K112586 · Becton, Dickinson and Company (BD) · JTN · Oct 18, 2011 · Microbiology
Device Facts
Record ID
K112586
Device Name
CEFTAROLINE 30 MICROGRAMS, BBL(TM) SENSI-DISC(TM)
Applicant
Becton, Dickinson and Company (BD)
Product Code
JTN · Microbiology
Decision Date
Oct 18, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1620
Device Class
Class 2
Intended Use
Antimicrobial Susceptibility Test Disks are used for semi-quantitative in vitro susceptibility testing by standardized agar diffusion test procedures. Ceftaroline 30μg BBL™ Sensi-Disc™ is intended for use in determining the susceptibility to Ceftaroline of a wide range of bacterial pathogens.
Device Story
Device consists of high-quality paper disks impregnated with 30µg of Ceftaroline. Used in clinical microbiology laboratories for semi-quantitative in vitro susceptibility testing via agar diffusion (Bauer-Kirby method). Disks are placed on inoculated Mueller Hinton Agar (or supplemented media for specific organisms). Following incubation, healthcare providers measure zones of inhibition surrounding disks. Measured diameters are compared against established zone size ranges from FDA-approved drug inserts and CLSI/NCCLS standards (M2, M100) to categorize isolates as susceptible, intermediate, or resistant. Results assist clinicians in selecting appropriate antimicrobial therapy for patients with bacterial infections.
Clinical Evidence
No clinical diagnostic studies were performed for this device. Performance is supported by clinical outcome studies conducted by the drug manufacturer for Ceftaroline approval, which established the interpretive criteria and quality control ranges used for disk diffusion testing.
Technological Characteristics
High-quality paper disks impregnated with 30μg Ceftaroline. Principle: Agar diffusion (Bauer-Kirby). Test medium: Mueller Hinton Agar (or supplemented variants). Standardized via CLSI M02-A10 and M100. Manual measurement of inhibition zones.
Indications for Use
Indicated for in vitro agar diffusion susceptibility testing to determine bacterial susceptibility to Ceftaroline. Active against Gram-positive (S. aureus, S. pyogenes, S. agalactiae, S. pneumoniae, S. dysgalactiae) and Gram-negative (E. coli, K. pneumoniae, K. oxytoca, H. influenzae, C. koseri, C. freundii, E. cloacae, E. aerogenes, M. catarrhalis, M. morganii, P. mirabilis, H. parainfluenzae) microorganisms.
Regulatory Classification
Identification
An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure by a disc-agar diffusion technique or a disc-broth elution technique the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents. In the disc-agar diffusion technique, bacterial susceptibility is ascertained by directly measuring the magnitude of a zone of bacterial inhibition around the disc on an agar surface. The disc-broth elution technique is associated with an automated rapid susceptibility test system and employs a fluid medium in which susceptibility is ascertained by photometrically measuring changes in bacterial growth resulting when antimicrobial material is eluted from the disc into the fluid medium. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
Predicate Devices
Other BBL™ Sensi-Disc™ (e.g., Ciprofloxacin 5 µg, BBL™ Sensi-Disc™)
Related Devices
K103538 — HARDYDISK AST CEFTAROLINE, 30 UG MODEL Z939 · Hardy Diagnostics · Dec 17, 2010
K100722 — TELAVANCIN 30 UG, BBL SENSI-DISC · Becton, Dickinson and Company · Mar 29, 2010
K965159 — CETIBUTEN, 30MCG, SENSI-DISC · Becton Dickinson Microbiology Systems · Feb 4, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
OCT 18 2011 K112580
510(k) Submission Ceftaroline 30µg BBL TM Sensi-Disc TM CONFIDENTIAL AND PROPRIETARY
## 510(k) SUMMARY
| SUBMITTED BY: | Becton Dickinson and Company<br>7 Loveton Circle<br>Sparks, MD 21152<br>Phone 410-316-4975<br>Fax: 410-316-4499 |
|------------------------|-----------------------------------------------------------------------------------------------------------------|
| CONTACT NAME: | Katie Coyle, Regulatory Affairs Specialist |
| DATE PREPARED: | September 2, 2011 |
| DEVICE TRADE NAME: | Ceftaroline 30µg, BBL™ Sensi-Disc™ Antimicrobial<br>Susceptibility Test Disks |
| DEVICE COMMON NAME: | Antimicrobial Susceptibility Test Disks |
| DEVICE CLASSIFICATION: | 21 CFR§866.1620, Class II (Product Code JTN),<br>Susceptibility Test Disks, Antimicrobial |
| PREDICATE DEVICE: | Other BBL™ Sensi-Disc™<br>(e.g., Ciprofloxacin 5 µg, BBL™ Sensi-Disc™) |
# INTENDED USE:
Antimicrobial Susceptibility Test Disks are used for semi-quantitative in vitro susceptibility testing by standardized agar diffusion test procedures. Ceftaroline 30μg BBL™ Sensi-Disc™ is intended for use in determining the susceptibility to Ceftaroline of a wide range of bacteria, as described in the "Indications for Use" section. Zone sizes used for interpretation of tests, including control organism limits, were determined by the antimicrobic manufacturer and received FDA approval under NDA Number 200327.
