MRSA SELECT - SKIN AND SOFT TISSUE WOUND SPECIMENS
Applicant
Bio-Rad
Product Code
JSO · Microbiology
Decision Date
Oct 29, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1700
Device Class
Class 2
Intended Use
MRSASelect™ is a selective and differential chromogenic medium for the qualitative detection of methicillin resistant Staphylococcus aureus (MRSA) from skin and soft-tissue wound specimens. The medium is indicated for use in conjunction with other laboratory tests and clinical data available to aid in the identification and diagnosis of MRSA from patients with skin and soft-tissue infections. Concomitant cultures and susceptibility testing are necessary for all skin and soft-tissue wound specimens. MRSASelect™ is not intended to guide, or monitor treatment for MRSA infection, or provides results of susceptibility to methicillin. Results can be interpreted after 18 to 28 hours incubation.
Device Story
MRSASelect is a selective/differential chromogenic culture medium for direct detection of MRSA from skin and soft-tissue wound specimens. Medium incorporates antifungal/antibiotic mixture and optimized salt concentration to inhibit non-MRSA growth (yeast, Gram-negative/positive bacteria). Identification relies on cleavage of chromogenic substrate by S. aureus enzymatic activity, producing pink colonies. Used in clinical microbiology laboratories; processed manually by technicians. Plates incubated 18-28 hours at 35-37°C. Results interpreted visually by laboratory personnel. Output aids clinicians in identifying MRSA in wound infections; requires concomitant culture/susceptibility testing. Does not monitor treatment or provide methicillin susceptibility results.
Clinical Evidence
Clinical study of 943 skin and soft-tissue wound samples across four US laboratories. Compared MRSASelect to routine culture (TSA/TSB) with confirmatory testing (Gram stain, Pastorex Staph Plus, mecA/Cefoxitin disk). Results: sensitivity 91.7% (95% CI: 87.3-94.7%), specificity 99.4% (95% CI: 98.5-99.8%). Analytical sensitivity (102 strains) was 97% after 24 hours. Reproducibility was 100%.
Technological Characteristics
Selective/differential chromogenic agar medium. Contains antibiotic/antifungal mixture and optimized salt concentration. Manual visual readout. Incubation 18-28 hours at 35-37°C. No electronic components or software.
Indications for Use
Indicated for qualitative detection of MRSA from skin and soft-tissue wound specimens in patients with skin and soft-tissue infections. Used as an adjunct to other lab tests and clinical data. Not for guiding/monitoring treatment or determining methicillin susceptibility. Concomitant cultures/susceptibility testing required.
Regulatory Classification
Identification
A culture medium for antimicrobial susceptibility tests is a device intended for medical purposes that consists of any medium capable of supporting the growth of many of the bacterial pathogens that are subject to antimicrobial susceptibility tests. The medium should be free of components known to be antagonistic to the common agents for which susceptibility tests are performed in the treatment of disease.
{0}------------------------------------------------
October 21, 2010 Page 39 of 42
# 510(k) SUMMARY
OCT 2 9 2010
| Date of Summary | October 21, 2010 |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | Bio-Rad MRSASelect -- Wound Specimens<br>A selective and differential chromogenic medium for the<br>qualitative detection of methicillin-resistant Staphylococcus aureus<br>from skin and soft-tissue wound specimens. Results can be<br>interpreted after 18 - 28 hours incubation. |
| Sponsor | Bio-Rad<br>3 Boulevard Raymond Poincaré<br>92430 Marnes-la-Coquette<br>France |
| Correspondent | MDC Associates, LLC<br>Fran White, Regulatory Consultant<br>180 Cabot Street<br>Beverly, MA 01915 |
## Substantially Equivalent Device
MRSASelect for skin and soft-tissue wounds is substantially equivalent to Bio-Rad MRSASelect extended incubation (reference 510(k) K081212). Products are thromogenic media used for the detection of MRSA direct from a specimen swab.
