← Product Code [JWY](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JWY) · K170670

# Etest Ceftolozane/ Tazobactam (0.016-256 ug/ml) (K170670)

_bioMerieux, Inc. · JWY · May 31, 2017 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JWY/K170670

## Device Facts

- **Applicant:** bioMerieux, Inc.
- **Product Code:** [JWY](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JWY.md)
- **Decision Date:** May 31, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1640
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

Etest® is a quantitative technique for determination of antimicrobial susceptibility of both nonfastidious Gram negative and Gram positive aerobic bacteria such as Enterobacteriaceae, Pseudomonas, Staphylococcus, and Enterococcus species and fastidious bacteria, such as anaerobes, N. gonorrhoeae, S. pneumoniae, Streptococcus and Haemophilus species. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC), in Ug/mL, of different antimicrobial agents against microorganisms as tested on agar media using overnight incubation. Etest® Ceftolozane / Tazobactam has been shown to be active against the Gram negative aerobic microorganisms listed below, according to the FDA label for this antimicrobial agent: Active both in vitro and in clinical infections: Enterobacter cloacae Escherichia coli Klebsiella oxytoca Klebsiella pneumoniae Proteus mirabilis Pseudomonas aeruginosa The following in vitro data are available, but clinical significance is unknown: Citrobacter koseri Morganella morganii Proteus vulgaris Providencia rettgeri Providencia stuartii Serratia liquefaciens Serratia marcescens

## Device Story

Etest is a thin, inert, non-porous plastic strip with a predefined antibiotic gradient on one side and an MIC reading scale on the other. Used in clinical microbiology laboratories; applied to inoculated agar surfaces. Antibiotic gradient transfers into agar matrix, forming a stable, continuous, exponential concentration gradient. Following overnight incubation, bacterial growth forms a symmetrical inhibition ellipse. The MIC value is determined by visual inspection where the ellipse intersects the strip. Provides quantitative susceptibility data to assist clinicians in selecting appropriate antimicrobial therapy for patients with bacterial infections.

## Clinical Evidence

Performance evaluated against CLSI M07-A10 broth microdilution reference method. Study used clinical isolates (fresh and stock) and challenge strains. Results demonstrated 97.2% overall Essential Agreement and 98.6% overall Category Agreement. Reproducibility and Quality Control testing met acceptance criteria.

## Technological Characteristics

Thin, inert, non-porous plastic strip; predefined antibiotic gradient; MIC reading scale. Manual antimicrobial susceptibility test system. No energy source required. Sterilization not specified. No software/algorithm.

## Regulatory Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Predicate Devices

- Etest® Ceftaroline ([K151873](/device/K151873.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Page 1 of 12

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
K170670

B. Purpose for Submission:
To obtain clearance for the addition of Ceftolozane/Tazobactam at concentrations 0.016-256μg/mL μg/mL to the Etest strip for susceptibility testing of Gram negative aerobic microorganisms.

C. Measurand:
Ceftolozane/Tazobactam 0.016 – 256μg/mL. Tazobactam concentration is fixed at 4 μg/mL in this combination.

D. Type of Test:
Quantitative AST growth based detection

E. Applicant:
bioMérieux, Inc.

F. Proprietary and Established Names:
Etest® Ceftolozane/Tazobactam (0.016 – 256 μg/mL)

G. Regulatory Information:
1. Regulation section:
866.1640 Antimicrobial Susceptibility Test Powder
2. Classification:
II
3. Product code:
JWY - Manual Antimicrobial Test Systems

{1}

Page 2 of 12

4. Panel:
83 - Microbiology

H. Intended Use:

1. Intended use(s):

Etest is a quantitative technique for determination of antimicrobial susceptibility of both non-fastidious Gram negative and Gram positive aerobic bacteria such as Enterobacteriaceae, Pseudomonas, Staphylococcus, and Enterococcus species and fastidious bacteria, such as anaerobes, N. gonorrhoeae, S. pneumoniae, Streptococcus and Haemophilus species. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC), in µg/mL, of different antimicrobial agents against microorganisms as tested on agar media using overnight incubation.