Becton, Dickinson and Company
{1}------------------------------------------------
## 510(k) SUMMARY
### Indications for Use:
Use of Ceftaroline 30µg BBL™ Sensi-Disc™ for in vitro agar diffusion susceptibility testing is indicated when there is a need to determine the susceptibility of bacteria to Ceftaroline. The concentration of 30ug has been shown to be active in vitro aqainst most strains of microorganisms listed below, as described in the FDA approved drug insert for this antimicrobic.
# Active In Vitro and in Clinical Infections Against:
# Gram-positive Microorganisms
Staphylococcus aureus (including methicillin-susceptible and -resistant isolates) Streptococcus pyogenes Streptococcus agalactiae Streptococcus pneumoniae
# Gram-negative Microorganisms
Escherichia coli Klebsiella pneumoniae Klebsiella oxytoca Haemophilus influenzae
## Active In Vitro Against:
Gram-positive Microorganisms Streptococcus dysgalactiae
#### Gram-negative Microorganisms
Citrobacter koseri Citrobacter freundii Enterobacter cloacae Enterobacter aerogenes Moraxella catarrhalis Morganella morqanii Proteus mirabilis Haemophilus parainfluenzae
{2}------------------------------------------------
## DEVICE DESCRIPTION:
Ceftaroline 30ug BBL" Sensi-Disc" is prepared by impregnating high quality paper with accurately determined amounts of Ceftaroline supplied by the drug manufacturer. Each Ceftaroline disk is clearly marked on both sides with the agent and drug content. Ceftaroline cartridges each contain 50 impregnated disks that are packed as either a single cartridge in a single box, or in a package containing ten cartridges. Ceftaroline disks are used for semi-quantitative in vitro susceptibility evaluations by the agar diffusion test method.
Agar diffusion susceptibility methods employing dried filter paper disks impregnated with specific concentrations of antimicrobial agents were developed in the 1940s. In order to eliminate or minimize variability in the testing. Bauer et al. developed a standardized procedure in which Mueller Hinton Aqar was selected as the test medium.
Various regulatory agencies and standards-writing organizations subsequently published standardized reference procedures based on the Bauer-Kirby method. Among the earliest and most widely accepted of these standardized procedures were those published by the U.S. Food and Drug Administration (FDA) and the World Health Organization (WHO). The procedure was adopted as a consensus standard by the Clinical and Laboratory Standards Institute (CLSI) [Formerly National Committee for Clinical Laboratory Standards (NCCLS)] and is periodically updated.
## DEVICE PRINCIPLE:
Disks containing a wide variety of antimicrobial agents are applied to the surface of Mueller Hinton Agar plates for Haemophilus Test Medium Aqar for Haemophilus influenzae or Mueller Hinton Agar with 5% Sheep Blood for Streptococcus species] inoculated with pure cultures of clinical isolates. Following incubation, the plates are examined and the zones of inhibition surrounding the disks are measured and compared with established zone size ranges for individual antimicrobial agents in order to determine the agent(s) most suitable for use in antimicrobial therapy. The categorical interpretation [susceptible (S), intermediate (I), or resistant (R)] for the organism being tested with the antimicrobial agent is made by comparing zone diameters to those found in the respective organism tables of the FDA drug insert and/or CLSI/NCCLS Document M2 ("Performance Standards for Antimicrobial Disk Susceptibility Tests) and of CLSI/NCCLS Document M100 ("Performance Standards for Antimicrobial Susceptibility Testing").
