K251630 · Siemens Healthcare Diagnostics, Inc. · LTJ · Jan 5, 2026 · Immunology
Device Facts
Record ID
K251630
Device Name
Atellica IM Total PSA II (tPSAII)
Applicant
Siemens Healthcare Diagnostics, Inc.
Product Code
LTJ · Immunology
Decision Date
Jan 5, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.6010
Device Class
Class 2
Indications for Use
Atellica® IM total PSA II (tPSAII) assay is for in vitro diagnostic use in the quantitative measurement of prostate-specific antigen (PSA) in human serum and plasma (EDTA and lithium-heparin) using the Atellica IM Analyzer. This assay is indicated as an aid in the detection of prostate cancer in conjunction with a digital rectal exam (DRE) in men aged 50 years and older. Prostate biopsy is required for diagnosis of prostate cancer. This assay is further indicated as an aid in the management (monitoring) of patients with prostate cancer.
Device Story
In vitro diagnostic assay for quantitative measurement of total prostate-specific antigen (PSA) in human serum and plasma; utilizes Atellica IM Analyzer. Principle of operation: sandwich immunoassay using monoclonal mouse anti-fPSA and anti-PSA antibodies labeled with acridinium ester and biotin-streptavidin paramagnetic particles. Used in clinical laboratories by trained personnel. Output provided as numerical concentration (ng/mL) to healthcare providers; assists in prostate cancer detection and patient monitoring. Benefits include standardized quantitative assessment of PSA levels to support clinical decision-making regarding biopsy and disease management.
Clinical Evidence
Prospective clinical study of 88 male subjects (mean age 67.7) with confirmed prostate cancer. 323 samples (baseline and follow-up) were analyzed. Clinical sensitivity was 54.5% (95% CI: 36.4%-71.9%) and specificity was 87.6% (95% CI: 83.3%-91.2%) for detecting disease progression (defined by 50% RCV). Comparison with the predicate device showed 91.8% PPA and 96.7% NPA. Analytical performance (precision, linearity, specificity) was established in a separate referenced submission (P240021).
Indicated for men aged 50+ as an aid in prostate cancer detection (with DRE) and management (monitoring). Prostate biopsy required for diagnosis.
Regulatory Classification
Identification
A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.
Special Controls
*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.
P240021 — Atellica IM Total PSA II (tPSAII) · Siemens Healthcare Diagnostics, Inc. · Dec 3, 2025
K974816 — ACCESS(R) PSA REAGENTS ON THE ACCESS(R) IMMUNOASSAY ANALYZER 33220, 33225, 33229 · Beckman Instruments, Inc. · Feb 20, 1998
P990056 — ELECSYS TOTAL PSA IMMUNOASSAY AND TOTAL PSA CALSET · Roche Diagnostics Corp. · Nov 22, 2000
K031343 — TPSA FLEX REAGENT CARTRIDGE · Dade Behring, Inc. · May 15, 2003
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K251630
B Applicant
Siemens Healthcare Diagnostics, Inc.
C Proprietary and Established Names
Atellica IM total PSA II (tPSAII)
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LTJ | Class II | 21 CFR 866.6010 - Tumor-Associated Antigen Immunological Test System | IM - Immunology |
## II Submission/Device Overview:
A Purpose for Submission:
Addition of intended use/indication for use of the device as aid in monitoring of patients with prostate cancer
B Measurand:
Total Prostate Specific Antigen (tPSA)
C Type of Test:
Quantitative, Chemiluminescent Immunoassay
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K251630 - Page 2 of 9
## III Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
Atellica IM total PSA II (tPSAII) assay is for in vitro diagnostic use in the quantitative measurement of total prostate-specific antigen (PSA) in human serum and plasma (EDTA and lithium-heparin) using the Atellica IM Analyzer.
This assay is indicated as an aid in the detection of prostate cancer in conjunction with a digital rectal exam (DRE) in men aged 50 years and older. Prostate biopsy is required for diagnosis of prostate cancer. This assay is further indicated as an aid in the management (monitoring) of patients with prostate cancer.
### C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
The Instructions for Use of the device contains the following warning statement:
The concentration of total PSA in a given specimen, as determined by assays from different manufacturers, can vary due to differences in assay methods and reagent specificity. The results reported by the laboratory to the physician must include the identity of the assay for total PSA used. Values obtained with different assay methods cannot be used interchangeably. If, in the course of monitoring a patient, the assay method used for determining serial levels of total PSA is changed, the laboratory must perform additional testing to confirm baseline values.
