavancin, Delafloxacin, Oritavancin Imipenem-Relebactam, and Ceftolozane-tazobactam has not been evaluated.
The performance of Ceftolozane-tazobactam was determined using the digital reading device (VIZION) reading method only. The use of an alternative reading method when testing Ceftolozane-tazobactam has not been evaluated.
Due to a lack of interpretive criteria other than susceptible for Ceftolozane-tazobactam, isolates of H. influenzae yielding MIC results other than Susceptible should be submitted to a reference laboratory for further testing.
K232310 - Page 2 of 10
{2}
D Special Instrument Requirements:
Sensititre AIM for device inoculation
Sensititre VIZION for plate reading
IV Device/System Characteristics:
A Device Description:
The device is an antimicrobial susceptibility test. Each plate is dosed with dried, stabilized antimicrobial agents at appropriate dilutions. It is a micro-version of the classic broth dilution method and can provide both qualitative and quantitative susceptibility results. After inoculation, plates are sealed with an adhesive seal, incubated at 34-36 °C for 20-24 hours and examined for bacterial growth.
B Principle of Operation:
The Sensititre Haemophilus influenzae/Streptococcus pneumoniae MIC Susceptibility plates are multi-well plastic microtiter plates that contain doubled dilutions of antibacterial agents. Each plate includes antimicrobial agents at appropriate dilutions. Results can be read using the digital reading device (VIZION) or by use of an automated reader (ARIS/OptiRead).
The digital reading device (VIZION) allows the panel image to be displayed on a touch screen directly from a video camera and allows the user to manually select MIC results. The Sensititre OptiRead utilizes fluorescence technology to read the microbroth dilution plates after 20 to 24 hours incubation. The technology involves the detection of bacterial growth by monitoring the activity of specific surface enzymes produced by the test organism. Growth is determined by generating a fluorescent product from a fluorogenic substrate. The non-fluorescent substrate is prepared by conjugating a fluorescent compound to the specific enzyme substrates with a bond which prevents fluorescence. The enzymatic action of the bacterial surface enzymes on the bound non-fluorescent substrate cleaves the bond releasing fluorescence. The amount of fluorescence detected is directly related to the activity of bacterial growth. The MIC is determined by observing the lowest dilution of antimicrobial agent that inhibits growth of the organism. The non-fluorescent (fluorogenic) substrate can be added to the inoculum broth which is dispensed into the test plate at the same time as the test organism, or the plates can be prepared with the substrate already added to each micro-well.
Streptococcus pneumoniae and Streptococcus spp. plates can either be read on the digital reading device (VIZION) or automatically on the Sensititre ARIS/OptiRead. H. influenzae can only be read on the digital reading device (VIZION).
V Substantial Equivalence Information:
A Predicate Device Name(s):
Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint, Susceptibility System with Lefamulin in the dilution range of 0.008-16ug/mL
K232310 - Page 3 of 10
{3}
B Predicate 510(k) Number(s):
K193024
C Comparison with Predicate(s):
| Device & Predicate Device(s): | Device: K232310 | Predicate: K193024 |
| --- | --- | --- |
| Device Trade Name | Sensititre 20-24 hour
Haemophilus / Streptococcus pneumoniae MIC or Breakpoint Susceptibility System with Ceftolozane-tazobactam in the dilution range of 0.03/4 - 64/4 μg/ml | Sensititre 20-24 hour
Haemophilus / Streptococcus pneumoniae (HP) MIC or Breakpoint Susceptibility System with Lefamulin in the dilution range of 0.008-16μg/mL |
| General Device Characteristic Similarities | | |
| Intended Use | The Sensititre Haemophilus influenzae/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenzae, Streptococcus pneumoniae and Streptococcus species. | Same |
| Test Panel | 96 well plate is dosed with selected antimicrobial agents and substrate for the fluorescent reads, then dried. The bacterial suspension in the appropriate broth is used to rehydrate the plate | Same |
| Incubation | 20-24 hours | Same |
| Reading method | Results can be read using the VIZION (digital viewing device) only | Same |
| General Device Characteristic Differences | | |
| Test Organisms | Haemophilus influenzae | Haemophilus influenzae and Streptococcus spp. |
K232310 - Page 4 of 10
{4}
| Device & Predicate Device(s): | Device: K232310 | Predicate: K193024 |
| --- | --- | --- |
| Antibiotic and Dilution Range | Ceftolozane-tazobactam 0.03/4 - 64/4 μg/ml | Lefamulin 0.008-16μg/ml |
VI Standards/Guidance Documents Referenced:
Guidance for Industry and FDA - Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems – August 28, 2009.
CLSI M7-A11: Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard – Eleventh Edition
CLSI M100-S26: Performance Standards for Antimicrobial Susceptibility Testing; Approved Standard – 26th Informational Supplement
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
A reproducibility study was performed at three sites using a panel of twenty-five (25) isolates of *H. influenzae* for a total of seventy-five (75) data points read using the digital viewing device (VIZION). The Sensititre AIM inoculator was used for Sensititre plate inoculation. The mode MIC value was determined, and the reproducibility was calculated based on MIC values falling within ±1 dilution of the mode MIC value. Reproducibility was greater than 97% for *H. influenzae* read using the digital reading device (VIZION) with Ceftolozane-tazobactam. Results were considered to be acceptable.
