Optilite Hevylite IgG Kappa is a quantitative in vitro assay performed on the Optilite analyser for the measurement of IgG kappa (IgG heavy chain and kappa light chain intact immunoglobulin) in serum. Measurement of Hevylite Human IgG Kappa is used alongside Hevylite Human IgG Lambda to calculate the IgG Kappa / IgG Lambda ratio. The Hevylite Human IgG Kappa / IgG Lambda ratio can be used when monitoring previously diagnosed IgG multiple myeloma and is used in conjunction with other laboratory tests and clinical evaluations. The assignment of complete response is reliant upon other tests including immunofixation, bone marrow and urine assessments. Optilite Hevylite IgG Lambda is a quantitative in vitro assay performed on the Optilite analyser for the measurement of IgG lambda (IgG heavy chain and lambda light chain intact immunoglobulin) in serum. Measurement of Hevylite Human IgG Lambda is used alongside Hevylite Human IgG Kappa to calculate the IgG Kappa / IgG Lambda ratio. The Hevylite Human IgG Kappa / IgG Lambda ratio can be used when monitoring previously diagnosed IgG multiple myeloma and is used in conjunction with other laboratory tests and clinical evaluations. The assignment of complete response is reliant upon other tests including immunofixation, bone marrow and urine assessments.
Device Story
The Optilite Hevylite IgG Kappa and IgG Lambda kits are in vitro diagnostic assays used on the Binding Site Optilite analyser. The device measures IgG heavy chain and light chain (Kappa or Lambda) concentrations in serum samples via turbidimetry. The system uses polyclonal monospecific sheep anti-human IgG antisera to form insoluble immune complexes with the target antigen; light scatter is measured to determine concentration. The resulting IgG Kappa/IgG Lambda ratio is used by clinicians to monitor disease status in patients previously diagnosed with IgG multiple myeloma. The device is intended for use in clinical laboratories by trained personnel. Results are used in conjunction with other clinical evaluations and laboratory tests (e.g., immunofixation, bone marrow, urine assessments) to assess patient response to therapy. The device provides quantitative data to assist in tracking disease progression or response to treatment, aiding clinical decision-making.
Clinical Evidence
Clinical validity established via comparison of HLC response categorization between the subject device and predicate. Study used 69 monitoring samples from 22 patients (10 IgG Kappa, 12 IgG Lambda) and mathematical modeling of 437 historical samples. Agreement evaluated using Cohen's Kappa (0.79; 95% CI 0.62–0.96) and weighted Kappa (0.87; 95% CI 0.74–0.98). Analytical performance included precision (total CVs 4.0–9.0%), linearity, and interference testing per CLSI guidelines.
Technological Characteristics
Turbidimetric immunoassay; reagents contain polyclonal monospecific sheep anti-human IgG antisera and 0.099% sodium azide preservative. Calibrators are single-level, autodiluted by the analyzer. Traceable to ERM-DA470k/IFCC. Designed for use on the Binding Site Optilite analyzer. Open-vial stability: 3 months at 2-8°C. On-board stability: 28 days.
Indications for Use
Indicated for the quantitative measurement of IgG kappa and IgG lambda in serum to calculate the IgG Kappa/IgG Lambda ratio for monitoring previously diagnosed IgG multiple myeloma in adult patients. For prescription use only.