Etest Ceftolozane / Tazobactam has been shown to be active against the Gram negative aerobic microorganisms listed below, according to the FDA label for this antimicrobial agent:

Active both in vitro and in clinical infections:
- Enterobacter cloacae,
- Escherichia coli,
- Klebsiella oxytoca,
- Klebsiella pneumoniae
- Proteus mirabilis
- Pseudomonas aeruginosa

The following in vitro data are available, but clinical significance is unknown
- Citrobacter koseri,
- Morganella morganii
- Proteus vulgaris
- Providencia rettgeri
- Providencia stuartii
- Serratia liquefaciens
- Serratia marcescens

2. Indication(s) for use:
Same as the Intended Use

3. Special conditions for use statement(s):
For prescription use only

{2}

Limitation(s):

"The ability of the Etest to detect resistance with the following combination (s) is unknown because resistant strains were not available at the time of comparative testing. If such a strain is observed, it should be submitted to a reference laboratory for further testing
- Ceftolozane/Tazobactam: Citrobacter koseri, Proteus vulgaris, and Serratia liquefaciens"

4. Special instrument requirements:

Manual readings only

I. Device Description:

Etest consists of a thin, inert and non-porous plastic strip, 5mm wide and 60 mm long. One side of the strip carries a two-letter code designating the identity of the antibiotic and is calibrated with MIC values in terms of µg/mL. On the other side, a predefined exponential gradient of the dried and stabilized antibiotic covers a continuous concentration range across 15 two-fold dilutions of a conventional MIC method.

J. Substantial Equivalence Information:

1. Predicate device name(s):

Etest Ceftaroline

2. Predicate 510(k) number(s):

K151873

Page 3 of 12

{3}

# 3. Comparison with predicate:

Table 1: Comparison with the Predicate Device

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device K170670 Etest Ceftolozane /Tazobactam (0.016-256 μg/mL) | Predicate K151873 Etest Ceftaroline (0.002-32μg/mL)  |
|  Intended Use | Etest is a quantitative technique for determination of antimicrobial susceptibility of both non-fastidious Gram negative and Gram positive aerobic bacteria such as Enterobacteriaceae, Pseudomonas, Staphylococcus, and Enterococcus species and fastidious bacteria, such as anaerobes, N. gonorrhoeae, S. pneumoniae, Streptococcus and Haemophilus species. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC), in μg/mL, of different antimicrobial agents against microorganisms as tested on agar media using overnight incubation. | Same  |
|  Incubation | 35°±2°C for 16-20 hours | 35°; 5% CO2; 20-24 hours  |
|  Inoculation | Isolated colonies from culture | Same  |
|  Result | MIC | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Claimed Organisms | Enterobacteriaceae and Pseudomonas aeruginosa | Community-Acquired Bacterial Pneumonia (CABP): Streptococcus pneumoniae Haemophilus influenzae Skin infections: Streptococcus agalactiae  |
|  Antimicrobial | Ceftolozane/Tazobactam | Ceftaroline  |
|  Antimicrobial Concentration Range | Ceftolozane MIC range: 0.016 – 256 μg/mL. (Tazobactam: 4 μg/mL) | 0.002-32 μg/mL  |
|  Reading time | Read at 16-20 hours incubation | Etest compared to Broth Microdilution reference method read at 18 and at 23hours  |

{4}

Page 5 of 12

## K. Standard/Guidance Document Referenced (if applicable):

Class II Special Controls Guidance Document: “Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA”;

CLSI M07-A10 “Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically”; Approved Standard, January 2015

CLSI M100-S25 “Performance Standards for Antimicrobial Susceptibility Testing”; Twenty-five Informational Supplement, Volume 35 No. 3, Jan 2015

