binx health io CT/NG Assay

K191352 · Binx Health, Inc. · QEP · Aug 8, 2019 · Microbiology

Device Facts

Record IDK191352
Device Namebinx health io CT/NG Assay
ApplicantBinx Health, Inc.
Product CodeQEP · Microbiology
Decision DateAug 8, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3393
Device ClassClass 2

Indications for Use

The binx health io CT/NG Assay, when tested using the binx health io Instrument, is a fully automated, rapid, qualitative test intended for use in point-of-care or clinical laboratory settings for the rapid detection of Chlamydia trachomatis and Neisseria gonorrhoeae DNA in female vaginal swab specimens collected either by a clinician or self-collected by a patient in a clinical setting, to aid in the diagnosis of symptomatic or asymptomatic infection in female patients with Chlamydia trachomatis and/or Neisseria gonorrhoeae.

Device Story

The binx io CT/NG Assay is a point-of-care, fully automated, rapid, qualitative in vitro diagnostic system. It consists of a benchtop instrument, a single-use assay-specific cartridge, a transfer pipet, and a vaginal swab collection kit. The user loads a vaginal swab specimen into the cartridge; the instrument then performs automated sample preparation (lysis, DNA isolation/purification), ultra-rapid PCR amplification of target DNA, and electrochemical detection. The instrument uses air pressure to control fluid movement within the cartridge. Results are displayed on the instrument's GUI in approximately 30 minutes. The system includes an internal process control (IPC) to validate results. The output is a qualitative 'Detected' or 'Not Detected' result for CT and NG, which aids clinicians in diagnosing infections. The device is designed for use in clinical settings by point-of-care personnel.

Clinical Evidence

Prospective clinical study at 9 U.S. sites (1,523 evaluable female subjects). Compared binx io results to a Composite Infected Status (CIS) derived from three FDA-cleared molecular assays. CT sensitivity 96.1% (91.3-98.3%), specificity 99.1% (98.4-99.5%). NG sensitivity 100.0% (92.4-100.0%), specificity 99.9% (99.5-100.0%). Study included both clinician-collected and self-collected vaginal swabs across symptomatic and asymptomatic populations.

Technological Characteristics

The system uses automated multiplex PCR with electrochemical detection. Reagents are contained in single-use cartridges with liquid blister packs and dried reagents. The instrument uses air pressure for fluidic control. It is a benchtop, standalone system. No specific material standards are cited. The assay targets genomic DNA for both CT and NG.

Indications for Use

Indicated for symptomatic or asymptomatic female patients to aid in the diagnosis of Chlamydia trachomatis and Neisseria gonorrhoeae infection using vaginal swab specimens.

Regulatory Classification

Identification

A device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s) is an in vitro diagnostic device intended for the detection and identification of nucleic acids from non-viral microorganism(s) and their associated resistance markers in clinical specimens collected from patients suspected of sexually transmitted infections. The device is intended to aid in the diagnosis of non-viral sexually transmitted infections in conjunction with other clinical and laboratory data. These devices do not provide confirmation of antibiotic susceptibility since mechanisms of resistance may exist that are not detected by the device.

Special Controls

A device to detect nucleic acids from non-viral microorganism(s) causing sexually transmitted infections and associated resistance marker(s) must comply with the following special controls: (1) The intended use for the 21 CFR 809.10 labeling must include a detailed description of targets the device detects, the results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended. (2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device: alternatively, the sample collection device must be cleared in a premarket submission as a part of this device. (3) The 21 CFR 809.10(b) labeling must include: (i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens; (ii) Detailed discussion of the performance characteristics of the device for all claimed specimen types based on analytical studies, including, but not limited to. Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, with-in lab precision, and reproducibility, as appropriate; (iii) Detailed descriptions of the test procedure, the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing. (iv) Limiting statements indicating that: (A)a negative test result does not preclude the possibility of infection; (B) the test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician; (C) reliable results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe procedures in any one of these steps can lead to incorrect results; and (D)if appropriate (e.g., recommended by CDC, by current well-accepted clinical guidelines, or by published peer reviewed research), that the clinical performance is inferior in a specific clinical subpopulation or for a specific claimed specimen type. (v) If the device is intended to detect antimicrobial resistance markers, limiting statements, as appropriate, indicating that: (A)negative results for claimed resistance markers do not indicate susceptibility of detected microorganisms, as resistance markers not measured by the assay or other potential mechanisms of antibiotic resistance may be present; (B) detection of resistance markers cannot be definitively linked to specific microorganisms and the source of a detected resistance marker may be an organism not detected by the assay, including colonizing flora; (C) detection of antibiotic resistance markers may not correlate with phenotypic gene expression; and (D) therapeutic failure or success cannot be determined based on the assay results, since nucleic acid may persist following appropriate antimicrobial therapy. (4) Design verification and validation must include: (i) Detailed device description documentation, including, but not limited to, methodology from obtaining sample to result, design of primer/probe sequences, rationale for target sequence selection, and computational path from collected raw data to reported result (e.g., how collected raw signals are converted into a reported result). (ii) Detailed documentation of analytical studies including but not limited to, Limit of Detection, inclusivity, cross-reactivity, microbial interference, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, with-in lab precision, and reproducibility, as appropriate. (iii) Detailed documentation and performance results from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, when determined to be appropriate by FDA, additional characterized clinical samples. The study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained from FDA accepted comparator methods. Documentation from the clinical studies must include the clinical study protocol (including a predefined statistical analysis plan) study report, testing results, and results of all statistical analyses. (iv) A detailed description of the impact of any software, including, but not limited to, software applications and hardware-based devices that incorporate software, on the device's functions.

*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use for the labeling required under § 809.10 of this chapter must include a detailed description of targets the device detects, the results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended. (2) Any sample collection device used must be FDA-cleared, -approved, or -classified as 510(k) exempt (standalone or as part of a test system) for the collection of specimen types claimed by this device; alternatively, the sample collection device must be cleared in a premarket submission as a part of this device. (3) The labeling required under § 809.10(b) of this chapter must include: (i) A detailed device description, including reagents, instruments, ancillary materials, all control elements, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens; (ii) Detailed discussion of the performance characteristics of the device for all claimed specimen types based on analytical studies, including Limit of Detection, inclusivity, cross-reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate; (iii) Detailed descriptions of the test procedure, the interpretation of test results for clinical specimens, and acceptance criteria for any quality control testing; (iv) Limiting statements indicating that: (A) A negative test result does not preclude the possibility of infection; (B) The test results should be interpreted in conjunction with other clinical and laboratory data available to the clinician; (C) Reliable results are dependent on adequate specimen collection, transport, storage, and processing. Failure to observe proper procedures in any one of these steps can lead to incorrect results; and (D) If appropriate ( *e.g.,* recommended by the Centers for Disease Control and Prevention, by current well-accepted clinical guidelines, or by published peer reviewed research), that the clinical performance is inferior in a specific clinical subpopulation or for a specific claimed specimen type; and(v) If the device is intended to detect antimicrobial resistance markers, limiting statements, as appropriate, indicating that: (A) Negative results for claimed resistance markers do not indicate susceptibility of detected microorganisms, as resistance markers not measured by the assay or other potential mechanisms of antibiotic resistance may be present; (B) Detection of resistance markers cannot be definitively linked to specific microorganisms and the source of a detected resistance marker may be an organism not detected by the assay, including colonizing flora; (C) Detection of antibiotic resistance markers may not correlate with phenotypic gene expression; and (D) Therapeutic failure or success cannot be determined based on the assay results, since nucleic acid may persist following appropriate antimicrobial therapy. (4) Design verification and validation must include: (i) Detailed device description documentation, including methodology from obtaining sample to result, design of primer/probe sequences, rationale for target sequence selection, and computational path from collected raw data to reported result ( *e.g.,* how collected raw signals are converted into a reported result).(ii) Detailed documentation of analytical studies, including, Limit of Detection, inclusivity, cross-reactivity, microbial interference, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, within lab precision, and reproducibility, as appropriate. (iii) Detailed documentation and performance results from a clinical study that includes prospective (sequential) samples for each claimed specimen type and, when determined to be appropriate by FDA, additional characterized clinical samples. The study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained from FDA accepted comparator methods. Documentation from the clinical studies must include the clinical study protocol (including a predefined statistical analysis plan) study report, testing results, and results of all statistical analyses. (iv) A detailed description of the impact of any software, including software applications and hardware-based devices that incorporate software, on the device's functions.