### D Special Instrument Requirements:
Atellica IM Analyzer
## IV Device/System Characteristics:
### A Device Description:
Materials provided:
The Atellica IM total PSA II (tPSAII) is available in a 1-pack (100 tests) kit and in a 5-pack (500 tests) kit. Both versions of the assay include:
- tPSAII ReadyPack primary reagent pack consisting of:
- Lite Reagent: 10.0 mL/reagent pack; Unlabeled monoclonal mouse anti-free-PSA (fPSA) antibody (~250 ng/mL); monoclonal mouse anti-PSA antibody
{2}
$(\sim 180 \mathrm{ng} / \mathrm{mL})$ labeled with acridinium ester; buffer; bovine serum albumin (BSA); preservative
- Solid Phase: 20.0 mL/reagent pack; Monoclonal mouse anti-PSA antibody (~3.5 μg/mL) labeled with biotin and bound to streptavidin paramagnetic particles; buffer; BSA, bovine gamma globulin (BGG); sodium azide (< 0.1%); preservative
- Atellica IM tPSAII master calibration curve
- tPSAII CAL low calibrator: 2.0 mL/vial; Purified PSA from human seminal fluid in buffer; BSA; sodium azide (< 0.1%)
- tPSAII CAL high calibrator: 2.0 mL/vial; Purified PSA from human seminal fluid in buffer; BSA; sodium azide (< 0.1%)
- Atellica IM tPSAII CAL calibrator assigned value sheet
Materials needed but not provided:
- Atellica IM Analyzer – with additional system fluids to operate the system including Atellica IM Wash, Atellica IM Acid, Atellica IM Base, and Atellica IM Cleaner
Optional materials may be used but not provided:
- Atellica IM Multi-Diluent 2
- Atellica IM tPSAII MCM (master curve material)
## B Principle of Operation:
The Atellica IM total PSA II (tPSAII) assay is a fully automated sandwich immunoassay using acridinium ester chemiluminescent technology. The assay uses three monoclonal mouse antibodies in the tPSAII ReadyPack primary reagent pack. The Lite Reagent contains a monoclonal anti-PSA antibody labeled with acridinium ester, and an unlabeled free-PSA-specific monoclonal mouse anti-PSA antibody. The Solid Phase contains a monoclonal mouse anti-PSA antibody labeled with biotin and bound to streptavidin paramagnetic latex particles. The sample is incubated with the Lite Reagent and Solid Phase simultaneously, and then the immune complex is washed. A direct relationship exists between the amount of analyte present in the patient sample and the amount of relative light units (RLUs) detected by the system.
## V Substantial Equivalence Information:
### A Predicate Device Name(s):
- Atellica IM PSA assay
### B Predicate 510(k) Number(s):
- P950021/S015
K251630 - Page 3 of 9
{3}
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K251630 | P950021/S015 |
| --- | --- | --- |
| Device Trade Name | Atellica IM tPSAII | Atellica IM PSA |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Atellica IM total PSA II (tPSAII) assay is for in vitro diagnostic use in the quantitative measurement of total prostate-specific antigen (PSA) in human serum and plasma (EDTA and lithium-heparin) using the Atellica IM Analyzer. This assay is indicated as an aid in the detection of prostate cancer in conjunction with a digital rectal exam (DRE) in men aged 50 years and older. Prostate biopsy is required for diagnosis of prostate cancer. This assay is further indicated as an aid in the management (monitoring) of patients with prostate cancer | The Atellica IM Prostate Specific Antigen (PSA) assay is for in vitro diagnostic use in the quantitative measurement of prostate specific antigen (PSA) in human serum using the Atellica IM Analyzer. This assay is indicated for the measurement of serum PSA in conjunction with a digital rectal exam (DRE) as an aid in the detection of prostate cancer in men aged 50 years and older. This assay is further indicated as an aid in the management (monitoring) of patients with prostate cancer. |
| Analyzer | Atellica IM | Same |
| Measurement | Quantitative | Same |
| Technology | Chemiluminescence | Same |
| Principle | Sandwich Immunoassay | Same |
| Calibration Levels | Two (low and high) | Same |
| General Device Characteristic Differences | | |
| Specimen | Serum and Plasma | Serum |
| Sample Volume | 30 μL | 35 μL |
| Analytical Measuring Interval | 0.009 – 50.00 ng/mL | 0.04–100.00 ng/mL |
| Detection Antibody | Unlabeled monoclonal mouse anti-fPSA antibody; monoclonal mouse anti-PSA antibody labeled with acridinium ester | Polyclonal goat anti-PSA antibody labeled with acridinium ester |
| Capture Antibody | Biotinylated mouse monoclonal anti-PSA antibody bound to streptavidin paramagnetic particles | Mouse monoclonal anti-PSA antibody covalently coupled to paramagnetic particles |
K251630 - Page 4 of 9
{4}
VI Standards/Guidance Documents Referenced:
The following Clinical and Laboratory Standards Institute (CLSI) guidelines were used:
- CLSI EP34 1st Edition, Establishing and Verifying an Extended Measuring Interval Through Specimen Dilution and Spiking
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
Analytical performance of the Atellica IM tPSAII has been established in P240021.
