← Product Code [JTN](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JTN) · K994067

# HARDY DISK AMPICILLIN/SULBACTAM (K994067)

_Hardy Diagnostics · JTN · Jan 14, 2000 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JTN/K994067

## Device Facts

- **Applicant:** Hardy Diagnostics
- **Product Code:** [JTN](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JTN.md)
- **Decision Date:** Jan 14, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1620
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

HardyDisk™ Antimicrobial Sensitivity Disks are used for semi-quantitative in vitro susceptibility testing by the agar diffusion test procedure (Kirby-Bauer) of rapidly growing and certain fastidious bacterial pathogens. Standardized methods for agar diffusion testing have been described for Enterobacteriaceae, Staphylococcus spp., Acinetobacter spp., Acinetobacter spp., Listeria monocytogenes, Enterococcus spp., other streptococci and, by modified procedures, Haemophilus influenzae, Neisseria gonorrhoeae and Streptococcus pneumoniae. HardyDisk™ Ampicillin/Sulbactam is indicated for in vitro activity against gram negative organisms including Escherichia coli, Klebsiella spp., Proteus spp., Providencia spp., Morganella morganii, Haemophilus spp. and Neisseria gonorrhoeae. HardyDisk™ Ampicillin/Sulbactam is also indicated for in vitro activity against gram positive organisms including Staphylococcus spp., enterococci, and Streptococcus spp., including S. pneumoniae.

## Device Story

HardyDisk™ Ampicillin/Sulbactam 10/10mcg are antimicrobial sensitivity disks used in clinical laboratories for semi-quantitative in vitro susceptibility testing. Device utilizes Kirby-Bauer agar diffusion method; disks are placed on inoculated agar plates. Antibiotic diffuses into medium; zone of inhibition measured after incubation. Results assist clinicians in selecting appropriate antibiotic therapy for bacterial infections. Device intended for professional use by laboratory personnel.

## Clinical Evidence

No clinical data provided. Device relies on established standardized agar diffusion (Kirby-Bauer) methodology for in vitro susceptibility testing.

## Technological Characteristics

Antimicrobial sensitivity disks containing 10/10mcg of Ampicillin/Sulbactam. Operates via passive diffusion on agar media. Standardized for Kirby-Bauer disk diffusion susceptibility testing.

## Regulatory Identification

An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure by a disc-agar diffusion technique or a disc-broth elution technique the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents. In the disc-agar diffusion technique, bacterial susceptibility is ascertained by directly measuring the magnitude of a zone of bacterial inhibition around the disc on an agar surface. The disc-broth elution technique is associated with an automated rapid susceptibility test system and employs a fluid medium in which susceptibility is ascertained by photometrically measuring changes in bacterial growth resulting when antimicrobial material is eluted from the disc into the fluid medium. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. In the center of the seal is a stylized image of three human figures.

## JAN 1 4 2000

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Melissa M. Traylor, RAC Director of Technical Services Regulatory Affairs Hardy Diagnostics 1430 West McCoy Lane Santa Maria, California 93455

K994067 Re:

> Trade Name: HardyDisk™ Ampicillin/Sulbactam 10/10mcg Regulatory Class: II Product Code: JTN Dated: November 29, 1999 Received: December 1, 1999

Dear Ms. Traylor:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{1}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Hardy Diagnostics. The logo consists of a circular graphic on the left and the text "HARDY DIAGNOSTICS" on the right. The word "HARDY" is on top of the word "DIAGNOSTICS".

mail: Sales@Hardydiannostics.com Website: Hardvdiaanostic

> Santa Maria, California 430 West McCov Lane Santa Maria, CA 93455 Tel: 18051 346-7766 Fax: 18051 346-7760

> > Salt Lake City, Utah 254 W Cottane Ave Sandv. UT 84070 Tel: (800) 995-4273 Fax: 18011562-3214

Phoenix, Arizona 535 W. Iron Ave., #105 Mesa A7 85210 Tel: 18001 995-8456 Fax: 16021 464-9828

510K Number K994067

## Indications for Use Statement-HardyDisk ™ Ampicillin/Sulbactam 10/10mcg

HardyDisk™ Antimicrobial Sensitivity Disks are used for semi-quantitative in vitro susceptibility testing by the agar diffusion test procedure (Kirby-Bauer) of rapidly growing and certain fastidious bacterial pathogens. Standardized methods for agar diffusion testing have been described for Enterobacteriaceae, Staphylococcus spp., Acinetobacter spp., Acinetobacter spp., Listeria monocytogenes, Enterococcus spp., other streptococci and, by modified procedures, Haemophilus influenzae, Neisseria gonorrhoeae and Streptococcus pneumoniae.

HardyDisk™ Ampicillin/Sulbactam is indicated for in vitro activity against gram negative organisms including Escherichia coli, Klebsiella spp., Proteus spp., Providencia spp., Morganella morganii, Haemophilus spp. and Neisseria gonorrhoeae.

HardyDisk™ Ampicillin/Sulbactam is also indicated for in vitro activity against gram positive organisms including Staphylococcus spp., enterococci, and Streptococcus spp., including S. pneumoniae.

<End>

Concurrence of CDRH-ODE

*Woody Dubois*
(Division Sign-off)

Division of Cli aboratory Devices 510(k) Number

Prescription Use
X
(per 21 CFR 801.109)

OR

Over-the Counter-Use (Optional format 1-2-96)

> Serving the laboratorian since 1 980 · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JTN/K994067](https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JTN/K994067)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).