# CHROMID VRE AGAR, MODEL: REF 43 851 (K091025)

_bioMerieux, Inc. · JSO · Mar 2, 2010 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JSO/K091025

## Device Facts

- **Applicant:** bioMerieux, Inc.
- **Product Code:** [JSO](/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JSO.md)
- **Decision Date:** Mar 2, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1700
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Intended Use

chromID™ VRE agar is a selective and differential chromogenic medium containing 8 µg/mL of vancomycin, for the qualitative detection of Enterococcus faecium and E. faecalis showing acquired vancomycin resistance (VRE) in stool specimens. chromID™ VRE agar can be used as an aid to identify, prevent and control VRE colonization in healthcare settings. chromID™ VRE agar is not intended to diagnose VRE infection nor to guide or monitor treatment for infections. Subculture to non-selective media (e.g. trypticase soy agar with 5% sheep blood) is needed for further identification, susceptibility testing and epidemiological typing.

## Device Story

chromID™ VRE agar is a selective, differential chromogenic culture medium used in clinical laboratories for the qualitative detection of vancomycin-resistant Enterococcus faecium and E. faecalis from stool specimens. The medium contains a nutritive base, vancomycin (8 µg/mL), and two chromogenic substrates. It functions by inhibiting non-resistant enterococci and most other bacteria while allowing VRE to grow. The principle of operation relies on enzymatic activity: β-galactosidase-producing VRE E. faecium colonies appear violet, while α-glucosidase-producing VRE E. faecalis colonies appear blue-to-green after 24-48 hours of aerobic incubation. Healthcare providers visually inspect the plates for these characteristic colors to identify potential VRE colonization. This information assists in infection control and epidemiological surveillance. The device is intended for professional use in clinical settings; it does not diagnose active infection or guide treatment. Subculture to non-selective media is required for further identification and susceptibility testing.

## Clinical Evidence

Performance evaluated at four laboratories using 1299 stool samples compared to conventional reference methods (Gram stain, catalase, VITEK 2, 16S-500 sequencing, agar dilution). At 24h, sensitivity was 97.1% and specificity 99.7%. At 48h, sensitivity was 96.9% and specificity 99.7%. Reproducibility was 100% after 48h. Interference, cross-reactivity, and recovery studies were also performed.

## Technological Characteristics

Selective/differential chromogenic medium; nutritive base with peptones, 8 µg/mL vancomycin, and chromogenic substrates. Form factor: agar plate. Manual reading of colony color (violet for E. faecium, blue-to-green for E. faecalis). Aerobic incubation at 35-37°C in the dark.

## Regulatory Identification

A culture medium for antimicrobial susceptibility tests is a device intended for medical purposes that consists of any medium capable of supporting the growth of many of the bacterial pathogens that are subject to antimicrobial susceptibility tests. The medium should be free of components known to be antagonistic to the common agents for which susceptibility tests are performed in the treatment of disease.

## Predicate Devices

- Remel Bile Esculin Azide Agar w/ 6ug/ml Vancomycin ([K972359](/device/K972359.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K091025

Image /page/0/Picture/1 description: The image shows the logo for bioMerieux. The logo consists of the word "BIOMERIEUX" in a stylized font, with a graphic above it. The graphic is a circle bisected by a curved line, with one half of the circle filled with vertical lines.

# 510(k) SUMMARY

A 510(k) Submission Information:

MAR - 222010

### bioMérieux, Inc. Submitter's Name: Address: 595 Anglum Road Hazelwood, MO 63042 Contact Person: Nancy Weaver Associate Director, Regulatory Affairs Phone Number: 314-731-8695 Fax Number: 314-731-8689 Date of Preparation: April 8, 2009 B. Device Name: chromID™ VRE Agar Formal/Trade Name: Classification Name: Culture Media, Antimicrobial Susceptibility Test, Excluding Mueller Hinton Agar. 21 CFR 866.1700

Common Name:

C. Predicate Device:

Excluding Mueller Hints

Culture media

Remel Bile Esculin Azide Agar w/6ug/ml Vancomycin, K972359

### D. 510(k) Summary:

### Intended Use:

chromID™ VRE agar is a selective and differential chromogenic medium containing 8 ug/mL of vancomycin, for the qualitative detection of Enterococcus faecium and E. faecalis showing acquired vancomycin resistance (VRE) in stool specimens, chromID™ VRE agar can be used as an aid to identify, prevent and control VRE colonization in healthcare settings. chromID™ VRE agar is not intended to diagnose VRE infection nor to guide or monitor treatment for infections. Subculture to non-selective media (e.g. trypticase soy agar with 5% sheep blood) is needed for further identification, susceptibility testing and epidemiological typing.