1. Precision/Reproducibility:
Refer to P240021
2. Linearity:
a. Linearity:
Refer to P240021
b. Dilution:
The Atellica IM tPSAII has automated dilution function when the patient sample results in the value exceed the upper limit of the AMI of the assay (result is “>50 ng/mL”). A study was performed according to CLSI EP34-Ed1 to evaluate the effect of auto-dilution to the assay performance with Atellica IM Multi-Diluent 2. A high PSA stock was prepared by spiking high concentration PSA (Scripps PSA-ACT) into normal female serum above the upper limits of AMI and few concentrations inside of the AMI. Data of the study verified the function of auto-dilutions 1:5, 1:10, 1:50, 1:100 and 1:500 with diluent Atellica IM Multi-Diluent 2.
3. Analytical Specificity/Interference:
Refer to P240021
4. Assay Reportable Range:
The Atellica IM tPSAII assay reportable range without dilution is 0.009–50.00 ng/mL.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Refer to P240021
K251630 - Page 5 of 9
{5}
6. Detection Capability:
For LoB, LoD and LoQ values, refer to P240021
7. Assay Cut-Off:
Refer to Clinical Cut-off
## B Comparison Studies:
1. Method Comparison with Predicate Device:
See Section C below
2. Matrix Comparison:
Refer to P240021
## C Clinical Studies:
1. Clinical Sensitivity and Specificity:
A prospective clinical study was conducted by enrolling subjects from six different collection sites to evaluate the clinical performance of the Atellica IM tPSAII as an aid in monitoring patients diagnosed with prostate cancer. A total of 88 male subjects that met the following inclusion and exclusion criteria are used in the study analysis:
### Inclusion Criteria:
- The subject must be a male above 50 years of age with a confirmed diagnosis of prostate cancer
- The specimen collected must be serum
- The subject must be monitored using local, standard of care PSA testing
- Specimens were collected at minimum of three visits per patient, with collections occurring within ±28 days of standard-of-care PSA draws.
### Exclusion Criteria:
- Subject previously diagnosed with a malignancy (non-melanoma skin cancer excluded) other than prostate cancer within five years prior to the study blood draw
Among 88 subjects, majority were Caucasian (n=73, 83.0%), with the remaining subjects being African Americans (n=15, 17.0%).
In the study, 13 subjects (14.8%) were from group 54-59 years old, 39 subjects (44.3%) were from age group 60-69 years old, 32 (36.4%) subjects were from age group 70-79 years old, and 4 (4.5%) subjects were from age group 80-86 years old. The mean age was 67.7 years ranging from 54 to 86 years.
K251630 - Page 6 of 9
{6}
The length of time (in days) over which the subjects were monitored ranged from 121 to 1014 days (median of 768 days). The median interval (in days) between successive visits was 182, ranging from 61 to 465 days.
# Results:
In this study, a total of 323 samples (including 88 baseline and 235 follow-up measurements) were obtained and measured using both the candidate tPSAII and predicate total PSA assays using the Atellica IM Analyzer platform. At each follow-up visit, the percentage change of the total PSA was calculated by comparing the test result to the results obtained from the previous visit. The performance of the tPSAII assay as an aid in the management of patients with prostate cancer subjects was evaluated by assessing the changes in PSA levels correlated to the clinical assessment at the time of assay. The reference change value (RCV) of $50.0\%$ of the visit compared to previous visit is used to determine the disease progression for both tPSAII and the predicate device. The change in the PSA concentration was compared to the change in the clinical status of the patients as determined by the physician based upon an examination of the subject for clinical signs and symptoms, including Digital Rectal Exam (DRE), medical notes and records, radiographic findings, and subject interviews. The following definition was used to categorize the patient's disease status:
- No Evidence of Disease (NED) – Based on one or more of the following diagnostic tools (total PSA, DRE, Biopsy, Ultrasound, MRI, CT, PET, bone scan) the physician determines that there is no tumor present.
- Stable/Persistent – Based on one or more of the following diagnostic tools (total PSA, DRE, Biopsy, Ultrasound, MRI, CT, PET, bone scan) the physician determines that tumor burden remains constant.
- Responding – Based on one or more of the following diagnostic tools (total PSA, DRE, Biopsy, Ultrasound, MRI, CT, PET, bone scan) the physician determines that tumor burden is decreasing.
