The EUROIMMUN Anti-PR3-hn-hr ELISA (IgG) test kit is intended for the qualitative or semiquantitative determination of IgG class autoantibodies against proteinase 3 (PR3) in human serum, EDTA plasma, lithium heparin plasma and citrate plasma. It is used as an aid in the differential diagnosis of Wegener's granulomatosis and other autoimmune vasculitides, in conjunction with other laboratory and clinical findings.
Device Story
The EUROIMMUN Anti-PR3-hn-hr ELISA (IgG) is an in vitro diagnostic test kit used in clinical laboratories. It detects IgG autoantibodies against proteinase 3 (PR3) in human serum or plasma samples. The device utilizes an enzyme-linked immunosorbent assay (ELISA) principle. Samples are incubated in microplate wells coated with PR3 antigen; if specific antibodies are present, they bind to the antigen. After washing, enzyme-labeled anti-human IgG is added, followed by a substrate to produce a colorimetric signal proportional to the antibody concentration. Results are interpreted by laboratory professionals to assist in the differential diagnosis of Wegener's granulomatosis and other autoimmune vasculitides. The output provides clinicians with diagnostic information to be used in conjunction with other clinical and laboratory findings.
Clinical Evidence
Clinical performance was evaluated using 1,177 samples, including patients with Wegener’s granulomatosis (WG), ANCA-associated vasculitides (AAV), other autoimmune diseases (SLE, SS, RA), and asymptomatic blood donors. The device demonstrated a clinical sensitivity of 93.8% (152/162) and a clinical specificity of 99.4% (1009/1015). Method comparison against the predicate device across 1,255 samples showed a positive percent agreement of 99.4% and a negative percent agreement of 96.7%.
Technological Characteristics
ELISA-based immunoassay. Microplate wells coated with PR3 antigen. Detection via enzyme-labeled anti-human IgG conjugate and colorimetric substrate. Quantitative/semi-quantitative output via spectrophotometric measurement. In vitro diagnostic use.
Indications for Use
Indicated for qualitative or semiquantitative detection of IgG autoantibodies against proteinase 3 (PR3) in human serum or plasma (EDTA, lithium heparin, citrate) to aid in the differential diagnosis of Wegener's granulomatosis and other autoimmune vasculitides.
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
K974702 — ORGENTEC ANTI-PR3 (C-ANCA) ELISA · American Laboratory Products Co., Ltd. · Feb 17, 1998
K983740 — IS-ANTI-PR3 IGG ELISA TEST SYSTEM · Columbia Bioscience, Inc. · Nov 18, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its head turned to the left. The eagle is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
EUROIMMUN US, LLC c/o Ms. Kathryn Kohl Managing Director 95 Washington St Morristown, NJ 07960
APR - 7 2009
Re: k082130
Trade/Device Name: EUROIMMUN Anti-PR3-hn-hr ELISA (IgG) Regulation Number: 21 CFR 866.5660 Regulation Name: Multiple autoantibodies immunological test systems Regulatory Class: Class II Product Code: MOB Dated: February 19, 2009 Received: February 23, 2009
Dear Ms. Kohl,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket
{1}------------------------------------------------
Page 2 – Ms. Kathryn Kohl
notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
m chan
Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
510(k) Number (if known): K082130
Anti-PR3-hn-hr ELISA (IgG) Device Name:
Indications For Use:
The EUROIMMUN Anti-PR3-hn-hr ELISA (IgG) test kit is intended for the qualitative or semiquantitative determination of IgG class autoantibodies against proteinase 3 (PR3) in human serum, EDTA plasma, lithium heparin plasma and citrate plasma. It is used as an aid in the differential diagnosis of Wegener's granulomatosis and other autoimmune vasculitides, in conjunction with other laboratory and clinical findings.
Prescription Use _ X Use (Part 21 CFR 801 Subpart D) C)
AND/OR
Over-The-Counter
(21 CFR 801 Subpart
日に、じて、モディスタートで、レディア・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
Cases of the state and the state and the states
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (OIVD)
Kaven K. Smaaff
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.