Liofilchem MIC Test Strip (MTS)- Linezolid 0.016 - 256ug/mL

K180330 · Liofilchem s.r.l. · JWY · Mar 15, 2018 · Microbiology

Device Facts

Record IDK180330
Device NameLiofilchem MIC Test Strip (MTS)- Linezolid 0.016 - 256ug/mL
ApplicantLiofilchem s.r.l.
Product CodeJWY · Microbiology
Decision DateMar 15, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2

Indications for Use

The Liofilchem® MIC Test Strip (MTS) is a quantitative method intended for the in vitro determination of antimicrobial susceptibility of bacteria. MTS consists of specialized paper impregnated with a pre-defined concentration gradient of an antimicrobial agent, which is used to determine inhibitory concentration (MIC) in ug/mL of antimicrobial agents against bacteria as tested on agar media using overnight incubation and manual reading procedures. The Linezolid MTS at concentrations of 0.016- 256 ug/mL should be interpreted at 16-20 hours of incubation. Linezolid has been shown to be active both clinically and in vitro agamst the non-fastidious bacteria listed below according to the FDA label: Staphylococcus aureus (including methicillin-resistant isolates) Enterococcus faecium (vancomycin resistant isolates only)

Device Story

Liofilchem MIC Test Strip (MTS) is a quantitative diagnostic tool for determining antimicrobial susceptibility. It consists of a specialized paper strip impregnated with a predefined concentration gradient of Linezolid (0.016–256 µg/mL). In a clinical laboratory setting, the strip is applied to inoculated agar media. Following 16–20 hours of overnight incubation, the user manually observes the zone of inhibition to determine the MIC value. This result assists clinicians in selecting appropriate antibiotic therapy for patients with specific bacterial infections. The device provides a standardized method for assessing bacterial resistance, supporting clinical decision-making for targeted treatment.

Clinical Evidence

Performance evaluated using 299 clinical and 77 challenge isolates (S. aureus, E. faecium). Compared to CLSI broth microdilution reference method. S. aureus (MSSA/MRSA) showed 99.0% EA and 99.5% CA. E. faecium showed 98.7% EA and 94.0% CA. Trending analysis indicated a slight one-dilution higher MIC bias compared to reference, deemed acceptable. Reproducibility study (3 sites, 10 isolates, 270 data points) demonstrated ≥95% reproducibility. QC testing with ATCC strains met expected ranges >95% of the time.

Technological Characteristics

Specialized paper strip impregnated with linezolid gradient; 15 two-fold dilutions. Manual reading of inhibition ellipse on Mueller Hinton agar. Incubation at 35°C for 16–20 hours. No electronic components, software, or connectivity. Single-use.