- Progression – Based on one or more of the following diagnostic tools (total PSA, DRE, Biopsy, Ultrasound, MRI, CT, PET, bone scan) the physician determines that tumor burden is increasing and/or metastasis may be present.
The following table represents the summary of all follow-ups $(n = 323)$ for subjects at which a clinical evaluation occurred for the tPSA II assay. A positive percentage change of the subjects at these clinical evaluations was defined by increase in $50\%$ RCV of the visit compared to previous visit.
| | Clinical Status | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | Progression | NED | Stable | Responding | Total |
| Change in Atellica IM tPSAII | Progression | 18 | 13 | 23 | 0 | 54 |
| | No Progression | 15 | 52 | 143 | 59 | 269 |
| | Total | 33 | 65 | 166 | 59 | 323 |
To evaluate clinical sensitivity and specific of the Atellica IM tPSAII, the clinical disease status was condensed into two categories: Progression and No-progression. Subjects with progression contained those monitoring events defined as progressive disease. Subjects with no-progression contained those monitoring events defined as NED, stable disease and
K251630 - Page 7 of 9
{7}
responding disease. The clinical performances of the Atellica IM tPSAII are summarized in the tables below:
| | Clinical Status | | | |
| --- | --- | --- | --- | --- |
| | | Progression | No-Progression | Total |
| Change in Atellica IM tPSAII | Progression | 18 | 36 | 54 |
| | No Progression | 15 | 254 | 269 |
| | Total | 33 | 290 | 323 |
| | Point Estimate | 95% CI* |
| --- | --- | --- |
| Sensitivity | 54.5% (18/33) | 36.4%-71.9% |
| Specificity | 87.6% (254/290) | 83.3%-91.2% |
| Positive Predictive Value (PPV) | 33.3% (18/54) | 21.1%-47.5% |
| Negative Predictive Value (NPV) | 94.4% (254/269) | 91.0%-96.9% |
| Positive Likelihood Ratio | 4.39 | 2.95-6.56 |
| Negative Likelihood Ratio | 0.52 | 0.34-0.79 |
| Prevalence | 10.22% (33/323) | |
*95% CI were calculate using bootstrap method
# 2. Comparison of the tPSAII and Predicate total PSA
All 323 follow-ups for 88 subjects were also tested with the predicate device, Atellica IM PSA assay. For the predicate device, a positive percentage change of the subjects at these clinical evaluations was also defined by increase in $50\%$ RCV of the visit compared to previous visit. The following tables summarized the performance estimates of the predicate device.
| | Clinical Status | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | Progression | No Progression | | | Total |
| | | | NED | Stable | Responding | |
| Change in Atellica IM PSA | Progression | 18 | 10 | 21 | 0 | 49 |
| | No Progression | 15 | 55 | 145 | 59 | 274 |
| | Total | 33 | 65 | 166 | 59 | 323 |
| | Point Estimate | 95% CI* |
| --- | --- | --- |
| Sensitivity | 54.5% (18/33) | 36.4%-71.9% |
| Specificity | 89.3% (259/290) | 85.3%-92.6% |
| Positive Predictive Value (PPV) | 36.7% (18/49) | 23.4%-51.7% |
| Negative Predictive Value (NPV) | 94.5% (259/274) | 91.2%-97.0% |
| Positive Likelihood Ratio | 5.10 | 3.32-7.81 |
| Negative Likelihood Ratio | 0.51 | 0.33-0.78 |
*95% CI were calculate using bootstrap method
Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) were evaluated between Atellica IM tPSAII and predicate total PSA and summarized in the table below:
K251630 - Page 8 of 9
{8}
| | Atellica IM PSA (Predicate) | | | |
| --- | --- | --- | --- | --- |
| | | Progression | No Progression | Total |
| Atellica IM tPSAII | Progression | 45 | 9 | 54 |
| | No Progression | 4 | 265 | 269 |
| | Total | 49 | 274 | 323 |
PPA : 91.8% (45/49), (95% CI : 80.4%-97.7%)
NPA : 96.7% (265/274), (95% CI : 93.9%-98.6%)
The overall results demonstrate substantial equivalence between Atellica IM tPSAII and predicate device for monitoring disease progression. Both assays exhibited identical sensitivity (54.5%) when compared against clinical progression status, with comparable specificity values (87.6% for the Atellica IM tPSAII vs. 89.3% for predicate total PSA). The direct comparison between the two assays with a positive percent agreement of 91.8% and negative percent agreement of 96.7%.
D Clinical Cut-Off:
Refer to P240021
E Expected Values/Reference Range:
Refer to P240021
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete.
K251630 - Page 9 of 9
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.