K200230 · Inova Diagnostics, Inc. · MVM · Aug 26, 2021 · Immunology
Device Facts
Record ID
K200230
Device Name
Aptiva Celiac Disease IgG Reagent
Applicant
Inova Diagnostics, Inc.
Product Code
MVM · Immunology
Decision Date
Aug 26, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5660
Device Class
Class 2
Attributes
AI/ML
Intended Use
The Aptiva Celiac Disease IgG Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semi-quantitative determination of anti-tissue transglutaminase IgG autoantibodies and anti-deamidated gliadin peptide IgG autoantibodies in human serum. The presence of these antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis, particularly in patients with selective IgA deficiency. The Aptiva Celiac Disease IgG Reagent is intended for use with the Inova Diagnostics Aptiva System.
Device Story
Aptiva Celiac Disease IgG Reagent is a semi-quantitative immunoassay for use on the Inova Diagnostics Aptiva System. It utilizes paramagnetic particles coated with recombinant tissue transglutaminase (tTG) and synthetic deamidated gliadin peptide (DGP), plus a control particle coated with goat anti-human IgG. The system dilutes patient serum, incubates with particles, performs wash cycles, and adds a phycoerythrin (PE)-labeled anti-human IgG conjugate. The system captures multiple images to identify particles and measure median fluorescent intensity (MFI). MFI is proportional to bound IgG concentration. Software calculates Fluorescent Light Units (FLU) using lot-specific master curves and instrument-specific working curves. Results are reported as positive or negative based on a 5 FLU cut-off. The device is intended for clinical laboratory use to aid in diagnosing celiac disease and dermatitis herpetiformis, particularly in IgA-deficient patients.
Clinical Evidence
Clinical validation used 515 samples: 171 celiac disease (CD), 20 IgA-deficient CD, 34 dermatitis herpetiformis (DH), and 290 controls (autoimmune/infectious diseases). For CD diagnosis, Aptiva DGP IgG showed 82.2% sensitivity and 97.9% specificity; Aptiva tTG IgG showed 60.7% sensitivity and 100% specificity. For DH diagnosis, DGP IgG showed 70.6% sensitivity and 97.9% specificity; tTG IgG showed 26.5% sensitivity and 100% specificity. Method comparison with predicates showed TPA of 86.7% (DGP) and 86.8% (tTG).
Technological Characteristics
Particle-based multi-analyte technology (PMAT) using paramagnetic particles. Reagents include tTG/DGP-coated particles, PE-labeled anti-human IgG tracer, and buffers. Energy source: Aptiva System instrumentation. Connectivity: RFID-labeled reagent cartridges for master curve data. Software: Automated image-based particle counting and fluorescence quantification. Sterilization: Not specified.
Indications for Use
Indicated for semi-quantitative detection of anti-tTG IgG and anti-DGP IgG in human serum to aid in diagnosis of celiac disease and dermatitis herpetiformis, specifically in patients with selective IgA deficiency.
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
Predicate Devices
QUANTA Flash DGP IgG (k113863)
QUANTA Flash h-tTG IgG (k101644)
Related Devices
K193604 — Aptiva Celiac Disease IgA Reagent · Inova Diagnostics, Inc. · Jun 16, 2021
K072632 — QUANTA PLEX CELIAC IGG PROFILE · Inova Diagnostics, Inc. · Feb 11, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
August 26, 2021
Inova Diagnostics, Inc. Andrea Seaman Manager, Research and Development 9900 Old Grove Road San Diego, California 92131
Re: K200230
Trade/Device Name: Aptiva Celiac Disease IgG Reagent Regulation Number: 21 CFR 866.5660 Regulation Name: Multiple Autoantibodies Immunological Test System Regulatory Class: Class II Product Code: MVM, MST Dated: January 27, 2020 Received: January 30, 2020
Dear Andrea Seaman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Ying Mao Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200230
Device Name Aptiva Celiac Disease IgG Reagents
#### Indications for Use (Describe)
The Aptiva Celiac Disease IgG Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semiquantitative determination of anti-tissue transglutaminase IgG autoantibodies and anti-deamidated gliadin peptide IgG autoantibodies in human serum. The presence of these antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis, particularly in patients with selective IgA deficiency.
The Aptiva Celiac Disease IgG Reagent is intended for use with the Inova Diagnostics Aptiva System.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Inova Diagnostics, a Werfen Company. The logo consists of a stylized checkmark in red and blue, followed by the text "Inova Diagnostics" in a sans-serif font. Below the company name is the text "A Werfen Company" in a smaller font size.