## L. Test Principle:

The Etest gradient technology is based on a combination of the concepts of dilution and diffusion test methods for susceptibility testing. Etest directly quantifies antimicrobial susceptibility in terms of discrete MIC values. When the Etest strip is applied to an inoculated agar plate, the antibiotic is immediately released from the plastic surface into the agar. A predefined, continuous gradient of antibiotic concentrations is created and maintained directly underneath the strip. After incubation whereby bacterial growth becomes visible, a symmetrical inhibition ellipse centered along the strip will be seen. The MIC value in $\mu \mathrm{g} / \mathrm{mL}$ is read where the ellipse edge intersects the strip. Since Etest generates MIC values which fall between two-fold dilutions for interpretation, the MIC value read must be recorded to the next two-fold dilution.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Precision/Reproducibility:

A reproducibility study was conducted at three external sites using 25 organisms of gram negative bacilli that were consistent with the intended use. The isolates tested in the reproducibility study included *E. cloacae* (one isolate), *E. coli* (seven isolates), *K. oxytoca* (three isolates), *K. pneumoniae* (five isolates), *P. aeruginosa* (nine isolates).

Results were within +/- one doubling dilution agreement as compared to the mode MIC value of Ceftolozane/Tazobactam for all organisms at all sites. All results were on scale. The reproducibility was 97.33%.

The reproducibility results were acceptable.

{5}

b. Linearity/assay reportable range:

Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

## Inoculum Density Check:

The inoculum was prepared to match a 0.5 McFarland. Colony counts were performed periodically at each site to demonstrate that the inoculum procedure results were in the expected CFU/mL.

Inoculum density checks were performed on *E. coli* ATCC 25922, quality control strain, reproducibility and 10% of the fresh clinical isolates. The mean inoculum density of *E. coli* ATCC 25922 was $2.6 \times 10^{5}$ CFU/mL.

## Quality Control (QC) Testing:

Organisms recommended by both FDA and CLSI, namely *E. coli* ATCC 25922, *E. coli* ATCC 35218, *Klebsiella pneumoniae* ATCC 700603, and *Pseudomonas aeruginosa* ATCC 27853 were tested against Ceftolozane/Tazobactam at four sites. These recommended QC organisms were tested a minimum of 20 times/site by both the Etest Ceftolozane/Tazobactam and reference method at each site. The Etest results demonstrate that the system can produce QC results in the recommended range &gt;95% of the time. Table 2 provides a summary of these QC results.

A second set of quality control results were obtained. This testing included one additional quality control organism, *Klebsiella pneumoniae* ATCC 700603. This organism was also tested against Ceftazidime alone to check for integrity of the plasmid-mediated resistance. Testing was conducted at four external sites plus one internal site. Results are presented in Table 3a/3b below.

*K. pneumoniae* ATCC 700603 is recommended to test against ceftazidime alone to ensure the presence of the plasmid encoding β-lactamase when testing Ceftolozane/Tazobactam. The footnote below is included in the labeling.

"*K. pneumoniae* ATCC 700603 should be used for routine QC of Ceftolozane/Tazobactam. This QC strain should also be tested against ceftazidime alone to ensure that the plasmid encoding the beta-lactamase has not been lost in this strain."

Page 6 of 12

{6}

Table 2: Quality Control Data for Ceftolozane/Tazobactam (First Set)

|  Organism | Conc (μg/mL) | Reference Frequency | Etest Frequency  |
| --- | --- | --- | --- |
|  E. coli ATCC 25922
Expected Range (0.12- 0.5 μg/mL) | ≤0.125 |  |   |
|   |  0.125 | 34 | 87  |
|   |  0.25 | 59 | 7  |
|   |  0.5 | 1 |   |
|   |  >0.5 |  |   |
|  E. coli
ATCC 35218 Expected Range (0.06-0.25 μg/mL) | ≤0.064 |  |   |
|   |  0.064 | 2 | 21  |
|   |  0.125 | 86 | 72  |
|   |  0.25 | 3 | 1  |
|   |  >0.25 |  |   |
|  K. pneumoniae
ATCC 700603
Expected Range (0.5- 2 μg/mL) | ≤0.5 |  |   |
|   |  0.5 | 1 |   |
|   |  1 | 72 | 62  |
|   |  2 | 20 | 32  |
|   |  >2 |  |   |
|  P. aeruginosa
ATCC 27853
Expected Range 0.25- 1 μg/mL | ≤0.25 |  |   |
|   |  0.25 |  | 3  |
|   |  0.5 | 90 | 85  |
|   |  1 | 4 | 1  |
|   |  >1 |  |   |

