REAADS II ANTI-CARDIOLIPIN IGG SEMI-QUANTITATIVE TEST KIT, MODEL 11139

K032868 · Corgenix, Inc. · MID · Oct 14, 2003 · Immunology

Device Facts

Record IDK032868
Device NameREAADS II ANTI-CARDIOLIPIN IGG SEMI-QUANTITATIVE TEST KIT, MODEL 11139
ApplicantCorgenix, Inc.
Product CodeMID · Immunology
Decision DateOct 14, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 866.5660
Device ClassClass 2

Intended Use

The REAADS® II Anti-Cardiolipin IgM Semi-Quantitative Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative determination of anti-cardiolipin IgM antibodies in human serum or plasma. For the detection and semi-quantitation of anti-cardiolipin antibodies in individuals with systemic lupus erythematosus (SLE) and lupus-like disorders (anti-phospholipid syndrome). The REAADS® II Anti-Cardiolipin IgM Semi-Quantitative Test Kit is intended to be used by clinical (hospital and reference) laboratories.

Device Story

REAADS test kits for anti-Cardiolipin (IgG/IgM/IgA), anti-Phosphatidylserine (IgG/IgM/IgA), anti-Beta 2 Glycoprotein I (IgG/IgM/IgA), and anti-Prothrombin (IgG/IgM) antibodies; modified to increase assay incubation periods (from 15/15/10 minutes to 30-40 minutes per step) and allow air blanking for plate readers; combined IgG/IgM kits separated into individual isotype kits; intended for semi-quantitative detection of specific antibodies; used in clinical laboratory settings; results aid clinicians in diagnosis/management of autoimmune conditions; modifications verified via design control activities and risk analysis to ensure fundamental scientific technology remains unchanged.

Clinical Evidence

No clinical data provided in the document; bench testing only.

Technological Characteristics

Enzyme-linked immunosorbent assay (ELISA) technology. Form factor: test kit for laboratory use. Analyte detection: semi-quantitative determination of specific autoantibodies. Sample types: human serum or citrated plasma (3.2% sodium citrate).

Indications for Use

Indicated for detection and semi-quantitation of anti-cardiolipin, anti-phosphatidylserine, anti-beta 2 glycoprotein I, and anti-prothrombin (IgG, IgM, IgA) antibodies in human serum or citrated plasma. Target population: individuals with systemic lupus erythematosus (SLE) and lupus-like disorders (anti-phospholipid syndrome).

