MTS Ampicillin-Sulbactam 0.016/0.008 - 256/128 µg/mL

K192345 · Liofilchem s.r.l. · JWY · Oct 31, 2019 · Microbiology

Device Facts

Record IDK192345
Device NameMTS Ampicillin-Sulbactam 0.016/0.008 - 256/128 µg/mL
ApplicantLiofilchem s.r.l.
Product CodeJWY · Microbiology
Decision DateOct 31, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2

Indications for Use

MTS (MIC Test Strip) Ampicillin-sulbactam 0.016/0.008-256/128 µg/mL is a quantitative method intended for the in vitro determination of antimicrobial susceptibility of bacteria. MTS consists of specialized paper impregnated with a pre-defined concentration gradient of an antimicrobial agent, which is used to determine the minimum inhibitory concentration (MIC) in µg/mL of antimicrobial agents against bacteria as tested on agar media using overnight incubation and manual reading procedures. MTS Ampicillin-sulbactam at concentration of 0.016/0.008-256/128 µg/mL should be interpreted at 16-20 hours of incubation.

Device Story

MTS Ampicillin-sulbactam is a quantitative antimicrobial susceptibility test (AST) strip; consists of specialized paper impregnated with a predefined exponential concentration gradient of ampicillin-sulbactam (15 two-fold dilutions). Used in clinical microbiology laboratories; manual procedure. Inoculated agar plates incubated 16-20 hours; antibiotic diffuses into agar forming a symmetrical inhibition ellipse. Clinician/technician manually reads MIC value at the intersection of the inhibition ellipse and the strip. Provides quantitative MIC (µg/mL) to guide antibiotic therapy decisions. Benefits patient by determining susceptibility of specific bacterial isolates to ampicillin-sulbactam, supporting targeted treatment.

Clinical Evidence

Performance evaluated using 444 clinical and challenge non-fastidious Gram-negative isolates (390 Enterobacteriaceae, 54 A. baumannii). Compared to CLSI broth microdilution reference method. Combined results showed 99.1% Essential Agreement and 91.7% Category Agreement. No major or very major discrepancies observed. Trending analysis identified higher MIC readings for specific species (E. cloacae, E. coli, K. aerogenes, M. morganii, P. mirabilis, P. stuartii, A. baumannii), which is addressed in device labeling.

Technological Characteristics

Specialized paper strip impregnated with predefined concentration gradient of ampicillin-sulbactam (0.016/0.008-256/128 µg/mL). Principle: antimicrobial diffusion into Mueller Hinton agar. Manual reading of inhibition ellipse. Dimensions: 15 two-fold dilutions. Sterilization: not specified. Software: none (manual system).