Table 3a: Quality Control Data for Ceftolozane/Tazobactam (Second Set)

|  Organism | Conc (μg/mL) | Reference Frequency | Etest Frequency  |
| --- | --- | --- | --- |
|  E. coli ATCC 25922
Expected Range (0.12- 0.5 μg/mL) | ≤0.125 |  |   |
|   |  0.125 | 11 | 93  |
|   |  0.25 | 90 | 7  |
|   |  0.5 |  |   |
|   |  >0.5 |  |   |
|  E. coli
ATCC 35218 Expected Range (0.06-0.25 μg/mL) | ≤0.064 |  |   |
|   |  0.064 |  | 14  |
|   |  0.125 | 90 | 87  |
|   |  0.25 | 11 |   |
|   |  >0.25 |  |   |
|  K. pneumoniae
ATCC 700603
Expected Range (0.5- 2 μg/mL) | ≤0.5 |  |   |
|   |  0.5 |  |   |
|   |  1 | 71 | 48  |
|   |  2 | 29 | 51  |
|   |  >2 |  | 1  |
|  P. aeruginosa
ATCC 27853
Expected Range 0.25- 1 μg/mL | ≤0.25 |  |   |
|   |  0.25 |  | 1  |
|   |  0.5 | 99 | 94  |
|   |  1 | 2 | 6  |
|   |  >1 |  |   |

Page 7 of 12

{7}

Table 3b: Quality Control Data for Ceftazidime

|  Ceftazidime | Conc (μg/mL) | Reference Frequency | Etest Frequency  |
| --- | --- | --- | --- |
|  K. pneumoniae*
ATCC 700603
Expected Range 16-64 μg/mL
(ancillary testing) | ≤16 | - | NA  |
|   |  16 | 4 | NA  |
|   |  32 | 83 | NA  |
|   |  64 | 14 | NA  |
|   |  >64 | - | NA  |

*Testing in accordance with:
- CLSI M100-S25 Performance Standards for Antimicrobial Susceptibility Testing: Twenty-Fifth Informational Supplement (version recognized at the beginning of clinical trial) (January 2015)-Table 5A
- CLSI M100-S26 Performance Standards for Antimicrobial Susceptibility Testing: Twenty-Sixth Informational Supplement (version recognized at the completion of clinical trial) (January 2016)-Table 5A

The quality control results are acceptable.

d. Detection limit:

Not Applicable

e. Analytical specificity:

Not Applicable

f. Assay cut-off:

Not Applicable

2. Comparison studies:

a. Method comparison with predicate device:

Results obtained with the Etest Ceftolozane/Tazobactam were compared to results obtained with the CLSI broth microdilution reference panel. The CLSI recommended broth microdilution reference panel was prepared and interpreted according to CLSI recommendations. Reference testing was in accordance with the CLSI Standard: CLSI Document M7-A10, Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard-Tenth Edition, Vol. 35, No.2; January 2015. The testing conditions for the reference method consisted of the following:

- Medium – Mueller Hinton broth with appropriate dilutions of antimicrobial solution added
- Inoculum – Direct colony suspension
- Incubation – 35°C ambient air incubator; 16-20 hours for all organisms

Page 8 of 12

{8}

Clinical testing was performed at five sites (four external and one internal). There were 966 (793Enterobacteriaceae and 173Pseudomonas aeruginosa) clinical isolates, of which 435 were stock (45%), and 531 were fresh (55%). All isolates grew in the Etest Ceftolozane/Tazobactam test.