# 510(k) Summary
## Aptiva Celiac Disease IgG Reagent
| Table of Contents | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Administrative data…………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………… | |
| Predicate device…………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………… | |
| Device description | |
| Intended use(s) | |
| Indications for use | |
| Substantial equivalence | |
| Comparison to predicate device | |
| Analytical performance characteristics | |
| Quantitation and units of measure | |
| Precision | |
| Reproducibility Studies | |
| Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ) | |
| Analytical Measuring Range (AMR) | |
| Auto-rerun function and reportable results | |
| High concentration hook effect | |
| Linearity | |
| Interference | |
| Sample Stability and Handling | |
| Reagent Stability | |
| Cut-off, reference range | |
| Clinical performance characteristics | |
| Clinical sensitivity, specificity | |
| Expected values | |
| Comparison with predicate device | |
| | |
{4}------------------------------------------------
This summary of the 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510K number is: K200230
## Administrative data
| Submitter: | Inova Diagnostics, Inc<br>9900 Old Grove Road,<br>San Diego, CA, 92131 | | |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|--|
| Purpose of submission: | New device | | |
| Device in the submission: | Aptiva Celiac Disease IgG Reagent | | |
| Revision Date: | August 23, 2021 | | |
| Scientific contact: | Andrea Seaman, Manager, Research and Development<br>Inova Diagnostics, Inc.<br>9900 Old Grove Road, San Diego, CA, 92131<br>Phone: 858-586-9900 x1395<br>Fax: 858-863-0025<br>Email: aseaman@inovadx.com | | |
| Quality Systems contact: | Ronda Elliott, VP, Quality Systems and RA<br>Inova Diagnostics, Inc<br>9900 Old Grove Road, San Diego, CA, 92131<br>Phone: 858-586-9900 x1381<br>Fax: 858-863-0025<br>Email: relliott@inovadx.com | | |
| Device name (kit): | Proprietary name: | Aptiva Celiac Disease IgG Reagent | |
| | Common name: | anti-deamidated gliadin peptide antibody<br>immunoassay, anti-tissue transglutaminase<br>antibody immunoassay | |
| | Classification name: | DGP IgG: Radioallergosorbent (RAST)<br>immunological test system<br>tTG IgG: Multiple autoantibodies immunological<br>test system | |
| Regulation Medical Specialty | Immunology | | |
| Review Panel | Immunology | | |
| Product Code | DGP IgG: MST | | |
{5}------------------------------------------------
| | tTG IgG: MVM |
|-------------------|--------------------|
| Regulation Number | 866.5750, 866.5660 |
| Device Class | 2 |
#### Predicate device
QUANTA Flash® DGP IgG, 510(k) number: k113863. Date declared: October 23, 2012. QUANTA Flash® h-tTG IgG, 510(k) number: k101644. Date declared: March 23, 2011.
#### Device description
The Aptiva Celiac Disease IgG reagent utilizes particle based multi-analyte technology (PMAT) in a cartridge format. Each analyte (tissue transglutaminase [tTG] and deamidated gliadin peptide [DGP]) in the Aptiva Celiac Disease IgG reagent is a solid phase immunoassay utilizing fluorescent microparticles. This technology allows each of the two analytes, along with a human IgG capture antibody (IgG Control Microparticle), to be coated onto three uniquely recognizable paramagnetic microparticles, which are combined into one tube.
The Aptiva instrument is a fully automated, random access analyzer. This platform is a closed system with continuous load and random-access capabilities that processes the samples, runs the reagent and reports results. It includes liquid handling hardware, optical module (OM), and integrated computer with proprietary software and touch screen user interface.
The two analyte microparticles, along with the control microparticle, are stored in the reagent cartridge under conditions that preserve the proteins in their reactive states. When the assay cartridge is ready to be used for the first time, the reagent tube seals are pierced using the cartridge lid. The reagent cartridge is then loaded onto the Aptiva instrument, where the microparticles are automatically rehydrated using buffer located within the cartridge.
A patient's serum is diluted 1:23 with Aptiva system rinse by the instrument in a disposable cuvette. A small amount of the diluted sample is combined with assay buffer and the microparticle suspension in a second cuvette, and mixed (final serum dilution: 1:230). This reaction cuvette is incubated for 9 ½ minutes at 37°C. The cuvette is then exposed to a small magnet that holds the microparticles in place. The liquid is aspirated, and the microparticles are resuspended as system rinse is added to the cuvette and the magnet is removed. This wash cycle is repeated one more time. During the third wash, no system rinse is added after the aspiration step. After the third wash, phycoerythrin conjugated polyclonal anti-human lgG (known as PE Tracer IgG) is added to the microparticles in the cuvette, and mixed. Again, the cuvette is incubated for 9 ½ minutes at 37℃. Three wash steps, as described above, are performed on the microparticles. Following the wash steps, the microparticles are transferred to the of the instrument, where a charge coupled device (CCD) camera takes multiple images in order to identify and count the three unique microparticle regions, as well as determine the amount of conjugate on the microparticles. The control microparticle, a third particle, coated with goat anti-human IgG, is included in the reagent in as a control to flag low concentrations of IgG the patient serum sample as an assay verification step. The median fluorescent intensity (MFI) is proportional to the amount of PE Tracer that is bound to the human IgG, which is proportional to the amount of IgG antibodies bound to the corresponding microparticle regions.
{6}------------------------------------------------
For quantitation, the DGP IgG and tTG IgG assays (together as part of the Aptiva Celiac Disease IgG Reagent) each utilizes a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge RFID tag. Every new lot of reagent cartridge must be calibrated before first use with the reagent specific calibrators. Based on the results obtained with the calibrators included in the Aptiva Celiac Disease IgG Calibrator kit (sold separately), an instrument specific Working Curve is created for each assay, which is used to calculate reported fluorescent light units (FLU) from the median fluorescent intensity (MFI) instrument signal obtained for each sample, on each of the two assays within the reagent.
Aptiva Celiac Disease IgG Calibrators and Aptiva Celiac Disease IgG Controls are sold separately.
The Aptiva Celiac Disease IgG Reagent kit contains the following materials:
One (1) Aptiva Celiac Disease IgG Reagent Cartridge, containing the following reagents for 200 determinations:
- a. Aptiva Celiac Disease IgG microparticle containing 3 unique microparticle regions coated with recombinant tissue transglutaminase, deamidated gliadin peptide, or goat antihuman IgG antibody.