Indications for Use

Indicated for in vitro antimicrobial susceptibility testing of non-fastidious bacteria, specifically Staphylococcus aureus (including methicillin-resistant isolates) and Enterococcus faecium (vancomycin-resistant isolates only), to determine the Minimum Inhibitory Concentration (MIC) of Linezolid.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K180330 B. Purpose for Submission: To obtain a substantial equivalence determination of the Liofilchem MIC Test Strip (MTS) containing linezolid at concentrations of 0.016 – 256 µg/mL for susceptibility testing of select gram positive organisms. C. Measurand: Linezolid 0.016 – 256 µg/mL D. Type of Test: Quantitative Antimicrobial Susceptibility growth-based detection E. Applicant: Liofilchem s.r.l. F. Proprietary and Established Names: Liofilchem MIC Test Strip (MTS), Linezolid 0.016 – 256 µg/mL G. Regulatory Information: 1. Regulation section: 21 CFR 866.1640 Antimicrobial Susceptibility Test Powder 2. Classification: Class II 3. Product code: JWY – Manual Antimicrobial Test Systems {1} 4. Panel: Microbiology (83) H. Intended Use: 1. Intended use(s): The Liofilchem MIC Test Strip (MTS) is a quantitative method intended for the in vitro determination of antimicrobial susceptibility of bacteria. MTS consists of specialized paper impregnated with a pre-defined concentration gradient of an antimicrobial agent, which is used to determine the minimum inhibitory concentration (MIC) in $\mu \mathrm{g} / \mathrm{mL}$ of antimicrobial agents against bacteria as tested on agar media using overnight incubation and manual reading procedures. The Linezolid MTS concentrations of $0.016 - 256\ \mu \mathrm{g/mL}$ should be interpreted after 16–20 hours of incubation. Linezolid has been shown to be active both clinically and in vitro against the non-fastidious bacteria listed below, according to the FDA label. Staphylococcus aureus (including methicillin-resistant isolates) Enterococcus faecium (vancomycin resistant isolates only) 2. Indication(s) for use: Same as intended use 3. Special conditions for use statement(s): For Prescription Use Only 4. Special instrument requirements: Manual reading only I. Device Description: The linezolid MIC Test Strip (MTS) consists of specialized paper impregnated with a predefined concentration gradient of linezolid across 15 two-fold dilutions like those of a conventional MIC method. One side of the strip is labelled with the linezolid code (LNZ); the MIC reading scale is in $\mu \mathrm{g} / \mathrm{mL}$. When the MTS, Linezolid is applied onto an inoculated agar surface, the preformed exponential gradient of antimicrobial agent is immediately transferred to the agar matrix. After 16–20 hours of incubation, a symmetrical inhibition ellipse centered along the strip is formed. The MIC is read directly from the scale in terms of $\mu \mathrm{g} / \mathrm{mL}$ at the point where the edge of the inhibition ellipse intersects the MTS, Linezolid. The MTS, 2 {2} Linezolid is single use only. Since the MTS Linezolid strip generates MIC values that may fall between two-fold dilutions designations, the MIC value is recorded as the next two-fold dilution value. # J. Substantial Equivalence Information: 1. Predicate device name(s): Liofilchem MTS, Vancomycin 2. Predicate $510(\mathrm{k})$ number(s): K153687 3. Comparison with predicate: Table 1. Comparison with the Predicate | Similarities | | | | --- | --- | --- | | Item | Device Liofilchem MTS, Linezolid (K180330) | Predicate Liofilchem MTS, Vancomycin (K153687) | | Intended Use | Quantitative susceptibility to antimicrobial agents | Same | | Media | Mueller Hinton agar | Same | | Inoculum | Isolated colonies from culture in suspension equivalent to a 0.5 McFarland turbidity standard. Inoculum is applied to agar with swab manually or with rotation plate. | Same | | Incubation | 35 °C, 16 – 20 hours | Same | | Reading | Manual: the point where the edge of the inhibition ellipse intersects the MIC Test Strip | Same | | Result | MIC (μg/mL) | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Antimicrobial Agent | Linezolid | Vancomycin | {3} 4 K. Standard/Guidance Document Referenced (if applicable): FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA (Issued August 28, 2009) CLSI M07-10, “Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard – Tenth Edition (January, 2015) CLSI M100-S27, “Performance Standards for Antimicrobial Susceptibility Testing”, Twenty-Seventh Informational Supplement (January 2017) L. Test Principle: The Liofilchem MIC Test Strips (MTS) are made of specialized paper impregnated with a predefined concentration gradient of antibiotic. When the MIC Test Strip is applied onto an inoculated agar surface, the preformed exponential gradient of antimicrobial agent is immediately transferred to the agar matrix. After 16-20 hours incubation, a symmetrical inhibition ellipse centered along the strip is formed. The minimum inhibitory concentration (MIC) is read directly from the scale in terms of $\mu \mathrm{g} / \mathrm{mL}$ at the point where the edge of the inhibition ellipse intersects the strip MIC Test Strip. Growth along the entire gradient (i.e., no inhibition ellipse) indicates that the MIC value is greater than or equal to $(\geq)$ the highest value on the scale. An inhibition ellipse that intersects below the lower end of the scale is read as less than $(<)$ the lowest value. An MIC of 0.125 $\mu \mathrm{g} / \mathrm{mL}$ is considered to be the same as $0.12~\mu \mathrm{g} / \mathrm{mL}$ for reporting purposes. An MTS MIC value which falls between standard two-fold dilutions must be rounded up to the next standard upper two-fold value before categorization. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: A reproducibility study testing the Liofilchem MIC Test Strip (MTS) containing linezolid was conducted at three clinical sites using ten isolates of gram-positive cocci consistent with the Intended Use. Testing was performed on three separate days and in triplicate for a total of 270 data points among the sites. The isolates tested in the reproducibility study included: S. aureus (MRSA) 3 isolates, S. aureus (MSSA) 4 isolates and E. faecium (vancomycin resistant) 3 isolates. The mode MIC was determined and the reproducibility was calculated based on MIC values that fell within $\pm$ one doubling dilution from the mode MIC value. Both intra-site and inter-site reproducibility for MTS, Linezolid was calculated. There were no off-scale results. {4} The combined reproducibility results for all three sites were acceptable and demonstrated $\geq 95\%$ reproducibility. b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): CLSI recommended QC organisms were tested throughout the comparative study at the three study sites. The organisms tested were S. aureus ATCC 29213 and E. faecalis ATCC 29212. These recommended QC strains were tested a minimum of 20 times at each site by both the MTS, Linezolid and the CLSI broth microdilution reference method. All results were within the expected range $>95\%$ of the time. Results are summarized in Table 2; all QC results were acceptable. Table 2. Quality Control Results Summary for Linezolid MTS | Organism | Expected Range (μg/mL) | Concentration (μg/mL) | Reference | MTS Linezolid | | --- | --- | --- | --- | --- | | S. aureus ATCC 29213 | 1 - 4 | 0.25 | 1 | 0 | | | | 0.5 | 0 | 1 | | | | 1 | 3 | 11 | | | | 2 | 52 | 57 | | | | 4 | 5 | 5 | | | | 8 | 0 | 0 | | E. faecalis ATCC 29212 | 1 - 4 | 0.25 | 0 | 0 | | | | 0.5 | 0 | 0 | | | | 1 | 20 | 0 | | | | 2 | 41 | 41 | | | | 4 | 0 | 30 | | | | 8 | 0 | 0 | Inoculum Density Check: The inoculum was prepared to achieve a turbidity equivalent to a 0.5 McFarland standard. Colony counts were performed as part of QC and reproducibility testing as well as during the clinical studies to demonstrate that the inoculum concentrations were within the expected CFU/mL. The colony counts were acceptable. Purity Checks. Purity checks were performed on all isolates following MTS inoculation. Only results from pure cultures were evaluated. Growth Failure. All isolates tested showed growth on the Mueller Hinton agar with {5} MTS, Linezolid. One *E. faecium* isolate did not grow in the reference method test; this isolate was excluded from the study. d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: The MTS, Linezolid was evaluated at three sites located within the United States. Each clinical isolate was tested one time by MTS, Linezolid and the reference method using the same initial standardized suspension. A total of 299 clinical isolates comprised of the following species were tested: *S. aureus* (MSSA, 75 isolates), *S. aureus* (MRSA, 105 isolates) and *E. faecium* (vancomycin resistant, 119 isolates). The 299 clinical gram-positive isolates tested included 77 (25.8%) fresh isolates that were tested within seven days of isolation, 104 (34.8%) recent isolates that were tested within one year of isolation and 118 (39.5%) stock isolates that were tested within three years of isolation. A total of 300 isolates grew on the Mueller Hinton plate with the Linezolid MTS strip; one isolate of *E. faecalis* did not grow in the broth microdilution reference plate and was removed from the study. A total of 77 challenge isolates were tested at a single site and included *S. aureus* (MSSA, one isolate), *S. aureus* (MRSA, 29 isolates), *E. faecium* (vancomycin resistant, 32 isolates) and coagulase negative staphylococci (15 isolates). Coagulase negative staphylococci are not included in the list of intended organisms and were not included in the determination of device performance. Results obtained with the Liofilchem MIC Test Strip (MTS), Linezolid were compared to results obtained with the CLSI broth microdilution reference panel. The reference panel contained two-fold serial dilutions of linezolid with a range of 0.016 – 256 µg/mL. The testing conditions for the reference method were consistent with CLSI guidelines as listed in the CLSI document M07-A10. Clinical sites were instructed to record if trailing/hazy end-points were observed and to read linezolid MICs at the point of 90% inhibition when such trailing was noted, as recommended in the Liofilchem Technical Sheet “Cidal-Static” found on the Liofilchem website. 