K122304 · The Binding Site Group , Ltd. · DBA · Apr 15, 2013 · Immunology
Device Facts
Record ID
K122304
Device Name
HUMAN C1 INACTIVATOR KIT FOR USE ON SPAPLUS
Applicant
The Binding Site Group , Ltd.
Product Code
DBA · Immunology
Decision Date
Apr 15, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5250
Device Class
Class 2
Intended Use
This Human C1 Inactivator kit for use on SPAPLUS is intended for the in vitro measurement of human C1 inactivator in human serum using the SPAPLUS Analyzer. Measurement of C1 inactivator levels in serum is an aid in the diagnosis of hereditary angioedema (HAE) in conjunction with other laboratory and clinical findings.
Device Story
The Human C1 Inactivator Kit is an in vitro diagnostic assay for use on the SPAPLUS Analyzer. It utilizes sheep anti-human C1 inactivator antibodies to react with C1 inactivator protein in human serum samples, forming immune complexes. These complexes scatter light; the intensity of scattered light is measured via turbidometry and is directly proportional to the C1 inactivator concentration. The SPAPLUS Analyzer automatically calculates protein concentration using a stored calibration curve. The device is intended for use in clinical laboratory settings by trained personnel. Results assist clinicians in diagnosing hereditary angioedema (HAE) when combined with other clinical and laboratory findings. The system supports automated sample dilution (1:10 or 1:20) for samples outside the standard measuring range.
Clinical Evidence
No clinical trials were performed. Substantial equivalence was established through bench testing, including precision (CV < 10%), linearity (R² = 0.9982), and a method comparison study (n=94) against the predicate device. The method comparison showed an overall agreement of 93.75% (95% CI: 88.91 - 98.59) including outliers, and 94.57% (95% CI: 90.13 - 99.01) excluding outliers.
Technological Characteristics
Quantitative immunoturbidimetric assay. Reagents include sheep anti-human C1 inactivator antiserum, human serum-based calibrators/controls, and reaction buffer containing sodium azide, EACA, EDTA, and benzamidine. Operates on the SPAPLUS Analyzer. Measuring range: 0.06–0.80 g/L. Calibration traceable to ERM-DA470k. Software performs automated calculations and dilutions.
Indications for Use
Indicated for the in vitro measurement of human C1 inactivator in human serum to aid in the diagnosis of hereditary angioedema (HAE) in conjunction with clinical and laboratory findings.
Regulatory Classification
Identification
A complement C1 inhibitor (inactivator) immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the complement C1 inhibitor (a plasma protein) in serum. Complement C1 inhibitor occurs normally in plasma and blocks the action of the C1 component of complement (a group of serum proteins which destroy infectious agents). Measurement of complement C1 inhibitor aids in the diagnosis of hereditary angioneurotic edema (increased blood vessel permeability causing swelling of tissues) and a rare form of angioedema associated with lymphoma (lymph node cancer).
Predicate Devices
Siemens N antiserum to C1 inhibitor (k960257)
Related Devices
K141100 — OPTILITE C1 INACTIVATOR KIT · The Binding Site Group , Ltd. · Jul 18, 2014
K965024 — N-ASSAY TIA C1-INACTIVATOR TEST KIT · Crestat Diagnostics, Inc. · Jul 14, 1997
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k122304
B. Purpose for Submission:
New device
C. Measurand:
Complement C1 inactivator (inhibitor)
D. Type of Test:
Quantitative immunonephelometry
E. Applicant:
The Binding Site
F. Proprietary and Established Names:
Human C1 Inactivator Kit for use on SPAPLUS
Complement C1 inhibitor
G. Regulatory Information:
1. Regulation section:
21 CFR 866.5250 Complement C1 inhibitor (inactivator) immunological test system
2. Classification:
Class II – Device and Calibrator
3. Product code:
DBA, Complement C1 inhibitor (inactivator) antigen, antiserum, control
{1}
1. Panel:
Immunology (82)
H. Intended Use:
1. Intended use(s):
This Human C1 Inactivator kit for use on SPAPLUS is intended for the in vitro measurement of human C1 inactivator in human serum using the SPAPLUS Analyzer. Measurement of C1 inactivator levels in serum is an aid in the diagnosis of hereditary angioedema (HAE) in conjunction with other laboratory and clinical findings
2. Indication(s) for use:
Same as Intended use.
3. Special conditions for use statement(s):
Prescription use only
4. Special instrument requirements:
SPAPLUS Analyzer (k040958)
I. Device Description:
The Human C1 Inactivator Kit for use on SPAPLUS is comprised of the following reagents:
Human C1 inactivator antiserum: This antiserum is monospecific for C1 inactivator and is supplied in stabilized liquid form. It contains 0.099% sodium azide, 0.1% EACA, 0.1% EDTA and 0.01% benzamidine as preservatives.
