K221826 · Becton, Dickinson and Company · JTN · Dec 21, 2022 · Microbiology
Device Facts
Record ID
K221826
Device Name
BD BBL Sensi-Disc Cefiderocol 30ug (FDC-30)
Applicant
Becton, Dickinson and Company
Product Code
JTN · Microbiology
Decision Date
Dec 21, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1620
Device Class
Class 2
Indications for Use
BD BBL Sensi-Disc Antimicrobial Susceptibility Test (AST) Discs are used in the semi-quantitative agar diffusion test method for in vitro susceptibility testing. BD BBL Sensi-Disc Cefiderocol Disc 30 µg (FDC30) can be used to determine susceptibility to Cefiderocol against the following bacteria, as described in the FDA-approved package insert for this antimicrobial agent. Active in vitro and in Clinical Infections Against: Complicated Urinary Tract Infections, Including Pyelonephritis (Gram-negative Bacteria: Escherichia coli, Enterobacter cloacae complex, Klebsiella pneumoniae, Proteus mirabilis, Pseudomonas aeruginosa); Hospital-acquired Bacterial Pneumonia and Ventilator-associated Bacterial Pneumonia (HABP/VABP) (Gram-negative Bacteria: Acinetobacter baumannii complex, Escherichia coli, Enterobacter cloacae complex, Klebsiella pneumoniae, Pseudomonas aeruginosa, Serratia marcescens). Active in vitro Against: Citrobacter freundii complex, Citrobacter koseri, Klebsiella aerogenes, Klebsiella oxytoca, Morganella morganii, Proteus vulgaris, Providencia rettgeri.
Device Story
Device consists of 6mm absorbent paper discs impregnated with 30 µg Cefiderocol; used in semi-quantitative agar diffusion (Kirby-Bauer) testing. Clinicians/technicians inoculate pure bacterial cultures onto agar media; place discs on surface. Antibiotic diffuses into agar; after incubation, zones of inhibition are measured. Measured zones compared against established interpretive criteria to determine susceptibility (S/I/R). Used in clinical microbiology laboratories to guide antimicrobial therapy decisions. Benefits include standardized susceptibility testing for Cefiderocol to inform treatment of serious Gram-negative infections.
Clinical Evidence
Bench testing only. Study evaluated 357 clinical and challenge isolates (Enterobacterales, P. aeruginosa, A. baumannii complex). Overall categorical agreement (CA) was 98.0% compared to the predicate. Reproducibility was 99.63% across readers and lots. Quality control testing met acceptance criteria (>95%).
Technological Characteristics
6mm absorbent paper disc impregnated with 30µg Cefiderocol. Principle: Kirby-Bauer agar diffusion. Manual measurement of zone of inhibition. No electronic components or software.
Indications for Use
Indicated for in vitro susceptibility testing of specific Gram-negative bacteria (Enterobacterales, P. aeruginosa, A. baumannii complex) against Cefiderocol. Used for patients with complicated urinary tract infections or hospital-acquired/ventilator-associated bacterial pneumonia. Contraindications: None stated, but results for certain species/zone diameters require interpretation with clinical/laboratory information due to potential false susceptible results.
Regulatory Classification
Identification
An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure by a disc-agar diffusion technique or a disc-broth elution technique the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents. In the disc-agar diffusion technique, bacterial susceptibility is ascertained by directly measuring the magnitude of a zone of bacterial inhibition around the disc on an agar surface. The disc-broth elution technique is associated with an automated rapid susceptibility test system and employs a fluid medium in which susceptibility is ascertained by photometrically measuring changes in bacterial growth resulting when antimicrobial material is eluted from the disc into the fluid medium. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K221826
B Applicant
BD Diagnostic Systems
Becton Dickinson, and Company
C Proprietary and Established Names
BD BBL Sensi-Disc Cefiderocol 30ug (FDC-30)
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JTN | Class II | 21 CFR 866.1620 - Antimicrobial Susceptibility Test Disc | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain substantial equivalence determination for Cefiderocol Antimicrobial Susceptibility Test Disc.
B Measurand:
Cefiderocol 30 µg (FDC30)
C Type of Test:
Antimicrobial Susceptibility Test Disc
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K221826 - Page 2 of 11
## III Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
BD BBL Sensi-Disc Antimicrobial Susceptibility Test (AST) Discs are used in the semi-quantitative agar diffusion test method for in vitro susceptibility testing.