Indications for Use

Indicated for in vitro antimicrobial susceptibility testing of specific Gram-negative bacteria (Enterobacter asburiae, E. cloacae, E. coli, Klebsiella spp., P. mirabilis, Acinetobacter baumannii/calcoaceticus complex, M. morganii, P. vulgaris, P. rettgeri, P. stuartii) to determine the minimum inhibitory concentration (MIC) of Ampicillin-sulbactam.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY ## I Background Information: A 510(k) Number K192345 B Applicant Liofilchem s. r. l. C Proprietary and Established Names MTS Ampicillin-Sulbactam 0.016/0.008 - 256/128 µg/mL D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | JWY | Class II | 21 CFR 866.1640 | MI-Microbiology | ## II Submission/Device Overview: A Purpose for Submission: To obtain a substantial equivalence determination for Ampicillin-sulbactam at concentrations of 0.016/0.008-256/128 µg/mL for susceptibility testing of non-fastidious Gram-negative organisms. B Measurand: MTS Ampicillin-sulbactam in the dilution range of 0.016/0.008-256/128 µg/mL. The concentrations represent the ratio of Ampicillin to Sulbactam in µg/mL. C Type of Test: Quantitative Antimicrobial Susceptibility Test growth-based detection K192345 - Page 1 of 11 {1} K192345 - Page 2 of 11 # III Intended Use/Indications for Use: ## A Intended Use(s): See Indications for Use below. ## B Indication(s) for Use: MTS (MIC Test Strip) Ampicillin-sulbactam 0.016/0.008-256/128 µg/mL is a quantitative method intended for the in vitro determination of antimicrobial susceptibility of bacteria. MTS consists of specialized paper impregnated with a pre-defined concentration gradient of an antimicrobial agent, which is used to determine the minimum inhibitory concentration (MIC) in µg/mL of antimicrobial agents against bacteria as tested on agar media using overnight incubation and manual reading procedures. MTS Ampicillin-sulbactam at concentration of 0.016/0.008-256/128 µg/mL should be interpreted at 16-20 hours of incubation. Ampicillin-sulbactam has been shown to be active both clinically and in vitro against these bacterial species according to the FDA drug approved label: - Gram-negative bacteria - Enterobacter asburiae - Enterobacter cloacae - Escherichia coli - Klebsiella aerogenes - Klebsiella oxytoca - Klebsiella pneumoniae - Proteus mirabilis - Acinetobacter baumannii/Acinetobacter calcoaceticus complex Ampicillin-sulbactam has been shown to be active in vitro only against the non-fastidious bacteria listed below according to the FDA drug approved label: - Gram-negative bacteria - Morganella morganii - Proteus vulgaris - Providencia rettgeri - Providencia stuartii {2} C Special Conditions for Use Statement(s): Rx - For Prescription Use Only D Special Instrument Requirements: Manual reading only IV Device/System Characteristics: A Device Description: The MIC Test Strip (MTS) consists of specialized paper impregnated with a predefined concentration gradient of ampicillin-sulbactam, across 15 two-fold dilutions similar to dilutions used by conventional MIC methods. One side of the strip is labelled with the ampicillin-sulbactam code (AMS) and the MIC reading scale is $\mu \mathrm{g} / \mathrm{mL}$. MIC values are determined by identifying the drug concentration at which growth of the ellipse ends. B Principle of Operation: MTS are made of specialized paper impregnated with a predefined concentration gradient of antibiotic, across 15 two-fold dilutions similar to dilutions used by conventional MIC methods. When the MIC Test Strip is applied onto an inoculated agar surface, the preformed exponential gradient of antimicrobial agent diffuses into the agar for over an hour. After 16-20 hours incubation, a symmetrical inhibition ellipse centered along the strip is formed. The MIC is read directly from the scale in terms of $\mu \mathrm{g} / \mathrm{mL}$ at the point where the edge of the inhibition ellipse intersects the strip MIC Test Strip. Growth along the entire gradient (i.e., no inhibition ellipse) indicates that the MIC value is greater than or equal to $(\geq)$ the highest value on the scale. An inhibition ellipse that intersects below the lower end of the scale is read as less than $(<)$ the lowest value. An