Calibrator set and Controls: These consist of pooled human serum and are supplied in stabilized liquid form. They contain 0.099% sodium azide, 0.1% EACA and 0.01% benzamidine as preservatives.
Reaction Buffer: Containing 0.099% sodium azide as a preservative
J. Substantial Equivalence Information:
1. Predicate device name(s) and 510(k) number(s):
Siemens N antiserum to C1 inhibitor k960257
{2}
2. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended use | Quantification of C1 inactivator in human serum. It is intended to aid in the diagnosis of hereditary angioedema in conjunction with other laboratory and clinical findings | Same |
| Assay type | Quantitative | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Test method | Turbidometry | Nephelometry |
| Specimen type | Serum | Serum and plasma (citrated and EDTA) |
| Detection antibody | Sheep anti-human C1 inactivator | Rabbit anti-human C1 inactivator |
| Sample dilution | 1:10 and 1:20 | 1:5 |
| Measuring range | 0.06 – 0.40 g/L (1/10 dilution)
0.12 – 0.80 g/L (1/20 dilution) | 0.03 to 0.40 g/L |
| Instrument | SPAPLUS Analyzer | Behring Nephelometer II System |
| Calibration | Six calibrators (provided):
0.006, 0.01, 0.015, 0.02, 0.03, 0.04 g/L | One N Protein Standard PY (not provided) serially diluted by the instrument |
| Control | Two controls (provided): high (0.3 g/L) and low (0.15g/L) | One N/T Protein Control PY (not provided) serially diluted by the instrument |
| Open vial stability | 3 months | 4 weeks |
| On-board stability | 30 days | 5 days at 8 hours per day |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline – Second Edition
CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
CLSI EP7-A2: Interference Testing in Clinical Chemistry, Approved Guideline – Second Edition
CSLI EP17-A: Protocols for Determination of Limits of Detection and Limits of
{3}
Quantitation; Approved Guideline
CLSI C28-A3: Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory
# L. Test Principle:
Human C1 inactivator protein contained in human plasma and serum samples reacts with sheep anti-human C1 inactivator to form immune complexes. These complexes scatter a beam of light passed through the sample and the intensity of the scattered light is proportional to the concentration of the protein in the sample. Protein concentration is automatically calculated by reference to a calibration curve stored within the SPAPLUS Analyzer.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
The studies followed CLSI EP5-A2, where three pooled sera samples with high $(0.359\mathrm{g / L})$ , mid $(0.162\mathrm{g / L})$ and low $(0.108\mathrm{g / L})$ analyte level were each tested in 2 runs per day (each of the 2 runs in duplicate) for over 21 days using 3 reagent lots on 3 analyzers. The same high sample was tested at $1/20$ dilution. Results met the Acceptance criteria for total precision $(\% \mathrm{CV} < 10\%)$ , within-run precision $(\% \mathrm{CV} < 7\%)$ , between-run precision $(\% \mathrm{CV} < 8\%)$ , and between-day precision $(\% \mathrm{CV} < 8\%)$ . The precision of the $1/20$ dilution is equivalent to that for the $1/10$ standard dilution.
| Sample | Low | Mid | High at 1/10 dilution | High at 1/20 dilution |
| --- | --- | --- | --- | --- |
| Mean concentration | 0.108 g/L | 0.162 g/L | 0.359 g/L | 0.370 g/L |
| Total precision | 0.0058 | 0.0070 | 0.0170 | 0.0171 |
| SD/%CV | 5.4% | 4.3% | 4.7% | 4.6% |
| Within-run precision | 0.0029 | 0.0023 | 0.0057 | 0.0055 |
| SD/%CV | 2.6% | 1.4% | 1.6% | 1.5% |
| Between-run precision | 0.0011 | 0.0029 | 0.0054 | 0.0057 |
| SD/%CV | 1.0% | 1.8% | 1.5% | 1.5% |
| Between-day precision | 0.0050 | 0.0059 | 0.0151 | 0.0151 |
| SD/%CV | 4.6% | 3.6% | 4.2% | 4.1% |
| Between-batch precision | 0.001 | 0.002 | 0.007 | 0.005 |
| SD/%CV | 0.62% | 1.51% | 1.94% | 1.33% |
| Between-instrument precision | 0.005 | 0.004 | 0.014 | 0.011 |
| SD/%CV | 4.31% | 2.37% | 3.89% | 2.98% |
{4}
# b. Linearity/assay reportable range:
The studies followed CLSI EP6-A, where eleven serial dilutions were prepared by diluting a high $(0.458\mathrm{g / L})$ in a low pool $(0.044\mathrm{g / L})$ of unprocessed serum. The resulting series of 13 samples (100, 90, 80, 70, 60, 50, 40, 30, 20, 10, 5, 2.5, $0\%$ of the high pool) was tested in 3 replicates (Rep). Results met the Acceptance criteria for $\mathrm{CV} < 8\%$ . The linear regression equation is $y = 0.962x + 0.0122$ and $\mathbf{R}^2$ value is 0.9982. The assay is linear between 0.044 - 0.458 g/L and the claimed measuring range is 0.06 - 0.4 g/L. The labeling states that for samples giving results outside of the measuring range at standard 1/10 sample dilution, auto dilution to 1/20 will be performed. Results which are still outside of the measuring range following dilution are reported as $< 0.06\mathrm{g / L}$ (from 1/10 dilution) or $>0.8\mathrm{g / L}$ (from 1/20 dilution).