- b. Assay buffer - colored pink, containing protein stabilizers and preservatives.
- C. PE Tracer IgG - phycoerythrin (PE) labeled anti-human IgG antibody, containing buffer, protein stabilizers and preservative.
- ರ. Rehydration Buffer - containing protein stabilizers and preservatives.
#### Intended use(s)
The Aptiva Celiac Disease IgG Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semi-quantitative determination of anti-tissue transglutaminase lgG autoantibodies and anti-deamidated gliadin peptide IgG autoantibodies in human serum. The presence of these antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis, particularly in patients with selective IgA deficiency.
The Aptiva Celiac Disease IgG Reagent is intended for use with the Inova Diagnostics Aptiva System.
#### Indications for use
Same as intended use.
#### Substantial equivalence
The Aptiva Celiac Disease IgG Reagent has the same intended use and assay principle as the predicate devices.
{7}------------------------------------------------
## Page 5 of 19
### Comparison to predicate device
Aptiva Celiac Disease IgG Reagent – DGP IgG Assay
| Similarities | | |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item | Aptiva Celiac Disease IgG Reagent (DGP IgG) | QUANTA Flash DGP IgG |
| Intended Use | The Aptiva Celiac Disease IgG Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semi-quantitative determination of anti-tissue transglutaminase IgG autoantibodies and anti-deamidated gliadin peptide IgG autoantibodies in human serum. The presence of these antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis, particularly in patients with selective IgA deficiency.<br><br>The Aptiva Celiac Disease IgG Reagent is intended for use with the Inova Diagnostics Aptiva System. | The QUANTA Flash DGP IgG is a chemiluminescent immunoassay (CIA) for the semi-quantitative determination of IgG anti-deamidated gliadin peptide (DGP) antibodies in human serum. The presence of IgG anti-DGP antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of celiac disease in IgA sufficient and IgA deficient patients, as well as dermatitis herpetiformis. |
| Assay Methodology | solid phase (heterogeneous) immunoassay | solid phase (heterogeneous) immunoassay |
| Antigen | deamidated gliadin peptide | deamidated gliadin peptide |
| Sample Type | human serum | human serum |
| Solid Phase | paramagnetic microparticles | paramagnetic microparticles |
| Differences | | |
| Item | Aptiva Celiac Disease IgG Reagent (DGP IgG) | QUANTA Flash DGP IgG |
| Detection/Operating Principle | fluorescent immunoassay | chemiluminescent immunoassay |
| Conjugate | phycoerythrin conjugated polyclonal anti-human IgG antibody | Isoluminol conjugated monoclonal anti-human IgG antibody |
| Units | fluorescent light units (FLU) | chemiluminescent units (CU) |
| Cut-off | 5.00 FLU | 20.0 CU |
| Analytical Measuring<br>Range | 0.56 FLU – 250.00 FLU | 2.8 CU – 1936.7 CU |
| Control | Controls have lot specific values<br>assigned.<br>Control 1<br>DGP IgG – 6.72 FLU<br>Control 2<br>DGP IgG – 14.12 FLU (8.47 – 19.77) | Controls have lot specific values<br>assigned.<br>Negative Control<br>DGP IgG – 10.9 CU (6.5 – 15.3)<br>Positive Control<br>DGP IgG – 52.4 CU (31.4 – 73.4) |
| Calibration | Lot specific Master Curve + 3<br>calibrators (sold separately) | Lot specific Master Curve + 2<br>calibrators (sold separately) |
{8}------------------------------------------------
#### Aptiva Celiac Disease IgG Reagent – tTG IgG Assay
| Similarities | | |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item | Aptiva Celiac Disease IgG Reagent (tTG IgG) | QUANTA Flash tTG IgG |
| Intended Use | The Aptiva Celiac Disease IgG Reagent is an immunoassay utilizing particle-based multi-analyte technology for the semi-quantitative determination of anti-tissue transglutaminase IgG autoantibodies and anti-deamidated gliadin peptide IgG autoantibodies in human serum. The presence of these antibodies, in conjunction with clinical findings and other laboratory tests, is an aid in the diagnosis of celiac disease and dermatitis herpetiformis, particularly in patients with selective IgA deficiency.<br>The Aptiva Celiac Disease IgG Reagent is intended for use with the Inova Diagnostics Aptiva System. | The QUANTA Flash h-tTG IgG is a chemiluminescent immunoassay (CIA) for the semi-quantitative detection of IgG anti-human tissue transglutaminase (h-tTG) antibodies in human serum. The presence of IgG anti-h-tTG antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of the gluten sensitive enteropathy celiac disease, particularly in patients with selective IgA deficiency. |