BD BBL Sensi-Disc Cefiderocol Disc 30 µg (FDC30) can be used to determine susceptibility to Cefiderocol against the following bacteria, as described in the FDA-approved package insert for this antimicrobial agent.
## Active in vitro and in Clinical Infections Against:
| Complicated Urinary Tract Infections, Including Pyelonephritis | Hospital-acquired Bacterial Pneumonia and Ventilator-associated Bacterial Pneumonia (HABP/VABP) |
| --- | --- |
| Gram-negative Bacteria | Gram-negative Bacteria |
| Escherichia coli | Acinetobacter baumannii complex |
| Enterobacter cloacae complex | Escherichia coli |
| Klebsiella pneumoniae | Enterobacter cloacae complex |
| Proteus mirabilis | Klebsiella pneumoniae |
| Pseudomonas aeruginosa | Pseudomonas aeruginosa |
| | Serratia marcescens |
## Active in vitro Against:
| Citrobacter freundii complex |
| --- |
| Citrobacter koseri |
| Klebsiella aerogenes |
| Klebsiella oxytoca |
| Morganella morganii |
| Proteus vulgaris |
| Providencia rettgeri |
## C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
## Limitations:
- The ability of the BD BBL Sensi Disc to detect resistance with the following combination(s) is unknown because an insufficient number of resistant strains were encountered at the time of comparative testing: Cefiderocol (30 µg) and Enterobacterales group, P. aeruginosa and Acinetobacter baumannii. If such a
{2}
strain is encountered, it should be submitted to a reference laboratory for further testing.
- While the categorical agreement for BD BBL Sensi-Disc Cefiderocol compared to FDA cleared Disk analysis was > 95% for Pseudomonas aeruginosa and Acinetobacter baumannii complex, categorical agreement for BD BBL Sensi-Disc Cefiderocol compared to historical BMD MIC values was below 90%, caused by the occurrence of false susceptible results and minor errors. Test results for Cefiderocol/P. aeruginosa which provide a disc zone of inhibition ≥ 22 mm, and test results for Cefiderocol/A. baumannii which provide a disc zone of inhibition ≥ 19 mm, should be interpreted in conjunction with other clinical and laboratory information.
- While the categorical agreement for BD BBL Sensi-Disc Cefiderocol compared to FDA cleared Disk analysis was > 95% for Enterobacterales group organisms, categorical agreement for BD BBL Sensi-Disc Cefiderocol compared to historical BMD MIC values resulted in false susceptible results. Citrobacter freundii complex, Citrobacter koseri, Klebsiella pneumoniae, Morganella morganii, and Proteus mirabilis isolates that provide a disc zone of inhibition ≥ 16 mm should be interpreted in conjunction with other clinical and laboratory information.
## D Special Instrument Requirements:
Not applicable.
## IV Device/System Characteristics:
### A Device Description:
The BD BBL Sensi-Disc Cefiderocol 30 µg (FDC30) device is comprised of 6 mm discs prepared by impregnating high quality absorbent paper with accurately determined amounts of Cefiderocol. Discs are clearly marked on both sides with the code FDC30. The code designates the agent Cefiderocol (FDC) and the drug content (30 µg).
BD BBL Sensi-Disc Antimicrobial Susceptibility Test Discs are supplied in cartridges containing 50 discs each. The last disc in each cartridge is marked "X" and contains the drug as coded. BD BBL Sensi-Disc Antimicrobial Susceptibility Test Discs can be dispensed using a BD BBL Sensi-Disc Dispenser.
### B Principle of Operation:
Filter paper discs impregnated with specified concentrations of antimicrobial agents are used to perform disk diffusion susceptibility testing. The test is performed by inoculating pure cultures of clinical isolates onto the suitable medium and placing the AST disc on the surface of the medium. The antibiotic within the disc diffuses into the agar. After incubation, the zones of
K221826 - Page 3 of 11
{3}
inhibition around the discs are measured and compared against recognized zone diameter ranges for the specific antimicrobial agent/organism combinations being tested.