MIC of 0.125 $\mu \mathrm{g} / \mathrm{mL}$ is considered to be the same as $0.12 \mu \mathrm{g} / \mathrm{mL}$ for reporting purposes. An MTS MIC value which falls between standard two-fold dilutions must be rounded up to the next standard upper two-fold value before categorization. V Substantial Equivalence Information: A Predicate Device Name(s): Liofilchem MIC Test Strip (MTS)-Vancomycin 0.016 -256 ug/mL B Predicate 510(k) Number(s): K153687 K192345 - Page 3 of 11 {3} # C Comparison with Predicate(s): Table 1. Comparison with the Predicate | Device & Predicate Device(s): | Device K192345 | Predicate K153687 | | --- | --- | --- | | Device Trade Name | MTS Ampicillin-sulbactam 0.016/0.008-256/128 μg/mL | Liofilchem MIC Test Strip (MTS)-Vancomycin 0.016 - 256 ug/mL | | General Device Characteristic Similarities | | | | Media | Mueller Hinton agar | Same | | MTS Strip Material | High quality paper impregnated with a predefined concentration of gradient Antimicrobial Agent | Same | | Inoculation | Isolated colonies from culture in a suspension equivalent to 0.5 McFarland. Inoculum is applied to agar with swab manually or with rotation plate for even distribution of inoculum. | Same | | Reading | Manual; the point where the edge of inhibition ellipse intersects the MIC Test Strip | Same | | Results | MIC (μg/mL) | Same | | General Device Characteristic Differences | | | | Intended Use/Indications for Use | Quantitative susceptibility to antimicrobial agents against Gram-negative organisms | Quantitative susceptibility to antimicrobial agents against Gram-positive organisms | | Antibiotic | Ampicillin-sulbactam (AMS) combination | Vancomycin (VA) | | Drug Concentration Range | 0.016/0.008-256/128 μg/mL | 0.016-256 μg/mL | | Incubation | 35°C ±2°C for 16-20 hours | 35°C ±2°C for 24 hours | # VI Standards/Guidance Documents Referenced: Guidance for Industry and FDA: "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems, August 28, 2009 - CLSI M07-A11 "Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically"; Approved Standard, Eleventh Edition, January 2018 - CLSI M100-29 $^{\text{th}}$ ed "Performance Standards for Antimicrobial Susceptibility Testing"; Approved Standard, Twenty-Ninth Edition, January 2019 K192345 - Page 4 of 11 {4} VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility: Reproducibility testing was conducted at three sites using ten-gram negative organisms. Each isolate was tested in triplicates over three days. The reproducibility panel included one *A. baumannii*, one *E. cloacae*, three *E. coli*, one *K. aerogenes*, one *K. oxytoca*, two *K. pneumoniae*, and one *P. mirabilis* isolates. The mode of MIC value was pre-determined, and the reproducibility was calculated based on the number of MIC values that fell within ±1 doubling dilution of the mode. All MIC results were on scale. The testing resulted in overall reproducibility of greater than 95%. The results were acceptable. 2. Linearity: Not applicable 3. Analytical Specificity/Interference: Not applicable 4. Assay Reportable Range: Not applicable 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): Inoculum Density Check: The inoculum was prepared to achieve turbidity equivalent to a 0.5 McFarland standard. Colony counts were performed periodically at each site for all QC replicates. Inoculum density checks were performed, and the colony counts obtained for each QC strain were within the recommended range of approximately 1× 10⁸ CFU/mL. Colony counts was also determined from one replicate of each reproducibility isolate on each of the three days of testing and from a minimum of 10% of the clinical strains tested. Purity Checks: Purity checks were performed on all isolates following MTS inoculation. All isolates were pure in both the broth microdilution reference panels and the MTS agar plates. Growth Failure Rate: None of the clinical and challenge strains in the study failed to grow with both the broth microdilution panels and the MTS agar plates. K192345 - Page 5 of 11 {5} # Quality control (QC) Testing: The QC strains recommended for routine testing by the CLSI for testing the Ampicillin-sulbactam, namely, $E.$ coli ATCC 35218, and $K.