Regulatory Classification
Identification
An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
Predicate Devices
Hevylite® Human IgG Kappa Kit and Hevylite® Human IgG Lambda Kit (K132555)
Related Devices
K132555 — HEVYLITE HUMAN IGG KAPPA KIT AND HEVYLITE HUMAN IGG LAMBDA KIT · The Binding Site · Dec 20, 2013
K161854 — Hevylite Human IgG Kappa Kit for use on the SPAPLUS, Hevylite Human IgG Lambda Kit for use on the SPAPLUS · The Binding Site Group , Ltd. · Oct 4, 2016
K160819 — Optilite Hevylite IgA Kappa Kit, Optilite Hevylite IgA Lambda Kit · The Binding Site Group , Ltd. · Jun 16, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
February 15, 2018
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The Department of Health & Human Services logo is a stylized depiction of a human figure, while the FDA part includes the acronym "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
The Binding Site Group Ltd. Andrea Thomas Regulatory Affairs Specialist 8 Calthorpe Road, Edgbaston Birmingham, B15 1QT Gb
Re: K172613
Trade/Device Name: Optilite Hevylite IgG Kappa Kit, Optilite Hevylite IgG Lambda Kit Regulation Number: 21 CFR 866.5510 Regulation Name: Immunoglobulins A, G, M, D, and E immunological test system Regulatory Class: Class II Product Code: PCN, PCO Dated: August 16, 2017 Received: August 31, 2017
Dear Andrea Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR
{1}------------------------------------------------
Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kelly Oliner-S
For, Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172613
#### Device Name
Optilite Hevylite IgG Kappa Kit (NK621.OPT) and Optilite Hevylite IgG Lambda Kit (NK622.OPT)
#### Indications for Use (Describe)
The Hevylite IgG Kappa kit is a quantitative in vitro assay intended for the measurement of IgG Kappa (IgG heavy chain and Kappa light chain intact immunoglobulin) in serum using the Optilite analyser. Measurement of Hevylite IgG Kappa is used alongside Hevylite IgG Lambda to calculate the IgG Lambda ratio. The Hevylite IgG Kappa / IgG Lambda ratio can be used when monitoring previously diagnosed IgG multiple myeloma patients and is used in conjunction with other laboratory tests and clinical evaluations. The assignment of complete response is reliant upon other tests including immunofixation, bone marrow and urine assessments.
The Hevylite IgG Lambda kit is a quantitative in vitro assay intended for the measurement of IgG heavy chain and Lambda light chain intact immunoglobulin) in serum using Site Optilite analyser. Measurement of Hevylite IgG Lambda is used alongside Hevylite IgG Kappa to calculate the IgG Kappa / IgG Lambda ratio. The Hevylite IgG Kappa / IgG Lambda ratio can be used when monitoring previously diagnosed IgG multiple myeloma patients and is used in conjunction with other laboratory tests and clinical evaluations. The assignment of complete response is reliant upon other tests including immunofixation, bone marrow and urine assessments.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| | <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--|---------------------------------------------------------------------------------------------------------------------|
| | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Optilite Hevylite Human IgG Kappa and IgG Lambda Kits Submission Summary
- A. 510(k) Number:
K172613
#### B. Purpose for Submission:
Modified IVD assay on previously cleared instrument
#### C. Measurand:
Immunoglobulin IgG Kappa (combined γ heavy and κ light chain) and Immunoglobulin IgG Lambda (combined y heavy and À light chain)
#### D. Type of Test:
Quantitative, Turbidimetry
- E. Applicant: The Binding Site Group, Ltd.
The Binding Site Group, Ltd.
- F. Proprietary and Established Names: Optilite® Hevylite® IgG Kappa Kit Optilite® Hevylite® IgG Lambda Kit
## G. Requlatory Information:
- 1. Regulatory Section: 21 CFR §866.5510, Immunoglobulins A, G, M, D and E Immunological Test System
- 2. Classification: Class II
- 3. Product Code:
PCN - IgG Kappa (Heavy and Light chain combined). Antigen, antiserum, control PCO - IgG Lambda (Heavy and Light chain combined). Antigen, antiserum, control
- 4. Panel:
Immunology (82)
#### H. Intended Use:
- 1. Intended use(s):
Optilite Hevylite IgG Kappa is a quantitative in vitro assay performed on the Optilite analyser for the measurement of IgG kappa (IgG heavy chain and kappa light chain intact immunoglobulin) in serum. Measurement of Hevylite Human IgG Kappa is used alongside Hevylite Human IgG Lambda to calculate the IgG Kappa / IgG Lambda ratio. The Hevylite Human IgG Kappa / IgG Lambda ratio can be used when monitoring previously diagnosed IgG multiple myeloma and is used in conjunction with other laboratory tests and clinical evaluations. The assignment of complete response is reliant upon other tests including immunofixation, bone marrow and urine assessments.