# V Substantial Equivalence Information:
# A Predicate Device Name(s):
HardyDisk AST Cefiderocol 30ug (FDC30)
# B Predicate 510(k) Number(s):
K193504
# C Comparison with Predicate:
| Device & Predicate Device: | Device: K221826 | Predicate: K193504 |
| --- | --- | --- |
| Device Trade Name | BD BBL Sensi-Disc Cefiderocol 30 μg (FDC30) | HardyDisk AST Cefiderocol 30 μg (FDC30) |
| General Device Characteristic Similarities | | |
| Regulation | 866.1620 | Same |
| Product Code | JTN | Same |
| Intended Use/Indications for Use | semi-quantitative agar diffusion test method for in vitro susceptibility testing. | Same |
| Antimicrobial Agent | Cefiderocol | Same |
| Antimicrobial Agent Concentration | 30 μg | Same |
| Interpretation | The user will interpret the zone diameter according to the established interpretive criteria for the drug. | Same |
| Methodology | Kirby-Bauer Disk Diffusion Susceptibility Test Protocol requires the user to determine categorical interpretations (S/I/R) using the measured zone diameters. | Same |
| Result Interpretation Method | Measurement of zone size. | Same |
| General Device Characteristic Differences | | |
| Manufacturing Specifications | BD's specifications | Hardy Diagnostics' specifications |
# VI Standards/Guidance Documents Referenced:
K221826 - Page 4 of 11
{4}
- CLSI. Performance Standards for Antimicrobial Susceptibility Testing. 31st ed. CLSI supplement M100. Clinical and Laboratory Institute; 2021.
- CLSI. Performance Standards for Antimicrobial Susceptibility Testing. 32nd ed. CLSI supplement M100. Clinical and Laboratory Institute; 2022.
- CLSI. Performance Standards for Antimicrobial Disk Susceptibility Tests. 13th ed. CLSI standard M02. Clinical and Laboratory Standards Institute; 2018.
# VII Performance Characteristics (if/when applicable):
# A Analytical Performance:
# 1. Precision/Reproducibility:
Reproducibility was conducted at one external site using 15 isolates, tested in triplicate with two disc lots on three separate days using one lot of BD Mueller Hinton agar (MHA) media. Each test was visually read by three independent readers with results blinded to avoid any bias results, resulting in 270 data points for evaluation (15 isolates x 2 disc lots x 3 days = 90 test results read by 3 independent readers = 270 data points).
The reproducibility study included the following species, indicated on the FDA-approved drug label: 11 Enterobacterales (1 Citrobacter koseri, 1 Enterobacter cloacae, 2 Escherichia coli, 1 Klebsiella aerogenes, 3 Klebsiella pneumoniae, 1 Morganella morganii, 1 Proteus mirabilis, 1 Proteus vulgaris, 1 Serratia marcescens), 2 Pseudomonas aeruginosa, and 1 Acinetobacter baumannii.
Reproducibility was calculated as the percent of results which were within $\pm 3\mathrm{mm}$ difference in zone diameter comparing test results with the modal zone diameter value. Summary results between disc lots and across readers are shown in Table 1.
Table 1: Reproducibility Summary
| Between Disc Lots | | | Across Readers | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Lot #1 | Lot #2 | All Lots | Reader #1 | Reader #2 | Reader #3 | All Readers |
| 100% | 99.26% | 99.63% | 100% | 100% | 99% | 99.63% |
| (135/135) | (134/135) | (269/270)a | (90/90) | (90/90) | (89/90) | (269/270) |
The reproducibility performance between disc lots and across readers is $>95\%$ and meets the acceptance criteria.
# 2. Linearity:
Not applicable.
# 3. Analytical Specificity/Interference:
Not applicable.
K221826 - Page 5 of 11
{5}
4. Assay Reportable Range:
Not applicable.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Quality Control (QC) Testing:
The CLSI-recommended quality control (QC) isolates, Escherichia coli (ATCC 25922), and Pseudomonas aeruginosa (ATCC 27853) were tested a sufficient number of times (i.e., a minimum of 20 replicates per lot per reader). One FDA cleared comparator disc lot and two BD disc lots were used. Each test was visually read by three independent readers, resulting in 148 data point (74 data points per lot) and 74 comparator disc data points. The performance is shown in Table 2.