Regulatory Classification

Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three heads facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Corgenix, Inc. c/o Ms. Nanci Dexter Director of Quality and Regulatory Affairs 12061 Tejon Street Westminster, Colorado 80234 OCT 2 0 2003 Re: k032868 Device Name: REAADS II anti-Cardiolipin IgG Semi-quantitative Test Kit REAADS II anti-Cardiolipin IgM Semi-quantitative Test Kit REAADS II anti-Cardiolipin IgA Semi-quantitative Test Kit REAADS II anti-Phosphatidylserine IgG Semiquantitative Test Kit REAADS II anti-Phosphatidylserine IgM Semiquantitative Test Kit REAADS II anti-Phosphatidylserine IgA Semiquantitative Test Kit REAADS II anti-Beta 2 Glycoprotein I IgG Semi-Quantitative Test Kit REAADS II anti-Beta 2 Glycoprotein I IgM Semi-Quantitative Test Kit REAADS II anti-Beta 2 Glycoprotein I IgA Semi-Quantitative Test Kit REAADS II anti-Prothrombin IgG Semi-Quantitative Test Kit REAADS II anti-Prothrombin IgM Semi-Quantitative Test Kit Regulation Number: 21 CFR § 866.5660 Regulation Name: Multiple Autoantibodies Immunological Test System Regulatory Class: II Product Code: MID, MSV, DHC Dated: September 12, 2003 Received: September 15, 2003 Dear Ms. Dexter: This letter corrects our substantially equivalent letter of October 14, 2003 regarding the incorrect company name. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ ## Page 2 - Ms. Nanci Dexter Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Butman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health {2}------------------------------------------------ ## 510(k) Number: K032868 #### Device Name: REAADS® II Anti-Cardiolipin IgM Semi-Quantitative Test Kit #### Indications for Use: The REAADS® II Anti-Cardiolipin IgM Semi-Quantitative Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative determination of anti-cardiolipin IgM antibodies in human serum or plasma. For the detection and semi-quantitation of anti-cardiolipin antibodies in individuals with systemic lupus erythematosus (SLE) and lupus-like disorders (anti-phospholipid syndrome). The REAADS® II Anti-Cardiolipin IgM Semi-Quantitative Test Kit is intended to be used by clinical (hospital and reference) laboratories. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) J. P. Rews for T. J. O'Leary Division Sign-Off 510(k) K032868 Prescription Use \$\underline{V}\$ {3}------------------------------------------------ ## 510(k) Number: K032868 ### Device Name: REAADS® II Anti-Cardiolipin IgG Semi-Quantitative Test Kit Indications for Use: The REAADS® II Anti-Cardiolipin IgG Semi-Quantitative Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative determination of anti-cardiolipin IgG antibodies in human serum or plasma. For the detection and semi-quantitation of anti-cardiolipin antibodies in individuals with systemic lupus erythematosus (SLE) and lupus-like disorders (anti-phospholipid syndrome). The REAADS® II Anti-Cardiolipin IgG Semi-Quantitative Test Kit is intended to be used by clinical (hospital and reference) laboratories. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) J.P. Reeves for T.J. Wilson/ Division Sign-Off 510(k) K032868 Prescription Use \$\checkmark\$ {4}------------------------------------------------ # 510(k) Number: K032868 ### Device Name: REAADS® II Anti-Cardiolipin IgA Semi-Quantitative Test Kit #### Indications for Use: The REAADS® II Anti-Cardiolipin IgA Semi-Quantitative Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative determination of anti-cardiolipin IgA antibodies in human serum or plasma. For the detection and semi-quantitation of anti-cardiolipin antibodies in individuals with systemic lupus erythematosus (SLE) and lupus-like disorders (anti-phospholipid syndrome). The REAADS® II Anti-Cardiolipin IgA Semi-Quantitative Test Kit is intended to be used by clinical (hospital and reference) laboratories. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) E.J.P.Leves for P.J.O'Leary Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k)_KO32868 Prescription Use ✓ {5}------------------------------------------------ ## 510(k) Number: K032868 #### Device Name: REAADS® II Anti-Phosphatidylserine IgG Semi-Quantitative Test Kit #### Indications for Use: The REAADS II Anti-Phosphatidylserine IgG Semi-Quantitative Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative determination of anti-phosphatidylserine IgG antibodies in human serum or citrated plasma (3.2% sodium citrate). Detection and semi-quantitation of anti-phosphatidylserine antibodies in individuals with systemic lupus erythematosus (SLE) and lupus-like disorders (anti-phospholipid syndrome). The REAADS® II Anti-Phosphatidylserine IgG Semi-Quantitative Test Kit is intended to be used by clinical (hospital and reference) laboratories. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) A.P. Reeves for D.L. U'Hearn Division Sign-Off. 510(k) K032868 Prescription Use \$\checkmark\$ {6}------------------------------------------------ ### 510(k) Number: K032868 #### Device Name: REAADS® II Anti-Phosphatidylserine IgM Semi-Quantitative Test Kit #### Indications for Use: The REAADS II Anti-Phosphatidylserine IgM Semi-Quantitative Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative determination of anti-phosphatidy/serine IgM antibodies in human serum or citrated plasma (3.2% sodium citrate). Detection and semi-quantitation of anti-phosphatidylserine antibodies in individuals with systemic lupus erythematosus (SLE) and lupus-like disorders (anti-phospholipid syndrome). The REAADS® II Anti-Phosphatidylserine IgM Semi-Quantitative Test Kit is intended to be used by clinical (hospital and reference) laboratories. # (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) AFReeve for T.J. O'leary Division Sign-Off **Division Sign-Off** 510(k) K032868 Prescription. Use \$\underline{ \checkmark }\$ {7}------------------------------------------------ ### 510(k) Number: K032868 ### Device Name: REAADS® II Anti-Phosphatidylserine IgA Semi-Quantitative Test Kit #### Indications for Use: The REAADS® II Anti-Phosphatidylserine IgA Semi-Quantitative Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative determination of anti-phosphatidylserine (aPS) IgA antibodies in human serum and citrated plasma (3.2% sodium citrate). Detection and semi-quantitation of anti-phosphatidylserine antibodies in individuals with systemic lupus erythematosus (SLE) and lupus-like disorders (anti-phospholipid syndrome). The REAADS® II Anti-Phosphatidylserine IgA Semi-Quantitative Test Kit is intended to be used by clinical (hospital and reference) laboratories. # (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) J.P. Reeves for T.J. O'Leary Division Sign-Off 510(k) K032868 Prescription Use \$\underline{ \checkmark }\$ {8}------------------------------------------------ ## 510(k) Number: K032868 ### Device Name: REAADS® II Anti-Beta 2 Glycoprotein I IgG Semi-Quantitative Test Kit #### Indications for Use: The REAADS® II Anti-Beta 2 Glycoprotein I IgG Semi-Quantitative Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative determination of anti-Beta 2 Glycoprotein 1 (B2GP1) IgG antibodies in human serum or citrated plasma (3.2% sodium citrate). For the detection and semi-quantitation of anti-ß2GPI IgG antibodies in individuals with systemic lupus erythematosus (SLE) and lupus-like disorders (anti-phospholipid syndrome). The REAADS® II Anti-Beta 2 Glycoprotein I IgG Semi-Quantitative Test Kit is intended to be used by clinical (hospital and reference) laboratories. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) A.P. Reeves for T.J. O'leary Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) KO32868 Prescription Use √ {9}------------------------------------------------ ### 510(k) Number: K032868 ## Device Name: REAADS® II Anti-Beta 2 Glycoprotein I IgM Semi-Quantitative Test Kit #### Indications for Use: The REAADS® II Anti-Beta 2 Glycoprotein I IgM Semi-Quantitative Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative determination of anti-Beta 2 Glycoprotein l (ß2GPI) IgM antibodies in human serum or citrated plasma (3.2% sodium citrate). For the detection and semi-quantitation of antibodies in individuals with systemic lupus erythematosus (SLE) and lupus-like disorders (anti-phospholipid syndrome). The REAADS® II Anti-Beta 2 Glycoprotein I IgM Semi-Quantitative Test Kit is intended to be used by clinical (hospital and reference) laboratories. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) A. Reus for T.J. O'leary Division Sign-Off 510(k) K032868 Prescription Use ✓ {10}------------------------------------------------ ## 510(k) Number: K032868 ## Device Name: REAADS® II Anti-Beta 2 Glycoprotein 1 IgA Semi-Quantitative Test Kit #### Indications for Use: The REAADS® II Anti-Beta 2 Glycoprotein I IgA Semi-Quantitative Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative determination of anti-Beta 2 Glycoprotein 1 (B2GPI) IgA antibodies in human serum or citrated plasma (3.2% sodium citrate). For the detection and semi-quantitation of antibodies in individuals with systemic lupus erythematosus (SLE) and lupus-like disorders (anti-phospholipid syndrome). The REAADS® II Anti-Beta 2 Glycoprotein I IgA Semi-Quantitative Test Kit is intended to be used by clinical (hospital and reference) laboratories. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) A.L. Reeves for T.J. O'leary Division Sign-Off 510(k) K032868 Prescription Use \/ {11}------------------------------------------------ ### 510(k) Number: K032868 ### Device Name: REAADS® II Anti-Prothrombin IgG Semi-Quantitative Test Kit #### Indications for Use: The REAADS® II Anti-Prothrombin IgG Semi-Quantitative Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative determination of anti-prothrombin (aPT) IgG antibodies in human serum or citrated plasma (3.2% sodium citrate). For the detection and semi-quantitation of anti-prothrombin (aPT) IgG antibodies with systemic lupus erythematosus (SLE) and lupus-like disorders (e.g., antiphospholipid syndrome). The REAADS® II Anti- Prothrombin IgG Semi-Quantitative Test Kit is intended to be used by clinical (hospital and reference) laboratories. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) J.P. Reeves for T.J. d'Leary Division Sign-Off 510(k) K032868 Prescription Use ✓ {12}------------------------------------------------ # 510(k) Number: K032868 ## Device Name: REAADS® II Anti-Prothrombin IgM Semi-Quantitative Test Kit #### Indications for Use: The REAADS® II Anti-Prothrombin IgM Semi-Quantitative Test Kit is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative determination of anti-prothrombin (aPT) IgM antibodies in human serum or citrated plasma (3.2% sodium citrate). For the detection and semi-quantitation of anti-prothrombin (aPT) IgM antibodies in individuals with systemic lupus erythematosus (SLE) and lupus-like disorders (e.g., antiphospholipid syndrome). The REAADS® II Anti- Prothrombin IgM Semi-Quantitative Test Kit is intended to be used by clinical (hospital and reference) laboratories. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) J.P. Reeves for A.J. O'Leary Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K032868 Preseription Use I
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...