A challenge set consisting of 90 isolates (51 Enterobacteriaceae, and 39 P. aeruginosa) was also evaluated at one external site. Each challenge isolate was tested by the Etest Ceftolozane/Tazobactam and the CLSI reference method. All of the challenge organisms grew in the Etest Ceftolozane/Tazobactam test.

The Etest Ceftolozane/Tazobactam results were compared to the CLSI broth microdilution reference method tested at each study site. Results from the 1056 clinical and challenge isolates are summarized in Table 4. The overall performance using the Etest Ceftolozane/Tazobactam was acceptable, with an EA of 97.2%, CA of 98.6%, a minor error (discrepancy) rate of 1.1% (12/1056), and a major error rate of 0.3% (3/928). There were no very major errors.

Table 4: Performance of Clinical and Challenge Isolates
|   | Tot | No. EA | EA % | Eval Tot | No. Eval EA | Eval EA % | No. CA | CA % | No. R | min | maj | vmj  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Enterobacteriaceae  |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 793 | 768 | 96.85 | 780 | 755 | 96.79 | 781 | 98.49 | 56 | 10 | 2 | 0  |
|  Challenge | 51 | 49 | 96.08 | 38 | 36 | 94.74 | 49 | 96.08 | 23 | 1 | 1 | 0  |
|  Combined | 844 | 817 | 96.8 | 818 | 791 | 96.7 | 830 | 98.34 | 79 | 11 | 3 | 0  |
|  Pseudomonas aeruginosa  |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 173 | 171 | 98.8 | 163 | 162 | 99.4 | 172 | 99.4 | 17 | 1 | 0 | 0  |
|  Challenge | 39 | 38 | 97.4 | 24 | 24 | 100 | 39 | 100 | 18 | 0 | 0 | 0  |
|  Combined | 212 | 209 | 98.6 | 187 | 186 | 99.5 | 211 | 99.5 | 35 | 1 | 0 | 0  |
|  All Organisms  |   |   |   |   |   |   |   |   |   |   |   |   |
|  Clinical | 966 | 939 | 97.2 | 943 | 917 | 97.2 | 953 | 98.6 | 73 | 11 | 2 | 0  |
|  Challenge | 90 | 87 | 96.7 | 62 | 60 | 96.8 | 88 | 97.8 | 41 | 1 | 1 | 0  |
|  Combined | 1056 | 1026 | 97.2 | 1005 | 977 | 97.2 | 1041 | 98.6 | 114 | 12 | 3 | 0  |

EA - Essential Agreement (+/- 2 dilutions)
CA - Category Agreement
EVAL - Evaluable isolates
R or NS - Resistant or non-susceptible isolates

min - minor discrepancies
maj - major discrepancies
vmj - very major discrepancies

Essential agreement (EA) is when the Etest results agree with the reference test panel results exactly or within one doubling dilution of the reference method. Category agreement (CA) is when the Etest result interpretation agrees exactly with the reference panel result interpretation.

Enzyme Groups Molecular Characterization:
Molecular characterization was conducted for Pseudomonas aeruginosa and for Enterobacteriaceae to determine the beta lactamase enzyme groups and

Page 9 of 12

{9}

other resistance markers noted in the FDA drug label. The performance of the Etest with Ceftolozane/Tazobactam included 38 Pseudomonas aeruginosa and 45 Enterobacteriaceae isolates that were characterized as detailed below.

## Pseudomonas aeruginosa:

Molecular characterization was conducted to determine beta lactamase enzyme resistance marker status for Pseudomonas aeruginosa isolates for the following: chromosomal AmpC, and loss of OprD. Enzyme characterization was also performed for Pseudomonas aeruginosa isolates that produce serine carbapenemases [K. pneumoniae carbapenemase (KPC)] and metallo-β-lactamases.

Pseudomonas isolates with MexXY and MexAB up-regulation enzymes were not evaluated as this information was not available at the time of testing. This was addressed by adding the following footnote in the labeling:

"Enzyme group characterization was not available for the following organisms at the time of comparative testing, and therefore the performance of the Etest with Ceftolozane/Tazobactam is unknown for: Pseudomonas aeruginosa (up-regulation of MexXY and MexAB) and Enterobacteriaceae (OXA)".

