Sensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System with Omadacycline in the dilution range of 0.008 - 32 ug/mL

K183324 · Thermo Fisher Scientific · JWY · Feb 15, 2019 · Microbiology

Device Facts

Record IDK183324
Device NameSensititre 20-24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System with Omadacycline in the dilution range of 0.008 - 32 ug/mL
ApplicantThermo Fisher Scientific
Product CodeJWY · Microbiology
Decision DateFeb 15, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2

Indications for Use

The Sensititre Haemophilus influenzae/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenzae, Streptococcus pneumoniae, and Streptococcus species.

Device Story

Sensititre MIC Susceptibility panels are multi-well microtiter plates containing dried, stabilized antimicrobials; miniaturized broth dilution method. Device takes bacterial isolates as input; plates are inoculated using Sensititre AIM autoinoculator; incubated at 34-36°C for 20-24 hours. Growth detection via VIZION digital viewing device (manual) or OptiRead automated reader (fluorescence-based). OptiRead monitors surface enzyme activity using fluorogenic substrates; fluorescence intensity correlates to bacterial growth. Output is Minimum Inhibitory Concentration (MIC) value. Used in clinical microbiology laboratories by technicians. Results assist clinicians in determining antimicrobial susceptibility for patient treatment decisions.

Clinical Evidence

No clinical data provided in the document; substantial equivalence is based on performance characteristics of the antimicrobial susceptibility test system.

Technological Characteristics

Microbroth dilution susceptibility test system. Consists of multi-well plates containing dehydrated antimicrobial agent (Omadacycline) in specific concentration ranges. Designed for 20-24 hour incubation. Operates via standard in vitro diagnostic laboratory procedures for bacterial growth inhibition measurement.