Manufacturer: Bio-Rad MRSASelect -- extended incubation Product:
| Product Attribute | Bio-Rad MRSASelect™<br>Extended Incubation | Bio-Rad MRSASelect™<br>Wound Specimen |
|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | MRSASelect is a selective and<br>differential chromogenic medium<br>for the qualitative detection of<br>MRSA direct from a nasal swab<br>for the detection of methicillin<br>resistant Staphylococcus aureus<br>(MRSA). | MRSASelect is a selective and differential<br>chromogenic medium for the<br>qualitative detection of methicillin resistant<br>Staphylococcus aureus (MRSA) from skin<br>and soft-tissue wound specimens. |
| Product format | Chromogenic agar | Chromogenic Agar |
| Read time | After 18-24 hour incubation | After 18-24 hours incubation |
| Quality Control | Daily with recommended | Daily - same recommended organisms |
Similarities
{1}------------------------------------------------
0
| | organisms | |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Differences | |
| Product Attribute | Bio-Rad MRSASelect™ Extended Incubation | Bio-Rad MRSASelect™ Wound Specimen |
| Intended use | MRSASelect is a screen for the detection of colonization of methicillin resistant Staphylococcus aureus (MRSA) to aid in the prevention and control of MRSA infections in healthcare settings. | MRSASelect is indicated for use in conjunction with other laboratory tests and clinical data available to aid in the identification and diagnosis of MRSA from patients with skin and soft-tissue infections. Concomitant cultures and susceptibility testing are necessary for all skin and soft-tissue wound specimens. MRSASelect™ is not intended to guide, or monitor treatment for MRSA infection, or provides results of susceptibility to methicillin. |
## Product Description
Methicillin-resistant Staphylococcus aureus is a major cause of nosocomial and life threatening infections which have been associated with significantly higher rates of mortality and morbidity. The Bio-Rad MRSASelect is a selective and differential chromogenic culture medium for the qualitative detection of MRSA from skin or soft-tissue wound specimens. Results can be interpreted after 18 - 28 hours incubation.
### Intended Use
MIRSASelect™ is a selective and differential chromogenic medium for the qualitative detection of methicillin resistant Staphylococcus aureus (MRSA) from skin and soft-tissue wound specimens. The MRSASelect™ is indicated for use in conjunction with other laboratory tests and clinical data available to aid in the identification and diagnosis of MRSA from patients with skin and soft-tissue infections. Concomitant cultures and susceptibility testing are necessary for all skin and soft-tissue wound specimens. MRSASelect™ is not intended to guide, or monitor treatment for MRSA infection, or provides results of susceptibility to methicillin. Results can be interpreted after 18 to 28 hours incubation.
### Summary of Technology
MRSASelect is a selective and differential chromogenic medium for the qualitative detection of methicillin resistant Staphylocccus aureus (MRSA) from skin and soft-tissue wound specimens. The selectivity of this medium is based on the presence of an antibiotic/antifungal mixture and an optimized salt concentration that inhibits the growth of yeast and the majority of Gram negative and Gram positive bacteria with the exception of methicillin-resistant staphylococci. Identification is based on the cleavage of a chromogenic substrate by a specific enzymatic activity of Staphylococcus aureus leading to a strong
{2}------------------------------------------------
October 21, 2010 Page 41 of 42
pink coloration of the Staphylococcus aureus colonies.
Within 18 - 28 hours incubation time methicillin-resistant Staphylococcus aureus produce small pink colonies on MRSASelect. Coagulase negative methicillin-resistant staphylococci that do not metabolize the chromogenic substrate appear as colorless or white colonies (possibly light pink). Methicillin sensitive staphylococci (MSS) are inhibited.
### Performance Data
### Interfering Substances
In order to evaluate the possible interference of common topical agents used in wound care samples of these agents were inoculated with MRSA and plated on MRSASelect media. Use of the following compounds has an inhibitory effect on MRSA growth that is unrelated to medium performance:
Bactine (Benzalkonium chloride 0.13%. Lidocaine hydrochloride 2.5%); Betadine (liquid) (Povidoneiodine 10%), Iodine Tincture (liquid) (Iodine 2%), Biseptine (liquid) (Chlorhexidine Gluconate 0.25%, Benzalkonium chloride 0.025%), Sodium hypochlorite (liquid), StaphAseptic (ointment) (Benzethonium Chloride 0.2%, Lidocaine HCl 2.5%), and silver chloride.
### Cross Reactivity Testing (Analytical Specificity)
To evaluate the analytical specificity of the MRSASelect™ media 35 bacterial and fungal strains found in wound or skin samples were cultured and inoculated onto MRSASelect™ plates at a concentration of ≥10° CFU/mL.
