HEVYLITE HUMAN IGM KAPPA AND HEVYLITE IGM LAMBDA KIT

K113823 · The Binding Site Group , Ltd. · CFN · Mar 27, 2013 · Immunology

Device Facts

Record IDK113823
Device NameHEVYLITE HUMAN IGM KAPPA AND HEVYLITE IGM LAMBDA KIT
ApplicantThe Binding Site Group , Ltd.
Product CodeCFN · Immunology
Decision DateMar 27, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5510
Device ClassClass 2

Intended Use

Hevylite Human IgM Lambda Kit for use on Siemens BN™II is intended for the in vitro quantification of IgM Lambda (combined μ heavy and λ light chain) concentration in human serum on the Siemens Behring Nephelometer™ II (BN™ II). The test result is to be used with previously diagnosed Waldenstom’s macroglobulinemia in conjunction with other clinical and laboratory findings. This assay has not been established for the diagnosis, monitoring and prognosis of Waldenstrom’s macroglobulinemia. Hevylite Human IgM Kappa Kit for use on Siemens BN™II is intended for the in vitro quantification of IgM Kappa (combined μ heavy and κ light chain) concentration in human serum on the Siemens Behring Nephelometer™ II (BN™ II). The test result is to be used with previously diagnosed Waldenstrom’s macroglobulinemia in conjunction with other clinical and laboratory findings. This assay has not been established for the diagnosis, monitoring and prognosis of Waldenstrom’s macroglobulinemia.

Device Story

In vitro diagnostic kits for quantitative measurement of IgM Kappa and IgM Lambda concentrations in human serum. Uses nephelometry on Siemens BN II systems. Input: patient serum sample; Reagents: polystyrene latex particles coated with polyclonal monospecific sheep anti-human IgM (anti-kappa or anti-lambda). Principle: light scattering by antigen-antibody immune complexes; intensity proportional to antigen concentration. Output: quantitative concentration values (g/L). Used in clinical laboratories by technicians. Results assist clinicians in assessing patients with previously diagnosed Waldenstrom’s macroglobulinemia by providing specific immunoglobulin light chain quantification.

Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP05-A2 (total CV 5.7-10.6%). Linearity (r² > 0.99) and LoD/LoQ established per CLSI EP-17A. Interference testing showed no significant impact from hemoglobin, bilirubin, or chyle. Method comparison performed on 186 samples (110 normal, 78 Waldenstrom’s) against Total IgM (r=0.94), SPE (97.9% overall agreement), and IFE (100% agreement on 60 samples).

Technological Characteristics

Nephelometric immunoassay; utilizes Siemens BN II system for measurement; quantifies IgM heavy chain combined with Kappa or Lambda light chains in human serum.

Indications for Use

Indicated for use in patients with previously diagnosed Waldenstrom’s macroglobulinemia to aid in clinical assessment in conjunction with other findings. Not for diagnosis, monitoring, or prognosis of the disease.

Regulatory Classification

Identification

An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/0/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features the HHS logo, which is a stylized caduceus, a symbol of medicine and healing. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the logo. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 27, 2013 The Binding Site Group, Ltd. c/o Dr. Jill Constatine, Regulatory Affairs Scientist 8 Calthorpe Road Edgbaston, Birmingham West Midlands UK B15 10T Re: k113823 Trade/Device Name: Hevylite™ Human IgM Kappa Kit for use on Siemens BN™ II Systems Hevylite™ Human IgM Lambda Kit for use on Siemens BNTM II Systems Regulation Number: 21 CFR 866.5510 Regulation Name: Immunoglobulins A, G, M, D, and E Immunological Test System Regulation Name: Immunoglobulins A. Regulatory Class: Class II Product Code: CFN. PDE. PDF Dated: March 26, 2013 Received: March 26, 2013 Dear Dr. Constantine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misJeading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ ## Page 2 - Dr. Jill Constantine Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number . (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Maria=M.Chan -S Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known): k113823 Device Name: Hevylite™ Human IgM Kappa Kit for use on Siemens BN™ II Hevylite™ Human IgM Lambda Kit for use on Siemens BN™ II Indications For Use: Hevylite Human IgM Kappa Kit for use on Siemens BN™II is intended for the in vitro quantification of IgM Kappa (combined u heavy and K light chain) concentration in human serum on the Siemens Behring Nephelometer™ II (BN™ II). The test result is to be used with previously diagnosed Waldenstoms Macrolobulinaemiain conjunction with other clinical and laboratory findings. This assay has not been established for the diagnosis, monitoring and prognosis of Waldenstroms macroglbulinaemia. Hevylite Human IgM Lambda Kit for use on Siemens BN™II is intended for the in vitro quantification of IgM Lambda (combined u heavy and À light chain) concentration in human serum on the Siemens Behring Nephelometer™ II (BN™ II). The test result is to be used with previously diagnosed Waldenstoms Macrolobulinaemia in conjunction with other clinical and laboratory findings. This assay has not been established for the diagnosis, monitoring and prognosis of Waldenstroms macrogibulinaemia. Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OIR) ## Maria Maria Maria - S Division Sign-Off Office of In Vitro Diagnostics and Radiological Health 510(k) k113823
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...