QUANTA LITE SS-A 52 ELISA

K063565 · Inova Diagnostics, Inc. · OBE · Apr 4, 2007 · Immunology

Device Facts

Record IDK063565
Device NameQUANTA LITE SS-A 52 ELISA
ApplicantInova Diagnostics, Inc.
Product CodeOBE · Immunology
Decision DateApr 4, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5100
Device ClassClass 2

Indications for Use

The QUANTA Lite™ SS-A 52 ELISA is a semiquantitative enzymelinked immunosorbent assay for the detection of IgG anti-SS-A 52 antibodies in patient sera. The presence of these antibodies, when considered in conjunction with other laboratory and clinical findings, is an aid in the diagnosis of systemic lupus erythematosus, Sjögren's syndrome, systemic sclerosis, polymyositis and dermatomyositis.

Device Story

Semi-quantitative ELISA for detection of IgG anti-SS-A 52 autoantibodies in patient serum; utilizes polystyrene microwell plates coated with purified recombinant SS-A 52 antigen. Patient serum added to wells; anti-SS-A 52 antibodies bind to immobilized antigen. Unbound material washed away; enzyme-labeled goat anti-human IgG conjugate added. Chromogenic substrate (TMB) added; color intensity measured spectrophotometrically at 450nm/620nm. Results compared to control wells to determine antibody presence; reported in arbitrary units. Used in clinical laboratories by trained personnel. Output aids clinicians in diagnosing systemic lupus erythematosus, Sjögren’s syndrome, systemic sclerosis, polymyositis, and dermatomyositis when combined with other clinical/laboratory findings.

Clinical Evidence

No clinical data provided in the document; bench testing only.

Technological Characteristics

Polystyrene microwell plate coated with purified recombinant SS-A 52 antigen. Detection via goat anti-human IgG horseradish peroxidase conjugate and TMB chromogen. Requires microwell plate reader (450nm/620nm). Manual/semi-automated ELISA workflow. No specific ASTM standards cited.

Indications for Use

Indicated for the detection of IgG anti-SS-A 52 antibodies in patient sera to aid in the diagnosis of systemic lupus erythematosus, Sjögren's syndrome, systemic sclerosis, polymyositis, and dermatomyositis.

Regulatory Classification

Identification

An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).