{4}------------------------------------------------
Optilite Hevylite IgG Lambda is a quantitative in vitro assay performed on the Optilite analyser for the measurement of IgG lambda (IgG heavy chain and lambda light chain intact immunoglobulin) in serum. Measurement of Hevylite Human IgG Lambda is used alongside Hevylite Human IgG Kappa to calculate the IgG Kappa / IgG Lambda ratio. The Hevylite Human IgG Kappa / IgG Lambda ratio can be used when monitoring previously diagnosed IgG multiple myeloma and is used in conjunction with other laboratory tests and clinical evaluations. The assignment of complete response is reliant upon other tests including immunofixation, bone marrow and urine assessments.
- 2. Indication(s) for use: Same as intended use.
- 3. Special conditions for use statement(s): This product is for in vitro diagnostic prescription use only.
Warning: The result of Hevylite Human IgG Kappa in a given specimen determined with assays with different manufacturers and different instrument platforms can vary due to differences in assay methods and reagent specificity. The results reported by the laboratory to the physician must include the identity of the Hevylite Human IgG Kappa assay used. Values obtained with different assay methods cannot be used interchangeably. If, in the course of serially monitoring a patient, the assay method used for determining Hevylite IgG Kappa levels is changed, additional sequential testing should be carried out. Prior to changing assays, the laboratory MUST confirm baseline values for patients being serially monitored.
Warning: The result of Hevylite Human IgG Lambda in a given specimen determined with assays with different manufacturers and different instrument platforms can vary due to differences in assay methods and reagent specificity. The results reported by the laboratory to the physician must include the identity of the Hevylite Human IgG Lambda assay used. Values obtained with different assay methods cannot be used interchangeably. If, in the course of serially monitoring a patient, the assay method used for determining Hevylite IgG Lambda levels is changed, additional sequential testing should be carried out. Prior to changing assays, the laboratory MUST confirm baseline values for patients being serially monitored.
- 4. Special instrument requirements: The Binding Site Optilite analyser
#### Device Description: l.
The Hevylite Human IgG Kappa and IgG Lambda Kits contain vials of ready-to-use polyclonal monospecific sheep anti-lgG antisera against combined γ heavy and κ light chain or combined y heavy and À light chain, calibrators (six levels), controls (low and high) and reaction buffer in liquid form. The reagents contain 0.099% sodium azide as preservative.
