ImmunoCAP Tryptase is an in vitro semi-quantitative assay for measurement of tryptase in human serum or plasma (EDTA, lithium heparin or sodium heparin). It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of patients with a suspicion of systemic mastocytosis in conjunction with other clinical and laboratory findings. ImmunoCAP Tryptase is to be used with the instruments Phadia 100, and Phadia 250.
Device Story
ImmunoCAP Tryptase is an in vitro diagnostic assay for measuring tryptase levels in human serum or plasma. The device utilizes anti-tryptase antibodies covalently coupled to an ImmunoCAP solid phase; patient samples are incubated with the solid phase, followed by the addition of enzyme-labeled anti-tryptase antibodies. After washing steps to remove unbound components, a developing agent is added; the resulting fluorescence is measured to quantify tryptase concentration. The assay is performed on Phadia 100 and Phadia 250 instruments, which automate processing and result calculation. Used in clinical laboratory settings by trained personnel, the output provides a semi-quantitative concentration value. This result assists clinicians in diagnosing systemic mastocytosis by providing objective biochemical data to support clinical evaluation.
Clinical Evidence
Clinical performance verified at two sites with 138 patients suspected of systemic mastocytosis. Study compared updated ImmunoCAP Tryptase against the currently cleared version. Results confirmed clinical performance remains unchanged for aiding diagnosis according to WHO criteria. Analytical performance verified via precision, lot-to-lot reproducibility, linearity, limit of detection, hook effect, analytical specificity, HAMA interference, recovery, and sample matrix studies.
Technological Characteristics
In vitro semi-quantitative immunoassay. Principle: Fluorescence enzyme immunoassay (FEIA). Components: Anti-tryptase coupled to ImmunoCAP, enzyme-labeled anti-tryptase antibodies. Instrumentation: Phadia 100 and Phadia 250. Sample types: Serum, EDTA plasma, lithium heparin plasma, sodium heparin plasma. Automated processing and calculation.
Indications for Use
Indicated for patients with suspected systemic mastocytosis, including adults and pediatric populations, to aid in clinical diagnosis via measurement of serum or plasma tryptase levels.
Regulatory Classification
Identification
A tryptase test system is a device that aids in the diagnosis of systemic mastocytosis. It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of patients with a suspicion of systemic mastocytosis in conjunction with other clinical and laboratory findings.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guideline entitled “Class II Special Controls Guideline: Tryptase Test System as an Aid in the Diagnosis of Systemic Mastocytosis.” For availability of the document, see § 866.1(e).
DEN120001 — IMMUNOCAP TRYPTASE, CALIBRATORS, CURVE CONTROL, CONJUGATE 50, CALIBRATOR STRIP, CURVE CONTROL STRIP, ANTI-TRYPTASE · Phadia US, Inc. · Feb 15, 2012
K133404 — IMMUNOCAP TOTAL IGE: CONJUGATE, ANTI-IGE, CALIBRATORS, CURVE CONTROLS, CONTROL LMH · Phadia AB · Feb 24, 2014
K190315 — ImmunoCAP Allergen e229, Allergen Component rCan f 4 Dog, ImmunoCAP Allergen e230, Allergen Component rCan f 6 Dog, ImmunoCAP Allergen e231, Allergen Component rFel d 7 Cat · Phadia AB · May 14, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
March 2, 2018
Phadia AB % Martin Mann Sr. Regulatory Affairs Manager Phadia US Inc 4169 Commercial Ave Portage, Michigan 49002
Re: K173704
Trade/Device Name: ImmunoCAP Tryptase, ImmunoCAP Tryptase Calibrators, ImmunoCAP Tryptase Curve Control, ImmunoCAP Tryptase Conjugate 50, ImmunoCAP Tryptase Calibrator Strip, ImmunoCAP Tryptase Curve Control Strip
Regulation Number: 21 CFR 866.5760 Regulation Name: Tryptase test system Regulatory Class: Class II Product Code: OYL Dated: December 1, 2017 Received: December 4, 2017
Dear Martin Mann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kelly Oliner -S
For, Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name ImmunoCAP Tryptase
#### Indications for Use (Describe)
ImmunoCAP Tryptase is an in vitro semi-quantitative assay for measurement of tryptase in human serum or plasma (EDTA, lithium heparin or sodium heparin). It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of patients with a suspicion of systemic mastocytosis in conjunction with other clinical and laboratory findings. ImmunoCAP Tryptase is to be used with the instrument Phadia 100.