Predicate Devices

Submission Summary (Full Text)

{1} Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ## I Background Information: A 510(k) Number K191352 B Applicant binx health, Inc. C Proprietary and Established Names binx health io CT/NG Assay D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | QEP, LSL, MKZ, NSU | Class II | 21 CFR 866.3393 - | MI - Microbiology | ## II Submission/Device Overview: A Purpose for Submission: Notification of intent to market a new device. B Measurand: 1. Chlamydia trachomatis genomic DNA 2. Neisseria gonorrhoeae genomic DNA C Type of Test: Nucleic acid amplification assay, qualitative. K191352 - Page 1 of 25 {2} III Intended Use/Indications for Use: A Intended Use(s): See Indications for Use below. B Indication(s) for Use: The binx health io CT/NG Assay, when tested using the binx health io Instrument, is a fully automated, rapid, qualitative test intended for use in point-of-care or clinical laboratory settings for the rapid detection of Chlamydia trachomatis and Neisseria gonorrhoeae DNA in female vaginal swab specimens collected either by a clinician or self-collected by a patient in a clinical setting, to aid in the diagnosis of symptomatic or asymptomatic infection in female patients with Chlamydia trachomatis and/or Neisseria gonorrhoeae. C Special Conditions for Use Statement(s): Rx - For Prescription Use Only D Special Instrument Requirements: binx health io Instrument IV Device/System Characteristics: A Device Description: The binx health io CT/NG Assay is a qualitative in vitro diagnostic system consisting of the following: 1. The binx health io Instrument for running the test Cartridge, 2. The binx health io CT/NG Cartridge; each Cartridge contains all the necessary reagents to test one patient sample, 3. A single-use, fixed-volume transfer pipet for transferring the sample to the Cartridge, A female Vaginal Swab Specimen Collection Kit consisting of a sterile flocked swab and a sample collection tube containing preservation medium, is provided separately. The Cartridges are packaged in cartons of 10 single use units. The assay results are displayed on the instrument screen. K191352 - Page 2 of 25 {3} | Assay Result | Interpretation of Assay Results | | --- | --- | | CT Not Detected | Chlamydia trachomatis target DNA **was not detected** in the patient specimen. The IPC passed specification. This is a valid result | | CT Detected | Chlamydia trachomatis target DNA **was detected** in the patient specimen. This is a valid result. | | NG Not Detected | At least one of the DNA targets for Neisseria gonorrhoeae **was not detected** in the patient specimen. The IPC passed specification. This is a valid result. | | NG Detected | Both DNA targets for Neisseria gonorrhoeae **were detected** in the patient specimen. This is a valid result. | | Test Invalid | The presence or absence of the DNA targets for Chlamydia trachomatis and/or Neisseria gonorrhoeae could not be ascertained in the patient specimen. If a CT or NG Not Detected result is obtained, and the IPC was outside the acceptable range, Test Invalid indicates a failure in the Assay process. The Assay should be repeated using the same patient specimen. | | User Aborted | A user cancelled the Assay. No result is given. | | Error | An internal fault occurred that terminated the Assay before it finished. No result is given. | B Principle of Operation: The binx health io CT/NG Assay detects and amplifies DNA from CT and NG organisms that may be present in female vaginal swabs. The binx io Cartridge is assay-specific and is intended for a single-use on a single patient. All reagents are contained in the Cartridge as a combination of liquid reagents in blister packs and dried reagents. The Cartridge contains reagents for three main assay steps: (i) sample preparation to isolate and purify target DNA, (ii) target DNA amplification by PCR, and (iii) a proprietary electrochemical detection to identify the presence of amplified DNA. The assay uses a vaginal swab collected in a preservation medium (eNAT). After the sample (0.5 mL of the swab eluate) is added to the port of the Cartridge using the included sample transfer pipet, the Cartridge is inserted into the Instrument and the assay begins. Each Cartridge label contains a barcode that includes the test type, batch information, and expiration date, which is automatically scanned by the Instrument. At the start of the assay, the instrument initiates the K191352 - Page 3 of 25 {4} release of reagents and, after extraction and purification, the sample is divided between two PCR chambers allowing for CT and NG DNA amplification by thermal cycling to be run separately. While only one genomic DNA target is used for the detection of CT, two genomic DNA targets are amplified for the NG and both must be present to generate a positive result for NG. In the next step, complementary electrochemically-labeled DNA probes hybridize to the amplified DNA followed by enzymatic digestion of the double stranded DNA-probe complexes releasing the electrochemical label. After application of voltage to a carbon electrode and oxidation of the released label, a current is generated which is measured to indicate the presence of CT and/or NG DNA. In negative specimens, the probes remain as single-stranded DNA and cannot be digested by the enzyme; no release of the electrochemical label occurs and no current is generated. C Instrument Description Information: 1. Instrument Name: binx health io Instrument 2. Specimen Identification: Manual entry of the specimen ID on the touchscreen, or Handheld barcode scanner, or Automatic assignment of unique ID. 3. Specimen Sampling and Handling: Requires binx health io Vaginal Swab Specimen Collection Kit. The specimen (swab eluate) is loaded onto the io test Cartridge which is inserted into the io Instrument. The sample is processed automatically by the Instrument. 4. Calibration: No calibration is required. 5. Quality Control: The assay incorporates a positive Internal Process Control (IPC) which is processed along with each patient sample, from the sample DNA extraction and purification through the detection. If the IPC measurement is outside of the specified range, the instrument will return an "Assay Invalid" message. A negative result for CT and/or NG will only be returned if the IPC measurement is within the acceptable range. Where CT and/or NG DNA is detected, the IPC will be disregarded as the detection of the CT and/or NG DNA target will verify that the assay functions as expected. External controls are not provided with the assay. The package insert recommends a commercial source of external controls. K191352 - Page 4 of 25 {5} V Substantial Equivalence Information: A Predicate Device Name(s): BD ProbeTec Chlamydia trachomatis (CT) Qx CT Amplified DNA Assay BD ProbeTec Neisseria gonorrhoeae (GC) Qx Amplified DNA Assay B Predicate 510(k) Number(s): K091724 K091730 C Comparison with Predicate(s): | Device & Predicate Device(s): | K191352 | K091724 | K091730 | | --- | --- | --- | --- | | Device Trade Name | binx health io CT/NG Assay | BD ProbeTec Qx CT | BD ProbeTec Qx GC | | General Device Characteristic Similarities | | | | | Intended Use/Indications for Use | The binx health io CT/NG Assay, when tested using the binx health io Instrument, is a fully automated, rapid, qualitative test intended for use in point-of-care or clinical laboratory settings for the rapid detection of Chlamydia trachomatis and Neisseria gonorrhoeae DNA in female vaginal swab specimens collected either by a clinician or self-collected by a patient in a clinical setting, to aid in the diagnosis of symptomatic or asymptomatic infection in female patients with Chlamydia trachomatis and/or Neisseria gonorrhoeae. | The BD ProbeTec Chlamydia trachomatis Qx Amplified DNA Assay, when