### Device Description:

chromID™ VRE agar is a selective chromogenic medium for the detection of E. faecalis showing acquired vancomycin resistance (VRE), in at risk patients11). The detection of this resistance is particularly important for the prevention and epidemiological surveillance of these infections and also to prevent the 
emergence of vancomycin-resistant Staphylococus aureus (VRSA), by t context, the use of chromID™ VRE agar contributes to the active surveillance for VRE agar is a selective and differential chromogenic medium for the qualitative detection of E. faecalis showing acquired vancomycin resistance (VRE), from stool specimens.

chromID™ VRE agar consists of a rich nutritive base including a variety of peptones. It also contains two chromogenic substrates and a mixture of antibiotics including vancomycin (8 ug/mL) which enable the specific and selective growth of VRE. After 24-48 hours incubation vancomycin-resistant E. faecium colonies are a violet color for ß-galactosidase-producing strains. Vancomycin-resistant E. faecalis colonies are a blue-togreen color for a-ducosidase-producing strains. The selective components in the medium inhibit enterpooccal strains that do not express acquired vancomycin resistance, enterococcal strains that express natural intinsic vancomycin resistance (vanC phenotype: E. gallinarum and E. casselflavus), as well as most Gram-negative and Gram-positive bacteria.

### bioMérieux, Inc.

595 Anglum Road, Hazelwood, Missouri 63042-2320, USA Phone: 314/731-8500 800/638-4835 Fax: 314/731-8700 http://www.biomerieux-usa.com

{1}------------------------------------------------

:

:

# Substantial Equivalence

chromID™ VRE agar is substantially equivalent to Remel Bile Esculin Azide Agar w/ 6 μg/ml
Vancomycin (K972359) .

# Device Comparison Table

|                       | Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate                                                                                                                                                                                                                                                                                                               |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                       | chromID™ VRE Agar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Remel Bile Esculin Azide Agar w/<br>6 µg/ml Vancomycin                                                                                                                                                                                                                                                                  |
| Similarities          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                         |
| Intended Use          | chromID™ VRE agar is a selective and<br>differential chromogenic medium<br>containing 8 µg/mL of vancomycin, for the<br>qualitative detection of <i>Enterococcus</i><br><i>faecium</i> and <i>E. faecalis</i> showing acquired<br>vancomycin resistance (VRE) in stool<br>specimens. chromID™ VRE agar can be<br>used as an aid to identify, prevent and<br>control VRE colonization in healthcare<br>settings. chromID™ VRE agar is not<br>intended to diagnose VRE infection nor to<br>guide or monitor treatment for infections.<br>Subculture to non-selective media (e.g.<br>trypticase soy agar with 5% sheep blood)<br>is needed for further identification,<br>susceptibility testing and epidemiological<br>typing. | REMEL Bile Esculin Azide Agar w/ 6<br>µg/ml Vancomycin is a solid medium<br>recommended for use in qualitative<br>procedures as a screening method for<br>primary isolation and presumptive<br>identification of vancomycin resistant<br>enterococci (VRE) from surveillance<br>cultures.                               |
| Test method           | Manual                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Manual                                                                                                                                                                                                                                                                                                                  |
| Inoculum              | Direct Specimen                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Direct Specimen                                                                                                                                                                                                                                                                                                         |
| Specimen              | Stool samples                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Urine, stool                                                                                                                                                                                                                                                                                                            |
| Differences           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                         |
| Detection method      | Two chromogenic substrates provide<br>for the direct detection of <i>E. faecium</i><br>and <i>E. faecalis</i> through characteristic<br>colony color.<br>- <i>E. faecium</i> : violet color for β-<br>galactosidase-producing strains,<br>- <i>E. faecalis</i> : blue-to-green color for α-<br>glucosidase-producing strains.                                                                                                                                                                                                                                                                                                                                                                                                | Organisms positive for esculin<br>hydrolysis hydrolyze the glycoside<br>esculin to esculetin and dextrose. The<br>esculetin reacts with the ferric citrate to<br>form a dark brown or black complex,<br>esculetin, which reacts with the ferric<br>ammonium citrate to produce a black-<br>brown complex in the medium. |
| Incubation Conditions | 35-37°C in aerobic conditions, in the<br>dark                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 33-37°C in aerobic conditions or in 5-<br>10% CO2                                                                                                                                                                                                                                                                       |