{3}------------------------------------------------
## DEVICE COMPARISON:
The BBL" Sensi-Disc" Antimicrobial Susceptibility Test Disks - Ceftaroline 30μg is similar to the BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disks – Ciprofloxacin 5 uq in that:
- Both methods are for antimicrobial susceptibility testing using paper disks . impregnated with an antimicrobial agent.
- . Both methods have the same intended use.
- Both methods provide the user with antimicrobic minimum inhibitory concentration . (MIC) results based on measurements of zone diameters.
- . Both methods require the user to determine categorical interpretations (S//R) using the measured zone diameters against CLSI/NCCLS Approved Standards M2 and M100.
- . Both methods use pure cultures of bacterial isolates.
The BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disks - Ceftaroline 30μg differs from the BBL" Sensi-Disc" Antimicrobial Susceptibility Test Disks - Ciprofloxacin 5 ug in that:
- BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disks Ceftaroline 30ug is a . susceptibility test that uses disks impregnated with the antimicrobic Ceftaroline at a concentration of 30pg while the BBL™ Sensi-Disc " Antimicrobial Susceptibility Test Disks - Ciprofloxacin 5 µg is a susceptibility test that uses disks impregnated with the antimicrobic Ciprofloxacin at a concentration of 5 µg.
- BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disk -- Ceftaroline 30 µg is a . susceptibility test used to test a different battery of microorganisms than the BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disk - Ciprofloxacin 5 μg.
# SUBSTANTIAL EQUIVALENCE TESTING DATA:
See the Ceftaroline drug package insert. "Microbiology".
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Becton Dickinson and Company c/o Ms. Katie Coyle Regulatory Affairs Specialist 7 Loveton Circle Sparks, MD 21152
OCT 1 8 2011
Re: K112586
Trade/Device Name: Ceftaroline 30μg, BBLTM Sensi -DiscTM Regulation Number: 21 CFR§866.1620 Regulation Name: Antimicrobial susceptibility test disc Regulatory Class: Class II Product Code: JTN Dated: September 2, 2011 Received: September 6, 2011
Dear Ms. Coyle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page- 2 - Ms. Katie Coyle
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Jal atozr
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices Radiological Health
Enclosure
{6}------------------------------------------------
### INDICATIONS FOR USE
510(k) Number (if known): K l ( 2 586
Device Name: BBL™ Sensi-Disc™ Antimicrobial Susceptibility Test Disks, Ceftaroline 30μg
#### Indications for Use:
Use of Ceftaroline 30μg, BBL™ Sensi-Disc™ for in vitro agar diffusion susceptibility testing is indicated when there is a need to determine the susceptibility of bacteria to Ceftaroline. The concentration of 30μg has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA approved drug insert for this antimicrobic.
#### Active In Vitro and in Clinical Infections Against:
| Gram-positive Microorganisms<br>Staphylococcus aureus (including methicillin-susceptible and<br>-resistant isolates)<br>Streptococcus pyogenes<br>Streptococcus agalactiae<br>Streptococcus pneumoniae | Gram-negative Microorganisms<br>Escherichia coli<br>Klebsiella pneumoniae<br>Klebsiella oxytoca<br>Haemophilus influenzae |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Active In Vitro Against: | |
| Gram-positive Microorganisms | Gram-negative Microorganisms |
رقش-انجو ative Micr Citrobacter koseri Citrobacter freundii Enterobacter cloacae Enterobacter aerogenes Moraxella catarrhalis Morganella morganii Proteus mirabilis Haemophilus parainfluenzae
Prescription Use V (Per 21 CFR 801 Subpart D)
Streptococcus dysgalactiae
AND/OR
Over-The-Counter Use (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Lindale L. Pope
**Division Sign-Off**
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K112586
Becton, Dickinson and Company
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.