- J. Substantial Equivalence Information:
{5}------------------------------------------------
- 1. Predicate device names and predicate 510(k) number: Hevylite Human IgG Kappa Kit and Hevylite Human IgG Lambda Kit for use on the Siemens BN™II (k132555)
- 2. Comparison with predicate:
| Similarities | | |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Item | Device | Predicate |
| Intended Use | Quantitative in vitro assay for the<br>measurement of IgG Kappa (IgG<br>heavy chain and kappa light chain<br>intact immunoglobulin) and IgG<br>Lambda (IgG heavy chain and<br>lambda light chain intact<br>immunoglobulin) in serum.<br>Measurement of Hevylite Human IgG<br>Kappa is used alongside Hevylite<br>Human IgG Lambda to calculate the<br>IgG Kappa/IgG Lambda ratio. The<br>Hevylite Human IgG Kappa/IgG<br>Lambda ratio can be used when<br>monitoring previously diagnosed IgG<br>multiple myeloma and is used in<br>conjunction with other laboratory<br>tests and clinical evaluations. The<br>assignment of complete response is<br>reliant upon other tests including<br>immunofixation, bone marrow and<br>urine assessments. | Same |
| Analyte | IgG Kappa and Lambda | Same |
| Antibody | Polyclonal monospecific sheep anti-<br>human combined γ heavy and κ light<br>chain or combined γ heavy and λ<br>light chain. | Same |
| Control | Binding Site High and Low Controls | Same |
| Traceability | DA470k | Same |
| Sample<br>Matrix | Serum | Same |
| Capture<br>Antibody | Sheep anti-human IgG combined | Same |
| Calibrator | Single level Binding Site Hevylite<br>calibrator autodiluted by the analyser<br>to six different concentrations | Same |
| Open Vial<br>Stability | 3 months | Same |
| Reference<br>Interval | IgGK: 4.03 - 9.78 g/L<br>IgGL: 1.97 - 5.71 g/L<br>IgGK/IgGL Ratio: 0.98 - 2.75 | Same |
{6}------------------------------------------------
| Differences | | |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item | Device | Predicate |
| Method | Turbidimetry | Nephelometry |
| Instrument | Binding Site Optilite | Siemens BN™ II Systems |
| On Board<br>Stability | 28 days | Not stated |
| Measuring<br>Range | IgGK: 2.3 – 30.0 g/L (1+19 dilution)<br>IgGL: 1.5 – 17.5 g/L (1+19 dilution) | IgGK: 1.72 – 27.5 g/L (1/100 dilution)<br>IgGL: 0.88 – 14.0 g/L (1/100 dilution) |
| | Extended Range for IgGK:<br>1+0 dilution: 0.115 – 1.5 g/L<br>1+4 dilution: 0.575 – 7.5 g/L<br>1+79 dilution: 9.2 - 120 g/L | Extended Range for IgGK:<br>1/5 dilution: 0.086 – 1.375 g/L<br>1/20 dilution: 0.344 – 5.500 g/L<br>1/400 dilution: 6.88 – 110.0 g/L<br>1/2000 dilution: 34.4 – 550 g/L |
| | Extended Range for IgGL:<br>1+0 dilution: 0.075 – 0.875 g/L<br>1+4 dilution: 0.375 – 4.375 g/L<br>1+79 dilution: 6 - 70 g/L<br>1+119 dilution: 9 – 105 g/L | Extended Range for IgGL:<br>1/5 dilution: 0.044 – 0.700 g/L<br>1/20 dilution: 0.175 – 2.800 g/L<br>1/100 dilution: 0.88 – 14.00 g/L<br>1/400 dilution: 3.50 – 56.0 g/L<br>1/2000 dilution: 17.5 – 280 g/L |
## K. Standard/Guidance Documents Referenced (if applicable):
CLSI EP17-A Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline.
CLSI EP7-A2 Interference Testing in Clinical Chemistry, Approved Guideline -Second Edition.
CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach.
CLSI EP5-A2 Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - Second Edition.
CLSI C28-A3: Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory.
#### L. Test Principle:
Evaluating the concentration of a soluble antigen (e.g. IgG lambda) by turbidimetry involves the addition of the test sample to a solution containing the appropriate antibody (anti-IgG lambda) in a reaction vessel or cuvette. A beam of light is passed through the cuvette and, as the antigen-antibody reaction proceeds, the light passing through the cuvette is scattered increasingly as insoluble immune complexes are formed. Light scatter is monitored by measuring the decrease in intensity of the incident beam of light. The antibody in the cuvette is in excess so the amount of immune complex formed is proportional to the antigen concentration. A series of calibrators of known antigen concentration are assayed initially to produce a calibration curve of measured light scatter versus antigen concentration. Samples of unknown antigen concentration can then be assayed and the results read from the calibration curve.
{7}------------------------------------------------
#### M. Performance Characteristics (if/when applicable):
## 1. Analytical performance:
- a. Precision/Reproducibility:
The within-run, between-run, between-day, between-lot and betweeninstrument precision were determined by testing six serum samples over 21 days with two runs per day on three different reagent lots on three analysers. Results are summarised below.
| Sample | Mean | Within Run | | Between<br>Run | | Between<br>Day | | Between<br>Lot | | Between<br>Instrument | | Total | |
|--------|--------|------------|-----|----------------|-----|----------------|-----|----------------|-----|-----------------------|-----|-------|-----|
| | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 4.260 | 0.184 | 4.3 | 0.209 | 4.9 | 0.162 | 3.8 | 0.140 | 3.3 | 0.033 | 0.8 | 0.323 | 7.6 |
| 2 | 3.148* | 0.048 | 1.5 | 0.104 | 3.3 | 0.124 | 3.9 | 0.098 | 3.1 | 0.036 | 1.1 | 0.169 | 5.4 |
| 3 | 5.601* | 0.145 | 2.6 | 0.263 | 4.7 | 0.308 | 5.5 | 0.205 | 3.7 | 0.099 | 1.8 | 0.430 | 7.7 |
| 4 | 7.621 | 0.178 | 2.3 | 0.263 | 3.5 | 0.413 | 5.4 | 0.426 | 5.6 | 0.216 | 2.8 | 0.521 | 6.8 |
| 5 | 13.574 | 0.218 | 1.6 | 0.385 | 2.8 | 0.879 | 6.5 | 0.851 | 6.3 | 0.253 | 1.9 | 0.984 | 7.2 |
| 6 | 16.824 | 0.383 | 2.3 | 0.352 | 2.1 | 0.992 | 5.9 | 0.991 | 5.9 | 0.608 | 3.6 | 1.120 | 6.7 |
IqGK Precision studies:
| Sample | Mean | Within Run | | Between<br>Run | | Between<br>Day | | Between<br>Lot | | Between<br>Instrument | | Total | |
|--------|--------|------------|-----|----------------|-----|----------------|-----|----------------|-----|-----------------------|-----|-------|-----|
| | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1 | 2.894 | 0.118 | 4.1 | 0.120 | 4.2 | 0.198 | 6.8 | 0.119 | 4.1 | 0.142 | 4.9 | 0.260 | 9.0 |
| 2 | 1.313* | 0.029 | 2.2 | 0.034 | 2.6 | 0.068 | 5.2 | 0.054 | 4.1 | 0.037 | 2.8 | 0.081 | 6.2 |
| 3 | 2.379* | 0.059 | 2.5 | 0.049 | 2.1 | 0.064 | 2.7 | 0.061 | 2.6 | 0.027 | 1.1 | 0.100 | 4.2 |
| 4 | 4.262 | 0.096 | 2.2 | 0.144 | 3.4 | 0.267 | 6.3 | 0.182 | 4.3 | 0.190 | 4.4 | 0.318 | 7.5 |
| 5 | 6.657 | 0.138 | 2.1 | 0.112 | 1.7 | 0.301 | 4.5 | 0.256 | 3.8 | 0.200 | 3.0 | 0.350 | 5.3 |
| 6 | 13.941 | 0.297 | 2.1 | 0.237 | 1.7 | 0.414 | 3.0 | 0.541 | 1.7 | 0.096 | 0.7 | 0.561 | 4.0 |
* sample run at 1+4 dilution.
- b. Linearity/assay reportable range:
A linearity study was performed following CLSI document Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach. The linearity of the IgGK and IgGL assays have been confirmed using serially diluted serum samples to cover the standard measuring ranges of 2.3 - 30.0 q/L and 1.5 - 17.5 g/L respectively. The results demonstrated that the IqGK and IgGL assays are linear over the ranges of 1.930 - 33.427 g/L and 1.380 -19.300 g/L at 1+19 dilution with deviation from linearity ≤ 10%.
- c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability: An Internal Reference material (IR) was assigned by comparison with the Reference Material DA470K.
