DIMENSION VISTA BETA 2 MICROGLOBULIN, DIMENSION VISTA PROTEIN 1 CALIBRATOR, DIMENSION VISTA PROTEIN 1 CONTROL M

K083463 · Siemens Healthcare Diagnostics · JZG · Mar 13, 2009 · Immunology

Device Facts

Record IDK083463
Device NameDIMENSION VISTA BETA 2 MICROGLOBULIN, DIMENSION VISTA PROTEIN 1 CALIBRATOR, DIMENSION VISTA PROTEIN 1 CONTROL M
ApplicantSiemens Healthcare Diagnostics
Product CodeJZG · Immunology
Decision DateMar 13, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5630
Device ClassClass 2

Intended Use

The B2MIC method is an in vitro diagnostic test for the quantitative measurement of ß2-microglobulin in human serum, heparinized plasma, EDTA plasma and urine using the B2MIC Flex® reagent cartridge on the Dimension Vista® Systems. Measurement of ß 2 -microglobulin aids in the diagnosis of active rheumatoid arthritis and kidney disease. PROT1 CAL is an in vitro diagnostic product for the calibration of the Dimension Vista® Systems for: αγ-Acid Glycoprotein (A1AG), α1-Antitrypsin (A1AT), a2-macroglobulin (A2MAC), IS2-Microglobulin (B2MIC, B2MU**), C3 Complement (C3), C4 Complement (C4), Ceruloplasmin (CER), Haptoglobin (HAPT),Hemopexin (HPX), Homocysteine (HCYS), Immunoglobulin A (IGA), Immunoglobulin E (IGE), Immunoglobulin G (IGG, IGG-C*, ICC-U**), Immunoglobulin G subclass 1(IGG1), Immunoglobulin G subclass 2 (IGG2), Immunoglobulin G subclass 3 (IGG3), Immunoglobulin G subclass 4 (IGG4), Immunoglobulin M (IGM), Prealbumin (PREALB), Retinol Binding Protein (RBP), soluble Transferrin Receptor (STFR), Transferrin (TRF). PROT1 CON L is an assayed, low level, intralaboratory quality control for assessment of precision and analytical bias on the Dimension Vista® Systems in the quantitative determination of: α-Acid Glycoprotein (A1AG), α-- Antitrypsin (A1AT), a2 -- Macroglobulin (A2MAC), ß2-Microglobulin (B2MIC-U **), C3 Complement (C3), C4 Complement (C4), Ceruloplasmin (CER), Haptoglobin (HAPT), Hemopexin (HPX), Homocysteine (HCYS), Immunoglobulin A (IGA), Immunoglobulin E (IGE), Immunoglobulin G (IGG),Immunoglobulin G Subclass , immunoglobulin G subclass 2, immunoglobulin G subclass 3, immunoglobulin G subclass 4, immunoglobulin M, prealburnin, retinol binding protein, homocysteine, soluble transferrin receptor and transferrin. PROT1 CON M is an assayed, mid-level, intralaboratory quality controls for assessment of precision and analytical bias on the Dimension Vista® System in the quantitative determination of: : α -- Acid Glycoprotein (A1AG), α -- Antitrypsin (A1AT), α 2 -- Macroglobulin (A2MAC), ß2-Microglobulin (B2MIC*,B2MIC-U**), C3 Complement (C3), C4 Complement (C4), Ceruloplasmin (CER), Haptoglobin (HAPT), Hemopexin (HPX), Homocysteine (HCYS), Immunoglobulin A (IGA), Immunoglobulin E (IGE), Immunoglobulin G (IGG), Immunoglobulin G subclass 1 (IGG1), lmmunoqlobulin G subclass 2 (IGG2), Immunoglobulin G subclass 3 (IGG3), Immunoglobulin G subclass 4 (IGG4), Immunoglobulin M (IGM), Prealbumin (PREALB), Retinol Binding Protein (RBP), specialty Albumin (sALB*), soluble Transferrin Receptor (STFR) and Transferrin (TRF).