6 {6} For S. aureus (MSSA and MRSA) the results from clinical and challenge testing demonstrated a combined EA of 99.0% and CA of 99.5% with one major error and no very major errors. All results were evaluable; the EA of evaluable results was 99.0%. For E. faecalis the results from clinical and challenge testing demonstrated a combined EA of 98.7% and CA of 94.0%. There were no major or very major errors. (See Table 3). Review of the data showed that hazy endpoints/trailing growth occurred at a frequency of 98% and 97% for S. aureus and E. faecium, respectively. Table 3. Performance of Clinical and Challenge Isolates, S. aureus (MSSA and MRSA) and E. faecium | | Tot | EA N | EA % | Eval Tot | Eval EA N | Eval EA % | CA Tot | CA % | No. R | No. S | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | S. aureus, MSSA | | | | | | | | | | | | | | | Clinical | 75 | 75 | 100 | 75 | 75 | 100 | 75 | 100 | 0 | 75 | NA* | 0 | 0 | | Challenge | 1 | 1 | 100 | 1 | 1 | 100 | 1 | 100 | 1 | 0 | NA* | 0 | 0 | | Total | 76 | 76 | 100 | 76 | 76 | 100 | 76 | 100 | 1 | 75 | NA* | 0 | 0 | | | | | | | | | | | | | | | | | S. aureus MRSA | | | | | | | | | | | | | | | Clinical | 105 | 104 | 99.0 | 105 | 104 | 99.0 | 105 | 100 | 0 | 105 | NA* | 0 | 0 | | Challenge | 29 | 28 | 96.6 | 29 | 28 | 96.6 | 28 | 96.6 | 25 | 4 | NA* | 1 | 0 | | Total | 134 | 132 | 98.5 | 134 | 132 | 98.5 | 133 | 99.3 | 25 | 109 | NA* | 1 | 0 | | | | | | | | | | | | | | | | | S. aureus, MSSA and MRSA | | | | | | | | | | | | | | | Clinical | 180 | 179 | 99.4 | 180 | 179 | 99.4 | 180 | 100 | 0 | 180 | NA* | 0 | 0 | | Challenge | 30 | 29 | 96.7 | 30 | 29 | 96.7 | 29 | 96.7 | 26 | 4 | NA* | 1 | 0 | | Total | 210 | 208 | 99.0 | 210 | 208 | 99.0 | 209 | 99.5 | 26 | 184 | NA* | 1 | 0 | | | | | | | | | | | | | | | | | E. faecium | | | | | | | | | | | | | | | Clinical | 119 | 117 | 98.3 | 119 | 117 | 98.3 | 112 | 94.1 | 6 | 110 | 7 | 0 | 0 | | Challenge | 32 | 32 | 100 | 32 | 32 | 100 | 30 | 93.8 | 24 | 1 | 2 | 0 | 0 | | Total | 151 | 149 | 98.7 | 151 | 149 | 98.7 | 142 | 94.0 | 30 | 111 | 9 | 0 | 0 | *NA – not applicable due to a lack of an intermediate breakpoint for this drug EA - Essential Agreement (+/- 1 dilution) CA - Category Agreement EVAL - Evaluable isolates R - Resistant isolates min - minor discrepancies maj - major discrepancies vmj - very major discrepancies S - Susceptible isolates Essential agreement (EA) occurs when the result of the reference method and that of the MTS, Linezolid are within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both the reference method and the MTS, Linezolid. Category agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the MTS, Linezolid. MIC Trends. Using the combined clinical and challenge data for S. aureus and E. faecium, an analysis of trending was conducted. This trending calculation considers MIC values that are determined to be one or more doubling dilutions lower or higher compared to the reference method irrespective of whether the device MIC values are on-scale or not. The trending analysis indicated that the MTS, Linezolid gives a one-dilution higher MIC when compared to the reference method; however, the difference was considered to be acceptable and no further notation was required in labeling. The evaluable data for trend analysis is presented in Table 4. {7} Table 4. Trending Analysis of Evaluable Clinical and Challenge Results | Organism | Total No. \(Isolates^a\) | Difference in MIC as Compared to the Reference Method | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | ≥ 2 Dilutions lower | 1 Dilution Lower | Exact | 1 Dilution Higher | ≥ 2 Dilutions Higher | | S. aureus | 210 | 0 | 17 | 136 | 55 | 2 | | | | \(17(8.10\%)^b\)95% CI (5.12 – 12.58) | | 64.76% | \(57(27.14\%)^b\)95% CI (21.58 – 33.53) | | | E. faecium | 151 | 0 | 13 | 94 | 42 | 2 | | | | \(13(8.61\%)^c\)95% CI (5.10 – 14.17) | | 62.25% | \(29.14\%^c\)95% CI (22.48 – 36.83) | | a Total number of evaluable results for trending analysis b Difference: 19.05%, 95% CI (11.90 - 26.09) c Difference: $20.53\%$ , $95\%$ CI (11.85 - 28.99) # Resistance Mechanisms The FDA approved pharmaceutical antimicrobial agent package insert indicates that point mutations in the 23S rRNA are associated with linezolid resistance. A total of 18 challenge MRSA isolates with the noted mutation were evaluated in this study. All isolates provided the expected resistant result. b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable {8} 5. Expected values/Reference range: Table 5. Interpretive Criteria for Linezolid | Organism | Interpretive Criteria for Linezolid (FDA and CLSI) | | | | --- | --- | --- | --- | | | S | I | R | | Staphylococcus spp. | ≤ 4 | - | ≥ 8 | | Enterococcus spp. | ≤ 2 | 4 | ≥ 8 | N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...