Access Ferritin

K234052 · Beckman Coulter, Inc. · JMG · Mar 20, 2024 · Immunology

Device Facts

Record IDK234052
Device NameAccess Ferritin
ApplicantBeckman Coulter, Inc.
Product CodeJMG · Immunology
Decision DateMar 20, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5340
Device ClassClass 2

Intended Use

The Access Ferritin assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of ferritin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. Ferritin is used as an aid in the diagnosis of iron deficiency or iron overload.

Device Story

Access Ferritin assay is a two-site immunoenzymatic sandwich assay; utilizes paramagnetic particles coated with goat anti-mouse IgG and mouse monoclonal anti-ferritin complexes; goat anti-ferritin-alkaline phosphatase conjugate binds to ferritin in patient serum or plasma. Magnetic field separates bound from unbound materials; chemiluminescent substrate (Lumi-Phos PRO) added; light generation measured by luminometer on DxI 9000 Access Immunoassay Analyzer. Light intensity proportional to ferritin concentration; concentration determined via stored calibration. Used in clinical laboratory settings by trained personnel. Output provides quantitative ferritin levels to assist clinicians in diagnosing iron deficiency or iron overload.

Clinical Evidence

Bench testing only. Precision/reproducibility evaluated per CLSI EP05-A3 (N=80 per sample for within-lab; N=75 for lot-to-lot and instrument-to-instrument). Linearity confirmed per CLSI EP06 (0.03–1698.57 ng/mL). LoB/LoD/LoQ established per CLSI EP17-A2 (LoQ 0.6 ng/mL). Method comparison (N=147) against predicate on Access 2 system showed R²=0.99, slope 0.96, intercept 0.23 via Passing-Bablok regression.

Technological Characteristics

Paramagnetic particle, chemiluminescent sandwich immunoassay. Automated system. Measuring range 0.6-1,500 ng/mL (up to 75,000 ng/mL with automated dilution). Traceable to WHO 94/572. Reagents include pack, calibrators, Lumi-Phos PRO substrate, and wash buffer. Standards: CLSI EP05-A3, EP06-2nd Ed, EP17-A2, EP09c.

Indications for Use

Indicated for the quantitative determination of ferritin in human serum and plasma to aid in the diagnosis of iron deficiency or iron overload.

Regulatory Classification

Identification

A ferritin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the ferritin (an iron-storing protein) in serum and other body fluids. Measurements of ferritin aid in the diagnosis of diseases affecting iron metabolism, such as hemochromatosis (iron overload) and iron deficiency amemia.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format. March 20, 2024 Beckman Coulter, Inc Kate Oelberg Senior Staff Quality and Regulatory Affairs 1000 Lake Hazeltine Drive Chaska, Minnesota 55318 Re: K234052 Trade/Device Name: Access Ferritin Regulation Number: 21 CFR 866.5340 Regulation Name: Ferritin Immunological Test System Regulatory Class: Class II Product Code: JMG Dated: December 21, 2023 Received: December 21, 2023 Dear Kate Oelberg: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {1}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Ying Mao -S Ying Mao, Ph.D. Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K234052 Device Name Access Ferritin Indications for Use (Describe) The Access Ferritin assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of ferritin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. Ferritin is used as an aid in the diagnosis of iron deficiency or iron overload. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span> </span> </div> | |----------------------------------------------|-----------------------------| | Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span> </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510 (k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. 510(k) Number___________K234052____________ #### Submitter Name and Address: Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, MN 55318 #### Primary Contact: Kate Oelberg, Senior Staff Quality and Regulatory Affairs Email: kmoelberg@beckman.com Phone: (612) 431-7315 #### Alternate Contact: Kuljeet Kaur, Requlatory Affairs Manager Email: kkaur@beckman.com Office Phone: (952) 465-1914 Trade Name: Access Ferritin Common Name: Ferritin Classification Regulation: 21 CFR 866.5340 Classification Product Code: JMG #### Predicate Device: Access Ferritin 510(k) Number K926221, k052082 #### Device Description The Access Ferritin assay is a sandwich immunoenzymatic assay. The Access Ferritin assay consists of the reagent pack and calibrators. Other items needed to run the assay include substrate and wash buffer. The Access Ferritin assay reagent pack, Access Ferritin assay calibrators, along with the UniCel Dxl Wash Buffer II are designed for use with the Dxl 9000 Access Immunoassay Analyzer in a clinical laboratory setting. #### Intended Use The Access Ferritin assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of ferritin levels in human serum and plasma (heparin) using the Access Immunoassay Systems. Ferritin is used as an aid in the diagnosis of iron deficiency or iron overload. {4}------------------------------------------------ # Comparison of Technological Characteristics to the Predicate (Assay) | System<br>Attribute/Characteristic | Predicate Access Ferritin<br>assay (k926221/ k052082)<br>run on the Access<br>Immunoassay System | Access Ferritin assay run on<br>the Dxl 9000 Access<br>Immunoassay Analyzer<br>Instrument | |----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use/<br>Indications for Use | The Access Ferritin assay is a<br>paramagnetic particle,<br>chemiluminescent<br>immunoassay for the<br>quantitative determination<br>of ferritin levels in human<br>serum and plasma (heparin)<br>using the Access<br>Immunoassay Systems. | The Access Ferritin assay is a<br>paramagnetic particle,<br>chemiluminescent immunoassay<br>for the quantitative determination<br>of ferritin levels in human serum<br>and plasma (heparin) using the<br>Access Immunoassay Systems.<br>Ferritin is used as an aid in the<br>diagnosis of iron deficiency or<br>iron overload. | | Analyte Measured | Ferritin | Same | | Traceable to | WHO 94/572 | Same | | Technology | Sandwich Immunoassay | Same | | Format | Chemiluminescent | Same | | Method | Automated | Same | | Calibration | Utilizes a stored calibration<br>curve | Same | | Sample Type | Serum/Plasma | Same | | Stability | 28 days after opening | Same | | Reagent Pack<br>formulation and<br>packaging | Access Reagent Pack<br>formulation and packaging. | Same | | Reagent Configurations | One Configuration:<br>100 determinations, 2 packs,<br>50 tests/pack | Two Configurations:<br>1) 100 determinations, 2<br>packs, 50 tests/pack (for<br>predicate and candidate<br>instrument)<br>2) 200 determinations, 2<br>packs, 100 tests/pack (for<br>candidate instrument only) | | Measuring Range | 0.2-1,500 ng/mL<br>Dil-Fer range<br>1,500-15,000 ng/mL | 0.6-1,500 ng/mL<br>Automated dilution: Up to<br>approximately 75,000 ng/mL<br>(µg/L) | | Sample Volume | 10 uL | 5 uL | | Instrument | Access Immunoassay system | Dxl 