#### Stability:
Open-vial stability was performed on three lots of Optilite Hevylite IgG Kappa and IgG Lambda kits with testing intervals at various time points up to 112 days. Data supports an open vial stability claim of 3 months at 2-8°C.
On-board stability was performed on three lots of Optilite Hevylite IgG Kappa and IgG Lambda kits with testing intervals at various time points up to 37 days for IgGK and
{8}------------------------------------------------
49 days for IgGL. Data supports an on-board stability claim of 28 days, provided that the power is left switched on as stated in the product insert.
- d. Detection limit:
The analytical sensitivity was determined in accordance with CLSI EP17-A. The Limit of Blank was based on 60 determinations of analyte depleted sample and was estimated at the 95th percentile of the distribution. The Limit of Detection was calculated from the LoB and the combined SD of the 5 LoQ samples. The LoQ was calculated from 5 independent samples (serum samples diluted with analyte depleted serum to achieve a concentration close to the bottom of the measuring range) tested twelve times over five days. The tabulated summary of results is shown below:
| | LoB | LoD | LoQ |
|------------|-----------|-----------|-----------|
| IgG Kappa | 0.000 g/L | 0.009 g/L | 0.115 g/L |
| IgG Lambda | 0.000 g/L | 0.005 g/L | 0.075 g/L |
- e. Analytical specificity:
#### Interference:
Interferences were assessed according to CLSI EP7-A2 by testing three serum samples with different IgG Kappa and IgG Lambda concentration ranges, including a sample close to the lower limit of the reference interval, a sample within the reference interval and an elevated serum sample. Each sample was spiked with interfering substances and tested in multiple replicates. The acceptance criterion was that the mean results from the spiked samples must be within ±10% of the mean of the control samples.
No significant assay interference effects were observed when the samples were tested with bilirubin at 200mg/L, haemoglobin at 5g/L, triglyceride at 1000mg/dL. intralipid at 125mg/dL or the 16 commonly used drugs at the concentrations given below.
| Substance | Concentration |
|------------------------------|--------------------------------------|
| Acetaminophen | 1324µmol/L |
| Acetylsalicylic Acid | 3.63mmol/L |
| Ascorbic Acid | 342µmol/L |
| Bortezomib | 6mg/mL |
| Caffeine | 308µmol/L |
| Cimetidine | 79.2µmol/L |
| Cyclophosphamide Monohydrate | 60µg/mL |
| Digoxin | 3.9nmol/L (IgGK)<br>7.8nmol/L (IgGL) |
| Furosemide | 181µmol/L |
| Ibuprofen | 2425µmol/L |
| Methotrexate | 2mmol/L |
| Penicillin | 75mg/L |
| Phenytoin | 198µmol/L |
| Pomalidomide | 100µg/mL |
| Prednisolone | 100µg/mL |
| Theophylline | 222µmol/L |
{9}------------------------------------------------
#### Cross reactivity:
No significant cross reaction was observed during testing for the predicate device. The specificity of the antisera is unchanged.
## Antigen Excess Detection:
The possibility of antigen excess occurring when using the device on the Binding Site Optilite was evaluated with 8 monoclonal IgG Kappa and 6 monoclonal IgG Lambda samples with concentrations above the respective standard measuring ranges. No antigen excess was observed up to 100.5q/L and 102.5g/L for IgG Kappa and IgG Lambda respectively.
- f. Assay cut-off:
Refer to Expected values.