$ pneumoniae ATCC 700603 were tested at three sites for a minimum of 20 times at each testing site. Ampicillin-sulbactam MIC results for these QC strains are summarized in Table 2. Table 2: QC Results for Ampicillin-Sulbactam with the CLSI Recommended QC Strain | Organism | Concentration (μg/mL) | Reference BMD (All Sites) | MTS (All Sites) | | --- | --- | --- | --- | | E. coli ATCC 35218 | 4 | 0 | 0 | | Expected Results: 8/4-32/16 μg/mL | 8 | 0 | 2 | | | 16 | 49 | 24 | | | 32 | 12 | 35 | | | 64 | 0 | 0 | | K. pneumoniae ATCC 700603 | 4 | 0 | 0 | | Expected Results: 8/4-32/16 μg/mL | 8 | 0 | 0 | | | 16 | 13 | 13 | | | 32 | 48 | 45 | | | 64 | 1a | 4a | aOne (1) BMD $(\mathrm{MIC} = 64\mu \mathrm{g / mL})$ and 4 MTS (MICs all at $48~\mu \mathrm{g / mL}$ ) outlier QC results for K. pneumoniae ATCC 700603 from site 1. There were 1 to 2 additional replicates from this strain from site 1 that were tested on the same day and results were in range for both BMD and MTS. Also, E. coli ATCC 35218 results on those days were within QC range. Therefore, MTS and BMD MIC results for clinical isolates tested on those days were acceptable. QC Strain Integrity Check: Additional QC testing was performed with ampicillin alone at the three sites to confirm the integrity of the quality control strains $K$ . pneumoniae ATCC 700603 as recommended by the CLSI (Table 2 A). Table 2A: Quality Control Summary for Ampicillin Alone | Organism | MIC (μg/mL) | Reference BMD (All Sites) | | --- | --- | --- | | K. pneumoniae ATCC 700603 | 64 | 0 | | Expected Results: >128 μg/mL | >128 | 46 | Auxiliary QC strains: In addition to testing the QC strains recommended for routine testing by CLSI (Table 2), $E.$ coli ATCC 25922 was also tested at each site by both the MTS and the reference BMD method. Even though this auxiliary QC strain generally is not relevant for verification of the activity of the ampicillin-sulbactam combination, it does provide verification of the activity of ampicillin component of the drug (Table 2B). K192345 - Page 6 of 11 {6} Table 2B. Quality Control Results for Ampicillin-Sulbactam with Auxiliary QC Strain | Organism^{b} | Concentration (μg/mL) | Reference BMD (All Sites) | MTS (All Sites) | | --- | --- | --- | --- | | E. coli ATCC 25922 | 1 | 0 | 0 | | | 2 | 1 | 1 | | Expected Results: | 4 | 58 | 46 | | | 8 | 3 | 15 | | 2/1-8/4 μg/mL | 16 | 0 | 0 | All QC results were acceptable. 6. Detection Limit: Not Applicable 7. Assay Cut-Off: Not Applicable B Comparison Studies: 1. Method Comparison with Predicate Device: Results obtained with Liofilchem MIC Test Strip (MTS) with ampicillin-sulbactam were compared to results obtained from frozen reference MIC panels. Reference panels were prepared with Muller Hinton broth as outlined in CLSI recommendations in M7-A11. Isolated colonies from an overnight blood agar plate were suspended in saline to achieve a 0.5 McFarland standard turbidity (approximately 10⁸ CFU/mL). Testing conditions consisted of incubation of the inoculated Mueller Hinton agar plates in an inverted position at 35°C ±2 for 16-20 hours. At the end of incubation, the MIC value where the edge of the inhibition ellipse intersects the strip was compared to MIC results obtained with the reference method. Clinical: Clinical testing was performed at three US sites with both MTS Ampicillin-sulbactam and the reference method using a total of 321 Enterobacteriaceae (12 Morganella morganii, 30 K. aerogenes, 31 E. cloacae, 8 E. asburiae, 90 E. coli, 30 K. oxytoca, 60 K. pneumoniae, 15 P. mirabilis, 15 P. vulgaris, 15 P. rettgeri and 15 P. stuartii) and 47 Acinetobacter baumannii clinical isolates. A total of 368 non-fastidious Gram-negative isolates tested of which 206 (56%) were tested within 6 months of isolation (contemporary isolates). Challenge: Challenge testing was performed at one internal site using MTS Ampicillin-sulbactam and the reference method. A total of 69 Enterobacteriaceae (4 Morganella morganii, 11 K. K192345 - Page 7 of 11 {7} aerogenes, 12 E. cloacae, 11 E. coli, 7 K. oxytoca, 11 K. pneumoniae, 7 P. mirabilis, 2 P. vulgaris, 2 P. rettgeri, 2 P. stuartii) and 7 Acinetobacter baumannii challenge isolates were tested. The total of 444 non-fastidious Gram-negative organisms (390 