9000 Access<br>Immunoassay Analyzer | | Substrate | Access Substrate | Lumi-Phos PRO substrate | {5}------------------------------------------------ # Standard/Guidance Document Referenced (if applicable): CLSI EP05-A3: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - Third Edition CLSI EP06-2nd Edition : Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition CLSI EP09c: Measurement Procedure Comparison and Bias Estimation Using Patient Samples-Third Edition ## Summarv of Studies: Method Comparison: A method comparison study was performed to compare the Access Ferritin assay on the Dxl 9000 Access Immunoassay Analyzer to the predicate device. A total of one hundred forty-seven (147) samples falling within the measuring range of the Access Ferritin assay were evaluated. The results of the within range method comparison study met the acceptance criteria of R2 ≥ 0.90 and slope 1.00 ± 0.09. | N | Concentration<br>Range*<br>(ng/mL[ug/L]) | Slope | Slope<br>95% Cl | Intercept | Intercept<br>95% Cl | R | |-----|------------------------------------------|-------|-----------------|-----------|---------------------|------| | 147 | 2.3-1471 | 0.96 | 0.95-0.97 | 0.23 | -0.34-1.1 | 0.99 | *Range is Access 2 values Precision: The assay was designed to have within-laboratory imprecision as listed below: ≤ 0.5 ng/mL (µg/L) SD at concentrations ≤ 5 ng/mL (µg/L) ≤ 10.0% CV at concentrations > 5 ng/mL (ug/L) A study based on CLSI EP05-A317 performed on the Dxl 9000 Access Immunoassay Analyzer tested multiple samples in duplicate in 2 runs per day for a minimum of 20 days. | Concentration (ng/mL[ug/L]) | | Repeatability<br>(Within-run) | | Between-run | | Between-day | | Within-<br>Laboratory<br>(Total) | | | |-----------------------------|----|-------------------------------|------|-------------|------|-------------|------|----------------------------------|------|------| | Sample | N | Mean | SD | %CV | SD | %CV | SD | %CV | SD | %CV | | SAMPLE 1 | 80 | 1.2 | 0.2 | 16.5 | 0.0 | 0.0 | 0.1 | 5.3 | 0.2 | 17.3 | | SAMPLE 2 | 80 | 13 | 0.4 | 3.1 | 0.2 | 1.3 | 0.4 | 3.2 | 0.6 | 4.7 | | SAMPLE 3 | 80 | 147 | 4.5 | 3.0 | 2.0 | 1.4 | 6.0 | 4.1 | 7.7 | 5.3 | | SAMPLE 4 | 80 | 289 | 8.6 | 3.0 | 0.0 | 0.0 | 9.1 | 3.1 | 12.5 | 4.3 | | SAMPLE 5 | 80 | 560 | 17.2 | 3.1 | 10.0 | 1.8 | 16.2 | 2.9 | 25.7 | 4.6 | | SAMPLE 6 | 80 | 1276 | 53.3 | 4.2 | 5.8 | 0.5 | 42.2 | 3.3 | 68.2 | 5.3 | Linearity: A verification study was performed to evaluate the linearity of the Access Ferritin assay on the Dxl 9000 Access Immunoassay Analyzer based on CLSI EP06-Ed2. The Access Ferritin assay is linear on the Dxl 9000 Access Immunoassay Analyzer throughout the analytical measuring interval of approximately 0.6-1,500 ng/mL Limit of Blank (LoB): In one study, LoB was tested using a protocol based on CLSI EP17-A2. A total of 75 replicates of native samples were measured using two reagent packs on two Dxl 9000 Access Immunoassay Analyzers. The claimed LoB for Access Ferritin assay is 0.2 ng/mL on Dxl 9000 Access Immunoassay Analyzer. {6}------------------------------------------------ Limit of Detection (LoD); In one study, LoD was tested using a protocol based on CLSI EP17-A2. Three Dxl 9000 Access Immunoassay Analyzers were used in the study design with three reagent lots and one calibrator lot. Five to nine serum samples containing low levels of Ferritin analyte were prepared. Samples were tested over five days with one run per day and nine replicates per run for each pack lot. This resulted in ≥ 40 replicates minimally required for LoD estimation for each sample on each pack lot tested. The claimed LoD estimate for the Access Ferritin assay is 0.4 ng/mL on Dxl 9000 Access Immunoassay Analyzer. Limit of Quantitation (LoQ): In one study, LoQ was tested using a protocol based on CLSI EP17-A2. For estimation of LoQ, eight to twelve serum samples containing low levels of Ferritin analyte were measured. Samples were tested in replicates of nine per run with one run per day and five total days on each pack lot and instrument. A minimum of 40 replicates for each sample on each pack lot was tested.The maximum claimed LoQ ((≤ 20% within-lab CV) determined for the Access Ferritin assay is 0.6 ng/mL on Dxl 9000 Access Immunoassay Analyzer). Other claims: The claims for the analytical specificity, reference intervals, matrix comparison are being transferred from file k926221. ## Substantial Equivalence Comparison Conclusion Beckman Coulter's Access Ferritin assay on the Dxl 9000 Access Immunoassay Analyzer is substantially equivalent to Ferritin assay on the Access Immunoassay System (K926221) as demonstrated through the information and data provided in this submission. The performance testing presented in this submission provides evidence that the device is safe and effective in its intended use.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...