## Enterobacteriaceae (ESBLs):

Molecular characterization was conducted to determine beta lactamase enzyme resistance marker status for claimed Enterobacteriaceae isolates for the following ESBLs: TEM, SHV and CTX-M. However, information on Enterobacteriaceae isolates with OXA enzymes was not available at the time of testing and was not provided. This was addressed by adding the footnote listed above in the labeling.

## Enterobacteriaceae (KPC and β-lactamases):

Enzyme characterization was also conducted for to determine carbapenemase enzyme resistance marker status for claimed Enterobacteriaceae isolates that produce serine carbapenemases [K. pneumoniae carbapenemase (KPC)]. However, MIC testing did not include Enterobacteriaceae isolates with known metallo-beta lactamase status with the exception of one Proteus mirabilis isolate tested that contain VEB metallo beta-lactamase. Tazobactam is not an inhibitor of carbapenemases such as KPC and metallo-β-lactamases such as IMP and VIM. In conformance with the Ceftolozane/Tazobactam Pharmaceutical Drug label, the following footnote was added in the labeling

"Ceftolozane/Tazobactam is not active against Enterobacteriaceae that produce KPC or metallo-beta lactamases".

Page 10 of 12

{10}

Page 11 of 12

# Resistant Organisms:

A total of 114 out of 1056 organisms tested (10.8%) were found to be resistant to Ceftolozane/Tazobactam by the reference method. However, the following indicated organisms had no resistant isolates available during comparative testing: *Citrobacter koseri*, *Proteus vulgaris*, and *Serratia liquefaciens*. This was addressed by adding the following limitation in the labeling:

"The ability of the Etest to detect resistance with the following combination (s) is unknown because resistant strains were not available at the time of comparative testing. If such a strain is observed, it should be submitted to a reference laboratory for further testing

- Ceftolozane/Tazobactam: *Citrobacter koseri*, *Proteus vulgaris*, and *Serratia liquefaciens*"

# MIC Trends:

Using the combined clinical and challenge data for *Enterobacteriaceae* an analysis of trending was conducted. This trending calculation takes into account MIC values that are determined to be one or more doubling dilution lower or higher compared to the reference method irrespective whether the device MIC values are on-scale or not. The combined data for 821 results constitute the evaluable data for trend analysis which is presented in Table 5.

Table 5. Trending in Combined Clinical and Challenge *Enterobacteriaceae*

|  Organism | Difference in MIC as Compared to the CLSI Reference Method  |   |   |   |
| --- | --- | --- | --- | --- |
|  *Enterobacteriaceae* | # Isolates | ≤1 dil lower | Exact | ≥1 dil. higher  |
|   |  821 | 282 (34.35%)* | 441 (53.71%) | 98 (11.94%)*  |

*Difference 22.41% 95%CI: 18.44% 26.30%

A trend towards lower MIC reading was observed in the overall performance of *Enterobacteriaceae* compared to the CLSI broth micro-dilution method, which raises concerns for potential occurrence of very major errors. This trending and the potential for occurrence of very major error(s) for Ceftolozane/Tazobactam was addressed in the labeling by adding the following footnote:

"Etest Ceftolozane/Tazobactam MIC values tended to be in exact agreement or at least one doubling dilution lower when testing Enterobacteriaceae compared to the reference broth micro-dilution".

b. Matrix comparison:

Not Applicable

{11}

Page 12 of 12

3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable

4. Clinical cut-off:
Not Applicable

5. Expected values/Reference range:

Table 6: FDA Interpretive Criteria for Ceftolozane/Tazobactam

|  Organism | MIC (μg/mL)  |   |   |
| --- | --- | --- | --- |
|   |  S | I | R  |
|  Enterobacteriaceae | ≤2/4 | 4/4 | ≥8/4  |
|  Pseudomonas aeruginosa | ≤4/4 | 8/4 | ≥16/4  |

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JWY/K170670](https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JWY/K170670)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