Indications for Use

Indicated for clinical susceptibility testing of fastidious isolates, specifically Streptococcus pneumoniae, Haemophilus influenzae, Streptococcus anginosus group (S. anginosus, S. intermedius, S. constellatus), and Streptococcus pyogenes, using Omadacycline.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K183324 B. Purpose for Submission: To obtain a substantial equivalence determination for the addition of Omadacycline at concentrations of 0.008 – 32 µg/mL to the Sensititre 20-24-hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System for testing H. influenzae and Streptococcus spp. C. Measurand: Omadacycline in the dilution range of 0.008 – 32 µg/mL. D. Type of Test: Quantitative Antimicrobial Susceptibility Test (AST), growth-based detection E. Applicant: ThermoFisher Scientific F. Proprietary and Established Names: Sensititre 20 - 24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System with Omadacycline in the dilution range of 0.008 - 32 µg/mL G. Regulatory Information: 1. Regulation section: 866.1640 Antimicrobial Susceptibility Test Powder 2. Classification: Class II 3. Product code: {1} JWY – Manual Antimicrobial Susceptibility Test System LRG – Instrument for Auto Reader and Instrumentation of Overnight Susceptibility Systems LTT – Panels, Test, Susceptibility, Antimicrobial 4. Panel: 83, Microbiology H. Intended Use: 1. Intended use(s): The Sensititre Haemophilus influenzae/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenzae, Streptococcus pneumoniae, and Streptococcus species. 2. Indication(s) for use: The Sensititre 20 - 24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System is an in vitro diagnostic product for clinical susceptibility testing of fastidious isolates. This 510(k) is for Omadacycline in the dilution range of 0.008 - 32 µg/mL for testing Streptococcus spp. and Haemophilus influenzae on the Sensititre 20 - 24 hour MIC panel. Omadacycline has been shown to be active both clinically and in vitro against the following organisms according to the FDA drug label: Streptococcus pneumoniae Haemophilus influenzae Streptococcus anginosus group (includes S. anginosus, S. intermedius, and S. constellatus) Streptococcus pyogenes 3. Special conditions for use statement(s): For prescription use only The following limitations are included in the labeling: The testing of omadacycline with Streptococcus spp. was performed using the AutoReader (OptiRead) and VIZION reading methods and Haemophilus influenzae was only read by VIZION manual method. The use of an alternative 2 {2} reading method when testing omadacycline has not been evaluated. The ability of the Sensititre system to detect resistance to Omadacycline in the following species is unknown because resistant strains were not available at the time of comparative testing: H. influenzae, S. pneumoniae, S. pyogenes, and S. anginosus. Isolates yielding omadacycline MIC results suggestive of a resistant interpretive category should be submitted to a reference laboratory for further testing. The performance of Omadacycline with Haemophilus influenzae and Streptococcus spp. was performed using the AIM autoinoculator. The use of an alternative inoculation system when testing Omadacycline has not been evaluated. Omadacycline MIC values tended to be in exact agreement or at least one dilution lower when testing S. pyogenes and S. pneumoniae with both OptiRead and VIZION reading methods compared to the CLSI reference broth microdilution. MIC values MIC values tended to be in exact agreement or one dilution higher when testing S. anginosus with both OptiRead and VIZION. Omadacycline MIC values tended to be in exact agreement or at least one dilution lower when testing H. influenzae with the VIZION only. 4. Special instrument requirements: - Sensititre AIM for device inoculation - Sensititre VIZION or OptiRead for plate reading I. Device Description: Sensititre MIC Susceptibility MIC panels are multi-well microtiter plates, dosed with dried, stabilized antimicrobials. It is a miniaturized version of the classic broth dilution method and can provide both qualitative and quantitative susceptibility results. After inoculation, plates are sealed with an adhesive seal, incubated at $34 - 36^{\circ}\mathrm{C}$ for $20 - 24$ hours and examined for bacterial growth. Antimicrobial susceptibility test results can be determined by reading growth using the digital viewing device (VIZION) or automatically on an autoreader (OptiRead) using fluorescence. J. Substantial Equivalence Information: 1. Predicate device name(s): Sensititre Haemophilus/Streptococcus pneumoniae (HP) MIC Plates 2. Predicate $510(\mathbf{k})$ number(s): {3} K040846 # 3. Comparison with predicate: Table 1. Comparison with the Predicate Device | Similarities | | | | --- | --- | --- | | Item | Device K183324 Sensititre