Device Story

Dimension Vista® B2MIC Flex® reagent cartridge performs in vitro quantitative measurement of ß2-microglobulin in human serum, plasma, and urine. Principle of operation: immunochemical reaction between proteins in body fluids and specific antibodies forms immune complexes; intensity of scattered light is proportional to protein concentration. Result evaluated by comparison with known standard. Used in clinical laboratory settings by trained personnel. Output provides quantitative protein concentration to aid clinicians in diagnosing active rheumatoid arthritis and kidney disease. System includes multi-analyte liquid human serum-based calibrators and low/mid-level quality controls to ensure analytical precision and bias assessment.

Clinical Evidence

Method comparison study performed comparing Dimension Vista® B2MIC assay to N Latex ß2-microglobulin on BN ProSpec® System. Evaluated 82 urine samples with concentrations ranging from 0.022 – 0.553 mg/dL. Regression analysis yielded slope of 0.952, intercept of -0.0008 mg/dL, and correlation coefficient of 0.988.

Technological Characteristics

Immunonephelometry assay; reagents include phosphate buffer, polyethylene glycol sorbitan monolaureate, polystyrene particles, and mouse monoclonal antibodies; 840 nm light source; liquid ready-to-use format; requires Dimension Vista® System; traceable to WHO 1st International Standard for β2-microglobulin; storage 2-8°C.

Indications for Use

Indicated for quantitative measurement of β2-microglobulin in human serum, heparinized plasma, EDTA plasma, and urine to aid in the diagnosis of active rheumatoid arthritis and kidney disease.