| Assay Methodology | solid phase (heterogeneous) immunoassay | solid phase (heterogeneous) immunoassay |
| Antigen | recombinant tissue transglutaminase | recombinant tissue transglutaminase |
| Sample Type | human serum | human serum |
| Solid Phase | paramagnetic microparticles | paramagnetic microparticles |
Page 6 of 19
{9}------------------------------------------------
| Item | Aptiva Celiac Disease IgG Reagent<br>(tTG IgG) | QUANTA Flash tTG IgG |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Detection/Operating<br>Principle | fluorescent immunoassay | chemiluminescent immunoassay |
| Conjugate | Phycoerythrin conjugated<br>polyclonal anti-human IgG<br>antibody | Isoluminol conjugated<br>monoclonal anti-human IgG<br>antibody |
| Units | fluorescent light units (FLU) | chemiluminescent units (CU) |
| Cut-off | 5.00 FLU | 20.0 CU |
| Analytical Measuring<br>Range | 0.82 FLU - 250.00 FLU | 3.75 CU - 2560.0 CU |
| Control | Controls have lot specific values<br>assigned.<br>Control 1<br>tTG IgA – 8.18 FLU<br>Control 2<br>tTG IgA - 16.39 FLU (9.83 – 22.95) | Controls have lot specific values<br>assigned.<br>Negative Control<br>h-tTG IgA - 11.4 CU (6.8 – 16.0)<br>Positive Control<br>h-tTG IgA - 50.4 CU (30.2 – 70.6) |
| Calibration | Lot specific Master Curve + 3<br>calibrators (sold separately) | Lot specific Master Curve + 2<br>calibrators (sold separately) |
## Analytical performance characteristics
## Quantitation and units of measure
For quantitation, the Aptiva Celiac Disease lgG reagent utilizes predefined lot specific Master Curves, one for tTG IgG and one for DGP IgG that is uploaded onto the instrument through the reagent cartridge RFID. The analyte specific Master Curves are generated at Inova for each reagent lot, where in-house Master Curve Standards with assigned FLU values are run multiple times. The resulting MFI values generated are used to create a unique 4 parameter logistic (4PL) curve for each of the two analytes. The lgG control bead is present in the reagent as a control to flag low concentration lgG in a patient serum sample as an assay verification step. This microparticle also has an in-house standard which is run each time a new reagent lot is manufactured. The MFI produced by this standard is used as the cut-off threshold for the IgG control microparticle for that reagent lot. These four parameters of the analyte curves, as well as the MFI cut-off for the IgG control microparticle are embedded in the reagent cartridge RFID.
| Material | Assigned Value (FLU) |
|------------------------------------|----------------------|
| Celiac IgG Master Curve Standard 1 | 0.00 |
| Celiac IgG Master Curve Standard 2 | 3.81 |
| Celiac IgG Master Curve Standard 3 | 8.97 |
| Celiac IgG Master Curve Standard 4 | 47.37 |
| Celiac IgG Master Curve Standard 5 | 250.07 |
List of Aptiva Celiac Disease IgG Master Curve Standards - DGP IgG:
List of Aptiva Celiac Disease IgG Master Curve Standards – tTG IgG
{10}------------------------------------------------
| Material | Assigned Value (FLU) |
|------------------------------------|----------------------|
| Celiac IgG Master Curve Standard 1 | 0.00 |
| Celiac IgG Master Curve Standard 2 | 2.62 |
| Celiac IgG Master Curve Standard 3 | 31.14 |
| Celiac IgG Master Curve Standard 4 | 73.23 |
| Celiac IgG Master Curve Standard 5 | 386.85 |
#### Precision
The precision of the Aptiva Celiac Disease lgG reagent was evaluated on 8 samples for DGP IgG and tTG lgG, containing various concentrations of antibodies in accordance with CLSI EPO5-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline. Samples were run in duplicates, twice a day, for 20 days.
Data were analyzed with the Analyse-it for Excel method evaluation software, and repeatability (withinrun), between run, between day and within-laboratory precision) were calculated. Results are summarized in the two tables below.
| DGP IgG Precision | | | Repeatability | | Between Run | | Between Day | | Within Laboratory | |
|-------------------|-------------------|---------------|---------------|------|-------------|------|-------------|------|-------------------|------|
| Sample | Replicates<br>(N) | Mean<br>(FLU) | SD<br>(FLU) | CV | SD<br>(FLU) | CV | SD<br>(FLU) | CV | SD<br>(FLU) | CV |
| 1 | 80 | 1.73 | 0.1 | 6.1% | 0.07 | 4.0% | 0.09 | 5.1% | 0.15 | 8.9% |
| 2 | 80 | 4.29 | 0.17 | 3.9% | 0.07 | 1.6% | 0.19 | 4.5% | 0.27 | 6.2% |
| 3 | 80 | 4.47 | 0.17 | 3.7% | 0.19 | 4.3% | 0.14 | 3.1% | 0.29 | 6.5% |