Enterobacteriaceae and 54 Acinetobacter baumannii) clinical and challenge isolates is summarized in Table 3. Table 3. Performance of MTS Ampicillin-Sulbactam | | Tot | EA N | EA % | Eval Tot | Eval EA N | Eval EA % | CA Tot | CA % | No. R | No. S | mi n | ma j | vm j | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Enterobacteriaceae a | | | | | | | | | | | | | | | Clinical | 321 | 318 | 99.1 | 218 | 215 | 98.6 | 289 | 90 | 194 | 87 | 32 | 0 | 0 | | Challenge | 69 | 68 | 98.6 | 13 | 12 | 92.3 | 69 | 100 | 63 | 3 | 0 | 0 | 0 | | Total | 390 | 386 | 99 | 231 | 227 | 98.3 | 358 | 91.8 | 257 | 90 | 32 | 0 | 0 | | Acinetobacter baumannii b | | | | | | | | | | | | | | | Clinical | 47 | 47 | 100 | 41 | 41 | 100 | 42 | 89.4 | 20 | 20 | 5 | 0 | 0 | | Challenge | 7 | 7 | 100 | 2 | 2 | 100 | 7 | 100 | 5 | 0 | 0 | 0 | 0 | | Total | 54 | 54 | 100 | 43 | 43 | 100 | 49 | 90.7 | 25 | 20 | 5 | 0 | 0 | | All Gram-negative | | | | | | | | | | | | | | | Clinical | 368 | 365 | 99.2 | 259 | 256 | 98.8 | 331 | 89.9 | 214 | 107 | 37 | 0 | 0 | | Challenge | 76 | 75 | 98.7 | 15 | 14 | 93.3 | 76 | 100 | 68 | 3 | 0 | 0 | 0 | | Combined | 444 | 440 | 99.1 | 274 | 270 | 98.5 | 407 | 91.7 | 282 | 110 | 37 | 0 | 0 | a Enterobacteriaceae is comprised of E. asburiae, E. cloacae, E. coli, K. aerogenes, K. oxytoca, K. pneumoniae, P. mirabilis, M.morganii, P. vulgaris, P. stuartii b Combined clinical and challenge A. baumannii (54) were tested as representative of the A. baumannii/A. calcoaceticus complex EA - Essential Agreement CA - Category Agreement EVAL - Evaluable isolates R or NS - Resistant or non-susceptible isolates min - minor discrepancies maj - major discrepancies vmj - very major discrepancies Essential Agreement (EA) is when the Liofilchem MIC Test Strip (MST) results agree exactly or within one doubling dilution of the reference broth microdilution results. Category Agreement (CA) is when the Liofilchem MIC Test Strip (MST) result interpretation agrees exactly with the reference broth microdilution result interpretation. The overall performance of Enterobacteriaceae isolates (Table 3) is acceptable with 99% EA and 91.8% CA. There were no major or very major discrepancies. The overall performance of Acinetobacter baumannii isolates (Table 3) is acceptable with 100% EA and 90.7% CA. There were no major or very major discrepancies. Results of comparative testing with clinical and challenge isolates combined demonstrated an EA of 99.1% and CA of 91.7 for all non-fastidious Gram-negative organism which was acceptable (Table 3). To address testing and reporting of MIC results for non-indicated species the sponsor included the following statement in the Precautions section of the device labeling: Per the FDA-Recognized Susceptibility Test Interpretive Criteria website, the safety and efficacy of antimicrobial drugs for which antimicrobial susceptibility is tested by this AST K192345 - Page 8 of 11 {8} device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labelling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved. ## Resistance Mechanism Characterization Challenge isolates of Enterobacteriaceae and A. baumannii harboring various molecular mechanisms of resistance were evaluated with MTS Ampicillin-sulbactam. Isolates harboring the following resistance mechanisms were evaluated: NDM, OXA, KPC, TEM, SHV, CTX-M, CMY, ACT, IMP, AmpC, ESBL, VIM, DHA, tet(A) ## MIC Trending Analysis Using the combined clinical and challenge data, an analysis of trending was conducted for both Enterobacteriaceae and A. baumannii. Results are stratified by species to determine if species-related trends were observed (Table 4). This trending calculation considers MIC values that are determined to be one or more doubling dilutions lower or higher compared to the reference method irrespective of whether the device MIC values are on-scale or not. Results that are not clearly at least one dilution lower, at least one dilution higher or in exact agreement with the CLSI reference method are not considered in the trending analysis. Species for which the difference between the percentage of isolates with higher vs. lower readings was $\geq 30\%$ and for which the confidence interval was determined to be