Regulatory Classification

Identification

A beta-2-microglobulin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques beta-2-microglobulin (a protein molecule) in serum, urine, and other body fluids. Measurement of beta-2-microglobulin aids in the diagnosis of active rheumatoid arthritis and kidney disease.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 866.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Ko 83463 **MAR. 1, 2009** #### 510(k) Summary for Dimension Vista® B2MIC Flex® reagent cartridge Dimension Vista® Protein 1 Calibrator Dimension Vista® Protein 1 Control L and M This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. 1. Manufacturer's Name, Address, Telephone, and Contact Person, Date of Preparation: Manufacturer: Siemens Healthcare Diagnostics Products Emil-von-Behring Str. 76 35041 Marburg, Germany Contact Information: Siemens Healthcare Diagnostics. 500 GBC Drive, M/S 514 Newark, Delaware 19702 Attn: Anna Marie Kathleen Ennis Tel: 302-632-9352 Fax: 302-631-6299 Preparation date: February 4, 2009 2. Device Name: Dimension Vista® B2MIC Flex® reagent cartridge Dimension Vista® PROT 1 CAL Dimension Vista® PROT 1 CON L and M Classification: Class II; Class II; Class I Product Code: JZG, JIX, JJY Panel: Immunology (82) and Clinical Chemistry (75) ### 3. Identification of the Legally Marketed Devices: Siemens N Latex ß2- microglobulin - K002731 {1}------------------------------------------------ #### N Protein Standard SL - K052788 N/T Protein Controls SL - K052788 #### 4. Device Descriptions: #### Dimension Vista® B2MIC Flex® reagent cartridge Proteins contained in human body fluids form immune complexes in an immunochemical reaction with specific antibodies. These complexes scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration. #### Dimension Vista® Protein 1 Calibrator Protein 1 Calibrator is a multi-analyte, liquid human serum based product containing: a+acid glycoprotein, a1 -antitrypsin, a1-macroglobulin, 132 -microglobulin, C3 complement, C4 complement, ceruloplasmin, haptoglobin, hemopexin, immunoglobulin A, immunoglobulin E, immunoglobulin G, immunoglobulin G subclass 1, immunoglobulin G subclass 2, immunoglobulin G subclass 3, immunodlobulin G subclass 4, immunoglobulin M, prealbumin, retinol binding protein, homocysteine, soluble transferrin receptor and transferrin. #### Dimension Vista® Protein 1 Control L Protein 1 Control L is a multi-analyte, low level liquid human serum based product containing : q-acid glycoprotein, α-- antitrypsin, α 2-macroglobulin, ß-microglobulin, C3 complement, C4 complement,ceruloplasmin, haptoglobin, hemopexin,immunoglobulin E, immunoglobulin A, immunoglobulin G, immunoqlobulin G Subclass , immunoglobulin G subclass 2, immunoglobulin G subclass 3, immunoglobulin G subclass 4, immunoglobulin M, prealburnin, retinol binding protein, homocysteine, soluble transferrin receptor and transferrin #### Dimension Vista® Protein 1 Control M Protein 1 Control M is a multi-analyte, mid level, liquid human serum based product containing: Q -- acid glycoprotein, a -- antitrypsin, as -macroglobulin, 13-microglobulin, C3 complement, C4 complement, ceruloplasmin, haptoglobin, hemopexin, immunoglobulin A, immunoglobulin E, immunoglobulin G, immunoglobulin G subclass 1, immunoglobulin G subclass 2, immunoglobulin G subclass 3. immunodobulin G subclass 4. immunoglobulin M, prealbumin, retinol binding protein, homocysteine, soluble transferrin receptor, and transferrin. #### 5. Device Intended Uses #### Dimension Vista® B2MIC Flex® reagent cartridge: The B2MU method is an in vitro diagnostic test for the quantitative measurement of ß₂microglobulin in human serum, heparinized plasma, EDTA plasma and urine using the B2MIC Flex® reagent cartridge on the Dimension Vista® Systems. Measurement of ß 2 -microglobulin aids in the diagnosis of active rheumatoid arthritis and kidney disease. #### Dimension Vista® PROT 1 CAL: {2}------------------------------------------------ PROT1 CAL is an in vitro diagnostic product for the calibration of the Dimension Vista® Systems for: αγ-Acid Glycoprotein (A1AG), α1-Antitrypsin (A1AT), a2-macroglobulin (A2MAC), IS2-Microglobulin (B2MIC, B2MU**), C3 Complement (C3), C4 Complement (C4), Ceruloplasmin (CER), Haptoglobin (HAPT),Hemopexin (HPX), Homocysteine (HCYS), Immunoglobulin A (IGA), Immunoglobulin E (IGE), Immunoglobulin G (IGG, IGG-C*, ICC-U**), Immunoglobulin G subclass 1(IGG1), Immunoglobulin G subclass 2 (IGG2), Immunoglobulin