| 4 | 80 | 5.42 | 0.23 | 4.3% | 0.30 | 5.5% | 0.22 | 4.0% | 0.44 | 8.1% |
| 5 | 80 | 16.17 | 0.38 | 2.3% | 0.48 | 2.9% | 0.81 | 5.0% | 1.01 | 6.3% |
| 6 | 80 | 31.27 | 0.81 | 2.6% | 1.00 | 3.2% | 0.90 | 2.9% | 1.57 | 5.0% |
| 7 | 80 | 131.92 | 3.25 | 2.5% | 6.32 | 4.8% | 0.94 | 0.7% | 7.16 | 5.4% |
| 8 | 80 | 216.46 | 9.76 | 4.5% | 7.11 | 3.3% | 12.23 | 5.6% | 17.19 | 7.9% |
Acceptance criteria: Total %CV: < 12% or SD < 0.6 FLU
| tTG IgG Precision | | | Repeatability | | Between Run | | Between Day | | Within Laboratory | |
|-------------------|-------------------|---------------|---------------|------|-------------|------|-------------|------|-------------------|------|
| Sample | Replicates<br>(N) | Mean<br>(FLU) | SD<br>(FLU) | CV | SD<br>(FLU) | CV | SD<br>(FLU) | CV | SD<br>(FLU) | CV |
| 1 | 80 | 1.88 | 0.13 | 7.1% | 0.00 | 7.1% | 0.14 | 3.0% | 0.14 | 7.7% |
| 2 | 80 | 2.43 | 0.15 | 6.2% | 0.04 | 1.7% | 0.12 | 5.1% | 0.20 | 8.2% |
| 3 | 80 | 4.53 | 0.23 | 5.1% | 0.05 | 1.2% | 0.18 | 3.9% | 0.30 | 6.5% |
| 4 | 80 | 5.29 | 0.27 | 5.1% | 0.00 | 0.0% | 0.15 | 2.8% | 0.31 | 5.8% |
| 5 | 80 | 11.49 | 0.42 | 3.7% | 0.12 | 1.0% | 0.44 | 3.8% | 0.62 | 5.4% |
| 6 | 80 | 46.27 | 1.40 | 3.0% | 1.04 | 2.3% | 1.02 | 2.2% | 2.02 | 4.4% |
| 7 | 80 | 117.59 | 6.73 | 5.7% | 2.14 | 1.8% | 2.51 | 2.1% | 7.49 | 6.4% |
| 8 | 80 | 210.59 | 14.77 | 7.0% | 0.00 | 0.0% | 9.17 | 4.4% | 17.38 | 8.3% |
{11}------------------------------------------------
### Reproducibility Studies
#### Reproducibility between sites (instruments)
Seven samples for DGP IgG and tTG IgG were tested according to CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline, at three different sites. Samples were run in replicates of 5, once a day, for 5 days, to generate 25 data points per site. Data were analyzed with the Analyse-it for Excel method evaluation software to calculate between site precision. Acceptance criteria: Reproducibility Between-Site %CV: < 12% or SD < 0.6 FLU Results are summarized in the tables below.
| Aptiva DGP IgG | | Repeatability | | Between-Day | | Within-Site | | Between-<br>Site/Instrument | | Reproducibility | | |
|----------------|----|---------------|-------------|-------------|-------------|-------------|-------------|-----------------------------|-------------|-----------------|-------------|-------|
| Sample | N | Mean<br>(FLU) | SD<br>(FLU) | CV | SD<br>(FLU) | CV | SD<br>(FLU) | CV | SD<br>(FLU) | CV | SD<br>(FLU) | CV |
| 1 | 75 | 2.26 | 0.25 | 11.2% | 0.00 | 0.0% | 0.25 | 11.2% | 0.14 | 6.2% | 0.29 | 12.8% |
| 2 | 75 | 5.41 | 0.27 | 5.0% | 0.32 | 5.9% | 0.42 | 7.7% | 0.27 | 5.1% | 0.50 | 9.2% |
| 3 | 75 | 6.42 | 0.35 | 5.4% | 0.38 | 5.9% | 0.51 | 8.0% | 0.24 | 3.8% | 0.57 | 8.9% |
| 4 | 75 | 12.40 | 0.57 | 4.6% | 0.86 | 7.0% | 1.04 | 8.4% | 0.00 | 0.0% | 1.04 | 8.4% |
| 5 | 75 | 44.17 | 1.40 | 3.2% | 2.34 | 5.3% | 2.73 | 6.2% | 2.16 | 4.9% | 3.48 | 7.9% |
| 6 | 75 | 119.06 | 6.15 | 5.2% | 5.38 | 4.5% | 8.17 | 6.9% | 6.72 | 5.6% | 10.58 | 8.9% |
| 7 | 75 | 160.22 | 10.82 | 6.8% | 4.54 | 2.8% | 11.74 | 7.3% | 9.78 | 6.1% | 15.28 | 9.5% |
| Aptiva tTG IgG | | Repeatability | | | Between-Day | | Within-Site | | Between-<br>Site/Instrument | | Reproducibility | |
|----------------|----|---------------|-------------|------|-------------|------|-------------|------|-----------------------------|------|-----------------|------|
| Sample | N | Mean<br>(FLU) | SD<br>(FLU) | CV | SD<br>(FLU) | CV | SD<br>(FLU) | CV | SD<br>(FLU) | CV | SD<br>(FLU) | CV |
| 1 | 75 | 2.33 | 0.12 | 5.0% | 0.11 | 4.5% | 0.16 | 6.7% | 0.14 | 6.2% | 0.21 | 9.1% |
| 2 | 75 | 5.33 | 0.25 | 4.5% | 0.19 | 3.4% | 0.31 | 5.7% | 0.32 | 5.8% | 0.45 | 8.1% |
| 3 | 75 | 6.03 | 0.25 | 4.1% | 0.18 | 3.0% | 0.30 | 5.0% | 0.34 | 5.6% | 0.45 | 7.5% |
| 4 | 75 | 12.12 | 0.38 | 3.1% | 0.23 | 1.9% | 0.44 | 3.7% | 0.45 | 3.7% | 0.63 | 5.2% |
| 5 | 75 | 30.43 | 1.35 | 4.4% | 0.55 | 1.8% | 1.46 | 4.8% | 1.12 | 3.7% | 1.84 | 6.0% |
| 6 | 75 | 96.08 | 5.94 | 6.2% | 5.19 | 5.4% | 7.88 | 8.2% | 4.31 | 4.5% | 8.98 | 9.4% |
| 7 | 75 | 192.87 | 13.07 | 6.8% | 4.58 | 2.4% | 13.85 | 7.2% | 8.86 | 4.6% | 16.44 | 8.5% |
#### Reproducibility between lots
Six samples for DGP IgG and seven samples for tTG IgG were tested according to CLSI EP05-A3 Evaluation of Precision of Quantitative Measurement Procedures: Approved Guideline, using three different lots. Samples were run in replicates of 5, once a day, for 5 days, to generate 25 data points per lot, 75 data points total for each sample. Data were analyzed with the Analyse-it for Excel method evaluation software to calculate between lot precision.