## 2. Comparison studies:
- a. Method comparison with predicate device:
## lqG Kappa:
A comparison study was performed by analysing 284 serum samples (including 135 IgG Kappa paraprotein and 60 IgG Lambda paraprotein samples, 60 donor samples and 29 other samples, covering the range 0.2 -47.69 g/L) using the Optilite Hevylite IgG Kappa kit and an alternative commercially available assay. Passing Bablok regression analysis generated the following results:
y = 1.10x - 0.68 g/L correlation coefficient r = 0.957 (y = Optilite, x = predicate analyser)
## lgG Lambda:
A comparison study was performed by analysing 172 serum samples (including 42 IqG Kappa paraprotein and 59 IgG Lambda paraprotein samples, 59 donor samples and 1 AL Amyloidosis sample, covering the range 0.09 - 40.65 q/L) using the Optilite Hevvlite IgG Lambda kit and an alternative commercially available assay. Passing Bablok regression analysis generated the following results:
(y = Optilite, x = predicate analyser) y = 0.95x + 0.00 g/L correlation coefficient r = 0.823
#### lgG Kappa/Lambda Ratio:
A comparison study was performed by analysing 143 serum samples (including 39 IgG Kappa paraprotein and 44 IgG Lambda paraprotein samples, 59 donor samples and 1 AL Amyloidosis sample, covering the range 0.01 - 277.50 g/L) using the Optilite Hevylite IgG Kappa and IgG Lambda kits and alternative commercially available assays. Passing Bablok regression analysis generated the following results:
(v = Optilite, x = predicate analyser) v = 1.12x - 0.17 correlation coefficient r = 0.901
{10}------------------------------------------------
- b. Matrix comparison: Not applicable
- 3. Clinical studies:
- a. Clinical Sensitivity/clinical specificity: Transformation of the BNII Study onto the Optilite
#### Purpose of the study
The purpose of this study was to compare the clinical HLC Response categorisation of the predicate and the Optilite Hevylite IgG Kappa and Lambda Kits obtained from samples taken at multiple time points from IgG Kappa and IgG Lambda multiple myeloma patients during the course of their disease. The sponsor generated Passing and Bablok regression equations for the comparison study of the Optilite kits against the predicate kits. The regression equation was then modelled with the existing monitoring sample results from the original BNII submission to evaluate the clinical validity of the new device.
| Complete Response<br>(CR) | HLC ratio within the normal range and negative urine immunofixation. |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Very Good Partial<br>Response (VGPR) | >91% reduction of HLC ratio from baseline and reduction in 24 hour<br>urinary M-protein to ≤100mg per 24 hours |
| Partial Response<br>(PR) | Reduction of HLC ratio from baseline between 47 - 91% and reduction in<br>24 hours urinary M-protein by ≥90% or to ≤200mg/24 hours. |
| Stable Disease (SD) | A change in HLC ratio from baseline < 32% increase but < 47% reduction. |
| Progressive Disease<br>(PD) | > 32% increase in HLC ratio from baseline (the absolute increase in<br>involved IgG must be ≥5g/L) or a ≥25% increase in urine M-protein from<br>baseline (the absolute increase must be ≥200mg/24 hours) |
| Relapse from CR | > 32% increase in HLC ratio from baseline (the absolute increase in<br>involved IgG must be ≥5g/L) |
#### Table 1: HLC Monitoring Response Category
#### Optilite HLC Response Category Study
Assignment of classification was based on the HLC Monitoring Response Category detailed in Table 1, using all assay data available. Responses were cateqorised in accordance with NCCN Guidelines v1.2011 by using the percentage change from baseline. Responses were characterised as progressive disease (PD), stable disease (SD), partial response (PR), very good partial response (VGPR), and complete response (CR). Kappa statistics were used to evaluate agreement between the test and predicate devices.