G subclass 3 (IGG3), Immunoglobulin G subclass 4 (IGG4), Immunoglobulin M (IGM), Prealbumin (PREALB), Retinol Binding Protein (RBP), soluble Transferrin Receptor (STFR), Transferrin (TRF) *For cerebrospinal fluid ** For urine #### Dimension Vista® Protein 1 Control L PROT1 CON L is an assayed, low level, intralaboratory quality control for assessment of precision and analytical bias on the Dimension Vista® Systems in the quantitative determination of: α-Acid Glycoprotein (A1AG), α-- Antitrypsin (A1AT), a2 -- Macroglobulin (A2MAC), ß2-Microglobulin (B2MIC-U **), C3 Complement (C3), C4 Complement (C4), Ceruloplasmin (CER), Haptoglobin (HAPT), Hemopexin (HPX), Homocysteine (HCYS), Immunoglobulin A (IGA), Immunoglobulin E (IGE), Immunoglobulin G (IGG),Immunoglobulin G subclass 1 (IGG1), Immunoglobulin G subclass 2 (IGG2), Immunoglobulin G subclass 3 (IGG3), Immunoglobulin G subclass 4 (IGG4), Immunodlobulin M (IGM), Prealbumin (PREALB), Retinol Binding Protein (RBP), specialty Albumin (sALB*), soluble Transferrin Receptor (STFR) and Transferrin (TRF). *For serum and plasma ** For Urine #### Dimension Vista® Protein 1 Control M PROT1 CON M is an assayed, mid-level, intralaboratory quality controls for assessment of precision and analytical bias on the Dimension Vista® System in the quantitative determination of: : α -- Acid Glycoprotein (A1AG), α -- Antitrypsin (A1AT), α 2 -- Macroglobulin (A2MAC), ß2-Microglobulin (B2MIC*,B2MIC-U**), C3 Complement (C3), C4 Complement (C4), Ceruloplasmin (CER), Haptoglobin (HAPT), Hemopexin (HPX), Homocysteine (HCYS), Immunoglobulin A (IGA), Immunoglobulin E (IGE), Immunoglobulin G (IGG), Immunoglobulin G subclass 1 (IGG1), lmmunoqlobulin G subclass 2 (IGG2), Immunoglobulin G subclass 3 (IGG3), Immunoglobulin G subclass 4 (IGG4), Immunoglobulin M (IGM), Prealbumin (PREALB), Retinol Binding Protein (RBP), specialty Albumin (sALB*), soluble Transferrin Receptor (STFR) and Transferrin (TRF), *For serum and plasma ** For urine ### 6. Medical device to which equivalence is claimed and comparison information: The Dimension Vista® B2MJC Flex® reagent cartridge, Dimension Vista® PROT 1 CAL and Dimension Vista® PROT 1 CON L and M, are substantially equivalent to the Siemens N Latex ßz-microglobulin assay (K002731), N/T Protein Standard SL (K052788) and N Protein Controls SL (K052788). The Dimension Vista® B2MIC assay, like Siemens N Latex (1 2-microglobulin assay is an in vitro diagnostic test for the quantitative measurement of ß2 -microglobulin in human serum, plasma and urine. #### 7. Device Performance Characteristics: {3}------------------------------------------------ ### Method Comparison Study The Dimension Vista® B2MU assay was compared to the N Latex to Human ß ₂-microglobulin on the BN ProSpec® System by evaluating urine samples with concentrations ranging from 0.022 – 0.553 mg/dL (0.223 – 5.53mg/L). Regression analysis of these results yielded the following equation: | Comparative<br>Method | Slope | Intercept<br>mg/dL (mg/L) | Correlation<br>Coefficient | n | |------------------------------------|-------|---------------------------|----------------------------|----| | β₂-microglobulin<br>on BN Prospec® | 0.952 | -0.0008 (-0.008) | 0.988 | 82 | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The eagle faces to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Seimens Healthcare Diagnostics Inc. c/o Ms. Anna Marie Kathleen Ennis Senior Regulatory Affairs and Compliance Specialist 500 GBC Drive P.O. Box 601 Newark, DE 19714-6101 US MAR 1 3 2009 Re: k083463 Trade/Device Name: Dimension Vista® B2MIC Flex® reagent cartridge Dimension Vista® Protein 1 Calibrator Dimension Vista® Protein 1 Control L Dimension Vista® Protein 1 Control M Regulation Number: 21 CFR 866.5630 Regulation Name: Beta-2-microglobulin immunological test system Regulatory Class: Class II Product Code: JZG, JIX, JJY Dated: February 4, 2009 Received: February 5, 2009 Dear Ms. Ennis, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {5}------------------------------------------------ Page 2 - Ms. Anna Marie Kathleen Ennis comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. ' mchen Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ### Indication for Use 510(k) Number (if known): KO83463 Device Name: Dimension Vista® B2MIC Flex Reagent Cartridge Indication for Use: The B2MIC method is an in vitro diagnostic test for the quantitative measurement of ß2-microglobulin in human serum, heparinized plasma, EDTA plasma