Acceptance criteria: Reproducibility Between-Lot %CV: < 12% or SD < 0.6 FLU
{12}------------------------------------------------
| Aptiva DGP IgG | | Repeatability | | Between-Day | | Within-Lot | | Between-Lot | | Reproducibility | | |
|----------------|----|---------------|-------------|-------------|-------------|------------|-------------|-------------|-------------|-----------------|-------------|-------|
| Sample | N | Mean<br>(FLU) | SD<br>(FLU) | CV | SD<br>(FLU) | CV | SD<br>(FLU) | CV | SD<br>(FLU) | CV | SD<br>(FLU) | CV |
| 1 | 75 | 2.95 | 0.13 | 4.4% | 0.30 | 10.3% | 0.33 | 11.2% | 0.00 | 0.0% | 0.33 | 11.2% |
| 2 | 75 | 5.79 | 0.25 | 4.3% | 0.45 | 7.7% | 0.51 | 8.9% | 0.40 | 7.0% | 0.65 | 11.3% |
| 3 | 75 | 8.88 | 0.41 | 4.6% | 0.80 | 9.0% | 0.89 | 10.0% | 0.26 | 2.9% | 0.93 | 10.5% |
| 4 | 75 | 24.09 | 0.80 | 3.3% | 1.36 | 5.7% | 1.58 | 6.6% | 0.00 | 0.0% | 1.58 | 6.6% |
| 5 | 75 | 93.99 | 2.60 | 2.8% | 5.68 | 6.0% | 6.25 | 6.7% | 5.13 | 5.5% | 8.09 | 8.6% |
| 6 | 75 | 122.73 | 5.36 | 4.4% | 5.23 | 4.3% | 7.49 | 6.1% | 10.70 | 8.7% | 13.06 | 10.6% |
Results are summarized in the tables below.
| Aptiva tTG IgG | | Repeatability | | Between-Day | | Within-Lot | | Between-Lot | | Reproducibility | | |
|----------------|----|---------------|-------------|-------------|-------------|------------|-------------|-------------|-------------|-----------------|-------------|------|
| Sample | N | Mean<br>(FLU) | SD<br>(FLU) | CV | SD<br>(FLU) | CV | SD<br>(FLU) | CV | SD<br>(FLU) | CV | SD<br>(FLU) | CV |
| 1 | 75 | 1.87 | 0.12 | 6.2% | 0.06 | 3.0% | 0.13 | 6.9% | 0.03 | 1.8% | 0.13 | 7.1% |
| 2 | 75 | 5.17 | 0.21 | 4.1% | 0.15 | 2.9% | 0.26 | 5.0% | 0.38 | 7.4% | 0.46 | 8.9% |
| 3 | 75 | 5.68 | 0.22 | 3.8% | 0.18 | 3.2% | 0.28 | 5.0% | 0.33 | 5.9% | 0.44 | 7.7% |
| 4 | 75 | 11.01 | 0.40 | 3.6% | 0.61 | 5.5% | 0.73 | 6.6% | 0.41 | 3.7% | 0.83 | 7.6% |
| 5 | 75 | 25.83 | 1.12 | 4.3% | 0.99 | 3.8% | 1.49 | 5.8% | 1.42 | 5.5% | 2.05 | 8.0% |
| 6 | 75 | 73.07 | 3.51 | 4.8% | 6.18 | 8.5% | 7.11 | 9.7% | 0.00 | 0.0% | 7.11 | 9.7% |
| 7 | 75 | 170.49 | 11.27 | 6.6% | 6.87 | 4.0% | 13.20 | 7.7% | 4.66 | 2.7% | 14.00 | 8.2% |
#### Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ)
The LoB, LoD, and LoQ of the DGP IgG and tTG IgG assays in the Aptiva Celiac Disease IgG Reagent were calculated separately by a study according to CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline- Second Edition.
#### Study protocol for LoB:
Eight blank samples (Aptiva System Rinse) from two different lots were run in replicates of five on two reagent lots, once per day, for 3 days, with 120 data points generated on each lot. The LoB was determined for each assay, on each reagent lot separately with the Analyse-it for Excel software's Reference Interval function, at the 95th percentile, using the non-parametric method for all 4 analyses (DGP IgG on two reagent lots, and tTG IgG on two reagent lots; all having a p-value = <0.0001)
The DGP IgG LoB for one reagent lot was determined as 0.01 FLU (41 MFI), and for the second reagent lot as 0.02 FLU (60 MFI). The final LoB value for DGP IgG is 0.02 FLU.
The tTG IgG LoB for one reagent lot was determined as 0.00 FLU (37 MFI), and for the second reagent lot as 0.02 FLU (34 MFI). The final LoB value for tTG IgG is 0.02 FLU.
{13}------------------------------------------------
#### Study protocol for LoD:
Four low level samples for each DGP IgG and tTG IgG assay (prepared by mixing human serum samples with high and low levels of antibodies) were run in replicates of five on two reagent lots, twice per day, for 3 days, with 120 data points generated on each assay, on each reagent lot. The LoD was determined separately for each assay, on each reagent lot.