#### Optilite Monitoring Study Design
A comparison of 69 monitoring samples from 10 IqG Kappa patients and 12 IgG Lambda patients was performed to compare the BNII HLC response category assigned to those observed with the Optilite Hevylite IgG Kappa and Lambda kits (Note: The Optilite monitoring response category is not to be used interchangeably with other manufacturer's assays or with any other instrument platform monitoring response category). The results of the comparison study using 69 monitoring samples vielding 43 response classifications are shown in the table below:
{11}------------------------------------------------
| Observed | | Predicate HLC response | | | | | |
|--------------------------|------|------------------------|----------|-----------|-----------|----------|-----------|
| | | CR | VGPR | PR | SD | PD | Total |
| Optilite HLC Response | CR | 0 | 0 | 0 | 0 | 0 | <b>0</b> |
| | VGPR | 0 | 1 | 0 | 0 | 0 | <b>1</b> |
| | PR | 0 | 0 | 11 | 1 | 0 | <b>12</b> |
| | SD | 0 | 0 | 1 | 23 | 1 | <b>25</b> |
| | PD | 0 | 0 | 0 | 2 | 3 | <b>5</b> |
| Total | | <b>0</b> | <b>1</b> | <b>12</b> | <b>26</b> | <b>4</b> | <b>43</b> |
| Kappa (95% CIs) | | 0.79 (0.62 – 0.96) | | | | | |
| Weighted Kappa (95% CIs) | | 0.87 (0.74 – 0.98) | | | | | |
These monitoring data were also supported by additional statistical regression equation modelling data.
#### Data Modelling:
A data modelling procedure was carried out on monitoring sample results from the original BNII submission. The Passing Bablok regression equations derived from the comparison study and the total imprecision values derived from the precision studies were used to mathematically transform the data. The results are summarised in the tables below.
| H/L* | | Predicate Assigned Response | | | | | |
|------------------------------|------|-----------------------------|------|-----|----|----|-------|
| | | CR | VGPR | PR | SD | PD | Total |
| Transformed<br>Data Response | CR | 29 | 5 | 2 | 0 | 0 | 36 |
| | VGPR | 24 | 72 | 27 | 0 | 0 | 123 |
| | PR | 7 | 4 | 165 | 4 | 0 | 180 |
| | SD | 0 | 1 | 4 | 92 | 0 | 97 |
| | PD | 0 | 0 | 0 | 0 | 1 | 1 |
| Total | | 60 | 82 | 198 | 96 | 1 | 437 |
| Kappa (95% CIs) | | 0.75 (0.70 – 0.80) | | | | | |
| Weighted Kappa (95% CIs) | | 0.86 (0.82 – 0.89) | | | | | |
* H/L: H = highest kappa, L = lowest lambda; the imprecision values were applied to generate the highest possible kappa and the lowest possible lambda results.
| L/H** | | Predicate Assigned Response | | | | | |
|------------------------------|------|-----------------------------|------|-----|----|----|-------|
| | | CR | VGPR | PR | SD | PD | Total |
| Transformed<br>Data Response | CR | 55 | 16 | 6 | 1 | 0 | 78 |
| | VGPR | 5 | 63 | 23 | 0 | 0 | 91 |
| | PR | 0 | 2 | 166 | 3 | 0 | 171 |
| | SD | 0 | 1 | 3 | 92 | 0 | 96 |
| | PD | 0 | 0 | 0 | 0 | 1 | 1 |
| Total | | 60 | 82 | 198 | 96 | 1 | 437 |
| Kappa (95% CIs) | | 0.81 (0.76 - 0.85) | | | | | |
| Weighted Kappa (95% CIs) | | 0.90 (0.86 - 0.93) | | | | | |
** L/H: L = lowest kappa, H = highest lambda; the imprecision values were applied to generate the lowest possible kappa and the highest possible lambda results.
- b. Other clinical supportive data (when a. is not applicable) Not applicable.
{12}------------------------------------------------
# 4. Clinical cut-off:
Refer to discussion above.
- 5. Expected values/Reference range:
The reference intervals were transferred from the predicate devices and were verified by testing 50 adult donor samples.
| | 95 Percentile Range |
|-----------------------------|---------------------|
| IgG kappa (g/L) | 4.03 – 9.78 g/L |
| IgG lambda (g/L) | 1.97 - 5.71 g/L |
| IgG kappa/ IgG lambda ratio | 0.98 - 2.75 |
# N. Proposed Labelling:
The labelling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
## O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.