and urine on the Dimension Vista® System. Measurements of beta-2-microglobulin, aid in the diagnosis of active rheumatoid arthritis and kidney disease. Prescription Use _ X And/Or Over the Counter Use (21 CFR Part 801 (21 CFR Part 801 Subpart D) Subpart C) (Please Do Not WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OMD)--- Maria In Chain Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) Kofs463 Page 1 of 1 {7}------------------------------------------------ ## Indications for Use # 510(k) Number (if known): Kod3 4 63 #### Dimension Vista® PROT 1 CON L Device Name: ### Indications For Use: PROT1 CON L is an assayed, low-level, intralaboratory quality control for the assessment of precision and analytical bias on the Dimension Vista® System in the quantitative determination of: | α₁-Acid Glycoprotein (A1AG) | Immunoglobulin G (IGG) | |----------------------------------|-------------------------------------| | α₁-Antitrypsin (A1AT) | Immunoglobulin G Subclass 1 (IGG1) | | α₂-Macroglobulin (A2MAC) | Immunoglobulin G Subclass 2 (IGG2) | | β₂-Microglobulin (B2MIC, B2MU**) | Immunoglobulin G Subclass 3 (IGG3) | | C3 Complement (C3) | Immunoglobulin G Subclass 4 (IGG4) | | C4 Complement (C4) | Immunoglobulin M (IGM) | | Ceruloplasmin (CER) | Prealbumin (PREALB) | | Haptoglobin (HAPT) | Retinol binding Protein (RBP) | | Hemopexin (HPX) | soluble Transferrin Receptor (STFR) | | Homocysteine (HCYS) | specialty Albumin (sALB*) | | Immunoglobulin A (IGA) | Transferrin (TRF) | | Immunoglobulin E (IGE) | | * For serum and plasma ** For urine Prescription Use _ × (Per 21 CFR 801 Subpart D) Over-The-Counter-Use_ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) AND/OR Concurrence of CDRH, Office of Device Evaluation Page 1 of Maria M'Chen. Division Slan- Office of In Vitro Diagnostic Device Evaluation and Safety 5100 {8}------------------------------------------------ ### Indications for Use 510(k) Number (if known): Ko 8 346 3 Device Name: * For Urine Dimension Vista® PROT 1 CAL Indications For Use: ### Dimension Vista® PROT 1 CAL: PROT1 CAL is an in vitro diagnostic product for the calibration of the Dimension Vista® System for: a1-Acid Glycoprotein (A1AG) a1-Antitrypsin (A1AT) a2- Macroglobulin (A2MAC) B2-Microglobulin (B2MIC, B2MU**) C3 Complement (C3) C4 Complement (C4) Ceruloplasmin (CER) Haptoglobin (HAPT) Hemopexin (HPX) Homocysteine (HCYS) Immunoglobulin A (IGA) * For cerebrospinal fluid (CSF) Immunoglobulin E (IGE) Immunoqlobulin G {IGG, IGG-C*, IGG-U**) Immunoglobulin G Subclass 1 (IGG1) Immunoglobulin G Subclass 2 (IGG2) Immunoglobulin G Subclass 3 (IGG3) Immunoglobulin G Subclass 4 (IGG4) lmmunoglobulin M (IGM) Prealbumin (PREALB) Retinol binding Protein (RBP) soluble Transferrin Receptor (STFR) Transferrin (TRF) Prescription Use × (Per 21 CFR 801 Subpart D) Over-The-Counter-Use AND/OR (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation . Page 1 of ia m chan **Division Sign-Off** Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K083463 {9}------------------------------------------------ ### Indications for Use # 510(k) Number (if known): Ko83463 #### Dimension Vista® PROT 1 CON M Device Name: #### Indications For Use: PROT1 CON M is an assayed, mid-level, intralaboratory quality control for the assessment of precision and analytical bias on the Dimension Vista® System in the quantitative determination of: | a₁-Acid Glycoprotein (A1AG) | Immunoglobulin G (IGG) | |----------------------------------|-------------------------------------| | a₁-Antitrypsin (A1AT) | Immunoglobulin G Subclass 1 (IGG1) | | α₂-Macroglobulin (A2MAC) | Immunoglobulin G Subclass 2 (IGG2) | | β₂-Microglobulin (B2MIC, B2MU**) | Immunoglobulin G Subclass 3 (IGG3) | | C3 Complement (C3) | Immunoglobulin G Subclass 4 (IGG4) | | C4 Complement (C4) | Immunoglobulin M (IGM) | | Ceruloplasmin (CER) | Prealbumin (PREALB) | | Haptoglobin (HAPT) | Retinol binding Protein (RBP) | | Hemopexin (HPX) | soluble Transferrin Receptor (STFR) | | Homocysteine (HCYS) | specialty Albumin (sALB*) | | Immunoglobulin A (IGA) | Transferrin (TRF) | | Immunoglobulin E (IGE) | | * For serum and plasma ** For urine Prescription Use × (Per 21 CFR 801 Subpart D) Over-The-Counter-Use_ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) AND/OR Concurrence of CDRH, Office of Device Evaluation Page 1 of Maria M Chan Division Sign-Off Office of in Vitro Diagnostic Device Evaluation and Safety K083463
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...