statistically significant were considered to show evidence of trending. Trending that provides higher or lower MIC values compared to the reference is addressed in labeling. No significant trending was observed for Enterobacteriaceae overall, however, a trend toward higher MIC reading was observed for E. cloacae, E. coli, K. aerogenes, M. morganii, P. mirabilis, P. stuartii and A. baumannii with MTS Ampicillin-sulbactam when compared to the reference method (Table 4). Table 4. Trending for MTS Ampicillin-Sulbactam with Gram-Negative Organisms | Organism | Total Evaluable for Trending | ≥ 1 Dilution lower No. (%) | Exact No. (%) | ≥ 1 Dilution Higher No. (%) | Percent Difference (CI)* | Trending Noted | | --- | --- | --- | --- | --- | --- | --- | | E. asburiae | 7 | 0 (0.00) | 4 (57.14) | 3 (42.86) | 42.9 | No | | E. cloacae | 22 | 1 (4.55) | 10 (45.45) | 11 (50.00) | 45.5 | Yes | | E. coli | 73 | 7 (9.59) | 36 (49.32) | 30 (41.10) | 31.5 | Yes | | K. aerogenes | 26 | 1 (3.85) | 11 (42.31) | 14 (53.85) | 50.0 | Yes | | K. oxytoca | 24 | 3 (12.50) | 19 (79.17) | 2 (8.33) | -4.2 | No | | K. pneumoniae | 35 | 3 (8.57) | 22 (62.86) | 10 (28.57) | 20 | No | | M. morganii | 13 | 1 (7.69) | 4 (30.77) | 8 (61.54) | 53.8 | Yes | | P. mirabilis | 20 | 1 (5.00) | 10 (50.00) | 9 (45.00) | 40.0 | Yes | | P. vulgaris | 17 | 6 (35.29) | 7 (41.18) | 4 (23.53) | -11.8 | No | | P. rettgeri | 15 | 5 (33.33) | 6 (40.00) | 4 (26.67) | -6.7 | No | | P. stuartii | 17 | 0 (0.00) | 10 (58.82) | 7 (41.18) | 41.2 | Yes | | Enterobacteriaceae | 269 | 28 (10.41) | 138 (51.30) | 103 (38.29) | 27.9 | No | | A.baumannii | 48 | 3 (6.25) | 24 (50.00) | 21 (43.75) | 37.5 | Yes | K192345 - Page 9 of 11 {9} * A percent difference ≥30% is considered significant trending; a positive percentage difference value in trending analysis indicates higher MIC observed with the device and could cause potential major discrepancies. A negative percentage difference value in trending analysis indicates lower MIC observed with the device and could cause potential very major discrepancies. To address the observed high trend, the following footnote has been added to the labeling in the performance characteristics section: MTS Ampicillin-sulbactam MIC values tended to be in exact agreement or at least one doubling dilution higher compared to the CLSI broth microdilution reference method when testing E. cloacae, E. coli, K. aeruginosa, M. morganii, P. mirabilis, P. stuartii and A. baumannii. 2. Matrix Comparison: Not Applicable C Clinical Studies: 1. Clinical Sensitivity: Not applicable 2. Clinical Specificity: Not Applicable 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not Applicable D Clinical Cut-Off: Not Applicable E Expected Values/Reference Range: The FDA recognized susceptibility interpretive criteria for Ampicillin-sulbactam are listed in Table 5. Table 5. FDA Identified Interpretive Criteria for Ampicillin-Sulbactam | Organism | Interpretive Criteria for Ampicillin-Sulbactam MIC (μg/mL) a | | | | --- | --- | --- | --- | | | Susceptible (S) | Intermediate (I) | Resistant (R) | | Enterobacteriaceae b | ≤8/4 | 16/8 | ≥32/16 | | Acinetobacter spp | ≤8/4 | 16/8 | ≥32/16 | a FDA STIC Webpage https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm b E. asburiae, E. cloacae, E. coli, K. aerogenes, K. oxytoca, K. pneumoniae, P. mirabilis, M. morganii, P. vulgaris, P. stuartii K192345 - Page 10 of 11 {10} K192345 - Page 11 of 11 VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a breakpoint change protocol that was reviewed and accepted by the FDA. This protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage (https://www.fda.gov/drugs/development-resources/antibacterial-susceptibility-test-interpretive-criteria). The protocol outlined the specific procedures and acceptance criteria that Liofilchem intends to use to evaluate the Liofilchem MIC test strip (MTS) when revised breakpoints for Ampicillin-sulbactam are published on the FDA STIC webpage. The breakpoint change protocol included with the submission indicated that if specific criteria are met, Liofilchem will update the Ampicillin-sulbactam device label to include (1) the new breakpoint, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...