xTAG Cystic Fibrosis 39 Kit v2

K163347 · Luminex Molecular Diagnostics, Inc. · NUA · Dec 15, 2016 · Immunology

Device Facts

Record IDK163347
Device NamexTAG Cystic Fibrosis 39 Kit v2
ApplicantLuminex Molecular Diagnostics, Inc.
Product CodeNUA · Immunology
Decision DateDec 15, 2016
DecisionSESK
Submission TypeSpecial
Regulation21 CFR 866.5900
Device ClassClass 2
AttributesPediatric

Indications for Use

The xTAG® Cystic Fibrosis 39 kit v2 is a device used to simultaneously detect and identify a panel of mutations and variants in the Cystic Fibrosis transmembrane conductance regulator (CFTR) gene in human blood specimens. The panel includes mutations and variants currently recommended by the American College of Medical Genetics and American College of Obstetricians and Gynecologists (ACMG/ACOG), plus some of the world's most common and North American-prevalent mutations.

Device Story

The xTAG Cystic Fibrosis 39 Kit v2 is an in vitro diagnostic test for the simultaneous detection of CFTR gene mutations and variants in human blood. The device utilizes multiplex PCR and allele-specific primer extension followed by hybridization to universal tag-coupled microspheres. The Luminex xMAP technology platform detects the fluorescently labeled microspheres to identify specific mutations. The test is intended for use by laboratory professionals to provide qualitative genotyping results. These results assist clinicians in carrier screening, newborn screening, and diagnostic confirmation. The device is not intended for stand-alone diagnosis or fetal/pre-implantation testing.

Clinical Evidence

No clinical data provided in the document. Substantial equivalence is based on analytical performance and technological comparison to the predicate device.

Technological Characteristics

The device uses multiplex PCR and allele-specific primer extension (ASPE) followed by hybridization to universal tag-coupled microspheres. Detection is performed using the Luminex xMAP platform. The system is a qualitative genotyping test for CFTR gene mutations. It is designed for laboratory use.

Indications for Use

Indicated for carrier testing in adults of reproductive age, as an aid in newborn screening, and in confirmatory diagnostic testing in newborns and children. Not indicated for fetal diagnostic, pre-implantation, or stand-alone diagnostic purposes.

Regulatory Classification

Identification

The CFTR gene mutation detection system is a device used to simultaneously detect and identify a panel of mutations and variants in the CFTR gene. It is intended as an aid in confirmatory diagnostic testing of individuals with suspected cystic fibrosis (CF), carrier identification, and newborn screening. This device is not intended for stand-alone diagnostic purposes, prenatal diagnostic, pre-implantation, or population screening.

Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: CFTR Gene Mutation Detection System.” See § 866.1(e) for the availability of this guidance document.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification OIR Decision Summary To: THE FILE RE: K163347 This 510(k) submission contains information/data on modifications made to the Luminex' own Class II devices requiring 510(k). The following items are present and acceptable: 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. xTAG® Cystic Fibrosis (CFTR) 39 kit v2 510(k) number: K083846 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S): New software threshold values for 2183/2184 variants. The increased thresholds modify the probability of no-call/incorrect call. The FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and {1} Similarities between New Device and Predicate | Item | Modified Device | Predicate Device xTAG® Cystic Fibrosis 39 | | --- | --- | --- | | Intended Use | The xTAG® Cystic Fibrosis 39 kit v2 is a device used to simultaneously detect and identify a panel of mutations and variants in the Cystic Fibrosis transmembrane conductance regulator (CFTR) gene in human blood specimens. The panel includes mutations and variants currently recommended by the American College of Medical Genetics and American College of Obstetricians and Gynecologists (ACMG/ACOG), plus some of the world's most common and North American-prevalent mutations. | Same | | Indications for Use | For carrier testing in adults of reproductive age, as an aid in newborn screening, and in confirmatory diagnostic testing in newborns and children. | Same | | Contra-Indications | The kit is not indicated for use in fetal diagnostic or pre- implantation testing. This kit is also not indicated for stand- alone diagnostic purposes. | Same | | Specimen Type | Peripheral human whole blood | Same | | Methodology | Multiplex PCR followed by multiplex allele specific primer extension for genotyping, hybridized to multiplex fluorescent microparticles, detected by flow cytometry | Same | 2 {2} 3 | Software | TDAS CFTR contains 1 template to detect a panel of 3 mutations and 4 variants. Software has a masking function where user can choose to display results for either the ACMG/ACOG 23 mutations or for the full panel of mutations. | Same | | --- | --- | --- | | Number of alleles detected | 39 mutations and 4 variants in the CFTR gene (Includes 23 ACMG/ACOG recommended mutations) | Same | | Instrument | Luminex 100/200 | Same | | Instrument Software | IS and xPONENT | Same | ## Differences between New Device and Predicate | Item | Modified Device | Predicate Device xTAG® Cystic Fibrosis 60 kit v2 (ICB387) | | --- | --- | --- | | Allelic ratio threshold for 2183/2184 tri-allelic variation (Full Panel) | Allelic ratio (AR) threshold for 2184delA mut allele set to 0.35 (LOW_HET_MUT2 threshold in TDAS CFTR). Process control threshold (PCT) for Presence of the 2184delA mut allele set to 0.40. | Allelic ratio (AR) threshold for 2184delA mut allele set to 0.20 (LOW_HET_MUT2 threshold in TDAS CFTR). Process control threshold (PCT) for Presence of the 2184delA mut allele set to 0.25. | | Allelic ratio threshold for 2183/2184 bi-allelic variation (ACMG Panel) | Lower bound HET mutant AR threshold set to 0.35 (LOW_HET_MUT threshold in TDAS CFTR). PCT of the lower bound HET mutant AR set to 0.40 | Lower bound HET mutant AR threshold set to 0.31 (LOW_HET_MUT threshold in TDAS CFTR). PCT of the lower bound HET mutant AR set to 0.36. | 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis. b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied. These {3} included (1) Report on the Process of Establishing Threshold Values for the 2183/2184 Variation of the xTAG® Cystic Fibrosis (CFTR) 39 kit v2; (2) Accuracy of mutation 2184delA with new thresholds in xTAG® Cystic Fibrosis (CFTR) 39 kit v2 6. A Truthful and Accurate Statement, a 510(k) Summary, and the Indications for Use Enclosure. The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device. 4
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...