Related Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k063565 B. Purpose for Submission: New device C. Measurand: Anti-SS-A 52 autoantibodies D. Type of Test: Semi-quantitative Enzyme-linked immunosorbent assay (ELISA) E. Applicant: INOVA Diagnostics, Inc. F. Proprietary and Established Names: QUANTA Lite™ SS-A 52 ELISA G. Regulatory Information: 1. Regulation section: CFR 866.5100 Antinuclear Antibody Immunological Test System 2. Classification: Class II 3. Product code: OBE Anti-SS-A 52 autoantibodies 4. Panel: IM 82 H. Intended Use: 1. Intended use: QUANTA Lite™ SS-A 52 ELISA is a semi-quantitative enzyme-linked immunosorbent assay (ELISA) for the detection of IgG anti-SS-A 52 antibodies in patient sera. The presence of these antibodies when considered in conjunction with other laboratory and clinical findings is an aid in the diagnosis of systemic lupus erythematosus, Sjögren’s syndrome, systemic sclerosis, polymyositis and dermatomyositis. 2. Indication(s) for use: Same as intended use 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: Microwell plate reader capable of measuring OD at 450nm (and 620nm for dual wavelength readings) I. Device Description: The QUANTA Lite SS-A 52 ELISA consists of a polystyrene microwell ELISA plate coated with purified recombinant SS-A 52 antigen; ELISA negative, low positive and high positive controls; sample diluent; wash concentrate; goat anti-human IgG horseradish peroxidase conjugate; TMB chromogen; and stop solution. {1} J. Substantial Equivalence Information: 1. Predicate device name(s): INOVA QUANTA Plex Profile 6 (anti-SS-A antibodies parameter) 2. Predicate 510(k) number(s): k031450 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | | SS-A 52 Antibody ELISA | Quanta Plex Profile 6 (SS-A bead) | | Indications for Use | To aid in the diagnosis of systemic lupus erythematosus, Sjögren’s syndrome, systemic sclerosis | To aid in the diagnosis of systemic lupus erythematosus, Sjögren’s syndrome, systemic sclerosis | | Analyte measured | Anti-SS-A 52 | Anti-SS-A 52 and 60 | | Calculation of results | Compared to the low positive control | Same | | Test matrix | Serum | Same | | Controls | Negative, low positive, and high positive | Same | | Sample diluent | Tris-buffered saline, Tween 20, protein stabilizers and preservative | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | | SS-A 52 Antibody ELISA | Quanta Plex Profile 6 (SS-A antibody) | | Indications for Use | Addition of polymyositis and dermatomyositis | Not included | | Capture antigen(s) | Purified recombinant SS-A 52 antigen | Native bovine SS-A 60 and recombinant human SS-A 52 antigens | | Units | Arbitrary ELISA Units | Luminex Units | | Result interpretation | ≤20 U/mL = negative; 20-39 = weak positive; 40-80 = mod. positive; >80 strong positive | ≤20 LU/mL = negative; 20-49 = weak positive; 50-100 = mod. positive; >100 strong positive | | Method | ELISA | Fluorescent immunoassay | | Sample dilution | 1:101 | 1:101, 5 μL sample is then diluted 1:10 for a final dilution of 1:1010 | {2} 3 | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Solid phase | Polystyrene microwell plate coated with purified recombinant SS-A 52 antigen | Polystyrene microwell plate containing 7 different beads, one of which is a SS-A bead coated with a mixture of native bovine SS-A 60 and a recombinant human SS-A 52 antigens | | Wash concentrate/buffer | Tris-buffered saline with Tween 20: 40X | No wash buffer | | IgG conjugate | Horseradish peroxidase conjugated goat anti-human IgG | Fluorescently labeled goat anti-human IgG | | Substrate | TMB Chromogen | None | | Stop solution | 0.344M sulfuric acid | None | ## K. Standard/Guidance Document Referenced (if applicable): None referenced ## L. Test Principle: Purified recombinant SS-A 52 antigen is bound to the wells of a polystyrene microwell plate under conditions that will preserve it in an antigenic state. Pre-diluted controls and diluted patient sera are added to separate wells, allowing any anti-SS-A 52 antibodies present to bind to the immobilized antigen. Unbound sample is washed away and an enzyme labeled goat anti-human IgG antibody is added to each well. A second incubation allows the enzyme labeled anti-human IgG to bind to any patient antibodies which have become attached to the microwells. After washing away any unbound enzyme labeled anti-human IgG, the remaining enzyme activity is measured by adding a chromogenic substrate and measuring the intensity of the color that develops. The assay is evaluated by spectrophotometrically measuring and comparing the color intensity that develops in the patient wells with the color in the control wells. Results determined with the assay are interpreted as negative or weak, moderate, or strong positive and are reported in arbitrary units. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: Intra-assay performance for the assay was evaluated by testing 9 specimens a total of 7 times each on 3 lots of kits. The samples tested ranged from 7 to 107 Units with %CVs ranging from 1% to 4%. | | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Mean units | 20 | 22 | 38 | 107 | 53 | 34 | 51 | 7 | 8 | | SD | 0.75 | 0.80 | 1.04 | 1.48 | 1.12 | 0.80 | 0.54 | 0.31 | 0.26 | | %CV | 4% | 4% | 3% | 1% | 2% | 2% | 1% | 4% | 3% | {3} Inter-assay variation was assessed by testing, in duplicate, a panel of 13 specimens twice in one day and once per day for 3 days, for a total of 5 different runs. The samples tested ranged from 2 to 134 U/mL. Percent CVs ranged from 2 to 47%. In the positive range including samples at the cut-off, percent CVs ranged from 2 to 5%. | | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | 13 | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Mean units | 2 | 9 | 9 | 6 | 9 | 22 | 24 | 24 | 45 | 134 | 65 | 42 | 59 | | SD | 0 | 4 | 3 | 1 | 1 | 0 | 0 | 1 | 1 | 4 | 1 | 2 | 1 | | %CV | 17% | 47% | 33% | 21% | 7% | 2% | 2% | 5% | 2% | 3% | 2% | 4% | 2% | b. Linearity/assay reportable range: No claims were made regarding linearity for the assay. Reactivity is related