The DGP IgG limit of detection for one reagent lot was determined as 0.15 FLU, and for the second reagent lot as 0.13 FLU. The final LoD value is 0.15 FLU.
The tTG IgG limit of detection for one reagent lot was determined as 0.13 FLU, and for the second reagent lot as 0.12 FLU. The final LoD value is 0.13 FLU.
#### Study protocol for LoQ:
Four low level samples for each DGP IgG and tTG IgG assay (prepared by mixing human serum samples with high and low levels of antibodies) were run in replicates of five on two reagent lots, twice per day, for 3 days, with 120 data points generated on each assay, on each reagent lot. The LoQ was determined separately for each assay, on each reagent lot. The LoQ was determined in each case by calculating the total imprecision of each sample (acceptance criteria: total imprecision <20%).
The DGP IgG limit of quantitation for one reagent lot was determined as 0.53 FLU, and for the second reagent lot as 0.56 FLU. The final LoQ value is 0.56 FLU, which has been set as the lower limit of the analytical measuring range of the DGP IgG assay.
The tTG IgG limit of quantitation for one reagent lot was determined as 0.80 FLU, and for the second reagent lot as 0.82 FLU. The final LoQ value is 0.82 FLU, which has been set as the lower limit of the analytical measuring range of the tTG IgG assay.
#### Analytical Measuring Range (AMR)
| Within the Aptiva Celiac Disease IgG Reagent: | |
|-----------------------------------------------|-----------------------|
| DGP IgG: | 0.56 FLU - 250.00 FLU |
| tTG IgG: | 0.82 FLU - 250.00 FLU |
#### Auto-rerun function and reportable results
The Aptiva software has an auto-rerun option available. If this option is selected, the instrument will automatically rerun any sample that has a result >250.00 FLU for DGP IgG or tTG IgG after performing an additional 10-fold dilution, thereby bringing the measured value within the AMR. The reported result will be calculated by the software factoring the additional dilution. As the highest value that can be measured is 2500.00 FLU for DGP IgG or tTG IgG.
{14}------------------------------------------------
#### High concentration hook effect
To assess hook effect, 3 samples for DGP lgG and 3 samples for tTG IgG were tested at two increasing 2fold serial dilutions from the standard 1:23 dilution used by the Aptiva Celiac Disease IgG Reagent. All FLU values above the analytical measuring ranges of the two assays are theoretical and were mathematically calculated using the 4 parameters of their respective calibration curves. All samples showed increase in FLU values as dilution factor became more concentrated, thereby confirming that high positive specimens above the AMR do not show hook effect up to 723.10 FLU for the DGP IgG assay and 465.52 FLU for the tTG IgG assay (theoretical values calculated) in the Aptiva Celiac Disease IgG Reagent.
#### Linearity
The Linearity of the AMR was calculated separately for DGP IgG as part of the Aptiva Celiac Disease IgG Reagent.
The linearity of the AMR of DGP IgG and tTG IgG was evaluated by a study according to CLSI EPO6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approved Guideline. Four human serum samples for DGP IgG and three human serum samples for tTG IgG with various antibody concentrations were serially diluted to obtain values that cover the entire AMR. The dilutions were assayed in duplicates. Results were analyzed according to the guideline performing regression analysis and identifying the best fitting polynomial.
Acceptance criteria:
- Best fitting polynomial is a linear one, otherwise, the difference between the best-fitting nonlinear and linear polynomial is less than 15% or ±0.75 FLU for low level samples (allowable nonlinearity).
| Sample | Test Range<br>(FLU) | Slope<br>(95% CI) | Y-Intercept<br>(95% CI) | R² | Average % Recovery |
|----------|---------------------|------------------------|----------------------------|------|--------------------|
| 1 | 54.85 - 274.25 | 1.04<br>(1.00 to 1.09) | -8.73<br>(-16.69 to -0.77) | 0.99 | 98.0% |
| 2 | 7.13 - 71.33 | 1.04<br>(1.02 to 1.06) | -2.29<br>(-3.27 to -1.31) | 1.00 | 95.1% |
| 3 | 4.82 - 48.20 | 1.04<br>(1.00 to 1.08) | -0.16<br>(-1.26 to 0.94) | 0.99 | 101.8% |
| 4 | 0.52 - 5.19 | 1.02<br>(0.99 to 1.05) | -0.15<br>(-0.24 to -0.05) | 1.00 | 95.4% |
| Combined | 0.52 - 274.25 | 1.00<br>(0.99 to 1.01) | -0.50<br>(-1.49 to 0.50) | 1.00 | 97.6% |
For DGP IgG, the best fitting polynomial found for samples 1, 2 and 4 was a linear one, while third order polynomial was found to provide the best fit for sample 3. All acceptance criteria were fulfilled.