No cross-reactivity was observed on any strains tested. Most strains showed no growth on MRSASelect™ with the exception of Corynebacterium jeikeium and Candida tropicalis. With both of these organisms pinpoint white colonies were observed: these are not representative of MRSA colonies, and therefore these are not cross-reactants. No variation was seen between 24 and 28 hour incubation time.
Analytical Sensitivity To evaluate the analytical sensitivity of the MRSASElect™, 102 strains of MRSA, including USA100, 200, 300, 500, 600, and 1000 were inoculated onto MRSASelect™ plates at concentrations of 10 to 10 CFU/mL. USA300-0114 was also tested. 97% (99/102) sensitivity was observed after 24 hours incubation.
### Reproducibility
A panel of 6 organisms, including MRSA, MSSA and S. epidermidis, was evaluated on MRSASelect™. It was performed at concentrations of 108 CFU/mL for MRSA, and 108 CFU/mL for non-MRSA. The panel was tested in triplicate each day for three clinical sites. Overall reproducibility was 100% after 24 hours incubation when testing this panel.
{3}------------------------------------------------
## Method Comparison
943 skin and soft-tissue wound samples were collected and tested at four clinical laboratories in the United States. Each sample was tested on MRSASelect™, Trypticase Soy Agar (TSA) with 5% Sheep's Blood, and Tryptic Soy Broth (TSB) with 6.5% NaCl. Samples that were positive on TSA or TSB were confirmed with Gram stain, Pastorex™ Staph Plus, and mecA mediated oxacillin resistance using 30 ug/mL Cefoxitin disk (R: ≤21 mm, S: ≥ 22 mm)
The following results were obtained: specificity 99.4% (95% CI: [98.5, 99.8]) and sensitivity 91.7% (95% CI: [87.3, 94.7}). The overall prevalence of MRSA in the study was 24.2% (95% CI: [21.5, 27.0]). MRSASelect™ vs. Routine Culture and TSB
| | | Routine Culture<br>&TSB | | |
|---------|-------|-------------------------|-----|-------|
| | | Pos | neg | total |
| MRSA | pos | 209 | 4 | 213 |
| Select™ | neg | 19* | 711 | 730 |
| | total | 228 | 715 | 943 |
* For 12/19 samples - MRSA were isolated only from TSB with 6.5% NaCl and were not isolated on initial direct culture.
99.4% [98.5, 99.8] Overall % agreement 97.6% [96.3, 98.4] Specificity Sensitivity 91.7% [87.3, 94.7]
### Statement of Safety and Efficacy
The data presented demonstrates the safety and efficacy of the Bio-Rad MRSASelect™ as compared to routine culture and identification when results are interpreted after 18 to 28 hours incubation in ambient air.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" arranged in a circular fashion around the symbol. The text is in all caps and is evenly spaced around the circle.
#### Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002
BIO-RAD c/o MDC Associates, LLC 180 Cabot Street Beverly, Massachusetts 01915 Attn: Fran White
OCT 2 9 2010
Re: K100589
| Trade/Device Name: MRSASelect |
|------------------------------------------------------------------------|
| Regulation Number: 21 CFR §866.1700 |
| Regulation Name: Culture medium for antimicrobial susceptibility tests |
| Regulatory Class: Class II |
| Product Code: JSO |
| Dated: October 21, 2010 |
| Received: October 25, 2010 |
#### Dear Ms. White:
We.have reviewed your Section 510(k) premarket notification of intent to market the device. referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Ms. Fran White
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Ledale hi-Poole, Ms
day
Sally A. Hojvat, M.Sc., Ph.D. Director
Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
October 21, 2010 Page 37 of 42
OCT 2 9 2010
#### Indications for Use
510(k) Number (if known): K100589
Device Name: MRSASelect
Indications for Use:
MRSASelect™ is a selective and differential chromogenic medium for the qualitative detection of methicillin resistant Staphylococcus aureus (MRSA) from skin and soft-tissue wound specimens. The MRSASelect™ is indicated for use in conjunction with other laboratory tests and clinical data available to aid in the identification and diagnosis of MRSA from patients with skin and soft-tissue infections. Concomitant cultures and susceptibility testing are necessary for all skin and soft-tissue wound specimens. MRSASelect™ is not intended to guide, or monitor treatment for MRSA infection, or provides results of susceptibility to methicillin. Results can be interpreted after 18 to 28 hours incubation.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)
Leedole. Podo, M.S.
---
Division Sign Off
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
KI66585 510(k)
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.