Table 2: Quality Control Performance of Cefiderocol (30 μg)
| QC Organism | Zone Diameter (mm) Range | BD Lot 1^{1} (N) | BD Lot 2^{1} (N) | Comparator Disc^{2} (N) |
| --- | --- | --- | --- | --- |
| Escherichia coli ATCC25922
Expected Range: 25-31 mm | 22 | | | |
| | 23 | | | |
| | 24 | | 2 | 2 |
| | 25 | 7 | 11 | 11 |
| | 26 | 18 | 12 | 16 |
| | 27 | 15 | 13 | 14 |
| | 28 | 21 | 21 | 20 |
| | 29 | 11 | 14 | 10 |
| | 30 | 1 | | |
| | 31 | | | |
| | 32 | | | |
| | 33 | | | |
| | 34 | | | |
| Pseudomonas aeruginosa ATCC 27853
Expected Range: 22-31 mm | 19 | | | |
| | 20 | | | |
| | 21 | | | |
| | 22 | | | 1 |
| | 23 | 2 | 3 | 4 |
| | 24 | 10 | 7 | 14 |
| | 25 | 23 | 17 | 28 |
| | 26 | 32 | 38 | 23 |
| | 27 | 6 | 8 | 4 |
| | 28 | 1 | 1 | |
| | 29 | | | |
| | 30 | | | |
| | 31 | | | |
| 32 | | | | |
K221826 - Page 6 of 11
{6}
| QC Organism | Zone Diameter (mm) Range | BD Lot 1^{1} (N) | BD Lot 2^{1} (N) | Comparator Disc^{2} (N) |
| --- | --- | --- | --- | --- |
| | 33 | | | |
| | 34 | | | |
ATCC = American Type Culture Collection
<sup>1</sup>Two BD disc lots were tested (Lot a and Lot 2).
<sup>2</sup>One FDA cleared comparator disc lot was tested.
The BD disc QC performance is > 95% and is acceptable.
## Inoculum Density Check:
Colony counts were conducted for all QC, reproducibility isolates and clinical isolates. All were within the expected range.
6. Detection Limit:
Not applicable.
7. Assay Cut-Off:
Not applicable.
## B Comparison Studies:
1. Method Comparison with Predicate Device:
The BD BBL Sensi-Disc Cefiderocol 30 µg (FDC30) was compared with an FDA cleared disc (of the same antimicrobial, mass/concentration, and content). The study was conducted at one external testing site. Three independent operators participated in reading of test results with isolates evenly distributed to mimic testing at multiple sites. Testing was performed with one lot of disc of each test manufacturer (BD and comparator FDA-cleared disc) utilizing MHA media following the method outlined in CLSI M02 13th edition method.
## Clinical:
Clinical testing was performed at one U.S. site with both the BD BBL Sensi-Disc Cefiderocol and the comparator FDA cleared disc using a total of 296 clinical isolates including 207 Enterobacterales (10 Citrobacter freundii, 2 Citrobacter koseri, 10 isolates of Enterobacter cloacae, 30 isolates of Escherichia coli, 3 isolates of Klebsiella aerogenes, 18 isolates of Klebsiella oxytoca, 62 isolates of Klebsiella pneumoniae, 17 isolates of Morganella morganii, 30 isolates of Proteus mirabilis, 2 isolates of Proteus vulgaris, 8 isolates of Providencia rettgeri, 15 isolates of Serratia marcescens), 52 Pseudomonas aeruginosa, 37 Acinetobacter baumannii complex.
## Challenge:
Challenge testing was performed at one U.S site. A total of 61 challenge isolates were tested which included 49 Enterobacterales (1 Citrobacter braakii, 3 Citrobacter freundii, 2 Citrobacter koseri, 1 Enterobacter asburiae, 3 Enterobacter cloacae, 6 Escherichia, 3 Klebsiella aerogenes, 2 Klebsiella oxytoca, 9 Klebsiella pneumoniae, 3 Morganella
K221826 - Page 7 of 11
{7}
morganii, 2 Proteus mirabilis, 5 Proteus vulgaris, 4 Providencia rettgeri, 1 Providencia stuartii, 4 Serratia marcescens), 5 Pseudomonas aeruginosa, 7 Acinetobacter baumannii complex.