tested with either the BD Viper™ System in Extracted Mode or the BD Viper LT System, uses Strand Displacement Amplification technology for the direct, qualitative detection of Chlamydia trachomatis DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens (both UPT and neat). The assay is also intended for use with gynecological specimens collected in BD SurePath™ | The BD ProbeTec™ Neisseria gonorrhoeae Qx Amplified DNA Assay, when tested with either the BD Viper™ System in Extracted Mode or the BD Viper LT System, uses Strand Displacement Amplification technology for the direct, qualitative detection of Neisseria gonorrhoeae DNA in clinician-collected female endocervical and male urethral swab specimens, patient-collected vaginal swab specimens (in a clinical setting), and male and female urine specimens (both UPT and Neat). The assay is also intended for use with gynecological specimens collected in BD SurePath™ | K191352 - Page 5 of 25 {6} K191352 - Page 6 of 25 | | | Preservative Fluid or PreservCyt™ Solution using an aliquot that is removed prior to processing for either the BD SurePath or ThinPrep™ Pap test. The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of chlamydial urogenital disease. | Preservative Fluid or PreservCyt™ Solution using an aliquot that is removed prior to processing for either the BD SurePath or ThinPrep™ Pap test. The assay is indicated for use with asymptomatic and symptomatic individuals to aid in the diagnosis of gonococcal urogenital disease. | | --- | --- | --- | --- | | Technology | Automated multiplex PCR amplification with electrochemical detection | Automated multiplex strand displacement amplification with fluorescent detection. | Automated multiplex strand displacement amplification with fluorescent detection. | | Instrument | binx health io Instrument | BD Viper | BD Viper | | Assay Results | Qualitative | Qualitative | Qualitative | | Analyte/Assay Targets | CT Genomic DNA NG Genomic DNA | CT Cryptic Plasmid DNA | NG Genomic DNA | | Nucleic Acid extraction | Yes | Yes | Yes | | Specimen | Vaginal Swab (clinician collected and self-collected in clinical setting) | • Vaginal Swab (clinician collected and self-collected in clinical setting) • Endocervical Swab • Male urethral swab Male urine • Female urine (both UPT and neat). • LBC specimens collected in BD SurePath™ Preservative Fluid or PreservCyt™ Solution using an aliquot that is removed prior to processing for either the BD SurePath or ThinPrep™ Pap test. | • Vaginal Swab (clinician collected and self-collected in clinical setting) • Endocervical Swab • Male urethral swab Male urine • Female urine (both UPT and neat). • LBC specimens collected in BD SurePath™ Preservative Fluid or PreservCyt™ Solution using an aliquot that is removed prior to processing for either the BD SurePath or ThinPrep™ Pap test. | | Collection Kit | binx health io Vaginal Swab Specimen | • Vaginal Specimen | • Vaginal Specimen | {7} K191352 - Page 7 of 25 VI Standards/Guidance Documents Referenced: The following voluntary standards were utilized in part in the studies carried out for the preparation of this 510(k). | Standards organization | Standard No. | Standard Title, | | --- | --- | --- | | CLSI | MM03-A2 | Molecular Diagnostic Methods for Infectious Diseases. | | CLSI | EP07-A2 | Interference Testing in Clinical Chemistry. | | CLSI | EP17-A2-17A | Protocols for Determination of Limits of Detection and Limits of Quantitation. | | CLSI | EP12-A2 | User Protocol for Evaluation of Qualitative Assay Performance | | ISTA | 3A | General Simulation Performance Tests | | ISO | 14155 | Clinical investigation of medical devices for human participants — Good clinical practice | | IEC | 60601-1-2 | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests | VII Performance Characteristics: A Analytical Performance: {8} # 1. Precision/Reproducibility: The reproducibility study was conducted at three POC testing sites, i.e., clinics where the testing was performed outside of a central CLIA certified laboratory, by non-laboratorians representing healthcare professionals that may be encountered at such sites. None of the users were trained in general laboratory procedures but were trained in the use of the binx io System. The study was conducted over 7 days, by 6 operators (2 at each site), with two io Readers at each site, using one lot of Cartridges. The reproducibility test panel consisted of 11 panel members prepared in pooled negative vaginal matrix collected in eNAT. Each test panel was run twice a day by each operator to generate 28 test results for each panel member at each of the three sites, for a total of 84 test results for each panel member across the three testing sites (924 Cartridges tested in total). The test samples were prepared to include three concentration levels for each organism, in combination and one sample negative for both organisms, i.e., unspiked negative swab matrix. Reproducibility Test Panel | Sample | Organism Concentration | CT | NG | | --- | --- | --- | --- | | LOW POS | Expected to test positive approximately 95% of the time | 1x LoD | 1x LoD | | MOD | Expected to test positive approximately 100% of the time | 3x LoD | 3x LoD | | HIGH | Expected to test positive 100% of the time (at a high end of medically relevant concentration) | 2.26 x 105GE/mL | 1.18 x 106GE/mL | | NEG | Expected to test negative approximately 95% of the time | No target | No target | The composition of the reproducibility test panel along with results of the testing (\% agreement with expected results) at each testing site and overall is shown below. Reproducibility of the binx health io CT/NG Test System | Sample | Concentration CT/NG | Analyte | Site 1 | | Site 2 | | Site 3 | | Overall Agreement | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | No. of Correct/Total No. Tested | % Agreement | No. of Correct/Total No. Tested | % Agreement | No. of Correct/Total No. Tested | % Agreement | No. of Correct/Total No. Tested | % Agreement | 95% CI | | 1 | CT: NEG | CT | 26/28 | 92.9 | 28/28 | 100.0 | 26/28 | 92.9 | 80/84 | 95.2 | 88.4 to 98.1 | | | NG: HIGH | NG | 28/28 | 100.0 | 28/28 | 100.0 | 28/28 | 100.0 | 84/84 | 100.0 | 95.6 to 100 | | 2 | CT: NEG | CT | 25/28 | 89.3 | 27/28 | 96.4 | 26/28 | 92.9 | 78/84 | 92.9 | 85.3 to 96.7 | | | NG: MOD | NG | 28/28 | 100.0 | 28/28 | 100.0 | 28/28 | 100.0 | 84/84 | 100.0 | 95.6 to 100 | | 3 | CT: NEG | CT | 28/28 | 100.0 | 28/28 | 100.0 | 26/28 | 92.9 | 82/84 | 97.6 | 91.7 to 99.3 | | | NG: LOD | NG | 28/28 | 100.0 | 28/28 | 100.0 | 27/28 | 96.4 | 83/84 | 98.8 | 93.6 to 99.8 | | 4 | CT: HIGH | CT | 28/28 | 100.0 | 28/28 | 100.0 | 28/28 | 100.0 | 84/84 | 100.0 | 95.6 to 100 | | | NG: NEG | NG | 28/28 | 100.0 | 28/28 | 100.0 | 28/28 | 100.0 | 84/84 | 100.0 | 95.6 to 100 | | 5 | CT: MOD | CT | 28/28 | 100.0 | 28/28 | 100.0 | 28/28 | 100.0 | 84/84 | 100.0 | 95.6 to 100 | K191352 - Page 8 of 25 {9} K191352 - Page 9 of 25 | | NG: NEG | NG | 28/28 | 100.0 | 28/28 | 100.0 | 28/28 | 100.0 | 84/84 | 100.0 | 95.6 to 100 | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | 6 | CT: LOD | CT | 27/28 | 96.4 | 26/28 | 92.9 | 28/28 | 100.0 | 81/84 | 96.4 | 90.0 to 98.8 | | | NG: NEG | NG | 28/28 | 100.0 | 28/28 | 100.0 | 28/28 | 100.0 | 84/84 | 100.0 | 95.6 to 100 | | 7 | CT: HIGH | CT | 28/28 | 100.0 | 28/28 | 100.0 | 28/28 | 100.0 | 84/84 | 100.0 | 95.6 to 100 | | | NG: HIGH | NG | 28/28 | 100.0 | 28/28 | 100.0 | 28/28 | 100.0 | 84/84 | 100.0 | 95.6 to 100 | | 8 | CT: HIGH | CT | 28/28 | 100.0 | 28/28 | 100.0 | 28/28 | 100.0 | 84/84 | 100.0 | 95.6 to 100 | | | NG: LOD | NG | 28/28 | 100.0 | 27/28 | 96.4 | 28/28 | 100.0 | 83/84 | 98.8 | 93.6 to 99.8 | | 9 | CT: LOD | CT | 28/28 | 100.0 | 27/28 | 96.4 | 27/28 | 96.4 | 82/84 | 97.6 | 91.7 to 99.3 | | | NG: HIGH | NG | 28/28 | 100.0 | 28/28 | 100.0 | 28/28 | 100.0 | 84/84 | 100.0 | 95.6 to 100 | | 10 | CT: LOD | CT | 28/28 | 100.0 | 26/28 | 92.9 | 28/28 | 100.0 | 82/84 | 97.6 | 91.7 to 99.3 | | | NG: LOD | NG | 28/28 | 100.0 | 28/28 | 100.0 | 28/28 | 100.0 | 84/84 | 100.0 | 95.6 to 100 | | 11 | CT: NEG | CT | 26/28 | 92.9 | 28/28 | 100.0 | 26/28 | 92.9 | 80/84 | 95.2 | 88.4 to 98.1 | | | NG: NEG | NG | 28/28 | 100.0 | 28/28 | 100.0 | 28/28 | 100.0 | 84/84 | 100.0 | 95.6 to 100 | *For the Negative panel member, the number of “Detected” samples indicates the number of correct (negative) results. 