| Sample | % High pool | Rep 1 (g/L) | Rep 2 (g/L) | Rep 3 (g/L) | Mean (g/L) | % Mean Recovery | SD | % CV |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| 1 | 100 | 0.465 | 0.456 | 0.454 | 0.458 | 100 | 0.0023 | 5.3 |
| 2 | 90 | 0.411 | 0.416 | 0.412 | 0.413 | 116.6 | 0.0020 | 3.2 |
| 3 | 80 | 0.380 | 0.370 | 0.361 | 0.370 | 112.3 | 0.00313 | 4.2 |
| 4 | 70 | 0.325 | 0.326 | 0.324 | 0.325 | 117.5 | 0.00363 | 3.6 |
| 5 | 60 | 0.290 | 0.292 | 0.289 | 0.290 | 109.8 | 0.0026 | 1.9 |
| 6 | 50 | 0.250 | 0.258 | 0.249 | 0.252 | 109.7 | 0.0032 | 1.7 |
| 7 | 40 | 0.212 | 0.212 | 0.226 | 0.217 | 103.4 | 0.0081 | 3.7 |
| 8 | 30 | 0.188 | 0.183 | 0.182 | 0.184 | 100.5 | 0.0049 | 2.0 |
| 9 | 20 | 0.137 | 0.142 | 0.138 | 0.139 | 99.3 | 0.0015 | 0.5 |
| 10 | 10 | 0.097 | 0.104 | 0.099 | 0.100 | 97.3 | 0.0010 | 0.3 |
| 11 | 5 | 0.075 | 0.073 | 0.069 | 0.072 | 98.7 | 0.0095 | 2.6 |
| 12 | 2.5 | 0.065 | 0.063 | 0.061 | 0.063 | 99.1 | 0.0026 | 0.6 |
| 13 | 0 | 0.045 | 0.045 | 0.041 | 0.044 | 100 | 0.0059 | 1.3 |
# c. Traceability, Stability, Expected values (controls, calibrators, or methods):
# i) Traceability:
The calibration of the assay is traceable to the international reference standard ERM-DA470k.
# ii) Kit Stability:
Real-time stability - A study to establish shelf-life stability (from the date of manufacture when stored at recommended temperature $2 - 8^{\circ}\mathrm{C}$ ) of the Human C1 Inactivator Kit is on-going. Currently available data supported a 3-month stability claim.
Open-vial stability - The Human C1 Inactivator Kit reagents can be stored, opened at $2 - 8^{\circ}\mathrm{C}$ for up to 3 months.
On-board stability - The Human C1 Inactivator Kit reagents can be stored onboard the SPAPLUS Analyzer at $8 - 12^{\circ}\mathrm{C}$ for at least 30 days.
{5}
# d. Detection limit:
The analytical sensitivity was determined in accordance with CLSI EP17-A. The Limit of Blank (LoB) was based on 60 determinations of a blank sample and was estimated as the $95\%$ percentile of the distribution. The Limit of Detection (LoD) was calculated according to the equation: the $\mathrm{LoB} + c_{\beta}$ x SD where SD, the standard deviation, was based on 40 determinations of a sample with analyte level near the lower limit of the reportable range. The LoQ was calculated and the bias $(-0.001~\mathrm{g / L})$ was within the maximum allowable bias.
| LoB | LoD | LoQ |
| --- | --- | --- |
| 0.0015 | 0.003 | 0.06 |
# e. Analytical specificity:
Interferences were assessed according to CLSI EP7-A2 by testing 3 serum samples with low (0.094 g/L), mid (0.222 g/L) and high (0.341 g/L) C1 inactivator concentrations. Each sample was spiked with interfering substances and tested in replicates of 7 (for high sample) or 3 (for mid and low samples). For non-interference to be claimed, the mean results from the spiked samples must be within $10\%$ of the mean of the neat samples. The data demonstrated that the assay was not affected by high levels of the following substances: hemoglobin (5 g/L), bilirubin (200 mg/L), and chyle (1500 formazine turbidity units or FTU).
# Hook effect
The possibility of a hook effect due to antigen excess was assessed as follows: 3 batches of a C1 Inactivator Internal Reference preparation with a value of $2.31\mathrm{g / L}$ which is 5.7 times the highest calibrator were assayed. The resulting OD values were higher than the normal curve top point for all three batches. The data demonstrated the assay is not susceptible to antigen excess up to a concentration of $0.231\mathrm{g / L}$ which is equivalent to $2.31\mathrm{g / L}$ at standard 1/10 dilution.
# f. Assay cut-off:
Not determined
# 2. Comparison studies:
# a. Method comparison with predicate device:
A total of 94 sera samples spanning the dynamic range were assayed in singleton by both the Human C1 Inactivator Kit on the SPAPLUS Analyzer and the N antiserum to C1 inhibitor test on the BN II System. The serum samples included 53 normal donors and 29 clinical samples (HAE (7), unknown diagnosis (18), recurrent periorbital edema (1), urticaria (1), IgG kappa paraprotein (1), SLE (1)). In addition, 4 contrived normal serum samples and 8 serum samples from patients with Kappa Myeloma (5) and Lambda Myeloma (3) with depressed levels of C1 inactivator were included to cover the lower end of the claimed measuring range. Regression statistics are based on the balance of the paired results and the inclusion or exclusion of 4 between-
{6}
method outliers outside 95% limits, and the data are as follows:
| Regression fit | Inclusion of outliers | Regression Equation | Slope (95% CI) | Intercept (95% CI) |
| --- | --- | --- | --- | --- |
| Ordinary Linear | Yes | y = 0.92 x + 0.02 | 0.86 to 0.98 | 0.00 to 0.03 |
| | No | y = 0.96 x + 0.01 | 0.91 to 1.01 | -0.01 to 0.02 |
| Passing-Bablok | Yes | y = 0.98 x | 0.93 to 1.04 | -0.01 to 0.02 |
| | No | y = 0.99 x | 0.94 to 1.05 | -0.02 to 0.01 |
| Deming | Yes | y = 0.96 x | 0.90 to 1.02 | -0.01 to 0.02 |
| | No | y = 0.98 x | 0.93 to 1.03 | -0.01 to 0.01 |
When considering the lower limit of the reference range as the medical decision point, the percentage positive and negative agreement in the serum samples (n=94) analyzed including all 4 outliers is as follows:
| Human C1 Inactivator Kit on the SPAPLUS Analyzer | Predicate | | |
| --- | --- | --- | --- |
| | Positive | Negative | Total |
| Positive | 17 | 4 | 21 |
| Negative | 2 | 73 | 75 |
| Total | 19 | 77 | 96 |
Positive percentage agreement: 89.47% (95% CI: 83.33 - 95.61)
Negative percentage agreement: 94.81% (95% CI: 90.37 - 99.25)
Overall agreement: 93.75% (95% CI: 88.91 - 98.59)
When considering the lower limit of the reference range as the medical decision point, the percentage positive and negative agreement in the serum samples (n=90) analyzed excluding all 4 outliers is as follows:
| Human C1 Inactivator Kit on the SPAPLUS Analyzer | Predicate | | |
| --- | --- | --- | --- |
| | Positive | Negative | Total |
| Positive | 17 | 4 | 21 |
| Negative | 1 | 70 | 71 |
| Total | 18 | 74 | 92 |
Positive percentage agreement: 94.44% (95% CI: 89.95 - 98.93)
Negative percentage agreement: 94.59% (95% CI: 90.16 - 99.02)
Overall agreement: 94.57% (95% CI: 90.13 - 99.01)
{7}
b. Matrix comparison:
None
3. Clinical studies:
a. Clinical Sensitivity:
None determined
b. Clinical specificity:
None determined
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Based on lower limit of reference range which is 0.21 g/L.
5. Expected values/Reference range:
The reference range (0.21 -0.38 g/L) selected is comparable to that of the predicate assay (0.21 -0.39 g/L) and those used by several US Reference Centers: Mayo Clinic (0.19-0.37 g/L); ARUP Laboratories (0.21-0.39 g/L); Duke University Hospital (0.21-0.39 g/L).
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.