{2}------------------------------------------------

### Performance

Performance of chrom!D™ VRE was evaluated at four geographically diverse laboratories. Stool specimens were inoculated on chromID™ VRE agar and bile esculin azide agar with 6 μ.g/ml vancomycin (BEAV). Both plates were observed for growth at 24 and 48 hours. Colonies with violet or blue-to-green pigment on chromID™ VRE agar and colonies on BEAV with brown to black pigment diffusing into the medium were identified with the following methods: Gram stain, catalase, VITEK® 2 GP and 16S-500 sequencing. Vancomycin resistance was confirmed by agar dilution.

A total of 1299 stool samples were evaluated with chromID™M VRE agar. Compared to the conventional reference methods described above chromID™ VRE identified 97.1% of the VRE positive samples and 99.7% of the VRE negative samples after 24 hours incubation. chromID™ VRE identified 96.9% of the VRE positive samples and 99.7% of the VRE negative samples after 48 hours incubation.

#### Performance vs. Conventional Methods

#### Table 1

|                                      | Positive<br>% Agreement                    | Negative<br>% Agreement                    |
|--------------------------------------|--------------------------------------------|--------------------------------------------|
| chromID™ VRE agar<br>@24h incubation | 97.1% (334/344)<br>(95% CI = 94.7%, 98.6%) | 99.7% (952/955)<br>(95% CI = 99.1%, 99.9%) |
| chromID™ VRE agar<br>@48h incubation | 96.9% (344/355)<br>(95% CI = 94.5%, 98.4%) | 99.7% (941/944)<br>(95% CI = 99.1%, 99.3%) |

### Table 2

|                               | Positive<br>% Agreement                     | Negative<br>% Agreement                       |
|-------------------------------|---------------------------------------------|-----------------------------------------------|
| BEAV agar @ 24h<br>incubation | 87.2% (300/344 )<br>(95% CI = 83.2%, 90.6%) | 100.0% (955/955).<br>(95% CI = 99.6%, 100.0%) |
| BEAV agar @ 48h<br>incubation | 91.8% (326/355)<br>(95% CI = 88.5%, 94.5%)  | 100.0% (944/944)<br>(95% CI = 99.6%, 100.0%)  |

# Table 3 - Performance vs. VITEK 2 ID: E. faecium

| <i>E. faeclum</i>                     | Positive<br>% Agreement                    | Negative<br>% Agreement                |
|---------------------------------------|--------------------------------------------|----------------------------------------|
| chromID™ VRE agar @<br>24h incubation | 94.3% (316/335)<br>(95% CI = 91.3%, 96.6%) | 11.1% (1/9)*<br>(95% CI = 0.3%, 42.3%) |
| chromID™ VRE agar @<br>48h incubation | 96.7% (324/335)<br>(95% CI = 94.2%, 98.4%) | 0.0% (0/9)*<br>(95% CI = 0.0%, 33.6%)  |

* Of the nine isolates that were not identified as E. faecium by VITEK 2, one isolate was negative on the chromID™ VRE at 24 hours but was false positive after 48 hours. The other eight isolates were false positive on chromID™ VRE at 24 and 48 hours.