2. Linearity:
Not applicable
3. Analytical Specificity/Interference:
Not applicable
4. Assay Reportable Range:
Not applicable
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
The quality control strain recommended by CLSI, namely *H. influenzae* ATCC 49247, was tested with Ceftolozane-tazobactam at three sites. The QC strain was tested a minimum of 20 times per site and read using the digital reading device (VIZION). The QC strain was also
K232310 - Page 5 of 10
{5}
tested with the reference method. The results demonstrate that the Sensititre Haemophilus influenzae/Streptococcus pneumoniae (HP) MIC Susceptibility plates with Ceftolozane-tazobactam produced quality control results in the recommended range >95% of time (Table 1).
Table 1. QC Results for H. influenzae with Ceftolozane-tazobactam with the Reference Method and the Digital Reading Device (Vizion)
| QC Organism | Expected Range (μg/mL) | Concentration (μg/mL) | Reference | Sensititre Digital Reading Device (VIZION) |
| --- | --- | --- | --- | --- |
| H. influenzae ATCC 49247 | 0.5/4 – 2/4 μg/mL | 0.25/4 | - | - |
| | | 0.5/4 | 3 | 17 |
| | | 1/4 | 50 | 39 |
| | | 2/4 | 7 | 4 |
| | | 4/4 | - | - |
Inoculum Density: Inoculum density checks were performed for all QC, reproducibility and challenge isolates and clinical isolates tested.
Purity Checks: Purity checks were performed each day for each clinical, challenge, reproducibility and QC strain tested. Only results from pure cultures were reported.
Growth Failure: There were no growth failures for H. influenzae.
6. Detection Limit:
Not applicable
7. Assay Cut-Off:
Not applicable
B Comparison Studies:
1. Method Comparison with Predicate Device:
Testing of the Sensititre Haemophilus influenzae/Streptococcus pneumoniae MIC Susceptibility plates with Ceftolozane-tazobactam was performed at two external sites and one internal site. Results were compared to results obtained with the CLSI broth microdilution reference panel. Sensititre panels were inoculated using only the AIM Autoinoculator and results were interpreted using only the digital reading device (VIZION) for H. influenzae. Reference panels were inoculated according to recommendations in the M07 CLSI document and results were interpreted manually using a mirrored reader.
No inoculation system other than the AIM Autoinoculator was used in the comparative study. To address the inoculation method limitation, the sponsor modified the existing method limitation in the device labeling to include H. influenzae with Ceftolozane-tazobactam (modifications in bold font):
K232310 - Page 6 of 10
{6}
The evaluation of Tedizolid and Dalbavancin, with Streptococcus spp. (Streptococcus pyogenes, S. agalactiae, and S. anginosus), Delafloxacin with Streptococcus pyogenes, S. agalactiae, S. anginosus, S. pneumoniae and H. influenzae, Imipenem-relebactam with H. influenzae, Ceftolozane-tazobactam with H. influenzae, and the evaluation of Oritavancin with Streptococcus spp. (Streptococcus pyogenes, S. agalactiae, S. dysgalactiae, and S. anginosus) was performed using the AIM autoinoculator. The use of an alternative inoculation system when testing Tedizolid, Dalbavancin, Delafloxacin, Oritavancin Imipenem-Relebactam, and Ceftolozane-tazobactam has not been evaluated.
No reading method other than the digital reading device (VIZION) was used in the comparative study. To address the read method limitation, the sponsor added a read method limitation in the device labeling:
The performance of Ceftolozane-tazobactam for H. influenzae was determined using the VIZION reading method only. The use of an alternative reading method when testing Ceftolozane-tazobactam has not been evaluated.
The testing conditions for the reference method consisted of the following:
- Media: per CLSI M07 guidelines for H. influenzae
- Inoculum: Inoculated per CLSI M07 guidelines
- Incubation: 34 - 36° C in a non-CO₂ incubator for 20 to 24 hours.
H. influenzae
- Media: CAMHBT and Haemophilus Test Medium (HTM)
- Inoculum: A suspension approximating a 05 McFarland standard was prepared with H. influenzae in 5 mL CAMHBT. A volume of 50 μL of the standardized suspension was added to 11 mL of HTM. Susceptibility panels were inoculated with 100 μL of the final organism suspension using the Sensititre AIM.
- Incubation: 34 - 36° C in a non-CO₂ incubator for 20 to 24 hours.
A total of 248 H. influenzae clinical isolates and 50 H. influenzae challenge isolates were evaluated. The sponsor evaluated 288 susceptible organisms and 10 non-susceptible organisms. For H. influenzae read using the digital reading device (VIZION), the combined clinical and challenge results (298 isolates) were acceptable at 97.7% and 99.3% for EA and CA, respectively. There was one potential major error (1/288 = 0.3%) and one potential very major error (1/10 = 10%).