Haemophilus/Streptococcus pneumoniae (HP) MIC Plates with Omadacycline | Predicate K040846 Sensititre Haemophilus/Streptococcus pneumoniae (HP) MIC Plates with Ertapenem | | Intended Use | The Sensititre Haemophilus influenzae/Streptococcus pneumoniae plates are in vitro diagnostic products for clinical susceptibility testing of Haemophilus influenzae, Streptococcus pneumoniae, and Streptococcus species. | Same | | Test Panel | 96 well plate is dosed with selected antimicrobial agents and substrate for the fluorescent reads, then dried. The bacterial suspension in the appropriate broth is used to rehydrate the plate. | Same | | Test Organism | Haemophilus influenzae, Streptococcus pneumoniae, and Streptococcus spp. | Same | | Reading Methods for Streptococcus spp | Results can be read using the following methods:1) Automatically with the OptiRead (fluorescent substrate technology)2) On the VIZION (digital viewing device) | Same | | Reading Methods for H. influenzae | Results can be read using the VIZION (digital viewing device) only | Same | | Incubation | 20-24 hours, 35 ± 1°C | Same | | Inoculation media | CAMHBT + LHB (Streptococcus spp.), HTM broth (H. influenzae) | Same | {4} | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Antimicrobial Agent | Omadacycline | Cefepime | | Concentration Range | 0.008 – 32 μg/mL | 0.008 – 16 μg/mL | ## K. Standard/Guidance Document Referenced (if applicable): Guidance for Industry and FDA - Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems – August 28, 2009. CLSI M100-S027: Performance Standards for Antimicrobial Susceptibility Testing; Twenty-Seventh Informational Supplement CLSI M7-A10: Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard – Tenth Edition ## L. Test Principle: The Sensititre 20 - 24 hour Haemophilus influenzae/Streptococcus pneumoniae system includes multi-well plastic microtiter plates that contain doubled dilution of antibacterial agents. Each plate includes antimicrobial agents at appropriate dilutions. Results can be read by the digital device, VIZION, or by use of an automated reader, OptiRead (Streptococcus spp. only). The VIZION allows the panel image to be displayed on a touch screen directly from a video camera and allows the user to visually determine MIC results. The Sensititre OptiRead utilizes fluorescence technology to read the microbroth dilution plates after 20 to 24 hours incubation. The technology involves the detection of bacterial growth by monitoring the activity of specific surface enzymes produced by the test organism. Growth is determined by generating a fluorescent product from a fluorogenic substrate. The substrate is prepared by conjugating a fluorescent compound to the specific enzyme substrates with a bond which prevents fluorescence. The enzymatic action of the bacterial surface enzymes on the substrate cleaves the bond releasing fluorescence. The amount of fluorescence detected is directly related to bacterial growth. The MIC is determined by observing the lowest dilution of antimicrobial agent that inhibits growth of the organism. The substrate can be added to the inoculum broth which is dispensed into the test plate at the same time as the test organism, or, the plates can be prepared with the substrate already added to each micro-well. {5} M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: A reproducibility study was performed at four sites using a panel comprised of 8 isolates of *Streptococcus* spp.: (*S. pyogenes* (one isolate), *S. pneumoniae* (three isolates), *S. anginosus* (two isolates), and *S. agalactiae* (two isolates)) and a panel comprised of 10 strains of *H. influenzae*. Of the *Streptococcus* spp., one non-indicated species was tested. All isolates were tested in triplicate over three days with each read method (i.e., VIZION and OptiRead for *Streptococcus* spp., and VIZION only for *H. influenzae*). The Sensititre Aim inoculator was used for plate inoculation. The mode MIC value was determined and the reproducibility was calculated based on MIC values falling within ±1 dilution of the mode MIC value. Reproducibility was greater than 95% for both read methods with *Streptococcus* spp. and VIZION with the *H. influenzae* panel and was considered to be acceptable. b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Quality control strains recommended by the CLSI were tested with Omadacycline at four sites. The QC organisms tested were *S. pneumoniae* ATCC 49619 and *H. influenzae* ATCC 49247. They were tested a minimum of 20 times per site and read using the VIZION and OptiRead for *S. pneumoniae* and VIZION only for *H. influenzae*. The results demonstrate that the Sensititre 20 - 24 hour *Haemophilus influenzae*/*Streptococcus pneumoniae* MIC or Breakpoint Susceptibility System with Omadacycline produced quality control results in the recommended range >95% of the time (Table 2). 