Table 3: Performance of the BD BBL Sensi-Disc Cefiderocol Dis vs. Comparator FDA Cleared Disc
| | Total | CA# | CA% | S(#) | I(#) | R(#) | VMJ | MAJ | MIN |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Enterobacterales Combined [Breakpoints (in mm): ≥16(S), 9-15(I), ≤8(R)] | | | | | | | | | |
| Clinical | 207 | 207 | 100.0% | 207 | 0 | 0 | 0 | 0 | 0 |
| Challenge | 49 | 45 | 91.8% | 37 | 10 | 2 | 0 | 0 | 4 |
| Combined | 256 | 252 | 98.4% | 244 | 10 | 2 | 0 | 0 | 4 |
| Pseudomonas aeruginosa [Breakpoints (in mm), ≥22(S), 13-21(I), ≤12(R)] | | | | | | | | | |
| Clinical | 52 | 51 | 98.1 | 51 | 1 | 0 | 0 | 0 | 1 |
| Challenge | 5 | 5 | 100.0 | 2 | 2 | 1 | 0 | 0 | 0 |
| Combined | 57 | 56 | 98.2 | 53 | 3 | 1 | 0 | 0 | 1 |
| Acinetobacter baumannii complex [Breakpoints (in mm), ≥19(S), 12-18(I), ≤11(R)] | | | | | | | | | |
| Clinical | 37 | 35 | 94.6 | 28 | 5 | 4 | 0 | 0 | 2 |
| Challenge | 7 | 7 | 100.0 | 2 | 2 | 3 | 0 | 0 | 0 |
| Combined | 44 | 42 | 95.5 | 30 | 7 | 7 | 0 | 0 | 2 |
CA – Category Agreement
MIN – minor errors
S – Susceptible isolates
MAJ – major errors
I – Intermediate isolates
VMJ – very major errors
R – Resistant isolates
Category Agreement (CA) is when the BD result interpretation agrees exactly with the comparator result interpretation.
Performance when testing the total 357 clinical and challenge isolates are shown in Table 3 and summarized below.
- The overall performance of the BD BBL Sensi-Disc Cefiderocol disc as compared to the FDA cleared comparator disc for Enterobacterales group (Table 3) is acceptable with 98.4% CA. There were 4 minor errors and no major or very major errors.
- The overall performance of the BD BBL Sensi-Disc Cefiderocol disc as compared to the FDA cleared comparator disc for P. aeruginosa group (Table 3) is acceptable with 98.2% CA. There was 1 minor error and no major or very major errors.
- The overall performance of the BD BBL Sensi-Disc Cefiderocol disc as compared to the FDA cleared comparator disc for Acinetobacter baumannii group (Table 3) is acceptable with 95.5% CA. There were 2 minor errors and no major or very major errors.
As required under 511A(b)(2)(C)(ii)(I) of the Federal Food, Drug and Cosmetic Act, the following statement is included in the 'Warnings and Precautions' section in the device labeling to address testing of non-indicated species:
"Per the FDA-Recognized Susceptibility Test Interpretive Criteria website, the safety and efficacy of antimicrobial drugs, for which antimicrobial susceptibility is tested by this AST device, may or may not have been established in adequate and well-controlled clinical trials for treating clinical infections due to microorganisms outside of those found in the indications and usage in the drug label. The clinical significance of susceptibility information in those instances is unknown. The approved labeling for specific antimicrobial drugs provides the uses for which the antimicrobial drug is approved."
K221826 - Page 8 of 11
{8}
# Resistance Isolates:
A total of 357 clinical and challenge isolates were tested when the BD BBL Sensi-Disc was compared to the FDA cleared comparator disc. However, an insufficient number of resistant isolates were available for testing. To address the lack of resistant strains encountered during the clinical evaluation, the following limitation was added in the device labeling:
"The ability of the BD BBL Sensi Disc to detect resistance with the following combination(s) is unknown because an insufficient number of resistant strains were encountered at the time of comparative testing: Cefiderocol (30μg) and Enterobacterales group, P. aeruginosa and Acinetobacter baumannii. If such a strain is encountered, it should be submitted to a reference laboratory for further testing".
# Secondary Analysis:
A secondary analysis of the results obtained with the 357 clinical and challenge isolates was also performed. The analysis was conducted to assess the qualitative categorical agreement (CA) of the BD BBL Sensi-Disc diffusion results compared to the reference broth microdilution (BMD) results (based on historical MIC data from BMD testing performed at the initial recovery of isolates during the drug clinical trial). Analysis of the BD BBL Sensi-Disc when tested against the historical MIC value for each isolate showed a high number of categorical errors. The results of the BD BBL Sensi-Disc Cefiderocol versus historical MIC values are summarized below.
- The analysis of the BD BBL Sensi-Disc Cefiderocol compared to historical BMD MIC values for *P. aeruginosa* had a low CA of 75.4% due to the high number of minor errors (6) and 8 very major errors (8/10 80.0%). There were no major errors. Refer to limitation below.