Due to the lack of an intermediate interpretive criterion (ceftolozane-tazobactam only has a "susceptible" category), further analysis of the categorical errors was performed and adjustments made by considering the MIC values where the errors occurred. The potential very major error was one doubling dilution from the reference and thus in essential agreement. Therefore, the adjusted very major error rate is 0% (0/10), which is acceptable. To address the adjustment of the potential very major error, the following performance footnote was included in the device labeling:
K232310 - Page 7 of 10
{7}
"The overall potential very major error rate for Ceftolozane-tazobactam when testing H. influenzae clinical and challenge isolates is 10% (1/10). Based on the essential agreement and lack of an intermediate breakpoint for Ceftolozane-tazobactam, the overall adjusted potential very major error rate for H. influenzae clinical and challenge isolates is 0% (0/10)."
To address the susceptible-only breakpoint for ceftolozane-tazobactam, the following limitation was included in the device labeling:
Due to a lack of interpretive criteria other than susceptible for Ceftolozane-tazobactam, isolates of H. influenzae yielding MIC results other than "Susceptible" should be submitted to a reference laboratory for further testing.
Table 2. Ceftolozane-tazobactam Performance with Clinical and Challenge H. influenzae Isolates
| | Tot | EA N | EA % | Eval Tot | Eval EA N | Eval EA % | CA Tot | CA % | No. NS | No. S | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| H. influenza (Susceptible ≤ 0.5/4 μg/mL) | | | | | | | | | | | | | |
| Clinical | 248 | 241 | 97.2 | 238 | 231 | 97.1 | 247 | 99.6 | 10 | 238 | - | 0 | 1 |
| Challenge | 50 | 50 | 100 | 46 | 46 | 100 | 49 | 98.0 | 0 | 50 | - | 1 | 0 |
| Total | 298 | 291 | 97.7 | 284 | 277 | 97.5 | 296 | 99.3 | 10 | 288 | - | 1 | 1 |
EA - Essential Agreement
CA - Categorical Agreement
S - Susceptible
Maj - Major Discrepancies
EVAL - Evaluable MICs
NS - Non-Susceptible
min - Minor Discrepancies
vmj - Very Major Discrepancies
Essential agreement (EA) occurs when the result of the reference method and that of the Sensititre panel are within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the reference method and the Sensititre panel or those in which an off-scale result is at least two doubling dilutions from the on-scale result. Category agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the Sensititre panel.
## Trending
A trending analysis was conducted using the combined data (clinical and challenge) obtained for the VIZION read method for H. influenzae. This trending calculation takes into account MIC values that are determined to be one or more doubling dilutions lower or higher than the reference method irrespective of whether the device MIC values are on-scale or not. Results that are not clearly at least one dilution lower at least one dilution higher or in exact agreement with the CLSI reference method are not considered in the trending analysis.
Species for which the difference between the percentage of isolates with higher vs. lower readings was > 30% and for which the confidence interval was determined to be statistically significant were considered to show evidence of trending. Trending that shows higher or lower MIC values compared to the reference is addressed in the labeling.
Evaluation of results for H. influenzae and Ceftolozane-tazobactam using the digital reading device (VIZION) did not indicate trending.
K232310 - Page 8 of 10
{8}
Table 3. Trending Analysis for H. influenzae with Ceftolozane-tazobactam
| Organism | Total Evaluable for Trending | ≥1 Dilution lower No. (%) | Exact No. (%) | ≥1 Dilution Higher No. (%) | Percent Difference (CI) | Trending Noted |
| --- | --- | --- | --- | --- | --- | --- |
| H. influenzae | 289 | 16 (5.5) | 179 (62.0) | 94 (32.5) | 27.0% (21% to 33%) | No |
Testing/Reporting MICs for Non-indicated Species.
For this review, the interpretive criteria are applied to the organisms/organism groups according to the FDA STIC website. As required under 511A(2)(2)(B) of the Federal Food, Drug and Cosmetic Act, the following statement is included in the Warnings and Precautions section of the device labeling to address testing and reporting of non-indicated species:
The safety and efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved.
2. Matrix Comparison: Not applicable
C Clinical Studies:
1. Clinical Sensitivity: Not applicable
2. Clinical Specificity: Not applicable
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not applicable
D Clinical Cut-Off: Not applicable
E Expected Values/Reference Range:
Table 4. FDA-Identified Interpretive Criteria for Ceftolozane-tazobactam
| Organism | Interpretive Criteria for Ceftolozane-tazobactama | | |
| --- | --- | --- | --- |
| | Susceptible | Intermediate | Resistant |
| H. influenzae | ≤0.5/4 | - | - |
aAccording to the FDA STIC Webpage
K232310 - Page 9 of 10
{9}
K232310 - Page 10 of 10
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a breakpoint change protocol that was reviewed and accepted by FDA. This protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage (https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm). The protocol outlined the specific procedures and acceptance criteria that ThermoFisher intends to use to evaluate the Sensititre 20–24-hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System with Ceftolozane-tazobactam when revised breakpoints for Ceftolozane-tazobactam are published on the FDA STIC webpage. The breakpoint change protocol included with the submission indicated that if specific criteria are met, ThermoFisher will update the Ceftolozane-tazobactam device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.