6 {6} Table 2. Quality Control Results for Sensititre 20 - 24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System with Omadacycline with the VIZION and OptiRead Methods. | QC Organism | Omadacycline Range (μg/mL) | Concentration (μg/mL) | Reference | Sensititre | | | --- | --- | --- | --- | --- | --- | | | | | | Read method | | | | | | | VIZION | OptiRead | | S. pneumoniae ATCC 49619 | 0.015 – 0.12 | 0.008 | 0 | 0 | 0 | | | | 0.015 | 6 | 16 | 17 | | | | 0.03 | 76 | 72 | 71 | | | | 0.06 | 6 | 1 | 1 | | | | 0.12 | 1 | 0 | 0 | | | | 0.25 | 0 | 0 | 0 | | H. influenzae ATCC 49247 | 0.5 - 2 | 0.25 | 0 | 0 | NA | | | | 0.5 | 8 | 18 | NA | | | | 1 | 29 | 68 | NA | | | | 2 | 44 | 5 | NA | | | | 4 | 0 | 0 | NA | Inoculum Density. Inoculum density checks were performed a sufficient number of times; all organism suspensions were in the acceptable range. Purity checks. Purity checks were performed on all isolates following plate inoculation. Only results from pure cultures were evaluated. Growth Failure: All H. influenzae and Streptococcus isolates tested showed growth in the Sensititre panels. d. Detection limit: Not applicable e. Analytical specificity: Not applicable {7} f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Results obtained with Sensititre 20 - 24 hour Haemophilus influenzae/Streptococcus pneumoniae MIC or Breakpoint Susceptibility System with Omadacycline were compared to results obtained with the CLSI broth microdilution reference panel. Drug dilutions were prepared using fresh Mueller Hinton broth with lysed horse blood (LHB) for Streptococcus spp. and Haemophilus Test Medium (HTM) broth for H. influenzae isolates as indicated in CLSI M07, 11th ed. Clinical testing was performed at four sites, three of which were in the U.S. A total of 352 clinical Streptococcus isolates were tested and read by both VIZION and OptiRead which were comprised of the following species: S. pneumoniae (199 isolates), S. pyogenes (100 isolates), and S. anginosus (53 isolates). In addition, a total of 393 H. influenzae isolates were tested of which results were read using the digital viewing device (VIZION) only. All of the clinical isolates tested were fresh isolates. During the course of the clinical trial, all Sensititre dried MIC panels were inoculated using the Sensititre Autoinoculator (AIM) and the same panel was read on both the VIZION and the OptiRead for Streptococcus isolates in a blinded manner. The sponsor added the following limitations to the device labeling to reflect these inoculation and read methods: The performance of Omadacycline with Haemophilus influenzae and Streptococcus spp. was performed using the AIM autoinoculator. The use of an alternative inoculation system when testing Omadacycline has not been evaluated. The testing of omadacycline with Streptococcus spp. was performed using the AutoReader (OptiRead) and VIZION reading methods and Haemophilus influenzae was only read by VIZION manual method. The use of an alternative reading method when testing omadacycline has not been evaluated. A total of 144 challenge isolates were tested at a single site. Species tested included S. pneumoniae (50 isolates), S. pyogenes (20 isolates), S. anginosus (24 isolates), and H. influenzae (50 isolates). For Streptococcus spp., results were evaluated for essential agreement (EA) and category agreement (CA). The susceptibility interpretive criteria (breakpoints) for Community Acquired Bacterial Pneumonia (CABP) were used to evaluate CA for S. pneumoniae isolates and H. influenzae. The breakpoints for Acute Bacterial Skin and Skin Structure Infections (ABSSSI) were used to evaluate CA for S. pyogenes and S. anginosus group. Breakpoints for both are noted on the FDA-Recognized Susceptibility Test Interpretive Criteria Website (STIC) 8 {8} (https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/u cm575163.htm). Performance results for *Streptococcus spp.* and *H. influenzae* using the VIZION are shown in Table 3. Performance results for *Streptococcus spp.* Using the OptiRead are shown in Table 4. *H. influenzae* was only read by VIZION method. The results from clinical and challenge testing determined with the VIZION for *H. influenzae* demonstrated a combined EA of 98% and CA of 98.9%. A total of 441 isolates were determined to have evaluable results; the EA of evaluable results was 98% (Table 3). Clinical and challenge isolate results for the *S. pneumoniae* determined with the VIZION demonstrated a combined EA of 97.2% and CA of 99.6%. A total of 248 isolates were determined to have evaluable results; the EA of evaluable results was 97.2% (Table 3). For MIC results determined with OptiRead, the combined results from clinical and challenge testing demonstrated an EA of 97.2% and CA of 99.6%; EA of evaluable results was 97.2% (Table 4). For *S. pyogenes*, the