- The analysis of the BD BBL Sensi-Disc Cefiderocol compared to historical BMD MIC values for *Acinetobacter baumannii* complex had a low CA of 61.4% due to the high number of minor errors (7), very major errors (9/20, 45.0%) and one major error (1/24, 4.2%). This single major error was accepted as a random error and did not require a limitation.
To address the low CA due to categorical errors obtained with Cefiderocol when tested with *P. aeruginosa* and *A. baumannii* complex, the following limitation was added to the device labeling:
"While the categorical agreement for BD BBL Sensi-Disc Cefiderocol compared to FDA cleared Disk analysis was >95% for *Pseudomonas aeruginosa* and *Acinetobacter baumannii* complex, categorical agreement for BD BBL Sensi-Disc Cefiderocol compared to historical BMD MIC values was below 90%, caused by the occurrence of false susceptible results and minor errors. Test results for Cefiderocol/P. aeruginosa which provide a disc zone of inhibition ≥22 mm, and test results for Cefiderocol/A. baumannii complex which provide a disc zone of
K221826 - Page 9 of 11
{9}
inhibition ≥ 19 mm, should be interpreted in conjunction with other clinical and laboratory information".
- The analysis of the BD BBL Sensi-Disc Cefiderocol compared to historical BMD MIC values for Enterobacterales is acceptable with 93.0% CA. However, the very major error (VMJ) rate was 41.2% (7/17) which does not meet FDA’s acceptance criteria of ≤ 2% VMJ error rate. There were 11 minor errors and no major errors. To address the categorical errors obtained with Cefiderocol when tested with Enterobacterales, the following limitation was added to the device labeling:
"While the categorical agreement for BD BBL Sensi-Disc Cefiderocol compared to FDA cleared Disk analysis was > 95% for Enterobacterales group organisms, categorical agreement for BD BBL Sensi-Disc Cefiderocol compared to historical BMD MIC values resulted in false susceptible results. Citrobacter freundii complex, Citrobacter koseri, Klebsiella pneumoniae, Morganella morganii, and Proteus mirabilis isolates that provide a disc zone of inhibition ≥ 16 mm should be interpreted in conjunction with other clinical and laboratory information.
2. Matrix Comparison:
Not applicable.
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
The FDA-identified interpretive criteria for Cefiderocol are listed in Table 4.
K221826 - Page 10 of 11
{10}
Table 4: FDA Identified Interpretive Criteria for Cefiderocol¹
| | Minimum Inhibitory Concentrations (μg/mL) | | | Disk Diffusion (zone diameters in mm) | | |
| --- | --- | --- | --- | --- | --- | --- |
| Pathogen | S | I | R | S | I | R |
| Enterobacterales²,³ | ≤4 | 8 | ≥16 | ≥16 | 9-15 | ≤8 |
| Pseudomonas aeruginosa | ≤1 | 2 | ≥4 | ≥22 | 13-21 | ≤12 |
| Acinetobacter baumannii complex | ≤1 | 2 | ≥4 | ≥19 | 12-18 | ≤11 |
¹ According to the FDA STIC Website.
² Clinical efficacy was shown for Escherichia coli, Klebsiella pneumoniae, Proteus mirabilis, and Enterobacter cloacae complex in patients with complicated urinary tract infections (cUTI).
³ Clinical efficacy was shown for Escherichia coli, Klebsiella pneumoniae, Enterobacter cloacae complex, and Serratia marcescens in patients with hospital-acquired bacterial pneumonia and ventilator-associated bacterial pneumonia (HABP/VABP).
S = Susceptible
I = Intermediate
R = Resistant
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
To support the implementation of changes to FDA-recognized susceptibility test interpretive criteria (i.e., breakpoints), this submission included a breakpoint change protocol that was reviewed and accepted by FDA. This protocol addresses future revisions to device labeling in response to breakpoint changes that are recognized on the FDA STIC webpage (https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm410971.htm). The protocol outlined the specific procedures and acceptance criteria that Becton Dickinson, and Company (BD Diagnostic Systems) intends to use to evaluate the BD BBL Sensi-Disc Cefiderocol 30μg (FDC-30) when revised breakpoints for Cefiderocol are published on the FDA STIC webpage. The breakpoint change protocol included with the submission indicated that if specific criteria are met, BD Diagnostic Systems will update the Cefiderocol 30μg device label to include (1) the new breakpoints, (2) an updated performance section after re-evaluation of data in this premarket notification with the new breakpoints, and (3) any new limitations as determined by their evaluation.
K221826 - Page 11 of 11
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.