{3}------------------------------------------------

### Table 4 - Performance vs. VITEK® 2 ID: E. faecalis

| E. faecalis                           | Positive<br>% Agreement                  | Negative<br>% Agreement |
|---------------------------------------|------------------------------------------|-------------------------|
| chromID™ VRE agar @<br>24h incubation | 87.1% (27/31)<br>(95% CI = 70.2%, 96.4%) | 0/0*                    |
| chromID™ VRE agar @<br>48h incubation | 96.8% (30/31)<br>(95% CI = 83.3%, 99.9%) | 0/0*                    |

All E. faecalls isolates were identified by VITEK 2 however four isolates did not produce the characteristic blue-to-green color at 24 hours. At 48h incubation, three of the four isolates developed the characteristic blue-to-green color.

| Table 5 - Performance vs. Vancomycin MIC: E. faecium |  |  |
|------------------------------------------------------|--|--|
|------------------------------------------------------|--|--|

| E. faecium                            | Positive<br>% Agreement                    | Negative<br>% Agreement                |
|---------------------------------------|--------------------------------------------|----------------------------------------|
| chromID™ VRE agar @<br>24h incubation | 94.4% (321/340)<br>(95% CI = 91.4%, 96.6%) | 25.0% (1/4)*<br>(95% CI = 0.6%, 80.6%) |
| chromID™ VRE agar @<br>48h incubation | 97.1% (330/340)<br>(95% CI = 94.7%, 98.6%) | 25.0% (1/4)*<br>(95% CI = 0.6%, 80.6%) |

* Three of four isolates were false positive as they produced violet colonies on chromID™ VRE but were confirmed to be vancomycin susceptible E. faecium by MIC. The fourth isolate did not produce violet pigmented colonies.

| <i>E. faecalis</i>                    | Positive<br>% Agreement                  | Negative<br>% Agreement               |
|---------------------------------------|------------------------------------------|---------------------------------------|
| chromID™ VRE agar @<br>24h incubation | 86.7% (26/30)<br>(95% CI = 69.3%, 96.3%) | 0.0% (0/1)*<br>(95% CI = 0.0%, 97.5%) |
| chromID™ VRE agar @<br>48h incubation | 96.7% (29/30)<br>(95% CI = 82.8%, 99.9%) | 0.0% (0/1)*<br>(95% CI = 0.0%, 97.5%) |

One sample produced blue to green colonies on chromID TM VRE and was vancomvoin susceptible by MIC.

#### Challenge Testing:

Seventy-five well-characterized challenge strains including vancomycin-resistant and vancomycinsusceptible E. faecalis and E. faecium, as well as Gram-positive microorganisms commonly isolated from stool, were evaluated on chromID™ VRE agar and produced the expected results. One challenge isolate, E. faecalis 13029 (vanB) produced blue-purple colonies after 48 hours incubation. This was due to a rare situation of activity of both enzymes («-glucosidase and ß-galactosidase) occurring in one strain. Both enzymes reacted with the chromogen in the chromID™ VRE agar and this mixture of activity produced the blue-purple color variation.

A second study was performed to evaluate the detection of enterococcal strains with low level vancomycin resistance. Forty-nine enterococci strains characterized by intermediate or low levels of resistance to vancomycin were evaluated. E. faecalis with a low level of vancomycin resistance, such as those carrying the vanB gene are detected on chromID™ VRE medium with their characteristic colony color if their vancomycin MIC is ≥ 4 ug/mL. For a majority of isolates, the characteristic colony coloration is observed after 48 h of incubation. E. casseliflavus strains are inhibited on chromID™ VRE. E. gallinarum

{4}------------------------------------------------

strains are inhibited on chromID™ VRE unless they have an acquired vanA gene. In this case, characteristic violet color is observed after 48 h.

### Interference Study

Commonly used medicinal substances, blood, Cary Blair Transport medium and physiological saline were evaluated for potential interference with the chromogenic reaction of the chromID™ VRE agar. None of the potentially interfering substances tested affected the performance of the chromogenic reaction of the chromID™ VRE medium but decreases in the quantity of growth were observed. Cary Blair Transport medium did not interfere with the chromogenic reaction of the chromID™ VRE agar or reduce the quantity of-growth on-the plate .- In-the-presence-of Preparation-H-Cream-or-Miconazole-7-the-detection-of-certainresistant E. faecalis or E. faecium isolates may be inhibited or delayed after 24 to 48 hours incubation on the chromID™ VRE medium. Other commonly used medicinal substances, blood and physiological saline may result in decreased growth.

#### Reproducibility

Ten reproducibility organisms comprised of a subset of challenge isolates representing Vancomycin susceptible and resistant E. faecalis and E. faecium (both vanA & vanB), as well as quality control reference strains, were evaluated by chromlD™ VRE in triplicate each day for three days at each clinical trial site and produced the expected results. Preliminary results after 24 hours of incubation show 80% reproducibility for this set of organisms. After the chromID™ VRE plates were incubated for the full 48 hours, overall reproducibility was 100% for this set of organisms.

#### Cross Reactivity Study

A cross reactivity study was performed to determine if strains other than vancomycin-resistant enterococci could grow on chromID™ VRE agar and develop violet or blue-to-green colonies. Ninety-nine organisms were included in the cross-reactivity study. Five strains developed violet or blue-to-green pigmented colonies. These included one strain each of Candida albicans, C. tropicalis, Citrobacter freundii, Enterocccus raffinosus and two strains of Klebsiella pneumoniae. Several micro-organisms other than enterococci (including veasts and Gram-negative bacili), may grow and or develop pigmented colonies on chromID™ VRE. On rare occasions colonies may produce violet or blue-to-green pigment.

### Recovery Study

A recovery study was performed to define the lowest number of colony forming units (CFU) of vancomycinresistant enterococci (VRE) that could grow on the chromID™ VRE medium. One strain each of Enterocccus faecalis (VRE) and E. faecium (VRE) were evaluated. A series of 10-fold serial dilutions were prepared in saline and then plated in duplicate onto chromID™ VRE medium. After 48 hours incubation, the number of CFU was counted on each plate. The number of CFU obtained on the two plates from each dilution was averaged. The lowest number of CFU that can be detected on chromID VRE is equivalent to a theoretical 100 CFU per mL of sample,

### Swab Studv

The impact of swabs on the detection of E. faecalis (VRE) with chromID™ VRE medium was evaluated. Rayon swabs and nylon flocked swabs were tested dry and with Amies transport medium. The study evaluated detection of VRE based on contact times of 1, 4 and 18 hours for each swab type both at noom temperature and at 2-8°C. The organism test set included 16 strains of vanA or vanB positive Enterococus faccalis and E. faecium with various vancomycin MIC values. Results indicated that after 1 h and 4 h of contact. there was no significant difference in the sensitivity of detection between dry swabs and those in Amies transport medium. After 18 h of contact at room temperature, the nylon flocked swab with Amies media had the best results followed by the nylon flocked dry swab and rayon swab w/ Amies transport medium. After 18 h of contact at 2-8°C, the nylon flocked swabs with and without Amies media had the best results followed by the rayon swabs with and without Amies transport medium.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" is arranged in a circular fashion around the eagle.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993

HAR 0 2 2010

Nancy Weaver Associate Director, Regulatory Affairs bioMerieux. Inc. 595 Anglum Road Hazelwood, MO 63042

Re: K091025

Trade/Device Name: chromID™ VRE Agar Regulation Number: 21 CFR 866.1700 Regulation Name: Culture Media, Antimicrobial Susceptibility Test, Excluding Mueller Hinton Agar Regulatory Class: Class II Product Code(s): JSO Dated: April 8, 2009 Received: April 10, 2009

Dear Ms. Weaver:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. ·

{6}------------------------------------------------

# Page 2 - Ms. Weaver

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Feraldine M. Poole

ally A. Hoivat. Ph.D. Director, Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

{7}------------------------------------------------

# Indications for Use

510(k) Number (if known): K09 1025

Device Name: chromID™ VRE Agar

Indications For Use:

chromID™ VRE agar is a selective and differential chromogenic medium containing 8 µg/mL of vancomycin, for the qualitative detection of Enterococcus faecium and E. faecalis showing acquired vancomycin resistance (VRE) in stool specimens. chromID™ VRE agar can be used as an aid to identify, prevent and control VRE colonization in healthcare settings. ChromID™ VRE agar is not intended to diagnose VRE infection nor to guide or monitor treatment for infections. Subculture to non-selective media (e.g. trypticase soy agar with 5% sheep blood) is needed for further identification, susceptibility testing and epidemiological typing.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Peraldi L. Pady
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K091025

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JSO/K091025](https://fda.innolitics.com/submissions/MI/subpart-b%E2%80%94diagnostic-devices/JSO/K091025)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).