to the quantity of antibody present in a non-linear fashion. While increases and decreases in patient antibody concentrations will be reflected in a corresponding rise or fall in reactivity, the change is not proportional, i.e. a doubling of the antibody concentration will not double the reactivity. c. Traceability, Stability, Expected values (controls, calibrators, or methods): There is no recognized standard or reference material for anti-SS-A 52 autoantibodies. Accelerated stability studies were conducted on 3 lots of antigen coated plates and 3 lots of low and high positive kit controls. Data supported a shelf life of one year. d. Detection limit: To determine the detection limit, a high positive sample was serially diluted until further dilutions did not cause a reduction in the O.D. reading. The lower limit was determined to be 2 Units. e. Analytical specificity: To demonstrate whether high levels of endogenous substances could cause a false positive, sera with known quantities of potential interferant were tested. To determine if the presence of these substances could cause a false negative, the substances were added to two sera with know quantities of anti-SS-A 52. The results were compared to mixing the positive sera with normal sera. | | Normal sera | Positive sera #1 | Recovery | Positive sera #2 | Recovery | | --- | --- | --- | --- | --- | --- | | Substance added: | | Units | % | Units | % | | Mixed with normal sera | — | 44 | | 99 | | | Hemoglobin 1000 mg/dL | 8 Units | 40 | 91% | 96 | 97% | | Bilirubin 29.7 mg/dL | 7 Units | 42 | 95% | 94 | 95% | {4} 5 | | Normal sera | Positive sera #1 | Recovery | Positive sera #2 | Recovery | | --- | --- | --- | --- | --- | --- | | Cholesterol 354 mg/dL | 5 Units | 40 | 91% | 98 | 99% | | Triglycerides 2173 mg/dL | 8 Units | 40 | 91% | 91 | 92% | The package insert contains the following recommendations: “Microbially contaminated, heat-treated, samples with visible particulate should not be used. Grossly hemolyzed or lipemic specimens should be avoided.” The testing of 136 sera with other autoimmune disease antibodies and infectious disease antibodies included the following: 6 samples with other autoimmune antibodies (RNP, Sm, Scl-70, SS-B, and Sm/RNP); 20 rheumatoid arthritis patients, 110 various infectious disease patient sera (HCV, HSV, CMV, toxo, rubella, and parvo). Nine (from the infectious disease group) of 136 (7%) were positive in the new assay. f. Assay cut-off: One hundred and nine normal blood donors and 20 patients with rheumatoid arthritis were tested for anti-SS-A 52 antibodies by the assay. Only one sample was positive at a value of 26 units meaning 99.2% were negative. The cut-off was also validated by ROC analysis. At the established cut-off of <20 Units = negative, the specificity was 95.8%. 2. Comparison studies: a. Method comparison with predicate device: One hundred and nine normal blood donors, 20 patients with rheumatoid arthritis (RA), 110 with high titers against infectious disease organisms, 31 with Sjögren’s syndrome (SS), 149 with systemic lupus erythematosus (SLE), and 176 with systemic sclerosis (SSc) were tested internally on both the new assay and the predicate. | QUANTA Lite SS-A 52 ELISA | QUANTA Plex SS-A Bead | | | | --- | --- | --- | --- | | | Positive | Negative | Total | | Positive | 78 | 36 | 114 | | Negative | 13 | 468 | 481 | | Total | 91 | 504 | 595 | Positive percent agreement: 78/91 = 85.7% (95% CI: 77%-92%) Negative percent agreement: 468/504 = 92.9% (95% CI: 90%-95%) Overall agreement: 546/595 = 91.8% b. Matrix comparison: Both assays use serum as the test matrix. {5} 6 3. Clinical studies: a. Clinical Sensitivity and Specificity: Sera from a total of 763 subjects were tested. These samples included 114 normal blood donors, 20 patients with RA, 110 with high titers against infectious disease organisms, 71 with SS, 149 with SLE, 176 with SSc and 123 with polymyositis (PM) or dermatomyositis (DM). | Groups (N=763) | N | SS-A 52 + | % + | SS-A 52 - | % - | | --- | --- | --- | --- | --- | --- | | Normal | 114 | | | 113/114 | 99.2 | | RA | 20 | | | 20/20 | 100 | | Infectious diseases | 110 | | | 101/110 | 91.8 | | Total disease neg. | 244 | | | 234/244 | 95.9 | | | | | | | | | SS | 71 | 60/71 | 84.5 | | | | SLE | 149 | 46/149 | 30.9 | | | | SSc | 176 | 34/176 | 19.3 | | | | PM (93)/DM (30) | 123 | 90/123 | 73.2 | | | | Total disease pos. | 519 | 230/519 | 44.3 | | | Clinical sensitivity and specificity calculations: | | Disease positive | Disease negative | Total | | --- | --- | --- | --- | | SS-A 52 positive | 230 | 10 | 240 | | SS-A 52 negative | 289 | 234 | 523 | | Total | 519 | 244 | 763 | Clinical sensitivity: 230/519 = 44.3% 95% CI = 40%-49% Clinical specificity: 234/244 = 95.9% 95% CI = 93%-98% b. Other clinical supportive data: In addition, 52 sera known to be positive for other myositis markers were tested. Seventy-five percent (39/52) were positive for SS-A 52 autoantibodies. Clinical study data versus literature data (based on labeling references # 2, 4, 7, 8, 9, and 10): | | Literature | | | INOVA QUANTA Lite™ SS-A 52 | | | | --- | --- | --- | --- | --- | --- | --- | | Disease | N | SS-A 52+ | Sensitivity | N | SS-A 52+ | Sensitivity | | SLE | 123 | 41 | 33% | 149 | 46 | 31% | | SS | 100 | 63 | 63% | 71 | 60 | 85% | | PM/DM*** | 262 | 98 | 37% | 123 | 90 | 73%*** | | Jo-1 | 64 | 42 | 66% | 32 | 25 | 78% | | SSc* | 263 | 31 | 12% | 176 | 34 | 19% | | Controls** | 102 | 1 | 1% | 129 | 1 | 1% | * includes 11 samples negative by Ouchterlony immunodiffusion. {6} ** includes healthy volunteers and rheumatoid arthritis patients ***This population includes a high percentage of patients who are positive for myositis specific antibodies. 4. Clinical cut-off: See assay cut-off and expected values. 5. Expected values/Reference range: The expected result in the normal population is negative (≤ 20 Units). N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is acceptable and supports a substantial equivalence decision. 7
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...