2. Linearity: Not applicable; this is a qualitative assay; 3. Detection Limit: The limit of detection (LoD) of the binx health io CT/NG Assay was evaluated using genomic DNA extracted from whole CT and NG cells. Two CT serovars (E and F) and two strains of NG were included in the study. The extracted CT and NG DNA was quantified using qPCR against a reference standard in genome equivalents (GE)/mL. The quantified stocks were diluted in pooled vaginal matrix using a range of concentrations (100 to 756 GE/mL for CT and 60 to 246 GE/mL for NG). The study was conducted using two distinct lots of Cartridges, testing each dilution with the binx health io CT/NG Assay in at least 20 replicates. At least five different concentrations (for each of the four organisms and for each of the two lots of Cartridges), covering the range around the putative LoD (0.01-99% detection rates) were tested. The LoDs for CT and for NG were determined by probit regression analysis to identify the concentration level that demonstrated a detection rate of 95%. The determined LoD was then confirmed by testing two additional preparations of each CT serovar and each NG strain using two lots of Cartridges (for a total of 40 Cartridges per serovar/strain per Cartridge lot). The results from the LoD study are summarized in the table below. Summary of the LoD Study Results | Organism | Cartridge Lot | Pre-determined LoD (GE/mL) | Verification Study Results | | | --- | --- | --- | --- | --- | | | | | POS/No. Tested | % Detection | | CT serovar E (ATCC-VR-348B) | Lot 1 | 407.4 | 40/40 | 100% | | | Lot 2 | 309.6 | 39/40 | 97.5% | | CT serovar F (ATCC-VR-346) | Lot 1 | 755.5* | 40/40 | 100% | | | Lot 2 | 734.0 | 40/40 | 100% | | | Lot 1 | 245.6** | 40/40 | 100% | {10} The final claimed LoD concentrations are shown below in GE/mL; the concentrations are also shown in IFU/ml or CFU/mL, as applicable, based on the concentrations stated on the Certificates of Analysis for the stocks used in the study. LoD for the binx health io CT/NG Assay in Vaginal Swab Samples | Organism | Serovar/Strain | GE/mL | IFU/mL | CFU/mL | | --- | --- | --- | --- | --- | | C. trachomatis | Serovar E (ATCC-VR-348B) | 407.4 | 5.6 | N/A | | | Serovar F (ATCC-VR-346) | 755.6 | 0.3 | N/A | | N. gonorrhoeae | Strain 1 (ATCC 49226) | 245.6 | N/A | 2.1 | | | Strain 2 (ATCC 700825) | 206.1 | N/A | 2.5 | ## 4. Analytical Reactivity (Inclusivity) The Assay's ability to detect additional 14 CT serovars, including the nvCT (Swedish variant) was evaluated by spiking each of the serovars (quantified in GE/mL) into vaginal matrix at 1x LoD and 0.5x LoD concentrations. Each dilution was tested with the binx health io CT/NG Assay in 20 replicates, using two lots of Cartridges. While four CT serovars (A, I, L1, and L3) were detected at the claimed LoD (755.5 GE/mL), the remaining 10 serovars were detected at 0.5x LoD (377.8 GE/mL). The summary of the inclusivity testing for CT is presented below. CT Serovars-Limit of Detection | CT Serovar | No. Positive/No. Tested at 1x LoD (755.5 GE/mL) | No. Positive/No. Tested at 0.5x LoD (377.8 GE/mL) | Claimed LoD (GE/mL) | | --- | --- | --- | --- | | A | 20/20 | 18/20 | 755.5 | | B | 20/20 | 19/20 | 377.8 | | Ba | 20/20 | 20/20 | 377.8 | | C | 20/20 | 20/20 | 377.8 | | D | 20/20 | 20/20 | 377.8 | | G | 20/20 | 20/20 | 377.8 | | H | 20/20 | 19/20 | 377.8 | | I | 20/20 | 18/20 | 755.5 | | J | 20/20 | 19/20 | 377.8 | | K | 20/20 | 20/20 | 377.8 | | L1 | 20/20 | 17/20 | 755.5 | | L2 | 20/20 | 20/20 | 377.8 | | L3 | 19/20 | 18/20 | 755.5 | | NvCT | 19/20 | 20/20 | 377.8 | K191352 - Page 10 of 25 {11} The assay's ability to detect multiple NG strains was evaluated by testing 30 unique NG strains, including two fluoroquinolone resistant isolates (quantified in genome copies by qPCR and spiked into vaginal matrix). At least 3 replicates of each strain were tested at 1x LoD concentration (245.6 GE/mL). If any of the 3 replicates yielded negative results, additional testing was performed until a concentration was reached that generated $\geq 19/20$ positive results. Of the 30 NG strains tested, 16 strains tested positive at 1x LoD in the first three replicates; the remaining 14 strains were tested at 5x LoD (1228 GE/mL) generating at least 19/20 positive results. For the strains detected in 3/3 replicates tested, the concentration is expressed as Reported Detectable Level (RDL); for the strains that were detected in at least 19/20 replicates, the concentration is expressed as LoD (Limit of Detection). The summary of the inclusivity testing for NG is presented below. Reactivity with NG Strains | NG Strain | No. Detected at 1x LOD (245.6 GE/mL) | No. Detected at 5x LOD (1228.0 GE/mL) | Detectable Concentration | | | --- | --- | --- | --- | --- | | | | | RDL*/LoD | GE/mL | | Alabama 201301 #06 | 7/9 | 20/20 | LoD | 1228.0 | | Alabama 201310 #01 | 6/9 | 20/20 | LoD | 1228.0 | | California 201302 #03 | 3/3 | N/A | RDL | 245.6 | | California 201302 #06 | 3/3 | N/A | RDL | 245.6 | | California 201304 #01 | 5/9 | 20/20 | LoD | 1228.0 | | California 201307 #02 | 7/9 | 20/20 | LoD | 1228.0 | | California 201301 #10 | 3/3 | N/A | RDL | 245.6 | | California 201301 #12 | 7/9 | 20/20 | LoD | 1228.0 | | California 201302 #10 | 6/9 | 20/20 | LoD | 1228.0 | | California 201303 #14 | 1/3 | 20/20 | LoD | 1228.0 | | California 201310 #04 | 3/3 | N/A | RDL | 245.6 | | Maryland 201301 #01 | 3/3 | N/A | RDL | 245.6 | | Maryland 201301 #02 | 17/20 | 20/20 | LoD | 1228.0 | | Alabama 201310 #05 | 1/3 | 20/20 | LoD | 1228.0 | | Maryland 201309 #21 | 1/3 | 19/20 | LoD | 1228.0 | | Maryland 201311 #16 | 6/9 | 20/20 | LoD | 1228.0 | | North Carolina 201302 #10 | 3/3 | N/A | RDL | 245.6 | | Pennsylvania 201301 #10 | 3/3 | N/A | RDL | 245.6 | | Pennsylvania 201301 #25 | 3/3 | N/A | RDL | 245.6 | | Pennsylvania 201312 #28 | 3/3 | N/A | RDL | 245.6 | | Virginia 201304 #01 | 7/9 | 20/20 | LoD | 1228.0 | | Virginia 201305 #01 | 3/3 | N/A | RDL | 245.6 | | ATCC 19424 | 17/20 | 20/20 | LoD | 1228.0 | | ATCC 31426 | 3/3 | N/A | RDL | 245.6 | | ATCC 43070 | 3/3 | N/A | RDL | 245.6 | | ATCC 49498 | 3/3 | N/A | RDL | 245.6 | | UK 8454 R1* | 3/3 | N/A | RDL | 245.6 | | UK 9155 R2* | 3/3 | N/A | RDL | 245.6 | | UK 9156 S1** | 1/3 | 20/20 | LoD | 1228.0 | K191352 - Page 11 of 25 {12} ATCC 19999 3/3 N/A RDL 245.6 RDL: Reported detectable level tested, 3/3 replicates tested positive LoD: Limit of Detection level, ≥19/20 replicates tested positive *Fluoroquinolone resistant mutant strain **Fluoroquinolone sensitive mutant strain # 5. Analytical Specificity/Interference: ## Cross-reactivity with other Microorganisms The analytical specificity of the binx health io CT/NG Assay was evaluated by spiking pooled vaginal swabs in eNAT buffer with relevant bacteria, fungi, protozoa, or viral organisms that may be present at the sampling site, such as commensal bacteria that may come into contact with the sampling device or genetically closely related species to *Chlamydia trachomatis* or *Neisseria gonorrhoeae*. The potential for cross-reactivity was evaluated with a panel of microorganisms, using cultured organisms at a concentration of 1x 10⁶ CFU/mL for bacteria, or 1x 10⁵ PFU/mL for viruses, or genome equivalents at 2 ng/mL of genomic DNA (≥1x 10⁵ GE/mL). The panel also included *H. sapiens* DNA at 5.7 x 10² GE/mL. Each organism was tested in three replicates. All tests generated expected results, i.e., negative for CT and NG in all 3 replicates, except for one organism, *N. sicca*, which tested positive for CT in one of the three replicates. Additional 17 replicates of the organism were tested, with no additional positive results (1/20). This organism is listed in the package insert as a possible cross-reactant for CT. To demonstrate that the two target organisms do not cross-react with each other, one NG strain and one CT strain (serovar F) were included in the study, generating expected results, i.e., samples containing *C. trachomatis* generated negative results for NG and positive results for CT, and samples containing *N. gonorrhoeae* generated negative results for CT and positive results for NG, in all three replicates. Two species, *Megasphaera*-type 1 and BVAB-2, were evaluated using *in silico* analysis of the genetic targets used in the binx health io CT/NG Assay against the published genome sequences for these organisms. *In silico* analysis demonstrated that neither of these organisms would be detected by the binx health io CT/NG Assay. The table below lists all the organisms evaluated for cross-reactivity with the binx health io CT/NG Assay. Microorganisms Evaluated for Cross-reactivity in the binx health io CT/NG Assay | Organism | Input/mL | Organism | Input/mL | | --- | --- | --- | --- | | Atopobium vaginae* | 1.3x10⁶ GE/mL | Mobiluncus mulieris* | 7.6 x 10⁵ GE/mL | | Bacteriodes fragilis* | 3.5x10⁵ GE/mL | Moraxella lacunata | 1 x 10⁶ CFU/mL | | Bacteriodes ureolyticus* | 1.1x10⁶ GE/mL | Mycoplasma genitalium* | 3.2 x 10⁶ GE/mL | K191352 - Page 12 of 25 {13} K191352 - Page 13 of 25 | Bifidobacterium longum* | 8.2x10^{5} GE/mL | Mycoplasma hominis* | 2.8 x 10^{6} GE/mL | | --- | --- | --- | --- | | BVAB-2 | in silico | Neisseria meningitidis Serogroup A* | 8.2 x 10^{5} GE/mL | | Candida albicans | 1 x 10^{6} CFU/mL | Neisseria meningitidis Serogroup B* | 8.2 x 10^{5} GE/mL | | Candida glabrata | 1 x10^{6} CFU/mL | Neisseria meningitidis Serogroup C* | 8.2 x 10^{5} GE/mL | | Candida parapsilosis | 1 x10^{6} CFU/mL | Neisseria meningitidis Serogroup C* | 8.2x10^{5} GE/mL | | Chlamydia pneumoniae* | 1.5 x10^{6} GE/mL | Neisseria meningitidis Serogroup W135* | 8.2x10^{5} GE/mL | | Chlamydia psittaci* | 1.6 x10^{6} GE/mL | Neisseria meningitidis Serogroup Y* | 8.2x10^{5} GE/mL | | Chlamydia trachomatis* | 2.3 x10^{7} GE/mL | Neisseria cinerea | 1 x 10^{6} CFU/mL | | Clostridium perfringens* | 5.7x10^{5} GE/mL | Neisseria denitrificans | 1 x 10^{6} CFU/mL | | Corynebacterium genitalium | 1 x 10^{6} CFU/mL | Neisseria elongata (4) | 1 x 10^{6} CFU/mL | | Corynebacterium xerosis | 1 x 10^{6} CFU/mL | Neisseria gonorrhoeae* | 1.2x10^{8} GE/mL | | Enterococcus faecalis | 1 x 10^{6} CFU/mL | Neisseria flava | 1 x 10^{6} CFU/mL | | Escherichia coli | 1 x 10^{6} CFU/mL | Neisseria flavescens (3) | 1 x 10^{6} CFU/mL | | Gardnerella vaginalis* | 1.1x10^{6} GE/mL | Neisseria lactamica (3) | 1 x 10^{6} CFU/mL | | Haemophilus ducreyi* | 1.1x10^{6} GE/mL | Neisseria mucosa (4) | 1 x 10^{6} CFU/mL | | Herpes simplex virus 1* | 1.2 x 10^{7} GE/mL | Neisseria perflava (2) | 1 x 10^{6} CFU/mL | | Herpes simplex virus 2 | 1.2 x 10^{7} GE/mL | Neisseria polysaccharea | 1 x 10^{6} CFU/mL | | Homo sapiens* | 5.7x10^{2} GE/mL | Neisseria sicca (4)* | 6.7x10^{5} GE/mL | | Human papilloma virus 16* | 1.8 x 10^{6} GE/mL | Neisseria subflava (2) | 1 x 10^{6} CFU/mL | | Kingella dentrificans | 1 x 10^{6} CFU/mL | Peptostreptococcus anaerobius* | 8.8 x 10^{5} GE/mL | | Kingella kingae | 1 x 10^{6} CFU/mL | Proteus mirabilis | 1 x 10^{6} CFU/mL | | Klebsiella pneumoniae | 1 x 10^{6} CFU/mL | Pseudomonas aeruginosa | 1 x 10^{6} CFU/mL | | Lactobacillus acidophilus | 1 x 10^{6} CFU/mL | Staphylococcus aureus | 1 x 10^{6} CFU/mL | {14} K191352 - Page 14 of 25 | Lactobacillus brevis | 1 x 10^{6} CFU/mL | Staphylococcus epidermidis | 1 x 10^{6} CFU/mL | | --- | --- | --- | --- | | Lactobacillus jensenii | 1 x 10^{6} CFU/mL | Streptococcus agalactiae | 1 x 10^{6} CFU/mL | | Lactobacillus lactis | 1 x 10^{6} CFU/mL | Trichomonas vaginalis | 1 x 10^{6} CFU/mL | | Megasphaera type 1 | in silico | Ureaplasma urealyticum* | 2.0x10^{6} GE/mL | | Mobiluncus curtisii* | 8.6 x 10^{5} GE/mL | Ureaplasma parvum* | 2.5x10^{6} GE/mL | (n) number of strains tested *Organisms tested with genomic DNA (GE = 2ng/mL) ## Interference with Substances Potentially Present at the Anatomic Collection Site The analytical performance of the binx health io CT/NG Assay was evaluated in the presence of a panel of potentially interfering substances that may be found in a vaginal swab sample. Each potential interferent was spiked into pooled vaginal matrix and subsequently spiked with a low concentration (2x LoD) of CT (serovar F) and NG (strain ATCC 49226). Additionally, each potential interferent was spiked into pooled vaginal matrix with no CT or NG present at the highest concentration expected to be found in a vaginal sample (i.e., clinically relevant). All tests generated expected results, demonstrating that the substances listed at the concentrations shown in the table below, do not interfere with the binx health io CT/NG Assay. ### Evaluation of Interference | Interfering Substance | Concentration | | --- | --- | | Human blood | 10% (v/v) | | Contraceptive Jelly | 0.25% (v/v) (Non-oxynol 9 – Gygel) | | Mucus | 0.8% (v/v) (Pooled vaginal fluid) | | Seminal fluid | 5.0% (v/v) | | Moisturizer | 0.25% (v/v) | | Anti-fungal cream | 0.25% (v/v) (Clotrimazole -Canesten) | | | 0.25% (v/v) (Miconazole nitrate - Daktarin) | | Vaginal lubricant | 0.25% (v/v) (KY Jelly) | | Feminine anti-itch cream | 0.25% (v/v) (2% lidocaine cream) | | Leukocytes | 1x 10^{6} cells/mL | | Progesterone | 7 mg/mL | | β-Estradiol | 0.25% (v/v) (Estrace cream) | | Anti-viral cream | 0.25% (v/v) (Acyclovir) | | Haemorrhoidal cream | 0.25% (v/v) | ## Microbial Interference The analytical performance of the binx health io CT/NG Assay was also evaluated for interference from microorganisms that may be encountered at the anatomical site of the swab {15} collection. The study was designed as above, where 10 microorganisms were added at 1 x 10⁵ CFU/mL into pooled vaginal matrix, however, for this study, low levels (2x LoD) of CT and NG organisms were subsequently added to the test samples. The samples were tested with the binx health io CT/NG Assay to check for interference. All tests generated expected results (CT Detected and NG Detected), demonstrating that the organisms listed in the table below, tested at the concentrations shown, do not interfere with the binx health io CT/NG Assay. ## Microbial Interference | Interfering Microorganism | Concentration Tested | | --- | --- | | Corynebacterium xerosis | 1x 10⁵ CFU/mL | | Lactobacillus acidophilus | 1x 10⁵ CFU/mL | | Lactobacillus jensenii | 1x 10⁵ CFU/mL | | Staphylococcus epidermidis | 1x 10⁵ CFU/mL | | Candida albicans | 1x 10⁵ CFU/mL | | Escherichia coli | 1x 10⁵ CFU/mL | | Lactobacillus brevis | 1x 10⁵ CFU/mL | | Lactobacillus lactis | 1x 10⁵ CFU/mL | | Streptococcus agalactiae | 1x 10⁵ CFU/mL | | Candida glabrata | 1x 10⁵ CFU/mL | ## 6. Assay Cut-Off: Thresholds for each of the assay targets were calculated from testing negative clinical specimens and low positive-spiked negative clinical specimens. The negative status of clinical samples was derived from testing with three FDA cleared CT/NG assays, using the two-out-of-three algorithm (any two negatives constituted a negative status). The goal was to balance sensitivity and specificity for CT, NG1 and NG2 targets to maximize the detection of infected subjects while minimizing false positive results. The calculations were based on the distribution of peak heights (in nA), such that the selected threshold provided a clear and distinct separation between positive and negative samples. The selection of the threshold was made with additional consideration of previously generated clinical data which included a large negative dataset. The following peak thresholds were selected: CT - 115nA NG1 - 50nA NG2 - 45nA The data set was also used to set appropriate IC (Internal Control) peak thresholds to allow for detection of significant reagent degradation. ## 7. Carry-Over: The risk of carryover from a high positive sample to a sample that sequentially follows testing was evaluated by testing 50 samples, alternating between high positive samples and negative samples. The test samples were prepared in female vaginal swab matrix in eNAT buffer. The positive samples were double spiked to a concentration of 1 x 10⁵ IFU/mL of CT (serovar F) and 1 x 10⁵ CFU/mL of one NG strain. The negative samples consisted of vaginal swab matrix in K191352 - Page 15 of 25 {16} eNAT buffer. The testing was performed using four binx io CT/NG instruments, with 50 tests (25 negative and 25 positives, run alternately), generating a total of 200 results. All 100 positive samples were correctly reported as CT Detected and NG Detected, and all 100 negative samples were correctly reported as CT Not Detected and NG Not Detected; no false positive results were observed. ## 8. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): ### Sample Stability The stability of vaginal swab samples in the Vaginal Swab Collection Kit was evaluated in an analytical study which used samples of pooled vaginal swab matrix spiked with CT organisms (serovar F) and NG organisms at 3x LoD concentration. Twenty replicates were tested at t=0; additional 20 aliquots were placed in controlled storage at 25°C and another 20 aliquots were placed at 2-8°C. Six replicates of the pooled vaginal matrix were tested as experimental negative controls at each timepoint. A positive and negative daily control was run on each instrument on each testing day. The samples were tested with the binx health io CT/NG Assay at the following storage timepoints: - After 25 hours when stored at 25°C - After 8 days when stored at 2-8°C All positive samples generated expected results, i.e., CT Detected/NG Detected. One of the six negative samples returned a positive result for CT after storage at 2-8°C and additional six replicates of negative swab matrix (stored at 2-8°C for 8 days) were tested, generating expected results for all replicates, for a total of one false positive CT in 12 tests. This suggested a low-level contamination of the test sample rather than contamination of the entire pooled matrix. The study supported a claim that C. trachomatis and N. gonorrhoeae organisms in vaginal swab samples collected in eNAT buffer are stable when stored for 24 hours at 25°C or for a week, when stored at 2-8°C. The summary of the data is shown below. | Timepoint | Temp. | | C. trachomatis | | N. gonorrhoeae | | | --- | --- | --- | --- | --- | --- | --- | | | | | POS | NEG | POS | NEG | | T=0 | Ambient | No. Valid Results | 19^{1} | 6 | 20/20 | 6 | | T=25 Hrs | 25°C | No. Correct Results | 19/19 | 6/6 | 20/20 | 6/6 | | | | % Correct | 100 | 100 | 100 | 100 | | | | 95% CI | 83.2-100 | 61.0-100 | 83.9-100 | 61.0-100 | | T=8 Days | 2-8°C | No. Correct Results | 20/20 | 11/12^{2} | 20/20 | 6/6 | | | | % Correct | 100 | 91.7 | 100 | 100 | | | | 95% CI | 83.9-100 | 64.6-98.5 | 83.9-100 | 61.0-100 | One test was invalid One false positive CT result K191352 - Page 16 of 25 {17} K191352 - Page 17 of 25 # Shipping Stability A study was carried out to verify the performance of the binx health io CT/NG Assay after the cartridges were subjected to simulated shipping conditions conducted using temperature-controlled chambers. Two lots of cartridges were stored for 5 days at 25°C. Subsequently, the cartridges were subjected to two cycles of the following conditions: - 9 hours at 10°C - 6 hours at 25°C - 2 hours at 35°C - 7 hours at 25°C Temperature cycling was carried out in monitored and controlled incubators. Twenty positive samples (at 4x LoD) and 20 negative (eNAT buffer only) samples were tested with the binx health io CT/NG assay using cartridges that were exposed to the above conditions. The positive samples were spiked with CT serovar F and NG strain ATCC 49226 at with 4x LoD; the negative samples consisted of eNAT buffer. Each sample type was tested in 20 replicates. All samples tested generated expected results, i.e., all positive samples generated CT Detected; NG Detected results; all negative samples generated CT Not Detected; NG Not Detected results. The results demonstrated that the binx io CT/NG Cartridges are stable following temporary shipping cycles as shown above and therefore will be suitable for shipping as airfreight. # ISTA 3A Testing ISTA (International Safe Transit Association) 3A testing was carried out to determine whether the packaging is suitable for storage and distribution by surface or air to prevent damage during transit when shipped from the manufacturer to the customer. One shipping container of five cartons of 10 binx io CT/NG Cartridges (50 Cartridges) from a single lot was sent for ISTA 3A testing. The cartons were packed by the manufacturer, shipped, and returned to the manufacturer. Upon receipt, the Cartridges were assessed and scored for damages from 1 to 5: 1. No visible damage 2. Minor damage to the outer packaging and/or slight misshapen appearance that still allows any flaps to open/close without significant re-shaping of the package by hand. No punctures. 3. Moderate damage, medium tears/breaches to the outer packaging and/or squashing of the packaging to the extent that the package can be re-shaped by hand to allow any flaps to open/close. 4. Significant damage/tearing/breach to the outer packaging but the integrity of the package is maintained such that it is still capable of retaining the contents in the 'this way up' orientation. 5. The integrity of the outer packaging is damaged to the extent that the packaging no longer holds the contents in any orientation. When the shipping container and each carton were inspected following shipping simulation, the following observations were documented: {18} K191352 - Page 18 of 25 Integrity During Shipping | Packaging | Score | Pass/Fail | | --- | --- | --- | | Shipping container (outer box) | 2 | Pass | | Carton 1 (inner box) | 2 | Pass | | Carton 2 | 1 | Pass | | Carton 3 | 2 | Pass | | Carton 4 | 2 | Pass | | Carton 5 | 2 | Pass | All 50 Cartridges were assessed individually and none showed any visible damage during shipping. The Cartridges were used in testing of positive and negative samples and all yielded expected result. The results showed that the packaging is adequate for transporting the Cartridges and that the performance of the binx health io CT/NG Assay was unaffected by shipping simulation. ## Shelf Life Stability A real-time stability study is being conducted to determine the shelf-life of the binx health io CT/NG Cartridge and its consumables. The study is performed with reagents that have been stored at the upper limit of the tests' claimed room temperature range (28-30°C) and at 2-8°C. The study is conducted with a panel of three samples at the following concentrations of the organism: a low positive sample, a moderate positive sample, and a negative sample, using two distinct lots of Cartridges. The testing will be conducted monthly for the first three months, then every two months, with the remaining test points at 12 months, 14 months, 18 months, and 20 months. The sponsor included the evaluation criteria and the acceptance criteria (lowest peak heights above threshold for each analyte target) in the 510(k) submission. At this time, the results of the testing support a shelf life claim for the binx health io CT/NG Cartridge of three months at less than 30°C (i.e., 25°C), and four months at 2-8°C. ## Daily Quality Controls One positive and one negative control were run daily on each instrument that was used to run female patient samples on each day of testing. A total of 1,096 daily controls were run throughout the study with 17 invalid results (8 invalids for Positive Control and 9 invalids for Negative Control), with an overall invalid rate of 1.55%. There were also five false negative results obtained for NG, which subsequently generated correct results upon repeat. ## B Comparison Studies: 1. Method Comparison with Predicate Device: See Clinical Studies section below. 2. Matrix Comparison: Not applicable. {19} C Clinical Studies: The clinical performance of the binx health io CT/NG Test System used with female vaginal swabs was evaluated in a prospective study conducted at nine clinical sites, of which seven performed testing with the binx health io CT/NG Assay and two sites served as patient enrollment and specimen collection sites only. The participating testing sites, at geographically different locations across the U.S., were representative of point-of-care (POC) environments and included STD clinics, family planning/OB-GYN clinics, and HIV clinics. Reference testing was conducted at one CLIA certified clinical laboratory. The study was conducted between September 2018 and March 2019 and consisted of prospective enrollment of 1634 female participants between the ages of 16 and 74 (median age of 27 years) with signs and symptoms of genitourinary infection or who were at risk of such an infection. All subjects enrolled in the study provided a signed informed consent form; however, 21 subjects were excluded due to ineligibility or withdrawal, leaving 1613 subjects available for testing. Specimens from 89 participants were excluded due to deviations from the study protocol, leaving 1524 subjects for testing with the binx health io CT/NG Assay. Approximately half of the vaginal samples tested with the binx assay were self-collected (736) and half were clinician-collected (788). There were 17 initial invalid results reported from testing patients with the binx health io CT/NG Assay (17/1524) for an overall invalid rate of 1.11%. However, one specimen (clinician-collected vaginal swab) was indeterminate after generating two successive invalid results on the binx health io CT/NG Assay, leaving a total of 1,523 evaluable vaginal swab specimens for inclusion in the final data analysis. The table below shows the excluded samples for each exclusion category. Excluded Samples | Exclusion Category from line list | Number excluded | | --- | --- | | Improper Specimen Collection | 12 | | Improper Testing | 6 | | Miscellaneous | 4 | | Participant Ineligible | 17 | | Participant Withdrew | 4 | | Specimen Handling | | | No temperature log during transit | 7 | | Sample leaked in transit | 4 | | Samples lost in transit | 3 | | Shipping temperature outside specifications | 53 | The testing with the binx health io CT/NG Assay was performed by 30 operators (test users), of which 26 were representative of an intended test user in POC environments, i.e., a member of the healthcare providing team without laboratory training. The users who participated in the binx study were trained to use the binx io CT/NG Assay. External Quality Control samples (positive and negative) were run on each testing day. A total of 1,142 daily controls were run throughout the study, using 29 io instruments across the testing sites. There was a total of 16 invalid results obtained from testing external QC samples. K191352 - Page 19 of 25 {20} Clinical Sensitivity and Specificity The clinical performance of the binx health io CT/NG Assay was assessed by comparing the test results obtained with the binx health io CT/NG Assay to a Composite Infected Status (CIS) for each subject, derived by an algorithm of test results obtained with three FDA-cleared assays (each with different DNA targets for CT and NG than the binx assay), testing vaginal samples. Participants were classified as "infected" if at least two of the three comparator assays provided positive results; all other participants were classified as "not infected." All female subjects provided four vaginal swab samples for use in the study. Three vaginal swabs from each participant were collected by a clinician, according to a pre-determined randomization scheme regarding the order of collection, for testing with three FDA cleared molecular assays (NAATs) detecting CT and NG. The fourth swab, either self-collected or clinician collected, was used for testing with the binx health io CT/NG Assay. For self-collected specimens, the binx vaginal swab sample was collected first and for clinician-collected specimens, the binx vaginal swab sample was collected last. The clinical sensitivity and specificity of the binx health io CT/NG Assay, when used with vaginal specimens, was calculated in comparison with the CIS comparator algorithm and is shown by symptom status, along with prevalence rates observed during the clinical study in the two tables below. Clinical Performance of the binx health io CT/NG Assay for C. trachomatis with Vaginal Swab Specimens | Symptom Status | N | TP | FN | TN | FP | Sensitivity (95% CI)¹ | Specificity (95% CI)¹ | Prevalence | PPV (95% CI)² | NPV (95% CI)² | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Asx | 706 | 65 | 2 | 634 | 5 | 97.0% (89.8; 99.2) | 99.2% (98.2; 99.7) | 9.5% | 92.7% (84.1; 97.6) | 99.7% (98.9; 100.0) | | Sx | 817 | 59 | 3 | 747 | 8 | 95.2% (86.7; 98.3) | 98.9% (97.9; 99.5) | 7.6% | 88.1% (77.8; 94.7) | 99.6% (98.8; 99.9) | | Total | 1523 | 124 | 5 | 1381 | 13 | 96.1% (91.3; 98.3) | 99.1% (98.4; 99.5) | 8.5% | 90.5% (84.3; 94.9) | 99.6% (99.2; 99.9) | Asx = Asymptomatic; Sx Symptomatic TP = True Positive FN = False Negative; TN = True Negative; FP = False Positive PPV = Positive Predictive Value; NPV = Negative Predictive Value ¹95% Confidence Intervals (CI) by Wilson's Score Method ²95% Confidence Intervals (CI) by Exact Method K191352 - Page 20 of 25 {21} Clinical Performance of the binx health io CT/NG Assay for N. gonorrhoeae with Vaginal Swab Specimens | Symptom Status | N | TP | FN | TN | FP | Sensitivity (95% CI)¹ | Specificity (95% CI)¹ | Prevalence | PPV (95% CI)² | NPV (95% CI)² | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Asx | 706 | 16 | 0 | 689 | 1 | 100.0% (80.6; 100.0) | 99.9% (99.2; 100.0) | 2.3% | 94.1% (71.3; 99.9) | 100.0% (99.5; 100.0) | | Sx | 817 | 29 | 0 | 787 | 1 | 100.0% (88.3; 100.0) | 99.9% (99.3; 100.0) | 3.5% | 96.7% (82.8; 99.9) | 100.0% (99.5; 100.0) | | Total | 1523 | 45 | 0 | 1476 | 2 | 100.0% (92.4; 100.0) | 99.9% (99.51; 100.0) | 3.0% | 95.7% (85.5; 99.5) | 100.0% (99.8; 100.0) | Asx = Asymptomatic; Sx Symptomatic TP = True Positive FN = False Negative; TN = True Negative; FP = False Positive PPV = Positive Predictive Value; NPV = Negative Predictive Value ¹95% Confidence Intervals (CI) by Wilson’s Score Method ²95% Confidence Intervals (CI) by Exact Method The frequency of test outcomes from the reference NAATs and the binx health io CT/NG Assay is summarized for each target (organism) in the two tables below. Frequency of Test Outcomes for Chlamydia trachomatis by Symptom Status | CIS | Comparator Test Results | | | binx io Result | Symptom Status | | Total | | --- | --- | --- | --- | --- | --- | --- | --- | | | NAAT1 | NAAT2 | NAAT3 | | Sx | Asx | | | NI | - | - | - | - | 731 | 620 | 1351 | | NI | - | - | + | - | 2 | 3 | 5 | | NI | - | + | - | - | 5 | 7 | 12 | | NI | + | - | - | - | 0 | 0 | 0 | | NI | - | - | IND | - | 2 | 1 | 3 | | NI | - | IND | - | - | 1 | 2 | 3 | | NI | IND | - | - | - | 6 | 1 | 7 | | NI | - | - | - | + | 8 | 5 | 13 | | NI | - | - | + | + | 0 | 0 | 0 | | NI | - | + | - | + | 0 | 0 | 0 | | NI | + | - | - | + | 0 | 0 | 0 | | NI | - | - | IND | + | 0 | 0 | 0 | | NI | - | IND | - | + | 0 | 0 | 0 | | NI | IND | - | - | + | 0 | 0 | 0 | | Total not Infected | | | | | 755 | 639 | 1394 | | I | + | + | + | + | 56 | 63 | 119 | | I | + | + | - | + | 0 | 1 | 1 | | I | + | - | + | + | 0 | 1 | 1 | | I | - | + | + | + | 1 | 0 | 1 | | I | + | + | IND | + | 0 | 0 | 0 | | I | + | IND | + | + | 1 | 0 | 1 | | I | IND | + | + | + | 1 | 0 | 1 | | I | + | + | + | - | 1 | 2 | 3 | | I | + | + | - | - | 2 | 0 | 2 | K191352 - Page 21 of 25 {22} K191352 - Page 22 of 25 | I | + | - | + | - | 0 | 0 | 0 | | --- | --- | --- | --- | --- | --- | --- | --- | | I | - | + | + | - | 0 | 0 | 0 | | I | + | + | IND | - | 0 | 0 | 0 | | I | + | IND | + | - | 0 | 0 | 0 | | I | IND | + | + | - | 0 | 0 | 0 | | Total Infected | | | | | 62 | 67 | 129 | CIS = Comparator Infected Status NI = Not Infected; I = Infected IND = Indeterminate Sx = Symptomatic; Asx = Asymptomatic ## Frequency of Test Outcomes for Neisseria gonorrhoeae by Symptom Status | CIS | Comparator Test Results | | | binx io Result | Symptom Status | | Total | | --- | --- | --- | --- | --- | --- | --- | --- | | | NAAT1 | NAAT2 | NAAT3 | | Sx | Asx | | | NI | - | - | - | - | 776 | 681 | 1457 | | NI | - | - | + | - | 3 | 4 | 7 | | NI | - | + | - | - | 0 | 0 | 0 | | NI | + | - | - | - | 0 | 0 | 0 | | NI | - | - | IND | - | 2 | 1 | 3 | | NI | - | IND | - | - | 0 | 2 | 2 | | NI | IND | - | - | - | 6 | 1 | 7 | | NI | - | - | - | + | 0 | 0 | 0 | | NI | - | - | + | + | 0 | 0 | 0 | | NI | - | + | - | + | 0 | 0 | 0 | | NI | + | - | - | + | 1 | 0 | 1 | | NI | - | - | IND | + | 0 | 0 | 0 | | NI | - | IND | - | + | 0 | 1 | 1 | | NI | IND | - | - | + | 0 | 0 | 0 | | Total not Infected | | | | | 788 | 690 | 1478 | | I | + | + | + | + | 24 | 15 | 39 | | I | + | + | - | + | 1 | 0 | 1 | | I | + | - | + | + | 2 | 1 | 3 | | I | - | + | + | + | 1 | 0 | 1 | | I | + | + | IND | + | 0 | 0 | 0 | | I | + | IND | + | + | 0 | 0 | 0 | | I | IND | + | + | + | 1 | 0 | 1 | | I | + | + | + | - | 0 | 0 | 0 | | I | + | + | - | - | 0 | 0 | 0 | | I | + | - | + | - | 0 | 0 | 0 | | I | - | + | + | - | 0 | 0 | 0 | | I | + | + | IND | - | 0 | 0 | 0 | | I | + | IND | + | - | 0 | 0 | 0 | | I | IND | + | + | - | 0 | 0 | 0 | | Total Infected | | | | | 29 | 16 | 45 | CIS = Comparator Infected Status NI = Not Infected; I = Infected IND = Indeterminate Sx = Symptomatic; Asx = Asymptomatic {23} D Clinical Cut-Off: Not applicable. E Expected Values/Reference Range: The prevalence of infection with CT and/or NG in patient populations depends on risk factors such as age, gender, type of clinic, and the sensitivity of the test used to detect infections. The observed positivity rates of the binx health io CT/NG Assay in females during the clinical study, testing vaginal swab samples, are shown below. Positivity Rates by binx health io CT/NG Assay in the Clinical Study | | | | | CT | | NG | | | --- | --- | --- | --- | --- | --- | --- | --- | | Site No. | Type of Clinic | US State | Total No. Samples | No. Positive by binx | % Positivity Rate | No. Positive by binx | % Positivity Rate | | 1 | STD Clinic | IN | 74 | 5 | 6.8% | 8 | 10.8% | | 2 | STD Clinic | MD | 39 | 2 | 5.1% | 1 | 2.6% | | 3 | STD Clinic | LA | 85 | 16 | 18.8% | 0 | 0.0% | | 4 | Family Planning | TX | 1053 | 99 | 9.4% | 26 | 2.5% | | 5 | Family Planning | PA | 73 | 3 | 4.1% | 1 | 1.4% | | 6 | STD Clinic | AL | 154 | 11 | 7.1% | 9 | 5.8% | | 7 | STD Clinic | NC | 14 | 0 | 0.0% | 2 | 14.3% | | 8 | OBGYN | MD | 9 | 1 | 11.1% | 0 | 0.0% | | 9 | STD Clinic | MS | 22 | 0 | 0.0% | 0 | 0.0% | | | Total | | 1523 | 137 | 9.0% | 47 | 3.0% | F Other Supportive Instrument Performance Characteristics Data: The sponsor performed usability studies as well as analytical studies to evaluate the robustness of the test system under sub-optimal operational scenarios in a point-of-care environment, when used by non-laboratory-trained personnel. Usability Studies The binx io System was designed for use in near-patient settings, such as a clinic or a medical practice, with laboratories conducting moderate complexity diagnostic testing under CLIA oversight. The sponsor developed the system with built-in risk-mitigation features specifically to meet the requirements of the intended users in those settings, including: - The design of the Cartridge includes a self-metering pipet for sample transfer and an invertible sample chamber. - All reagents are enclosed in the test Cartridge. K191352 - Page 23 of 25 {24} - The instrument is simple to set up and requires no calibration. - The screen display is interactive and informative. - The software provides prompts for correct sequence of operations, including specimen identification. - Two levels of user credentials are employed to include basic users and administrators. - No access to the operating system is possible, to avoid malicious access. The usability of the system was evaluated prior to production by a group of users that represented the intended users in POC settings and was found acceptable. # Operational Environment The sponsor evaluated the io CT/NG Test System operation to assess whether the assay can run reliably within the specified temperature and humidity limits and whether it aborts due to marginal conditions (rather than giving incorrect results). The sponsor considered the following: Instrument operating range (specifications): $10^{\circ}\mathrm{C} - 35^{\circ}\mathrm{C} / 0\% -80\%$ RH Typical ambient working environment: $20^{\circ}\mathrm{C} - 25^{\circ}\mathrm{C} / 20\% -60\%$ RH The testing was performed at the outermost limits of the potential environmental conditions: $9^{\circ}\mathrm{C} / 40\%$ RH $37^{\circ}\mathrm{C} / 40\%$ RH $20^{\circ}\mathrm{C} / 83\%$ RH $30^{\circ}\mathrm{C} / 83\%$ RH $20^{\circ}\mathrm{C} / 15\%$ RH The instrument and cartridges were equilibrated to the set conditions, followed by testing 3 consecutive positive samples (spiked with CT and NG organisms at $4\mathrm{x}$ LoD). All tests returned valid positive results. No instrument failures were observed. The summary of the study conditions and results is shown below. | Condition | Temperature | Relative Humidity | CT/NG Positive Sample | CT/NG Negative Sample | Pass/Fail | | --- | --- | --- | --- | --- | --- | | 1 | 9°C | 40% | 3/3 | 3/3 | Pass | | 2 | 37°C | 40% | 3/3 | 3/3 | Pass | | 3 | 20°C | 83% | 3/3 | 3/3 | Pass | | 4 | 30°C | 83% | 3/3 | 3/3 | Pass | | 5 | 20°C | 15% | 3/3 | 3/3 | Pass | External controls were included with each test condition under evaluation. One positive control generated a false negative result; a positive result was obtained upon a retest. Performing Testing with Cold (out of refrigerator) Reagents. To demonstrate that binx health io CT/NG Cartridges and/or samples can be run immediately following refrigerator storage, a study was carried out to evaluate the performance of binx io CT/NG Cartridges when tested immediately after removing from storage (for a minimum of 12 K191352 - Page 24 of 25 {25} hours) at 2-8°C, using samples that were either at room temperature or stored refrigerated at 2-8°C for a minimum of 12 hours. Vaginal swab matrix was spiked to 4x LoD concentration with one CT strains (serovar F) and one NG strain (ATCC 49226). Three positive samples and three negative samples (unspiked negative matrix) were tested for each condition. All tests generated expected results. | Cartridge Storage | Sample Storage | No. Correct Results | | | --- | --- | --- | --- | | | | CT/NG Positive | CT/NG Negative | | 2-8°C | Ambient | 3/3 | 3/3 | | 2-8°C | 2-8°C | 3/3 | 3/3 | ## VIII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. ## IX Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K191352 - Page 25 of 25
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...