{15}------------------------------------------------
| Sample | Test Range<br>(FLU) | Slope<br>(95% CI) | Y-Intercept<br>(95% CI) | R² | Average % Recovery |
|----------|---------------------|------------------------|---------------------------|------|--------------------|
| 1 | 32.78 - 327.80 | 0.93<br>(0.88 to 0.99) | 13.20<br>(1.59 to 24.82) | 0.99 | 103.5% |
| 2 | 7.48 - 74.77 | 1.00<br>(0.96 to 1.05) | 1.40<br>(-0.72 to 3.53) | 0.99 | 104.8% |
| 3 | 0.99 - 9.91 | 1.03<br>(1.00 to 1.06) | -0.22<br>(-0.42 to -0.02) | 1.00 | 96.3% |
| Combined | 0.99 - 327.80 | 0.98<br>(0.96 to 1.00) | 2.32<br>(-0.08 to 4.73) | 0.99 | 101.6% |
For tTG IgG, the best fitting polynomial found for samples 1, 2 and 3 was a linear one. All acceptance criteria were fulfilled.
These data demonstrate the linearity of the analytical measuring range (0.56 FLU – 250.00 FLU) of the DGP lgG assay and the analytical measuring range 0.82 FLU - 250.00 FLU) of the tTG IgG assay, both as part of the Aptiva Celiac Disease IgG Reagent.
#### Interference
The interference study was performed according to CLSI EPO7-A2, Interference Testing in Clinical Chemistry; Approved Guideline - Second Edition. A set of three human serum specimens, one positive, one near the cutoff and one negative sample were tested using the following interfering substances (bilirubin, hemoglobin, triglycerides, cholesterol, rheumatoid factor IgM and human IgG). All interferents were spiked into every serum specimen and the resulting samples were assessed in triplicates with the Aptiva Celiac Disease IgG assays. Recovery of the unit values was calculated compared to control samples. Acceptance criteria for the interference studies were 85% - 115% recovery, or ± 20% of the cut-off (±1.0 FLU) difference, whichever is greater.
Less than 15% of interference was observed for DGP or tTG IgG with the following interferents: bilirubin up to 1 mg/mL (recovery: from 96.0%-101.3% and from 97.7% to 102.5% for DGP lgG and tTG IgG respectively), hemoglobin up to 2 mg/mL (recovery: from 97.6% to 104.2% and from 96.9% to 100.8% for DGP IgG and tTG IgG respectively), triglycerides up to 1000 mg/dL (recovery: from 93.8% to 102.5% and from 94.5% to 100.8% for DGP IgG and tTG IgG respectively), cholesterol up to 332.5 mg/dL (recovery: from 85.4% to 98.7% and from 89.2% to 100.4% for DGP IgG and tTG IgG respectively), RF IgM up to 250 IU/mL (recovery: 95.4% to 97.6 and 92.1% to 97.3% for DGP IgG and tTG IgG, respectively) and human lgG up to 35 mg/mL (recovery: from 105.0% to 114.1% or -0.80 FLU and from 101.7 to 110.8% or 0.47 FLU for DGP IgG and tTG IgG, respectively).
#### Sample Stability and Handling
For the DGP IgG assay, six test samples were tested and for tTG IgG assay seven test samples were tested. The samples used for this study were achieved by combining high and low antibody level to yield their desired reactivity. Test samples covered the analytical measuring ranges of each analyte. All samples were tested in duplicates for up to 21 days while stored at 2-8℃, up to 48 hours while stored at room
{16}------------------------------------------------
temperature, and after repeated freeze/thaw cycles up to 6 cycles. Results were compared to those obtained on control samples (time zero / zero cycles).
Acceptance criteria: percent recovery is between 85-115% for positive samples, and between 80-120% for negative samples (<5.00 FLU).
All samples fulfilled the acceptance criteria at each time point for each condition. Based on these results, we recommend that samples may be stored up to 48 hours at room temperature, up to 14 days at 2-8°C and can be subjected to up to 5 freeze/thaw cycles.
#### Reagent Stability
#### Shelf life
To establish the initial claim for shelf life, accelerated stability studies were performed for 5 weeks at 37°C ± 3°C, where one week is equal to six months at 5 ± 3°C.
Accelerated stability testing was performed on each of the following sealed components to establish initial stability claim:
- . Aptiva Celiac Disease IgG microparticle (bead) - 3 lots
- PE Tracer IgG - 3 lots
Each week a new sealed component was placed in the incubator, and all components were tested at the end of the experiment together with the one that was stored at 5 ± 3°C. The recovery of the measured values was calculated for each time point (compared to those obtained with 5 ± 3°C stored reagent). All calculations were performed by comparing results of sealed components stored at 5 ± 3℃ (control) to those stored at 37 ± 3℃ (test) for 1, 2, 3, 4, and 5 weeks, where one week is equal to six months at 5 ± 3°C. Linear regression analysis was performed between recovery values and the number of days. For each component tested, linear regression analysis was performed separately on each bead in the Aptiva Celiac Disease IgG Reagent (DGP IgG, tTG IgG, IgG Control Bead).
Acceptance criteria for two-year preliminary expiration dating: With regression analysis, the lower and upper 95% Cl interval of the regression line is between 80% and 120% recovery at day 28 (week 4).
All components tested fulfilled the acceptance criteria above, therefore, two-year expiration dating was assigned to each component.
## In-use (onboard) stability
#### Reagent Cartridge
To establish the in-use stability of the Aptiva Celiac Disease lgG reagent cartridges, one lot of reagent cartridge was tested using up to 12 human serum samples (with different reactivity levels). The specimens were tested periodically for 31 days. At day 15 the reagent cartridge was recalibrated, and a cartridge specific Working Curve was generated. Percent recoveries were calculated compared to the day zero
{17}------------------------------------------------
average values, and linear regression analysis was performed by plotting percent recovery against the number of days. The claim was established using the following criteria (using the one that is fulfil…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.