ImmunoCAP Tryptase is an in vitro semi-quantitative assay for measurement of tryptase in human serum or plasma (EDTA, lithium heparin or sodium heparin). It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of patients with a suspicion of systemic mastocytosis in conjunction with other clinical and laboratory findings. ImmunoCAP Tryptase is to be used with the instrument Phadia 250.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Traditional 510(k) Submission, Update of ImmunoCAP Tryptase A.6 510k Summary
# 510(k) Summary
This 510(k) Summary is prepared in accordance with the requirements of 21 CFR Part 807.92.
| Date of Summary Preparation: | February 7, 2018 |
|--------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Phadia AB<br>Rapsgatan 7P<br>P.O. Box 6460<br>751 37 Uppsala, Sweden |
| US Distributor: | Phadia US Inc.<br>4169 Commercial Avenue<br>Portage, MI 49002 |
| Company Contact Person: | Martin Mann<br>Sr. Regulatory Affairs Manager<br>Phadia US Inc.<br>4169 Commercial Avenue<br>Portage, MI 49002<br>269-207-7049<br>martin.mann@thermofisher.com |
| Device Name: | ImmunoCAP Tryptase |
| Common Name: | In vitro diagnostic test system for semi-<br>quantitative measurement of tryptase |
| Classification:<br>Regulation section<br>Classification<br>Class II Special Controls Guideline<br>Product name<br>Product code | 21 CFR §866.5760 – Tryptase test system<br>Class II<br>Tryptase Test System as an Aid in the<br>Diagnosis of Systemic Mastocytosis<br>ImmunoCAP Tryptase<br>OYL, tryptase assay system |
| Substantial Equivalence to: | ImmunoCAP Tryptase DEN120001 (k103039) |
Premarket Notification 510(k) No: K173704
{4}------------------------------------------------
Traditional 510(k) Submission, Update of ImmunoCAP Tryptase A.6 510k Summary
### Indications for use statement
ImmunoCAP Tryptase is an in vitro semi-quantitative assay for measurement of tryptase in human serum or plasma (EDTA, lithium heparin or sodium heparin). It is intended for in vitro diagnostic use as an aid in the clinical diagnosis of patients with a suspicion of systemic mastocytosis in conjunction with other clinical and laboratory findings. ImmunoCAP Tryptase is to be used with the instruments Phadia 100 and Phadia 250.
### Device Description
#### Reagents
ImmunoCAP Tryptase reagents are modular in concept and are available individually in different package sizes dependent on instrument system used.
The reagents needed to perform the ImmunoCAP Tryptase assay are listed below. ImmunoCAP Tryptase/ImmunoCAP Tryptase Conjugate 50 (10-9525-02/10-9522-02) ImmunoCAP Tryptase Calibrators/ImmunoCAP Tryptase Calibrator Strip (10-9526-01/10-9523-01) ImmunoCAP Tryptase Curve Control/ImmunoCAP Tryptase Curve Control Strip (10-9527-01/10-9524-01) ImmunoCAP Tryptase Anti-Tryptase (14-4518-01), ImmunoCAP Tryptase Control (10-9370-01)
### Instrument System
Phadia 100 and Phadia 250 instruments with associated software process all steps of the assay and calculate results automatically after the assay is completed.
## Test Principle, ImmunoCAP Tryptase
Anti-tryptase, covalently coupled to ImmunoCAP, reacts with the tryptase in the patient sample. After washing, enzyme labeled antibodies against tryptase are added to form a complex. After incubation, unbound enzyme-anti-tryptase is washed away and the bound complex is then incubated with a developing agent. After stopping the reaction, the fluorescence of the eluate is measured. The higher the response value, the more tryptase is present in the sample. To evaluate the test results, the responses of the patient samples are transformed into concentrations with the use of a calibration curve.
#### Reason for Submission
This submission comprises changes in product raw material made to assay reagents ImmunoCAP Tryptase Conjugate and ImmunoCAP Tryptase Calibrators. The changes do not affect the Intended use or the Indications for use statements.
## Performance Characteristics
The updated ImmunoCAP Tryptase was compared with the currently cleared ImmunoCAP Tryptase with the use of samples with and without clinical documentation, as well as samples from healthy individuals. The analytical performance characteristics of the updated ImmunoCAP Tryptase were verified through studies of precision, lot-to-lot reproducibility, linearity, limit of detection, hook, analytical specificity, interference of HAMA, recovery and sample matrix.
{5}------------------------------------------------
Traditional 510(k) Submission, Update of ImmunoCAP Tryptase A.6 510k Summary
The clinical performance was verified through studies performed at two sites, including in total 138 patients with a suspicion of systemic mastocytosis. The results showed that the clinical performance of ImmunoCAP Tryptase has not changed when used as an aid in the routine clinical diagnosis of systemic mastocytosis according to WHO criteria.
### Conclusion
The analytical and clinical data included in this 510(k) submission support that the updated ImmunoCAP Tryptase is substantially equivalent to the currently cleared product.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches