EPISCREEN

K970357 · Epitope, Inc. · PJD · Mar 10, 1997 · Clinical Chemistry

Device Facts

Record IDK970357
Device NameEPISCREEN
ApplicantEpitope, Inc.
Product CodePJD · Clinical Chemistry
Decision DateMar 10, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1675
Device ClassClass 2

Indications for Use

EpiScreen is intended to collect oral fluid specimens, contain those specimens, and preserve the specimens after collection and during transport from the collection area to the laboratory.

Device Story

EpiScreen device consists of treated absorbent cotton fiber pad on nylon stick and preservative solution in plastic vial. Pad impregnated with salts and gelatin creates hypertonic environment; induces osmotic gradient across buccal/gingival mucosa; enhances flow of mucosal transudate onto cotton fibers. Pad placed between lower gum and cheek for collection period; removed and placed into vial containing preservative to inhibit oral microorganism growth. Vial sealed and transported to laboratory for processing. Used by clinicians/technicians to facilitate oral fluid specimen collection for diagnostic testing.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Components: treated absorbent cotton fiber pad, nylon stick, plastic specimen vial, preservative solution. Principle: osmotic gradient induction via hypertonic pad impregnation. Form factor: manual collection applicator and transport container. No electronic or software components.

Indications for Use

Indicated for the collection, preservation, and transport of oral fluid specimens for laboratory analysis. No specific patient population, age, or gender restrictions are defined.

Regulatory Classification

Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} Epitope, Inc. 8505 SW Creekside Place Beaverton, OR 97008 503-641-6115 EpiScreen® Oral Specimen Collection Device January 29, 1997 # 510(k) Summary **Trade Name:** EpiScreen™ **Common Name:** Oral Specimen Collection Device **MAR 10 1997** 14970357 ## Device Description The EpiScreen device consists of a treated absorbent cotton fiber pad affixed to a nylon stick (Collection Pad) and a preservative solution in a plastic container (Specimen Vial). The Collection Pad is impregnated with a mixture of common salts and gelatin, creating a hypertonic environment which produces an osmotic gradient across the buccal and gingival mucosa. The pad is placed in contact with the gingival mucosa (between the lower gum and cheek) which enhances the flow of mucosal transudate onto the absorptive cotton fibers of the pad. Following the collection period, the Collection Pad is removed from the mouth and placed into a Specimen Vial. The vial contains a preservative solution which serves to inhibit the growth of oral microorganisms recovered on the Collection Pad. The vial is sealed with a plastic cap and transported to a laboratory for processing and testing. ## Intended Use EpiScreen is intended to collect oral fluid specimens, contain those specimens, and preserve the specimens after collection and during transport from the collection area to the laboratory. ## Classification EpiScreen is classified as, or is substantially equivalent to, one or more of the following: - Specimen Container (Class I, 21 CFR §880.6175, General Hospital and Personal Use Devices, General Hospital Panel, Product Code 80FMH); - Specimen Transport and Storage Container (Class I, 21 CFR §880.3250, Hematology and Pathology Devices, Pathology Panel, Product Code 88KDT); - Absorbent Tipped Applicator (Class I, 21 CFR §880.6025, General Hospital and Personal Use Devices, General Hospital Panel, Product Code 80FOR); - Microbiological Specimen Collection and Transport Device (Class I, 21 CFR §866.2900), Immunology and Microbiology Devices, Microbiology Panel, Product Code 83LIO). Like EpiScreen, these devices are used for taking specimens, and for containing those specimens, and preserving (with or without a fixative solution) the viability and integrity of the specimens during transport from the collecting area to the laboratory. EpiScreen 510(k) Summary Epitope, Inc. {1} Epitope, Inc. 8505 SW Creekside Place Beaverton, OR 97008 503-641-6115 EpiScreen® Oral Specimen Collection Device January 29, 1997 # Comparison of Technological Characteristics EpiScreen is substantially equivalent to the following devices: - Saliva Sampler manufactured by Saliva Diagnostics Systems of Vancouver, WA. - Salivette manufactured by Sarstedt, Inc. of Newton, NC. - Saliva Sac manufactured by BioQuant, Inc. of Ann Arbor, MI. Like EpiScreen, the Saliva Sampler, Salivette, and Saliva Sac are intended for collecting an oral fluid specimen, and for containing and transporting that specimen. The following table lists the indications and components for EpiScreen and the three equivalent devices. Note that all four devices share the same indication as well as major components (collection apparatus and transport container). While EpiScreen has different technological characteristics than the predicate devices discussed here, EpiScreen does not raise different questions of safety and efficacy. ## Comparison Table: EpiScreen, Saliva Sampler, Saliva Sac, and Salivette | | EpiScreen | Saliva Sampler | Saliva Sac | Salivette | | --- | --- | --- | --- | --- | | Indication | Collection/transport of oral fluid specimens | Collection/transport of oral fluid specimens | Collection/transport of oral fluid specimens | Collection/transport of oral fluid specimens | | Collection Apparatus | ☑ | ☑ | ☑ | ☑ | | Transport Container | ☑ | ☑ | ☑ | ☑ | | Fixative Solution | ☑ | ☑ | N/A | N/A | | Suspension Insert | N/A | N/A | N/A | ☑ | | Volume Indicator | N/A | ☑ | N/A | N/A | EpiScreen 510(k) Summary Epitope, Inc. {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center – WO66-G609 Silver Spring, MD 20993-0002 Epitope, Inc. Caroline R. Sayre, RAC Regulatory Affairs Supervisor 8505 SW CREEKSIDE PLACE BEAVERTON, OREGON 97008 FEB 06 2015 Re: K970357 Trade/Device Name: EpiScreen™ Oral Specimen Collection Device Regulation Number: 21 CFR 862.1675 Regulation Name: Blood specimen collection device Regulatory Class: II Product Code: PJD Dated: January 29, 1997 Received: January 30, 1997 Dear Ms. Sayre: This letter corrects our previous Substantially Equivalent (SE) letter of March 10, 1997. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act {3} Page 2 – Ms. Sayre or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH’s Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ![img-0.jpeg](img-0.jpeg) Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {4} 510(k) Number (if known): ___________________________ Device Name: EpiScreen™ Oral Specimen Collection Device Indications for Use: EpiScreen is intended for use in the collection, preservation, and transport of oral fluid specimens. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Davon Hansen (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number K920357 Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-96) 3
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...