combined MIC results from clinical and challenge testing determined with the VIZION demonstrated EA of 99.2% and CA 97.5%; EA of evaluable results was 99.2% (Table 3). For MIC results determined with OptiRead, the combined results from clinical and challenge testing demonstrated an EA of 99.2% and CA of 100%; EA of evaluable results was 99.2% (Table 4). *S. anginosus* group isolates were not speciated during the course of the study and, therefore, performance for each species was not evaluated. There were six results for both VIZION and OptiRead that were off-scale by the reference method, however, these results were deemed evaluable for essential agreement determination given that all results were at least two dilutions from the lowest dilution of the reference. The combined MIC results from clinical and challenge testing determined with the VIZION demonstrated EA of 90.9% and CA 96.1%; EA of evaluable results was 90.1% (Table 3). For MIC results determined with OptiRead, the combined results from clinical and challenge testing demonstrated an EA of 90.0% and CA of 96.1%; EA of evaluable results was 88.7% (Table 4). Only one isolate was considered resistant by the reference method. The OptiRead gave a false susceptible result with this isolate, resulting in one VMJ error. The OptiRead gave no major errors. Given that results for *S. anginosus* trended towards major errors and not very major errors, this one very major error was considered a random error and deemed acceptable. For all organisms tested, an insufficient number of resistant isolates were encountered during the clinical evaluation. The sponsor included the following limitation in the device labeling: The ability of the Sensititre system to detect resistance to *Omadacycline* in the following species is unknown because resistant strains were not available at the {9} time of comparative testing: H. influenzae, S. pneumoniae, S. pyogenes, and S. anginosus. Isolates yielding omadacycline MIC results suggestive of a resistant interpretive category should be submitted to a reference laboratory for further testing. Table 3. Performance of Streptococcus spp. and H. influenzae Clinical and Challenge Isolates, CABP and ABSSSI Breakpoints Read Using VIZION | | Tot | EA N | EA % | Eval Tot | Eval EA N | Eval EA % | CA Tot | CA % | No. R | No. S | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | | | | | | | | | | | Streptococcus pneumoniae CABP breakpoints (≤0.12, 0.25, ≥0.5) | | | | | | | | | | | | | | | Clinical | 199 | 197 | 99.0 | 198 | 196 | 99.0 | 199 | 100 | 0 | 198 | 0 | 0 | 0 | | Challenge | 50 | 45 | 90.0 | 50 | 45 | 90.0 | 49 | 98.0 | 0 | 49 | 1 | 0 | 0 | | Total | 249 | 242 | 97.2 | 248 | 241 | 97.2 | 248 | 99.6 | 0 | 247 | 1 | 0 | 0 | | | | | | | | | | | | | | | | | Streptococcus pyogenes ABSSSI breakpoints (≤0.12, 0.25, ≥0.5) | | | | | | | | | | | | | | | Clinical | 100 | 99 | 99.0 | 100 | 99 | 99 | 98 | 98.0 | 1 | 97 | 2 | 0 | 0 | | Challenge | 20 | 20 | 100 | 20 | 20 | 100 | 19 | 95.0 | 1 | 19 | 1 | 0 | 0 | | Total | 120 | 119 | 99.2 | 120 | 119 | 99.2 | 117 | 97.5 | 2 | 116 | 3 | 0 | 0 | | | | | | | | | | | | | | | | | Streptococcus anginosus group ABSSSI breakpoints (≤0.12, 0.25, ≥0.5) | | | | | | | | | | | | | | | Clinical | 53 | 46 | 86.8 | 47 | 40 | 85.1 | 52 | 98.1 | 0 | 52 | 1 | 0 | 0 | | Challenge | 24 | 24 | 100 | 24 | 24 | 100 | 22 | 91.7 | 1 | 21 | 2 | 0 | 0 | | Total | 77 | 70 | 90.9 | 71 | 64 | 90.1 | 74 | 96.4 | 1 | 73 | 3 | 0 | 0 | | | | | | | | | | | | | | | | | H. influenzae CABP breakpoints (≤2, 4, ≥8) | | | | | | | | | | | | | | | Clinical | 393 | 389 | 99.0 | 391 | 387 | 99.0 | 388 | 98.7 | 2 | 385 | 5 | 0 | 0 | | Challenge | 50 | 45 | 95.0 | 50 | 45 | 95.0 | 50 | 100 | 0 | 49 | 0 | 0 | 0 | | Total | 443 | 434 | 98.0 | 441 | 432 | 98.0 | 438 | 98.9 | 2 | 434 | 5 | 0 | 0 | EA - Essential Agreement (+/- 1 dilution) CA - Category Agreement EVAL - Evaluable isolates R - Resistant isolates min - minor discrepancies maj - major discrepancies vmj - very major discrepancies Essential agreement (EA) occurs when the result of the reference method and that of the Sensititre panel are within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the reference method and the Sensititre panel. Category agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the Sensititre panel. Table 4. Performance of Streptococcus spp. Clinical and Challenge Isolates, CABP and ABSSSI Breakpoints Read Using OptiRead | | Tot | EA N | EA % | Eval Tot | Eval EA N | Eval EA % | CA Tot | CA % | No. R | No. S | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | | | | | | | | | | | | Streptococcus pneumoniae CABP breakpoints (≤0.12, 0.25, ≥0.5) | | | | | | | | | | | | | | | Clinical | 199 | 197 | 99.0 | 99 | 99 | 100 | 199 | 100 | 0 | 198 | 0 | 0 | 0 | | Challenge | 50 | 45 | 90.0 | 50 | 45 | 90.0 | 49 | 98.0 | 0 | 49 | 1 | 0 | 0 | | Total | 249 | 242 | 97.2 | 248 | 241 | 97.2 | 248 | 99.6 | 0 | 247 | 1 | 0 | 0 | | | | | | | | | | | | | | | | | Streptococcus pyogenes ABSSSI breakpoints (≤0.12, 0.25, ≥0.5) | | | | | | | | | | | | | | {10} | | Tot | EA N | EA % | Eval Tot | Eval EA N | Eval EA % | CA Tot | CA % | No. R | No. S | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Clinical | 100 | 99 | 99.0 | 100 | 99 | 99 | 98 | 98.0 | 1 | 97 | 2 | 0 | 0 | | Challenge | 20 | 20 | 100 | 20 | 20 | 100 | 19 | 95.0 | 1 | 19 | 1 | 0 | 0 | | Total | 120 | 119 | 99.2 | 120 | 119 | 99.2 | 117 | 97.5 | 2 | 116 | 3 | 0 | 0 | | | | | | | | | | | | | | | | | Streptococcus anginosus group ABSSSI breakpoints (≤0.12, 0.25, ≥0.5) | | | | | | | | | | | | | | | Clinical | 53 | 46 | 86.8 | 47 | 40 | 85.1 | 52 | 98.1 | 0 | 52 | 1 | 0 | 0 | | Challenge | 24 | 23 | 95.8 | 24 | 23 | 95.8 | 22 | 91.7 | 1 | 21 | 2 | 0 | 1 | | Total | 77 | 69 | 90.0 | 71 | 63 | 88.7 | 74 | 96.1 | 1 | 73 | 3 | 0 | 0 | To address the testing of non-indicated species the following footnote was included in the precautions section of the device labeling: The safety and efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved. ## MIC Trending An analysis of trending was conducted using the combined clinical and challenge data for each organism group. This trending calculation takes into account MIC values that are determined to be one or more doubling dilutions lower or higher compared to the reference method irrespective of whether the device MIC values are on-scale or not. Trending results are shown in Table 5 for $H.$ influenzae and for Streptococcus spp.. The acceptable percent difference between higher and lower dilution readings is $<30\%$. {11} Table 5. Trending for All Fastidious Clinical and Challenge Isolates | Read Method | Organism | Total evaluable for trending | ≥1 dilution lower No. (%) | Exact No (%) | ≥1 dilution higher No (%) | | --- | --- | --- | --- | --- | --- | | VIZION | S. pyogenes\(^a\) | 120 | 60 (50) | 55 (45.8) | 5 (4.17) | | | S. pneumoniae\(^b\) | 249 | 157 (63.1) | 84 (33.7) | 8 (3.21) | | | S. anginosus\(^c\) | 74 | 5 (6.8) | 29 (39.2) | 40 (54.1) | | | S. agalactiae\(^d\) | 126 | 85 (67.5) | 36 (28.6) | 5 (4.0) | | | H. influenzae\(^e\) | 342 | 260 (76.0) | 132 (38.6) | 50 (14.62) | | OptiRead | S. pyogenes\(^f\) | 120 | 56 (46.7) | 57 (47.5) | 7 (5.8) | | | S. pneumoniae\(^g\) | 248 | 166 (66.9) | 77 (31.1) | 5 (2.0) | | | S. anginosus\(^h\) | 74 | 14 (18.9) | 23 (31.1) | 37 (50) | | | S. agalactiae\(^i\) | 126 | 106 (84.1) | 15 (11.9) | 5 (4.0) | a Difference between the higher and lower dilutions for S. pyogenes is: -45.8% b Difference between the higher and lower dilutions for S. pneumoniae is: -59.8% c Difference between the higher and lower dilutions for S. anginosus group is: $47.3\%$ d Difference between the higher and lower dilutions for S. agalactiae is: -63.5% e Difference between the higher and lower dilutions for H. influenzae is: -61.4% f Difference between the higher and lower dilutions for S. pyogenes is: -40.8% g Difference between the higher and lower dilutions for S. pneumoniae is: -64.9% h Difference between the higher and lower dilutions for S. anginosus is: 31.08% 1 Difference between the higher and lower dilutions for S. agalactiae is: -80.16% A trend toward lower MIC readings was observed for S. pyogenes and S. pneumoniae with both VIZION and OptiRead. In addition, a trend toward higher MIC readings was observed for S. anginosus group isolates for both VIZION and OptiRead. Furthermore, a trend toward lower MIC readings was observed for $H.$ influenzae with the VIZION. The sponsor included the following footnotes to the performance table to address the trending observed for Omadacycline. Omadacycline MIC values tended to be in exact agreement or at least one dilution lower when testing S. pyogenes and S. pneumoniae with both OptiRead and VIZION reading methods compared to the CLSI reference broth microdilution. MIC values MIC values tended to be in exact agreement or one dilution higher when testing S. anginosus with both OptiRead and VIZION. Omadacycline MIC values tended to be in exact agreement or at least one dilution lower when testing H. influenzae with the VIZION only. b. Matrix comparison: Not applicable {12} 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Table 6. Interpretive Criteria for Omadacycline | Organism | Infection Type | FDA-Recognized Interpretive Criteria for Omadacycline, MIC (μg/mL) | | | | --- | --- | --- | --- | --- | | | | S | I | R | | H. influenzae | CABP | ≤2 | 4 | ≥8 | | S. pyogenes | ABSSSI | ≤0.12 | 0.25 | ≥0.5 | | S. anginosus group¹ | ABSSSI | ≤0.12 | 0.25 | ≥0.5 | | S. pneumoniae | CABP | ≤0.12 | 0.25 | ≥0.5 | ¹S. anginosus group includes S. anginosus, S. intermedius, and S. constellatus. N. Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...